The Gamma Camera
Components:
NB:
1. HVL of lead for Tc-99m = 0.3 mm
2. Scattered radiation that pass through the collimator will be rejected later by PHA
Collimator:
• Consists of lead
• Average Thickness: 25 mm
• Average Diameter : 400 mm
• About 20000 closely packed circular or hexagonal holes, each is about 2.5 mm in diameter separated by 0.3 mm thick septa
N.B. The collimator acts as a lens, its purpose is for spatial mapping. It does not reject scatter.
Features of the collimator
-The most common, Used with general purpose camera, Do not change the image size, In air sensitivity are the same at all distances from the collimator face
-Used with smaller diameter cameras (mobile cameras), Large FOV (Imaging large organs e.g. lung), Minification of image
NB: allow the use of a small crystal to image a large field of view
-Holes converge to a point inside the patient, Small FOV (children & imaging small organs), Magnification of image, Spatial resolution deteriorates towards the edge of the field
-Cone of lead with single hole at its apex, Used with small & superficial organs e.g. thyroid, Magnification & Inversion of image
NB: Divergent & Convergent – Suffer from geometrical distortion (with deterioration of spatial resolution at the edge of the field) – FOV & in air sensitivity differs with distance
Hole formation:
The holes can be created by:
Septal thickness:
Collimator sensitivity:
Detector:
1) Scintillation crystal
NB:
2) Perspex slab (light pipe / light guide)
3) Photomultiplier tubes (PMT)
Evacuated glass envelope.
Mechanism
1. The light photons hit a photocathode at the entrance to the PMT → releases electrons in proportion to the amount of light that hits it (one electron per 5-10 light photons)
2. Electrons are accelerated toward a positive anode, en route electrons impinge on a series of dynodes (electrodes) connected to progressively increasing positive potentials
3. When each electron strikes a dynode it knocks out 3-4 electrons → accelerated to strike the next dynode
4. After 10 stages, the electrons have been multiplied by a factor 410 ≈ 106
5. The total electrons hit the final anode and the current produced forms the signal received by the pre-amplifier.
6. Thus each initial flash of light produces a pulse of charge or voltage large enough to be measured electronically
NB: There are between 10–12 dynodes in the chain and each has a potential difference of about 100 v. This results in a total potential difference across the tube of 1–1.2 kV.
Pre-amplifier
Image Formation
1) Energy calculation (pulse height spectrum)
For each scintillation formed, the calculated absorbed energy (Z value) that caused it depends on the energy of the gamma photon that was emitted from the patient and the proportion of the energy that was absorbed into the crystal.
The gamma photon energy absorbed by the scintillation crystal depends on its interaction with that photon which results in a spectrum of Z values. If:
1. All energy is absorbed: gamma photon interacts with crystal via photoelectric effect
2. Part of the energy is absorbed: photon undergoes one or more Compton interactions
The spectrum has:
The photopeak width:
NB: This range of energies in the photopeak is due to statistical fluctuations in both:
1. Number of light photons produced in the crystal by each gamma ray photon.
2. Number of electrons produced in the photomultiplier by each light photon.
This value is used to calculate the energy resolution of the crystal, which is given as a percentage:
Energy resolution = FWHM (keV) / photopeak energy (keV) x 100
2) Scatter rejection (Pulse height analyser):
NB:
1. cosmic rays
2. as result of pulse pile up
3) Image formation
- Preset number of counts obtained (0.5-1 million counts are acquired for each image frame)
- Preset time passed
NB:
Position of the radioactive atom in the body:
1. The horizontal "X" and vertical "Y" coordinates →determined by the hole through which the gamma ray has passed and produce light flash in the crystal
2. The Z-axis (depth) → determined by photon energy of the original γ ray.
4) Image display
- Each pixel corresponding to a memory location in the matrix (X & Y coordinates)
- Brightness / colour scale corresponding to the count number in that location (Z axis)
- Smoothing to reduce noise
- Windowing to increase contrast
- Interpolation increases the display matrix relative to the acquisition matrix which spreads the counts and makes the pixels less apparent
- Adding and subtracting images to extract quantified information
NB: Separate images of 2 radionuclides can be obtained in the same time by setting two different PHA windows
Planar Imaging:
Planar imaging is the acquisition of 2D nuclear images, similar to plain x-ray films imaging.
Types:
1) Static
- Organ size, shape and position
- Regions of increased or decreased uptake
Examples: DMSA renal scan, bone scan, lung perfusion scan
2) Dynamic
Examples: MAG3 renal scan, gallbladder emptying scan, gastric emptying scan
3) Gated
Sehr informativ😀
Assistant Professor Cum Training & Placement Coordinator | CT University 💼 🩺 J&K Registered Radiologic Technologist ✨ ISRT Lifetime Member
10moVery informative