Air Sampling and Air Flow Adjustment, Time to Change the Dialogue to Modern Terms

Air Sampling and Air Flow Adjustment, Time to Change the Dialogue to Modern Terms

The issues: the terms many learned academically or through industry training such as, calibration, primary/secondary standards are no longer appropriate based on modern standards. We do not calibrate sampling pumps we are setting and verifying flow rates.

Within the science of metrology and flow measurement, the concepts that many of us grew up with, primary (bubble burette) and secondary standard (as with rotameters [rotometers])…those terms are now effectively antiquated even though they are very commonly used. We are not calibrating air sampling pumps. In International Standards Organization/International Electrotechnical Commission (ISO/IEC) standard 17025, calibration means a service from a qualified, accredited firm to provide those services. Think of those certified firms calibrating balances in a laboratory as a simple example. We are determining and adjusting flow rates, not calibrating.

We are determining and adjusting flow rates, not calibrating

With sampling pumps, we are measuring flow into a sampler. There are no primary and secondary calibrators, only traceable (through annual calibration) [BIOS Defender, TSI calibrator etc.] or non-traceable (as with rotameters) "working standards."

The ASTM standard D5337 – 23, Standard Practice for Setting and Verifying the Flow Rate of Personal Sampling Pumps, is an international standard practice that defines these issues and terminology. A link to the standard practice is below. It must be purchased if not a member with download ability.

ASTM D5337: Standard Practice for Setting and Verifying the Flow Rate of Personal Sampling Pumps

Notice in the title: “…for Setting and Verifying the Flow Rate…” this has been changed over the years from “calibration.” This standard practice focuses on personal sampling pumps but the concepts are the same for the high volume [“hi-vol”] pumps used in asbestos work.

Traceability means "the property of a measurement result whereby the result can be related to a reference through a documented unbroken chain of calibrations, each contributing to the measurement uncertainty."

Reference: Joint Committee for Guides in Metrology (2012) International vocabulary of metrology — Basic and general concepts and associated terms (VIM) (International Bureau of Weights and Measures (BIPM), Sèvres, France), 3rd Ed. BIPM, IEC, IFCC, ILAC, ISO, IUPAC, IUPAP and OIML, JCGM 200:2012 (2017 version with minor corrections and informative annotations) Available at https://jcgm.bipm.org/vim/en/.

Examples of traceable working standards, there are others in the marketplace:

Examples of traceable working standards. Photos, left from MesaLabs and right from Zefon

The examples above are traceable to meet ISO/IEC requirements. The most important aspect is that these devices are required to be sent back to the manufacturer for service to maintain traceability. If one were to go past the date required for service, they should not be used until the manufacturer re-certifies the device.

While we can use the traceable standards in the field, not all are robust enough for rough handling or being dropped to the floor. Significant damage can occur. Should this happen, it necessitates servicing before continued use.

Examples of the most commonly used non-traceable working standards:

Non-traceable working standards. Photos, Zefon

With the devices we need to verify flow measurements with a traceable standard. That is, to know what the actual air flow may be for the given divisions on the rotameters. These cannot be used as is from the manufacturer. For the high flow rotameter on the left, I do not recommend using the lower divisions for personal sampling pumps. Use the rotameter seen on the right.

These devices are perfectly acceptable should the be verified with a traceable standard periodically. There are no federal rules on this in the U.S., but not less than annually is a common procedure. In the pic above, the hi-flow rotometer has a metal ball as the flow indicator. This likely has more impact on the interior surface than the glass bead in the low flow rotometer. If used with regularity, the hi-flow rotometer may need to be revalidated more often, perhaps at every 6 months.

In-line calibration with the TSI traceable standard. On the left of the pic is a filter device that must be present when using this device.
In-line flow measurement with a rotometer. The short tubing on the pump is to accommodate the photo.
In-line flow measurement with a hi-vol pump
When observing the indicator ball in a rotometer to determine the flow rate – the device must not be held at any angle otherwise the observed flow rate will be incorrect. Most use the middle of the ball to observe the given flow measurement.

Determining the flow rates for the divisions on the rotometer with a traceable working standard.

An example of an Excel form to perform a linear regression to determine flow positions on a rotameter.

A rotameter with determined positions of given flow rates. This rotameter is capable of low and high flow rate determinations, most are not. If you have a rotameter with a 2003 date as its last flow determination, that is NOT defensible.

Of great importance, your firm must have a standard operating procedure (SOP) for this work...from determining the flow positions on a rotameter to determining and verifying flow rates on sampling pumps with traceable or non-traceable working standards. Without this and service records for your chosen traceable standard, it will not go well should your data be challenged. It is very common in depositions for defendant's experts to ask these questions, and in detail.

Here's an issue that we need to visit...the open face method of using a rotameter. Unfortunately this is not appropriate to the science of flow measurements despite that many of us were taught to "calibrate like you sample." This method was also described in an EPA guidance document we call the Silver Book (EPA 600/4-85-049, November, 1985). Unfortunately as much as I and others taught this method for years, its not the most accurate nor is it technically correct.

From the Silver Book...likely the source for many to use the open face rotameter process
The open-face method of flow measurement. While utilized by many, for a variety of reasons, it's time to move away from this method and use the in-line methods

So, the old terminology of primary and secondary standards will likely persist for years, but these terms are not in the lexicon of international standards such as ASTM D-5337. We are not calibrating air sampling pumps or rotameters. So, we need to change old habits and think about terms like “setting and verifying the flow rate” and “traceable or non-traceable working standards.”


Hope this helps...I know it's a big change, but it's time to come in line with modern analytical terms and technology expectations.

All the best!


Copyright, Thomas G. Laubenthal, July 2025. Use for information exchange is OK. No other uses without permission from the author.

I have always argued that the setup shown with your caption "Determining the flow rates for the divisions on the rotometer with a traceable working standard." would lead to slightly different flow rates than actual use. Because changing hose length will change actual flow rates, having the rotometer and Standard set up in series creates a longer hose path, and one with more resistance due to the ball in the rotometer.. IMHO we should swap out the Traceable Standard and the field rotometer for each calibrated value, i.e. only one meter in the sampling chain at a time, and each should have the same hose length between the meter and the cassette. A similar argument to why the "Silver Book" method is less accurate and incorrect! Real problems in actual field monitoring likely lead to much bigger discrepancies and thus errors: failing to warm up the pumps properly, having kinked hoses, sloppy time keeping, sucking up dust bunnies with a loose hose on the floor, or having encapsulant mist clogging up the filter during the sampling run.

Chris Nielsen

Vice President at ECOH

2mo

Excellent information Tom - thanks for sharing.

Mollie Littrell Struzick

Senior Safety & EHS Leader | Driving Operational Excellence & Compliance | Expert in Nuclear, Industrial Hygiene, OSHA & Risk Mitigation | Data-Driven Solutions for Incident Reduction | Building Proactive Safety Cultures

2mo

Thanks for sharing, Tom

Paul Kirby

Vice President at Enpuricon, Inc.

2mo

Thanks for sharing, Tom

John P.

Third-Party Environmental Consultant, Industrial Hygienist

2mo

It's going to take a while to get those phrases "air samples were calibrated"..out of my mind.

To view or add a comment, sign in

Others also viewed

Explore content categories