Flash chromatography
Prepared by :
SHINDE GANESH SHASHIKANT
PRAVARA RURAL COLLEGE OF
PHARMACY,PRAVARANAGAR
FLASH CHROMATOGRAPHY
 Also called as “medium pressure chromatography”
 “An air pressure driven hybrid of medium and short
column chromatography optimized for rapid separation"
 Popularized by Clark Still of Columbia University
 An alternative to slow and often inefficient gravity-fed
chromatography
Introduction..
 Differs from the conventional technique in 2 ways:
 Slightly smaller silica gel particles (250-400 mesh)
are used, and
 Due to restricted flow of solvent caused by the small
gel particles, pressurized gas (10-15 psi) used to
drive the solvent through the column of stationary
phase
The net result is a rapid “over in a flash” and high
resolution chromatography.
Column vs Flash Chromatography
Traditional Column
chromatography
• Glass columns with silica
gel
• Separation is very slow
(typically many hours)
• End of the run, silica gel
must be removed,
cleaned, dried and re-
packed
• Both time consuming
and hazardous
Flash chromatography
• Pre-packed plastic
cartridges
• Solvent is pumped
through the cartridge
• Much quicker and more
reproducible
• Remaining solvent
flushed out of the column
using pressurized gas
Column chromatography Flash chromatography
Flashchromatography 111201084630-phpapp01
Flashchromatography 111201084630-phpapp01
Modern flash chromatography systems are sold as pre-packed
plastic cartridges, and the solvent is pumped through the
cartridge.
Systems may also be linked with detectors and fraction
collectors providing automation.
The introduction of gradient pumps resulted in quicker
separations and less solvent usage.
Flash chromatography is not expected to provide the resolution
or reproducibility of HPLC; it is a technique that can quickly
improve the purity of samples to an acceptable level.
SELECTION OF STATIONARY PHASE
• Smaller particles (higher mesh values {70-230} ) are used for flash
chromatography; larger particles (lower mesh values {230-400}) are
used for gravity chromatography.
• The amount of silica gel depends on the Rf difference of the
compounds to be separated, and on the amount of sample.
Adsorbents
• Silica: Slightly acidic medium. Best for ordinary compounds, good
separation is achieved.
• Florosil- acidic/neutral
• Alumina: Basic or neutral medium. Can be effective for easy
separations, and purification of amines.
• Reverse phase silica: The most polar compounds elute fastest, the
most non-polars lowest.
Quantity of silica gel required..
 40-63 µm silica gel particles are used
 Amount depends on 2 factors :
– Rf difference of the compounds to be separated
– Amount of sample
 ↑ silica gel -↑ the length of time for chromatography
 For,
– Easier separations, ratios closer to 30 : 1 are effective
– Difficult separations, more silica gel is often required
Selection of solvent system
Solvent system
Compound should have TLC Rf of 0.15 to 0.20 in the solvent
system
 Binary solvent system –
 Polarity can be adjusted
 Rate of elution can be determined
Common solvents used :
dichloromethane/hexane, ether/hexane, hexane/ethyl
acetate, and dichloromethane/methanol
High polarity of solvents increase the rate of elution of all
compounds.
Solvent Systems
• Flash column chromatography is usually carried out with a mixture
of two solvents, with a polar and a nonpolar component.
• 1.Hydrocarbons: pentane, petroleum ether , hexanes.
• 2. Ether and dichloromethane (very similar polarity)
• 3. Ethyl acetate
Two-component Solvent System
• Ether/Petroleum Ether
• Ether/Hexane
• Ether/Pentane
• Ethyl Acetate/Hexane
• Methanol/Dichloromethane
.
Columns
 Glass Columns :-
• A wide range of columns offer maximum flexibility for every
situation.
• Depending on the nature and the quantity of the sample offers a
series of column types which vary in form, size and performance.
.
COLUMN
 Plastic +Glass Column :-
• Plastic+Glass-coated Glass Columns are available for larger amounts
of samples and higher pressure applications on a high safety level.
.
Packing the column
 Glass column / plastic cartridges
 Has either a glass frit or a plug of cotton wool directly above the
stopcock (To prevent the silica gel from escaping from the
column through the stopcock)
 ~1/2 inches layer of clean sand above the plug of glass wool
 Make sure that surface is flat
 Pour in the silica gel using a funnel.(Do this step in hood)
Method
• A solvent is chosen which gives good separation and moves the
desired component to Rf = 0.20 on analytical TLC
• A column of the appropriate diameter is selected and filled with 5-6
in. of dry 40-63 µ silica gel
• Column is solvated
– The column is filled with solvent and pressure is used to rapidly
push all the air from the silica gel
• The sample is applied and the column is refilled with solvent and
eluted at a flow rate of 2 in./min
• Top of the column should never run dry
Procedure for Microscale Flash Column
Chromatography
 In microscale flash chromatography, the column need neither a
pinchclamp or a stopcock at the bottom of the column to control
the flow, nor does it need air-pressure connections at the top of
the column.
 Instead, the solvent flows very slowly through the column by
gravity until you apply air pressure at the top of the column with
an ordinary Pasteur pipet bulb.
(1) Prepare the column.
1. Plug a Pasteur pipet with a small
amount of cotton; use a wood applicator
stick to tamp it down lightly.
2. Take care that you do not use either
too much cotton or pack it too tightly. You
just need enough to prevent the
adsorbent from leaking out.
Add dry silica gel adsorbent, 230-400
mesh -- usually the jar is labeled "for
flash chromatography." One way to fill
the column is to invert it into the jar of
silica gel and scoop it out . . .
. . . then tamp it down before scooping
more out
Another way to fill the column is to pour the
gel into the column using a 10 mL beaker.
(2) Pre-elute the column.
(3) Load the sample onto the silica gel column
Two different methods are used to load the column: the wet
method and the dry method:
Wet loading method
The sample to be purified (or separated into components) is dissolved in a small amount
of solvent, such as hexanes, acetone, or other solvent. This solution is loaded onto the
column.
Dry loading method
(4) Elute the column.
5) Elute the column with the second elution solvent.
If separating a mixture of one or more compounds, at this point
change the eluting solvent to a more polar system, Elution would
proceed as in step (4).
(6) Analyze the fractions.
If the fractions are colored, simply combine like-colored fractions,
although TLC before combination is usually advisable.
If the fractions are not colored, they are analyzed by TLC
(usually). Once the composition of each fraction is known, the
fractions containing the desired compound(s) are combined.
Flashchromatography 111201084630-phpapp01
Instrumentation of Flash Chromatography
Advantages
• Large quantities of the sample can be separated (0.5-2g)
• Fast ( 1o to 15 minutes)
• Cost efficient
• Elaborate equipment and the purchase of expensive
equipment is not necessary
• If high resolution is required, flash chromatography is
carried out before HPLC to avoid contamination of the
expensive plates
Applications
• Purification of various peptides, antibiotics
• Separation of closely related organic compounds
• Purification of closely related drug intermediates
• High speed fractionation of natural products – tocopherols,
alkaloids, lignans, xanthones, stilbenes, flavonoids
• Drug discovery
• Agrochemistry
• Petrochemistry
Conclusion

More Related Content

PPTX
Flash chromatography
PPTX
Flash chromatography
PPTX
counter current chromatography
PPTX
HPTLC PPT.pptx
PPTX
Flash chromatography
PPTX
Countercurrent Chromatography
PPTX
Bioassay.pptx
PPT
high performance thin layer chromatography [HPTLC]
Flash chromatography
Flash chromatography
counter current chromatography
HPTLC PPT.pptx
Flash chromatography
Countercurrent Chromatography
Bioassay.pptx
high performance thin layer chromatography [HPTLC]

What's hot (20)

PPT
Thin Layer Chromatography.ppt
PPTX
RATE TH,PLATE SYS SUTIABILITY.pptx
PPTX
PPT
Theory of chromatographic separations
PPTX
High Performance Thin Layer Chromatography (HPTLC) - Anas Shaikh - 13FET1006
PDF
Counter current chromatography
PDF
1.column chromatography
PPTX
Introduction of chromatography.pptx
PPTX
Gas Chromatography.pptx principle, instrumentation
PPTX
High Performance Thin Layer Chromatography(HPTLC)
PPTX
Tlc (Thin layer chromatography)
PPTX
Instrumentation of column and gas chromatography
PPTX
High performance thin layer chromatography HPTLC.pptx
PPT
Rheology
PDF
High performance liquid chromatography
PPTX
Thin layer chromatography. (TLC)
PPTX
Plate theory of Chromatography
PPTX
Fluorimetery.
PPTX
Sample injection system in hplc
PPTX
Instrumentation Of HPLC
Thin Layer Chromatography.ppt
RATE TH,PLATE SYS SUTIABILITY.pptx
Theory of chromatographic separations
High Performance Thin Layer Chromatography (HPTLC) - Anas Shaikh - 13FET1006
Counter current chromatography
1.column chromatography
Introduction of chromatography.pptx
Gas Chromatography.pptx principle, instrumentation
High Performance Thin Layer Chromatography(HPTLC)
Tlc (Thin layer chromatography)
Instrumentation of column and gas chromatography
High performance thin layer chromatography HPTLC.pptx
Rheology
High performance liquid chromatography
Thin layer chromatography. (TLC)
Plate theory of Chromatography
Fluorimetery.
Sample injection system in hplc
Instrumentation Of HPLC
Ad

Similar to Flashchromatography 111201084630-phpapp01 (20)

PPTX
Flash chromatography
PPTX
flash chromatography.pptx
PPTX
Sna chromatography column chromatography
PPTX
Flash chromatography is also known as medium pressure chromatography
PPTX
Thin layer chromatography ( Horizontal)
PPTX
Chromatography
PPT
Flash Chromatography by GBNigade
PDF
Flash Chromatography.pdf
PPTX
Flash cromatography
PPTX
Flash and chiral chromatography
PPTX
Flash and chiral chromatography
PPTX
PHM 316 column and flash chromatography.pptx
PDF
Thin Layer Chromatography for Biotechnology & Botany Sem-2 Students
PPTX
Thin layer chromatography- Advances of Paper Chromatography
PPTX
Silica Gel | Aluminium Oxide Column chroamtography
PPT
Column Chromatography ppt
PPTX
6. columnchromatography jntu pharmacy
DOCX
assignment on column chromatography
Flash chromatography
flash chromatography.pptx
Sna chromatography column chromatography
Flash chromatography is also known as medium pressure chromatography
Thin layer chromatography ( Horizontal)
Chromatography
Flash Chromatography by GBNigade
Flash Chromatography.pdf
Flash cromatography
Flash and chiral chromatography
Flash and chiral chromatography
PHM 316 column and flash chromatography.pptx
Thin Layer Chromatography for Biotechnology & Botany Sem-2 Students
Thin layer chromatography- Advances of Paper Chromatography
Silica Gel | Aluminium Oxide Column chroamtography
Column Chromatography ppt
6. columnchromatography jntu pharmacy
assignment on column chromatography
Ad

More from Ganesh Shinde (20)

PPTX
Gout.pptx
PPTX
TYPOID.pptx
PPTX
Tuberculosis.pptx
PPTX
Powders.pptx
PPTX
posology.pptx
PPTX
Nmr interpration
PPTX
2 d nmr
PPTX
Pharmacognosy orientation ppt
PPTX
Nuclear magnetic resonanace
PPTX
PPTX
IR Spectra Interpretation
PPTX
Ir spectroscopy
PPTX
Gas chromatography
PPTX
Gas chromatography
DOCX
Column chromatography
DOCX
THIN LAYER CHROMATOGRAPHY
DOCX
Chrometography
DOCX
Notes on mass spectroscopy
PPTX
Mass specctroscopy and interpretation
PPTX
Ir mass interpration
Gout.pptx
TYPOID.pptx
Tuberculosis.pptx
Powders.pptx
posology.pptx
Nmr interpration
2 d nmr
Pharmacognosy orientation ppt
Nuclear magnetic resonanace
IR Spectra Interpretation
Ir spectroscopy
Gas chromatography
Gas chromatography
Column chromatography
THIN LAYER CHROMATOGRAPHY
Chrometography
Notes on mass spectroscopy
Mass specctroscopy and interpretation
Ir mass interpration

Recently uploaded (20)

PDF
grade 11-chemistry_fetena_net_5883.pdf teacher guide for all student
PPTX
Final Presentation General Medicine 03-08-2024.pptx
PDF
Classroom Observation Tools for Teachers
PDF
3rd Neelam Sanjeevareddy Memorial Lecture.pdf
PPTX
Institutional Correction lecture only . . .
PDF
Complications of Minimal Access Surgery at WLH
PDF
OBE - B.A.(HON'S) IN INTERIOR ARCHITECTURE -Ar.MOHIUDDIN.pdf
PPTX
human mycosis Human fungal infections are called human mycosis..pptx
PDF
A GUIDE TO GENETICS FOR UNDERGRADUATE MEDICAL STUDENTS
PPTX
Final Presentation General Medicine 03-08-2024.pptx
PDF
O7-L3 Supply Chain Operations - ICLT Program
PPTX
Lesson notes of climatology university.
PDF
Saundersa Comprehensive Review for the NCLEX-RN Examination.pdf
PPTX
Presentation on HIE in infants and its manifestations
PDF
The Lost Whites of Pakistan by Jahanzaib Mughal.pdf
PDF
Abdominal Access Techniques with Prof. Dr. R K Mishra
PPTX
school management -TNTEU- B.Ed., Semester II Unit 1.pptx
PDF
O5-L3 Freight Transport Ops (International) V1.pdf
PPTX
GDM (1) (1).pptx small presentation for students
PDF
GENETICS IN BIOLOGY IN SECONDARY LEVEL FORM 3
grade 11-chemistry_fetena_net_5883.pdf teacher guide for all student
Final Presentation General Medicine 03-08-2024.pptx
Classroom Observation Tools for Teachers
3rd Neelam Sanjeevareddy Memorial Lecture.pdf
Institutional Correction lecture only . . .
Complications of Minimal Access Surgery at WLH
OBE - B.A.(HON'S) IN INTERIOR ARCHITECTURE -Ar.MOHIUDDIN.pdf
human mycosis Human fungal infections are called human mycosis..pptx
A GUIDE TO GENETICS FOR UNDERGRADUATE MEDICAL STUDENTS
Final Presentation General Medicine 03-08-2024.pptx
O7-L3 Supply Chain Operations - ICLT Program
Lesson notes of climatology university.
Saundersa Comprehensive Review for the NCLEX-RN Examination.pdf
Presentation on HIE in infants and its manifestations
The Lost Whites of Pakistan by Jahanzaib Mughal.pdf
Abdominal Access Techniques with Prof. Dr. R K Mishra
school management -TNTEU- B.Ed., Semester II Unit 1.pptx
O5-L3 Freight Transport Ops (International) V1.pdf
GDM (1) (1).pptx small presentation for students
GENETICS IN BIOLOGY IN SECONDARY LEVEL FORM 3

Flashchromatography 111201084630-phpapp01

  • 1. Flash chromatography Prepared by : SHINDE GANESH SHASHIKANT PRAVARA RURAL COLLEGE OF PHARMACY,PRAVARANAGAR
  • 2. FLASH CHROMATOGRAPHY  Also called as “medium pressure chromatography”  “An air pressure driven hybrid of medium and short column chromatography optimized for rapid separation"  Popularized by Clark Still of Columbia University  An alternative to slow and often inefficient gravity-fed chromatography
  • 3. Introduction..  Differs from the conventional technique in 2 ways:  Slightly smaller silica gel particles (250-400 mesh) are used, and  Due to restricted flow of solvent caused by the small gel particles, pressurized gas (10-15 psi) used to drive the solvent through the column of stationary phase The net result is a rapid “over in a flash” and high resolution chromatography.
  • 4. Column vs Flash Chromatography Traditional Column chromatography • Glass columns with silica gel • Separation is very slow (typically many hours) • End of the run, silica gel must be removed, cleaned, dried and re- packed • Both time consuming and hazardous Flash chromatography • Pre-packed plastic cartridges • Solvent is pumped through the cartridge • Much quicker and more reproducible • Remaining solvent flushed out of the column using pressurized gas
  • 8. Modern flash chromatography systems are sold as pre-packed plastic cartridges, and the solvent is pumped through the cartridge. Systems may also be linked with detectors and fraction collectors providing automation. The introduction of gradient pumps resulted in quicker separations and less solvent usage. Flash chromatography is not expected to provide the resolution or reproducibility of HPLC; it is a technique that can quickly improve the purity of samples to an acceptable level.
  • 9. SELECTION OF STATIONARY PHASE • Smaller particles (higher mesh values {70-230} ) are used for flash chromatography; larger particles (lower mesh values {230-400}) are used for gravity chromatography. • The amount of silica gel depends on the Rf difference of the compounds to be separated, and on the amount of sample.
  • 10. Adsorbents • Silica: Slightly acidic medium. Best for ordinary compounds, good separation is achieved. • Florosil- acidic/neutral • Alumina: Basic or neutral medium. Can be effective for easy separations, and purification of amines. • Reverse phase silica: The most polar compounds elute fastest, the most non-polars lowest.
  • 11. Quantity of silica gel required..  40-63 µm silica gel particles are used  Amount depends on 2 factors : – Rf difference of the compounds to be separated – Amount of sample  ↑ silica gel -↑ the length of time for chromatography  For, – Easier separations, ratios closer to 30 : 1 are effective – Difficult separations, more silica gel is often required
  • 12. Selection of solvent system Solvent system Compound should have TLC Rf of 0.15 to 0.20 in the solvent system  Binary solvent system –  Polarity can be adjusted  Rate of elution can be determined Common solvents used : dichloromethane/hexane, ether/hexane, hexane/ethyl acetate, and dichloromethane/methanol High polarity of solvents increase the rate of elution of all compounds.
  • 13. Solvent Systems • Flash column chromatography is usually carried out with a mixture of two solvents, with a polar and a nonpolar component. • 1.Hydrocarbons: pentane, petroleum ether , hexanes. • 2. Ether and dichloromethane (very similar polarity) • 3. Ethyl acetate
  • 14. Two-component Solvent System • Ether/Petroleum Ether • Ether/Hexane • Ether/Pentane • Ethyl Acetate/Hexane • Methanol/Dichloromethane .
  • 15. Columns  Glass Columns :- • A wide range of columns offer maximum flexibility for every situation. • Depending on the nature and the quantity of the sample offers a series of column types which vary in form, size and performance. .
  • 16. COLUMN  Plastic +Glass Column :- • Plastic+Glass-coated Glass Columns are available for larger amounts of samples and higher pressure applications on a high safety level. .
  • 17. Packing the column  Glass column / plastic cartridges  Has either a glass frit or a plug of cotton wool directly above the stopcock (To prevent the silica gel from escaping from the column through the stopcock)  ~1/2 inches layer of clean sand above the plug of glass wool  Make sure that surface is flat  Pour in the silica gel using a funnel.(Do this step in hood)
  • 18. Method • A solvent is chosen which gives good separation and moves the desired component to Rf = 0.20 on analytical TLC • A column of the appropriate diameter is selected and filled with 5-6 in. of dry 40-63 µ silica gel • Column is solvated – The column is filled with solvent and pressure is used to rapidly push all the air from the silica gel • The sample is applied and the column is refilled with solvent and eluted at a flow rate of 2 in./min • Top of the column should never run dry
  • 19. Procedure for Microscale Flash Column Chromatography  In microscale flash chromatography, the column need neither a pinchclamp or a stopcock at the bottom of the column to control the flow, nor does it need air-pressure connections at the top of the column.  Instead, the solvent flows very slowly through the column by gravity until you apply air pressure at the top of the column with an ordinary Pasteur pipet bulb.
  • 20. (1) Prepare the column. 1. Plug a Pasteur pipet with a small amount of cotton; use a wood applicator stick to tamp it down lightly. 2. Take care that you do not use either too much cotton or pack it too tightly. You just need enough to prevent the adsorbent from leaking out. Add dry silica gel adsorbent, 230-400 mesh -- usually the jar is labeled "for flash chromatography." One way to fill the column is to invert it into the jar of silica gel and scoop it out . . .
  • 21. . . . then tamp it down before scooping more out Another way to fill the column is to pour the gel into the column using a 10 mL beaker.
  • 22. (2) Pre-elute the column.
  • 23. (3) Load the sample onto the silica gel column Two different methods are used to load the column: the wet method and the dry method: Wet loading method The sample to be purified (or separated into components) is dissolved in a small amount of solvent, such as hexanes, acetone, or other solvent. This solution is loaded onto the column.
  • 25. (4) Elute the column.
  • 26. 5) Elute the column with the second elution solvent. If separating a mixture of one or more compounds, at this point change the eluting solvent to a more polar system, Elution would proceed as in step (4). (6) Analyze the fractions. If the fractions are colored, simply combine like-colored fractions, although TLC before combination is usually advisable. If the fractions are not colored, they are analyzed by TLC (usually). Once the composition of each fraction is known, the fractions containing the desired compound(s) are combined.
  • 28. Instrumentation of Flash Chromatography
  • 29. Advantages • Large quantities of the sample can be separated (0.5-2g) • Fast ( 1o to 15 minutes) • Cost efficient • Elaborate equipment and the purchase of expensive equipment is not necessary • If high resolution is required, flash chromatography is carried out before HPLC to avoid contamination of the expensive plates
  • 30. Applications • Purification of various peptides, antibiotics • Separation of closely related organic compounds • Purification of closely related drug intermediates • High speed fractionation of natural products – tocopherols, alkaloids, lignans, xanthones, stilbenes, flavonoids • Drug discovery • Agrochemistry • Petrochemistry