FDA Expectations for

             in Device & Diagnostic Design




                        John Avellanet
                        Cerulean Associates LLC
                        www.CeruleanLLC.com
                        © Cerulean Associates LLC all rights reserved
Acknowledgements
                  Tim Ulatowski
                   Dan O’Leary
                 Anita Fauchier
                    Mike Weber
                  Nancy Singer
               Karen Steinberg
                Kerry McCarter
                    Akos Bartha
                 Carl Anderson
               Jackie Cassada
                   John Lincoln
                      Jim Shore
                  Tom Colonna
                 Terry Winchell
             Chris Szustkiewicz
                 Kim Trautman
               Larry Nicholson
                     Jan Welch
                    David Elder
             Annamarie Kempic
            Doug Throckmorton
                  Jonathan Lee




                                  2
Agenda
                   fda requirements
                practical challenges
lean compliance solutions framework




     This is not legal advice. Information in this presentation draws upon a variety of sources, including published FDA warning letters, personal experiences,
interviews and research, all or any of which may or may not have been prepared or conducted by Cerulean Associates LLC. Cerulean Associates LLC does not
    provide a warranty concerning the accuracy of the information contained in this presentation. The contents of this presentation are intended for general
   information only and should not be construed as legal advice. Cerulean Associates LLC assumes no liability for actions taken or not taken as a result of the
                                         information in this presentation. This presentation is copyrighted 2011 by Cerulean Associates LLC, all rights reserved.

                                                                                                                                                                    3
FDA Requirements
       traceability regulations
recent enforcement examples
 example inspector questions
           other requirements




                                  4
21 CFR 820.30
(a) “General. (1) Each manufacturer…shall establish and maintain
procedures to control the design of the device in order to ensure that
specified design requirements are met.”



Translation for FDA inspectors
   “The purpose of the design control subsystem is to
   control the design process to assure that devices meet
   user needs, intended uses, and specified requirements:
       • Inputs must be documented
       • Outputs must be documented
       • Confirm that device outputs are traceable to design inputs”

                                                                         5
21 CFR 803.18
(b) (1) (i) “…including all documentation of your deliberations and
decisionmaking processes used to determine if a device-related death,
serious injury, or malfunction was or was not reportable….”



Translation for FDA investigators
   “A firm must demonstrate that it exercised “good faith” in
   any attempts to obtain required data…. In addition, the
   Center believes that the parameters of good faith must,
   at a minimum, comport with the level of risk/nature of
   the device….”


                                                                        6
21 CFR 806.20
(b) (4) “Justification for not reporting the correction or removal action to
the FDA, which shall contain conclusions and any followups….”



Translation for FDA investigators
   “Verify that non-reported device corrections or removals
   meet the following criteria based on design controls:
       • Risk is not increased
       • Repairs are not unexpected
       • Part replacement is not earlier than expected”




                                                                               7
Warning Letter Excerpts
“ You have failed to establish and maintain a
  design history file (DHF) to contain and
  reference the records necessary to
  demonstrate that the design was developed in
  accordance with the approved design plan and
  the requirements of 21 CFR 820.”
                    - warning letter to ASI Medical, 20 April 2011




                                                                     8
Warning Letter Excerpts
“ Failure to maintain adequate device history
  records…. For example, a review of (b)(4)
  Meniscal Insert DHRs found one DHR
  contained an inaccurate date.”
     - warning letter to Advanced Surgical Design & Manufacture, 1 December 2010




                                                                                   9
Warning Letter Excerpts
“ When requested, design output requirements
  for the upgrade from Version (b)(4) to Version
  (b)(4) done by (b)(4)*…could not be
  provided.”
                 - warning letter to 3CPM Company, 25 March 2010




                 *This change was done 7 years prior in 2003!


                                                                   10
Example Inspector Questions
•   were design characteristic acceptance criteria established and documented
    prior to design approval?
•   were design outputs verified and documented against design input
    requirements?
•   was risk analysis conducted (and documented) on the design prior to its
    final design approval?
•   do design outputs draw a distinction between those characteristics essential
    for functioning of the device versus those non-essential (e.g., aesthetic)
    characteristics? How are these verified?
•   review design outputs linkages wherein design outputs for one stage
    become design inputs for the next – how are these made clear in the
    documentation?
•   verify that changes to device design were documented and approved – are
    there date discrepancies between the actual change and its approval?
•   how were design changes tested to assess impact to risk or to adjacent
    design inputs or outputs? How was this documented?
•   is there a documented traceability analysis or matrix linking product design
    requirements, design specifications, risks and controls, and tests?


                                                                                   11
Other Requirements
FDA Guidance Documents
•   Human Factors Points to Consider for IDE Devices (1996)
•   Human Factors Implications of the New GMP Rule: Overall Requirements of
    the New Quality System Regulations (1997)
•   Design Control Guidance for Medical Device Manufacturers (1997)
•   Medical Device Use-Safety: Incorporating Human Factors Engineering into
    Risk Management (2000)
•   General Principles of Software Validation (2002)
•   Guidance for the Content of Premarket Submissions for Software Contained
    in Medical Devices (2005)
•   Radiation Safety Considerations for X-Ray Equipment Designed for Hand-
    Held Use (2008)
•   Technical Considerations for Pen, Jet, and Related Injectors Intended for
    Use with Drugs and Biological Products (2009)
•   Clinical Investigations of Devices Indicated for the Treatment of Urinary
    Incontinence (2011)


                                                                                12
Other Requirements
GHTF Guidelines
• SG1 N:068 Essential Principles of Safety and Performance of
  Medical Devices (2010 draft)
• SG1 N:11 Summary Technical Documentation for
  Demonstrating Conformity to the Essential Principles of
  Safety and Performance of Medical Devices (STED) (2008)
• SG3 N:15 Implementation of Risk Management Principles and
  Activities Within a Quality Management System (2005)
• SG5 N:2 Clinical Evaluation (2007)
• SG5 N:3 Clinical Investigations (2010)
• AHWG N:2 Unique Device Identification (UDI) System for
  Medical Devices (2010 draft)


                                                                13
Practical Constraints
     industry confusion
                   costs
             ownership
   development lifecycle




                           14
Industry Confusion
   Design Requirements   Issues/Defects/Anomalies




          Tests               Risk Controls




                                    Source:
                                    Life Sciences Product Development Artifacts
                                    Survey White Paper (Seapine Software, 2011)
                                                                                  15
Costs                  tracking, verifying earlier tests, add’l tests, etc.)
Cost of Change (incl. documentation,




                                                                                                                                                                                           Perception
                                                                                                                                                                                           Reality




                                                                                               Requirements   Early Design   Protype Production   Clinical   Final Production

                                                                                                                    Time (development cycle)
                                                                                                                                                                   Source:
                                                                                                                                                                   Adapted from Scott Ambler, Agile Modeling (2002)
                                                                                                                                                                                                                16
Ownership

                       Quality



     Validation /
                                      Engineering
       Testing




              Design             Production


                                                    17
Development Lifecycle
             start traceability …?




     ?

         ?
                             ?
              ?




                                     Source:
                                     Get to Market Now! Turn FDA Compliance
                                     into a Competitive Edge (2010, Logos Press)
                                                                                   18
Practically Speaking
• address records sought by inspectors

• capture logic through traceability

• need a systematic approach to traceability
     •   cost-effective
     •   generates records
     •   allows easy reporting
     •   maintain clear linkages

                                           19
Rationale (Traceability)
• relationship to safety

• relationship to effectiveness

• relationship to risk mitigation

• relationship to testing
     • corrective actions
     • design changes/fixes

                                    20
Lean Compliance Strategy
               do’s v. don’ts
          eight-step process
         lifecycle placement




                                21
Traceability Do’s
• Form a cross-functional traceability team
• Clearly define accountabilities v. responsibilities in your SOPs
• Start traceability efforts when you make the “go/no-go” decision on
  a project
• Ensure that each specific feature  risk/request  tests 
  production characteristics (suppliers, processes, etc.)
• Plan for at least one significant change during product development
• Verify progress (and identify gaps) with a mock FDA audit




                                                                        22
Traceability Don’ts
• Assume that you can remember dates, times and people involved in
  design decisions 4, 5, 7+ years ago
• Rely only on internal audits by your Quality department to catch
  missing traceability records
• Assume FDA inspector will only go back a few months in design
  history
• Forget that traceability records help you prove safety, efficacy, and
  compliance
• Lose sight of traceability costs – design changes include
  documentation changes, test changes, test re-work, etc.




                                                                          23
Eight-Step Process
                     design control SOPs
         1
                     product concept sheet
     8       2
                     product specs
 7               3
                     risk controls
                     traceability matrix
     6       4       QC “yea/nay”
         5
                     management review
                     audit

                                             24
1
                          8       2



Design Control SOPs   7               3



                          6       4
                              5




                                          25
1
                                          8                      2



Product Concept Sheet                 7                               3



                                          6                      4
                                                     5



              • focus on:
                   • intended use
                   • desired features
                   • distinguishing
                     characteristics


              • development
                considerations
                            Source:
                            Get to Market Now! Turn FDA Compliance
                            into a Competitive Edge (2010, Logos Press)
                                                                          26
1
                                                    8            2



Product Specs                                   7                    3



                                                    6            4
                                                             5


Product Component   Requirement
seal integrity      seal integrity specifications
                       1. does not allow loss of product or
                          moisture over shelf life
                       2. retain sterility over shelf life
                       3. does not allow alteration to
                          finished product over shelf life


stopper             stopper specifications
                       1. biocompatible
                       2. not contain silicon
                       3. 20mm serum stopper
                                                                         27
1
                                                                      8       2



Risk Controls                                                    7                3



                                                                      6       4
                                                                          5


Product Specifications                 Control for
seal integrity specifications          seal integrity controls
1. does not allow loss of product or
                                          USP <1207>
   moisture over shelf life
2. retain sterility over shelf life       stability study (ICH Q6A)
3. does not allow alteration to           USP <381>
   finished product over shelf life       USP <87>


stopper                                stopper controls
1. biocompatible                          USP <88> or USP <381>
2. not contain silicon                    purchasing control
3. 20mm serum stopper                     see drawing
                                                                                      28
1
                                    8       2



Traceability Matrix             7               3



                                    6       4
                                        5


• manual (logs, spreadsheets)

• automated (database)
    • clear linkages
    • systemic reviews
    • red flag assessments




                                                    29
1
                                   8       2



Traceability Matrix (cont’d)   7               3



                                   6       4
                                       5




                                                   30
1
                                                 8       2



QC “Yea/Nay”                                 7               3



                                                 6       4
                                                     5


• are the test results…?
     • within specification ranges


• have a “Plan B”
     • if results indicate failure, what will you do?
     • decide potential actions ahead of time based on
       design and risk



                                                                 31
1
                                                    8        2



Management Review                               7                3



                                                    6        4
                                                         5


• during design transfer

• relate traceability data to…
     • risk assessments
     • design specifications for materials / suppliers


• do not forget: relate to cost
     • senior management must factor in ROI
                                                                     32
1
                                                    8       2



Audit                                          7                3



                                                    6       4
                                                        5


• audit traceability records prior to
     •   IDE submission (if any)
     •   market application submission (510(k))
     •   any licensing agreements (due diligence)
     •   patent applications / changes
     •   long-term archival


• every 2 years otherwise
     • consider a mock FDA audit and gap analysis
                                                                    33
Development Lifecycle & 8 Steps

                                                          (until longer of two
              Formal traceability                         years after end of
                                                          device/life of design
                                                          OR statute of
                                                          limitations)




      2   3           6     7

      1   4       5

              8                     8

                                        Adapted from:
                                        Get to Market Now! Turn FDA Compliance
                                        into a Competitive Edge (2010, Logos Press)
                                                                                      34
Key Takeaway Review
       core regulatory requirements
           four practical constraints
eight-step lean compliance strategy




                                        35
Want More?
Traceability Resources
http://www.seapine.com/traceability.html

Six Exercises To Strengthen Traceability
http://downloads.seapine.com/pub/papers/SixExercisesStrengthenTraceability.pdf

How To Have A Painless FDA Audit
http://downloads.seapine.com/pub/papers/PainlessFDAAudit.pdf

The Seapine View Blog
http://blogs.seapine.com/

Seapine Life Science Solutions
http://www.seapine.com/lifesciences.html
                                                                             36
About Your Presenter
                        John Avellanet solves compliance problems for clients with
                        practical solutions. Winner of the 2009 Best of Business award by
                        the Small Business Commerce Association, Mr. Avellanet has
                        earned international acclaim for his business-savvy, pragmatic
                        compliance advice.

                        His latest book, Get to Market Now! Turn FDA Compliance into
                        a Competitive Edge in the Era of Personalized Medicine, has
                        earned multiple five-star reviews from industry publications,
                        blogs, Amazon.com readers, and former FDA officials.

                        He has a breadth of experience designing, implementing, and
                        being accountable for quality systems and compliance programs
                        for FDA, the ICH, GHTF, and ISO. For more than 15 years, John
                        served as an executive accountable for compliance, records
www.Ceruleanllc.com     management, and information technology, most recently as a C-
                        level executive for a Fortune 50 combination medical device and
www.Get2MarketNow.com   biotech subsidiary.

                        In 2006, Mr. Avellanet founded his private FDA compliance
                        consulting and training firm, Cerulean Associates LLC.
                                                                                      37
Picture Credits

Photos, images and clip art that appear on these slides have been used to enhance this presentation and may NOT
be used for commercial or promotional purposes without permission from copyright holders. Do not remove or copy
                                                                                           from this presentation.



                                                                                                        Contact:
                                                                                                iStockphoto.com
                                                                                           Microsoft Corporation
                                                                                               Seapine Software
                                                                                        Cerulean Associates LLC




                                                                                                                     38

More Related Content

PPTX
An Overview for Software as a Medical Device (SaMD)
PDF
Understanding IEC 62304
PDF
IEC 62304: SDLC Conformance and Management
PDF
Cybersecurity and Software Updates in Medical Devices.pdf
 
PDF
IEC 62304 Action List
PDF
Usability Validation Testing of Medical Devices and Software
PDF
FDA Design Controls: What Medical Device Makers Need to Know
DOCX
Design and development of medical device
An Overview for Software as a Medical Device (SaMD)
Understanding IEC 62304
IEC 62304: SDLC Conformance and Management
Cybersecurity and Software Updates in Medical Devices.pdf
 
IEC 62304 Action List
Usability Validation Testing of Medical Devices and Software
FDA Design Controls: What Medical Device Makers Need to Know
Design and development of medical device

What's hot (20)

PDF
The 510(k) Process
PPTX
ISO 62304 & TIR 45
PPTX
US FDA Medical Device or Equipment
PDF
Classification of In Vitro Diagnostic Devices per FDA and IVDR Rules
PPTX
Regulation of Medical Devices in US
PPTX
Quality Control for Medical Device Software - It Arena Lviv Presentation
PDF
Presentation: Software as a Medical Device: Regulatory insights and Q & A
PDF
The European Medical Device Regulations - analysis of the final text
 
PPTX
Iso 13485:2016
PDF
Global unique device identification database (gudid) and udi
PDF
The In vitro diagnostic medical devices regulation (EU) 2017/746: what will c...
PPTX
Post Marketing Survelliance.pptx
PDF
Implementing a Global Unique Device Identification (UDI) Solution: Regional U...
PDF
Postmarket Surveillance Medical Devices
PPTX
Medical Device Regulations - 510(k) Process
PDF
Fda quality system regulation 21 CFR820_Medical devices_k_trautman
PDF
Software as a Medical Device (SaMD) - IMDRF Definition and Categorisation
 
PPT
Process Validation Master Planning DMAIC Fusion
PPTX
How to Comply with IDMP Regulatory Requirements
PPTX
ISO IDMP: Practical considerations from XEVMPD experience
The 510(k) Process
ISO 62304 & TIR 45
US FDA Medical Device or Equipment
Classification of In Vitro Diagnostic Devices per FDA and IVDR Rules
Regulation of Medical Devices in US
Quality Control for Medical Device Software - It Arena Lviv Presentation
Presentation: Software as a Medical Device: Regulatory insights and Q & A
The European Medical Device Regulations - analysis of the final text
 
Iso 13485:2016
Global unique device identification database (gudid) and udi
The In vitro diagnostic medical devices regulation (EU) 2017/746: what will c...
Post Marketing Survelliance.pptx
Implementing a Global Unique Device Identification (UDI) Solution: Regional U...
Postmarket Surveillance Medical Devices
Medical Device Regulations - 510(k) Process
Fda quality system regulation 21 CFR820_Medical devices_k_trautman
Software as a Medical Device (SaMD) - IMDRF Definition and Categorisation
 
Process Validation Master Planning DMAIC Fusion
How to Comply with IDMP Regulatory Requirements
ISO IDMP: Practical considerations from XEVMPD experience
Ad

Viewers also liked (20)

PDF
Design control FDA requirements
PDF
Systems Engineering and Requirements Management in Medical Device Product Dev...
PDF
Common Mistakes in the Medical Device Development Continuum
PDF
Good design practice for medical devices
PDF
Medical device development lifecycle
PDF
MEDICAL DEVICE DEVELOPMENT_0327
PDF
Life Sciences Product Development
PDF
Medical Product Development cycle
PPTX
One size does not fit all unique study management challenges for diagnostic ...
PPTX
Beyond FDA Compliance Webinar: 5 Hidden Benefits of Your Traceability Matrix
PDF
requirements_traceability_matrix_sample
PPTX
Personalised Medicine in the EU— Evolving Landscape and New HTA Considerations
PPTX
Devices Sponsor Information Day: 0 - Developments in medical device regulation
PDF
The FDA - Mobile, and Fixed Medical Devices Cybersecurity Guidance
PPTX
Regulatory approval process for invitro diagnostics in us
PDF
Poster Presentation
PDF
Medical Device Regulation
PDF
BioMedical Strategy Medical Devices Workshop Presentations
PDF
FDA’s approach to regulation of in vitro diagnostic tests
PDF
Systems Engineering in Medical Devices
Design control FDA requirements
Systems Engineering and Requirements Management in Medical Device Product Dev...
Common Mistakes in the Medical Device Development Continuum
Good design practice for medical devices
Medical device development lifecycle
MEDICAL DEVICE DEVELOPMENT_0327
Life Sciences Product Development
Medical Product Development cycle
One size does not fit all unique study management challenges for diagnostic ...
Beyond FDA Compliance Webinar: 5 Hidden Benefits of Your Traceability Matrix
requirements_traceability_matrix_sample
Personalised Medicine in the EU— Evolving Landscape and New HTA Considerations
Devices Sponsor Information Day: 0 - Developments in medical device regulation
The FDA - Mobile, and Fixed Medical Devices Cybersecurity Guidance
Regulatory approval process for invitro diagnostics in us
Poster Presentation
Medical Device Regulation
BioMedical Strategy Medical Devices Workshop Presentations
FDA’s approach to regulation of in vitro diagnostic tests
Systems Engineering in Medical Devices
Ad

Similar to FDA Expectations for Traceability in Device & Diagnostic Design (20)

PPTX
Medical Device Development - Automating Traceability and Documentation
PDF
Know the Global Regulations for Change Control
PDF
Regulatory Considerations In Medical Electronics G Nobis Apr 20 2010
PDF
Ce Marking Ho
PDF
Pharmaceutical Quality by Design: Review of Progress and Challenges
PPTX
Agile Practices Proven in Highly Regulated Environments by Craig Langenfeld
PDF
Appealing to users: developing usable & desirable devices
PDF
WhitepaperMedicalSystems
PDF
Extraordinary design considerations e bacon 2011 08-05
PDF
Will Biosimilars Be A Driver For Innovation
PPTX
The FDA Early Feasibility Study Pilot and the Innovation Pathway
PDF
Need specification v1
PDF
Thinking Beyond Compliance Medical Device Whitepaper
PDF
Consulting Brochure
PDF
Agile in an FDA Regulated Environment
PPTX
Design & Development.pptx
PDF
GMPs for the 21st Century May 2008
PDF
Ensuring reliability in lean new product development part2of2 24
PDF
2010 10 07 Octane Ofi Presentation
PDF
FDA OCRA Product Safety 09
Medical Device Development - Automating Traceability and Documentation
Know the Global Regulations for Change Control
Regulatory Considerations In Medical Electronics G Nobis Apr 20 2010
Ce Marking Ho
Pharmaceutical Quality by Design: Review of Progress and Challenges
Agile Practices Proven in Highly Regulated Environments by Craig Langenfeld
Appealing to users: developing usable & desirable devices
WhitepaperMedicalSystems
Extraordinary design considerations e bacon 2011 08-05
Will Biosimilars Be A Driver For Innovation
The FDA Early Feasibility Study Pilot and the Innovation Pathway
Need specification v1
Thinking Beyond Compliance Medical Device Whitepaper
Consulting Brochure
Agile in an FDA Regulated Environment
Design & Development.pptx
GMPs for the 21st Century May 2008
Ensuring reliability in lean new product development part2of2 24
2010 10 07 Octane Ofi Presentation
FDA OCRA Product Safety 09

More from Seapine Software (20)

PPTX
Breaking the Barriers to Agile Adoption in Safety- and Quality-Critical Envir...
PPTX
Closing the Gap Between Risk and Requirements
PDF
Traceability Trends in Medical Device Development
PPTX
FMEA RCA Risk Management webinar
PDF
Seapine driverless car safety infographic
PDF
Test-Driven Development Reference Card
PDF
Seapine Scrum Reference Card
PDF
The Road to Better Traceability
PDF
2013 State of medical device webinar
PPTX
Sw qual joint webinar deck (5)
PPT
Pragmatic Device Risk Management
PPT
Use the Windshield, Not the Mirror Predictive Metrics that Drive Successful ...
PPT
Leveraging Reusability and Traceability in Medical Device Development
PPTX
Webinar - Maximizing Requirements Value Throughout the Product Lifecycle
PDF
Take the Pain (and Weekends) Out of Test Case Management
PDF
Understanding the Business Case for Agile
PPTX
When Requirements Change
PDF
Agile at Seapine (University of Cincinnati 2011)
PPTX
Monitoring Quality Metrics to Know When to Ship
PPTX
Improving the Development Lifecycle with a Quality-Centric Approach
Breaking the Barriers to Agile Adoption in Safety- and Quality-Critical Envir...
Closing the Gap Between Risk and Requirements
Traceability Trends in Medical Device Development
FMEA RCA Risk Management webinar
Seapine driverless car safety infographic
Test-Driven Development Reference Card
Seapine Scrum Reference Card
The Road to Better Traceability
2013 State of medical device webinar
Sw qual joint webinar deck (5)
Pragmatic Device Risk Management
Use the Windshield, Not the Mirror Predictive Metrics that Drive Successful ...
Leveraging Reusability and Traceability in Medical Device Development
Webinar - Maximizing Requirements Value Throughout the Product Lifecycle
Take the Pain (and Weekends) Out of Test Case Management
Understanding the Business Case for Agile
When Requirements Change
Agile at Seapine (University of Cincinnati 2011)
Monitoring Quality Metrics to Know When to Ship
Improving the Development Lifecycle with a Quality-Centric Approach

Recently uploaded (20)

PPT
Teaching material agriculture food technology
PPTX
A Presentation on Artificial Intelligence
PPTX
Detection-First SIEM: Rule Types, Dashboards, and Threat-Informed Strategy
PDF
Spectral efficient network and resource selection model in 5G networks
PPTX
Understanding_Digital_Forensics_Presentation.pptx
PDF
Electronic commerce courselecture one. Pdf
PPTX
PA Analog/Digital System: The Backbone of Modern Surveillance and Communication
PDF
Bridging biosciences and deep learning for revolutionary discoveries: a compr...
PPTX
MYSQL Presentation for SQL database connectivity
PDF
NewMind AI Weekly Chronicles - August'25 Week I
PDF
Machine learning based COVID-19 study performance prediction
PDF
Blue Purple Modern Animated Computer Science Presentation.pdf.pdf
PDF
Modernizing your data center with Dell and AMD
PDF
Mobile App Security Testing_ A Comprehensive Guide.pdf
PDF
KodekX | Application Modernization Development
PDF
Agricultural_Statistics_at_a_Glance_2022_0.pdf
PDF
cuic standard and advanced reporting.pdf
PDF
Encapsulation_ Review paper, used for researhc scholars
PPTX
VMware vSphere Foundation How to Sell Presentation-Ver1.4-2-14-2024.pptx
PPTX
KOM of Painting work and Equipment Insulation REV00 update 25-dec.pptx
Teaching material agriculture food technology
A Presentation on Artificial Intelligence
Detection-First SIEM: Rule Types, Dashboards, and Threat-Informed Strategy
Spectral efficient network and resource selection model in 5G networks
Understanding_Digital_Forensics_Presentation.pptx
Electronic commerce courselecture one. Pdf
PA Analog/Digital System: The Backbone of Modern Surveillance and Communication
Bridging biosciences and deep learning for revolutionary discoveries: a compr...
MYSQL Presentation for SQL database connectivity
NewMind AI Weekly Chronicles - August'25 Week I
Machine learning based COVID-19 study performance prediction
Blue Purple Modern Animated Computer Science Presentation.pdf.pdf
Modernizing your data center with Dell and AMD
Mobile App Security Testing_ A Comprehensive Guide.pdf
KodekX | Application Modernization Development
Agricultural_Statistics_at_a_Glance_2022_0.pdf
cuic standard and advanced reporting.pdf
Encapsulation_ Review paper, used for researhc scholars
VMware vSphere Foundation How to Sell Presentation-Ver1.4-2-14-2024.pptx
KOM of Painting work and Equipment Insulation REV00 update 25-dec.pptx

FDA Expectations for Traceability in Device & Diagnostic Design

  • 1. FDA Expectations for in Device & Diagnostic Design John Avellanet Cerulean Associates LLC www.CeruleanLLC.com © Cerulean Associates LLC all rights reserved
  • 2. Acknowledgements Tim Ulatowski Dan O’Leary Anita Fauchier Mike Weber Nancy Singer Karen Steinberg Kerry McCarter Akos Bartha Carl Anderson Jackie Cassada John Lincoln Jim Shore Tom Colonna Terry Winchell Chris Szustkiewicz Kim Trautman Larry Nicholson Jan Welch David Elder Annamarie Kempic Doug Throckmorton Jonathan Lee 2
  • 3. Agenda fda requirements practical challenges lean compliance solutions framework This is not legal advice. Information in this presentation draws upon a variety of sources, including published FDA warning letters, personal experiences, interviews and research, all or any of which may or may not have been prepared or conducted by Cerulean Associates LLC. Cerulean Associates LLC does not provide a warranty concerning the accuracy of the information contained in this presentation. The contents of this presentation are intended for general information only and should not be construed as legal advice. Cerulean Associates LLC assumes no liability for actions taken or not taken as a result of the information in this presentation. This presentation is copyrighted 2011 by Cerulean Associates LLC, all rights reserved. 3
  • 4. FDA Requirements traceability regulations recent enforcement examples example inspector questions other requirements 4
  • 5. 21 CFR 820.30 (a) “General. (1) Each manufacturer…shall establish and maintain procedures to control the design of the device in order to ensure that specified design requirements are met.” Translation for FDA inspectors “The purpose of the design control subsystem is to control the design process to assure that devices meet user needs, intended uses, and specified requirements: • Inputs must be documented • Outputs must be documented • Confirm that device outputs are traceable to design inputs” 5
  • 6. 21 CFR 803.18 (b) (1) (i) “…including all documentation of your deliberations and decisionmaking processes used to determine if a device-related death, serious injury, or malfunction was or was not reportable….” Translation for FDA investigators “A firm must demonstrate that it exercised “good faith” in any attempts to obtain required data…. In addition, the Center believes that the parameters of good faith must, at a minimum, comport with the level of risk/nature of the device….” 6
  • 7. 21 CFR 806.20 (b) (4) “Justification for not reporting the correction or removal action to the FDA, which shall contain conclusions and any followups….” Translation for FDA investigators “Verify that non-reported device corrections or removals meet the following criteria based on design controls: • Risk is not increased • Repairs are not unexpected • Part replacement is not earlier than expected” 7
  • 8. Warning Letter Excerpts “ You have failed to establish and maintain a design history file (DHF) to contain and reference the records necessary to demonstrate that the design was developed in accordance with the approved design plan and the requirements of 21 CFR 820.” - warning letter to ASI Medical, 20 April 2011 8
  • 9. Warning Letter Excerpts “ Failure to maintain adequate device history records…. For example, a review of (b)(4) Meniscal Insert DHRs found one DHR contained an inaccurate date.” - warning letter to Advanced Surgical Design & Manufacture, 1 December 2010 9
  • 10. Warning Letter Excerpts “ When requested, design output requirements for the upgrade from Version (b)(4) to Version (b)(4) done by (b)(4)*…could not be provided.” - warning letter to 3CPM Company, 25 March 2010 *This change was done 7 years prior in 2003! 10
  • 11. Example Inspector Questions • were design characteristic acceptance criteria established and documented prior to design approval? • were design outputs verified and documented against design input requirements? • was risk analysis conducted (and documented) on the design prior to its final design approval? • do design outputs draw a distinction between those characteristics essential for functioning of the device versus those non-essential (e.g., aesthetic) characteristics? How are these verified? • review design outputs linkages wherein design outputs for one stage become design inputs for the next – how are these made clear in the documentation? • verify that changes to device design were documented and approved – are there date discrepancies between the actual change and its approval? • how were design changes tested to assess impact to risk or to adjacent design inputs or outputs? How was this documented? • is there a documented traceability analysis or matrix linking product design requirements, design specifications, risks and controls, and tests? 11
  • 12. Other Requirements FDA Guidance Documents • Human Factors Points to Consider for IDE Devices (1996) • Human Factors Implications of the New GMP Rule: Overall Requirements of the New Quality System Regulations (1997) • Design Control Guidance for Medical Device Manufacturers (1997) • Medical Device Use-Safety: Incorporating Human Factors Engineering into Risk Management (2000) • General Principles of Software Validation (2002) • Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (2005) • Radiation Safety Considerations for X-Ray Equipment Designed for Hand- Held Use (2008) • Technical Considerations for Pen, Jet, and Related Injectors Intended for Use with Drugs and Biological Products (2009) • Clinical Investigations of Devices Indicated for the Treatment of Urinary Incontinence (2011) 12
  • 13. Other Requirements GHTF Guidelines • SG1 N:068 Essential Principles of Safety and Performance of Medical Devices (2010 draft) • SG1 N:11 Summary Technical Documentation for Demonstrating Conformity to the Essential Principles of Safety and Performance of Medical Devices (STED) (2008) • SG3 N:15 Implementation of Risk Management Principles and Activities Within a Quality Management System (2005) • SG5 N:2 Clinical Evaluation (2007) • SG5 N:3 Clinical Investigations (2010) • AHWG N:2 Unique Device Identification (UDI) System for Medical Devices (2010 draft) 13
  • 14. Practical Constraints industry confusion costs ownership development lifecycle 14
  • 15. Industry Confusion Design Requirements Issues/Defects/Anomalies Tests Risk Controls Source: Life Sciences Product Development Artifacts Survey White Paper (Seapine Software, 2011) 15
  • 16. Costs tracking, verifying earlier tests, add’l tests, etc.) Cost of Change (incl. documentation, Perception Reality Requirements Early Design Protype Production Clinical Final Production Time (development cycle) Source: Adapted from Scott Ambler, Agile Modeling (2002) 16
  • 17. Ownership Quality Validation / Engineering Testing Design Production 17
  • 18. Development Lifecycle start traceability …? ? ? ? ? Source: Get to Market Now! Turn FDA Compliance into a Competitive Edge (2010, Logos Press) 18
  • 19. Practically Speaking • address records sought by inspectors • capture logic through traceability • need a systematic approach to traceability • cost-effective • generates records • allows easy reporting • maintain clear linkages 19
  • 20. Rationale (Traceability) • relationship to safety • relationship to effectiveness • relationship to risk mitigation • relationship to testing • corrective actions • design changes/fixes 20
  • 21. Lean Compliance Strategy do’s v. don’ts eight-step process lifecycle placement 21
  • 22. Traceability Do’s • Form a cross-functional traceability team • Clearly define accountabilities v. responsibilities in your SOPs • Start traceability efforts when you make the “go/no-go” decision on a project • Ensure that each specific feature  risk/request  tests  production characteristics (suppliers, processes, etc.) • Plan for at least one significant change during product development • Verify progress (and identify gaps) with a mock FDA audit 22
  • 23. Traceability Don’ts • Assume that you can remember dates, times and people involved in design decisions 4, 5, 7+ years ago • Rely only on internal audits by your Quality department to catch missing traceability records • Assume FDA inspector will only go back a few months in design history • Forget that traceability records help you prove safety, efficacy, and compliance • Lose sight of traceability costs – design changes include documentation changes, test changes, test re-work, etc. 23
  • 24. Eight-Step Process design control SOPs 1 product concept sheet 8 2 product specs 7 3 risk controls traceability matrix 6 4 QC “yea/nay” 5 management review audit 24
  • 25. 1 8 2 Design Control SOPs 7 3 6 4 5 25
  • 26. 1 8 2 Product Concept Sheet 7 3 6 4 5 • focus on: • intended use • desired features • distinguishing characteristics • development considerations Source: Get to Market Now! Turn FDA Compliance into a Competitive Edge (2010, Logos Press) 26
  • 27. 1 8 2 Product Specs 7 3 6 4 5 Product Component Requirement seal integrity seal integrity specifications 1. does not allow loss of product or moisture over shelf life 2. retain sterility over shelf life 3. does not allow alteration to finished product over shelf life stopper stopper specifications 1. biocompatible 2. not contain silicon 3. 20mm serum stopper 27
  • 28. 1 8 2 Risk Controls 7 3 6 4 5 Product Specifications Control for seal integrity specifications seal integrity controls 1. does not allow loss of product or USP <1207> moisture over shelf life 2. retain sterility over shelf life stability study (ICH Q6A) 3. does not allow alteration to USP <381> finished product over shelf life USP <87> stopper stopper controls 1. biocompatible USP <88> or USP <381> 2. not contain silicon purchasing control 3. 20mm serum stopper see drawing 28
  • 29. 1 8 2 Traceability Matrix 7 3 6 4 5 • manual (logs, spreadsheets) • automated (database) • clear linkages • systemic reviews • red flag assessments 29
  • 30. 1 8 2 Traceability Matrix (cont’d) 7 3 6 4 5 30
  • 31. 1 8 2 QC “Yea/Nay” 7 3 6 4 5 • are the test results…? • within specification ranges • have a “Plan B” • if results indicate failure, what will you do? • decide potential actions ahead of time based on design and risk 31
  • 32. 1 8 2 Management Review 7 3 6 4 5 • during design transfer • relate traceability data to… • risk assessments • design specifications for materials / suppliers • do not forget: relate to cost • senior management must factor in ROI 32
  • 33. 1 8 2 Audit 7 3 6 4 5 • audit traceability records prior to • IDE submission (if any) • market application submission (510(k)) • any licensing agreements (due diligence) • patent applications / changes • long-term archival • every 2 years otherwise • consider a mock FDA audit and gap analysis 33
  • 34. Development Lifecycle & 8 Steps (until longer of two Formal traceability years after end of device/life of design OR statute of limitations) 2 3 6 7 1 4 5 8 8 Adapted from: Get to Market Now! Turn FDA Compliance into a Competitive Edge (2010, Logos Press) 34
  • 35. Key Takeaway Review core regulatory requirements four practical constraints eight-step lean compliance strategy 35
  • 36. Want More? Traceability Resources http://www.seapine.com/traceability.html Six Exercises To Strengthen Traceability http://downloads.seapine.com/pub/papers/SixExercisesStrengthenTraceability.pdf How To Have A Painless FDA Audit http://downloads.seapine.com/pub/papers/PainlessFDAAudit.pdf The Seapine View Blog http://blogs.seapine.com/ Seapine Life Science Solutions http://www.seapine.com/lifesciences.html 36
  • 37. About Your Presenter John Avellanet solves compliance problems for clients with practical solutions. Winner of the 2009 Best of Business award by the Small Business Commerce Association, Mr. Avellanet has earned international acclaim for his business-savvy, pragmatic compliance advice. His latest book, Get to Market Now! Turn FDA Compliance into a Competitive Edge in the Era of Personalized Medicine, has earned multiple five-star reviews from industry publications, blogs, Amazon.com readers, and former FDA officials. He has a breadth of experience designing, implementing, and being accountable for quality systems and compliance programs for FDA, the ICH, GHTF, and ISO. For more than 15 years, John served as an executive accountable for compliance, records www.Ceruleanllc.com management, and information technology, most recently as a C- level executive for a Fortune 50 combination medical device and www.Get2MarketNow.com biotech subsidiary. In 2006, Mr. Avellanet founded his private FDA compliance consulting and training firm, Cerulean Associates LLC. 37
  • 38. Picture Credits Photos, images and clip art that appear on these slides have been used to enhance this presentation and may NOT be used for commercial or promotional purposes without permission from copyright holders. Do not remove or copy from this presentation. Contact: iStockphoto.com Microsoft Corporation Seapine Software Cerulean Associates LLC 38