SlideShare a Scribd company logo
ARCCHECK:
COMPREHENSIVE EVALUATION OF THE
DIODE ARRAY PHANTOM RULES OF THUMB FOR PHANTOM USE
FOR QA
By
Vibha Chaswal, Ph.D.
ACKNOWLEDGEMENT
  Dr.

Nilendu Gupta,
  Dr. Arnab Chakrabarti
  Dr. Yi Rong,
  Michael Weldon
The Ohio State University Radiation Oncology
Department and the James Cancer Center.
Conflicts of Interest:
None, that author is aware of
Disclosures:
This work of conducted with my role as a ‘volunteer
medical physicist’ as an independent researcher
and collaborator with OSU James CC.
A detailed journal paper in JACMP is accepted for
publication in JACMP (Feb-18-2014):
(V.Chaswal, Michael Weldon et. al., “Comprehensive Commissioning and
Evaluation of the ArcCheck cylindrical diode array for VMAT patient pre-treatment
delivery QA.” )
WHY TEST?
• 
• 
• 

New software and hardware upgrades
Relatively less literature on device’s characteristics
Prior studies* reported:
=> angular and directional dependencies
=> field size dependencies
=> limitation for dosimetry of fixed arcs due to peripheral
placement of all the detector diodes (miserable failures of
narrow arcs dosimetry, as low as, globalγ(3%/3mm)<5%)

“device’s capability to catch realistic and clinically relevant
dose errors is a subject of future work.”
*(Feygelman V et al 2011, Kozalka J et al 2011,
Neilson C et al 2013)
WHY TEST?
  To

set limits on VMAT patient QA results
  Publish a comprehensive “procedures and testing
results” document for the Medical Physics
Community to refer.
  To understand the device’s nature before its
deployment as a primary QA tool for VMAT at
OSU – know its mind!
TESTING: MATERIALS
• 
• 
• 
• 
• 
• 

Linac: TrueBeam™ STx accelerator (Varian
Medical Systems, Palo Alto, CA)
Beam Energy: 6 MV beam with and without
flattening filter
Phantom: ArcCHECK phantom (Console version
1.6)
Axilliary Software: SNC Patient (version 6.2.3)
Varian Eclipse treatment planning system (TPS)
was used (version 10.0.39)
Reference dose grids: symmetric 3D dose grid size
of 2 mm x 2 mm x 2 mm with angular resolution
for both conformal arc and VMAT is set to 4
degree.
EVALUATION TESTING
• 
• 
• 
• 
• 
• 
• 
• 
• 
• 

Linac dose rate dependency
Instantaneous dose rate response of the diode,
Radiation field size dependency
Angular dependency
Couch insertion dependency
Scatter dose characterization
Stability and consistency of response
Symmetry of response
Dosimetry accuracy for fixed arcs and
Dosimetry accuracy of VMAT patient plans.
MEASUREMENT GEOMETRY
• 

• 
• 

ArcCHECK phantom set in SAD geometry for all
measurements
Central plug inserted, PMMA density assigned
Measurements made for 6MV and 6FFF beam
ANALYSIS
Procedures:
- 

Composite gamma analysis 3%/3mm and 2%/2mm
global and local gamma comparisons of TPS reference
dose grids with ArcCHECK measured dose grids,
using 10% LD threshold

- 

Diode by Diode absolute dose comparison by looking
at ‘same co-ordinate diode(s)’ in the TPS and SNC
Patient SW:
- average dose from 6 central diodes
- LHS diode versus RHS diode versus TPS
- diodes at off-axis distances etc
I.

• 
• 
• 

• 

• 

LINAC DOSE RATE DEPENDENCY
10x10 cm2 field size, 100 cm SAD geometry
50 MU for 6X and 100 MU for 6F
6X dose rates tested (MU/min): 20, 40, 200, 400,
600
6F dose rates tested (MU/min): 600, 800, 1000,
1200, 1400
Analysis: comparison of measured dose with
reference dose using average dose from 6 central
diodes
I.

LINAC DOSE RATE DEPENDENCY: RESULTS:

  The

dose output measured by ArcCHECK diodes is
stable (within 1%) over the whole range of dose rates
(20–600 MU/min for 6X and 600–1400MU/min for 6F
beam).

Rule(s) of Thumb (RoT) for QA:
ArcCHECK shows No significant linac dose-rate based
dependency.
=> Dose output stays stable for different dose-rates
II. INSTANTANEOUS DOSE RATE RESPONSE
OF THE DIODE
  10x10

cm2 field size, on a varying SAD geometry
(90, 100, 110, and 120 cm), 100MU for both 6X
and 6F energies

  Analysis:

comparison of measured dose with
reference dose using average dose from 6 central
diodes
II. INSTANTANEOUS DOSE RATE RESPONSE:
RESULT

RoT:
• 

• 

Diodes over-respond (under-estimate dose) when ArcCHECK
closer to the radiation source, by nearly 0.3%/cm on an
average
Diodes Under-respond (over-estimate dose) when farther by
nearly 0.15%/cm
III. RADIATION FIELD SIZE DEPENDENCY
  100

cm SAD geometry and 100 MU delivery for
four static field sizes (5x5, 10x10, 15x15 and
20x20 cm2)

  Analysis:

comparison of measured dose with
reference dose using average dose from 6 central
diodes
III. RADIATION FIELD SIZE DEPENDENCY:
RESULTS
RoT:
No significant
field size based
dependence.
Measurements
match TPS
within 1%
BEAM ANGLE DEPENDENCY
  Dose

data for beam-angle dependency derived
from FS data using beam’s divergence angle
  Considers angles ranging between 0.86° to nearly
6° - derived from field sizes ranging between 5x5
cm2 and 20x20 cm2
  These angles comprise the full range of the
clinically significant BEV-diode based geometry
  Data analysis using diode by diode based
comparison
  Inverse-square dependence not factored out
BEAM ANGLE DEPENDENCY: RESULTS

RoT:
-  New correction factors for angular dependence work; No
significant angular dependence different from TPS seen.
-  Within 0.86° to 6° beam incidence, the difference from TPS
is ~ 3%
COUCH INSERTION DEPENDENCY
Measurements using a 10x10 cm2 arc delivery
compared with the dose calculated in the TPS with
and without the couch insertion
Result: γ(3%/3mm) increased from 89.5% to 100%,
and γ(2%/2mm) increased from 73.1% to 95.5%, when
couch was inserted
RoT:
Always perform dose calculations in the TPS with the
couch ROI contoured.
SCATTER DOSE CHARACTERIZATION
  Diodes

over-respond to low energy photons
  measured for varying amount of scatter derived
from irradiations of field sizes of 5x5, 10x10,
15x15 and 20x20 cm2 field sizes
  at distances varying between 1 cm to 8 cm from
the field’s edge
  both axial and transverse directions.
6MV SCATTER
CHARACTERIZATION
RoT:
- Always an over
response as compared to
the TPS

- as high as 13% to 40%,
at 1 cm from field edge

-Eclipse underestimates Out of field
doses
- Scatter doses may
differ from TPS, but
may not be drastically
different from actual
doses
- Be vary of conclusions
you make!
6FFF SCATTER
CHARACTERIZATION
RoT:
- Always an over
response as compared to
the TPS

- as high as 23% to 40%,
at 1 cm from field edge

-Eclipse underestimates Out of field
doses
- Scatter doses may
differ from TPS, but
may not be drastically
different from actual
doses
- Be vary of conclusions
you make!
STABILITY AND CONSISTENCY OF RESPONSE
• 
• 

• 

A continuing procedure
Measurements acquired using static 10x10 cm2 fields
at four cardinal angles
Two-fold benefit:
a) dosimetric set-up accuracy testing and
b) device constancy measurements
Result: global-γ(3%/3mm) = 100% over 4 months of
measurements! 
SYMMETRY OF RESPONSE
  Classic

phantom flip test
  Irradiations using wide open arcs are used
  Two wide field arcs of field sizes 10x25 cm2 and
25x25 cm2, each spanning 358°, are used in the
detector flip test, 400 MU/arc are delivered while
phantom in SAD set-up
SYMMETRY OF RESPONSE: RESULT

RoT:
- Highly axially-symmetric response over long axis
- On a busy evening when you set the phantom the
wrong side in, don’t pull your hair! Not worth it.
DOSIMETRY ACCURACY FOR FIXED
NARROW ARCS

RoT:
- Narrow arcs have limited dosimetry accuracy,
more chances of a failed or just pass plan
- One dimension< 5 cm is a narrow arc
DOSIMETRY ACCURACY: WIDE OPEN ARCS

RoT:
High dosimetry accuracy demonstrated by both local
and global gamma passing rates at 3/3 and 2/2 levels
DOSIMETRY ACCURACY: VMAT PATIENT
PLANS

• 

• 

• 

All Plans a
clinical pass
RoT: 6FFF
has greater
passing rates
than 6 MV
Why RPCspine and HN
case low at
2%/2mm?
THANK YOU!!

More Related Content

PPTX
PPT
Immobilization in radiotherapy
PPTX
Total Body Irradiation (TBI) Planning
PPTX
challenges of small field dosimetry
PPTX
Immobilization techniques in SRS and SBRT
PDF
planning systems in radiotherapy
PDF
Tomotherapy
Immobilization in radiotherapy
Total Body Irradiation (TBI) Planning
challenges of small field dosimetry
Immobilization techniques in SRS and SBRT
planning systems in radiotherapy
Tomotherapy

What's hot (20)

PDF
Enhancement of clinical outcome using OBI and Cone Beam CT in Radiotherapy
PPTX
Radiotherapy machines upto cobalt 60
PPTX
Therapeutic nuclear medicine
PPT
Dosimetric calculations
PPTX
Brachytherapy dosimetry
PDF
Icru 71 electrons
PPT
Seminar 2 .beam modification devices
PPTX
Total skin electron irradiation
PPTX
IMMOBILIZATION IN RADIOTHERAPY
PPTX
Sealed radionuclides
PDF
Shielding calculation
PPTX
TRS 398 (Technical Report Series)
PDF
Linear Accelerator Acceptance, Commissioning and Annual QA
PPTX
Modern Radiotherapy
ODP
Helical Tomotherapy
PPT
Intra Operative Radiotherapy
PDF
TG-51 and Clinical Monitor Unit Calculations
PDF
Volumetric Modulated Arc Therapy
PPT
clinical applications of ldr and hdr brachytherapy
PPTX
Hemi body irradiation
Enhancement of clinical outcome using OBI and Cone Beam CT in Radiotherapy
Radiotherapy machines upto cobalt 60
Therapeutic nuclear medicine
Dosimetric calculations
Brachytherapy dosimetry
Icru 71 electrons
Seminar 2 .beam modification devices
Total skin electron irradiation
IMMOBILIZATION IN RADIOTHERAPY
Sealed radionuclides
Shielding calculation
TRS 398 (Technical Report Series)
Linear Accelerator Acceptance, Commissioning and Annual QA
Modern Radiotherapy
Helical Tomotherapy
Intra Operative Radiotherapy
TG-51 and Clinical Monitor Unit Calculations
Volumetric Modulated Arc Therapy
clinical applications of ldr and hdr brachytherapy
Hemi body irradiation
Ad

Similar to ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QA (20)

PPTX
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
PPTX
ARC CHECK USE FOR PATIENT SPECIFIC Quality assurance
PPTX
SRS PLAN METRICS BY DR KANHU CHARAN PATRO
PPTX
DR-DDR(QUALITY CONTROL TESTDigital Radiography).pptx
PPTX
Quality assurance
PDF
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
PDF
Imrt&amp;vmat
PDF
Physical and clinical aspects of brachytherapy
PDF
QUALITY-ASSURANCE---COMPUTED-TOMOGRAPHY.pdf
PPTX
Manchester system_Dr.Supriya
PPTX
Intensity Modulated Radiotherapy (IMRT) - Dr. S. Sachin
PDF
SRT AND SRS - Arun.pdf
PPTX
SRS SBRT WORKFLOW.pptx
PPTX
QUALITY CONTROL PDF.pptx
PPTX
PDF
Implementation of IAEA-AAPM code of practice for the dosimetry
PPTX
QUALITY ASSUARENCE AND CONTROL RADIOLOGY.pptx
PPTX
A rotational reference cell for high-accuracy real-time spectroscopic trace-g...
PPTX
6 Lecture 6 Control of scatter radiation.pptx
PDF
Icru recommendation; thomas macki rockwell
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
ARC CHECK USE FOR PATIENT SPECIFIC Quality assurance
SRS PLAN METRICS BY DR KANHU CHARAN PATRO
DR-DDR(QUALITY CONTROL TESTDigital Radiography).pptx
Quality assurance
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
Imrt&amp;vmat
Physical and clinical aspects of brachytherapy
QUALITY-ASSURANCE---COMPUTED-TOMOGRAPHY.pdf
Manchester system_Dr.Supriya
Intensity Modulated Radiotherapy (IMRT) - Dr. S. Sachin
SRT AND SRS - Arun.pdf
SRS SBRT WORKFLOW.pptx
QUALITY CONTROL PDF.pptx
Implementation of IAEA-AAPM code of practice for the dosimetry
QUALITY ASSUARENCE AND CONTROL RADIOLOGY.pptx
A rotational reference cell for high-accuracy real-time spectroscopic trace-g...
6 Lecture 6 Control of scatter radiation.pptx
Icru recommendation; thomas macki rockwell
Ad

More from Miami Cancer Institute (14)

PPT
Advances in Brachytherapy Treatment Planning and Delivery
PDF
IGRT: MVCBCT Calibration and Acceptance testing Procedure
PDF
Ion chambers Acceptance Testing and Evaluation
PDF
Shielding Design
PDF
Stereotactic Radiosurgery at UIHC
PDF
Special Procedures: TBI, TSET and IORT
PDF
TPS modeling and Beam commissioning at UIHC Radiation Oncology
PPT
Radiotherapy Treatment Simulation
PDF
Image Contrast, Noise, Resolution
PDF
Mutual Information Algorithm applied to rigid registration
PDF
Causes of Noise in PET imaging
PPT
Distortion Artifacts in MRI and their correction
PDF
Imaging for Radiotherapy delivery and verification
PDF
Clinical Reference Dosimetry
Advances in Brachytherapy Treatment Planning and Delivery
IGRT: MVCBCT Calibration and Acceptance testing Procedure
Ion chambers Acceptance Testing and Evaluation
Shielding Design
Stereotactic Radiosurgery at UIHC
Special Procedures: TBI, TSET and IORT
TPS modeling and Beam commissioning at UIHC Radiation Oncology
Radiotherapy Treatment Simulation
Image Contrast, Noise, Resolution
Mutual Information Algorithm applied to rigid registration
Causes of Noise in PET imaging
Distortion Artifacts in MRI and their correction
Imaging for Radiotherapy delivery and verification
Clinical Reference Dosimetry

Recently uploaded (20)

PDF
ChatGPT for Dummies - Pam Baker Ccesa007.pdf
PDF
Supply Chain Operations Speaking Notes -ICLT Program
PPTX
Introduction to Building Materials
PDF
Trump Administration's workforce development strategy
PPTX
Chinmaya Tiranga Azadi Quiz (Class 7-8 )
PDF
medical_surgical_nursing_10th_edition_ignatavicius_TEST_BANK_pdf.pdf
PPTX
Radiologic_Anatomy_of_the_Brachial_plexus [final].pptx
PPTX
UNIT III MENTAL HEALTH NURSING ASSESSMENT
PDF
Computing-Curriculum for Schools in Ghana
PDF
Hazard Identification & Risk Assessment .pdf
PDF
LDMMIA Reiki Yoga Finals Review Spring Summer
PDF
Classroom Observation Tools for Teachers
PDF
1_English_Language_Set_2.pdf probationary
PPTX
202450812 BayCHI UCSC-SV 20250812 v17.pptx
PPTX
Unit 4 Skeletal System.ppt.pptxopresentatiom
PDF
Practical Manual AGRO-233 Principles and Practices of Natural Farming
PDF
What if we spent less time fighting change, and more time building what’s rig...
PPTX
Lesson notes of climatology university.
PPTX
History, Philosophy and sociology of education (1).pptx
PPTX
CHAPTER IV. MAN AND BIOSPHERE AND ITS TOTALITY.pptx
ChatGPT for Dummies - Pam Baker Ccesa007.pdf
Supply Chain Operations Speaking Notes -ICLT Program
Introduction to Building Materials
Trump Administration's workforce development strategy
Chinmaya Tiranga Azadi Quiz (Class 7-8 )
medical_surgical_nursing_10th_edition_ignatavicius_TEST_BANK_pdf.pdf
Radiologic_Anatomy_of_the_Brachial_plexus [final].pptx
UNIT III MENTAL HEALTH NURSING ASSESSMENT
Computing-Curriculum for Schools in Ghana
Hazard Identification & Risk Assessment .pdf
LDMMIA Reiki Yoga Finals Review Spring Summer
Classroom Observation Tools for Teachers
1_English_Language_Set_2.pdf probationary
202450812 BayCHI UCSC-SV 20250812 v17.pptx
Unit 4 Skeletal System.ppt.pptxopresentatiom
Practical Manual AGRO-233 Principles and Practices of Natural Farming
What if we spent less time fighting change, and more time building what’s rig...
Lesson notes of climatology university.
History, Philosophy and sociology of education (1).pptx
CHAPTER IV. MAN AND BIOSPHERE AND ITS TOTALITY.pptx

ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QA

  • 1. ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM RULES OF THUMB FOR PHANTOM USE FOR QA By Vibha Chaswal, Ph.D.
  • 2. ACKNOWLEDGEMENT   Dr. Nilendu Gupta,   Dr. Arnab Chakrabarti   Dr. Yi Rong,   Michael Weldon The Ohio State University Radiation Oncology Department and the James Cancer Center.
  • 3. Conflicts of Interest: None, that author is aware of Disclosures: This work of conducted with my role as a ‘volunteer medical physicist’ as an independent researcher and collaborator with OSU James CC. A detailed journal paper in JACMP is accepted for publication in JACMP (Feb-18-2014): (V.Chaswal, Michael Weldon et. al., “Comprehensive Commissioning and Evaluation of the ArcCheck cylindrical diode array for VMAT patient pre-treatment delivery QA.” )
  • 4. WHY TEST? •  •  •  New software and hardware upgrades Relatively less literature on device’s characteristics Prior studies* reported: => angular and directional dependencies => field size dependencies => limitation for dosimetry of fixed arcs due to peripheral placement of all the detector diodes (miserable failures of narrow arcs dosimetry, as low as, globalγ(3%/3mm)<5%) “device’s capability to catch realistic and clinically relevant dose errors is a subject of future work.” *(Feygelman V et al 2011, Kozalka J et al 2011, Neilson C et al 2013)
  • 5. WHY TEST?   To set limits on VMAT patient QA results   Publish a comprehensive “procedures and testing results” document for the Medical Physics Community to refer.   To understand the device’s nature before its deployment as a primary QA tool for VMAT at OSU – know its mind!
  • 6. TESTING: MATERIALS •  •  •  •  •  •  Linac: TrueBeam™ STx accelerator (Varian Medical Systems, Palo Alto, CA) Beam Energy: 6 MV beam with and without flattening filter Phantom: ArcCHECK phantom (Console version 1.6) Axilliary Software: SNC Patient (version 6.2.3) Varian Eclipse treatment planning system (TPS) was used (version 10.0.39) Reference dose grids: symmetric 3D dose grid size of 2 mm x 2 mm x 2 mm with angular resolution for both conformal arc and VMAT is set to 4 degree.
  • 7. EVALUATION TESTING •  •  •  •  •  •  •  •  •  •  Linac dose rate dependency Instantaneous dose rate response of the diode, Radiation field size dependency Angular dependency Couch insertion dependency Scatter dose characterization Stability and consistency of response Symmetry of response Dosimetry accuracy for fixed arcs and Dosimetry accuracy of VMAT patient plans.
  • 8. MEASUREMENT GEOMETRY •  •  •  ArcCHECK phantom set in SAD geometry for all measurements Central plug inserted, PMMA density assigned Measurements made for 6MV and 6FFF beam
  • 9. ANALYSIS Procedures: -  Composite gamma analysis 3%/3mm and 2%/2mm global and local gamma comparisons of TPS reference dose grids with ArcCHECK measured dose grids, using 10% LD threshold -  Diode by Diode absolute dose comparison by looking at ‘same co-ordinate diode(s)’ in the TPS and SNC Patient SW: - average dose from 6 central diodes - LHS diode versus RHS diode versus TPS - diodes at off-axis distances etc
  • 10. I. •  •  •  •  •  LINAC DOSE RATE DEPENDENCY 10x10 cm2 field size, 100 cm SAD geometry 50 MU for 6X and 100 MU for 6F 6X dose rates tested (MU/min): 20, 40, 200, 400, 600 6F dose rates tested (MU/min): 600, 800, 1000, 1200, 1400 Analysis: comparison of measured dose with reference dose using average dose from 6 central diodes
  • 11. I. LINAC DOSE RATE DEPENDENCY: RESULTS:   The dose output measured by ArcCHECK diodes is stable (within 1%) over the whole range of dose rates (20–600 MU/min for 6X and 600–1400MU/min for 6F beam). Rule(s) of Thumb (RoT) for QA: ArcCHECK shows No significant linac dose-rate based dependency. => Dose output stays stable for different dose-rates
  • 12. II. INSTANTANEOUS DOSE RATE RESPONSE OF THE DIODE   10x10 cm2 field size, on a varying SAD geometry (90, 100, 110, and 120 cm), 100MU for both 6X and 6F energies   Analysis: comparison of measured dose with reference dose using average dose from 6 central diodes
  • 13. II. INSTANTANEOUS DOSE RATE RESPONSE: RESULT RoT: •  •  Diodes over-respond (under-estimate dose) when ArcCHECK closer to the radiation source, by nearly 0.3%/cm on an average Diodes Under-respond (over-estimate dose) when farther by nearly 0.15%/cm
  • 14. III. RADIATION FIELD SIZE DEPENDENCY   100 cm SAD geometry and 100 MU delivery for four static field sizes (5x5, 10x10, 15x15 and 20x20 cm2)   Analysis: comparison of measured dose with reference dose using average dose from 6 central diodes
  • 15. III. RADIATION FIELD SIZE DEPENDENCY: RESULTS RoT: No significant field size based dependence. Measurements match TPS within 1%
  • 16. BEAM ANGLE DEPENDENCY   Dose data for beam-angle dependency derived from FS data using beam’s divergence angle   Considers angles ranging between 0.86° to nearly 6° - derived from field sizes ranging between 5x5 cm2 and 20x20 cm2   These angles comprise the full range of the clinically significant BEV-diode based geometry   Data analysis using diode by diode based comparison   Inverse-square dependence not factored out
  • 17. BEAM ANGLE DEPENDENCY: RESULTS RoT: -  New correction factors for angular dependence work; No significant angular dependence different from TPS seen. -  Within 0.86° to 6° beam incidence, the difference from TPS is ~ 3%
  • 18. COUCH INSERTION DEPENDENCY Measurements using a 10x10 cm2 arc delivery compared with the dose calculated in the TPS with and without the couch insertion Result: γ(3%/3mm) increased from 89.5% to 100%, and γ(2%/2mm) increased from 73.1% to 95.5%, when couch was inserted RoT: Always perform dose calculations in the TPS with the couch ROI contoured.
  • 19. SCATTER DOSE CHARACTERIZATION   Diodes over-respond to low energy photons   measured for varying amount of scatter derived from irradiations of field sizes of 5x5, 10x10, 15x15 and 20x20 cm2 field sizes   at distances varying between 1 cm to 8 cm from the field’s edge   both axial and transverse directions.
  • 20. 6MV SCATTER CHARACTERIZATION RoT: - Always an over response as compared to the TPS - as high as 13% to 40%, at 1 cm from field edge -Eclipse underestimates Out of field doses - Scatter doses may differ from TPS, but may not be drastically different from actual doses - Be vary of conclusions you make!
  • 21. 6FFF SCATTER CHARACTERIZATION RoT: - Always an over response as compared to the TPS - as high as 23% to 40%, at 1 cm from field edge -Eclipse underestimates Out of field doses - Scatter doses may differ from TPS, but may not be drastically different from actual doses - Be vary of conclusions you make!
  • 22. STABILITY AND CONSISTENCY OF RESPONSE •  •  •  A continuing procedure Measurements acquired using static 10x10 cm2 fields at four cardinal angles Two-fold benefit: a) dosimetric set-up accuracy testing and b) device constancy measurements Result: global-γ(3%/3mm) = 100% over 4 months of measurements! 
  • 23. SYMMETRY OF RESPONSE   Classic phantom flip test   Irradiations using wide open arcs are used   Two wide field arcs of field sizes 10x25 cm2 and 25x25 cm2, each spanning 358°, are used in the detector flip test, 400 MU/arc are delivered while phantom in SAD set-up
  • 24. SYMMETRY OF RESPONSE: RESULT RoT: - Highly axially-symmetric response over long axis - On a busy evening when you set the phantom the wrong side in, don’t pull your hair! Not worth it.
  • 25. DOSIMETRY ACCURACY FOR FIXED NARROW ARCS RoT: - Narrow arcs have limited dosimetry accuracy, more chances of a failed or just pass plan - One dimension< 5 cm is a narrow arc
  • 26. DOSIMETRY ACCURACY: WIDE OPEN ARCS RoT: High dosimetry accuracy demonstrated by both local and global gamma passing rates at 3/3 and 2/2 levels
  • 27. DOSIMETRY ACCURACY: VMAT PATIENT PLANS •  •  •  All Plans a clinical pass RoT: 6FFF has greater passing rates than 6 MV Why RPCspine and HN case low at 2%/2mm?