Thermal Characterization of
Polymers/Polymeric Materials
By
Dr Shahzad Ali Shahid Chatha
Professor of Analytical Chemistry
Chapter 10 Thermal Analysis of Polymers.ppt
Chapter 10 Thermal Analysis of Polymers.ppt
THERMAL ANALYSIS
THERMAL ANALYSIS
THERMAL METHODS OF ANALYSIS
THERMAL METHODS OF ANALYSIS
GENERAL THERMODYNAMIC RELATIONSHIPS
Thermal analyses are usually run under conditions of constant pressure, the
underlying thermodynamic equation is the Gibbs-Helmholtz expression:
G0
=H0
-TS0
where G=free energy of the system, H=enthalpy of the system, S=entropy
of the system, T=temperature in kelvins
Consider the following general chemical reaction
aA+bBcC+dD
This reaction is spontaneous as written if G<0, at equilibrium if G=0,
and does not proceed if G >0.
Thermal analysis involves the monitoring of spontaneous reaction
Differentiating the Gibbs-Helmholtz equation with respect to
temperature (assuming S and H not vary with temperature):
S
dT
G
d



 )
(
Show how to move from a stable situation (G>>0) to one
where reaction will occur.
S>0, an increase in temperature cause G<0,
S<0, decreasing the temperature will achieve the desired
spontaneous reaction.
Once the reaction is made to occur, thermal analysis may be used
to detect the process, yielding different and complementary
information.
Thermal Events
Thermal Events
•Thermogravimetry (TG) is a technique in which weight changes
of a specimen as a function of temperature or time is studied.
• Thermogravimetric analysis (TGA) is useful strictly for
transformations involving the absorption or evolution of gases
from a specimen consisting of a condensed phase.
•A plot of mass versus temperature (thermogravimetric curves or
TG curves) permits evaluation of thermal stabilities, rate of
reaction, reaction processes, and sample composition.
Measurements of changes in sample mass with temperature are
made using thermobalance. The balance should be in a suitably
enclosed system so that the atmosphere can be controlled.
2. Thermogravimetry
General considerations
Suitable samples for TG are solids that undergo one of the two
general types of reaction:
Reactant(s)  Product(s)+Gas (a mass loss)
Gas+Reactant(s)  Product(s) (a mass gain)
Processes occuring without change in mass (e.g., the melting of a
sample) obviously cannot be studied by TG.
Thermogravimetric instrumentation should include several basic
components to provide the flexibility necessary for the
production of useful analytical data:
a) A balance,
b) A heating device,
c) A unit for temperature measurement and control,
d) A means for automatically recording the mass and
temperature changes,
e) A system to control the atmosphere around the sample.
2.1 Instrumentation
LINSEIS L81
Two typical designs of the thermobalance are shown in the following:
2.1.1 The Thermobalance
* Null-deflection weighing mechanisms are favoured in TG as
they ensure that the sample remains in the same zone of the
furnace irrespective of changes in mass.
* Sensitivity of balance  1g for a 1g maximum load balance.
* The output weight signal may be differentiated electronically to
give a derivative thermogravimetric curve (DTG)
Balances must remain precise and accurate continuously
under extreme temperature and atmosphere conditions and
should deliver a signal suitable for continuous recording.
The furnace is normally an electrical resistive heater;
Some basic requirements of the heating chamber are :
be non-inductively wound
be capable of reaching 100 to 200°C above the maximum desired working
temperature
have a uniform hot-zone of reasonable length
reach the required starting temperature as quickly as possible
not affect the balance mechanism through radiation or convection
In order to overcome the problem of possible temperature gradient, infrared
or microwave radiation have been used in some equipment.
infrared heating : use halogen lamp, temperature up to 1400°C, heating rate
can be as high as 1000°C/min, accuracy is about ±0.5°C.
Microwave heating : large sample can be used because uniform heating
generated within sample, but temperature measurement and power control are
difficult.
2.1.2 The Heating Chamber
Constant heating rate
Constant heating rate: lag behind of the sample temperature
During heating a temperature
difference between the furnace and the
sample temperature appears which
means that the sample temperature lags
always behind the furnace temperature.
Measurement of the melting point
of Di-tert.-biphenyle at different
heating rates.
Gradual raise of temperature
Thermal equilibrium is better reached by gradual raise of the
temperature.
2.1.3 The atmosphere
Sort, pressure and flow rate of the gas in the sample chamber
influence the following parameters:
•Sample reaction
Sample reactions with the gas (oxidation in the presence of oxygen).
•Heat transitions
Different heat conductivity of the gases used in an experiment.
•Buoyancy and current effects
Different density and flow rate of the gases used in an experiment.
For all thermoanalytical investigations it is very important to
report the sort, the pressure and the flow rate of the gases used in
the experiment.
Thermal decomposition temperatures for CaCO3 in different
gas atmospheres
 Thermbalance are normally housed in
glass or metal system to allow for operation
at pressures ranging from high vacuum (<
10-4
Pa) to high pressure (>3000 kPa) of
inert, oxidizing, reducing or corrosive
gases.
 Care must be taken to correct for
buoyancy arising from the lack of symmetry
in the weighing system
 Thermal convection is responsible for
noise in the signal of TG. The use of dense
carrier gases at high pressures in hot zones
with large temperature gradients give the
most noise. Fitting of convoluted baffles
was found to be most successful in reducing
thermal convection.
Sample form, defect content, porosity and surface properties
has influence on the behaviour on heating, e. g. single crystal
sample give different response from powdered sample
Large sample size cause problems like heat transfer, and gas
exchange with the surrounding is reduced; in general, the use
of small (~ 20 mg) specimen is preferable if sensitivity of
balance permits
Sample should be powdered and spread thinly and uniformly
in the container
2.1.4 The sample
Crucibles
Decomposition temperatures of CaCO3 as function of crucibles
2.1.5 Temperature measurement and calibration
Platinum resistance thermometers or
thermocouples are used for temperature
measurement.
Large difference between sample
temperature (Ts
) and furnace temperature
(Tf
) can exist, sometime as high as 30°C.
Calibration is thus needed.
The difference or lag is more marked
when operating in vacuum or in fast
flowing atmosphere and with high heating
rate.
A series of high purity wires may be
suspend in the region where the
specimen crucible would normally be
located. If the furnace temperature is
slowly raised through the melting point
of a particular wire, a significant
weight loss will be recorded when the
wire melts.
A series of fusible wire, such as :
indium (156.63°C), lead (327.5°C),
zinc (419.58°C), aluminium
(660.37°C), silver (961.93°C), and gold
(1064.42°C) should give a reasonable
calibration curve.
furnace
hanger of sample pan
different metal wires
thermocouple
Temperature calibration for small furnace can be done by making use
of the melting point or Curie points of a range of metals and alloys.
Calibration can also be done by placing a series of
ferromagnetic materials in the specimen basket and a magnet
below or above it, external to the furnace. When each material
goes through its Curie temperature (ferro- to paramagnetic
transition), a sharp ‘weight’ change will be indicated.
2.2 Interpretation of TG and
DTG curves i.The sample undergoes no decomposition with
loss of volatile products over the temperature
range shown but solid phase transformation,
melting ,etc can not be detected by TG,
ii.The rapid initial mass loss is characteristic of
desorption or drying. If it is true, then re-run the
sample should result in type (i) curves,
iii.Single stage decomposition,
iv.Multi-stage decomposition with relatively
stable intermediates : provide information on the
temperature limit of stability of reactants and
intermediate products and also stoichiometry,
v.Multi-stage decomposition with no stable
intermediate product. However heating-rate
effect must be considered. At low heating rate,
type (v) resemble type (iv). At high heating rate,
type (iv) and (v) resemble type (iii) and lose all
the details,
vi.Gain in mass due to reaction with atmosphere,
e.g. oxidation of metals,
vii.Oxidation product decompose again at higher
temperature; this is not often encountered.
Resolution of stages can be improved by
recording DTG or by digital differentiation
of TG data.
2.3 Preparing the measurement
General advices:
•Exact characterization of the starting materials (purity, grain size)!
•Large amount of the starting material for repeated and further
measurements
•Removal of absorbed water by drying (m must be constant)
•Use samples with narrow grain size distribution (Sieving)
•For measurement in vacuum no sample with a grain size below 60
mesh (0.25 mm) (a part of the sample can be lost)
2.4 Applications of TG
TG curve for CuSO4
5H2
O TG curve for CaSO4
2H2
O at
different water-vapour pressure
knowledge of thermal stability can give information on problems like
the hazards of storing explosives, shelf life of drugs, etc.
TG curves can also be used for 'fingerprinting' purpose.
The thermal balance in a TG equipment can also be used to measure
vapour pressure of a sample and magnetic susceptibility, etc.
ATTN:
Three factors should be noted when you get a TG curve:
1. General shape,
2. The particular temperatures at which changes in mass occur
(severely affected by many experimental conditions),
3. The magnitudes of the mass changes (much more use directly
related to the specific stoichiometries of the reactions,
independent of the many factors that affect the shape of the
curves. Can be used for precise quantitative analysis).
Analytical calculations
Under controlled and reproducible conditions, quantitative data
can be extracted from the relevant TG curves. Most commonly,
the mass change is related to sample purity or composition.
Example: A pure compound may be either MgO, MgCO3, or MgC2O4. A
thermogram of the substance shows a loss of 91.0 mg from a total of 175.0 mg
used for analysis. What is the formula of the compound? The relevant possible
reactions are
MgO  No reaction
MgCO3  MgO+CO2
MgC2O4  MgO+CO2+CO
Solution: % Mass loss Sample=(91.0/175.0)(100%)=52.0
% Mass loss if MgCO3=(44/84.3)(100%)=52.2
% Mass loss if MgC2O4=((44+28)/112.3)(100%)=64.1
If the preparation was pure, the compound present is MgCO3.
3. Dynamic Mechanical Analysis
3.1 Viscoelastic Properties of Polymers
 A polymer may exhibit mechanical behaviour characteristic
of either an elastic solid or a viscous liquid, depending upon
temperature, in relation to the glass-transition temperature
(Tg
) of the polymer and the time scale of the deformation.
 Two extremes types of stress-strain curves are those for
elastic solid ( , Hooke’s law) and fluid (  d/dt,
Newton’s law)
relationship between moduli: E = 3B(1-2) = 2(1+)G
E: Young’s modulus; B: bulk modulus; G: shear modulus; : Poisson’s ratio
Chapter 10 Thermal Analysis of Polymers.ppt
Chapter 10 Thermal Analysis of Polymers.ppt
*For polymer, if d/dt = constant, a curve like the following will be observed
 In DMA, the sample is subjected to a sinusoidally varying
stress of angular frequency . The strain is also sinusoidal but
out of phase with the stress by an angle  due to internal
damping effects.
3.2 Periodic stress and DMA
The response of the sample to this treatment can provide information on the
stiffness of the material (quantified by its elastic moduli) and its ability to
dissipate energy (measured by its damping). For a viscoelastic material, the
strain resulting from the periodic stress will also be periodic, but will be out
of phase with the applied stress owing to energy dispersion as heat, or
damping.
If an elastic sample is vibrated over a range of frequencies and the
amplitude of vibration is measured, the resonance frequency is that which
produces a maximum in a plot of amplitude against frequency. Young's
modulus (of elasticity), E, is related to the square of the resonance
frequency, Vr.
2
2
4
/ d
v
cL
E r


where c is a constant, L is the sample length between clamps, d is the sample
thickness and  is the sample density.
3.3 The resonance frequency
If a sample is vibrated over a range of frequencies and the the amplitude of
vibration can be measured, the resonance frequency is that which produces a
maximum in a plot of amplitude against frequency. Modulus is related to the
resonance frequency.
Free oscillation with damping
Damping: log10(A1/A2)=log10(A2/A3)
For elastic materials, the modulus E is simply the constant
ratio between the stress and the resulting strain, but for
viscoelastic materials, the modulus is a complex quantity:
E*
= E' + iE"
where E' is the storage modulus or in-phase component
and E" is the loss modulus or out-of-phase component. The
ratio E" / E' is the tangent of the phase angle, .
 For this forced-vibration situation, complex variables (i.e. ) is used for
analysis The modulus can also be written as G* = G + iG where G is
called the storage modulus and G is called the loss modulus.
 The outputs of the test are usually temperature variation plots of either tan
, G and/or G or some other combinations of these parameters.
DMA response of
polystyrene cross-linked
with 2% divinyl benzene
DMA spectrum of polysulfone.
o : storage_modulus;  : loss-modulus. (Tg
480K)
3.3 Apparatus
 The sample is set in cyclic tensile load, a linear variable differential transformer
(LVDT) is used to monitor the frequency and the amplitude of vibration.
 The preset oscillation amplitude is maintained by a feedback control loop and the
driving force required to do so is a measure of the energy dissipation of the sample
3.4 Applications
 change in E (or G) indicate changes in rigidity and hence
strength of the sample (cure behaviour)
damping measurements give practical information on glass
transitions, change in crystallinity, the occurrence of cross-
linking and also show up the features of polymer chains
 damping information can be useful in studies of vibration
dissipation impact resistance and noise abatement.
 Stress relaxation behaviour of polymer
Typical DMA results on two different samples of polyethylene
(b) branched polyethylene.
(a) Linear polyethylene
The damping curve for linear polyethylene (a) shows peaks at -95°C and 65°C. The lower temperature
peak has been attributed to long chain (-CH2-)n crankshaft relaxations in the amorphous phase and the
higher temperature peak to similar motion in the crystalline phase. The temperatures and relative sizes
of the two peaks can be related to the degree of crystallinity of the sample.
The damping curve for branched polyethylene (b) has features at -112°C, -9°C and 45°C. The -112°C
and 45°C peaks are explained as above, while the -9°C peak is attributed to (-CH3) relaxations in the
amorphous phase.
The thermal behaviour of styrene-butadiene-rubber (SBR)
Various formulations of SBR are used in tyre manufacture. Different styrene-
butadiene ratios may be used, or different butadiene isomers, or different
additives e.g. carbon black. A high cis-butadiene content (a) lowers the glass
transition temperature, Tg, (to as much as -110°C compared to -50°C) giving
greater flexibility at low temperatures. The addition of carbon black (c) increases
the modulus of elasticity. The Tg is also slightly increased. The complex
damping curve at low temperatures indicates polymer-carbon black interactions
and may lead to adverse properties e.g. heat build-up.
4. Differential Scanning Calorimetry
In power-compensated DSC, the sample and a reference material
are maintained at the same temperature throughout the controlled
temperature programme. The difference in the independent
energy supplies to the sample and the reference is then recorded
against the programme temperature
DSC can be used to study heats of reaction, kinetics, heat
capacities, phase transitions, thermal stabilities, sample
composition and purity, critical points, and phase diagrams.
Circuitry of a DSC
Two separate heating circuits:
•The average-heating controller
(the temperatures of the sample
(Ts) and reference (Tr) are
measured and averaged and the
heat output is automatically
adjusted to increase the average
temperature of the sample and
reference in a linear rate)
•Differential-heating circuit
(monitor the difference in Ts and
Tr, and automatically adjust the
power to either the reference or
sample chambers to keep the
temperatures equal)
x-axis: temperature, y-axis: the difference in power supplied to the
two differential heater (calories per unit time).
 Thermal events in the sample appear as deviation from the DSC baseline,
in either an endothermic or exothermic direction (marked on DSC curves). In
DSC, endothermic responses are usually represented as being positive, i.e.
above the base line.
DSC trace of poly(ethylene terephthalate-co-p-oxbenzoate)
Power
difference
4.2 Sample containers and sampling
DSC cell
 T<500°C : usually contained in aluminium sample pans
which can be sealed either by crimping or by cold-welding for
holding volatile samples
 T>500°C : use quartz, alumina (Al2
O3
), gold or graphite
pans
 the reference material in most DSC applications is simply an
empty sample pan
 purging of gas into the DSC sample holder is possible, e.g.
N2
, O2
, etc.
 the mass of (sample+pan+lid) should be recorded before and
after a run so that further information about the processes can
be deduced
The reference sample
For all difference methods (DTA, DSC) reference samples like Al2O3 are
needed to ensure that the heat flow from the furnace to the sample and from
the furnace to the reference sample is identical!
The thermal behavior of the reference sample is included in the measured
signal.
Requirements for the reference sample:
•Known temperature behavior
•No discontinuity in the temperature curve
•If possible a similar thermal behavior as the sample (similar heat capacity)
For small weights of the sample and when no precise measurements are
required an experiment without a reference sample is possible.
In such case an empty crucible can be used as reference.
 aim at correlating the features recorded with the thermal
events taking place in the sample
 after baseline correction, the peak area is proportional to
enthalpy change,
where K is a constant and m is the mass of the sample
K can be obtained by melting a known amount of a pure metal
4.3 Interpretation of DSC curves
H
AK
m

Chapter 10 Thermal Analysis of Polymers.ppt
 reversibility can be monitored by cooling and reheating
Heating and cooling curves for a partially
crystallized polymer.
Chapter 10 Thermal Analysis of Polymers.ppt
4.4 Applications
4.4.1 Measurement of heat capacity
 
H C dT S
C
T
dT
p
p
T
T
T
T
 







 1
2
1
2
G = H - TS
Chapter 10 Thermal Analysis of Polymers.ppt
4.4.2 Measurement of thermal conductivity
 The temperature at the
bottom of the sample (T1
) is
measured via the output of
the DSC, while the
temperature at the top of
sample is measured with a
separate thermocouple in
the contact rod.
 The DSC cell is brought to the desired measurement
temperature, T, and when the output to the recorder is steady
with time, the temperature difference across the sample Ts
and
the displacement of baseline, hs
, is recorded. The same is then
go through for a standard calibrant, e.g. a standard glass.
where Ri
= recorder sensitivity
li
= length of sample/calibrant
di
= diameter of sample/calibrant











s
s
c
c
c
c
c
s
s
s
c
s
T
d
l
R
h
T
d
l
R
h
2
2


4.4.3 Determination of phase diagrams
 melting point can be determined from the DSC curve
 Melting point of pure components are easily determined
 DSC curves for slow cooling of mixture
 If heating is done instead of cooling, the curve should
ideally be endothermic mirror image of that shown and the
problem of supercooling is avoided.
 From a series of this kind of curves, a phase diagram can be
constructed.
 Temperature and enthalpy changes for the thermal events
enthalpy  area of peak after baseline correction
4.4.4 General Applications
Corresponding
TG curve
 Detection of solid-solid phase transition and the measurement
of H for these transitions
DSC curve of carbon tetrachloride
 Tracing the ferromagnetic to paramagnetic transformation.
Most rewarding applications is in study of polymer
•Most solid polymers are formed by rapid cooling to low temperatures
(quenching) are thus in glassy state; by heating above Tg
, glass transition,
with change in cp
but no change in enthalpy, is observed, therefore no peak
is observed, only discontinuity results
•degradation or oxidation of polymers can be study with DSC in isothermal
mode
•for recycling plastics, identification is important and DSC curves provide
'fingerprint'of the materials.
 Introdcution to Thermal Analysis, M.E. Brown -- Chapman
and Hall
 Thermal Analysis - Techniques and Applications, ed. E.L.
Charsley and S.B. Warrington -- Royal Society of Chemistry
 Thermal Analysis of Materials, Robert F. Speyer –Marcel
Dekker, Inc.
References:

More Related Content

PPT
PPT NSUT measurement techniques in research.ppt
PPTX
THERMOGRAVIMETRY ANALYSIS [TGA] AS PER PCI
PPTX
THERMOGRAVIMETRY ANALYSIS [TGA] AS PER PCI[M.PHARM]
PPTX
THERMOGRAVIMETRIC ANALYSIS ppt by devika.pptx
PPTX
Thermogravimetric analysis with curveTGA (1).pptx
PPTX
Thermo Gravimetric analysis Principles and Design.pptx
PPTX
TGA TECHNIQUES.pptxhhhjjkkkhhhbbbbbbbnjjjn
PPT NSUT measurement techniques in research.ppt
THERMOGRAVIMETRY ANALYSIS [TGA] AS PER PCI
THERMOGRAVIMETRY ANALYSIS [TGA] AS PER PCI[M.PHARM]
THERMOGRAVIMETRIC ANALYSIS ppt by devika.pptx
Thermogravimetric analysis with curveTGA (1).pptx
Thermo Gravimetric analysis Principles and Design.pptx
TGA TECHNIQUES.pptxhhhjjkkkhhhbbbbbbbnjjjn

Similar to Chapter 10 Thermal Analysis of Polymers.ppt (20)

PPTX
M PHARMACY MPAT NOTES ON THERMOGRAVIMETRIC ANAYSIS
PPTX
TGA.pptx principle, instrumentation, theory
PPTX
Thermogravimetric analysis
PPTX
Thermo analytical methods by srk
PPTX
Thermogravimetric analysis methods .....
PPTX
A presentation on Thermal gravimetry analysis (TGA)
PPTX
Instrumentation
PDF
Thermogravimetric analysis (TGA) By Thermogravimetric analysis(TGA) by Vikr...
PPTX
TGA. Thermogravimetric Analysis topic of instrumentation pptx
PPTX
Thermal analysis - TGA & DTA
PPTX
PPTX
thermogravimetric analysis
PPTX
Thermogravimetric analysis ppt
PPTX
TGA.pptx
PPTX
Thermo gravimetric analysis
PPTX
Thermo gravimetric analysis(tga)
PPTX
PPTX
Thermo gravimetric Analysis
PPTX
Exploring Thermal Gravimetric Analysis: Applications, Techniques, and Insights
M PHARMACY MPAT NOTES ON THERMOGRAVIMETRIC ANAYSIS
TGA.pptx principle, instrumentation, theory
Thermogravimetric analysis
Thermo analytical methods by srk
Thermogravimetric analysis methods .....
A presentation on Thermal gravimetry analysis (TGA)
Instrumentation
Thermogravimetric analysis (TGA) By Thermogravimetric analysis(TGA) by Vikr...
TGA. Thermogravimetric Analysis topic of instrumentation pptx
Thermal analysis - TGA & DTA
thermogravimetric analysis
Thermogravimetric analysis ppt
TGA.pptx
Thermo gravimetric analysis
Thermo gravimetric analysis(tga)
Thermo gravimetric Analysis
Exploring Thermal Gravimetric Analysis: Applications, Techniques, and Insights
Ad

Recently uploaded (20)

PDF
medical_surgical_nursing_10th_edition_ignatavicius_TEST_BANK_pdf.pdf
PDF
GENETICS IN BIOLOGY IN SECONDARY LEVEL FORM 3
DOC
Soft-furnishing-By-Architect-A.F.M.Mohiuddin-Akhand.doc
PPTX
Final Presentation General Medicine 03-08-2024.pptx
PDF
Practical Manual AGRO-233 Principles and Practices of Natural Farming
PPTX
1st Inaugural Professorial Lecture held on 19th February 2020 (Governance and...
PPTX
Final Presentation General Medicine 03-08-2024.pptx
PPTX
Radiologic_Anatomy_of_the_Brachial_plexus [final].pptx
PPTX
Onco Emergencies - Spinal cord compression Superior vena cava syndrome Febr...
PDF
What if we spent less time fighting change, and more time building what’s rig...
PDF
احياء السادس العلمي - الفصل الثالث (التكاثر) منهج متميزين/كلية بغداد/موهوبين
PDF
SOIL: Factor, Horizon, Process, Classification, Degradation, Conservation
PDF
Hazard Identification & Risk Assessment .pdf
PDF
Complications of Minimal Access Surgery at WLH
PDF
Trump Administration's workforce development strategy
PPTX
Chinmaya Tiranga Azadi Quiz (Class 7-8 )
PDF
Chinmaya Tiranga quiz Grand Finale.pdf
PDF
LNK 2025 (2).pdf MWEHEHEHEHEHEHEHEHEHEHE
PDF
ChatGPT for Dummies - Pam Baker Ccesa007.pdf
PPTX
Introduction-to-Literarature-and-Literary-Studies-week-Prelim-coverage.pptx
medical_surgical_nursing_10th_edition_ignatavicius_TEST_BANK_pdf.pdf
GENETICS IN BIOLOGY IN SECONDARY LEVEL FORM 3
Soft-furnishing-By-Architect-A.F.M.Mohiuddin-Akhand.doc
Final Presentation General Medicine 03-08-2024.pptx
Practical Manual AGRO-233 Principles and Practices of Natural Farming
1st Inaugural Professorial Lecture held on 19th February 2020 (Governance and...
Final Presentation General Medicine 03-08-2024.pptx
Radiologic_Anatomy_of_the_Brachial_plexus [final].pptx
Onco Emergencies - Spinal cord compression Superior vena cava syndrome Febr...
What if we spent less time fighting change, and more time building what’s rig...
احياء السادس العلمي - الفصل الثالث (التكاثر) منهج متميزين/كلية بغداد/موهوبين
SOIL: Factor, Horizon, Process, Classification, Degradation, Conservation
Hazard Identification & Risk Assessment .pdf
Complications of Minimal Access Surgery at WLH
Trump Administration's workforce development strategy
Chinmaya Tiranga Azadi Quiz (Class 7-8 )
Chinmaya Tiranga quiz Grand Finale.pdf
LNK 2025 (2).pdf MWEHEHEHEHEHEHEHEHEHEHE
ChatGPT for Dummies - Pam Baker Ccesa007.pdf
Introduction-to-Literarature-and-Literary-Studies-week-Prelim-coverage.pptx
Ad

Chapter 10 Thermal Analysis of Polymers.ppt

  • 1. Thermal Characterization of Polymers/Polymeric Materials By Dr Shahzad Ali Shahid Chatha Professor of Analytical Chemistry
  • 8. GENERAL THERMODYNAMIC RELATIONSHIPS Thermal analyses are usually run under conditions of constant pressure, the underlying thermodynamic equation is the Gibbs-Helmholtz expression: G0 =H0 -TS0 where G=free energy of the system, H=enthalpy of the system, S=entropy of the system, T=temperature in kelvins Consider the following general chemical reaction aA+bBcC+dD This reaction is spontaneous as written if G<0, at equilibrium if G=0, and does not proceed if G >0. Thermal analysis involves the monitoring of spontaneous reaction
  • 9. Differentiating the Gibbs-Helmholtz equation with respect to temperature (assuming S and H not vary with temperature): S dT G d     ) ( Show how to move from a stable situation (G>>0) to one where reaction will occur. S>0, an increase in temperature cause G<0, S<0, decreasing the temperature will achieve the desired spontaneous reaction. Once the reaction is made to occur, thermal analysis may be used to detect the process, yielding different and complementary information.
  • 12. •Thermogravimetry (TG) is a technique in which weight changes of a specimen as a function of temperature or time is studied. • Thermogravimetric analysis (TGA) is useful strictly for transformations involving the absorption or evolution of gases from a specimen consisting of a condensed phase. •A plot of mass versus temperature (thermogravimetric curves or TG curves) permits evaluation of thermal stabilities, rate of reaction, reaction processes, and sample composition. Measurements of changes in sample mass with temperature are made using thermobalance. The balance should be in a suitably enclosed system so that the atmosphere can be controlled. 2. Thermogravimetry
  • 13. General considerations Suitable samples for TG are solids that undergo one of the two general types of reaction: Reactant(s)  Product(s)+Gas (a mass loss) Gas+Reactant(s)  Product(s) (a mass gain) Processes occuring without change in mass (e.g., the melting of a sample) obviously cannot be studied by TG.
  • 14. Thermogravimetric instrumentation should include several basic components to provide the flexibility necessary for the production of useful analytical data: a) A balance, b) A heating device, c) A unit for temperature measurement and control, d) A means for automatically recording the mass and temperature changes, e) A system to control the atmosphere around the sample.
  • 16. Two typical designs of the thermobalance are shown in the following: 2.1.1 The Thermobalance
  • 17. * Null-deflection weighing mechanisms are favoured in TG as they ensure that the sample remains in the same zone of the furnace irrespective of changes in mass. * Sensitivity of balance  1g for a 1g maximum load balance. * The output weight signal may be differentiated electronically to give a derivative thermogravimetric curve (DTG) Balances must remain precise and accurate continuously under extreme temperature and atmosphere conditions and should deliver a signal suitable for continuous recording.
  • 18. The furnace is normally an electrical resistive heater; Some basic requirements of the heating chamber are : be non-inductively wound be capable of reaching 100 to 200°C above the maximum desired working temperature have a uniform hot-zone of reasonable length reach the required starting temperature as quickly as possible not affect the balance mechanism through radiation or convection In order to overcome the problem of possible temperature gradient, infrared or microwave radiation have been used in some equipment. infrared heating : use halogen lamp, temperature up to 1400°C, heating rate can be as high as 1000°C/min, accuracy is about ±0.5°C. Microwave heating : large sample can be used because uniform heating generated within sample, but temperature measurement and power control are difficult. 2.1.2 The Heating Chamber
  • 20. Constant heating rate: lag behind of the sample temperature During heating a temperature difference between the furnace and the sample temperature appears which means that the sample temperature lags always behind the furnace temperature. Measurement of the melting point of Di-tert.-biphenyle at different heating rates.
  • 21. Gradual raise of temperature Thermal equilibrium is better reached by gradual raise of the temperature.
  • 22. 2.1.3 The atmosphere Sort, pressure and flow rate of the gas in the sample chamber influence the following parameters: •Sample reaction Sample reactions with the gas (oxidation in the presence of oxygen). •Heat transitions Different heat conductivity of the gases used in an experiment. •Buoyancy and current effects Different density and flow rate of the gases used in an experiment. For all thermoanalytical investigations it is very important to report the sort, the pressure and the flow rate of the gases used in the experiment.
  • 23. Thermal decomposition temperatures for CaCO3 in different gas atmospheres
  • 24.  Thermbalance are normally housed in glass or metal system to allow for operation at pressures ranging from high vacuum (< 10-4 Pa) to high pressure (>3000 kPa) of inert, oxidizing, reducing or corrosive gases.  Care must be taken to correct for buoyancy arising from the lack of symmetry in the weighing system  Thermal convection is responsible for noise in the signal of TG. The use of dense carrier gases at high pressures in hot zones with large temperature gradients give the most noise. Fitting of convoluted baffles was found to be most successful in reducing thermal convection.
  • 25. Sample form, defect content, porosity and surface properties has influence on the behaviour on heating, e. g. single crystal sample give different response from powdered sample Large sample size cause problems like heat transfer, and gas exchange with the surrounding is reduced; in general, the use of small (~ 20 mg) specimen is preferable if sensitivity of balance permits Sample should be powdered and spread thinly and uniformly in the container 2.1.4 The sample
  • 27. Decomposition temperatures of CaCO3 as function of crucibles
  • 28. 2.1.5 Temperature measurement and calibration Platinum resistance thermometers or thermocouples are used for temperature measurement. Large difference between sample temperature (Ts ) and furnace temperature (Tf ) can exist, sometime as high as 30°C. Calibration is thus needed. The difference or lag is more marked when operating in vacuum or in fast flowing atmosphere and with high heating rate.
  • 29. A series of high purity wires may be suspend in the region where the specimen crucible would normally be located. If the furnace temperature is slowly raised through the melting point of a particular wire, a significant weight loss will be recorded when the wire melts. A series of fusible wire, such as : indium (156.63°C), lead (327.5°C), zinc (419.58°C), aluminium (660.37°C), silver (961.93°C), and gold (1064.42°C) should give a reasonable calibration curve. furnace hanger of sample pan different metal wires thermocouple Temperature calibration for small furnace can be done by making use of the melting point or Curie points of a range of metals and alloys.
  • 30. Calibration can also be done by placing a series of ferromagnetic materials in the specimen basket and a magnet below or above it, external to the furnace. When each material goes through its Curie temperature (ferro- to paramagnetic transition), a sharp ‘weight’ change will be indicated.
  • 31. 2.2 Interpretation of TG and DTG curves i.The sample undergoes no decomposition with loss of volatile products over the temperature range shown but solid phase transformation, melting ,etc can not be detected by TG, ii.The rapid initial mass loss is characteristic of desorption or drying. If it is true, then re-run the sample should result in type (i) curves, iii.Single stage decomposition, iv.Multi-stage decomposition with relatively stable intermediates : provide information on the temperature limit of stability of reactants and intermediate products and also stoichiometry, v.Multi-stage decomposition with no stable intermediate product. However heating-rate effect must be considered. At low heating rate, type (v) resemble type (iv). At high heating rate, type (iv) and (v) resemble type (iii) and lose all the details, vi.Gain in mass due to reaction with atmosphere, e.g. oxidation of metals, vii.Oxidation product decompose again at higher temperature; this is not often encountered. Resolution of stages can be improved by recording DTG or by digital differentiation of TG data.
  • 32. 2.3 Preparing the measurement General advices: •Exact characterization of the starting materials (purity, grain size)! •Large amount of the starting material for repeated and further measurements •Removal of absorbed water by drying (m must be constant) •Use samples with narrow grain size distribution (Sieving) •For measurement in vacuum no sample with a grain size below 60 mesh (0.25 mm) (a part of the sample can be lost)
  • 33. 2.4 Applications of TG TG curve for CuSO4 5H2 O TG curve for CaSO4 2H2 O at different water-vapour pressure
  • 34. knowledge of thermal stability can give information on problems like the hazards of storing explosives, shelf life of drugs, etc. TG curves can also be used for 'fingerprinting' purpose. The thermal balance in a TG equipment can also be used to measure vapour pressure of a sample and magnetic susceptibility, etc. ATTN: Three factors should be noted when you get a TG curve: 1. General shape, 2. The particular temperatures at which changes in mass occur (severely affected by many experimental conditions), 3. The magnitudes of the mass changes (much more use directly related to the specific stoichiometries of the reactions, independent of the many factors that affect the shape of the curves. Can be used for precise quantitative analysis).
  • 35. Analytical calculations Under controlled and reproducible conditions, quantitative data can be extracted from the relevant TG curves. Most commonly, the mass change is related to sample purity or composition. Example: A pure compound may be either MgO, MgCO3, or MgC2O4. A thermogram of the substance shows a loss of 91.0 mg from a total of 175.0 mg used for analysis. What is the formula of the compound? The relevant possible reactions are MgO  No reaction MgCO3  MgO+CO2 MgC2O4  MgO+CO2+CO Solution: % Mass loss Sample=(91.0/175.0)(100%)=52.0 % Mass loss if MgCO3=(44/84.3)(100%)=52.2 % Mass loss if MgC2O4=((44+28)/112.3)(100%)=64.1 If the preparation was pure, the compound present is MgCO3.
  • 36. 3. Dynamic Mechanical Analysis 3.1 Viscoelastic Properties of Polymers  A polymer may exhibit mechanical behaviour characteristic of either an elastic solid or a viscous liquid, depending upon temperature, in relation to the glass-transition temperature (Tg ) of the polymer and the time scale of the deformation.  Two extremes types of stress-strain curves are those for elastic solid ( , Hooke’s law) and fluid (  d/dt, Newton’s law) relationship between moduli: E = 3B(1-2) = 2(1+)G E: Young’s modulus; B: bulk modulus; G: shear modulus; : Poisson’s ratio
  • 39. *For polymer, if d/dt = constant, a curve like the following will be observed
  • 40.  In DMA, the sample is subjected to a sinusoidally varying stress of angular frequency . The strain is also sinusoidal but out of phase with the stress by an angle  due to internal damping effects. 3.2 Periodic stress and DMA
  • 41. The response of the sample to this treatment can provide information on the stiffness of the material (quantified by its elastic moduli) and its ability to dissipate energy (measured by its damping). For a viscoelastic material, the strain resulting from the periodic stress will also be periodic, but will be out of phase with the applied stress owing to energy dispersion as heat, or damping. If an elastic sample is vibrated over a range of frequencies and the amplitude of vibration is measured, the resonance frequency is that which produces a maximum in a plot of amplitude against frequency. Young's modulus (of elasticity), E, is related to the square of the resonance frequency, Vr. 2 2 4 / d v cL E r   where c is a constant, L is the sample length between clamps, d is the sample thickness and  is the sample density.
  • 42. 3.3 The resonance frequency If a sample is vibrated over a range of frequencies and the the amplitude of vibration can be measured, the resonance frequency is that which produces a maximum in a plot of amplitude against frequency. Modulus is related to the resonance frequency. Free oscillation with damping Damping: log10(A1/A2)=log10(A2/A3)
  • 43. For elastic materials, the modulus E is simply the constant ratio between the stress and the resulting strain, but for viscoelastic materials, the modulus is a complex quantity: E* = E' + iE" where E' is the storage modulus or in-phase component and E" is the loss modulus or out-of-phase component. The ratio E" / E' is the tangent of the phase angle, .
  • 44.  For this forced-vibration situation, complex variables (i.e. ) is used for analysis The modulus can also be written as G* = G + iG where G is called the storage modulus and G is called the loss modulus.
  • 45.  The outputs of the test are usually temperature variation plots of either tan , G and/or G or some other combinations of these parameters. DMA response of polystyrene cross-linked with 2% divinyl benzene DMA spectrum of polysulfone. o : storage_modulus;  : loss-modulus. (Tg 480K)
  • 46. 3.3 Apparatus  The sample is set in cyclic tensile load, a linear variable differential transformer (LVDT) is used to monitor the frequency and the amplitude of vibration.  The preset oscillation amplitude is maintained by a feedback control loop and the driving force required to do so is a measure of the energy dissipation of the sample
  • 47. 3.4 Applications  change in E (or G) indicate changes in rigidity and hence strength of the sample (cure behaviour) damping measurements give practical information on glass transitions, change in crystallinity, the occurrence of cross- linking and also show up the features of polymer chains  damping information can be useful in studies of vibration dissipation impact resistance and noise abatement.  Stress relaxation behaviour of polymer
  • 48. Typical DMA results on two different samples of polyethylene (b) branched polyethylene. (a) Linear polyethylene The damping curve for linear polyethylene (a) shows peaks at -95°C and 65°C. The lower temperature peak has been attributed to long chain (-CH2-)n crankshaft relaxations in the amorphous phase and the higher temperature peak to similar motion in the crystalline phase. The temperatures and relative sizes of the two peaks can be related to the degree of crystallinity of the sample. The damping curve for branched polyethylene (b) has features at -112°C, -9°C and 45°C. The -112°C and 45°C peaks are explained as above, while the -9°C peak is attributed to (-CH3) relaxations in the amorphous phase.
  • 49. The thermal behaviour of styrene-butadiene-rubber (SBR) Various formulations of SBR are used in tyre manufacture. Different styrene- butadiene ratios may be used, or different butadiene isomers, or different additives e.g. carbon black. A high cis-butadiene content (a) lowers the glass transition temperature, Tg, (to as much as -110°C compared to -50°C) giving greater flexibility at low temperatures. The addition of carbon black (c) increases the modulus of elasticity. The Tg is also slightly increased. The complex damping curve at low temperatures indicates polymer-carbon black interactions and may lead to adverse properties e.g. heat build-up.
  • 50. 4. Differential Scanning Calorimetry In power-compensated DSC, the sample and a reference material are maintained at the same temperature throughout the controlled temperature programme. The difference in the independent energy supplies to the sample and the reference is then recorded against the programme temperature DSC can be used to study heats of reaction, kinetics, heat capacities, phase transitions, thermal stabilities, sample composition and purity, critical points, and phase diagrams.
  • 51. Circuitry of a DSC Two separate heating circuits: •The average-heating controller (the temperatures of the sample (Ts) and reference (Tr) are measured and averaged and the heat output is automatically adjusted to increase the average temperature of the sample and reference in a linear rate) •Differential-heating circuit (monitor the difference in Ts and Tr, and automatically adjust the power to either the reference or sample chambers to keep the temperatures equal) x-axis: temperature, y-axis: the difference in power supplied to the two differential heater (calories per unit time).
  • 52.  Thermal events in the sample appear as deviation from the DSC baseline, in either an endothermic or exothermic direction (marked on DSC curves). In DSC, endothermic responses are usually represented as being positive, i.e. above the base line. DSC trace of poly(ethylene terephthalate-co-p-oxbenzoate) Power difference
  • 53. 4.2 Sample containers and sampling DSC cell
  • 54.  T<500°C : usually contained in aluminium sample pans which can be sealed either by crimping or by cold-welding for holding volatile samples  T>500°C : use quartz, alumina (Al2 O3 ), gold or graphite pans  the reference material in most DSC applications is simply an empty sample pan  purging of gas into the DSC sample holder is possible, e.g. N2 , O2 , etc.  the mass of (sample+pan+lid) should be recorded before and after a run so that further information about the processes can be deduced
  • 55. The reference sample For all difference methods (DTA, DSC) reference samples like Al2O3 are needed to ensure that the heat flow from the furnace to the sample and from the furnace to the reference sample is identical! The thermal behavior of the reference sample is included in the measured signal. Requirements for the reference sample: •Known temperature behavior •No discontinuity in the temperature curve •If possible a similar thermal behavior as the sample (similar heat capacity) For small weights of the sample and when no precise measurements are required an experiment without a reference sample is possible. In such case an empty crucible can be used as reference.
  • 56.  aim at correlating the features recorded with the thermal events taking place in the sample  after baseline correction, the peak area is proportional to enthalpy change, where K is a constant and m is the mass of the sample K can be obtained by melting a known amount of a pure metal 4.3 Interpretation of DSC curves H AK m 
  • 58.  reversibility can be monitored by cooling and reheating Heating and cooling curves for a partially crystallized polymer.
  • 60. 4.4 Applications 4.4.1 Measurement of heat capacity   H C dT S C T dT p p T T T T           1 2 1 2 G = H - TS
  • 62. 4.4.2 Measurement of thermal conductivity  The temperature at the bottom of the sample (T1 ) is measured via the output of the DSC, while the temperature at the top of sample is measured with a separate thermocouple in the contact rod.
  • 63.  The DSC cell is brought to the desired measurement temperature, T, and when the output to the recorder is steady with time, the temperature difference across the sample Ts and the displacement of baseline, hs , is recorded. The same is then go through for a standard calibrant, e.g. a standard glass. where Ri = recorder sensitivity li = length of sample/calibrant di = diameter of sample/calibrant            s s c c c c c s s s c s T d l R h T d l R h 2 2  
  • 64. 4.4.3 Determination of phase diagrams  melting point can be determined from the DSC curve
  • 65.  Melting point of pure components are easily determined  DSC curves for slow cooling of mixture
  • 66.  If heating is done instead of cooling, the curve should ideally be endothermic mirror image of that shown and the problem of supercooling is avoided.  From a series of this kind of curves, a phase diagram can be constructed.
  • 67.  Temperature and enthalpy changes for the thermal events enthalpy  area of peak after baseline correction 4.4.4 General Applications Corresponding TG curve
  • 68.  Detection of solid-solid phase transition and the measurement of H for these transitions DSC curve of carbon tetrachloride
  • 69.  Tracing the ferromagnetic to paramagnetic transformation. Most rewarding applications is in study of polymer •Most solid polymers are formed by rapid cooling to low temperatures (quenching) are thus in glassy state; by heating above Tg , glass transition, with change in cp but no change in enthalpy, is observed, therefore no peak is observed, only discontinuity results
  • 70. •degradation or oxidation of polymers can be study with DSC in isothermal mode •for recycling plastics, identification is important and DSC curves provide 'fingerprint'of the materials.
  • 71.  Introdcution to Thermal Analysis, M.E. Brown -- Chapman and Hall  Thermal Analysis - Techniques and Applications, ed. E.L. Charsley and S.B. Warrington -- Royal Society of Chemistry  Thermal Analysis of Materials, Robert F. Speyer –Marcel Dekker, Inc. References: