Introduction
The United States Environmental Protection
Agency (USEPA) Photochemical Assessment
Monitoring Stations (PAMS) are air quality
monitoring locations that are used to
monitor pollutant compounds in air that contribute to ground level (bad) ozone. These nonpolar
compounds are typically found in the 0.1 to tens ppbv, if not greater in some cases. Since these
levels are too low to detect directly, a concentration technique such as cryotrapping or chemical
adsorbent trapping needs to be used. One of the main issues with any sub-ambient temperature
method is the negative effect of moisture in the air sample. Quantities up to tens of milligrams of
water may be present and this can cause havoc with the chromatographic system, such as the
water freezing with the formation of ice causing a flow restriction or blockage. Water can also
cause poor peak shape/retention of analytes on certain columns such as Alumina PLOT. Several
different techniques can be used for the effective removal of water from a gas sampling stream,
prior to preconcentration, such as, drying agents, semipermeable membranes and reduced
temperatures (used to condense out the water). One well accepted technique uses a Nafion®
(TM,
DuPont™
) membrane to dry the incoming sample stream, and is very effective. The membrane
permeability is very dependent on the analyte, generally speaking, and nonpolars are retained while
water and certain polars are lost across the membrane.
This work is focused on humidity effects on low level sub-ppbv (0.3-1 ppbv) PAMS targets compared
to the work done by G. Broadway et al¹ who studied humidity effects at a higher 10 ppbv.
PAMS: A Study of
Performance in Low and
High Humidity Environments
A P P L I C A T I O N N O T E
Author:
PerkinElmer Canadian
Chromatographic Application Lab
PerkinElmer, Inc.
Woodbridge, ON Canada
Gas Chromatography
2
Experimental
PAMS standard was purchased from Linde Canada at the USEPA
recommended concentration of 3-10 ppbv and was used as the
stock standard. The PAMS components dilutions were dynamically
created in real time using a custom made on-line diluter, designed
and built by the author, using UHP Nitrogen (LN2) as the diluent.
The gas was purified using an advanced triple bed carrier gas
purifier filter.
The on-line diluter utilized two 2000 mL/min. MKS Mass-Flo Mass
flow controllers (MFC). One was used to control pure nitrogen
diluent and the other was to control the flow to a 10 um stainless
steel frit, inserted into a 2L HPLC solvent bottle, ¾ filled with 18
megohm DI water. The bubbler MFC outlet was teed in to the
pure nitrogen MFC output. To eliminate the possibility of trace
level contaminates in the water and/or associated tubing/fittings/
MFC, since we are running at sub-ppbv levels, the total diluent
flow (both Humid and Dry) was scrubbed of organics by the use
of two Soil Vapour Intrusion tubes (SVI, thermal desorption) in
series post blending of the diluent. The now clean diluent was
blended with the PAMS standard (via a 200 mL/min. MKS MFC)
into a ¼” SS tubing. To reduce the effects of laminar flow and
promote mixing, the tubing was reduced to 1/8” SS (6 ft). This
tubing was terminated into a SS 1/8” Tee with one arm going
toward the on-line accessory approx. 1 ft. away with a 15 mL/
min. flow. The remaining arm was plumbed with ¼” Bev-A-Line,
as it was downstream of the sampling accessory, and at the end
of this line a custom designed and manufactured humidity
accessory was installed. A block diagram of the system is shown
in Figure 1.
The custom designed humidity accessory utilized a Honeywell
Humidcon™ HIH9130-021 with a 0-100% RH range, with
+/-1.7 accuracy, +/-0.6 °C temperature reading communication
via I2C. The I2C data was read via a customized Atmel®
328P
microprocessor and the resulting data was displayed via I2C with
a 16 x 2 LCD screen. The % RH and temperature was updated at
10 Hz. Similarly, the MFC were controlled by a custom programmed
Atmel 328P, using SPI controlled MCP4821 12-bit DAC. Since the
MCP4821 max output is 4096 mV and the MFC requires 0-5000
mV, a TLC2272 Rail to Rail Op Amp was used to scale the DAC
0-4096 mV output by 1.25. A rotary encoder with momentary
push buttons was used for the setpoint and MFC selection. The
MFC readback was read by the ATMega328P 10-bit internal ADC
directly without buffering. The readbacks/setpoints were displayed
on a 16 x 2 LCD via I2C. An electronic block diagram showing
the basic components of the MFC control is shown in Figure 2.
Figure1.Diagramofadynamicdiluter.
Figure2.SchematicofMFCcontrolelectronics.
Table1.SystemConditionsforOzonePrecursorAnalysis.
GC Clarus®
580 GC with heartcut option
Column Ozone
Precursor Column Set
50 m x 0.22 mm x 1 μm dimethyl siloxane column
Part number 04978179
50 m x 0.32 mm alumina/Na2SO4 PLOT column
Part number N6301107
Oven Conditions
46 °C for 15 min., then 5 °C/min. to 170 °C,
then 15 °C/minute to 200 °C and hold for 6 min.
(48-min. chromatogram)
Detector Dual FID at 250 °C
Carrier Gas 42 psig helium at the ATD
16 psig helium at the midpoint
ATD
Thermal Desorber TurboMatrix™
TD with online
sampling accessory
Trap
Air Toxics trap packed with carbonaceous adsorbents
Low -30 °C
High 325 °C
Transfer Line 200 °C
Operating Mode On-Line at 15 mL/min. for 40 min.
Split
Inlet Off
Outlet 2 mL/min.
Data Handling TotalChrom™
3
MFC Settings and Flows
MFC settings were 10 bit (0-1023) for the 0-5 V analog input
for the MFC via the 12 bit DAC. The MFC flows were calibrated
using a NIST traceable PE1000 electronic flow meter. The MFCs
used were identified as MFC#1 (for Humid Flow), MFC#2 (Dry
Flow) and MFC#3 (Gas Standard). Unless otherwise noted the
MFC were set as follows:
MFC#1, set to 100, produced a measured flow of 228 mL/min.
direct at MFC output, and 263 mL/min. of humid flow at 77% RH
after humidification. Since the extra volume of the water vapor
at RT contributes, it is believed the difference in flow with the
humidifier attached is a result of the extra backpressure due
to humidifier.
MFC#2, set to 100, produced a flow of 193 mL/min.
MFC#3, set to 100, produced a flow of 34.7 mL/min.
Dilution of standard with dry nitrogen =
(193+34.7)/34.7 = 1:6.56
Dilution of standard with humidified nitrogen =
(263+34.7)/34.7 = 1:8.58
Humidity Notes
Since the generated humidity is passed through two SVI tubes
in series, the humidity needs time to stabilize as each tube has a
moisture capacity of approx. 35 mg each. This effect is described
in detail by Snow et al². In this study, typically one hour
stabilization time was used prior to sampling.
Results
Both the 77% RH and dry dilutions of the PAMS standard give
similar chromatograms in both the PLOT and BP-1 separations as
shown in Figures 3 and 4 respectively. Figure 5 is an overlay of
four sequential chromatographic runs highlighting excellent
repeatability at low levels.
Figure3.PLOTchromatogramofPAMSstandardatapprox.0.3-1ppbv.Topchromatogramis77%RHandbottomisapproximately0%RH.
4
Figure4.BP-1chromatogramofPAMSstandardatapprox.0.3-1ppbv.Topchromatogramis77%RHandbottomisapproximately0%RH.
Figure5.ExpandedsectionofBP-1columnseparationwithfouroverlaid77%RHsamples.
For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs
Copyright ©2015, PerkinElmer, Inc. All rights reserved. PerkinElmer®
is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.
012297_01	PKI
PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA	
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com
References
1.	“Ozone Precursor Analysis Using a Thermal Desorption-GC
System” by G. Broadway and A. Tipler available for download at
www.perkinelmer.com.
2.	Patient Application US 20130180315 A1 Devices, systems and
methods for purging and loading sorbent tubes by Snow et al.
Conclusion
The ozone precursor system performance at the 0.3 ppbv-1 ppbv,
with a PAMS standard under high humidity conditions, are virtually
identical to that under low/zero humidity. The chromatographic
peak shapes and intensities show that high humidity has little/no
negative effect on the analysis of PAMS target compounds at low
levels. This enables the use of dry gaseous standards to be used
without the laborious preparation of humidified standards.

More Related Content

PPT
Steam water analysis system
PPTX
Handling Difficult Samples in Karl Fischer Analysis
PDF
Analysis of Volatile Organic Compounds (VOCs) in Air Using U.S. EPA Method TO-17
PDF
Multipoint dilution and permeation gas calibrators
PPTX
New High-quality, Cost-Efficient Headspace GC Autosampler
PDF
Stamolys ca71 cod endress+hauser datasheet-cod analyzer
PDF
Stamolys ca71 al endress+hauser datasheet-aluminium analyser
Steam water analysis system
Handling Difficult Samples in Karl Fischer Analysis
Analysis of Volatile Organic Compounds (VOCs) in Air Using U.S. EPA Method TO-17
Multipoint dilution and permeation gas calibrators
New High-quality, Cost-Efficient Headspace GC Autosampler
Stamolys ca71 cod endress+hauser datasheet-cod analyzer
Stamolys ca71 al endress+hauser datasheet-aluminium analyser

What's hot (18)

PDF
AICHE Senior Design Project
PDF
Stamolys ca71 ph endress+hauser datasheet-phosphate analyzer
PDF
Stamolys ca71 no endress+hauser datasheet-nitrite analyzer
PDF
Stamolys ca71 mn endress+hauser datasheet-manganese analyser
PDF
Stamolys ca71 fe endress+hauser datasheet-iron analyzer
PDF
Liquid chlorine incident
PDF
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
DOCX
AIChE Final Design Report Excerpt
PDF
Clora-Online
PDF
Scion instruments refinery gas analyzer saranalab mandiri analitika
DOC
Toc analyser, microprocessor controlled
PPTX
Karl fischer
PPTX
Types of Industrial Process Analyzers
PDF
Scion instruments transformer oil gas analyzer by sarana lab mandiri analitika
DOCX
A Quantitative Analysis of Ethanol in Beer using Gas Liquid Chromatography – ...
PPTX
Unit 6 Water Content Determination and Moisture analysis
PPTX
PHARMACEUTICAL APPLICATIONS OF GAS CHROMATOGRAPHY(GC), PHARMACEUTICAL ANALYS...
PDF
Water analysis from Metrohm
AICHE Senior Design Project
Stamolys ca71 ph endress+hauser datasheet-phosphate analyzer
Stamolys ca71 no endress+hauser datasheet-nitrite analyzer
Stamolys ca71 mn endress+hauser datasheet-manganese analyser
Stamolys ca71 fe endress+hauser datasheet-iron analyzer
Liquid chlorine incident
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
AIChE Final Design Report Excerpt
Clora-Online
Scion instruments refinery gas analyzer saranalab mandiri analitika
Toc analyser, microprocessor controlled
Karl fischer
Types of Industrial Process Analyzers
Scion instruments transformer oil gas analyzer by sarana lab mandiri analitika
A Quantitative Analysis of Ethanol in Beer using Gas Liquid Chromatography – ...
Unit 6 Water Content Determination and Moisture analysis
PHARMACEUTICAL APPLICATIONS OF GAS CHROMATOGRAPHY(GC), PHARMACEUTICAL ANALYS...
Water analysis from Metrohm
Ad

Viewers also liked (7)

PDF
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
PDF
PDF
The british regulatory system – global mba brazil
PDF
Analysis of Additives in Biodegradable Polymers Using DSA/TOF Mass Spectrometry
PDF
Detection and Identification of Contaminations in the Manufacturing Process U...
PPTX
Cairo Brokerage Conference- Global MBA Brazil
PDF
Optimization of Cannabis Grows Using Fourier Transform Mid-Infrared Spectroscopy
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
The british regulatory system – global mba brazil
Analysis of Additives in Biodegradable Polymers Using DSA/TOF Mass Spectrometry
Detection and Identification of Contaminations in the Manufacturing Process U...
Cairo Brokerage Conference- Global MBA Brazil
Optimization of Cannabis Grows Using Fourier Transform Mid-Infrared Spectroscopy
Ad

Similar to PAMS: A Study of Performance in Low and High Humidity Environments (20)

PDF
American Ecotech VOC PAMS Webinar
PDF
EPA CFR 40. Appendix A-2 to part 50..pdf
PDF
Specifications gc
PPTX
GAS CHROMATOGRAPHY.pptx
PPTX
Gas chromatography and its instrumentation
PDF
MRU Nova Plus Emissions Analyzer | Portable Flue Gas Analyzer
PPTX
Gas chromatography
PPTX
GAS CHROMATOGRAPHY
PPTX
PPTX
Gas Chromatography
PPT
chapter-24.ppty5y54y645654645645645y5t55
PPT
Cromatografía de gases principios y fundamentos
PDF
ATIs-F12-D-Toxic-Gas-Transmitter.pdf
PPTX
GAS CHROMATOGRAPHY in pharmaceutical product development
PDF
MG 2030 CEM Datasheet
PPTX
Gas chromatography by Dr.Rashmi Agarwal
PPTX
GAS CHROMATOGRAPHY.pptx
PPT
Gas chromatography
PPTX
Mahesh g ame
PDF
Sulfur dioxide monitoring
American Ecotech VOC PAMS Webinar
EPA CFR 40. Appendix A-2 to part 50..pdf
Specifications gc
GAS CHROMATOGRAPHY.pptx
Gas chromatography and its instrumentation
MRU Nova Plus Emissions Analyzer | Portable Flue Gas Analyzer
Gas chromatography
GAS CHROMATOGRAPHY
Gas Chromatography
chapter-24.ppty5y54y645654645645645y5t55
Cromatografía de gases principios y fundamentos
ATIs-F12-D-Toxic-Gas-Transmitter.pdf
GAS CHROMATOGRAPHY in pharmaceutical product development
MG 2030 CEM Datasheet
Gas chromatography by Dr.Rashmi Agarwal
GAS CHROMATOGRAPHY.pptx
Gas chromatography
Mahesh g ame
Sulfur dioxide monitoring

More from PerkinElmer, Inc. (20)

PDF
Detectando a Pré Eclampsia no 1º Trimestre
PDF
É A VEZ DO PRIME
PDF
TREC PARA A VIDA
PDF
As Tecnologias que Definem o Desempenho
PDF
GCMS Portátil: Análises em 5 minutos!
PDF
Nós transformamos complexidade em simplicidade
PDF
Janus G3: Rapidez e Automação que você precisa.
PDF
Como a PerkinElmer contribui para diagnósticos mais rápidos?
PDF
Como a PerkinElmer contribui para diagnósticos mais rápidos?
PDF
Aspirina Vs Pré Eclampsia
PDF
O Mercado da Carne no Brasil
PDF
KITS de Preparo para Biblioteca NGS
PDF
O que é X Frágil?
PDF
Escolha um parceiro e não um fornecedor
PDF
Eletroforese Capilar Automatizada
PDF
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
PDF
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
PDF
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
PDF
CG Clarus® All In One
PDF
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
Detectando a Pré Eclampsia no 1º Trimestre
É A VEZ DO PRIME
TREC PARA A VIDA
As Tecnologias que Definem o Desempenho
GCMS Portátil: Análises em 5 minutos!
Nós transformamos complexidade em simplicidade
Janus G3: Rapidez e Automação que você precisa.
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Aspirina Vs Pré Eclampsia
O Mercado da Carne no Brasil
KITS de Preparo para Biblioteca NGS
O que é X Frágil?
Escolha um parceiro e não um fornecedor
Eletroforese Capilar Automatizada
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
CG Clarus® All In One
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing

Recently uploaded (20)

PPTX
EME Aerospace.pptx basics of mechanical engineering
PPTX
STL Academy - Highlights & Impact 2020-21-v2 (1).pptx
DOCX
Biogas Balloon for Bio CNG Plants An efficient solution for biogas storage..docx
PPTX
Biodiversity of nature in environmental studies.pptx
PDF
BD4E4- DISASTER MANAGEMENT BY A.R.SIVANESH.pdf
PPTX
Environmental pollutants for natural res
PPTX
computer of health my name i d kussta lpaggyhsgd
DOCX
Double Membrane Roofs for Anaerobic Digesters Essential for capturing and sto...
PPTX
Air_Pollution_Thesis_Presentation (1).pptx
PDF
Lesson_1_Readings.pdfjjjjjjjjjjjjjjjjjjjjjjjjjjjjj
PDF
Harnessing the Earth's Heat with Geothermal Systems Installation
PPTX
Importance of good air quality and different pollutants.
PPTX
Plant Production 7.pptx in grade 7 students
PDF
The European Green Deal (EU Green Deal)
PPTX
Climate_Change_Renewable_and_Energy.pptx
PPTX
Microbial-Pathogens-and-Parasites-Their-Impact-on-Plant-Health.pptx
PDF
PAKAM TECHNOLOGY LIMTED PITCH DECK pptx.pdf
PDF
1748933543SJA_41_2_826-834 SJA Ihsan ullha.pdf
PPTX
he document discusses solid waste management. It defines different types of s...
PPTX
Definition, Causes And Effects Of Greenhouse.pptx
EME Aerospace.pptx basics of mechanical engineering
STL Academy - Highlights & Impact 2020-21-v2 (1).pptx
Biogas Balloon for Bio CNG Plants An efficient solution for biogas storage..docx
Biodiversity of nature in environmental studies.pptx
BD4E4- DISASTER MANAGEMENT BY A.R.SIVANESH.pdf
Environmental pollutants for natural res
computer of health my name i d kussta lpaggyhsgd
Double Membrane Roofs for Anaerobic Digesters Essential for capturing and sto...
Air_Pollution_Thesis_Presentation (1).pptx
Lesson_1_Readings.pdfjjjjjjjjjjjjjjjjjjjjjjjjjjjjj
Harnessing the Earth's Heat with Geothermal Systems Installation
Importance of good air quality and different pollutants.
Plant Production 7.pptx in grade 7 students
The European Green Deal (EU Green Deal)
Climate_Change_Renewable_and_Energy.pptx
Microbial-Pathogens-and-Parasites-Their-Impact-on-Plant-Health.pptx
PAKAM TECHNOLOGY LIMTED PITCH DECK pptx.pdf
1748933543SJA_41_2_826-834 SJA Ihsan ullha.pdf
he document discusses solid waste management. It defines different types of s...
Definition, Causes And Effects Of Greenhouse.pptx

PAMS: A Study of Performance in Low and High Humidity Environments

  • 1. Introduction The United States Environmental Protection Agency (USEPA) Photochemical Assessment Monitoring Stations (PAMS) are air quality monitoring locations that are used to monitor pollutant compounds in air that contribute to ground level (bad) ozone. These nonpolar compounds are typically found in the 0.1 to tens ppbv, if not greater in some cases. Since these levels are too low to detect directly, a concentration technique such as cryotrapping or chemical adsorbent trapping needs to be used. One of the main issues with any sub-ambient temperature method is the negative effect of moisture in the air sample. Quantities up to tens of milligrams of water may be present and this can cause havoc with the chromatographic system, such as the water freezing with the formation of ice causing a flow restriction or blockage. Water can also cause poor peak shape/retention of analytes on certain columns such as Alumina PLOT. Several different techniques can be used for the effective removal of water from a gas sampling stream, prior to preconcentration, such as, drying agents, semipermeable membranes and reduced temperatures (used to condense out the water). One well accepted technique uses a Nafion® (TM, DuPont™ ) membrane to dry the incoming sample stream, and is very effective. The membrane permeability is very dependent on the analyte, generally speaking, and nonpolars are retained while water and certain polars are lost across the membrane. This work is focused on humidity effects on low level sub-ppbv (0.3-1 ppbv) PAMS targets compared to the work done by G. Broadway et al¹ who studied humidity effects at a higher 10 ppbv. PAMS: A Study of Performance in Low and High Humidity Environments A P P L I C A T I O N N O T E Author: PerkinElmer Canadian Chromatographic Application Lab PerkinElmer, Inc. Woodbridge, ON Canada Gas Chromatography
  • 2. 2 Experimental PAMS standard was purchased from Linde Canada at the USEPA recommended concentration of 3-10 ppbv and was used as the stock standard. The PAMS components dilutions were dynamically created in real time using a custom made on-line diluter, designed and built by the author, using UHP Nitrogen (LN2) as the diluent. The gas was purified using an advanced triple bed carrier gas purifier filter. The on-line diluter utilized two 2000 mL/min. MKS Mass-Flo Mass flow controllers (MFC). One was used to control pure nitrogen diluent and the other was to control the flow to a 10 um stainless steel frit, inserted into a 2L HPLC solvent bottle, ¾ filled with 18 megohm DI water. The bubbler MFC outlet was teed in to the pure nitrogen MFC output. To eliminate the possibility of trace level contaminates in the water and/or associated tubing/fittings/ MFC, since we are running at sub-ppbv levels, the total diluent flow (both Humid and Dry) was scrubbed of organics by the use of two Soil Vapour Intrusion tubes (SVI, thermal desorption) in series post blending of the diluent. The now clean diluent was blended with the PAMS standard (via a 200 mL/min. MKS MFC) into a ¼” SS tubing. To reduce the effects of laminar flow and promote mixing, the tubing was reduced to 1/8” SS (6 ft). This tubing was terminated into a SS 1/8” Tee with one arm going toward the on-line accessory approx. 1 ft. away with a 15 mL/ min. flow. The remaining arm was plumbed with ¼” Bev-A-Line, as it was downstream of the sampling accessory, and at the end of this line a custom designed and manufactured humidity accessory was installed. A block diagram of the system is shown in Figure 1. The custom designed humidity accessory utilized a Honeywell Humidcon™ HIH9130-021 with a 0-100% RH range, with +/-1.7 accuracy, +/-0.6 °C temperature reading communication via I2C. The I2C data was read via a customized Atmel® 328P microprocessor and the resulting data was displayed via I2C with a 16 x 2 LCD screen. The % RH and temperature was updated at 10 Hz. Similarly, the MFC were controlled by a custom programmed Atmel 328P, using SPI controlled MCP4821 12-bit DAC. Since the MCP4821 max output is 4096 mV and the MFC requires 0-5000 mV, a TLC2272 Rail to Rail Op Amp was used to scale the DAC 0-4096 mV output by 1.25. A rotary encoder with momentary push buttons was used for the setpoint and MFC selection. The MFC readback was read by the ATMega328P 10-bit internal ADC directly without buffering. The readbacks/setpoints were displayed on a 16 x 2 LCD via I2C. An electronic block diagram showing the basic components of the MFC control is shown in Figure 2. Figure1.Diagramofadynamicdiluter. Figure2.SchematicofMFCcontrolelectronics. Table1.SystemConditionsforOzonePrecursorAnalysis. GC Clarus® 580 GC with heartcut option Column Ozone Precursor Column Set 50 m x 0.22 mm x 1 μm dimethyl siloxane column Part number 04978179 50 m x 0.32 mm alumina/Na2SO4 PLOT column Part number N6301107 Oven Conditions 46 °C for 15 min., then 5 °C/min. to 170 °C, then 15 °C/minute to 200 °C and hold for 6 min. (48-min. chromatogram) Detector Dual FID at 250 °C Carrier Gas 42 psig helium at the ATD 16 psig helium at the midpoint ATD Thermal Desorber TurboMatrix™ TD with online sampling accessory Trap Air Toxics trap packed with carbonaceous adsorbents Low -30 °C High 325 °C Transfer Line 200 °C Operating Mode On-Line at 15 mL/min. for 40 min. Split Inlet Off Outlet 2 mL/min. Data Handling TotalChrom™
  • 3. 3 MFC Settings and Flows MFC settings were 10 bit (0-1023) for the 0-5 V analog input for the MFC via the 12 bit DAC. The MFC flows were calibrated using a NIST traceable PE1000 electronic flow meter. The MFCs used were identified as MFC#1 (for Humid Flow), MFC#2 (Dry Flow) and MFC#3 (Gas Standard). Unless otherwise noted the MFC were set as follows: MFC#1, set to 100, produced a measured flow of 228 mL/min. direct at MFC output, and 263 mL/min. of humid flow at 77% RH after humidification. Since the extra volume of the water vapor at RT contributes, it is believed the difference in flow with the humidifier attached is a result of the extra backpressure due to humidifier. MFC#2, set to 100, produced a flow of 193 mL/min. MFC#3, set to 100, produced a flow of 34.7 mL/min. Dilution of standard with dry nitrogen = (193+34.7)/34.7 = 1:6.56 Dilution of standard with humidified nitrogen = (263+34.7)/34.7 = 1:8.58 Humidity Notes Since the generated humidity is passed through two SVI tubes in series, the humidity needs time to stabilize as each tube has a moisture capacity of approx. 35 mg each. This effect is described in detail by Snow et al². In this study, typically one hour stabilization time was used prior to sampling. Results Both the 77% RH and dry dilutions of the PAMS standard give similar chromatograms in both the PLOT and BP-1 separations as shown in Figures 3 and 4 respectively. Figure 5 is an overlay of four sequential chromatographic runs highlighting excellent repeatability at low levels. Figure3.PLOTchromatogramofPAMSstandardatapprox.0.3-1ppbv.Topchromatogramis77%RHandbottomisapproximately0%RH.
  • 5. For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs Copyright ©2015, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 012297_01 PKI PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com References 1. “Ozone Precursor Analysis Using a Thermal Desorption-GC System” by G. Broadway and A. Tipler available for download at www.perkinelmer.com. 2. Patient Application US 20130180315 A1 Devices, systems and methods for purging and loading sorbent tubes by Snow et al. Conclusion The ozone precursor system performance at the 0.3 ppbv-1 ppbv, with a PAMS standard under high humidity conditions, are virtually identical to that under low/zero humidity. The chromatographic peak shapes and intensities show that high humidity has little/no negative effect on the analysis of PAMS target compounds at low levels. This enables the use of dry gaseous standards to be used without the laborious preparation of humidified standards.