1- In atomic spectroscopy, samples are vaporized at 2 000–8 000 K
and decompose into atoms.
2- Concentrations of atoms in the vapor are measured by emission or
absorption of characteristic wavelengths of radiation.
3- Atomic spectroscopy is a principal tool of analytical chemistry
because of its high sensitivity, its ability to distinguish one element
from another in a complex sample, its ability to perform simultaneous
multi-element analyses, and the ease with which many samples can be
automatically analyzed.
4- The precision of atomic spectroscopy, typically 1–2%, is not as
good as that of some wet chemical methods.
5- The equipment is expensive, but widely available.
1- An important difference between atomic and molecular
spectroscopy is the width of absorption or emission bands. Spectra of
liquids and solids typically have bandwidths of 100 nm.
2- Spectra of gaseous atoms consist of sharp lines with widths of
0.001 nm.
3- Lines are so sharp that there is usually little overlap between the
spectra of different elements in the same sample. Therefore, some
instruments can measure more than 70 elements simultaneously.
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
4- In atomic fluorescence experiment. Atoms in the flame are
irradiated by a laser to promote them to an excited electronic state
from which they can fluoresce to return to the ground state.
5- Atomic fluorescence is potentially a thousand times more sensitive
than atomic absorption, but equipment for atomic fluorescence is not
common.
6- In atomic emission, collisions in the very hot plasma promote some
atoms to excited electronic states from which they can emit photons to
return to lower-energy states.
7- In atomic emission, no lamp is required.
8- Emission intensity is proportional to the concentration of the
element in the sample. Emission from atoms in a plasma is now the
dominant form of atomic spectroscopy.
Flame Atomization: In a flame atomizer, a solution of
the sample is nebulized by a flow of gaseous oxidant,
mixed with a gaseous fuel, and carried into a flame
where atomization occurs. The following processes then
occur in the flame.
• Desolvation (produce a solid molecular aerosol)
• Dissociation (leads to an atomic gas)
• Ionization (to give cations and electrons)
• Excitation (giving atomic, ionic, and molecular
emission)
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
Types of Flames:
Several common fuels and oxidants can be
employed in flame spectroscopy depending on
temperature needed. Temperatures of 1700o
C to
2400o
C are obtained with the various fuels
when air serves as the oxidant. At these
temperature, only easily decomposed samples
are atomized. For more refractory samples,
oxygen or nitrous oxide must be employed as
the oxidant. With the common fuels these
oxidants produce temperatures of 2500o
C to
3100o
C.
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
Burning Velocity:
The burning velocities are of considerable
importance because flames are stable in certain
ranges of gas flow rates only. If the gas flow rate
does not exceed the burning velocity, the flame
propagates itself back in to the burner, giving
flashback. As the flow rate increases, the flame
rises until it reaches a point above the burner
where the flow velocity and the burning velocity
are equal. This region is where the flame is stable.
At higher flow rates, the flame rises and
eventually reaches a point where it blows off of
the burner.
• Flame Structure:
Important regions of a flame include:
1. primary combustion zone
2. interzonal region
3. secondary combustion zone
1. Primary combustion zone: Thermal
equilibrium is ordinarily not reached in
this region, and it is, therefore, seldom
used for flame spectroscopy.
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
2. Interzonal region: This area is relatively
narrow in stoichiometric hydrocarbon
flames, is often rich in free atoms and is
the most widely used part of the flame for
spectroscopy.
3. Secondary combustion zone: In the
secondary reaction zone, the products of
the inner core are converted to stable
molecular oxides that are then dispersed
into the surroundings.
Temperature Profiles:
A temperature profile of a typical flame for
atomic spectroscopy is shown in Fig. 9-3.
The maximum temperature is located in the
flame about 1 cm above the primary
combustion zone. It is important–
particularly for emission methods – to focus
the same part of the flame on the entrance
slit for all calibrations and analytical
measurements.
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
Flame absorbance Profiles:
Fig. 9-4 shows typical absorption profiles for
three elements. Magnesium exhibits a
maximum in absorbance at the middle of the
flame. The behavior of silver, which is not
readily oxidized, is quite different, a
continuous increase in the number of atoms,
and thus the absorbance, is observed from the
base to the periphery of the flame. Chromium,
which forms very stable oxides, shows a
continuous decrease in absorbance beginning
close to the burner tip.
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
Flame Atomizers:
Figure 9-5 is a diagram of a typical
commercial laminar flow burner that
employs a concentric tube nebulizer. The
aerosol is mixed with fuel. The aerosol,
oxidant, and fuel are then burned in a
slotted burner that provides a flame that is
usually 5 or 10 cm in length.
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
Advantages:
1. Uniform dropsize
2. Homogeneous flame
3. Quiet flame and a long path length
Disadvantages:
1. Flash back if Vburning > Vflow
2. ~90% of sample is lost
3. Large mixing volume
Performance Characteristics
Of Flame Atomizers
1- In terms of reproducible behavior, flame
atomization appears to be superior to all other
methods for liquid sample introduction.
2- In terms of sampling efficiency and thus
sensitivity, however, other atomization
methods are markedly better.
3-A large portion of the sample flows down the
drain and the residence time of individual
atoms in the optical path in the flame is brief
(~10-4
s).
Electrothermal Atomization
1- It provides enhanced sensitivity because the entire
sample is atomized in a short period, and the
average residence time of the atoms in the optical
path is a second or more.
2- A few microliters of sample are first evaporated at
a low temperature and then ashed at a somewhat
higher temperature in an electrically heated graphite
tube or in a graphite cup. Then the current is rapidly
increased to several hundred amperes, which caused
the temperature to soar to perhaps 2000o
C to
3000o
C; atomization of the sample occurs in a
period of a few milliseconds to seconds.
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
Performance Characteristics:
1-Electrothermal atomizers offer the advantage
of unusually high sensitivity for small
volumes of sample.
2- Typically, sample volumes between 0.5 and
10 L are used;
3- absolute detection limits lie in the range of
10-10
to 10-13
g of analyte.
4- A final disadvantage is that the analytical
range is low, being usually less than two
orders of magnitude.
ATOMIC ABSORPTION
INSTRUMENTATION
Instruments for atomic absorption
spectrometry (AAS) consist of a radiation
source, a sample holder, a wavelength
selector, a detector, and a signal
processor and readout. The sample holder
in atomic absorption instruments is the
atomizer cell that contains the gaseous
atomized sample.
1. Radiation Sources:
It is necessary that band width of the
radiation source must be narrow relative to
the width of an absorption peak. The
problem created by limited width of atomic
absorption peaks has been solved by the
use of line sources with bandwidths even
narrower than absorption peaks.
Fig. 9-10a shows the emission spectrum of a typical
atomic lamp source. With a suitable filter or
monochromator, all but one of these lines are
removed. Fig. 9-10b shows the absorption spectrum
for the analyte between wavelengths 1 and 2.
Passage of the line from the source through the
flame reduces its intensity from P0 to P; the
absorbance is then given by log(Po/P), which is
linearly related to the concentration of the analyte
in the sample. A disadvantage of the procedure is
that separate lamp source is needed for each
element.
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
Hollow Cathode Lamps:
It is the most common source for atomic
absorption measurements. This lamp consists
of a tungsten anode and a cylindrical cathode
sealed in a glass tube that is filled with neon or
argon at a pressure of 1 to 5 torr. The cathode
is constructed of the metal whose spectrum is
desired. Ionization of the inert gas occurs
when a potential on the order of 300 V is
applied across the electrodes, which generates
a current of about 5 to 15 mA.
… Hollow Cathode Lamps continued…
If the potential is sufficiently large, the
gaseous cation acquire enough kinetic energy
to dislodge some of the metal atoms from the
cathode surface and produce an atomic cloud
in a process called sputtering. A portion of the
sputtered metal atom are in excited states and
thus emit their characteristic radiation as they
return to the ground state. Eventually, the
metal atoms diffuse back to the cathode
surface or to the glass walls of the tube and are
redeposited.
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
Electrodeless Discharge Lamps (EDLs):
These provide radiant intensities that are
usually one to two orders of magnitude
greater than hollow cathode lamps. A
typical lamp is constructed from a sealed
quartz tube containing a few torr of an inert
gas such as argon and a small quantity of
the metal (or its salt) whose spectrum is of
interest.
…Electrodeless Discharge Lamps (EDLs) continued…
The lamp is energized by an intense field of
radio-frequency or microwave radiation.
Ionization of the argon occurs to give ions
that are accelerated by the high-frequency
component of the field until they gain
sufficient energy to excite the atoms of the
metal whose spectrum is sought.
Electrodeless discharge lamps are available
commercially for 15 or more elements.
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
Single-Beam Instruments: A typical
single-beam instrument, consists of several
hollow cathode sources, an atomizer, and
simple grating spectrophotometer with a
photomultiplier transducer. The 100% T
adjustment is then made while a blank is
aspirated into the flame. Finally, the
transmittance is obtained with the sample
replacing the blank.
Double-Beam Instruments: In double-beam
instrument the beam from the hollow cathode
source is split by a mirrored chopper, one half
passing through the flame and the other half
around it. The two beams are then recombined
by a half-silvered mirror and passed into a
grating monochromator; a photomultiplier
tube serves as the transducer. The ratio
between the reference and sample signal is
then amplified and fed to the readout, which
may be a digital meter or a signal recorder.
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
40
2. Double beam (most commercial instruments)
– Light is split and directed towards both
reference cell (blank) and sample cell
– Two detectors; electronics measure ratio (i.e.,
measure/calculate absorbance)
– Advantages:
• Compensates for fluctuations in source
intensity and drift in detector
• Better design for continuous recording of
spectra
41
Merits of Double Beam Instruments
1.Compensate for all but the most short term
fluctuation in radiant output of the source
2.Compensate drift in transducer and amplifier
3.Compensate for wide variations in source
intensity with wavelength
INTERFERENCES IN ATOMIC
ABSORPTION SPECTROSCOPY
1. Spectral Interferences:
(I) Spectral interference can occur due to overlapping
lines. e.g. a vanadium line at 3082.11Å interferes in an
analysis based upon the aluminum absorption line at
3082.15 Å. This type of interference can be avoid by
employing the aluminum line at 3092.7 Å instead.
(II) Spectral interferences result from the presence of
combustion products that exhibit broadband absorption or
particulate products that scatter radiation. Both diminish
the power of the transmitted beam. A blank can be
aspirated into the flame to make the correction.
…Spectral Interferences continued…
(III) Source of absorption or scattering can be
originated in the sample matrix. An example of
a potential matrix interference due to
absorption occurs in the determination of
barium in alkaline earth mixture. The
wavelength of Ba line used for atomic
absorption analysis appears in the center of a
broad absorption band for CaOH. The effect
can be eliminated by substituting nitrous oxide
for air as the oxidant which yields a higher
temperature that decomposed the CaOH and
eliminates the absorption band.
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
…Spectral Interferences continued…
(IV) Concentrated solution of elements such
as Ti, Zr and W which form refractory
oxides can cause spectral interference due to
scattering.
(V) Organic solvent or organic impurities in
the sample can cause scattering interference
from carbonaceous particle because of
incomplete combustion of the organic
matrix.
2. Chemical Interferences:
(I) Formation of Compounds of Low Volatility:
The most common type of interference is by
anions that form compounds of low volatility with
the analyte and thus reduce the rate at which the
analyte is atomized. The decrease in calcium
absorbance that is observed with increasing
concentrations of sulfate or phosphate. Example of
cation interference have also been recognized.
Aluminum is found to cause low results in the
determination of magnesium, apparently as a
result of the formation of a heat-stable
aluminum/magnesium compound.
…Formation of Compounds of Low Volatility continued…
Interference due to formation of species of low
volatility can often be eliminated or moderated by
use of higher temperatures. Releasing agents
which are cations that react preferentially with the
interferant and prevent its interaction with the
analyte, can be employed. Protective agents
prevent interference by forming stable but volatile
species with the analyte. Three common reagents
for this purpose are EDTA, 8-hydroxyquinoline,
and APDC (ammonium salt of 1-
pyrrolidinecarbodithioic acid).
…Chemical Interferences continued…
(II) Dissociation Equilibria: Gaseous environment
of a flame or a furnace, numerous dissociation and
association reactions lead to conversion of the
metallic constituents to the elemental state. Some of
these reactions are reversible
MO M + O
M(OH)2 M + 2OH
Where M is the analyte atom.
VOx V + Ox
AlOx Al + Ox
TiOx Ti + Ox
…Chemical Interferences continued…
(III) Ionization Equilibria: Ionization of atoms and
molecules is small in combustion mixtures that involve
air as the oxidant, and generally can be neglected. In
higher temperatures of flames where oxygen or nitrous
oxide serves as the oxidant, however, ionization
becomes important, and a significant concentration of
free electrons exists as a consequence of the
equilibrium
M M+
+ e-
The equilibrium constant K for this reaction takes the
form
K= [M+
][e-
]
[M]
Inductively Coupled Plasmas
1- The inductively coupled plasma is twice as hot as a combustion
flame (Figure 21-11).
2- The high temperature, stability, and relatively inert Ar environment
in the plasma eliminate much of the interference encountered with
flames.
3- Simultaneous multi-element analysis is routine for inductively
coupled plasma atomic emission spectroscopy, which has replaced
flame atomic absorption.
4- The plasma instrument costs more to purchase and operate than a
flame instrument.
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
5- The concentration of analyte needed for adequate signal is reduced
by an order of magnitude with an ultrasonic nebulizer, in which sample
solution is directed onto a piezoelectric crystal oscillating at 1 MHz.
6- The vibrating crystal creates a fine aerosol that is carried by a
stream of Ar through a heated tube, where solvent evaporates. The
stream then passes through a refrigerated zone in which solvent
condenses and is removed. Analyte reaches the plasma flame as an
aerosol of dry, solid particles.
7- Plasma energy is not needed to evaporate solvent, so more energy is
available for atomization. Also, a larger fraction of the sample reaches
the plasma than with a conventional nebulizer.
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt
Chapter  21-Atomic Spectroscopy رلبلاب (1).ppt

More Related Content

PPTX
Atomic absorption spectroscopy
PPTX
CHM260 - AAS (i)
PPTX
Chapter 13_Part1.pptx mineral engineering
PPTX
Atomic absorption Spectrophotometry
PPT
1606732404-atomic-absorption-emission-2.ppt
PPTX
Atomic absorption spectroscopy lec
PPTX
Atomic absorption spectroscopy
PPT
1606732404-atomic-absorption-emission-2.ppt
Atomic absorption spectroscopy
CHM260 - AAS (i)
Chapter 13_Part1.pptx mineral engineering
Atomic absorption Spectrophotometry
1606732404-atomic-absorption-emission-2.ppt
Atomic absorption spectroscopy lec
Atomic absorption spectroscopy
1606732404-atomic-absorption-emission-2.ppt

Similar to Chapter 21-Atomic Spectroscopy رلبلاب (1).ppt (20)

PPT
Atomic spectroscopy
PPTX
Atomic absorption spectrophotometry
PPTX
Atomic absorption spectroscopy, History, atomization techniques, and instrume...
PPTX
Atomic absorption Spectroscopy Instrumentation .pptx
PDF
Atomic absorption-spectrometry
PDF
Atomic absorption-spectrometry
PDF
Atomic absorption spectroscopy
PPT
1606732404-atomic-absorption-emission-2.ppt
PDF
Atomic_Spectroscopyyyy2024-25new pdf.pdf
PDF
PPTX
Atomic absorption spectroscopy (AAS).pptx
PPT
Atomic Absorption Emissionnnnnnnnnnnnnn
PDF
Instru booklet
PPTX
Presantation of aas
PPT
Atomic_Absorption_Emission.ppt
PPTX
Flame and atomic abosrption spectrophometry
PDF
Atomic Absorption Spectroscopy (AAS)
PPTX
flame emission and AAS.pptx
PPTX
AAS mdnovjnsvrgewrgerwgerger 4egfw4rg wrg4rg4wrg wreuvnuPPT.pptx
PPTX
Atomic Absorption Spectroscopy
Atomic spectroscopy
Atomic absorption spectrophotometry
Atomic absorption spectroscopy, History, atomization techniques, and instrume...
Atomic absorption Spectroscopy Instrumentation .pptx
Atomic absorption-spectrometry
Atomic absorption-spectrometry
Atomic absorption spectroscopy
1606732404-atomic-absorption-emission-2.ppt
Atomic_Spectroscopyyyy2024-25new pdf.pdf
Atomic absorption spectroscopy (AAS).pptx
Atomic Absorption Emissionnnnnnnnnnnnnn
Instru booklet
Presantation of aas
Atomic_Absorption_Emission.ppt
Flame and atomic abosrption spectrophometry
Atomic Absorption Spectroscopy (AAS)
flame emission and AAS.pptx
AAS mdnovjnsvrgewrgerwgerger 4egfw4rg wrg4rg4wrg wreuvnuPPT.pptx
Atomic Absorption Spectroscopy
Ad

More from salimaabajja1 (20)

PPTX
Thermodynamicrererererereerererrrer.pptx
PPTX
Reactions of Alkyl Halidesfdfdffffdfdffdd.pptx
PPTX
acid base catalysisvcvccvcvxvxvxvxxvxvxvxv.pptx
PPT
386630757667676667676767667766776660.ppt
PPT
5A-SURFACE-CHEMISTRYtjrwqiooritiwutt.ppt
PPT
5084798SSFETGFFGFGFGFGFGFGFGFGFGFGFG.ppt
PPT
GPIMASSsoooooooooooooooooo444444oooo.ppt
PPTX
Arab223-Week1,;kjjh;jhjhlkljllkjkkl3.pptx
PPTX
QNIS-2019-Prof-Yellowle,m,m,m,mmmes.pptx
PPTX
nmr&mass phramHKJ';JLJK'KLKKLKLKKL;.pptx
PPTX
Electrochemistry_corrosio7898عمنانn.pptx
PPT
catalystsdddu5euruwetweuwtyweyweryrwey.ppt
PPT
chapter 16-5ggdgdggdgdgdgdgdddfdfdfd.ppt
PPTX
Enzymesteryreeryryyryrr645445454454.pptx
PPT
Chapter 21-Atomic Spectroscopymnkkk.ppt
PPT
534453hfgjfghfghhfhsdhdhsdhgdhsdhdhsdhsd3.ppt
PPTX
chemical shvxvxvxvxvxvxvxvxvxvxvxvift.pptx
PPTX
essential oil.ppt lkfjdlgkjaldgkjfakajakfa'flkd
PPT
NMR spectroscopy.ppfkjsgasglkagklahskasdgl;sl
PPT
Chem7JD;SLKGJSDLKJSKGJSFLKJG;LASDKGASLGLA250p2.ppt
Thermodynamicrererererereerererrrer.pptx
Reactions of Alkyl Halidesfdfdffffdfdffdd.pptx
acid base catalysisvcvccvcvxvxvxvxxvxvxvxv.pptx
386630757667676667676767667766776660.ppt
5A-SURFACE-CHEMISTRYtjrwqiooritiwutt.ppt
5084798SSFETGFFGFGFGFGFGFGFGFGFGFGFG.ppt
GPIMASSsoooooooooooooooooo444444oooo.ppt
Arab223-Week1,;kjjh;jhjhlkljllkjkkl3.pptx
QNIS-2019-Prof-Yellowle,m,m,m,mmmes.pptx
nmr&mass phramHKJ';JLJK'KLKKLKLKKL;.pptx
Electrochemistry_corrosio7898عمنانn.pptx
catalystsdddu5euruwetweuwtyweyweryrwey.ppt
chapter 16-5ggdgdggdgdgdgdgdddfdfdfd.ppt
Enzymesteryreeryryyryrr645445454454.pptx
Chapter 21-Atomic Spectroscopymnkkk.ppt
534453hfgjfghfghhfhsdhdhsdhgdhsdhdhsdhsd3.ppt
chemical shvxvxvxvxvxvxvxvxvxvxvxvift.pptx
essential oil.ppt lkfjdlgkjaldgkjfakajakfa'flkd
NMR spectroscopy.ppfkjsgasglkagklahskasdgl;sl
Chem7JD;SLKGJSDLKJSKGJSFLKJG;LASDKGASLGLA250p2.ppt
Ad

Recently uploaded (20)

PDF
The scientific heritage No 166 (166) (2025)
PPTX
POULTRY PRODUCTION AND MANAGEMENTNNN.pptx
PDF
ELS_Q1_Module-11_Formation-of-Rock-Layers_v2.pdf
PDF
Formation of Supersonic Turbulence in the Primordial Star-forming Cloud
PDF
CHAPTER 3 Cell Structures and Their Functions Lecture Outline.pdf
PDF
An interstellar mission to test astrophysical black holes
PPT
Heredity-grade-9 Heredity-grade-9. Heredity-grade-9.
PDF
Unveiling a 36 billion solar mass black hole at the centre of the Cosmic Hors...
PPTX
Pharmacology of Autonomic nervous system
PPTX
Hypertension_Training_materials_English_2024[1] (1).pptx
PPTX
ognitive-behavioral therapy, mindfulness-based approaches, coping skills trai...
PDF
Biophysics 2.pdffffffffffffffffffffffffff
PPTX
Application of enzymes in medicine (2).pptx
PDF
GROUP 2 ORIGINAL PPT. pdf Hhfiwhwifhww0ojuwoadwsfjofjwsofjw
PDF
Placing the Near-Earth Object Impact Probability in Context
PDF
Cosmic Outliers: Low-spin Halos Explain the Abundance, Compactness, and Redsh...
PDF
BET Eukaryotic signal Transduction BET Eukaryotic signal Transduction.pdf
PPTX
Fluid dynamics vivavoce presentation of prakash
PDF
Looking into the jet cone of the neutrino-associated very high-energy blazar ...
PDF
CAPERS-LRD-z9:AGas-enshroudedLittleRedDotHostingaBroad-lineActive GalacticNuc...
The scientific heritage No 166 (166) (2025)
POULTRY PRODUCTION AND MANAGEMENTNNN.pptx
ELS_Q1_Module-11_Formation-of-Rock-Layers_v2.pdf
Formation of Supersonic Turbulence in the Primordial Star-forming Cloud
CHAPTER 3 Cell Structures and Their Functions Lecture Outline.pdf
An interstellar mission to test astrophysical black holes
Heredity-grade-9 Heredity-grade-9. Heredity-grade-9.
Unveiling a 36 billion solar mass black hole at the centre of the Cosmic Hors...
Pharmacology of Autonomic nervous system
Hypertension_Training_materials_English_2024[1] (1).pptx
ognitive-behavioral therapy, mindfulness-based approaches, coping skills trai...
Biophysics 2.pdffffffffffffffffffffffffff
Application of enzymes in medicine (2).pptx
GROUP 2 ORIGINAL PPT. pdf Hhfiwhwifhww0ojuwoadwsfjofjwsofjw
Placing the Near-Earth Object Impact Probability in Context
Cosmic Outliers: Low-spin Halos Explain the Abundance, Compactness, and Redsh...
BET Eukaryotic signal Transduction BET Eukaryotic signal Transduction.pdf
Fluid dynamics vivavoce presentation of prakash
Looking into the jet cone of the neutrino-associated very high-energy blazar ...
CAPERS-LRD-z9:AGas-enshroudedLittleRedDotHostingaBroad-lineActive GalacticNuc...

Chapter 21-Atomic Spectroscopy رلبلاب (1).ppt

  • 2. 1- In atomic spectroscopy, samples are vaporized at 2 000–8 000 K and decompose into atoms. 2- Concentrations of atoms in the vapor are measured by emission or absorption of characteristic wavelengths of radiation. 3- Atomic spectroscopy is a principal tool of analytical chemistry because of its high sensitivity, its ability to distinguish one element from another in a complex sample, its ability to perform simultaneous multi-element analyses, and the ease with which many samples can be automatically analyzed. 4- The precision of atomic spectroscopy, typically 1–2%, is not as good as that of some wet chemical methods. 5- The equipment is expensive, but widely available.
  • 3. 1- An important difference between atomic and molecular spectroscopy is the width of absorption or emission bands. Spectra of liquids and solids typically have bandwidths of 100 nm. 2- Spectra of gaseous atoms consist of sharp lines with widths of 0.001 nm. 3- Lines are so sharp that there is usually little overlap between the spectra of different elements in the same sample. Therefore, some instruments can measure more than 70 elements simultaneously.
  • 6. 4- In atomic fluorescence experiment. Atoms in the flame are irradiated by a laser to promote them to an excited electronic state from which they can fluoresce to return to the ground state. 5- Atomic fluorescence is potentially a thousand times more sensitive than atomic absorption, but equipment for atomic fluorescence is not common. 6- In atomic emission, collisions in the very hot plasma promote some atoms to excited electronic states from which they can emit photons to return to lower-energy states. 7- In atomic emission, no lamp is required. 8- Emission intensity is proportional to the concentration of the element in the sample. Emission from atoms in a plasma is now the dominant form of atomic spectroscopy.
  • 7. Flame Atomization: In a flame atomizer, a solution of the sample is nebulized by a flow of gaseous oxidant, mixed with a gaseous fuel, and carried into a flame where atomization occurs. The following processes then occur in the flame. • Desolvation (produce a solid molecular aerosol) • Dissociation (leads to an atomic gas) • Ionization (to give cations and electrons) • Excitation (giving atomic, ionic, and molecular emission)
  • 10. Types of Flames: Several common fuels and oxidants can be employed in flame spectroscopy depending on temperature needed. Temperatures of 1700o C to 2400o C are obtained with the various fuels when air serves as the oxidant. At these temperature, only easily decomposed samples are atomized. For more refractory samples, oxygen or nitrous oxide must be employed as the oxidant. With the common fuels these oxidants produce temperatures of 2500o C to 3100o C.
  • 12. Burning Velocity: The burning velocities are of considerable importance because flames are stable in certain ranges of gas flow rates only. If the gas flow rate does not exceed the burning velocity, the flame propagates itself back in to the burner, giving flashback. As the flow rate increases, the flame rises until it reaches a point above the burner where the flow velocity and the burning velocity are equal. This region is where the flame is stable. At higher flow rates, the flame rises and eventually reaches a point where it blows off of the burner.
  • 13. • Flame Structure: Important regions of a flame include: 1. primary combustion zone 2. interzonal region 3. secondary combustion zone 1. Primary combustion zone: Thermal equilibrium is ordinarily not reached in this region, and it is, therefore, seldom used for flame spectroscopy.
  • 15. 2. Interzonal region: This area is relatively narrow in stoichiometric hydrocarbon flames, is often rich in free atoms and is the most widely used part of the flame for spectroscopy. 3. Secondary combustion zone: In the secondary reaction zone, the products of the inner core are converted to stable molecular oxides that are then dispersed into the surroundings.
  • 16. Temperature Profiles: A temperature profile of a typical flame for atomic spectroscopy is shown in Fig. 9-3. The maximum temperature is located in the flame about 1 cm above the primary combustion zone. It is important– particularly for emission methods – to focus the same part of the flame on the entrance slit for all calibrations and analytical measurements.
  • 18. Flame absorbance Profiles: Fig. 9-4 shows typical absorption profiles for three elements. Magnesium exhibits a maximum in absorbance at the middle of the flame. The behavior of silver, which is not readily oxidized, is quite different, a continuous increase in the number of atoms, and thus the absorbance, is observed from the base to the periphery of the flame. Chromium, which forms very stable oxides, shows a continuous decrease in absorbance beginning close to the burner tip.
  • 20. Flame Atomizers: Figure 9-5 is a diagram of a typical commercial laminar flow burner that employs a concentric tube nebulizer. The aerosol is mixed with fuel. The aerosol, oxidant, and fuel are then burned in a slotted burner that provides a flame that is usually 5 or 10 cm in length.
  • 22. Advantages: 1. Uniform dropsize 2. Homogeneous flame 3. Quiet flame and a long path length Disadvantages: 1. Flash back if Vburning > Vflow 2. ~90% of sample is lost 3. Large mixing volume
  • 23. Performance Characteristics Of Flame Atomizers 1- In terms of reproducible behavior, flame atomization appears to be superior to all other methods for liquid sample introduction. 2- In terms of sampling efficiency and thus sensitivity, however, other atomization methods are markedly better. 3-A large portion of the sample flows down the drain and the residence time of individual atoms in the optical path in the flame is brief (~10-4 s).
  • 24. Electrothermal Atomization 1- It provides enhanced sensitivity because the entire sample is atomized in a short period, and the average residence time of the atoms in the optical path is a second or more. 2- A few microliters of sample are first evaporated at a low temperature and then ashed at a somewhat higher temperature in an electrically heated graphite tube or in a graphite cup. Then the current is rapidly increased to several hundred amperes, which caused the temperature to soar to perhaps 2000o C to 3000o C; atomization of the sample occurs in a period of a few milliseconds to seconds.
  • 26. Performance Characteristics: 1-Electrothermal atomizers offer the advantage of unusually high sensitivity for small volumes of sample. 2- Typically, sample volumes between 0.5 and 10 L are used; 3- absolute detection limits lie in the range of 10-10 to 10-13 g of analyte. 4- A final disadvantage is that the analytical range is low, being usually less than two orders of magnitude.
  • 27. ATOMIC ABSORPTION INSTRUMENTATION Instruments for atomic absorption spectrometry (AAS) consist of a radiation source, a sample holder, a wavelength selector, a detector, and a signal processor and readout. The sample holder in atomic absorption instruments is the atomizer cell that contains the gaseous atomized sample.
  • 28. 1. Radiation Sources: It is necessary that band width of the radiation source must be narrow relative to the width of an absorption peak. The problem created by limited width of atomic absorption peaks has been solved by the use of line sources with bandwidths even narrower than absorption peaks.
  • 29. Fig. 9-10a shows the emission spectrum of a typical atomic lamp source. With a suitable filter or monochromator, all but one of these lines are removed. Fig. 9-10b shows the absorption spectrum for the analyte between wavelengths 1 and 2. Passage of the line from the source through the flame reduces its intensity from P0 to P; the absorbance is then given by log(Po/P), which is linearly related to the concentration of the analyte in the sample. A disadvantage of the procedure is that separate lamp source is needed for each element.
  • 31. Hollow Cathode Lamps: It is the most common source for atomic absorption measurements. This lamp consists of a tungsten anode and a cylindrical cathode sealed in a glass tube that is filled with neon or argon at a pressure of 1 to 5 torr. The cathode is constructed of the metal whose spectrum is desired. Ionization of the inert gas occurs when a potential on the order of 300 V is applied across the electrodes, which generates a current of about 5 to 15 mA.
  • 32. … Hollow Cathode Lamps continued… If the potential is sufficiently large, the gaseous cation acquire enough kinetic energy to dislodge some of the metal atoms from the cathode surface and produce an atomic cloud in a process called sputtering. A portion of the sputtered metal atom are in excited states and thus emit their characteristic radiation as they return to the ground state. Eventually, the metal atoms diffuse back to the cathode surface or to the glass walls of the tube and are redeposited.
  • 34. Electrodeless Discharge Lamps (EDLs): These provide radiant intensities that are usually one to two orders of magnitude greater than hollow cathode lamps. A typical lamp is constructed from a sealed quartz tube containing a few torr of an inert gas such as argon and a small quantity of the metal (or its salt) whose spectrum is of interest.
  • 35. …Electrodeless Discharge Lamps (EDLs) continued… The lamp is energized by an intense field of radio-frequency or microwave radiation. Ionization of the argon occurs to give ions that are accelerated by the high-frequency component of the field until they gain sufficient energy to excite the atoms of the metal whose spectrum is sought. Electrodeless discharge lamps are available commercially for 15 or more elements.
  • 37. Single-Beam Instruments: A typical single-beam instrument, consists of several hollow cathode sources, an atomizer, and simple grating spectrophotometer with a photomultiplier transducer. The 100% T adjustment is then made while a blank is aspirated into the flame. Finally, the transmittance is obtained with the sample replacing the blank.
  • 38. Double-Beam Instruments: In double-beam instrument the beam from the hollow cathode source is split by a mirrored chopper, one half passing through the flame and the other half around it. The two beams are then recombined by a half-silvered mirror and passed into a grating monochromator; a photomultiplier tube serves as the transducer. The ratio between the reference and sample signal is then amplified and fed to the readout, which may be a digital meter or a signal recorder.
  • 40. 40 2. Double beam (most commercial instruments) – Light is split and directed towards both reference cell (blank) and sample cell – Two detectors; electronics measure ratio (i.e., measure/calculate absorbance) – Advantages: • Compensates for fluctuations in source intensity and drift in detector • Better design for continuous recording of spectra
  • 41. 41 Merits of Double Beam Instruments 1.Compensate for all but the most short term fluctuation in radiant output of the source 2.Compensate drift in transducer and amplifier 3.Compensate for wide variations in source intensity with wavelength
  • 42. INTERFERENCES IN ATOMIC ABSORPTION SPECTROSCOPY 1. Spectral Interferences: (I) Spectral interference can occur due to overlapping lines. e.g. a vanadium line at 3082.11Å interferes in an analysis based upon the aluminum absorption line at 3082.15 Å. This type of interference can be avoid by employing the aluminum line at 3092.7 Å instead. (II) Spectral interferences result from the presence of combustion products that exhibit broadband absorption or particulate products that scatter radiation. Both diminish the power of the transmitted beam. A blank can be aspirated into the flame to make the correction.
  • 43. …Spectral Interferences continued… (III) Source of absorption or scattering can be originated in the sample matrix. An example of a potential matrix interference due to absorption occurs in the determination of barium in alkaline earth mixture. The wavelength of Ba line used for atomic absorption analysis appears in the center of a broad absorption band for CaOH. The effect can be eliminated by substituting nitrous oxide for air as the oxidant which yields a higher temperature that decomposed the CaOH and eliminates the absorption band.
  • 45. …Spectral Interferences continued… (IV) Concentrated solution of elements such as Ti, Zr and W which form refractory oxides can cause spectral interference due to scattering. (V) Organic solvent or organic impurities in the sample can cause scattering interference from carbonaceous particle because of incomplete combustion of the organic matrix.
  • 46. 2. Chemical Interferences: (I) Formation of Compounds of Low Volatility: The most common type of interference is by anions that form compounds of low volatility with the analyte and thus reduce the rate at which the analyte is atomized. The decrease in calcium absorbance that is observed with increasing concentrations of sulfate or phosphate. Example of cation interference have also been recognized. Aluminum is found to cause low results in the determination of magnesium, apparently as a result of the formation of a heat-stable aluminum/magnesium compound.
  • 47. …Formation of Compounds of Low Volatility continued… Interference due to formation of species of low volatility can often be eliminated or moderated by use of higher temperatures. Releasing agents which are cations that react preferentially with the interferant and prevent its interaction with the analyte, can be employed. Protective agents prevent interference by forming stable but volatile species with the analyte. Three common reagents for this purpose are EDTA, 8-hydroxyquinoline, and APDC (ammonium salt of 1- pyrrolidinecarbodithioic acid).
  • 48. …Chemical Interferences continued… (II) Dissociation Equilibria: Gaseous environment of a flame or a furnace, numerous dissociation and association reactions lead to conversion of the metallic constituents to the elemental state. Some of these reactions are reversible MO M + O M(OH)2 M + 2OH Where M is the analyte atom. VOx V + Ox AlOx Al + Ox TiOx Ti + Ox
  • 49. …Chemical Interferences continued… (III) Ionization Equilibria: Ionization of atoms and molecules is small in combustion mixtures that involve air as the oxidant, and generally can be neglected. In higher temperatures of flames where oxygen or nitrous oxide serves as the oxidant, however, ionization becomes important, and a significant concentration of free electrons exists as a consequence of the equilibrium M M+ + e- The equilibrium constant K for this reaction takes the form K= [M+ ][e- ] [M]
  • 50. Inductively Coupled Plasmas 1- The inductively coupled plasma is twice as hot as a combustion flame (Figure 21-11). 2- The high temperature, stability, and relatively inert Ar environment in the plasma eliminate much of the interference encountered with flames. 3- Simultaneous multi-element analysis is routine for inductively coupled plasma atomic emission spectroscopy, which has replaced flame atomic absorption. 4- The plasma instrument costs more to purchase and operate than a flame instrument.
  • 52. 5- The concentration of analyte needed for adequate signal is reduced by an order of magnitude with an ultrasonic nebulizer, in which sample solution is directed onto a piezoelectric crystal oscillating at 1 MHz. 6- The vibrating crystal creates a fine aerosol that is carried by a stream of Ar through a heated tube, where solvent evaporates. The stream then passes through a refrigerated zone in which solvent condenses and is removed. Analyte reaches the plasma flame as an aerosol of dry, solid particles. 7- Plasma energy is not needed to evaporate solvent, so more energy is available for atomization. Also, a larger fraction of the sample reaches the plasma than with a conventional nebulizer.