ECG
•  Interpretation of
  the electrical activity of
  the heart over a period of time
• Electrical manifestation of the
  activity of the heart. Recorded
  with surface electrodes on the
  limbs or chest.
• ECG is used to measure the rate
  and regularity of heartbeats, the
  presence of any damage to the
  heart.




                                       Page 2
Page 3
Page 4
ECG is detected by Einthoven's triangle
               method

                                          Page 5
THE HEART
• It is a 4 chambered muscular
  organ consist of 2 atriums
  and 2 ventricles.
• It function in a regular
  fashion to pump blood
  thought the body.
• Average heart rate of a
  human being is 72beats/min




                                   Page 6
• SA node- sinoartial
  node- acts like a
  function generator,
  called pacemaker of
  heart
• AV node-
  artioventricular node-
  acts like a delay device




                             Page 7
Page 8
HEART WAVES




                                                        QRS
 P wave- SA
                                                   wave- cumulati
    node
                                      Time               ve
genetrated, du   Amplitude- 0.1                                     Amplitude-   Time duration-
                                  duration- 60-8   repolarization
     e to          to 0.2 mV                                          1mV            80ms
                                       0ms           of atria &
contraction of
                                                   depolarization
    atria.
                                                    of ventricle.




                                                                                    Page 9
T wave-
                   repolarization of
                      ventricle




Amplitude-                                   Amplitude-
 0.01mV                                       0.3mV




       U wave- after
                                   Time duration-
        potential in
                                     120-160ms
         ventricle
                                                          Page 10
Page 11
DETECTION OF QRS EVENT


Using PAN tompkins    Recognises QRS on
                          the bases of
     algorithm
                     slope, width, amplitude




                                               Page 12
Band-pass filter




    In order to
 attenuate noise       Attenuates low
   the signal is      frequency of P &
passed through a     T waves, baseline
 bandpass filter,          drift, high
 composed of a          frequencies of
 high pass filter     EMG and power
 and a low pass       line interference
       filter.

                                      Page 13
Page 14
High pass   It has a DC
filter      gain of 32




                          Page 15
Differentiation




In differentiation, technique
   for finding slopes that
distinguish QRS wave from
      other ECG wave.




                                          Page 16
Page 17
Squaring


                      Makes all data
Point-by-point
                       positive and
squaring of the
                   accentuates higher
signal samples
                   frequencies (QRS)




                                Page 18
Page 19
Adaptive thresholding




Adaptive thresholding
                          It updates the peak
 is done on the basis
                           level & noise level
of BPF waveforms &
                         and determine if it is a
    moving-window
                            QRS event or not
       integral




                                             Page 20
Artifacts in ECG

                   High
                   frequency
                   noise

                   Wandering
                   base line

                   Power line
                   interference




                               Page 21
High frequency noise




Maybe formed by instrumentation
   amplifier, recording device or
electrodes may pickup ambient EM
         signals by cables




                                    Page 22
Power line interface

                            Odd harmonics are
Maybe formed if ECG
                         generated which would be
signal is not properly
                         represented as spikes on
     sinusoidal
                              ECG waveform




                                           Page 23
Wandering Base line


               Isoelectric line is not fixed,
                   noise in the signal




   Patient needs to be
                                   Also known as motion
static, electrodes should
                                          artifact
         be clean



               Maybe arise due to patient
                movement, movement of
                external source, patient
               body’s (inside) movement.
                                                          Page 24
Page 25
Page 26

More Related Content

DOC
various type of artifacts in ECG signal & how it's remove
PDF
Amplifiers, filters and digital recording systems
PPTX
ECG Machine
PPT
cardiac output measurment
PPTX
Cardiac Pacemaker
PPTX
Vector cardiography
PPTX
Phonocardiography(PCG)
PPT
Pacemaker
various type of artifacts in ECG signal & how it's remove
Amplifiers, filters and digital recording systems
ECG Machine
cardiac output measurment
Cardiac Pacemaker
Vector cardiography
Phonocardiography(PCG)
Pacemaker

What's hot (20)

PDF
Sources of bioelectric potentials
PPTX
Implantable Cardioverter Defibrillators -ICD
PPTX
Jp's pacemaker
PDF
Medical Electronics - Hints for Slow Learner
PPTX
fundamentals of pacemaker
PPT
Cardiac pace makerspart 1
PPT
Unit 3 biomedical
PPT
Presentation paper
PDF
ARTIFACTS ON ECG RECORDING
PPT
Pacemaker
PPT
BASICS OF ECG
PPTX
Impedance plethysmography Technique.pptx
PPT
Electrical testing of pacemaker
PPTX
Defibrillator Biomedical instrumentation.pptx
PPTX
Holter monitoring
PPTX
Tilt Table Test
PPTX
Pacemaker
PDF
Biopotential electrodes
Sources of bioelectric potentials
Implantable Cardioverter Defibrillators -ICD
Jp's pacemaker
Medical Electronics - Hints for Slow Learner
fundamentals of pacemaker
Cardiac pace makerspart 1
Unit 3 biomedical
Presentation paper
ARTIFACTS ON ECG RECORDING
Pacemaker
BASICS OF ECG
Impedance plethysmography Technique.pptx
Electrical testing of pacemaker
Defibrillator Biomedical instrumentation.pptx
Holter monitoring
Tilt Table Test
Pacemaker
Biopotential electrodes
Ad

Viewers also liked (20)

PPTX
Ecg Signal Processing
PPT
PPT
Basics of ECG
PPT
Basics of ECG.ppt dr.k.subramanyam
PPT
ECG interpretation: the basics
PDF
Shortcut to ECG
PPTX
ELECTROCARDIOGRAPHY (ECG)
PDF
ECG easy way
PPTX
12 Leads Made Easy
PPTX
ECG Noise cancelling
PPTX
Systematic ECG Interpretation
PPTX
ECG basics
PPTX
ECG interpretation
PPTX
EEG artifacts
PPTX
Defibrillator (ppt)
PDF
A Simple and Robust Algorithm for the Detection of QRS Complexes
PDF
ANALYSIS OF ECG SIGNAL USING SELF ORGANISING MAP
DOCX
Difference between tdm & fdm
Ecg Signal Processing
Basics of ECG
Basics of ECG.ppt dr.k.subramanyam
ECG interpretation: the basics
Shortcut to ECG
ELECTROCARDIOGRAPHY (ECG)
ECG easy way
12 Leads Made Easy
ECG Noise cancelling
Systematic ECG Interpretation
ECG basics
ECG interpretation
EEG artifacts
Defibrillator (ppt)
A Simple and Robust Algorithm for the Detection of QRS Complexes
ANALYSIS OF ECG SIGNAL USING SELF ORGANISING MAP
Difference between tdm & fdm
Ad

Similar to Ecg ppt (20)

PPTX
basicsofecg-150707084411-lva1-app6892.pptx
PPT
Arrythmias and ek gs 1
PPTX
Pacemaker & cabg
PPTX
Essentials of ecg interpretation aphrs
PDF
Electrocardiography
PPTX
Pacemaker
PPT
PPT
Basic technical concepts in cardiac pacing
PPT
NEED OF PHYSIOLOGICAL AND ELECTROTHERAPY EQUIPMENTS
PDF
Intelligent Heart Disease Recognition using Neural Networks
PPT
ECG Workshop
PPTX
PDF
ECG 2.pdf
PDF
PPTX
ECG(EKG) BASICS
PPTX
Ecg ppt
PPTX
ECG Fundamentals.pptx
PPTX
Ecg ppp.pptx 2
PPT
basicsofecg-150707084411-lva1-app6892.pptx
Arrythmias and ek gs 1
Pacemaker & cabg
Essentials of ecg interpretation aphrs
Electrocardiography
Pacemaker
Basic technical concepts in cardiac pacing
NEED OF PHYSIOLOGICAL AND ELECTROTHERAPY EQUIPMENTS
Intelligent Heart Disease Recognition using Neural Networks
ECG Workshop
ECG 2.pdf
ECG(EKG) BASICS
Ecg ppt
ECG Fundamentals.pptx
Ecg ppp.pptx 2

More from Priya_Srivastava (20)

PPTX
Vitualisation
PPTX
PPTX
PPTX
Environment
PPTX
Environmental
PPTX
PPTX
PPTX
Technical ppt
PPTX
PPTX
Project ppt on Rapid Battery Charger using Fuzzy Controller
PPTX
Project ppt
PPTX
PPTX
PPTX
PPTX
PPTX
Seminar ppt...; )
PPTX
Technical presentation on
PPTX
PPTX
Vitualisation
Environment
Environmental
Technical ppt
Project ppt on Rapid Battery Charger using Fuzzy Controller
Project ppt
Seminar ppt...; )
Technical presentation on

Ecg ppt

  • 2. ECG • Interpretation of the electrical activity of the heart over a period of time • Electrical manifestation of the activity of the heart. Recorded with surface electrodes on the limbs or chest. • ECG is used to measure the rate and regularity of heartbeats, the presence of any damage to the heart. Page 2
  • 5. ECG is detected by Einthoven's triangle method Page 5
  • 6. THE HEART • It is a 4 chambered muscular organ consist of 2 atriums and 2 ventricles. • It function in a regular fashion to pump blood thought the body. • Average heart rate of a human being is 72beats/min Page 6
  • 7. • SA node- sinoartial node- acts like a function generator, called pacemaker of heart • AV node- artioventricular node- acts like a delay device Page 7
  • 9. HEART WAVES QRS P wave- SA wave- cumulati node Time ve genetrated, du Amplitude- 0.1 Amplitude- Time duration- duration- 60-8 repolarization e to to 0.2 mV 1mV 80ms 0ms of atria & contraction of depolarization atria. of ventricle. Page 9
  • 10. T wave- repolarization of ventricle Amplitude- Amplitude- 0.01mV 0.3mV U wave- after Time duration- potential in 120-160ms ventricle Page 10
  • 12. DETECTION OF QRS EVENT Using PAN tompkins Recognises QRS on the bases of algorithm slope, width, amplitude Page 12
  • 13. Band-pass filter In order to attenuate noise Attenuates low the signal is frequency of P & passed through a T waves, baseline bandpass filter, drift, high composed of a frequencies of high pass filter EMG and power and a low pass line interference filter. Page 13
  • 15. High pass It has a DC filter gain of 32 Page 15
  • 16. Differentiation In differentiation, technique for finding slopes that distinguish QRS wave from other ECG wave. Page 16
  • 18. Squaring Makes all data Point-by-point positive and squaring of the accentuates higher signal samples frequencies (QRS) Page 18
  • 20. Adaptive thresholding Adaptive thresholding It updates the peak is done on the basis level & noise level of BPF waveforms & and determine if it is a moving-window QRS event or not integral Page 20
  • 21. Artifacts in ECG High frequency noise Wandering base line Power line interference Page 21
  • 22. High frequency noise Maybe formed by instrumentation amplifier, recording device or electrodes may pickup ambient EM signals by cables Page 22
  • 23. Power line interface Odd harmonics are Maybe formed if ECG generated which would be signal is not properly represented as spikes on sinusoidal ECG waveform Page 23
  • 24. Wandering Base line Isoelectric line is not fixed, noise in the signal Patient needs to be Also known as motion static, electrodes should artifact be clean Maybe arise due to patient movement, movement of external source, patient body’s (inside) movement. Page 24