HIGH PERFORMANCE LIQUID
CHROMATOGRAPHY
(HPLC)
A PRACTICAL GUIDE FOR
THE BEGINNER USERS
A. Prof. Sherif M. Taha
Tel: 01004724944
sherif2taha@gmail.com
‫ميحرلا نمحرلا هللا مسب‬
INTRODUCTION
Chromatography:
Is the separation and detection of certain compounds in a sample (Food, water, blood, urine,
chemicals, …).
based on their different chemical-physical properties
based on different physical interactions between two immiscible phases
Such separation may be accomplished in a liquid-liquid extraction
system or by its transfer through an immobilized stationary phase by a
carrier gas (GC) or a mobile phase (LC).
INTRODUCTION
Mikhail Semenovich Tswett
https://bitesizebio.com/29947/basics-chromatography-column/
THE BASIC COMPONENTS OF
HPLC SYSTEM
MOBILE PHASE
COLUMN
DETECTOR
Sample loading with the mobile phase
(Pump role)
A
C
D
B
B
A
D C
Mobile phase
Stationary phase, column
Detector
At HPLC, The sample is partitioning between
the mobile phase and the stationary phase
(not just being carried by a carrier gas, GC).
THE BASIC COMPONENTS OF
HPLC SYSTEM
Pilar Campíns-Falcó; Rosa Herráez-Hernández; Pascual Serra-Mora, Liquid Chromatography-Instrumentation
MOBILE PHASE
COLUMN
DETECTOR
THE BASIC COMPONENTS OF
HPLC SYSTEM
HPLC separation
Adsorption
Polar St phase
Normal phase
Partition
Non polar St phase,
Reversed phase
size-exclusion Ion Exchange Chiral
THE BASIC COMPONENTS OF
HPLC SYSTEM
B
B
A
C
A
D
x
H
x
x
x
H
Adsorption partition size-exclusion Ion Exchange Chiral
C
E
B
A
A- Analytes are in an adsorption interaction with functional gps on surface of stationary phase
B-Analytes partitions between the mobile phase and the stationary phase depending on its solubility
B-Strong ads, late eluted
A- Weak ads, early eluted
PARTITION
CONSTANT;
(KOW )
KOW =
[𝑆𝑜𝑙𝑢𝑡𝑒 𝑛𝑜𝑡 𝑖𝑜𝑛𝑖𝑠𝑒𝑑 𝑖𝑛 𝑜𝑐𝑡𝑎𝑛𝑜𝑙]
[𝑆𝑜𝑙𝑢𝑡𝑒 𝑛𝑜𝑡 𝑖𝑜𝑛𝑖𝑠𝑒𝑑 𝑖𝑛 𝑤𝑎𝑡𝑒𝑟]
http://dx.doi.org/10.1016/B978-0-12-385013-3.00009-4
STATIONARY
PHASE;
(RP)
C8 RP
https://www.chromacademy.com/lms/sco4/Theory_Of_HPLC_Column_Chemistry.pdf
C18 and C 8 are the commonly
used ones in HPLC analyses.
Polar
Stationary phase
Non polar
Stationary phase
Analyte Highly polar Moderately to non
polar
Mobile phase for
loading
Low polar solvent High polar solvent
Mobile phase for
eluting
High polar solvent Low polar solvent STATIONARY
PHASE;
(RP // NP)
• Improvement of the obtained selectivity by using a specific
separation condition (NP or RP) was carried out firstly by
selecting proper mobile phase conditions and then changing the
stationary phase.
• It is also preferred to use previously published conditions
(especially on the used column) and then modify it according to
your own need.
COLUMN
SPECIFICATION
C18, 250 mm, 4,6 mm, 5 µm, 300 A
Bonded phase
Particle size ,
Smaller PZ introduce
• higher surface area,
• higher peak resolution
• Shorter run time
• But, higher back
pressure too
Particle ‘s Pores size
• Below 60 A
• 60-150 A
• 150-300 A (proteins)
Depending on target
analytes and matrix
Inter diameter
Smaller ID
• Higher peak
resolution
• But, higher back
pressure too
Length
Increasing L
• Higher peak
resolution
• But, higher back
pressure too
Injection volume, Vinj;
Vinj < 0.14 𝐝2
𝐋 x dp
d, L; internal diameter and length of the used column
dp diameter of solid phase particles
High Vinj and analyst Conc.
May cause fronting of un-retained molecules
COLUMN
SPECIFICATION
C18, 5 µm, 150 mm × 4.6 mm column
C18, 3 µm, 150 mm × 4.6 mm column
C18, 2.7 µm, 100 mm × 4.6 mm
HPLC-FLD chromatograms of aflatoxins
COLUMN
SPECIFICATION
Pilar Campíns-Falcó; Rosa Herráez-Hernández; Pascual Serra-Mora, Liquid Chromatography-Instrumentation
SAMPLE FILTERING
AND COLUMN
CLOGGING
https://www.sigmaaldrich.com/content/dam/sigmaaldrich/pro
duct8/135/p000543.tif/_jcr_content/renditions/p000543-
medium.jpg
https://encrypted-tbn0.gstatic.com/images?q=tbn:ANd9GcQVL_oKnG-
Wr4kJeF2xVFsIjmlToXWu8g0mJ6ChKcHczDpHUaJ05Q
Sample particle size ≤ pores PZ in sintered frit <Stationary phase PZ
e.g. Acrodisck of 0.2µm< frit of 0.5µm< particles size inside the column 3µm
MOBILE PHASE, KOW
THE ELUTION STRENGTH APPLING RP-HPLC ;
• HEXANE(3.13)>TERT-BUTANOL(0.54),ISOPROPANOL(0.25), ACETONE (0.11),
ETHANOL (-0.16)ACETONITRILE (-0.34), METHANOL(-0.77), WATER (-1.38).
first loading of sample
components in RP HPLC.
changing between two immiscible
solvents. It is also used for long
storage of the used columns
Kow Polarity
water and a miscible organic solvent are commonly used mobile phase
mixtures in HPLC since they give a new Kow that rapid the chromatographic
separation run.
Log Kow (A+B)= 1 − 𝑥 𝐿𝑜𝑔 𝐾𝑜𝑤 𝐴 + 𝑋 𝐿𝑜𝑔 𝐾𝑜𝑤 𝐵
A& B; used solvents.
MOBILE PHASE, KOW
Kow Polarity
• Acetone has a strong eluting properties in
RP-HPLC but it has a high UV cut-off
value.
• Methylene chloride, chloroform have
medium polarity but in RP HPLC it is
immiscible with water.
• Acetonitrile has a very low UV cut-off,
THE BASIC COMPONENTS OF
HPLC SYSTEM
 Gradient elution results in; minimizing the run time, good shape for eluted peaks, and
cleaning the used column for each run.
includes three basic steps:
 Short isocratic start (Solution A, lower elution)
 Gradient program (gradually increase solution B, higher elution).
 Short isocratic for cleaning the used column ( Highest percent of solution B).
 Return to the initial conditions (column conditioning, very critical step).
MOBILE PHASE,
GRADIENT ELUTION
A (H2O:MeOH 5%, pH 4.6). B (MeOH)
Time
10 % B
0.01-5.00
% B
70
5.01-7.00
10 % B
7.01-10.00
BENZOIC AND SORBIC ACID
DETERMINATION IN JUICE, MILK
PRODUCTS,…
BUFFER, HPLC MOBILE PHASE
 In RP- HPLC Solvent A is mainly water as it:
 The highest polar solvent (weakest eluent), suitable for sample loading.
 Buffers can be easily prepared in water at different concentrations.
 The purity of water for HPLC is very important, especially when using MS/ MS.
 It is preferred to use a percent (About 10 %) of a suitable organic solvent in water
(Solvent A);
 To prevent algae formation.
 Enhance the evaporation in ESI ionization unit (LC-MS/MS).
BUFFER, HPLC MOBILE PHASE
 For separation of acidic or basic compounds, a buffer should be
used in solvent A to keep the ionizable compounds in neutral
form as possible. Where pH of the mobile phase A (using
buffer) should be acidic for basic compounds and the reverse.
 Changing pH by 2 units (lower or higher than its pKa) create a
reverse partitioning condition for this compound.
pH= pKa+ Log (
[𝐴−]
[𝐻𝐴]
)
At pH= 2,
-2.2= Log (
[𝐴−]
[𝐻𝐴]
), [HA]=166 [𝐴−
]
http://dx.doi.org/10.1016/B978-0-12-803684-6.00004-4
BUFFER, HPLC MOBILE PHASE
pH= pKa+ Log (
[𝐴−]
[𝐻𝐴]
)
http://dx.doi.org/10.1016/B978-0-12-803684-6.00004-4
Hexane
MeOH (basic mix)
MeOH (Acidic mix)
Ricinoic acid (castor Oil)
BUFFER, HPLC MOBILE PHASE
 A buffer is
a solution of a weak acid (HA) and its conjugate base (𝐴−
) &
a solution of a weak base (B) and its conjugated acid (𝐵𝐻+
)
 Such buffer system has the capability to resistance changing in pH upon the addition of
small amounts from a strong acid or base.
pH= pKa+ Log (
[𝐴−]
[𝐻𝐴]
)
Cbuf = [𝐴−] + [𝐻𝐴]
Buffer
(pH & Concentration)
BUFFER, HPLC MOBILE PHASE
 Buffer capacity, β:
The number of added moles (dn) of a strong acid or a base that change the pH of one-liter
buffer solution.
β=
𝑑 𝑛
𝑝𝐻
Since, the addition of n moles from a base (NaOH) leads to the formation of [𝐴−
Na] or CNaA
β=[𝐴−
b]
𝑝𝐻
http://dx.doi.org/10.1016/B978-0-12-803684-6.00004-4
https://www.kisspng.com/png-bouncer-clip-
art-bodyguard-security-guard-royalty-
5953930/
β
BUFFER, HPLC MOBILE PHASE
https://www.kisspng.com/png-bouncer-clip-
art-bodyguard-security-guard-royalty-
5953930/
β
Fixed
http://dx.doi.org/10.1016/B978-0-12-803684-6.00004-4
pH= pKa+ Log (
[𝐴−]
[𝐻𝐴]
)
Cbuf = [𝐴−
] + [𝐻𝐴]
BUFFER, HPLC MOBILE PHASE
• BUFFER CONCENTRATION IN HPLC UV IS USUALLY MADE WITH A CONCENTRATION OF 10-
200 MM. (LOWER CONCENTRATIONS FOR LC MSMS <50MM, VOLATILE SALTS ARE MORE
FAVORABLE).
• THE SOLUBILITY OF INORGANIC SALTS DEPENDS MAINLY ON THE NATURE OF THE CATION,
AND THEIR SOLUBILITY TREND IN ORGANIC SOLVENTS FOLLOWS (THE SAME AS IN
WATER): NH4+ >𝐾+>𝑁𝑎+.
• A HIGHER CONTENT OF ORGANIC PHASE SHOULD BE AVOIDED NOT TO PRECIPITATE THE
BUFFER SALTS.
_____________________________________________________
• FOR PREPARING A BUFFER AT PH 4.5 YOU SHOULD USE:
A WEAK ACID OF PKA CLOSE TO THE REQUIRED PH (……ACID PKA =….)
THE CONJUGATE BASE FOR THIS ACID WILL BE ITS (NA, K, AMMONIUM) SALT
WHAT IS THE TOTAL BUFFER CONCENTRATION AND HOW TO PREPARE ?
BUFFER, HPLC MOBILE PHASE
BUFFER, HPLC MOBILE PHASE
http://dx.doi.org/10.1016/B978-0-12-811732-3.00006-6
Volatile buffers that can be used for LC MS
HPLC DETECTORS
HPLC
UV-Vis
MS
Others
Florescence
SOLVENT OF THE SAMPLE
The solvent of the sample should be with;
- Same polarity as the mobile phase (loading solvent, A)
or weaker (increase solvent compressing, enhance the
retention of solutes).
- pH close to that of the mobile phase (loading solvent, A)
These points are more critical, especially for higher injection volume
and for ionizable solutes solutes
Zorbax Eclipse XDB-
C18, 150 mm, 4.6 mm, 5 µm
with a mobile phase:
35% ACN and 65% H2O (0.2% H3PO4);
flow-rate: 2 mL/min.
A CHROMATOGRAM AND A MASS SPECTRUM
COLUMN EFFICIENCY
Jesús Lozano-Sánchez; Isabel Borrás-Linares; Agnes Sass-Kiss; Antonio Segura-Carretero, Chapter 13 - Chromatographic
Technique: High-Performance Liquid Chromatography (HPLC)
Dead time, Void time
T0 = Time at which mobile phase pass through the chromatographic column
Depending on column length and flow rate
Actual Rt (X)= Obtained Rt (X)+ t0
COLUMN EFFICIENCY
Plate Number (N)
N= (𝑅𝑡/ )2
N= 16(𝑅𝑡/Wb)2
N= 5.45 (𝑅𝑡/W 0.5)2
Wb= 4*x Standard deviation ( )
Selectivity
=
𝑅𝑡 (𝑥)
𝑅𝑡 (𝑦)
Resolution
R =
2 [𝑅𝑡 𝑥 −𝑅𝑡 𝑦 ]
(𝑊𝑏 𝑥 +𝑊𝑏 (𝑦)
R=
Δ𝑅𝑡
Column efficiency:
 High Plate N per unit length (N/L)
or
Height of the theoretical plate H= 𝐿/𝑁≃ 2dp
dp diameter of solid phase particles
 Selectivity (Separation Factor)
A measurement for the separation
of two compounds X and Y
 Resolution
Separation of apexes of two adjacent peaks
A. Prof. Sherif M. Taha
Tel: 01004724944
sherif2taha@gmail.com
Thank You

More Related Content

PDF
Mass spectrometry, Ion motion in the commonly used mass analyzer.pdf
PDF
Gas chromatography, an introduction.pdf
PDF
Commonly used ionizations techniques in mass spectrometry.pdf
PPT
Types chrom
PPTX
Gas chromotography (Histroy, Importance, Classification)
DOCX
Chromatography( HPLC, ion exchange chromatography, Gas chromatography)
PPT
Chromatography - General and Theory.ppt
PDF
Karl Fischer Titration (KF titration)
Mass spectrometry, Ion motion in the commonly used mass analyzer.pdf
Gas chromatography, an introduction.pdf
Commonly used ionizations techniques in mass spectrometry.pdf
Types chrom
Gas chromotography (Histroy, Importance, Classification)
Chromatography( HPLC, ion exchange chromatography, Gas chromatography)
Chromatography - General and Theory.ppt
Karl Fischer Titration (KF titration)

What's hot (20)

PPTX
CHROMATOGRAPHY
PPTX
Chiral hplc intro
PPTX
Chromatography & Band Broadening
PPT
Solvent extraction
PDF
High performance liquid chromatography
PPTX
orthogonal hplc methods
PPTX
Chromatography
PPTX
Hplc introduction
PPTX
Gas Chromatography
PPTX
PPT
2. liquid chromatography
PDF
chiral chromatography
PPT
Good chromatographic practices
PDF
Introduction to hplc
PPTX
Dry column vacuum chromatography (dcvc)
PPTX
Chiral chromatography
PPTX
Principles in chromatography
 
PPTX
Chromatography
PDF
Gas Chromatography
PPTX
Thin layer chromatography
CHROMATOGRAPHY
Chiral hplc intro
Chromatography & Band Broadening
Solvent extraction
High performance liquid chromatography
orthogonal hplc methods
Chromatography
Hplc introduction
Gas Chromatography
2. liquid chromatography
chiral chromatography
Good chromatographic practices
Introduction to hplc
Dry column vacuum chromatography (dcvc)
Chiral chromatography
Principles in chromatography
 
Chromatography
Gas Chromatography
Thin layer chromatography
Ad

Similar to HPLC_A practical guide for the beginner users.pdf (20)

PDF
HPLC method development by DrA.S Charan
PDF
PPTX
Introduction to RP-HPLC
PPT
High Pressure Liquid Chromatography presenataion
PPTX
Hplc seminar
PPTX
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY
PPT
HPLC-NIAB BS chemistry, University Of Agriculture Faisalabad.ppt
PPTX
HPLC
PPTX
High Performance Liquid Chromatography
PPTX
PPTX
HPLC-(High Performance liquid Chromatography)Lecture-3.pptx
PPT
Hplc presentation for class
PPTX
Instrumentation Of HPLC
DOCX
hplc
PPTX
High Performance Liquid Chromatography.pptx
PPT
Lect. 3 intro hplc
PDF
Good Chromatographic Practices
PPTX
Demostration of HPLC Chromatography
PPTX
HPLC.pptx
HPLC method development by DrA.S Charan
Introduction to RP-HPLC
High Pressure Liquid Chromatography presenataion
Hplc seminar
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY
HPLC-NIAB BS chemistry, University Of Agriculture Faisalabad.ppt
HPLC
High Performance Liquid Chromatography
HPLC-(High Performance liquid Chromatography)Lecture-3.pptx
Hplc presentation for class
Instrumentation Of HPLC
hplc
High Performance Liquid Chromatography.pptx
Lect. 3 intro hplc
Good Chromatographic Practices
Demostration of HPLC Chromatography
HPLC.pptx
Ad

Recently uploaded (20)

PPTX
Final Presentation General Medicine 03-08-2024.pptx
PPTX
Lesson notes of climatology university.
PDF
ChatGPT for Dummies - Pam Baker Ccesa007.pdf
PDF
A systematic review of self-coping strategies used by university students to ...
PDF
What if we spent less time fighting change, and more time building what’s rig...
PPTX
school management -TNTEU- B.Ed., Semester II Unit 1.pptx
PDF
The Lost Whites of Pakistan by Jahanzaib Mughal.pdf
PDF
Updated Idioms and Phrasal Verbs in English subject
PPTX
202450812 BayCHI UCSC-SV 20250812 v17.pptx
PPTX
Final Presentation General Medicine 03-08-2024.pptx
PDF
grade 11-chemistry_fetena_net_5883.pdf teacher guide for all student
PDF
LDMMIA Reiki Yoga Finals Review Spring Summer
PDF
RTP_AR_KS1_Tutor's Guide_English [FOR REPRODUCTION].pdf
PDF
Complications of Minimal Access Surgery at WLH
PDF
Module 4: Burden of Disease Tutorial Slides S2 2025
PDF
Paper A Mock Exam 9_ Attempt review.pdf.
PDF
Anesthesia in Laparoscopic Surgery in India
PDF
Yogi Goddess Pres Conference Studio Updates
PDF
RMMM.pdf make it easy to upload and study
PDF
Chinmaya Tiranga quiz Grand Finale.pdf
Final Presentation General Medicine 03-08-2024.pptx
Lesson notes of climatology university.
ChatGPT for Dummies - Pam Baker Ccesa007.pdf
A systematic review of self-coping strategies used by university students to ...
What if we spent less time fighting change, and more time building what’s rig...
school management -TNTEU- B.Ed., Semester II Unit 1.pptx
The Lost Whites of Pakistan by Jahanzaib Mughal.pdf
Updated Idioms and Phrasal Verbs in English subject
202450812 BayCHI UCSC-SV 20250812 v17.pptx
Final Presentation General Medicine 03-08-2024.pptx
grade 11-chemistry_fetena_net_5883.pdf teacher guide for all student
LDMMIA Reiki Yoga Finals Review Spring Summer
RTP_AR_KS1_Tutor's Guide_English [FOR REPRODUCTION].pdf
Complications of Minimal Access Surgery at WLH
Module 4: Burden of Disease Tutorial Slides S2 2025
Paper A Mock Exam 9_ Attempt review.pdf.
Anesthesia in Laparoscopic Surgery in India
Yogi Goddess Pres Conference Studio Updates
RMMM.pdf make it easy to upload and study
Chinmaya Tiranga quiz Grand Finale.pdf

HPLC_A practical guide for the beginner users.pdf

  • 1. HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC) A PRACTICAL GUIDE FOR THE BEGINNER USERS A. Prof. Sherif M. Taha Tel: 01004724944 sherif2taha@gmail.com ‫ميحرلا نمحرلا هللا مسب‬
  • 2. INTRODUCTION Chromatography: Is the separation and detection of certain compounds in a sample (Food, water, blood, urine, chemicals, …). based on their different chemical-physical properties based on different physical interactions between two immiscible phases Such separation may be accomplished in a liquid-liquid extraction system or by its transfer through an immobilized stationary phase by a carrier gas (GC) or a mobile phase (LC).
  • 4. THE BASIC COMPONENTS OF HPLC SYSTEM MOBILE PHASE COLUMN DETECTOR Sample loading with the mobile phase (Pump role) A C D B B A D C Mobile phase Stationary phase, column Detector At HPLC, The sample is partitioning between the mobile phase and the stationary phase (not just being carried by a carrier gas, GC).
  • 5. THE BASIC COMPONENTS OF HPLC SYSTEM Pilar Campíns-Falcó; Rosa Herráez-Hernández; Pascual Serra-Mora, Liquid Chromatography-Instrumentation MOBILE PHASE COLUMN DETECTOR
  • 6. THE BASIC COMPONENTS OF HPLC SYSTEM HPLC separation Adsorption Polar St phase Normal phase Partition Non polar St phase, Reversed phase size-exclusion Ion Exchange Chiral
  • 7. THE BASIC COMPONENTS OF HPLC SYSTEM B B A C A D x H x x x H Adsorption partition size-exclusion Ion Exchange Chiral C E B A A- Analytes are in an adsorption interaction with functional gps on surface of stationary phase B-Analytes partitions between the mobile phase and the stationary phase depending on its solubility B-Strong ads, late eluted A- Weak ads, early eluted
  • 8. PARTITION CONSTANT; (KOW ) KOW = [𝑆𝑜𝑙𝑢𝑡𝑒 𝑛𝑜𝑡 𝑖𝑜𝑛𝑖𝑠𝑒𝑑 𝑖𝑛 𝑜𝑐𝑡𝑎𝑛𝑜𝑙] [𝑆𝑜𝑙𝑢𝑡𝑒 𝑛𝑜𝑡 𝑖𝑜𝑛𝑖𝑠𝑒𝑑 𝑖𝑛 𝑤𝑎𝑡𝑒𝑟] http://dx.doi.org/10.1016/B978-0-12-385013-3.00009-4
  • 10. Polar Stationary phase Non polar Stationary phase Analyte Highly polar Moderately to non polar Mobile phase for loading Low polar solvent High polar solvent Mobile phase for eluting High polar solvent Low polar solvent STATIONARY PHASE; (RP // NP) • Improvement of the obtained selectivity by using a specific separation condition (NP or RP) was carried out firstly by selecting proper mobile phase conditions and then changing the stationary phase. • It is also preferred to use previously published conditions (especially on the used column) and then modify it according to your own need.
  • 11. COLUMN SPECIFICATION C18, 250 mm, 4,6 mm, 5 µm, 300 A Bonded phase Particle size , Smaller PZ introduce • higher surface area, • higher peak resolution • Shorter run time • But, higher back pressure too Particle ‘s Pores size • Below 60 A • 60-150 A • 150-300 A (proteins) Depending on target analytes and matrix Inter diameter Smaller ID • Higher peak resolution • But, higher back pressure too Length Increasing L • Higher peak resolution • But, higher back pressure too Injection volume, Vinj; Vinj < 0.14 𝐝2 𝐋 x dp d, L; internal diameter and length of the used column dp diameter of solid phase particles High Vinj and analyst Conc. May cause fronting of un-retained molecules
  • 12. COLUMN SPECIFICATION C18, 5 µm, 150 mm × 4.6 mm column C18, 3 µm, 150 mm × 4.6 mm column C18, 2.7 µm, 100 mm × 4.6 mm HPLC-FLD chromatograms of aflatoxins
  • 13. COLUMN SPECIFICATION Pilar Campíns-Falcó; Rosa Herráez-Hernández; Pascual Serra-Mora, Liquid Chromatography-Instrumentation
  • 15. MOBILE PHASE, KOW THE ELUTION STRENGTH APPLING RP-HPLC ; • HEXANE(3.13)>TERT-BUTANOL(0.54),ISOPROPANOL(0.25), ACETONE (0.11), ETHANOL (-0.16)ACETONITRILE (-0.34), METHANOL(-0.77), WATER (-1.38). first loading of sample components in RP HPLC. changing between two immiscible solvents. It is also used for long storage of the used columns Kow Polarity water and a miscible organic solvent are commonly used mobile phase mixtures in HPLC since they give a new Kow that rapid the chromatographic separation run. Log Kow (A+B)= 1 − 𝑥 𝐿𝑜𝑔 𝐾𝑜𝑤 𝐴 + 𝑋 𝐿𝑜𝑔 𝐾𝑜𝑤 𝐵 A& B; used solvents.
  • 16. MOBILE PHASE, KOW Kow Polarity • Acetone has a strong eluting properties in RP-HPLC but it has a high UV cut-off value. • Methylene chloride, chloroform have medium polarity but in RP HPLC it is immiscible with water. • Acetonitrile has a very low UV cut-off,
  • 17. THE BASIC COMPONENTS OF HPLC SYSTEM  Gradient elution results in; minimizing the run time, good shape for eluted peaks, and cleaning the used column for each run. includes three basic steps:  Short isocratic start (Solution A, lower elution)  Gradient program (gradually increase solution B, higher elution).  Short isocratic for cleaning the used column ( Highest percent of solution B).  Return to the initial conditions (column conditioning, very critical step).
  • 18. MOBILE PHASE, GRADIENT ELUTION A (H2O:MeOH 5%, pH 4.6). B (MeOH) Time 10 % B 0.01-5.00 % B 70 5.01-7.00 10 % B 7.01-10.00 BENZOIC AND SORBIC ACID DETERMINATION IN JUICE, MILK PRODUCTS,…
  • 19. BUFFER, HPLC MOBILE PHASE  In RP- HPLC Solvent A is mainly water as it:  The highest polar solvent (weakest eluent), suitable for sample loading.  Buffers can be easily prepared in water at different concentrations.  The purity of water for HPLC is very important, especially when using MS/ MS.  It is preferred to use a percent (About 10 %) of a suitable organic solvent in water (Solvent A);  To prevent algae formation.  Enhance the evaporation in ESI ionization unit (LC-MS/MS).
  • 20. BUFFER, HPLC MOBILE PHASE  For separation of acidic or basic compounds, a buffer should be used in solvent A to keep the ionizable compounds in neutral form as possible. Where pH of the mobile phase A (using buffer) should be acidic for basic compounds and the reverse.  Changing pH by 2 units (lower or higher than its pKa) create a reverse partitioning condition for this compound. pH= pKa+ Log ( [𝐴−] [𝐻𝐴] ) At pH= 2, -2.2= Log ( [𝐴−] [𝐻𝐴] ), [HA]=166 [𝐴− ] http://dx.doi.org/10.1016/B978-0-12-803684-6.00004-4
  • 21. BUFFER, HPLC MOBILE PHASE pH= pKa+ Log ( [𝐴−] [𝐻𝐴] ) http://dx.doi.org/10.1016/B978-0-12-803684-6.00004-4 Hexane MeOH (basic mix) MeOH (Acidic mix) Ricinoic acid (castor Oil)
  • 22. BUFFER, HPLC MOBILE PHASE  A buffer is a solution of a weak acid (HA) and its conjugate base (𝐴− ) & a solution of a weak base (B) and its conjugated acid (𝐵𝐻+ )  Such buffer system has the capability to resistance changing in pH upon the addition of small amounts from a strong acid or base. pH= pKa+ Log ( [𝐴−] [𝐻𝐴] ) Cbuf = [𝐴−] + [𝐻𝐴] Buffer (pH & Concentration)
  • 23. BUFFER, HPLC MOBILE PHASE  Buffer capacity, β: The number of added moles (dn) of a strong acid or a base that change the pH of one-liter buffer solution. β= 𝑑 𝑛 𝑝𝐻 Since, the addition of n moles from a base (NaOH) leads to the formation of [𝐴− Na] or CNaA β=[𝐴− b] 𝑝𝐻 http://dx.doi.org/10.1016/B978-0-12-803684-6.00004-4 https://www.kisspng.com/png-bouncer-clip- art-bodyguard-security-guard-royalty- 5953930/ β
  • 24. BUFFER, HPLC MOBILE PHASE https://www.kisspng.com/png-bouncer-clip- art-bodyguard-security-guard-royalty- 5953930/ β Fixed http://dx.doi.org/10.1016/B978-0-12-803684-6.00004-4 pH= pKa+ Log ( [𝐴−] [𝐻𝐴] ) Cbuf = [𝐴− ] + [𝐻𝐴]
  • 25. BUFFER, HPLC MOBILE PHASE • BUFFER CONCENTRATION IN HPLC UV IS USUALLY MADE WITH A CONCENTRATION OF 10- 200 MM. (LOWER CONCENTRATIONS FOR LC MSMS <50MM, VOLATILE SALTS ARE MORE FAVORABLE). • THE SOLUBILITY OF INORGANIC SALTS DEPENDS MAINLY ON THE NATURE OF THE CATION, AND THEIR SOLUBILITY TREND IN ORGANIC SOLVENTS FOLLOWS (THE SAME AS IN WATER): NH4+ >𝐾+>𝑁𝑎+. • A HIGHER CONTENT OF ORGANIC PHASE SHOULD BE AVOIDED NOT TO PRECIPITATE THE BUFFER SALTS. _____________________________________________________ • FOR PREPARING A BUFFER AT PH 4.5 YOU SHOULD USE: A WEAK ACID OF PKA CLOSE TO THE REQUIRED PH (……ACID PKA =….) THE CONJUGATE BASE FOR THIS ACID WILL BE ITS (NA, K, AMMONIUM) SALT WHAT IS THE TOTAL BUFFER CONCENTRATION AND HOW TO PREPARE ?
  • 27. BUFFER, HPLC MOBILE PHASE http://dx.doi.org/10.1016/B978-0-12-811732-3.00006-6 Volatile buffers that can be used for LC MS
  • 29. SOLVENT OF THE SAMPLE The solvent of the sample should be with; - Same polarity as the mobile phase (loading solvent, A) or weaker (increase solvent compressing, enhance the retention of solutes). - pH close to that of the mobile phase (loading solvent, A) These points are more critical, especially for higher injection volume and for ionizable solutes solutes Zorbax Eclipse XDB- C18, 150 mm, 4.6 mm, 5 µm with a mobile phase: 35% ACN and 65% H2O (0.2% H3PO4); flow-rate: 2 mL/min.
  • 30. A CHROMATOGRAM AND A MASS SPECTRUM
  • 31. COLUMN EFFICIENCY Jesús Lozano-Sánchez; Isabel Borrás-Linares; Agnes Sass-Kiss; Antonio Segura-Carretero, Chapter 13 - Chromatographic Technique: High-Performance Liquid Chromatography (HPLC) Dead time, Void time T0 = Time at which mobile phase pass through the chromatographic column Depending on column length and flow rate Actual Rt (X)= Obtained Rt (X)+ t0
  • 32. COLUMN EFFICIENCY Plate Number (N) N= (𝑅𝑡/ )2 N= 16(𝑅𝑡/Wb)2 N= 5.45 (𝑅𝑡/W 0.5)2 Wb= 4*x Standard deviation ( ) Selectivity = 𝑅𝑡 (𝑥) 𝑅𝑡 (𝑦) Resolution R = 2 [𝑅𝑡 𝑥 −𝑅𝑡 𝑦 ] (𝑊𝑏 𝑥 +𝑊𝑏 (𝑦) R= Δ𝑅𝑡 Column efficiency:  High Plate N per unit length (N/L) or Height of the theoretical plate H= 𝐿/𝑁≃ 2dp dp diameter of solid phase particles  Selectivity (Separation Factor) A measurement for the separation of two compounds X and Y  Resolution Separation of apexes of two adjacent peaks
  • 33. A. Prof. Sherif M. Taha Tel: 01004724944 sherif2taha@gmail.com Thank You