PET Imaging
PET is a form of tomographic nuclear imaging.
However, PET relies on simultaneous detection of the pair of gamma photons that are released from annihilation of a positron and an electron.
Annihilation
1. Positron decay
A positron (represented as e+, β+ or e) is released:
Radionuclides:
The energy difference between the parent and daughter nuclei must exceed 1.022 MeV (2 x 0.511 MeV) for positron decay to occur.
2. Positron travel through matter
1. As it travels, it collides with atoms losing energy and causing ionisation (main method of radiation dose deposition in patient)
2. As it collides, it is deflected - tortuous path with length depends upon the number of collisions and the starting energy of the positron.
This means that the distance between where the positron is emitted and where it annihilates is variable
3. Annihilation
PET radiopharmaceuticals
NB: Other useful radionuclides are 68Ga (68min) and 82Rh (1 min), and these two are produced by radionuclide generators
PET scanner
Blocks of scintillation crystals (detector blocks) are arranged in a circle mounted on a gantry in one or two rows (10000–20000 detectors for each PET scanner).
Scintillation crystal
LAC not enough for the annihilation photons which have a higher energy of 511 keV
- Bismuth Germanate (BGO) - Most commonly used scintillator in PET imaging, but the light output and light decay time are inferior to NaI
- Newer materials (lutetium oxyorthosilicate (LSO) / gadolinium oxyorthosillicate (GSO))
Ideal properties of PET scintillation detectors:
Forming an Image
N.B. this is how collimation is achieved in PET. The lead collimator grids that are used in planar imaging and SPECT are not required.
1) Data acquisition
2D vs 3D acquisition:
NB:
Unwanted coincidence rejection:
Unwanted coincidences cause artefactual lines of response to be calculated which do not correspond to the true location of the annihilations.
Cause:
Increased scatter coincidence, which occurs with:
Increased random coincidence, which occurs with:
How to reduce random and scatter events
Narrow lead or tungsten septa (1 mm thick , 10 mm deep) are used between each ring of detectors (in Z plane) i.e. act as antiscatter grid
2) Data correction
Attenuation correction
Attenuation is greater in PET > SPECT due to longer path the photon must travel through the patient.
Correction:
1. Assume cross-sectional shape and uniform LAC of tissue at 511 keV
2. Measure the transmission of 511 keV photons through the patient for each line of response. A radioactive rod source (gallium-68) that gives rise to annihilation radiation is rotated around inside the detector gantry without the patient and then with the patient. This allows a calculation to be made correcting for the attenuation by the patient.
Normalisation
Individual detector elements differ in dimensions and fraction of scintillation light photons that reach the PMTs -Same radiation source may not produce same response in every detector element.
Solution:
Rotating rod source used without object in the scanner to calculate the correction factor required for differences in the individual detector elements
Correction factor
= measured counts for line of response / average counts for all lines
Dead time (Temporal resolution)
NB: → if two flashes of light overlap and they are treated as one.
Solution:
Dead time measured and mathematical algorithms that take into account detector behaviour applied to extrapolate from measured counts
Radioactive decay
Solution:
Counts corrected for radioactive decay
3) Data reconstruction
2D acquisitions are reconstructed using filtered back projection or iterative reconstruction
PET-CT scanning:
Advantage:
• The combined PET‐CT images are particularly useful in oncology, both for diagnosis and for accurate tumor localisation and follow‐up.
• Carefully gating image acquisition to the cardiac cycle can also produce useful fused images in cardiology.
Difficulties:
• Adjustment of matrix size • Matching the transverse planes for co-registration
• Fusion software (complicated by different patient’s position and movement)
• Integrated PET CT scanners (mounted adjacent to each other)
– Perfect matching is produced
– Patient table move to acquire the CT scan, then the table return to position to acquire PET data
– CT scan can be also used for PET attenuation correction → whole scan in <30 min
– When used in cardiology: cardiac gating may be mandatory
NB: Use of gamma camera as PET scanners
Dual headed conventional gamma camera can be used as PET scanner
Conditions:
– Integration of coincidence circuity
– Rotation of the cameras around the patient without the collimators
– Thicker crystals of NaI must be used
Disadvantages:
– Higher background noise (decrease contrast)
– Poorer spatial resolution
Attended Mewar University
10moVery informative
Sehr informativ, Danke 😀
Scientific researcher | Medical physicist
10moInsightful
RT @AHMC and FGAE Adama Model Clinic
10moInteresting