Micro Sample Measurement with the
Shimadzu UV-2600 and Standard Optical Microscope
C.M. Talbott; Sudhir Dahal; Shani Samuel
Introduction
In the past two decades, there have been significant advances in the
miniaturization of electronic components utilized in industrial and consumer
products. Now it is possible to obtain a complete computer system packaged
to be the size of a small cell phone.
This drive for miniaturization has not only been focused on the electronic
components but also on the other items that make up these advanced
products. A prime example of this symbiotic miniaturization is the wide
assortment of cameras now found on most cell phones and laptop
computers.
The small cameras not only have miniature electronics, but also contain very
small lens systems to focus the incoming light onto the array. Lenses used in
such systems can be very small in size, measuring to just a few millimeters in
diameter.
Introduction
The ability to accurately characterize the transmission and reflection of small
optical components is critical to optimal function of these systems. The ability
to reproducibly mount items that are less than 3 mm in diameter in the
standard UV-Vis sample compartment can be challenging.
This paper describes the use of a Shimadzu UV-2600 UV-Vis
Spectrophotometer coupled to a standard optical microscope via an optical
fiber relay system to evaluate small filter samples approximately 1–2 mm in
width.
The Harrick FiberMate II along with two 100 µm Vis/NIR fibers (0.37 NA
PCO4) were used to connect the Shimadzu UV-2600 Spectrophotometer to a
standard optical microscope.
The biggest challenge in connecting a microscope to a spectrophotometer is
the reduction in light throughput that occurs at each lens/mirror element.
Discussion
A Zeiss Universal microscope equipped with
Epi Illumination was selected because of its
ability to minimize lens elements in the light
path. In addition, the epi illuminator can be used
with Zeiss EpiPlan HD 4X0.65 objectives, which
have a reduced number of lens elements.
Fiber optic couplers were fabricated to connect the
fibers to the microscope; one each for the source
lamp input and one for the trinocular camera port.
In addition, the lower tilt-in substage condenser
element of the microscope was removed from the
light path, as was the polarizer/analyzer, the blue tint
filter, and the epi reflection assembly to eliminate
reflection losses.
Small sections of optical filter material
(red, yellow, and blue) were excised from larger
filters to be approximately 1–2 mm at their greatest
width.
These small filter pieces were placed on a glass
microscope slide for analysis.
Discussion
The Shimadzu UV-2600 was set up for medium scan speed between 375 nm
and 875 nm, a 0.5 nm sampling pitch, and a 5 nm bandwidth. In addition, a
wire mesh attenuator (Shimadzu PN 206-82299-92) was placed in the
reference light path to compensate for the reduction in sample beam intensity
by the fibers and microscope. Visible light at 575 nm was used to position
and focus the samples. A background was acquired on the focused empty
slide.
Discussion
Microscope images of excised filters (blue, yellow and green)
A blank scan (empty slide) and scans of each filter material were acquired and are shown
below. The blank scan shows low noise transmission between 380 nm and 840 nm. Below
380 nm, the glass elements in the microscope and microscope slide absorb and block
throughput. Similarly, between 800 and 900 nm, the response profile of the PMT detector
begins to decline and noise in the baseline begins to be apparent as the S/N is reduced.
The acquired spectra for each film are also shown below. The transmission spectrum for
each individual film is of high quality with good S/N. In addition, the individual film’s
transmission profiles are well characterized by the spectra.
Results
Spectral scans of small filters
using the Shimadzu UV-2600
Spectrophotometer fiber-
connected to a standard bench
optical microscope; small green
filter (green), small blue filter
(blue), and small yellow filter
(yellow).
In a similar fashion, silicon nanospheres (approximately 3 µm in diameter) were
analyzed for their absorption spectra. The spheres were dispersed onto a glass
microscope slide and spread very thin by leveling between two slides.
The acquired absorption spectra of the two samples are shown below. The spectra
show good S/N even with the low absorbance values. In addition, the spectra show
differences between the two sphere samples.
Results
Acquired spectra of silicon
nano-spheres with different
surface treatments.
Conclusion
The advanced LO-RAY-LIGH® gratings of the UV-2600 UV-Vis
Spectrophotometer allow for the photometric range and low baseline noise
required to perform accurate analysis using fiber optic systems.
When the spectrophotometer is combined with a standard optical microscope,
very small samples can be routinely, accurately, and successfully characterized.
Thank you for viewing this presentation. Should you have any
questions or require additional information about our research,
products or services, please visit our support page:
www.ssi.shimadzu.com/support/
@shimadzussiFollow us on Twitter
Need More Info?

More Related Content

PPT
Thin Film Analysis Using Shimadzu UV-2600
PPT
Absolute Total Reflectance Spectra of Very Small Objects
PPT
Variable Angle Reflectance & Transmittance of Thin Film Coatings
PDF
When to Use a Measuring Microscope: And How to Further Enhance its Capabilities
PPT
Dental
PPTX
characteristics of films - graininess, density, speed, contrast
PPTX
Lasers in Dentistry
PPTX
Anti reflection coating
Thin Film Analysis Using Shimadzu UV-2600
Absolute Total Reflectance Spectra of Very Small Objects
Variable Angle Reflectance & Transmittance of Thin Film Coatings
When to Use a Measuring Microscope: And How to Further Enhance its Capabilities
Dental
characteristics of films - graininess, density, speed, contrast
Lasers in Dentistry
Anti reflection coating

What's hot (20)

PPTX
geometric Factors
PPTX
Talbot Interferometry to estimate focal length
PPTX
Optics v3 2
PPT
Photomask Fabrication
PPTX
LASERS IN ENDODONTICS....... Dr Jagadeesh Kodityala
PPTX
Lasers in prosthodontics
PPTX
DIODE LASER MECHANISM OF ACTION
DOCX
EUV Lithography Final
PPT
Femtosecond Machining
PDF
Characterization of materials lec 17
PPTX
LASERS USED IN OPERATIVE DENTISTRY
PPT
What are the advantages of laser driver diodes
PPTX
Lasers in Prosthetic Dentistry
PPTX
L40 film and film less techniques
PPTX
LASERS IN PROSTHODONTICS
PPTX
Optical coating
PPT
Laser in dentistry
PDF
UT Space Institute femtosecond laser micromachining workstation
PPTX
Photoresist coating
PPTX
Mask fabrication process
geometric Factors
Talbot Interferometry to estimate focal length
Optics v3 2
Photomask Fabrication
LASERS IN ENDODONTICS....... Dr Jagadeesh Kodityala
Lasers in prosthodontics
DIODE LASER MECHANISM OF ACTION
EUV Lithography Final
Femtosecond Machining
Characterization of materials lec 17
LASERS USED IN OPERATIVE DENTISTRY
What are the advantages of laser driver diodes
Lasers in Prosthetic Dentistry
L40 film and film less techniques
LASERS IN PROSTHODONTICS
Optical coating
Laser in dentistry
UT Space Institute femtosecond laser micromachining workstation
Photoresist coating
Mask fabrication process
Ad

Similar to Micro Sample Measurement Using UV-Vis and Standard Optical Microscope (20)

PPTX
9.-FIBER-OPTICS.pptx
PDF
Compendium.imaging.physics
DOCX
Calibrationfinal
PPTX
Fibre optics
PDF
Small catadioptric microscope optics 2003
PDF
basic of angiography physics and equipement.pdf
PDF
Infrared Eye Tracking Contact Lens | Whitepaper
PDF
A biologically inspired cmos image sensor
PDF
A biologically inspired cmos image sensor
PPTX
X-Ray Filters and X-Ray Beam Restrictors.pptx
PPTX
Cone beam computed tomography (2)
PPT
UV visible spectropy introduction and instrument
PDF
Closed Loop Control of Gimbal-less MEMS Mirrors for Increased Bandwidth in Li...
PDF
UV SPECTROSCOPY AND INSTRUMENTATION
PPTX
ASOCT NEW SLIDES.pptx
PDF
"OPTICAL FIBRES" -Physics Investigatory Project (Class XII)
PDF
whitelight efpi
PPTX
CBCT in orthodontics
PPT
Spetrophotometer
PPTX
My Final Year Project Final Report.pptx
9.-FIBER-OPTICS.pptx
Compendium.imaging.physics
Calibrationfinal
Fibre optics
Small catadioptric microscope optics 2003
basic of angiography physics and equipement.pdf
Infrared Eye Tracking Contact Lens | Whitepaper
A biologically inspired cmos image sensor
A biologically inspired cmos image sensor
X-Ray Filters and X-Ray Beam Restrictors.pptx
Cone beam computed tomography (2)
UV visible spectropy introduction and instrument
Closed Loop Control of Gimbal-less MEMS Mirrors for Increased Bandwidth in Li...
UV SPECTROSCOPY AND INSTRUMENTATION
ASOCT NEW SLIDES.pptx
"OPTICAL FIBRES" -Physics Investigatory Project (Class XII)
whitelight efpi
CBCT in orthodontics
Spetrophotometer
My Final Year Project Final Report.pptx
Ad

More from Shimadzu Scientific Instruments (20)

PDF
A Fast and Comprehensive Potency Determination of 21 Cannabinoids in Hemp Mat...
PDF
Determination of Heavy Metals in Baby Foods by ICP-MS
PDF
Application of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
PDF
Analysis of Diesel Range Organics (DRO) and Motor/Lube Oil Range Organics (OR...
PDF
Analysis of PFAS Compounds in Fish Tissue Using Offline Supercritical Fluid E...
PDF
Illicit Drugs and Homemade Explosives Identification using FTIR Microscopy
PDF
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
PDF
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
PDF
Ion Chromatography with Electrolytically Regenerated Suppressor for Water Ana...
PDF
Exploring the Layered Structure of Composite Films Using an FTIR Microscope
PDF
From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111
PDF
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
PDF
The Potency Determination of 15 Cannabinoids using the Hemp Analyzer
PDF
Can we keep the cost of analysis of haloacetic acids (HAAs) down by using an ...
PDF
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
PDF
Microplastics Detection and Characterization using FTIR Microscopy
PDF
Sulfur Analysis by GC-SCD using Shimadzu’s SCD-2030 for ASTM D5504, D5623, an...
PDF
Extractables and Leachables Analysis of Common Household Food Storage Product...
PPTX
The PFAS conundrum: Mass spectrometry solutions for addressing it.
PDF
Blood Alcohol Content (BAC) Analysis Utilizing GCMS and FID Combined with Ful...
A Fast and Comprehensive Potency Determination of 21 Cannabinoids in Hemp Mat...
Determination of Heavy Metals in Baby Foods by ICP-MS
Application of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
Analysis of Diesel Range Organics (DRO) and Motor/Lube Oil Range Organics (OR...
Analysis of PFAS Compounds in Fish Tissue Using Offline Supercritical Fluid E...
Illicit Drugs and Homemade Explosives Identification using FTIR Microscopy
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
Ion Chromatography with Electrolytically Regenerated Suppressor for Water Ana...
Exploring the Layered Structure of Composite Films Using an FTIR Microscope
From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
The Potency Determination of 15 Cannabinoids using the Hemp Analyzer
Can we keep the cost of analysis of haloacetic acids (HAAs) down by using an ...
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
Microplastics Detection and Characterization using FTIR Microscopy
Sulfur Analysis by GC-SCD using Shimadzu’s SCD-2030 for ASTM D5504, D5623, an...
Extractables and Leachables Analysis of Common Household Food Storage Product...
The PFAS conundrum: Mass spectrometry solutions for addressing it.
Blood Alcohol Content (BAC) Analysis Utilizing GCMS and FID Combined with Ful...

Recently uploaded (20)

PPTX
Digital-Transformation-Roadmap-for-Companies.pptx
PPTX
TLE Review Electricity (Electricity).pptx
PPTX
1. Introduction to Computer Programming.pptx
PDF
project resource management chapter-09.pdf
PPTX
TechTalks-8-2019-Service-Management-ITIL-Refresh-ITIL-4-Framework-Supports-Ou...
PPTX
Programs and apps: productivity, graphics, security and other tools
PPTX
Group 1 Presentation -Planning and Decision Making .pptx
PPTX
OMC Textile Division Presentation 2021.pptx
PDF
Encapsulation_ Review paper, used for researhc scholars
PPTX
KOM of Painting work and Equipment Insulation REV00 update 25-dec.pptx
PDF
Assigned Numbers - 2025 - Bluetooth® Document
PDF
A comparative analysis of optical character recognition models for extracting...
PDF
NewMind AI Weekly Chronicles - August'25-Week II
PDF
Building Integrated photovoltaic BIPV_UPV.pdf
PDF
Zenith AI: Advanced Artificial Intelligence
PPTX
A Presentation on Artificial Intelligence
PDF
Hindi spoken digit analysis for native and non-native speakers
PDF
DASA ADMISSION 2024_FirstRound_FirstRank_LastRank.pdf
PDF
From MVP to Full-Scale Product A Startup’s Software Journey.pdf
PPTX
cloud_computing_Infrastucture_as_cloud_p
Digital-Transformation-Roadmap-for-Companies.pptx
TLE Review Electricity (Electricity).pptx
1. Introduction to Computer Programming.pptx
project resource management chapter-09.pdf
TechTalks-8-2019-Service-Management-ITIL-Refresh-ITIL-4-Framework-Supports-Ou...
Programs and apps: productivity, graphics, security and other tools
Group 1 Presentation -Planning and Decision Making .pptx
OMC Textile Division Presentation 2021.pptx
Encapsulation_ Review paper, used for researhc scholars
KOM of Painting work and Equipment Insulation REV00 update 25-dec.pptx
Assigned Numbers - 2025 - Bluetooth® Document
A comparative analysis of optical character recognition models for extracting...
NewMind AI Weekly Chronicles - August'25-Week II
Building Integrated photovoltaic BIPV_UPV.pdf
Zenith AI: Advanced Artificial Intelligence
A Presentation on Artificial Intelligence
Hindi spoken digit analysis for native and non-native speakers
DASA ADMISSION 2024_FirstRound_FirstRank_LastRank.pdf
From MVP to Full-Scale Product A Startup’s Software Journey.pdf
cloud_computing_Infrastucture_as_cloud_p

Micro Sample Measurement Using UV-Vis and Standard Optical Microscope

  • 1. Micro Sample Measurement with the Shimadzu UV-2600 and Standard Optical Microscope C.M. Talbott; Sudhir Dahal; Shani Samuel
  • 2. Introduction In the past two decades, there have been significant advances in the miniaturization of electronic components utilized in industrial and consumer products. Now it is possible to obtain a complete computer system packaged to be the size of a small cell phone. This drive for miniaturization has not only been focused on the electronic components but also on the other items that make up these advanced products. A prime example of this symbiotic miniaturization is the wide assortment of cameras now found on most cell phones and laptop computers. The small cameras not only have miniature electronics, but also contain very small lens systems to focus the incoming light onto the array. Lenses used in such systems can be very small in size, measuring to just a few millimeters in diameter.
  • 3. Introduction The ability to accurately characterize the transmission and reflection of small optical components is critical to optimal function of these systems. The ability to reproducibly mount items that are less than 3 mm in diameter in the standard UV-Vis sample compartment can be challenging. This paper describes the use of a Shimadzu UV-2600 UV-Vis Spectrophotometer coupled to a standard optical microscope via an optical fiber relay system to evaluate small filter samples approximately 1–2 mm in width.
  • 4. The Harrick FiberMate II along with two 100 µm Vis/NIR fibers (0.37 NA PCO4) were used to connect the Shimadzu UV-2600 Spectrophotometer to a standard optical microscope. The biggest challenge in connecting a microscope to a spectrophotometer is the reduction in light throughput that occurs at each lens/mirror element. Discussion A Zeiss Universal microscope equipped with Epi Illumination was selected because of its ability to minimize lens elements in the light path. In addition, the epi illuminator can be used with Zeiss EpiPlan HD 4X0.65 objectives, which have a reduced number of lens elements.
  • 5. Fiber optic couplers were fabricated to connect the fibers to the microscope; one each for the source lamp input and one for the trinocular camera port. In addition, the lower tilt-in substage condenser element of the microscope was removed from the light path, as was the polarizer/analyzer, the blue tint filter, and the epi reflection assembly to eliminate reflection losses. Small sections of optical filter material (red, yellow, and blue) were excised from larger filters to be approximately 1–2 mm at their greatest width. These small filter pieces were placed on a glass microscope slide for analysis. Discussion
  • 6. The Shimadzu UV-2600 was set up for medium scan speed between 375 nm and 875 nm, a 0.5 nm sampling pitch, and a 5 nm bandwidth. In addition, a wire mesh attenuator (Shimadzu PN 206-82299-92) was placed in the reference light path to compensate for the reduction in sample beam intensity by the fibers and microscope. Visible light at 575 nm was used to position and focus the samples. A background was acquired on the focused empty slide. Discussion Microscope images of excised filters (blue, yellow and green)
  • 7. A blank scan (empty slide) and scans of each filter material were acquired and are shown below. The blank scan shows low noise transmission between 380 nm and 840 nm. Below 380 nm, the glass elements in the microscope and microscope slide absorb and block throughput. Similarly, between 800 and 900 nm, the response profile of the PMT detector begins to decline and noise in the baseline begins to be apparent as the S/N is reduced. The acquired spectra for each film are also shown below. The transmission spectrum for each individual film is of high quality with good S/N. In addition, the individual film’s transmission profiles are well characterized by the spectra. Results Spectral scans of small filters using the Shimadzu UV-2600 Spectrophotometer fiber- connected to a standard bench optical microscope; small green filter (green), small blue filter (blue), and small yellow filter (yellow).
  • 8. In a similar fashion, silicon nanospheres (approximately 3 µm in diameter) were analyzed for their absorption spectra. The spheres were dispersed onto a glass microscope slide and spread very thin by leveling between two slides. The acquired absorption spectra of the two samples are shown below. The spectra show good S/N even with the low absorbance values. In addition, the spectra show differences between the two sphere samples. Results Acquired spectra of silicon nano-spheres with different surface treatments.
  • 9. Conclusion The advanced LO-RAY-LIGH® gratings of the UV-2600 UV-Vis Spectrophotometer allow for the photometric range and low baseline noise required to perform accurate analysis using fiber optic systems. When the spectrophotometer is combined with a standard optical microscope, very small samples can be routinely, accurately, and successfully characterized.
  • 10. Thank you for viewing this presentation. Should you have any questions or require additional information about our research, products or services, please visit our support page: www.ssi.shimadzu.com/support/ @shimadzussiFollow us on Twitter Need More Info?

Editor's Notes

  • #2: Font style and color: Myriad, Myriad Pro or Arial, Black Font size in slide title and author: Can be adjusted depending on the text length. Slide title and author in left alignment Characters should not be overlapped on global strings. No slide page number shall be indicated on the cover page.
  • #3: Font style: Myriad, Myriad Pro or Arial Headline: 32 pt within one line, 28pt in two lines, Black, Texts within two lines Subhead: 28pt for first level bullet (subhead), 20pt for the second level bullet or contents However, sizes can be adjusted depending on the text length. The color in subhead can be changed with that used in descriptive texts for visual quality. When changing the font color in subhead, grey color is recommended to be used as same with the bullet. 4. Characters or graphics/images should not be overlapped on global strings. 5. It is recommended to indicate the total slide page numbers by manual using slide master function as shown in this template.