WHAT IS OCT?
Optical Coherence Tomography (OCT) is an optical imaging modality
that uses near-infrared light to create high-resolution images of tissue
microstructure.
Optical  of or relating to light (visual)
Coherence  a measure of the correlation between phases of a wave
or waves
Tomography  method of producing images from a series of single
planes or slices (Greek “TOMOS=Slice”)
4
WHAT IS INTRAVASCULAR OCT?
OCT is an optical imaging modality that uses near-infrared light
(1250-1350nm) for high-resolution imaging of vessel anatomy,
tissue microstructure and stents.
Key Features:
Uses light, not sound
Does not use X-ray
Image acquisition is fast
Images acquired are sharp,
detailed and easy to interpret
5
INTRAVASCULAR OCT
Flexible fiber-optic catheter used for light delivery
Fiber rotates to create image frames
Fiber pulls back to map vessel segment
• The more frames per mm, or high frame density, the higher
the resolution
6
IMAGE DISPLAY
8
“L-Mode”
longitudinal view
“B-Mode”
cross-sectional view
9
Current Generation PCI Optimization Systems
2011-12 2013-15 2015 2016
ILUMIEN™ System ILUMIEN™ OPTIS™
System
OPTIS™ Integrated
System
OPTIS™ Mobile
System
• Commercially
available 2011
• 100 fps / 54 mm
pullback
• Combined FFR and
OCT
• Wireless FFR
• 180 fps/75 and 54 mm
pullback
• Advanced software
tools
• Real time 3D
reconstruction
• Pullback initialization
from DOC
• Integrated in the
Cath lab
• Angio Co-
Registration With
tableside FFR/ OCT
controller
• Integrates with
Multiple Cath labs
• Angio Co-
Registration With
tableside FFR/ OCT
controller
FFR and OCT System 2nd Gen FFR and OCT
System
Cath lab integrated FFR
and OCT system
Latest system with 3D
OCT, FFR, ACR & MSO
Standard available for bug fixes only
Note: All systems having MSO must be updated to E.4.1 version by Service team.
10
2016/17 OCT PRODUCT LAUNCHES
OPTIS™ Mobile System OPTIS™ Mobile Upgrade Kit
(Ilumien™ OPTIS™ Systems only)
OPTIS™ Metallic Stent
Optimization Software-
MSO
(Software Version E.4)
OPTIS™ Mobile
Workstation
(Accessory for OPTIS™
Integrated System)
Mobile System integrated into
Cath lab during procedure to
offer angio co-registration
• Angio Co-registration
• Wireless Tableside Controller
• Mobile cart connect into Cath
Lab during procedure
• Enables the use in multiple
labs
• Will include E.4 after LMR
Enables angio co-registration
function on Ilumien™ OPTIS
System
• Angio Co-registration (ACR)
• Wireless Tableside Controller
• Mobile cart connect into Cath
Lab during procedure
• Enables the use of OCT-ACR
and FFR in multiple labs
• Leverage options to include E.4
New clinically relevant
software enhancements for
pre and post PCI
• Stent roadmap on angio co-
registration
• New default view with
automated measurements
from lumen profile
• Acute Stent detection and
apposition assessment
• Expanded 3D imaging
Offered as another point of
control to operate the
OPTIS Integrated system
• Optional control point for
non-sterile personnel to
perform all input/viewing
functions available on
OPTIS Integrated System
Proprietary and confidential — do not distribute
OPTICAL COHERENCE TOMOGRAPHY - SYSTEM OVERVIEW
Proprietary and confidential — do not distribute
DOC (Drive Motor & Optical Controller)
CPU (Central Processing Unit)
ENGINE (Infrared Light Source)
Physician and Operator Monitors
Monitors
DOC ( DRIVE MOTOR AND OPTICAL CONTROLLER )
1 Emergency shutoff – stops rotation, pullback and
scanning
2 Enable scanning; second push will start
pullback in manual trigger mode
3 Toggle between Live and Standby
4 Unload catheter
1
2
3
4
DragonflyTM
Duo /
DragonflyTM
Optis
OCT Catheter
OCT
Wireless
FFR
PressureWireTM
AerisTM
PressureWireTM
X
ILUMIEN / ILUMIEN OPTIS™ / OPTIS™ RANGE – BOTH OCT & FFR
ILUMIEN™ Console
Ilumien Optis Console
COMPARISON TO ILUMIEN ENGINE
ILUMIEN ILUMIEN OPTIS/OPTISi & Mobile
Parameter Nominal
Settings
Long
Pullback High-resolution Pullback
Engine speed
100 frames/sec
20 mm/sec
5 frames/mm
54 mm
180 frames/sec 180 frames/sec
Pullback speed 36 mm/sec 18 mm/sec
Frame density 5 frames/mm 10 frames/mm
Pullback length 75 mm 54 mm
Pullback time
2.7 sec
2.1 sec 3.0 sec
Contrast (Max)
14 ml at 4 ml/sec
10 ml at 4 ml/sec 14 ml at 4 ml/sec
File size 271 frames
270 MB
375 frames
375 MB
540 frames
540 MB
Key Features of Abbott - OCT
More
Information
Same Time
Same Contrast
ILUMIEN OPTIS & MSO SOFTWARE
Reviewing an acquired Image
– Entering note
– Bookmark
– Calibration
Performing the Measurements
– Lumen Profile
– MLA /AS% / DS% Mean Diameter
– Dynamic stent planning
Automated Measurements
Stent Roadmap
Apposition Indicator
3D Bifurcation Mode
 Side-branch detection
 Carina view
Expanded 3D Navigation
 3D flythrough view
 Object segmentation
17
OPTIS™ Stent Optimization Software
broadens the clinical utility for OCT to
guide daily decision making for complex
PCI both pre and post stenting though:
CLINICAL APPLICATIONS OF OCT
18
• Measurement of reference segments
• MLA/MLD measurement and identification
• Tissue differentiation
• Lumen morphology
• Potential culprit lesion
Pre-PCI and
Lesion
Assessment
• Lesion coverage and deployment
mapping
• Underexpansion and apposition
• Lumen optimization
• Identification of sub-optimal stent
results
PCI Guidance
• Restenosis measurements
• Tissue coverage assessment
• Visualization of absorption and remodeling
of bioabsorbable stents
• Lesion progression
Stent Follow-up
19
OCT DRAGONFLY RANGE OF
CATHETERS
Proprietary and confidential — do not distribute
QUESTION
What are the
different Catheters
of OCT ?
21
Usable length: 135 cm
Outer diameter: 2.7 F (distal)
Wire lumen: 0.014”
Hydrophilic Coating
DRAGONFLY™ DUO IMAGING CATHETER
SPECIFICATIONS
Lens
INTRODUCING LATEST OCT CATHETER
DragonflyTM
OPTISTM
Simplifies the OCT procedure
Lens Marker
Facilitates Angio Co‐Registration
DUAL LUMEN for GUIDEWIRE EXIT
Easier loading ; reduce potential for lens damage
and kinking
TIP MARKER
©2018 Abbott. All Rights Reserved. IN CP0007‐EN 02/18
Dragonfly
Duo
Dragonfl
y
OPTIS
Dragonfly™ OPTIS™ changes vs. Dragonfly™ Duo Catheters
1. Dual lumen
2. Smoother transition zone
3. Shorter rail
4. Dedicated purge exit
TIP DESIGN DRAGONFLY™ OPTIS CATHETER
Catheter design changes:
• Rapid exchange tip
– Improves catheter performance
– Reduces risk of damage and kinking
• Support for continuous calibration*
– Reduces steps necessary during calibration
• Shaft guide marker at 90 cm
– Aids in understanding distal tip position during
radial procedures
• Gray hub
– Differentiates Dragonfly OPTIS catheter from
previous Dragonfly catheters
*Refer to D.2 or E.2 Software
Note: The Dragonfly OPTIS imaging catheter is designed for use with the ILUMIEN OPTIS PCI Optimization and OPTIS
integrated systems equipment. It should not be used with older model OCT systems.
DRAGONFLY™ OPTIS™ IMAGING CATHETER
ILUMIEN™, ILUMIEN™ OPTIS™ PCI
Optimization and OPTIS™ integrated
systems software
Dragonfly™ OPTIS™ imaging
catheter

More Related Content

PPTX
CORONARY BALLOONS PRACTICAL ASPECTS.pptx
PPTX
CORONARY ANOMALY
PPTX
Coronary Intravascular Lithotripsy
PDF
Echo assessment of aortic valve disease
PPTX
Longitudinal stent deformation in PCI
PPTX
PPTX
BMV balloons- FINAL.pptx
CORONARY BALLOONS PRACTICAL ASPECTS.pptx
CORONARY ANOMALY
Coronary Intravascular Lithotripsy
Echo assessment of aortic valve disease
Longitudinal stent deformation in PCI
BMV balloons- FINAL.pptx

What's hot (20)

PPTX
Guiding catheter in coronary intervention
PPTX
VSD devices
PPT
Shunt Detection And Quantification
PPTX
Ffr guided coronarY intervention
PDF
CONTRAST ECHOCARDIOGRAPHY
PPTX
CALCIFIED CORONARY ARTERY LESIONS
PPTX
OCT slides
PPTX
FFR(fractional flow reserve)
PPTX
PPTX
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHAN
PPTX
Right heart catheters
PPTX
Lv systolic function
PPTX
His bundle pacing by Dr. Vaibhav Yawalkar MD DM Cardiology, Consultant Interv...
PDF
Contrast echocardiography
PDF
Angiography.pdf
PPTX
FRACTIONAL FLOW RESERVE
PPTX
Echo in hypertrophic obstructive cardiomyopathies
PPTX
Doppler echocardiography
Guiding catheter in coronary intervention
VSD devices
Shunt Detection And Quantification
Ffr guided coronarY intervention
CONTRAST ECHOCARDIOGRAPHY
CALCIFIED CORONARY ARTERY LESIONS
OCT slides
FFR(fractional flow reserve)
BMV,PTMC,BALLOON MITRAL VALVOTOMY, BAL, VIRBHAN BALAI, DR VIRBHAN
Right heart catheters
Lv systolic function
His bundle pacing by Dr. Vaibhav Yawalkar MD DM Cardiology, Consultant Interv...
Contrast echocardiography
Angiography.pdf
FRACTIONAL FLOW RESERVE
Echo in hypertrophic obstructive cardiomyopathies
Doppler echocardiography
Ad

Similar to INTRO TO OCT & SYSYTEM.pptx (20)

PPTX
New OCT Slides for revascularisation and decision making
PPTX
PPT
OCT Presentation with St. Jude. Medical System.
PPT
Optical Coherence Tomography dr md toufiqur rahman cardiologist
PPTX
OCT overview and tissue characteristaion-Mohan.pptx
PDF
OCT in coronary PCI
PPTX
OPTICAL COHERENCE TOMOGRAPHY
PDF
Intracoronary optical coherence tomography
PPTX
OCT product materials-ST.JUDE MEDICAL.pptx
PPTX
OCT -INTERPRETATION.pptx
PPTX
optical-coherence-tomography-8889-l2YBJiS.pptx
PPTX
OCT in intervention cardiology managment.pptx
PDF
Intracoronary Optical Coherence Tomography (2)
PPTX
cse Optical coherence tomography ppt.pptx
PPTX
Optical coherence tomography ppt
PPTX
Intracoronary optical coherence tomography
PDF
Oct angiography
PPTX
PPTX
The Eye as the Window to the Heart OCT in Coronary Artery Disease A New Era o...
New OCT Slides for revascularisation and decision making
OCT Presentation with St. Jude. Medical System.
Optical Coherence Tomography dr md toufiqur rahman cardiologist
OCT overview and tissue characteristaion-Mohan.pptx
OCT in coronary PCI
OPTICAL COHERENCE TOMOGRAPHY
Intracoronary optical coherence tomography
OCT product materials-ST.JUDE MEDICAL.pptx
OCT -INTERPRETATION.pptx
optical-coherence-tomography-8889-l2YBJiS.pptx
OCT in intervention cardiology managment.pptx
Intracoronary Optical Coherence Tomography (2)
cse Optical coherence tomography ppt.pptx
Optical coherence tomography ppt
Intracoronary optical coherence tomography
Oct angiography
The Eye as the Window to the Heart OCT in Coronary Artery Disease A New Era o...
Ad

Recently uploaded (20)

PPTX
Nancy Caroline Emergency Paramedic Chapter 11
PPTX
Newer Technologies in medical field.pptx
PDF
health promotion and maintenance of elderly
DOCX
ch 9 botes for OB aka Pregnant women eww
PPTX
Nancy Caroline Emergency Paramedic Chapter 13
PDF
01. Histology New Classification of histo is clear calssification
PPTX
Theories and Principles of Nursing Management
PPTX
Diabetes_Pathology_Colourful_With_Diagrams.pptx
PDF
Introduction to Clinical Psychology, 4th Edition by John Hunsley Test Bank.pdf
PPTX
PE and Health 7 Quarter 3 Lesson 1 Day 3,4 and 5.pptx
PPTX
FUNCTIONS OF BLOOD PART I AND PART 2 WHOLE
PPTX
General Pharmacology by Nandini Ratne, Nagpur College of Pharmacy, Hingna Roa...
PPTX
Understanding The Self : 1Sexual health
PDF
Back node with known primary managementt
PPTX
Acute renal failure.pptx for BNs 2nd year
PPTX
Nepal health service act.pptx by Sunil Sharma
PDF
Essentials of Hysteroscopy at World Laparoscopy Hospital
PPTX
Arthritis Types, Signs & Treatment with physiotherapy management
PPTX
POSTURE.pptx......,............. .........
PPTX
Nancy Caroline Emergency Paramedic Chapter 4
Nancy Caroline Emergency Paramedic Chapter 11
Newer Technologies in medical field.pptx
health promotion and maintenance of elderly
ch 9 botes for OB aka Pregnant women eww
Nancy Caroline Emergency Paramedic Chapter 13
01. Histology New Classification of histo is clear calssification
Theories and Principles of Nursing Management
Diabetes_Pathology_Colourful_With_Diagrams.pptx
Introduction to Clinical Psychology, 4th Edition by John Hunsley Test Bank.pdf
PE and Health 7 Quarter 3 Lesson 1 Day 3,4 and 5.pptx
FUNCTIONS OF BLOOD PART I AND PART 2 WHOLE
General Pharmacology by Nandini Ratne, Nagpur College of Pharmacy, Hingna Roa...
Understanding The Self : 1Sexual health
Back node with known primary managementt
Acute renal failure.pptx for BNs 2nd year
Nepal health service act.pptx by Sunil Sharma
Essentials of Hysteroscopy at World Laparoscopy Hospital
Arthritis Types, Signs & Treatment with physiotherapy management
POSTURE.pptx......,............. .........
Nancy Caroline Emergency Paramedic Chapter 4

INTRO TO OCT & SYSYTEM.pptx

  • 1. WHAT IS OCT? Optical Coherence Tomography (OCT) is an optical imaging modality that uses near-infrared light to create high-resolution images of tissue microstructure. Optical  of or relating to light (visual) Coherence  a measure of the correlation between phases of a wave or waves Tomography  method of producing images from a series of single planes or slices (Greek “TOMOS=Slice”) 4
  • 2. WHAT IS INTRAVASCULAR OCT? OCT is an optical imaging modality that uses near-infrared light (1250-1350nm) for high-resolution imaging of vessel anatomy, tissue microstructure and stents. Key Features: Uses light, not sound Does not use X-ray Image acquisition is fast Images acquired are sharp, detailed and easy to interpret 5
  • 3. INTRAVASCULAR OCT Flexible fiber-optic catheter used for light delivery Fiber rotates to create image frames Fiber pulls back to map vessel segment • The more frames per mm, or high frame density, the higher the resolution 6
  • 5. 9 Current Generation PCI Optimization Systems 2011-12 2013-15 2015 2016 ILUMIEN™ System ILUMIEN™ OPTIS™ System OPTIS™ Integrated System OPTIS™ Mobile System • Commercially available 2011 • 100 fps / 54 mm pullback • Combined FFR and OCT • Wireless FFR • 180 fps/75 and 54 mm pullback • Advanced software tools • Real time 3D reconstruction • Pullback initialization from DOC • Integrated in the Cath lab • Angio Co- Registration With tableside FFR/ OCT controller • Integrates with Multiple Cath labs • Angio Co- Registration With tableside FFR/ OCT controller FFR and OCT System 2nd Gen FFR and OCT System Cath lab integrated FFR and OCT system Latest system with 3D OCT, FFR, ACR & MSO
  • 6. Standard available for bug fixes only Note: All systems having MSO must be updated to E.4.1 version by Service team. 10 2016/17 OCT PRODUCT LAUNCHES OPTIS™ Mobile System OPTIS™ Mobile Upgrade Kit (Ilumien™ OPTIS™ Systems only) OPTIS™ Metallic Stent Optimization Software- MSO (Software Version E.4) OPTIS™ Mobile Workstation (Accessory for OPTIS™ Integrated System) Mobile System integrated into Cath lab during procedure to offer angio co-registration • Angio Co-registration • Wireless Tableside Controller • Mobile cart connect into Cath Lab during procedure • Enables the use in multiple labs • Will include E.4 after LMR Enables angio co-registration function on Ilumien™ OPTIS System • Angio Co-registration (ACR) • Wireless Tableside Controller • Mobile cart connect into Cath Lab during procedure • Enables the use of OCT-ACR and FFR in multiple labs • Leverage options to include E.4 New clinically relevant software enhancements for pre and post PCI • Stent roadmap on angio co- registration • New default view with automated measurements from lumen profile • Acute Stent detection and apposition assessment • Expanded 3D imaging Offered as another point of control to operate the OPTIS Integrated system • Optional control point for non-sterile personnel to perform all input/viewing functions available on OPTIS Integrated System
  • 7. Proprietary and confidential — do not distribute OPTICAL COHERENCE TOMOGRAPHY - SYSTEM OVERVIEW
  • 8. Proprietary and confidential — do not distribute DOC (Drive Motor & Optical Controller) CPU (Central Processing Unit) ENGINE (Infrared Light Source) Physician and Operator Monitors Monitors
  • 9. DOC ( DRIVE MOTOR AND OPTICAL CONTROLLER ) 1 Emergency shutoff – stops rotation, pullback and scanning 2 Enable scanning; second push will start pullback in manual trigger mode 3 Toggle between Live and Standby 4 Unload catheter 1 2 3 4
  • 10. DragonflyTM Duo / DragonflyTM Optis OCT Catheter OCT Wireless FFR PressureWireTM AerisTM PressureWireTM X ILUMIEN / ILUMIEN OPTIS™ / OPTIS™ RANGE – BOTH OCT & FFR ILUMIEN™ Console Ilumien Optis Console
  • 11. COMPARISON TO ILUMIEN ENGINE ILUMIEN ILUMIEN OPTIS/OPTISi & Mobile Parameter Nominal Settings Long Pullback High-resolution Pullback Engine speed 100 frames/sec 20 mm/sec 5 frames/mm 54 mm 180 frames/sec 180 frames/sec Pullback speed 36 mm/sec 18 mm/sec Frame density 5 frames/mm 10 frames/mm Pullback length 75 mm 54 mm Pullback time 2.7 sec 2.1 sec 3.0 sec Contrast (Max) 14 ml at 4 ml/sec 10 ml at 4 ml/sec 14 ml at 4 ml/sec File size 271 frames 270 MB 375 frames 375 MB 540 frames 540 MB Key Features of Abbott - OCT More Information Same Time Same Contrast
  • 12. ILUMIEN OPTIS & MSO SOFTWARE Reviewing an acquired Image – Entering note – Bookmark – Calibration Performing the Measurements – Lumen Profile – MLA /AS% / DS% Mean Diameter – Dynamic stent planning Automated Measurements Stent Roadmap Apposition Indicator 3D Bifurcation Mode  Side-branch detection  Carina view Expanded 3D Navigation  3D flythrough view  Object segmentation 17 OPTIS™ Stent Optimization Software broadens the clinical utility for OCT to guide daily decision making for complex PCI both pre and post stenting though:
  • 13. CLINICAL APPLICATIONS OF OCT 18 • Measurement of reference segments • MLA/MLD measurement and identification • Tissue differentiation • Lumen morphology • Potential culprit lesion Pre-PCI and Lesion Assessment • Lesion coverage and deployment mapping • Underexpansion and apposition • Lumen optimization • Identification of sub-optimal stent results PCI Guidance • Restenosis measurements • Tissue coverage assessment • Visualization of absorption and remodeling of bioabsorbable stents • Lesion progression Stent Follow-up
  • 14. 19 OCT DRAGONFLY RANGE OF CATHETERS
  • 15. Proprietary and confidential — do not distribute QUESTION What are the different Catheters of OCT ?
  • 16. 21
  • 17. Usable length: 135 cm Outer diameter: 2.7 F (distal) Wire lumen: 0.014” Hydrophilic Coating DRAGONFLY™ DUO IMAGING CATHETER SPECIFICATIONS Lens
  • 18. INTRODUCING LATEST OCT CATHETER DragonflyTM OPTISTM Simplifies the OCT procedure Lens Marker Facilitates Angio Co‐Registration DUAL LUMEN for GUIDEWIRE EXIT Easier loading ; reduce potential for lens damage and kinking TIP MARKER ©2018 Abbott. All Rights Reserved. IN CP0007‐EN 02/18
  • 19. Dragonfly Duo Dragonfl y OPTIS Dragonfly™ OPTIS™ changes vs. Dragonfly™ Duo Catheters 1. Dual lumen 2. Smoother transition zone 3. Shorter rail 4. Dedicated purge exit TIP DESIGN DRAGONFLY™ OPTIS CATHETER
  • 20. Catheter design changes: • Rapid exchange tip – Improves catheter performance – Reduces risk of damage and kinking • Support for continuous calibration* – Reduces steps necessary during calibration • Shaft guide marker at 90 cm – Aids in understanding distal tip position during radial procedures • Gray hub – Differentiates Dragonfly OPTIS catheter from previous Dragonfly catheters *Refer to D.2 or E.2 Software Note: The Dragonfly OPTIS imaging catheter is designed for use with the ILUMIEN OPTIS PCI Optimization and OPTIS integrated systems equipment. It should not be used with older model OCT systems. DRAGONFLY™ OPTIS™ IMAGING CATHETER ILUMIEN™, ILUMIEN™ OPTIS™ PCI Optimization and OPTIS™ integrated systems software Dragonfly™ OPTIS™ imaging catheter