Chapter 10
Measurement of hydrogen ion concentration (pH)
1. General
pH is a way of expressing the hydrogen ion concentration in water. It is related to the
acidic or alkaline nature of water. Consideration of hydrogen ion concentration is important
in almost all uses of water. In particular, pH balance is important in maintaining desirable
aquatic ecological conditions in natural waters. pH is also maintained at various levels for
efficient operation of water and wastewater treatment systems such as coagulation,
disinfecting, softening, anaerobic decomposition of wastes, etc. The pH of most natural
waters lies between 6.5 and 8.
2. The pH scale
Pure water dissociates to yield 10"7
moles/L of H+
at 25 °C:
H 2 0 ~ H +
+ 0 H " (1)
Since water dissociates to produce one OH" ion for each H+
ion, it is obvious that 10"7
OH"
ions are produced simultaneously.
The product of [H+
] and [OH"] always remains constant. When the value for one of the
species changes the other also changes accordingly.
[H+
] x [OH] = 10"14
(2)
The large bracket sign, [ ], indicates molar concentration.
The concentration of H+
ions can be increased when compounds are added which release
H+
ions such as H2SO4:
H2SO4—• 2H+
+ S04
2
" (3)
Or preferentially combine with OH" ions when added to water, such as FeCl3.
FeCI3 + 3H2 0 —• Fe(OH)3 + 3H+
+ 3CI" (4)
Note that in either case the product of /H+
7and [OH"] ions remains constant at 10"14
Likewise the concentration of H+
ions can be decreased when compounds are added
which release OH" ions, such as NaOH, or preferentially combine with H+
ions, such as
Na2C03.
Expression of the molar concentration of hydrogen ions is rather cumbersome because of
the extremely small values and large variations. To overcome this difficulty, the
concentration is expressed in terms of pH value, which is negative logarithm of the
concentration in moles/L.
pH = - log [H+
] (5)
or [H+
] = 10pH
(6)
Thus the pH value of a sample of water containing 10"7 5
moles/L H+
is 7.5.
Since the product of [H+
] and [OH"] is always a constant, Eq. (2) becomes
log[H+
] +log[OH"]= -14
Or -log[H+
] - log [OH1 = +14
The pH scale is usually represented as ranging from 0 to 14, with pH 7 at 25 °C
representing neutrality. Acid conditions increase as pH values decrease and alkaline
conditions increase as pH values increase.
Acid range Neutral Alkaline range
1 2 3 4 5 6 7 8 9 10 11 12 13 14
3. Calculating pH
Example calculation
Calculate the pH value for the following cases:
A: [H+
]= a.0.0001
b.0.00005 and
B: [OH"] a 10"8
b. 0.00005 and
c.0.00001
A: [H+
} a. 0.0001
A. 0.00005
pH= -log 10"3
=3
pH= -log 0.00005= -log (5 x10"5
)= - 0.7 +5=4.3
B:[OH"] a. 10"8
b. 0.00005
c. 0.00001
[H+
] x 10"8
= 10"14
, or [H+
] =10"6
Therefore pH=6
—log[H+
]—log [OH"] =14
Therefore pH=14-4.3 =9.7
pH=14 + Iog10"5
=14- 5 =9
4. pH indicators
A number of naturally occurring or synthetic organic compounds undergo definite colour
changes in well-defined pH ranges. A number of indicators that are useful for various pH
ranges are listed in Table 1. The indicators assume different hues within the specified pH
range. These are used as liquid solutions or some as pH papers. The change in colour
occurs over a wide range of pH change and therefore pH value cannot be measured
accurately. Further, the turbidity and colour of the sample may cause interference.
Table 1
Indicator Acid colour Base colour pH
range
Methyl orange Red Yellow orange 3.1-4.6
Methyl red Red Yellow 4.4-6.2
Litmus Red Blue 4.5-8.3
Thymol blue Yellow Blue 8.0-9.6
Phenolphthalein Colourless Pink 8.2-9.8
Alizarin yellow Yellow Lilac 10.1-
11.1
5. pH meter
The use of colour indicators for pH measurements has, to some extent, been superseded
by development of glass electrode. pH meters employing glass-indicating electrodes and
saturated calomel reference electrodes are now commonly used. Such meters are capable
of measuring pH within + 0.1 pH unit. The electrodes, connected to the pH meters are
immersed in the sample and the meter measures the potential developed at the glass
electrode due to hydrogen ion concentration in the sample and displays it directly in pH
units. The pH meters are also equipped with a temperature-compensation adjustment.
The electrodes should be carefully handled and should not be scratched by butting against
the sides of the beaker containing the sample. Follow the manufacturer's instructions for
their care during storage and use. For short-term storage, the glass electrode may be left
immersed in pH 4 buffer solution., Saturated KCI is preferred for the reference electrode.
The glass electrode needs to be soaked in water for at least 12 hours before it is used for
pH measurement.
Buffer solutions & instrument calibration
pH meters have to be calibrated against solutions of known pH values. Standard buffers
are used for this purpose. Buffers are solutions of chemicals of known pH which do not
change their pH value upon dilution and resist change of pH when small amounts of acid
or alkali are added to them. Buffer solutions usually contain mixtures of weak acids and
their salts (conjugate bases) or weak bases and their salts (conjugate acids). Buffer has
importance for life forms that usually can survive only within a narrow pH range. Table 2
gives the composition of some commonly used buffers. Buffers of different pH values are
also available commercially.
Table 2 Buffer solutions of known pH
S. No Buffer solution pH at
25°C
Amount of salt to be dissolved in
1000 ml freshly boiled and cooled
and cooled distilled water
1. 0.05 M potassium hydrogen phthalate 4.00 10.12 g KHC8H404
2. 0.025 M potassium dihydrogen
phosphate + 0.025 M disodium
hydrogen phosphate
6.86 3.387 g KH2P04 and 3.533 g Na2HP04
3. 0.01 M sodium borate decahydrate 9.18 3.80 g Na2B407.10 H2 0
It is a good practice to calibrate the pH meter using two buffers. After calibrating with an
initial buffer, use a second buffer within 2 pH units of the sample pH.
OPERATION OF THE pH METER
How to measure the pH
1. Connect the pH electrode & temperature sensor to the measuring instrument.
2. Press the on/off Key.(display test appears briefly on the display)
3. Select pH value or redox voltage (mv) with <M>.
4. Immerse the pH electrode in water sample .
5. Press <AR>to switch on the drift control. Wait until measured value is stable and AR
stops flashing.
6. To cancel auto read at any time press <Run/Enter>.
Storing the data
1.Press <STO>Key in the measuring mode(display No. with the number of the next free
memory location.
2. Press <Run/Enter>.
3. Enter the Identification number with < V> <A>.
4. Terminate the save with <Run/Enter>.
How to see data memory
1. Press the <RCL> key in the measuring mode.(display Sto disp)
2. Press <Run/Enter>key (display number at which data store)
3. Press <Run/Enter>key(display identification number)
4. Press <Run/Enter>key (display day .month)
5. Press <Run/Enter>key (display time)
6. Press <M>key to return in measuring mode.
Calibration
1. The pH electrode to the measuring instrument.
2. Press the <Cal>key repeatedly untill the display Ct1 and the function display Auto
Cal TEC appears. The sensor symbol displays the evaluation of the last calibration.
3. Immerse the pH electrode in to the first buffer solution (pH-7.0)
4. Press <Run/Enter>key.(the auto read measurement begins. If the measured value
is stable, Ct2 appears.
Note:--At this point, the auto cal TEC calibration can be terminated with <M>.
1. Immerse the two pH electrode in the second buffer solution to continue.
2. Press <Run/Enter>key(the auto read measurement begins.) If the measured value
is stable, the instrument displays the value of the slope and the evaluation of the
two point calibration.
3. Press<Run/Enter>key.(the instrument displays the value of the asymmetry)
4. Switch to the measuring mode with<M>.
5. To set calibration interval, while pressing the <M>key, press <ON/OFF>Key.(display
CAL disp)
6. Press <Run/Enter>key.(display time of calibration display)
7. Set the interval of Calibration with <A > <V >.
Precautions
1. Always keep the electrode wet(with electrolyte in the cap of electrode),if you use it
after 2 or 3 days interval.
2. Keep it in ordinary water if you use continuously.
3. Clean properly with absorbent paper before and after use.
4. Need to Calibrate after a fixed interval.

More Related Content

PDF
Measurement of hydrogen ion concentration (p h)
PPT
Indicators
PPT
P h indicators
PPTX
Ph indicator-----(Pharmaceutics)
PPSX
Indicator
PPTX
P h indicator
PPTX
Acid base indicator
PPTX
PH measurement in Food and drink
Measurement of hydrogen ion concentration (p h)
Indicators
P h indicators
Ph indicator-----(Pharmaceutics)
Indicator
P h indicator
Acid base indicator
PH measurement in Food and drink

What's hot (19)

PPT
Lab acid titrate
DOC
Experiment5 chem210
PDF
Lab 3 acid base titration curves and acid_base indicators
PPTX
Beetroot as an indicator
PPTX
chemical Indicators
PPTX
Theory of Indicators/choice of indicator/acid base indicator
PDF
Table top microprocessor p h meters
PDF
Susb012
PPTX
Theory of indicator
PPTX
pH indicator
PPTX
Ph and ph meter and application in foods
PPTX
Titration
PPTX
Manish ppt
PPTX
UNIT 5TH - pH, buffers and Isotonic solutions
PPTX
Acid-Base Titrations
PPT
Titration ppt
PPS
Indicators
PPTX
Acid base titrations
PPTX
P h indicator
Lab acid titrate
Experiment5 chem210
Lab 3 acid base titration curves and acid_base indicators
Beetroot as an indicator
chemical Indicators
Theory of Indicators/choice of indicator/acid base indicator
Table top microprocessor p h meters
Susb012
Theory of indicator
pH indicator
Ph and ph meter and application in foods
Titration
Manish ppt
UNIT 5TH - pH, buffers and Isotonic solutions
Acid-Base Titrations
Titration ppt
Indicators
Acid base titrations
P h indicator
Ad

Similar to Chapter 10 (20)

PDF
Introduction to-ph
PPTX
PDF
Potentio lab report
PDF
Drilling Fluid Engineering-PH Lab-Drilling Lab#9.pdf
PPTX
Preparation of buffers and estimation of pH by using pH meter.pptx
PDF
Volumetric Analysis By Dr.A. Amsavel
PDF
Volumetric Analysis by Dr. A. Amsavel
PPTX
buffer and_ ph _meter_biochemistry .pptx
PPT
1. Acid- Base Titration.ppt
PPTX
Ph,buffers,balance
PDF
Chemistry Lab Report 2
PDF
Analisis de agua merck (1)
PPT
pH - MEASUREMENT (ph meter, ph determination).ppt
PPT
pH - MEASUREMENT EXPERIMENT PROCEDURE .ppt
PPT
Analytical Measurements: Troubleshooting, Maintenance and the Future
PPT
pH - MEASUREMENT .ppt
PPTX
Unit 5 pH, buffers.pRESENTATION SEMII B.
DOCX
AC lab 7 cover and report
PDF
P h technical handbook
DOCX
Exp 1 (A)
Introduction to-ph
Potentio lab report
Drilling Fluid Engineering-PH Lab-Drilling Lab#9.pdf
Preparation of buffers and estimation of pH by using pH meter.pptx
Volumetric Analysis By Dr.A. Amsavel
Volumetric Analysis by Dr. A. Amsavel
buffer and_ ph _meter_biochemistry .pptx
1. Acid- Base Titration.ppt
Ph,buffers,balance
Chemistry Lab Report 2
Analisis de agua merck (1)
pH - MEASUREMENT (ph meter, ph determination).ppt
pH - MEASUREMENT EXPERIMENT PROCEDURE .ppt
Analytical Measurements: Troubleshooting, Maintenance and the Future
pH - MEASUREMENT .ppt
Unit 5 pH, buffers.pRESENTATION SEMII B.
AC lab 7 cover and report
P h technical handbook
Exp 1 (A)
Ad

More from ECRD IN (20)

PDF
Spectrophotometer Meter
PDF
Quality Assurance and Quality Control
PDF
Basic Chemistry Concepts
PDF
Measurement of Respirable Suspended Pm10 Particles
PDF
Measurement of Hydrogen Ion Concentration (pH)
PDF
Determination of Suspended PM in Atmosphere
PDF
Basic Statistics Concepts
PDF
Air Pollution, Sources and Characteristics
PDF
Air Pollution Its Origin and Effects
PDF
Air Pollution and Meteorology
PDF
Dust-Caused Respiratory Occupational Diseases
PDF
Air Quality Monitoring
PDF
Approach & Strategy to Meet new Ambient Air Quality Standards
PDF
Monitoring CO -NDIR Method
PDF
Appropriate Instruments & techniques for Complying with Air Quality Standards
PDF
Techniques of Measurement of Organic Pollutants
PDF
Envirotech APM 460 BL
PDF
Monitoring of Gaseous Pollutants
PDF
Filter Extraction of Heavy Metal & Benzo Pyrene
PDF
Maintenance of High Volume Dust Repairable Sampler
Spectrophotometer Meter
Quality Assurance and Quality Control
Basic Chemistry Concepts
Measurement of Respirable Suspended Pm10 Particles
Measurement of Hydrogen Ion Concentration (pH)
Determination of Suspended PM in Atmosphere
Basic Statistics Concepts
Air Pollution, Sources and Characteristics
Air Pollution Its Origin and Effects
Air Pollution and Meteorology
Dust-Caused Respiratory Occupational Diseases
Air Quality Monitoring
Approach & Strategy to Meet new Ambient Air Quality Standards
Monitoring CO -NDIR Method
Appropriate Instruments & techniques for Complying with Air Quality Standards
Techniques of Measurement of Organic Pollutants
Envirotech APM 460 BL
Monitoring of Gaseous Pollutants
Filter Extraction of Heavy Metal & Benzo Pyrene
Maintenance of High Volume Dust Repairable Sampler

Recently uploaded (20)

DOCX
Biogas Tank for Breweries & Food Processing The main container for biogas dig...
DOCX
Aluminum Dome Roofs for Drinking Water Tanks Shield Water from Debris and Pol...
PPTX
EME Aerospace.pptx basics of mechanical engineering
PDF
PAKAM TECHNOLOGY LIMTED PITCH DECK pptx.pdf
PDF
Pollution Fees in case of Imperfect Competition
DOCX
Biogas Balloon for Bio CNG Plants An efficient solution for biogas storage..docx
PPTX
-Case-Study 2 What Happened in the Cavite Mutiny (1).pptx
PPTX
14.1 Opinion Essay (Writing). to teach opinion
PDF
Lesson_1_Readings.pdfjjjjjjjjjjjjjjjjjjjjjjjjjjjjj
PPTX
Climate_Change_Renewable_and_Energy.pptx
DOCX
Double Membrane Roofs for Biogas Digesters A sealed cover for biogas producti...
DOCX
Biogas Balloon for Pig Farm Plants Holds biogas from hog farm waste..docx
PPTX
Understanding Socialism and people. Revised.pptx
PDF
Biomass cookstoves: A review of technical aspects
PDF
rainfed swc for nature and agroforestrys
PPTX
Advances in Integrated Nutrient and Insect-pest Management in Vegetable Crops...
DOCX
The Ripple Effect: Understanding Extreme Weather Patterns and Geomagnetic Dyn...
PPTX
SCADAhjknvbxfbgmmmmmmmmmmmmmmmmmmmmmmm.pptx
PPTX
computer of health my name i d kussta lpaggyhsgd
PPTX
Biodiversity PPT by Gaithanlung Gonmei.pptx
Biogas Tank for Breweries & Food Processing The main container for biogas dig...
Aluminum Dome Roofs for Drinking Water Tanks Shield Water from Debris and Pol...
EME Aerospace.pptx basics of mechanical engineering
PAKAM TECHNOLOGY LIMTED PITCH DECK pptx.pdf
Pollution Fees in case of Imperfect Competition
Biogas Balloon for Bio CNG Plants An efficient solution for biogas storage..docx
-Case-Study 2 What Happened in the Cavite Mutiny (1).pptx
14.1 Opinion Essay (Writing). to teach opinion
Lesson_1_Readings.pdfjjjjjjjjjjjjjjjjjjjjjjjjjjjjj
Climate_Change_Renewable_and_Energy.pptx
Double Membrane Roofs for Biogas Digesters A sealed cover for biogas producti...
Biogas Balloon for Pig Farm Plants Holds biogas from hog farm waste..docx
Understanding Socialism and people. Revised.pptx
Biomass cookstoves: A review of technical aspects
rainfed swc for nature and agroforestrys
Advances in Integrated Nutrient and Insect-pest Management in Vegetable Crops...
The Ripple Effect: Understanding Extreme Weather Patterns and Geomagnetic Dyn...
SCADAhjknvbxfbgmmmmmmmmmmmmmmmmmmmmmmm.pptx
computer of health my name i d kussta lpaggyhsgd
Biodiversity PPT by Gaithanlung Gonmei.pptx

Chapter 10

  • 1. Chapter 10 Measurement of hydrogen ion concentration (pH) 1. General pH is a way of expressing the hydrogen ion concentration in water. It is related to the acidic or alkaline nature of water. Consideration of hydrogen ion concentration is important in almost all uses of water. In particular, pH balance is important in maintaining desirable aquatic ecological conditions in natural waters. pH is also maintained at various levels for efficient operation of water and wastewater treatment systems such as coagulation, disinfecting, softening, anaerobic decomposition of wastes, etc. The pH of most natural waters lies between 6.5 and 8. 2. The pH scale Pure water dissociates to yield 10"7 moles/L of H+ at 25 °C: H 2 0 ~ H + + 0 H " (1) Since water dissociates to produce one OH" ion for each H+ ion, it is obvious that 10"7 OH" ions are produced simultaneously. The product of [H+ ] and [OH"] always remains constant. When the value for one of the species changes the other also changes accordingly. [H+ ] x [OH] = 10"14 (2) The large bracket sign, [ ], indicates molar concentration. The concentration of H+ ions can be increased when compounds are added which release H+ ions such as H2SO4: H2SO4—• 2H+ + S04 2 " (3) Or preferentially combine with OH" ions when added to water, such as FeCl3. FeCI3 + 3H2 0 —• Fe(OH)3 + 3H+ + 3CI" (4) Note that in either case the product of /H+ 7and [OH"] ions remains constant at 10"14 Likewise the concentration of H+ ions can be decreased when compounds are added which release OH" ions, such as NaOH, or preferentially combine with H+ ions, such as Na2C03. Expression of the molar concentration of hydrogen ions is rather cumbersome because of
  • 2. the extremely small values and large variations. To overcome this difficulty, the concentration is expressed in terms of pH value, which is negative logarithm of the concentration in moles/L. pH = - log [H+ ] (5) or [H+ ] = 10pH (6) Thus the pH value of a sample of water containing 10"7 5 moles/L H+ is 7.5. Since the product of [H+ ] and [OH"] is always a constant, Eq. (2) becomes log[H+ ] +log[OH"]= -14 Or -log[H+ ] - log [OH1 = +14 The pH scale is usually represented as ranging from 0 to 14, with pH 7 at 25 °C representing neutrality. Acid conditions increase as pH values decrease and alkaline conditions increase as pH values increase. Acid range Neutral Alkaline range 1 2 3 4 5 6 7 8 9 10 11 12 13 14 3. Calculating pH Example calculation Calculate the pH value for the following cases: A: [H+ ]= a.0.0001 b.0.00005 and B: [OH"] a 10"8 b. 0.00005 and c.0.00001 A: [H+ } a. 0.0001 A. 0.00005 pH= -log 10"3 =3 pH= -log 0.00005= -log (5 x10"5 )= - 0.7 +5=4.3 B:[OH"] a. 10"8 b. 0.00005 c. 0.00001 [H+ ] x 10"8 = 10"14 , or [H+ ] =10"6 Therefore pH=6 —log[H+ ]—log [OH"] =14 Therefore pH=14-4.3 =9.7 pH=14 + Iog10"5 =14- 5 =9
  • 3. 4. pH indicators A number of naturally occurring or synthetic organic compounds undergo definite colour changes in well-defined pH ranges. A number of indicators that are useful for various pH ranges are listed in Table 1. The indicators assume different hues within the specified pH range. These are used as liquid solutions or some as pH papers. The change in colour occurs over a wide range of pH change and therefore pH value cannot be measured accurately. Further, the turbidity and colour of the sample may cause interference. Table 1 Indicator Acid colour Base colour pH range Methyl orange Red Yellow orange 3.1-4.6 Methyl red Red Yellow 4.4-6.2 Litmus Red Blue 4.5-8.3 Thymol blue Yellow Blue 8.0-9.6 Phenolphthalein Colourless Pink 8.2-9.8 Alizarin yellow Yellow Lilac 10.1- 11.1 5. pH meter The use of colour indicators for pH measurements has, to some extent, been superseded by development of glass electrode. pH meters employing glass-indicating electrodes and saturated calomel reference electrodes are now commonly used. Such meters are capable of measuring pH within + 0.1 pH unit. The electrodes, connected to the pH meters are immersed in the sample and the meter measures the potential developed at the glass electrode due to hydrogen ion concentration in the sample and displays it directly in pH units. The pH meters are also equipped with a temperature-compensation adjustment. The electrodes should be carefully handled and should not be scratched by butting against the sides of the beaker containing the sample. Follow the manufacturer's instructions for their care during storage and use. For short-term storage, the glass electrode may be left immersed in pH 4 buffer solution., Saturated KCI is preferred for the reference electrode. The glass electrode needs to be soaked in water for at least 12 hours before it is used for pH measurement. Buffer solutions & instrument calibration pH meters have to be calibrated against solutions of known pH values. Standard buffers are used for this purpose. Buffers are solutions of chemicals of known pH which do not
  • 4. change their pH value upon dilution and resist change of pH when small amounts of acid or alkali are added to them. Buffer solutions usually contain mixtures of weak acids and their salts (conjugate bases) or weak bases and their salts (conjugate acids). Buffer has importance for life forms that usually can survive only within a narrow pH range. Table 2 gives the composition of some commonly used buffers. Buffers of different pH values are also available commercially. Table 2 Buffer solutions of known pH S. No Buffer solution pH at 25°C Amount of salt to be dissolved in 1000 ml freshly boiled and cooled and cooled distilled water 1. 0.05 M potassium hydrogen phthalate 4.00 10.12 g KHC8H404 2. 0.025 M potassium dihydrogen phosphate + 0.025 M disodium hydrogen phosphate 6.86 3.387 g KH2P04 and 3.533 g Na2HP04 3. 0.01 M sodium borate decahydrate 9.18 3.80 g Na2B407.10 H2 0 It is a good practice to calibrate the pH meter using two buffers. After calibrating with an initial buffer, use a second buffer within 2 pH units of the sample pH.
  • 5. OPERATION OF THE pH METER How to measure the pH 1. Connect the pH electrode & temperature sensor to the measuring instrument. 2. Press the on/off Key.(display test appears briefly on the display) 3. Select pH value or redox voltage (mv) with <M>. 4. Immerse the pH electrode in water sample . 5. Press <AR>to switch on the drift control. Wait until measured value is stable and AR stops flashing. 6. To cancel auto read at any time press <Run/Enter>. Storing the data 1.Press <STO>Key in the measuring mode(display No. with the number of the next free memory location. 2. Press <Run/Enter>. 3. Enter the Identification number with < V> <A>. 4. Terminate the save with <Run/Enter>. How to see data memory 1. Press the <RCL> key in the measuring mode.(display Sto disp) 2. Press <Run/Enter>key (display number at which data store) 3. Press <Run/Enter>key(display identification number) 4. Press <Run/Enter>key (display day .month) 5. Press <Run/Enter>key (display time) 6. Press <M>key to return in measuring mode.
  • 6. Calibration 1. The pH electrode to the measuring instrument. 2. Press the <Cal>key repeatedly untill the display Ct1 and the function display Auto Cal TEC appears. The sensor symbol displays the evaluation of the last calibration. 3. Immerse the pH electrode in to the first buffer solution (pH-7.0) 4. Press <Run/Enter>key.(the auto read measurement begins. If the measured value is stable, Ct2 appears. Note:--At this point, the auto cal TEC calibration can be terminated with <M>. 1. Immerse the two pH electrode in the second buffer solution to continue. 2. Press <Run/Enter>key(the auto read measurement begins.) If the measured value is stable, the instrument displays the value of the slope and the evaluation of the two point calibration. 3. Press<Run/Enter>key.(the instrument displays the value of the asymmetry) 4. Switch to the measuring mode with<M>. 5. To set calibration interval, while pressing the <M>key, press <ON/OFF>Key.(display CAL disp) 6. Press <Run/Enter>key.(display time of calibration display) 7. Set the interval of Calibration with <A > <V >. Precautions 1. Always keep the electrode wet(with electrolyte in the cap of electrode),if you use it after 2 or 3 days interval. 2. Keep it in ordinary water if you use continuously. 3. Clean properly with absorbent paper before and after use. 4. Need to Calibrate after a fixed interval.