Measuring Cardiac Contraction Velocity Using M-Mode
Ultrasonography with Digital Image Processing
Niken Larasati*, Suryono**, Kusworo Adi***
*Physics Graduate Program, Diponegoro University, Kota Semarang, Indonesia
nikenlrst@st.fisika.undip.ac.id
**, ***Physics Department, Diponegoro University, Kota Semarang, Indonesia
suryono@undip.ac.id, kusworoadi@undip.ac.id
Manuscript History
Number: IJIRAE/RS/Vol.04/Issue06/JNAE10085
Received: 27, April 2017
Final Correction: 19, May 2017
Final Accepted: 27, May 2017
Published: June 2017
Citation: Larasati, N.; Suryono & Adi, K. (2017), 'Measuring Cardiac Contraction Velocity Using M-Mode
Ultrasonography with Digital Image Processing', Master's thesis, Diponegoro University, Kota Semarang, Indonesia.
Editor: Dr.A.Arul L.S, Chief Editor, IJIRAE, AM Publications, India
Copyright: ©2017 This is an open access article distributed under the terms of the Creative Commons Attribution License, Which
Permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Abstract— Measurement of cardiac contraction velocity using M-mode ultrasonography with the help of digital
image processing has been carried out. Analysis of organ movement rate is of great importance in medical imaging
as it relates to diagnosis for diseases. This research is aimed at measuring the rate of heart contraction using digital
image processing that employs the active contour segmentation method. The sample used is a video of cardiac check
using M-mode USG that lasts 30 seconds and has a frame rate of 30 frames/seconds. This image was then extracted as
to get 900 USG images each at 480 x 360 pixels. Measurement of the edges of cardiac images was conducted using
MATLAB software. Results show that the cardiac in the video shifted farthest at 73/30 seconds, to 10.34 mm against
the x-axis and at 28/30 seconds, to 14.00 mm from the y-axis. Meanwhile, the contraction velocity fluctuated.
Keywords— Velocity, Cardiac, USG Image, Digital Image Processing
I.INTRODUCTION
Analysis of organ movement velocity is of great importance in medical imaging as it relates to diagnosis for
diseases. Shifts in motion velocity can also be used to compress digital images using blocks based on motion
estimation techniques [7]. Measuring organs’ motion velocity is necessary in order to know the position of
certain organs at particular times, as this also relates to the proper therapy the organs may need [17].
Over the past few years there have been studies concerning analyses of organ motion velocity. Organ motion
measurement is used in the analysis of breast deformation to know tissue elasticity and provide indications for
tissue stiffness by calculating the images’ relative strain and young modulus [10]. Some elasticity imaging
techniques have also been developed to quantitatively measure tissue elasticity with the help of ultrasound.
Examples for these include detection and diagnosis of cancer in breast, prostate, and liver. Other examples
include related clinical applications that measure the elastic property of soft tissues. Hence, the development of
elastography USG took place. This technique shows great promise due to its real-time ability and its ease of use.
Elastography USG, when it comes to B-mode sonography, proves to be a reliable method in confirming benign
breast lesion. Moreover, differentiation of malignancy category using elastography equals to that using
conventional USG [12].
International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: 2349-2163
Issue 06, Volume 4 (June 2017) www.ijirae.com
________________________________________________________________________________________________
IJIRAE: Impact Factor Value – SJIF: Innospace, Morocco (2016): 3.916 | PIF: 2.469 | Jour Info: 4.085 |
ISRAJIF (2016): 3.715 | Indexcopernicus: (ICV 2015): 47.91
IJIRAE © 2014- 17, All Rights Reserved Page -58
Ultrasonography is also utilized in estimating and visualizing longitudinal movements of muscles [11]. USG is
used in velocimetry imaging of human vitreous as well [18]. In 2007, USG was used to measure the velocity of
blood flow using two methods of cross-correlation and de-correlation [16].
Ultrasound is considered the cheapest method of all. This shows its routine use in hospitals and clinics to
diagnose all sorts of diseases. This is the instrument of choice for midwifery and cardiology due to its safety and
real-time image processing ability. Nowadays, USG images are already in the form of digital image known as
Digital Image Communication in Medicine (DICOM). Each digital image has its own blurring level that may affect
an image’s sharpness [6]. Generally, a digital image comes with noise that requires image processing for its
improvement, as to allow proper interpretation.
II. METHOD
The research was carried out at the Instrumentation and Electronics Laboratory of the Faculty of Science and
Mathematics in Diponegoro University, Semarang. Analyses were made on a video of a cardiac check using M-
mode USG with the help of a PC equipped with MATLAB R2014b.
The procedures involved were; video extraction, image segmentation, morphology operation, determination of
image spatial resolution, and graphical analyses. Extraction was carried out on a video of a cardiac check that
lasts 30 seconds and with a frame rate of 30 frames/second. This process yielded 900 images of 480 x 360 pixel.
From these, the number to be analyzed was 150, or the cardiac contraction for the first 5 seconds.
Image segmentation is the processing of images that separates the region of the object with that of its
background, as to make it easier to do analyses. This is necessary as object recognition requires a lot of visual
perceptions. The segmentation method done here includes; active contouring, which a segmentation method is
using a loop curve that can either narrows or expands [4].
The initial process of this stage is by determining the initial masking, represented in a loop curve and then
iteratively modifying this mask that will allow the narrowing and expanding operation until a desired object
shape is obtained, as depicted in Figure 1.
Figure 1. The mask used in segmentation
Measurement of spatial resolution is carried out by calculating the number of pixels within a 10 mm distance.
This research analyzed two dimensional cardiac contraction. Therefore, the upper horizontal edge is taken as the
x-axis, and the vertical edge of the image is taken as the y-axis. Subsequently, two graphs will be analyzed; the
graph between the edge of the heart to the x-axis, and the graph between the edge of the heart to the y-axis.
The first step to do is determining a random point in the x and y axis as a reference. This reference for the x-axis
is at point 360 and for the y-axis is at point 150. The next step is to draw a line from the reference point to the
edge of the cardiac image and then measure its distance. Once the distance is known, the following step is
plotting this distance data against time in a graph. The flow chart of the research is given in Figure 2 as follows:
International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: 2349-2163
Issue 06, Volume 4 (June 2017) www.ijirae.com
________________________________________________________________________________________________
IJIRAE: Impact Factor Value – SJIF: Innospace, Morocco (2016): 3.916 | PIF: 2.469 | Jour Info: 4.085 |
ISRAJIF (2016): 3.715 | Indexcopernicus: (ICV 2015): 47.91
IJIRAE © 2014- 17, All Rights Reserved Page -59
Figure 2. Research Flow Chart
III.RESULT AND DISCUSSION
Segmentation is aimed at separating the object region from the background area as to make it easier for object
analysis using visual perception. This process involves the use of the loop curve model that can either narrows or
expands by minimizing image energy using external force. This process is also influenced by the image’s
characteristics such as line or edge.
(a) Original image (b) The mask used
(c) Result of active contouring segmentation (d)Visual of segmentation result
Figure 3. Segmentation process using the active contouring method on frame 1
International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: 2349-2163
Issue 06, Volume 4 (June 2017) www.ijirae.com
________________________________________________________________________________________________
IJIRAE: Impact Factor Value – SJIF: Innospace, Morocco (2016): 3.916 | PIF: 2.469 | Jour Info: 4.085 |
ISRAJIF (2016): 3.715 | Indexcopernicus: (ICV 2015): 47.91
IJIRAE © 2014- 17, All Rights Reserved Page -60
Figure 3 depicts results of active contouring on the image at frame 1. Figure 3(a) shows the original image from
video grabbing of a cardiac check image that used M-mode USG in frame 1. Figure 3(b) is the mask used in the
segmentation process using active contour. The mask can either narrows or expands in line with the cardiac
image. Figure 3(c) reveals result of active contouring of the cardiac image. Figure 3(d) depicts result of active
contouring for the heart visualized on the first frame of the cardiac image.
Figure 4 shows result of active contouring on frame 55. Like in Figure 4, Figure 4(a) depicts the real image from
video grabbing from the image of a cardiac check using M-mode USG on frame 55. 4(b) is the mask used in active
contouring segmentation. The mask can either narrows or expands in line with the cardiac image. Figure 4(c)
reveals result of active contouring of the cardiac image. Figure 4(d) depicts result of active contouring of the
cardiac image visualized on frame 55 of the cardiac image.
(a) Original image (b) The mask used
(c) Result of active contouring segmentation (d) Visual of segmentation result
Figure 4. Segmentation process using the active contouring method on Frame 55
The subsequent step is conducting morphological operation on segmentation results. This process is aimed at
making up for the lacks from segmentation. The operation employed here is known as areal opening. Results of
this operation are the loss of smaller objects without compromising the area of the target object. After some
trials and errors, it is determined that the area of more than 2750 is considered the cardiac image and the area
less than this value is going to be erased.
Figure 3(a) shows result of active contouring segmentation of the cardiac image on frame 1. Figure 3(b) is Figure
3(a) after undergoing morphology operation. Both pictures do not reveal any differences as there is no noise on
frame 1. Meanwhile, Figure 4(a) shows result of active contouring segmentation of the cardiac image on frame
55. Unlike frame 1, there is noise on frame 55, just below the cardiac image. Once morphology is carried out, the
noise is gone and the leftover is the cardiac image as depicted in Figure 4(b).
(a)Result of active contouring segmentation
(b)Result of morphology operation
Figure 5. Morphology operation on frame 1
International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: 2349-2163
Issue 06, Volume 4 (June 2017) www.ijirae.com
________________________________________________________________________________________________
IJIRAE: Impact Factor Value – SJIF: Innospace, Morocco (2016): 3.916 | PIF: 2.469 | Jour Info: 4.085 |
ISRAJIF (2016): 3.715 | Indexcopernicus: (ICV 2015): 47.91
IJIRAE © 2014- 17, All Rights Reserved Page -61
(a) Result of active contouring segmentation (b) Result of morphology operation
Figure 6. Morphology operation on frame 55
CALCULATION OF IMAGE SPATIAL RESOLUTION
Measurements show that within a 10 mm distance, there are 15 pixels. The magnitude of image spatial
resolution of USG probe can be known from the following formula:
Spatial resolution =
tan ( )
numberofpixel
dis ce mm
Spatial Rsolution =
15
10
pixel
mm
Resolution olution = 1.5 pixels per mm
Once measurement of the cardiac edge against the x-axis from frame 1 to frame 150 is done, distance for each
frame is calculated to figure out the distance covered in 1/30 seconds. Results of calculation of the distance
covered by the cardiac against time on the x-axis are given in Figure 7. The reference point use is x = 360 mm. It
can be seen in Figure 7 that some points show the cardiac does not shift from its position. The cardiac has
maximum contraction at 73/30 seconds, for as far as 10.34 mm.
Figure 7. The distance covered by the cardiac (mm) against time (s) on the x-axis
Results of measurement for the distance covered by the cardiac against time on the y-axis are given in Figure 8.
The reference point used is at y = 150 mm. It can be seen in Figure 8 that some points are of 0 (zero) value, which
indicates no displacement. The farthest shift takes place on the 28/30 seconds, for as far as 14.00 mm.
Measurements of distance on the y-axis does not reveal any contraction pattern as is the case for the x-axis.
Therefore, the average contraction velocity cannot be calculated.
International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: 2349-2163
Issue 06, Volume 4 (June 2017) www.ijirae.com
________________________________________________________________________________________________
IJIRAE: Impact Factor Value – SJIF: Innospace, Morocco (2016): 3.916 | PIF: 2.469 | Jour Info: 4.085 |
ISRAJIF (2016): 3.715 | Indexcopernicus: (ICV 2015): 47.91
IJIRAE © 2014- 17, All Rights Reserved Page -62
Figure 8. The distance covered by the cardiac (mm) against time (s) on the y-axis
IV. CONCLUSIONS
It can be inferred from the results here that cardiac contraction fluctuates in its rate. The cardiac studied here has
the farthest shift on 73/30 seconds, for as far as 10.34 mm on the x-axis, and on the 28/30 seconds, for as far as
14.00 mm on the y-axis.
REFERENCES
1. Burns, P.N, 2005, Introduction to the Physical Principles of Ultrasound Imaging and Doppler, Journal Medical
Biophysics-MBP1007/1008, pp 5.
2. Bushberg, J., T., and J. Anthony S., 2002, The Essential Physics of Medical Imaging, second edition,
Philadelphia, USA, Lippincott Williams & Wilkins.
3. Cameron, J., R., and Skofronick, J.G, 1978, Medical Physics, pp 253-287, John Wiley and Sons Inc., New York.
4. Casseles, V., Kimmel R. and Sapiro G., 1997, Geodesic Active Contour. International Journal of Computer
Vision.
5. Destyningtias B., Heranurweni S. and T. Nurhayati. 2010. Segmentasi Citra dengan Metode Pengambangan.
Jurnal Elektrika. Vol.2, No.1, 2010: 39 – 49.
6. Dougherty, G., 2009, Digital Image Processing for Medical Applications, Cambridge University Press, New
York.
7. Gao, L., K. J. Parker, R. M. Lerner, and S. F. Levinson, 1996, Ultrasound Medical Biology, pp. 959–977.
8. Goodsitt, M.M., and Carson, P.L, 1998, Real-time B-mode ultrasound quality control test procedures, Report
of AAPM Ultrasound Task Group No.1, Department of Radiology, University of Michigan.
9. Kist, W.J., 2011, Sonography-Based Automated Volume Count to estimate fetal urine production in twin-to-
twin transfusion syndrome, comparison with Virtual Organ Computer-aided analysis, American Journal of
Obstetrics Gynecol, Leiden University Medical Center, Leiden, The Netherlands.
10. Kontogeorgakis, C., M. G. Strintzis, N. Maglaveras, and I. Kokkinidis, 1994, Tumor Detection in Ultrasound B-
mode Image Through Estimation Using Texture Detection Algorithm, Proceedings 1994 Computer Cardiology
Conference, pp. 117–120.
11. Li, Jizhou, 2013, Estimation and Visualization of Longitudinal Muscle Motion Using Ultrasonography: A
feasibility study, Ultrasonic 54 (2014) 779–788.
12. Lu, Minhua, 2014, A Real Time Displacement Algorithm for Ultrasound Elastography, COMIND-2613; No. of
Pages 11.
13. Maysanjaya, I., and M.Dendi, 2013, Pengembangan system identifikasi jenis kelamin janin pada citra USG,
ISSN 2089-8673, ,Volume 2, Nomor 1, Jurnal Nasional Pendidikan Teknik Informatika (JANAPATI).
14. Munir, R., 2004, Pengolahan Citra Digital dengan Pendekatan Algoritma, Penerbit Informatika, Bandung.
15. Osman, M.Y., and Tahab F.A., 2005, Quality Control Program of Real Time Medical Ultrasound Machines In
Sudan, Sudan Atomic Energy commission, Khartoum, Department of Applied Physics, Faculty of Applied
sciences, Sudan.
International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: 2349-2163
Issue 06, Volume 4 (June 2017) www.ijirae.com
________________________________________________________________________________________________
IJIRAE: Impact Factor Value – SJIF: Innospace, Morocco (2016): 3.916 | PIF: 2.469 | Jour Info: 4.085 |
ISRAJIF (2016): 3.715 | Indexcopernicus: (ICV 2015): 47.91
IJIRAE © 2014- 17, All Rights Reserved Page -63
16. Pan, T., and Huihua Kenny Chiang, 2007, Ultrasound Low-Velocity Flow Estimations Using Cross-Correlation
and Decorrelation: A thread phantom study, Medical Engineering & Physics 29 (2007) 602–614.
17. Ranjit ,S.S., A.F. Tuani Ibrahim, S.I. Salim and S.K. Subramaniam, 2009, Analysis of Motion Velocity in
Ultrasounds Image, International Journal of Video & Image Processing and Network Security Vol: 9 No: 9.
18. Rossi, T., and Giorgio Querzoli, 2012, Ultrasound Imaging Velocimetry of The Human Vitreous, Experimental
Eye Research 99 (2012) 98e104.
19. Smith, S.W., 1999, The Scientist and Engineer’s Guide to Diginal Signal Processing Second Edition, California
Technical Publishing, California.
20. Tole, N.M. and Ostensen, H., 2005, Basic Physics of Ultrasonographic Imaging, WHO press, Geneva.
21. Vincent, Wang J., 2004, Transducer.
22. Zhang, Junhua, 2008, Computer-Aided diagnosis of Cervical Lymph Nodes on Ultrasonography, Fudan
University, Shanghai, China.

More Related Content

PDF
FOOT ANTHROPOMETRY DEVICE AND SINGLE OBJECT IMAGE THRESHOLDING
PDF
A theoretical study on partially automated method
PDF
A theoretical study on partially automated method for (prostate) cancer pinpo...
PDF
Gait Recognition for Security and Surveillance System- A Review
PDF
International Journal of Engineering Research and Development (IJERD)
PDF
Lec5: Pre-Processing Medical Images (III) (MRI Intensity Standardization)
PDF
Classification of appendicitis based on ultrasound image
PDF
Left Ventricle Volume Measurement on Short Axis MRI Images Using a Combined R...
FOOT ANTHROPOMETRY DEVICE AND SINGLE OBJECT IMAGE THRESHOLDING
A theoretical study on partially automated method
A theoretical study on partially automated method for (prostate) cancer pinpo...
Gait Recognition for Security and Surveillance System- A Review
International Journal of Engineering Research and Development (IJERD)
Lec5: Pre-Processing Medical Images (III) (MRI Intensity Standardization)
Classification of appendicitis based on ultrasound image
Left Ventricle Volume Measurement on Short Axis MRI Images Using a Combined R...

Similar to Measuring Cardiac Contraction Velocity Using M-Mode Ultrasonography with Digital Image Processing (20)

PDF
CALCULATION of the CEREBRAL HEMORRHAGE VOLUME USING ANALYSIS of COMPUTED TOMO...
PDF
CALCULATION OF AREA, CENTER AND DISTANCE OF CERVICAL CANCER FROM ORGAN AT RIS...
PDF
CHEST X-RAY SEGMENTATION to CALCULATE PLEURAL EFFUSION INDEX in PATIENT with ...
PDF
BREAST TUMOR DETECTION USING HAAR-LIKE FEATURE METHOD ON ULTRASONOGRAPHY (USG...
PDF
DISTANCE MEASUREMENT WITH A STEREO CAMERA
PDF
3D Body Scanning for Human Anthropometry
PDF
A Survey on Needle Tip Estimation in Ultrasound Images
PDF
IMAGINARY ANATOMY LINE MEASUREMENTS ON THE THORAX
PDF
FPGA Implementation for Image Edge Detection using Xilinx System Generator
PDF
DETERMINATION OF BREAST CANCER AREA FROM MAMMOGRAPHY IMAGES USING THRESHOLDIN...
PDF
IMPLEMENTATION OF LOSSLESS HYPERSPECTRAL AND MULTISPECTRAL IMAGE COMPRESSION ...
PDF
IRJET-V9I1137.pdf
PDF
Segmentation of MR Images using Active Contours: Methods, Challenges and Appl...
PDF
An Innovative Deep Learning Framework Integrating Transfer- Learning And Extr...
PDF
IRJET - 3D Reconstruction and Modelling of a Brain MRI with Tumour
PDF
GRADIENT CALCULATION on the PLEURAL EFFUSION and NORMAL LUNGS IMAGE
PDF
PROFILE DOSE AND PDD ANALYSIS IN SMALL PHOTON FIELD WITH PTW PINPOINT CHAMBER...
PDF
Dorsal hand vein pattern authentication by hough peaks
PDF
Real time voting system using face recognition for different expressions and ...
PDF
Advanced Audio Aid for Visually Impaired using Raspberry Pi
CALCULATION of the CEREBRAL HEMORRHAGE VOLUME USING ANALYSIS of COMPUTED TOMO...
CALCULATION OF AREA, CENTER AND DISTANCE OF CERVICAL CANCER FROM ORGAN AT RIS...
CHEST X-RAY SEGMENTATION to CALCULATE PLEURAL EFFUSION INDEX in PATIENT with ...
BREAST TUMOR DETECTION USING HAAR-LIKE FEATURE METHOD ON ULTRASONOGRAPHY (USG...
DISTANCE MEASUREMENT WITH A STEREO CAMERA
3D Body Scanning for Human Anthropometry
A Survey on Needle Tip Estimation in Ultrasound Images
IMAGINARY ANATOMY LINE MEASUREMENTS ON THE THORAX
FPGA Implementation for Image Edge Detection using Xilinx System Generator
DETERMINATION OF BREAST CANCER AREA FROM MAMMOGRAPHY IMAGES USING THRESHOLDIN...
IMPLEMENTATION OF LOSSLESS HYPERSPECTRAL AND MULTISPECTRAL IMAGE COMPRESSION ...
IRJET-V9I1137.pdf
Segmentation of MR Images using Active Contours: Methods, Challenges and Appl...
An Innovative Deep Learning Framework Integrating Transfer- Learning And Extr...
IRJET - 3D Reconstruction and Modelling of a Brain MRI with Tumour
GRADIENT CALCULATION on the PLEURAL EFFUSION and NORMAL LUNGS IMAGE
PROFILE DOSE AND PDD ANALYSIS IN SMALL PHOTON FIELD WITH PTW PINPOINT CHAMBER...
Dorsal hand vein pattern authentication by hough peaks
Real time voting system using face recognition for different expressions and ...
Advanced Audio Aid for Visually Impaired using Raspberry Pi
Ad

More from AM Publications (20)

PDF
DEVELOPMENT OF TODDLER FAMILY CADRE TRAINING BASED ON ANDROID APPLICATIONS IN...
PDF
TESTING OF COMPOSITE ON DROP-WEIGHT IMPACT TESTING AND DAMAGE IDENTIFICATION ...
PDF
THE USE OF FRACTAL GEOMETRY IN TILING MOTIF DESIGN
PDF
TWO-DIMENSIONAL INVERSION FINITE ELEMENT MODELING OF MAGNETOTELLURIC DATA: CA...
PDF
USING THE GENETIC ALGORITHM TO OPTIMIZE LASER WELDING PARAMETERS FOR MARTENSI...
PDF
ANALYSIS AND DESIGN E-MARKETPLACE FOR MICRO, SMALL AND MEDIUM ENTERPRISES
PDF
REMOTE SENSING AND GEOGRAPHIC INFORMATION SYSTEMS
PDF
EVALUATE THE STRAIN ENERGY ERROR FOR THE LASER WELD BY THE H-REFINEMENT OF TH...
PDF
HMM APPLICATION IN ISOLATED WORD SPEECH RECOGNITION
PDF
PEDESTRIAN DETECTION IN LOW RESOLUTION VIDEOS USING A MULTI-FRAME HOG-BASED D...
PDF
INTELLIGENT BLIND STICK
PDF
EFFECT OF SILICON - RUBBER (SR) SHEETS AS AN ALTERNATIVE FILTER ON HIGH AND L...
PDF
UTILIZATION OF IMMUNIZATION SERVICES AMONG CHILDREN UNDER FIVE YEARS OF AGE I...
PDF
REPRESENTATION OF THE BLOCK DATA ENCRYPTION ALGORITHM IN AN ANALYTICAL FORM F...
PDF
OPTICAL CHARACTER RECOGNITION USING RBFNN
PDF
DETECTION OF MOVING OBJECT
PDF
SIMULATION OF ATMOSPHERIC POLLUTANTS DISPERSION IN AN URBAN ENVIRONMENT
PDF
PREPARATION AND EVALUATION OF WOOL KERATIN BASED CHITOSAN NANOFIBERS FOR AIR ...
PDF
ANALYSIS ON LOAD BALANCING ALGORITHMS IMPLEMENTATION ON CLOUD COMPUTING ENVIR...
PDF
A MODEL BASED APPROACH FOR IMPLEMENTING WLAN SECURITY
DEVELOPMENT OF TODDLER FAMILY CADRE TRAINING BASED ON ANDROID APPLICATIONS IN...
TESTING OF COMPOSITE ON DROP-WEIGHT IMPACT TESTING AND DAMAGE IDENTIFICATION ...
THE USE OF FRACTAL GEOMETRY IN TILING MOTIF DESIGN
TWO-DIMENSIONAL INVERSION FINITE ELEMENT MODELING OF MAGNETOTELLURIC DATA: CA...
USING THE GENETIC ALGORITHM TO OPTIMIZE LASER WELDING PARAMETERS FOR MARTENSI...
ANALYSIS AND DESIGN E-MARKETPLACE FOR MICRO, SMALL AND MEDIUM ENTERPRISES
REMOTE SENSING AND GEOGRAPHIC INFORMATION SYSTEMS
EVALUATE THE STRAIN ENERGY ERROR FOR THE LASER WELD BY THE H-REFINEMENT OF TH...
HMM APPLICATION IN ISOLATED WORD SPEECH RECOGNITION
PEDESTRIAN DETECTION IN LOW RESOLUTION VIDEOS USING A MULTI-FRAME HOG-BASED D...
INTELLIGENT BLIND STICK
EFFECT OF SILICON - RUBBER (SR) SHEETS AS AN ALTERNATIVE FILTER ON HIGH AND L...
UTILIZATION OF IMMUNIZATION SERVICES AMONG CHILDREN UNDER FIVE YEARS OF AGE I...
REPRESENTATION OF THE BLOCK DATA ENCRYPTION ALGORITHM IN AN ANALYTICAL FORM F...
OPTICAL CHARACTER RECOGNITION USING RBFNN
DETECTION OF MOVING OBJECT
SIMULATION OF ATMOSPHERIC POLLUTANTS DISPERSION IN AN URBAN ENVIRONMENT
PREPARATION AND EVALUATION OF WOOL KERATIN BASED CHITOSAN NANOFIBERS FOR AIR ...
ANALYSIS ON LOAD BALANCING ALGORITHMS IMPLEMENTATION ON CLOUD COMPUTING ENVIR...
A MODEL BASED APPROACH FOR IMPLEMENTING WLAN SECURITY
Ad

Recently uploaded (20)

PDF
. Radiology Case Scenariosssssssssssssss
PDF
VARICELLA VACCINATION: A POTENTIAL STRATEGY FOR PREVENTING MULTIPLE SCLEROSIS
PDF
Sciences of Europe No 170 (2025)
PDF
Phytochemical Investigation of Miliusa longipes.pdf
PDF
AlphaEarth Foundations and the Satellite Embedding dataset
PDF
SEHH2274 Organic Chemistry Notes 1 Structure and Bonding.pdf
PDF
lecture 2026 of Sjogren's syndrome l .pdf
PPTX
Protein & Amino Acid Structures Levels of protein structure (primary, seconda...
PPTX
BIOMOLECULES PPT........................
PPTX
DRUG THERAPY FOR SHOCK gjjjgfhhhhh.pptx.
PPTX
ANEMIA WITH LEUKOPENIA MDS 07_25.pptx htggtftgt fredrctvg
PPTX
Introduction to Cardiovascular system_structure and functions-1
PPTX
EPIDURAL ANESTHESIA ANATOMY AND PHYSIOLOGY.pptx
PPTX
cpcsea ppt.pptxssssssssssssssjjdjdndndddd
PPT
POSITIONING IN OPERATION THEATRE ROOM.ppt
PPTX
Introduction to Fisheries Biotechnology_Lesson 1.pptx
PPTX
neck nodes and dissection types and lymph nodes levels
PPTX
2. Earth - The Living Planet earth and life
PPTX
Comparative Structure of Integument in Vertebrates.pptx
PPTX
Derivatives of integument scales, beaks, horns,.pptx
. Radiology Case Scenariosssssssssssssss
VARICELLA VACCINATION: A POTENTIAL STRATEGY FOR PREVENTING MULTIPLE SCLEROSIS
Sciences of Europe No 170 (2025)
Phytochemical Investigation of Miliusa longipes.pdf
AlphaEarth Foundations and the Satellite Embedding dataset
SEHH2274 Organic Chemistry Notes 1 Structure and Bonding.pdf
lecture 2026 of Sjogren's syndrome l .pdf
Protein & Amino Acid Structures Levels of protein structure (primary, seconda...
BIOMOLECULES PPT........................
DRUG THERAPY FOR SHOCK gjjjgfhhhhh.pptx.
ANEMIA WITH LEUKOPENIA MDS 07_25.pptx htggtftgt fredrctvg
Introduction to Cardiovascular system_structure and functions-1
EPIDURAL ANESTHESIA ANATOMY AND PHYSIOLOGY.pptx
cpcsea ppt.pptxssssssssssssssjjdjdndndddd
POSITIONING IN OPERATION THEATRE ROOM.ppt
Introduction to Fisheries Biotechnology_Lesson 1.pptx
neck nodes and dissection types and lymph nodes levels
2. Earth - The Living Planet earth and life
Comparative Structure of Integument in Vertebrates.pptx
Derivatives of integument scales, beaks, horns,.pptx

Measuring Cardiac Contraction Velocity Using M-Mode Ultrasonography with Digital Image Processing

  • 1. International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: 2349-2163 Issue 06, Volume 4 (June 2017) www.ijirae.com ________________________________________________________________________________________________ IJIRAE: Impact Factor Value – SJIF: Innospace, Morocco (2016): 3.916 | PIF: 2.469 | Jour Info: 4.085 | ISRAJIF (2016): 3.715 | Indexcopernicus: (ICV 2015): 47.91 IJIRAE © 2014- 17, All Rights Reserved Page -57 Measuring Cardiac Contraction Velocity Using M-Mode Ultrasonography with Digital Image Processing Niken Larasati*, Suryono**, Kusworo Adi*** *Physics Graduate Program, Diponegoro University, Kota Semarang, Indonesia nikenlrst@st.fisika.undip.ac.id **, ***Physics Department, Diponegoro University, Kota Semarang, Indonesia suryono@undip.ac.id, kusworoadi@undip.ac.id Manuscript History Number: IJIRAE/RS/Vol.04/Issue06/JNAE10085 Received: 27, April 2017 Final Correction: 19, May 2017 Final Accepted: 27, May 2017 Published: June 2017 Citation: Larasati, N.; Suryono & Adi, K. (2017), 'Measuring Cardiac Contraction Velocity Using M-Mode Ultrasonography with Digital Image Processing', Master's thesis, Diponegoro University, Kota Semarang, Indonesia. Editor: Dr.A.Arul L.S, Chief Editor, IJIRAE, AM Publications, India Copyright: ©2017 This is an open access article distributed under the terms of the Creative Commons Attribution License, Which Permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited Abstract— Measurement of cardiac contraction velocity using M-mode ultrasonography with the help of digital image processing has been carried out. Analysis of organ movement rate is of great importance in medical imaging as it relates to diagnosis for diseases. This research is aimed at measuring the rate of heart contraction using digital image processing that employs the active contour segmentation method. The sample used is a video of cardiac check using M-mode USG that lasts 30 seconds and has a frame rate of 30 frames/seconds. This image was then extracted as to get 900 USG images each at 480 x 360 pixels. Measurement of the edges of cardiac images was conducted using MATLAB software. Results show that the cardiac in the video shifted farthest at 73/30 seconds, to 10.34 mm against the x-axis and at 28/30 seconds, to 14.00 mm from the y-axis. Meanwhile, the contraction velocity fluctuated. Keywords— Velocity, Cardiac, USG Image, Digital Image Processing I.INTRODUCTION Analysis of organ movement velocity is of great importance in medical imaging as it relates to diagnosis for diseases. Shifts in motion velocity can also be used to compress digital images using blocks based on motion estimation techniques [7]. Measuring organs’ motion velocity is necessary in order to know the position of certain organs at particular times, as this also relates to the proper therapy the organs may need [17]. Over the past few years there have been studies concerning analyses of organ motion velocity. Organ motion measurement is used in the analysis of breast deformation to know tissue elasticity and provide indications for tissue stiffness by calculating the images’ relative strain and young modulus [10]. Some elasticity imaging techniques have also been developed to quantitatively measure tissue elasticity with the help of ultrasound. Examples for these include detection and diagnosis of cancer in breast, prostate, and liver. Other examples include related clinical applications that measure the elastic property of soft tissues. Hence, the development of elastography USG took place. This technique shows great promise due to its real-time ability and its ease of use. Elastography USG, when it comes to B-mode sonography, proves to be a reliable method in confirming benign breast lesion. Moreover, differentiation of malignancy category using elastography equals to that using conventional USG [12].
  • 2. International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: 2349-2163 Issue 06, Volume 4 (June 2017) www.ijirae.com ________________________________________________________________________________________________ IJIRAE: Impact Factor Value – SJIF: Innospace, Morocco (2016): 3.916 | PIF: 2.469 | Jour Info: 4.085 | ISRAJIF (2016): 3.715 | Indexcopernicus: (ICV 2015): 47.91 IJIRAE © 2014- 17, All Rights Reserved Page -58 Ultrasonography is also utilized in estimating and visualizing longitudinal movements of muscles [11]. USG is used in velocimetry imaging of human vitreous as well [18]. In 2007, USG was used to measure the velocity of blood flow using two methods of cross-correlation and de-correlation [16]. Ultrasound is considered the cheapest method of all. This shows its routine use in hospitals and clinics to diagnose all sorts of diseases. This is the instrument of choice for midwifery and cardiology due to its safety and real-time image processing ability. Nowadays, USG images are already in the form of digital image known as Digital Image Communication in Medicine (DICOM). Each digital image has its own blurring level that may affect an image’s sharpness [6]. Generally, a digital image comes with noise that requires image processing for its improvement, as to allow proper interpretation. II. METHOD The research was carried out at the Instrumentation and Electronics Laboratory of the Faculty of Science and Mathematics in Diponegoro University, Semarang. Analyses were made on a video of a cardiac check using M- mode USG with the help of a PC equipped with MATLAB R2014b. The procedures involved were; video extraction, image segmentation, morphology operation, determination of image spatial resolution, and graphical analyses. Extraction was carried out on a video of a cardiac check that lasts 30 seconds and with a frame rate of 30 frames/second. This process yielded 900 images of 480 x 360 pixel. From these, the number to be analyzed was 150, or the cardiac contraction for the first 5 seconds. Image segmentation is the processing of images that separates the region of the object with that of its background, as to make it easier to do analyses. This is necessary as object recognition requires a lot of visual perceptions. The segmentation method done here includes; active contouring, which a segmentation method is using a loop curve that can either narrows or expands [4]. The initial process of this stage is by determining the initial masking, represented in a loop curve and then iteratively modifying this mask that will allow the narrowing and expanding operation until a desired object shape is obtained, as depicted in Figure 1. Figure 1. The mask used in segmentation Measurement of spatial resolution is carried out by calculating the number of pixels within a 10 mm distance. This research analyzed two dimensional cardiac contraction. Therefore, the upper horizontal edge is taken as the x-axis, and the vertical edge of the image is taken as the y-axis. Subsequently, two graphs will be analyzed; the graph between the edge of the heart to the x-axis, and the graph between the edge of the heart to the y-axis. The first step to do is determining a random point in the x and y axis as a reference. This reference for the x-axis is at point 360 and for the y-axis is at point 150. The next step is to draw a line from the reference point to the edge of the cardiac image and then measure its distance. Once the distance is known, the following step is plotting this distance data against time in a graph. The flow chart of the research is given in Figure 2 as follows:
  • 3. International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: 2349-2163 Issue 06, Volume 4 (June 2017) www.ijirae.com ________________________________________________________________________________________________ IJIRAE: Impact Factor Value – SJIF: Innospace, Morocco (2016): 3.916 | PIF: 2.469 | Jour Info: 4.085 | ISRAJIF (2016): 3.715 | Indexcopernicus: (ICV 2015): 47.91 IJIRAE © 2014- 17, All Rights Reserved Page -59 Figure 2. Research Flow Chart III.RESULT AND DISCUSSION Segmentation is aimed at separating the object region from the background area as to make it easier for object analysis using visual perception. This process involves the use of the loop curve model that can either narrows or expands by minimizing image energy using external force. This process is also influenced by the image’s characteristics such as line or edge. (a) Original image (b) The mask used (c) Result of active contouring segmentation (d)Visual of segmentation result Figure 3. Segmentation process using the active contouring method on frame 1
  • 4. International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: 2349-2163 Issue 06, Volume 4 (June 2017) www.ijirae.com ________________________________________________________________________________________________ IJIRAE: Impact Factor Value – SJIF: Innospace, Morocco (2016): 3.916 | PIF: 2.469 | Jour Info: 4.085 | ISRAJIF (2016): 3.715 | Indexcopernicus: (ICV 2015): 47.91 IJIRAE © 2014- 17, All Rights Reserved Page -60 Figure 3 depicts results of active contouring on the image at frame 1. Figure 3(a) shows the original image from video grabbing of a cardiac check image that used M-mode USG in frame 1. Figure 3(b) is the mask used in the segmentation process using active contour. The mask can either narrows or expands in line with the cardiac image. Figure 3(c) reveals result of active contouring of the cardiac image. Figure 3(d) depicts result of active contouring for the heart visualized on the first frame of the cardiac image. Figure 4 shows result of active contouring on frame 55. Like in Figure 4, Figure 4(a) depicts the real image from video grabbing from the image of a cardiac check using M-mode USG on frame 55. 4(b) is the mask used in active contouring segmentation. The mask can either narrows or expands in line with the cardiac image. Figure 4(c) reveals result of active contouring of the cardiac image. Figure 4(d) depicts result of active contouring of the cardiac image visualized on frame 55 of the cardiac image. (a) Original image (b) The mask used (c) Result of active contouring segmentation (d) Visual of segmentation result Figure 4. Segmentation process using the active contouring method on Frame 55 The subsequent step is conducting morphological operation on segmentation results. This process is aimed at making up for the lacks from segmentation. The operation employed here is known as areal opening. Results of this operation are the loss of smaller objects without compromising the area of the target object. After some trials and errors, it is determined that the area of more than 2750 is considered the cardiac image and the area less than this value is going to be erased. Figure 3(a) shows result of active contouring segmentation of the cardiac image on frame 1. Figure 3(b) is Figure 3(a) after undergoing morphology operation. Both pictures do not reveal any differences as there is no noise on frame 1. Meanwhile, Figure 4(a) shows result of active contouring segmentation of the cardiac image on frame 55. Unlike frame 1, there is noise on frame 55, just below the cardiac image. Once morphology is carried out, the noise is gone and the leftover is the cardiac image as depicted in Figure 4(b). (a)Result of active contouring segmentation (b)Result of morphology operation Figure 5. Morphology operation on frame 1
  • 5. International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: 2349-2163 Issue 06, Volume 4 (June 2017) www.ijirae.com ________________________________________________________________________________________________ IJIRAE: Impact Factor Value – SJIF: Innospace, Morocco (2016): 3.916 | PIF: 2.469 | Jour Info: 4.085 | ISRAJIF (2016): 3.715 | Indexcopernicus: (ICV 2015): 47.91 IJIRAE © 2014- 17, All Rights Reserved Page -61 (a) Result of active contouring segmentation (b) Result of morphology operation Figure 6. Morphology operation on frame 55 CALCULATION OF IMAGE SPATIAL RESOLUTION Measurements show that within a 10 mm distance, there are 15 pixels. The magnitude of image spatial resolution of USG probe can be known from the following formula: Spatial resolution = tan ( ) numberofpixel dis ce mm Spatial Rsolution = 15 10 pixel mm Resolution olution = 1.5 pixels per mm Once measurement of the cardiac edge against the x-axis from frame 1 to frame 150 is done, distance for each frame is calculated to figure out the distance covered in 1/30 seconds. Results of calculation of the distance covered by the cardiac against time on the x-axis are given in Figure 7. The reference point use is x = 360 mm. It can be seen in Figure 7 that some points show the cardiac does not shift from its position. The cardiac has maximum contraction at 73/30 seconds, for as far as 10.34 mm. Figure 7. The distance covered by the cardiac (mm) against time (s) on the x-axis Results of measurement for the distance covered by the cardiac against time on the y-axis are given in Figure 8. The reference point used is at y = 150 mm. It can be seen in Figure 8 that some points are of 0 (zero) value, which indicates no displacement. The farthest shift takes place on the 28/30 seconds, for as far as 14.00 mm. Measurements of distance on the y-axis does not reveal any contraction pattern as is the case for the x-axis. Therefore, the average contraction velocity cannot be calculated.
  • 6. International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: 2349-2163 Issue 06, Volume 4 (June 2017) www.ijirae.com ________________________________________________________________________________________________ IJIRAE: Impact Factor Value – SJIF: Innospace, Morocco (2016): 3.916 | PIF: 2.469 | Jour Info: 4.085 | ISRAJIF (2016): 3.715 | Indexcopernicus: (ICV 2015): 47.91 IJIRAE © 2014- 17, All Rights Reserved Page -62 Figure 8. The distance covered by the cardiac (mm) against time (s) on the y-axis IV. CONCLUSIONS It can be inferred from the results here that cardiac contraction fluctuates in its rate. The cardiac studied here has the farthest shift on 73/30 seconds, for as far as 10.34 mm on the x-axis, and on the 28/30 seconds, for as far as 14.00 mm on the y-axis. REFERENCES 1. Burns, P.N, 2005, Introduction to the Physical Principles of Ultrasound Imaging and Doppler, Journal Medical Biophysics-MBP1007/1008, pp 5. 2. Bushberg, J., T., and J. Anthony S., 2002, The Essential Physics of Medical Imaging, second edition, Philadelphia, USA, Lippincott Williams & Wilkins. 3. Cameron, J., R., and Skofronick, J.G, 1978, Medical Physics, pp 253-287, John Wiley and Sons Inc., New York. 4. Casseles, V., Kimmel R. and Sapiro G., 1997, Geodesic Active Contour. International Journal of Computer Vision. 5. Destyningtias B., Heranurweni S. and T. Nurhayati. 2010. Segmentasi Citra dengan Metode Pengambangan. Jurnal Elektrika. Vol.2, No.1, 2010: 39 – 49. 6. Dougherty, G., 2009, Digital Image Processing for Medical Applications, Cambridge University Press, New York. 7. Gao, L., K. J. Parker, R. M. Lerner, and S. F. Levinson, 1996, Ultrasound Medical Biology, pp. 959–977. 8. Goodsitt, M.M., and Carson, P.L, 1998, Real-time B-mode ultrasound quality control test procedures, Report of AAPM Ultrasound Task Group No.1, Department of Radiology, University of Michigan. 9. Kist, W.J., 2011, Sonography-Based Automated Volume Count to estimate fetal urine production in twin-to- twin transfusion syndrome, comparison with Virtual Organ Computer-aided analysis, American Journal of Obstetrics Gynecol, Leiden University Medical Center, Leiden, The Netherlands. 10. Kontogeorgakis, C., M. G. Strintzis, N. Maglaveras, and I. Kokkinidis, 1994, Tumor Detection in Ultrasound B- mode Image Through Estimation Using Texture Detection Algorithm, Proceedings 1994 Computer Cardiology Conference, pp. 117–120. 11. Li, Jizhou, 2013, Estimation and Visualization of Longitudinal Muscle Motion Using Ultrasonography: A feasibility study, Ultrasonic 54 (2014) 779–788. 12. Lu, Minhua, 2014, A Real Time Displacement Algorithm for Ultrasound Elastography, COMIND-2613; No. of Pages 11. 13. Maysanjaya, I., and M.Dendi, 2013, Pengembangan system identifikasi jenis kelamin janin pada citra USG, ISSN 2089-8673, ,Volume 2, Nomor 1, Jurnal Nasional Pendidikan Teknik Informatika (JANAPATI). 14. Munir, R., 2004, Pengolahan Citra Digital dengan Pendekatan Algoritma, Penerbit Informatika, Bandung. 15. Osman, M.Y., and Tahab F.A., 2005, Quality Control Program of Real Time Medical Ultrasound Machines In Sudan, Sudan Atomic Energy commission, Khartoum, Department of Applied Physics, Faculty of Applied sciences, Sudan.
  • 7. International Journal of Innovative Research in Advanced Engineering (IJIRAE) ISSN: 2349-2163 Issue 06, Volume 4 (June 2017) www.ijirae.com ________________________________________________________________________________________________ IJIRAE: Impact Factor Value – SJIF: Innospace, Morocco (2016): 3.916 | PIF: 2.469 | Jour Info: 4.085 | ISRAJIF (2016): 3.715 | Indexcopernicus: (ICV 2015): 47.91 IJIRAE © 2014- 17, All Rights Reserved Page -63 16. Pan, T., and Huihua Kenny Chiang, 2007, Ultrasound Low-Velocity Flow Estimations Using Cross-Correlation and Decorrelation: A thread phantom study, Medical Engineering & Physics 29 (2007) 602–614. 17. Ranjit ,S.S., A.F. Tuani Ibrahim, S.I. Salim and S.K. Subramaniam, 2009, Analysis of Motion Velocity in Ultrasounds Image, International Journal of Video & Image Processing and Network Security Vol: 9 No: 9. 18. Rossi, T., and Giorgio Querzoli, 2012, Ultrasound Imaging Velocimetry of The Human Vitreous, Experimental Eye Research 99 (2012) 98e104. 19. Smith, S.W., 1999, The Scientist and Engineer’s Guide to Diginal Signal Processing Second Edition, California Technical Publishing, California. 20. Tole, N.M. and Ostensen, H., 2005, Basic Physics of Ultrasonographic Imaging, WHO press, Geneva. 21. Vincent, Wang J., 2004, Transducer. 22. Zhang, Junhua, 2008, Computer-Aided diagnosis of Cervical Lymph Nodes on Ultrasonography, Fudan University, Shanghai, China.