SlideShare a Scribd company logo
MRI Made Simple HH Schild
History Felix Bloch & Edward Purcell Discovered magnetic resonance phenomenon in 1946 Received Nobel Prize in 1952 Used for chemical and physical molecule analysis Raymond Damadian in 1971 showed magnetic relaxation of tumor and normal tissue differed
MRI  Patient is put in a magnet RF signal is sent RF signal is switched off Patient emits a signal which is used for  Reconstruction of image
Proton are little magnet Proton  spin  around its axis like a planet, has positive changer and electrical field around it, and moving electrical field has a magnetic field like a bar magnet
Proton alignment Protons align to external magnetic field parallel or anti-parallel. Both have different energy level Like walking on feet require less energy than on hand, parallel state is preferred to anti-parallel Difference is 10,000,000/10,000,007
Precession  Proton moves in like spinning top In two axis wobbling motion called Precession, depends on magnetic  field strength
Precession frequency Precession frequency is dependent on strength of external magnet field It is determined by Larmor Equation Wo = yBo Wo is precession frequency in Hz or MHz Bo in magnetic field strength in Tesla y is gyro-magnetic ratio, for proton is 42.5MHz/Tesla Stronger the external magnetic field higher the precession frequency
Coordinate system Representation of magnetic force in Z axis, Proton vector as red arrow
Net magnetic force Proton pointing in opposite direction cancels each others magnetic effect in respective direction. 9 proton align up and 5 down, resulting in 4 proton up force
Net magnetic force As there are more protons aligned parallel to the external magnetic filed, there is a net magnetic movement aligned with or longitudinal to the external magnetic field
Human magnetic vector In a strong external magnetic field a new magnetic vector is induced in the patient, who becomes a magnet himself. This new magnetic vector is aligned with the external magnetic field
Measuring magnetization Magnetization along an external magnetic field cannot be measured For this a magnetization transverse to the external magnetic field is  necessary
RF pulse frequency After protons aligned along the external magnetic force, RF pulse is  send for energy exchange. Which is only possible if RF pulse has same  frequency as protons precession frequency
Effect of RF pulse The RF pulse exchange energy with the protons (a), and some of them are lifted to a higher level of energy, pointing down (b), in effect the magnetization along the z-axis decreases, as the protons which point down neutralize the same number of proton pointing up
Whip like action of RF pulse
Effect of RF pulse It lifts some protons to higher energy ( they point down), and it also causes the protons to precess in step, in phase. The former result in decreasing the magnetization along the Z-axis, the longitudinal magnetization The later establishes a new magnetization in the x-y-phase (->), a new transversal magnetization, which moves around with the precessing protons A ship, passengers being distributed randomly all over the deck, the ship in a normal position When all passengers walk in equal step around the railing The ship is leaning towards the side where the people are, a new force is established, and become visible
RF pulse on proton In a strong external magnetic field a new magnetic vector along the external field is established Sending in an RF pulse causes a new transveral magnetization while longitudinal magnetization decreases Depending on the RF pulse , longitudinal magnetization may even totally disappear
Result of transverse magnetization The new transversal magnetization moves around with the precessing protons Thus for an external observer, transversal magnetization constantly changes its direction, and can induce a signal in an antenna
Longitudinal relaxation After the RF pulse is switched off, protons go back from their higher to the lower state of energy. Longitudinal magnetization increases and grows back to its original value Energy of RF pulse is handed over to the surrounding lattice This process is call longitudinal relaxation or spin lattice relaxation
T1 Curve Recovery of longitudinal relaxation vs. time curve
Transverse relaxation After the RF pulse is switched off, protons lose phase coherence, they get out of step.
T2 Curve Transverse magnetization vs. time after RF pulse is switched off
T1 > T2  Coupling of a T1 and T2- curve resembles a mountain with a slope T1- 300-2000 msec. T2- 30-150 msec. It takes longer to climb a mountain than to slide or jump down.
Relaxation time It is difficult to pinpoint the end of the longitudinal and transverse relaxation exactly. Thus T1 and T2 were not defined as the time when relaxation is completed.  Instead T1 was defined as the time when about 63% of the original longitudinal magnetization is reached. T2 is the time when transverse magnetization decreases to 37% of the original value.  These percentages are derived equation 1-1/3 = 63%, 1/e = 37% 1/T1 is also called longitudinal relaxation rate 1/T2 transverse relaxation rate
Liquids have long T1 & T2
Fat has short T1 & T2
What influences T1/T2? T1 is longer than T2 T1 varies with the magnetic field strength It is longer in strong magnetic field Water has a long T1 and fat has a short T1 T2 of water is longer than the T2 of impure liquids containing larger molecules.
90 degree RF pulse If after the RF pulse, the number of protons on the higher energy level equals the number of protons on the lower energy level, longitudinal magnetization has disappeared, and there is only transverse magnetization due to phase coherence. The magnetic vector seems to have been tilted 90degree to the side The corresponding RF pulse is thus also called a 90degree pulse
90degree RF pulse Before RF pulse After 90d RF pulse, longitudinal magnetization decreases, proton also precess in phase, causes new transversal magnetization  After the RF pulse is switched off longitudinal magnetization increases, and transversal magnetization disappears,
Magnetizing vector Before RF pulse there is only longitudinal magnetization After the 90dgr RF pulse there is only transversal magnetization and this is spinning around With time after the removal of RF pulse the transversal magnetization decreases and longitudinal magnetization increases in spiral motion
Free induction decay The sum magnetic vector constantly, its direction and magnitude, while it performs its spiraling motion The sum vector induces an electrical current in an antenna, the MR signal. This is greatest immediately after RF signal is switched off and then decreases The signal intensity reduces but frequency is constant  This is called free induction decay (FID) signal
Long and Short TR A and B are two tissue with different relaxation times. Frame 0 shows the situation before, frame 1 immediately after a 90degree pulse. When we wait for a long time (TR long) the longitudinal magnetization of both tissue will have totally recovered (frame 5). A second 90d pulse after this time results in the same amount of transversal magnetization (frame 6) for both tissue. When we do not wait as long as in a, but send a second RF pulse after a shorter time (TR short), longitudinal magnetization of tissue B which has the longer T1, has not recovered as much as that of tissue A with shorter T1. The transversal magnetization of the two tissue after the second RF pulse will then be different (frame 5). Thus by changing the time between successive RF pulse we can influence and modify magnetization and the signal intensity of tissue.
Tissue signal with short TR Brain has a shorter longitudinal relaxation time than CSF. With a short TR the signal intensities of brain CSF differ more than after a long TR
T2 - Weighted image After the RF pulse is switched off, the protons dephase (a-c).  The 180deg pulse causes them to precess in the opposite direction and so they rephase again (d -f)
180 turn  When a rabbit and a turtle run in one direction for a certain time, then turn around and run in the opposite direction with the same speed for the same time they will arrive at the starting point at the same time.
T2 effect The 180 0  pulse refocuses the dephasing protons which results in a stronger signal, the spin echo after the time TE. The protons then dephase again and can be refocussed another time by a 180 0  pulse and so on. Thus it is possible to obtain more than one signal, more than one spin echo. The spin echo, however differ in intensity due to so-called T 2  – effect. A curve connecting the spin echo intensities is the T 2  curve. If we did not use the 180 0  pulse, the signal intensity would decay much faster. A curve describing the signal intensity in that case the T * 2  (T 2  star) curve.
T2-effect If we record the signal of two buses as the buses drives away. The signal vanish due to extrinsic (bus speed) and intrinsic (exhaustion of the passengers)  Without having the bus come back (i.e. a 180 0  pulse), it is impossible to say whether a decrease in signal intensity is due to inherent tissue properties , or due to external influence, i.e. different bus speed.
T 2 - weighted sequence T2-curve of the two tissue with different transversal relaxation times Tissue A has a shorter T2 than tissue B, thus looses transversal magnetization faster. With short TE the difference in signal intensity is less pronounced than after a long TE
Spin Echo sequence The spin echo sequence consists of a 90 0  and a 180 0  pulse After the 90 0  pulse protons are dephasing due to external and internal magnetic fields inhomogeneities The 180 0  pulse rephase the dephasing protons, and stronger signal , the spin echo results. The 180 0  pulse serve to “neutralize” the external magnetic field inhomogeneities Signal decrease from one echo to the next, when using multiple 180 0  pulse is due to internal T2-effects By choosing  different TEs this signal can be T2 weighted in varying degrees with very short TEs, T2-effects have not yet had time to really show up With longer TEs, the signal intensity difference between tissue will be depending very much on their T2s , their transversal relaxation times With very long TEs, there should be even more T2-weighing, however , signal intensity as such would be so small, that at best it can just barely be distinguished from the background noise
Spin Echo sequence (90 0  – TE/2 – 180 0  – TE/2 -> record signal at TE) after TR (time from the beginning of one 90 0  pulse) follows another pulse cycle and signal measurement. (90 0  – TE/2 – 180 0  – TE/2 -> record signal at TE)
 

More Related Content

PPTX
BASIC PHYSICS OF MRI AND ITS ADVANCEMENTS BY Joginder.pptx
PDF
Lecture12 mri
PPTX
Basics of mri
PDF
Magnetic Resonance Imaging
PPTX
Presentation1, mr pulse sequences.
PDF
Mri basics
PPTX
Mri basic principle and sequences
PPT
Mri physics
BASIC PHYSICS OF MRI AND ITS ADVANCEMENTS BY Joginder.pptx
Lecture12 mri
Basics of mri
Magnetic Resonance Imaging
Presentation1, mr pulse sequences.
Mri basics
Mri basic principle and sequences
Mri physics

What's hot (20)

PDF
MRI Physics
PPTX
Mri physics
PPTX
Basic principles of mri
PPT
MRI physics
PPTX
MRI (Magnetic resonance imaging)
PPTX
Basic of mri
PPTX
PARAMETERS AND ASSOCIATED Trade-Offs, MRI
PPTX
Encoding and image formation
PPTX
MR IMAGE QUALITY PARAMETERS.pptx
PPTX
Gradient Recalled Echo(GRE)
PPT
Mri image quality gamal mahdaly
PPTX
MRI - magnetic Imaging resonance
PDF
Basic principles of magnetic resonance imaging for beginner - Dr Sanjana Ravi...
PPTX
Basic Pulse Sequences In MRI
PPTX
Mri physics PART-1
PPTX
FACTORS AFFECTING THE SIGNAL-TO-NOISE RATIO.pptx
PPTX
Mri brain sequences
PPTX
Mri sequences
PPTX
Gradient echo pulse sequence swastik
PPTX
Basics of mri physics Dr. Muhammad Bin Zulfiqar
MRI Physics
Mri physics
Basic principles of mri
MRI physics
MRI (Magnetic resonance imaging)
Basic of mri
PARAMETERS AND ASSOCIATED Trade-Offs, MRI
Encoding and image formation
MR IMAGE QUALITY PARAMETERS.pptx
Gradient Recalled Echo(GRE)
Mri image quality gamal mahdaly
MRI - magnetic Imaging resonance
Basic principles of magnetic resonance imaging for beginner - Dr Sanjana Ravi...
Basic Pulse Sequences In MRI
Mri physics PART-1
FACTORS AFFECTING THE SIGNAL-TO-NOISE RATIO.pptx
Mri brain sequences
Mri sequences
Gradient echo pulse sequence swastik
Basics of mri physics Dr. Muhammad Bin Zulfiqar
Ad

Viewers also liked (6)

PPT
BASICS OF MRI
PPT
MRI sequences
PPT
Basics Of MRI
PPT
Magnetic Resonance Imaging
PPTX
MRI Procedure of Brain
BASICS OF MRI
MRI sequences
Basics Of MRI
Magnetic Resonance Imaging
MRI Procedure of Brain
Ad

Similar to Mri basics (20)

PDF
mribasicprinciples-19091912334002451.pdf
PPTX
Mri basic principles
PPTX
MRI PHYSICS FOR RADIOLOGY RESIDENTS INDIA
PPTX
MRI PHYSICS.pptx
PPTX
Basics of MRI Physics
PPT
Registry Part 3
PPTX
MRI-Chilkulwar.pptx
PPTX
MRI obstetric practice part 1 Basic Physics
PPTX
mriphysics-1610311406mmmmmmmmmmmm05.pptx
PPTX
Understanding mri in neonate
PPTX
mri physics.pptx
PPTX
MAGNETIC RESONANCE IMAGING PHYSICS.pptxx
PPTX
MRI PHYSICS.pptxMRI PHYSICS.pptxMRI PHYSICS.pptx
PPTX
The Physics of MRI
PDF
Magnetic Resonance Imaging: Basic Concepts and its applications in Dental Fie...
PPTX
Magnetic Resonance Imaging: Basic Concepts and its applications in Dental Fie...
PPTX
Mri for identifying types of fistulae
PPTX
MRI physics part 1: Basic principle by GKM
PPTX
Magnetic Resonance Imaging of Knee Diseases
mribasicprinciples-19091912334002451.pdf
Mri basic principles
MRI PHYSICS FOR RADIOLOGY RESIDENTS INDIA
MRI PHYSICS.pptx
Basics of MRI Physics
Registry Part 3
MRI-Chilkulwar.pptx
MRI obstetric practice part 1 Basic Physics
mriphysics-1610311406mmmmmmmmmmmm05.pptx
Understanding mri in neonate
mri physics.pptx
MAGNETIC RESONANCE IMAGING PHYSICS.pptxx
MRI PHYSICS.pptxMRI PHYSICS.pptxMRI PHYSICS.pptx
The Physics of MRI
Magnetic Resonance Imaging: Basic Concepts and its applications in Dental Fie...
Magnetic Resonance Imaging: Basic Concepts and its applications in Dental Fie...
Mri for identifying types of fistulae
MRI physics part 1: Basic principle by GKM
Magnetic Resonance Imaging of Knee Diseases

More from PS Deb (20)

PPTX
Lead poisoning in neurology
PPTX
Spinal cord disorders Anatomical Approach
PPTX
Should we allow natural death?
PPTX
Chronic progressive external ophthalmoplegia
PDF
Motivating hospital workers
PPTX
Brain stem 2014
PPTX
Muscle tone
PPTX
Cerebellum Anatomy and Physiology
PPTX
Corticospinal system
PPTX
Motor paralysis clinical
PPTX
Basal Ganglia Clinical Anatomy Physiology
PPTX
Myoclonus
PPTX
Athetosis and dystonia
PPTX
Tic disorder
PPTX
Chorea and ballismus
PPTX
Management of Tremor
PPTX
Hypertension and stroke
PPTX
Rapidly Progressive Fatal Neuromyositis
PPTX
Stroke management
PPTX
Neurologic manifestation of HIV/AIDS
Lead poisoning in neurology
Spinal cord disorders Anatomical Approach
Should we allow natural death?
Chronic progressive external ophthalmoplegia
Motivating hospital workers
Brain stem 2014
Muscle tone
Cerebellum Anatomy and Physiology
Corticospinal system
Motor paralysis clinical
Basal Ganglia Clinical Anatomy Physiology
Myoclonus
Athetosis and dystonia
Tic disorder
Chorea and ballismus
Management of Tremor
Hypertension and stroke
Rapidly Progressive Fatal Neuromyositis
Stroke management
Neurologic manifestation of HIV/AIDS

Mri basics

  • 1. MRI Made Simple HH Schild
  • 2. History Felix Bloch & Edward Purcell Discovered magnetic resonance phenomenon in 1946 Received Nobel Prize in 1952 Used for chemical and physical molecule analysis Raymond Damadian in 1971 showed magnetic relaxation of tumor and normal tissue differed
  • 3. MRI Patient is put in a magnet RF signal is sent RF signal is switched off Patient emits a signal which is used for Reconstruction of image
  • 4. Proton are little magnet Proton spin around its axis like a planet, has positive changer and electrical field around it, and moving electrical field has a magnetic field like a bar magnet
  • 5. Proton alignment Protons align to external magnetic field parallel or anti-parallel. Both have different energy level Like walking on feet require less energy than on hand, parallel state is preferred to anti-parallel Difference is 10,000,000/10,000,007
  • 6. Precession Proton moves in like spinning top In two axis wobbling motion called Precession, depends on magnetic field strength
  • 7. Precession frequency Precession frequency is dependent on strength of external magnet field It is determined by Larmor Equation Wo = yBo Wo is precession frequency in Hz or MHz Bo in magnetic field strength in Tesla y is gyro-magnetic ratio, for proton is 42.5MHz/Tesla Stronger the external magnetic field higher the precession frequency
  • 8. Coordinate system Representation of magnetic force in Z axis, Proton vector as red arrow
  • 9. Net magnetic force Proton pointing in opposite direction cancels each others magnetic effect in respective direction. 9 proton align up and 5 down, resulting in 4 proton up force
  • 10. Net magnetic force As there are more protons aligned parallel to the external magnetic filed, there is a net magnetic movement aligned with or longitudinal to the external magnetic field
  • 11. Human magnetic vector In a strong external magnetic field a new magnetic vector is induced in the patient, who becomes a magnet himself. This new magnetic vector is aligned with the external magnetic field
  • 12. Measuring magnetization Magnetization along an external magnetic field cannot be measured For this a magnetization transverse to the external magnetic field is necessary
  • 13. RF pulse frequency After protons aligned along the external magnetic force, RF pulse is send for energy exchange. Which is only possible if RF pulse has same frequency as protons precession frequency
  • 14. Effect of RF pulse The RF pulse exchange energy with the protons (a), and some of them are lifted to a higher level of energy, pointing down (b), in effect the magnetization along the z-axis decreases, as the protons which point down neutralize the same number of proton pointing up
  • 15. Whip like action of RF pulse
  • 16. Effect of RF pulse It lifts some protons to higher energy ( they point down), and it also causes the protons to precess in step, in phase. The former result in decreasing the magnetization along the Z-axis, the longitudinal magnetization The later establishes a new magnetization in the x-y-phase (->), a new transversal magnetization, which moves around with the precessing protons A ship, passengers being distributed randomly all over the deck, the ship in a normal position When all passengers walk in equal step around the railing The ship is leaning towards the side where the people are, a new force is established, and become visible
  • 17. RF pulse on proton In a strong external magnetic field a new magnetic vector along the external field is established Sending in an RF pulse causes a new transveral magnetization while longitudinal magnetization decreases Depending on the RF pulse , longitudinal magnetization may even totally disappear
  • 18. Result of transverse magnetization The new transversal magnetization moves around with the precessing protons Thus for an external observer, transversal magnetization constantly changes its direction, and can induce a signal in an antenna
  • 19. Longitudinal relaxation After the RF pulse is switched off, protons go back from their higher to the lower state of energy. Longitudinal magnetization increases and grows back to its original value Energy of RF pulse is handed over to the surrounding lattice This process is call longitudinal relaxation or spin lattice relaxation
  • 20. T1 Curve Recovery of longitudinal relaxation vs. time curve
  • 21. Transverse relaxation After the RF pulse is switched off, protons lose phase coherence, they get out of step.
  • 22. T2 Curve Transverse magnetization vs. time after RF pulse is switched off
  • 23. T1 > T2 Coupling of a T1 and T2- curve resembles a mountain with a slope T1- 300-2000 msec. T2- 30-150 msec. It takes longer to climb a mountain than to slide or jump down.
  • 24. Relaxation time It is difficult to pinpoint the end of the longitudinal and transverse relaxation exactly. Thus T1 and T2 were not defined as the time when relaxation is completed. Instead T1 was defined as the time when about 63% of the original longitudinal magnetization is reached. T2 is the time when transverse magnetization decreases to 37% of the original value. These percentages are derived equation 1-1/3 = 63%, 1/e = 37% 1/T1 is also called longitudinal relaxation rate 1/T2 transverse relaxation rate
  • 25. Liquids have long T1 & T2
  • 26. Fat has short T1 & T2
  • 27. What influences T1/T2? T1 is longer than T2 T1 varies with the magnetic field strength It is longer in strong magnetic field Water has a long T1 and fat has a short T1 T2 of water is longer than the T2 of impure liquids containing larger molecules.
  • 28. 90 degree RF pulse If after the RF pulse, the number of protons on the higher energy level equals the number of protons on the lower energy level, longitudinal magnetization has disappeared, and there is only transverse magnetization due to phase coherence. The magnetic vector seems to have been tilted 90degree to the side The corresponding RF pulse is thus also called a 90degree pulse
  • 29. 90degree RF pulse Before RF pulse After 90d RF pulse, longitudinal magnetization decreases, proton also precess in phase, causes new transversal magnetization After the RF pulse is switched off longitudinal magnetization increases, and transversal magnetization disappears,
  • 30. Magnetizing vector Before RF pulse there is only longitudinal magnetization After the 90dgr RF pulse there is only transversal magnetization and this is spinning around With time after the removal of RF pulse the transversal magnetization decreases and longitudinal magnetization increases in spiral motion
  • 31. Free induction decay The sum magnetic vector constantly, its direction and magnitude, while it performs its spiraling motion The sum vector induces an electrical current in an antenna, the MR signal. This is greatest immediately after RF signal is switched off and then decreases The signal intensity reduces but frequency is constant This is called free induction decay (FID) signal
  • 32. Long and Short TR A and B are two tissue with different relaxation times. Frame 0 shows the situation before, frame 1 immediately after a 90degree pulse. When we wait for a long time (TR long) the longitudinal magnetization of both tissue will have totally recovered (frame 5). A second 90d pulse after this time results in the same amount of transversal magnetization (frame 6) for both tissue. When we do not wait as long as in a, but send a second RF pulse after a shorter time (TR short), longitudinal magnetization of tissue B which has the longer T1, has not recovered as much as that of tissue A with shorter T1. The transversal magnetization of the two tissue after the second RF pulse will then be different (frame 5). Thus by changing the time between successive RF pulse we can influence and modify magnetization and the signal intensity of tissue.
  • 33. Tissue signal with short TR Brain has a shorter longitudinal relaxation time than CSF. With a short TR the signal intensities of brain CSF differ more than after a long TR
  • 34. T2 - Weighted image After the RF pulse is switched off, the protons dephase (a-c). The 180deg pulse causes them to precess in the opposite direction and so they rephase again (d -f)
  • 35. 180 turn When a rabbit and a turtle run in one direction for a certain time, then turn around and run in the opposite direction with the same speed for the same time they will arrive at the starting point at the same time.
  • 36. T2 effect The 180 0 pulse refocuses the dephasing protons which results in a stronger signal, the spin echo after the time TE. The protons then dephase again and can be refocussed another time by a 180 0 pulse and so on. Thus it is possible to obtain more than one signal, more than one spin echo. The spin echo, however differ in intensity due to so-called T 2 – effect. A curve connecting the spin echo intensities is the T 2 curve. If we did not use the 180 0 pulse, the signal intensity would decay much faster. A curve describing the signal intensity in that case the T * 2 (T 2 star) curve.
  • 37. T2-effect If we record the signal of two buses as the buses drives away. The signal vanish due to extrinsic (bus speed) and intrinsic (exhaustion of the passengers) Without having the bus come back (i.e. a 180 0 pulse), it is impossible to say whether a decrease in signal intensity is due to inherent tissue properties , or due to external influence, i.e. different bus speed.
  • 38. T 2 - weighted sequence T2-curve of the two tissue with different transversal relaxation times Tissue A has a shorter T2 than tissue B, thus looses transversal magnetization faster. With short TE the difference in signal intensity is less pronounced than after a long TE
  • 39. Spin Echo sequence The spin echo sequence consists of a 90 0 and a 180 0 pulse After the 90 0 pulse protons are dephasing due to external and internal magnetic fields inhomogeneities The 180 0 pulse rephase the dephasing protons, and stronger signal , the spin echo results. The 180 0 pulse serve to “neutralize” the external magnetic field inhomogeneities Signal decrease from one echo to the next, when using multiple 180 0 pulse is due to internal T2-effects By choosing different TEs this signal can be T2 weighted in varying degrees with very short TEs, T2-effects have not yet had time to really show up With longer TEs, the signal intensity difference between tissue will be depending very much on their T2s , their transversal relaxation times With very long TEs, there should be even more T2-weighing, however , signal intensity as such would be so small, that at best it can just barely be distinguished from the background noise
  • 40. Spin Echo sequence (90 0 – TE/2 – 180 0 – TE/2 -> record signal at TE) after TR (time from the beginning of one 90 0 pulse) follows another pulse cycle and signal measurement. (90 0 – TE/2 – 180 0 – TE/2 -> record signal at TE)
  • 41.  

Editor's Notes

  • #3: Magnetic resonance imaging (MRI) is an imaging technique used primarily in medical settings to produce high quality images of the inside of the human body. MRI is based on the principles of nuclear magnetic resonance (NMR), a spectroscopic technique used by scientists to obtain microscopic chemical and physical information about molecules. The technique was called magnetic resonance imaging rather than nuclear magnetic resonance imaging (NMRI) because of the negative connotations associated with the word nuclear in the late 1970's. MRI started out as a tomographic imaging technique, that is it produced an image of the NMR signal in a thin slice through the human body. MRI has advanced beyond a tomographic imaging technique to a volume imaging technique. This package presents a comprehensive picture of the basic principles of MRI. Before beginning a study of the science of MRI, it will be helpful to reflect on the brief history of MRI. Felix Bloch and Edward Purcell, both of whom were awarded the Nobel Prize in 1952, discovered the magnetic resonance phenomenon independently in 1946. In the period between 1950 and 1970, NMR was developed and used for chemical and physical molecular analysis. In 1971 Raymond Damadian showed that the nuclear magnetic relaxation times of tissues and tumors differed, thus motivating scientists to consider magnetic resonance for the detection of disease. In 1973 the x-ray-based computerized tomography (CT) was introduced by Hounsfield. This date is important to the MRI timeline because it showed hospitals were willing to spend large amounts of money for medical imaging hardware. Magnetic resonance imaging was first demonstrated on small test tube samples that same year by Paul Lauterbur. He used a back projection technique similar to that used in CT. In 1975 Richard Ernst proposed magnetic resonance imaging using phase and frequency encoding, and the Fourier Transform. This technique is the basis of current MRI techniques. A few years later, in 1977, Raymond Damadian demonstrated MRI of the whole body. In this same year, Peter Mansfield developed the echo-planar imaging (EPI) technique. This technique will be developed in later years to produce images at video rates (30 ms / image). Edelstein and coworkers demonstrated imaging of the body using Ernst's technique in 1980. A single image could be acquired in approximately five minutes by this technique. By 1986, the imaging time was reduced to about five seconds, without sacrificing too much image quality. The same year people were developing the NMR microscope, which allowed approximately 10 mm resolution on approximately one cm samples. In 1987 echo-planar imaging was used to perform real-time movie imaging of a single cardiac cycle. In this same year Charles Dumoulin was perfecting magnetic resonance angiography (MRA), which allowed imaging of flowing blood without the use of contrast agents. In 1991, Richard Ernst was rewarded for his achievements in pulsed Fourier Transform NMR and MRI with the Nobel Prize in Chemistry. In 1993 functional MRI (fMRI) was developed. This technique allows the mapping of the function of the various regions of the human brain. Six years earlier many clinicians thought echo-planar imaging's primary applications was to be in real-time cardiac imaging. The development of fMRI opened up a new application for EPI in mapping the regions of the brain responsible for thought and motor control. In 1994, researchers at the State University of New York at Stony Brook and Princeton University demonstrated the imaging of hyperpolarized 129Xe gas for respiration studies. MRI is clearly a young, but growing science.
  • #15: A radiofrequency pulse that has the same frequency as the precessing protons, can cause resonance, transfer energy to protons. This result in more protons being anti-parallel and thus neutralizing more protons in the opposite direction. Consequence the longitudinal magnetization decreases.
  • #16: The RF pulse also causes the protons to precess in synch, in phase this result in new magnetic vector the transverse magnetization.
  • #20: When the RF pulse is switched off 1. Longitudinal magnetization increases again; this longitudinal relaxation is described by a time constant T1, the longitudinal relaxation time
  • #21: The time that it takes for the longitudinal magnetization to recover, to go back to its original value , is described by the longitudinal relaxation time also called T1 . This actually is not the exact time it takes but a time constant describing how fast this process goes. This is like taking time for one round at a car race. The time gives you an idea of how long the race may take, but not the exact time. Or T1 is a time constant.
  • #22: Transverse magnetization decreases and disappears . This transversal relaxation is described by a time constants T2 the transversal relaxation time.
  • #23: Transverse relaxation time is also a time constant describing disappearance of transverse magnetization.
  • #25: Previously it was believed that measuring the relaxation times would give tissue characteristic result, and thus enable exact tissue typing. But there is quite some overlap of time range, and also T1 is dependent on the magnetic field strength used for examination. What is
  • #28: Actually, T1 depends on tissue composition, structure and surroundings. T1 relaxation is due to exchange of thermal energy to the surrounding, the lattice. The precessing protons have a magnetic field, that constantly changes directions, that constantly fluctuate with the Larmor frequency. The lattice also has its own magnetic fields. The protons now want to hand over energy over the lattice to relax. This can be done very effectively when the fluctuation of the magnetic fields in the lattice occur with a frequency. When the lattice consists of pure liquid/water, it is difficult for the protons to get rid of their energy, as a small water molecules move too rapidly. And the protons cannot hand over their energy over to the lattice quickly, they will only slowly go back to their lower energy level, their longitudinal alignment. Thus it takes a long time for the longitudinal magnetization to show up again, and this means that liquid /water have long T1s. When the lattice consists of medium size molecules that move and have fluctuating magnetic fields near the Larmor frequency of the precessing protons, energy can be transferred much faster, thus T1 is short. The carbon bonds at the ends of the fatty acids have frequencies near the Larmor frequency, thus resulting in effective energy transfer. It is easy to imagine that in a stronger magnetic field it takes more energy for the proton to align against it. Thus these protons have more energy to hand down to the lattice, and this takes longer than handing down just a small amount of energy. Though it is logical it is wrong. The precessing frequency depends on magnetic field strength. If we have a stronger magnetic field then the protons precess faster. And when they precess faster , they have more problem handling down the energy to a lattice with more slowly fluctuating magnetic fields. T2 relaxation comes about when protons get out of phase, which has two causes; inhomogeneities of the external magnetic field, and the local magnetic fields within the tissues. As water molecules move around very fast, their local magnetic fields fluctuate fast, thus kind of average each other out, so there are no big net differences in internal magnetic fields from place to place. And if there are no big difference in magnetic field strength inside of a tissue the protons stay in step for a long time, and so T2 is longer. With impure liquids, e.g. those containing some larger molecules, there are bigger variation in the local magnetic fields. The larger molecules do not move around as fast so their local magnetic fields do not cancel each other out as much. These larger difference s in local magnetic fields consequently cause larger differences in precessing frequencies, thus protons get out of phase faster T2 is shorter.
  • #30: We have 6protons pointing up; we send in a RF pulse, which lifts up 3 of them to higher energy level. The result we no longer nave a longitudinal, but a transversal magnetization When RF pulse is switched off protons go back to their lower state of energy and they lose phase coherence. Both process occur simultaneously and independently. First one proton goes back to the lower energy state resulting in 4 protons up and two pointing down. In net effect we now have a longitudinal magnetization of 2. Then the next proton goes back up, now 5 protons point up and one down resulting in a net longitudinal magnetization of 4. After the next proton goes up, longitudinal magnetization equals 6. At the same time, transversal magnetization decreases.
  • #32: A changing magnetic force can induce an electrical current, which was the signal that we receive and use in MR. If we put up an antenna we get a signal as illustrated If you think of the antenna as a microphone, and the sum magnetic vector as having a bell at its tip. The further the vector goes away from the microphone m the less loud the sound. The frequency of the sound however, remains the same because the sum vector spins with the precessing frequency. This type of signal is called FID signal from free induction decay. 90degree pulse gives a very good strong signal. The magnetic vector directly determine the MRI signal and signal intensity by inducing electrical currents in the antenna. Instead of term longitudinal and transversal magnetization we can also use the term “signal or signal intensity “ at the axis of our T1 and T2 curves.
  • #33: I figure a we have two tissue, A and B, which have different relaxation times (tissue A has a shorter transversal as well as longitudinal relaxation time). We send in a 90d RF pulse, and wait a certain time TR long. After the time TR long tissue A and tissue B have regained all of their longitudinal magnetization (frame 5), the transversal magnetization after the second pulse will be the same for both tissues, as tit was in frame 1. If we give RF pulse early, as in fig b, after frame 4 TR short. At this time tissue A has regained more of its longitudinal magnetization than tissue B. When the second 90d pulse now “tilts” the longitudinal magnetization 90d. The transversal magnetic vector of tissue A is larger than that of tissue B. And when this vector of A is larger , it will reach closer to our antenna; thus the imaginary bell at the tip of vector A will cause a louder, stronger signal in our “microphone”, the antenna, than vector B. The difference in signal intensity in this experiment depends on the difference in longitudinal magnetization, and this means on the difference in T1 between the tissue. Using these two pulses, we can now differentiate tissue A from tissue B, which might have been impossible choosing only one 90d pulse or two 90d pulses that are a long time apart. When you use more than one RF pulse –a succession of RF pulses – you use a so called pulse sequence . As you can use different pulses, e.g. 90d or 180d pulse, and the time intervals between successive pulses can be different, there can be many different pulse sequence. As we saw the choice of a pulse sequence will determine what kind of signal you get out of a tissue. So it is necessary to carefully chose and also describe the pulse sequence for a specific study. The pulse sequence that we used was made up of one type of pulse only , the 90d pulse . This was repeated after a certain time , which is called TR = time to repeat. With long TR we got similar signal from both tissues, both would appear the same on a MR picture. Using a shorter TR there was a difference in signal intensity between the tissues, determined by their difference in T1. The resulting picture is called T1 weighted picture. This means that the difference of signal intensity between tissues in that picture , the tissue contrast is mainly due to their difference in T1. A TR is less than 500mse is considered to be short, a TR greater than 1500 msec to be long.
  • #34: Why are the signals after a very long time TR between pulses not identical? When we wait a long time , T1 does not influence the tissue contrast any more, however, there may still be a possible difference in the proton density of the tissues in question. And when we wait a very long time TR in our experiment from above figure, the difference in signal is mainly due to different proton densities, we have a so called proton density, weighed image.
  • #35: After 90degree pulse, the longitudinal magnetization is tilted we get a transversal magnetization. After the pulse is switched off, longitudinal magnetization starts to reappear, the transversal magnetization however starts to disappear due to loss of proton phase coherence. Protons are in phase a, but increasingly spread out, as they have different precession frequencies b and c . The loss of phase coherence results in decreasing transversal magnetization and thus loss of signal. After a certain time (TE/2, half of TE) we send a 180degree pulse. The protons turn around and precess in exactly opposite direction. The result is that the faster precessing protons are now behind the slower ones. If we wait another time TE/2, the faster ones will have caught up with the slower ones. At that time the protons are nearly in phase again, which results in a stronger transversal magnetization, and thus in a stronger signal again. A little later however, the faster precessing protons will be ahead again, with the signal decreasing again.
  • #36: A race between a rabbit and a turtle starting at the same line. After a certain time (TE/2), the rabbit is ahead of the turtle. When you make the competitors run in the opposite direction for the same length of time , they will be both back at the starting line exactly the same time. The 180degree pulse in our experiment act like a wall, from which the protons bounce back , like a mountain reflecting sound wave as echoes. This is why the resulting strong signal is also called a echo or spin echo.
  • #37: After we have our signal , our spin echo, the protons lose phase coherence again, the faster ones getting ahead as we have heard. We naturally can perform the experiment again with another 180degree pulse , and another and another… If we plot time vs. signal intensity, we get a curve like above fig. From this curve we can see that the spin echo, the resulting signal, decreases with time. Responsible for this is the fact that out 180 0 pulse only “neutralizes” effects that influences the protons in a constant manner, and these are the constant inhomogeneities form local magnetic fields inside of the tissue cannot be “ evened out” , as they may influence some protons before the 180 0 pulse differently than after the 180 0 pulse. So some of the protons may still be behind or in front of the majority of the protons, that will reach the starting line at the same time. So from echo to echo the intensity of the signal goes down due to the so-called T 2 effects.
  • #38: Imagine two buses full of people ,e.g. after a soccer game. They are standing at a starting line. With two microphone , you record the signal (e.g. the signal from the crowd) that is coming from each bus . Then the bus leaves in the same direction . Recording the signal, you may recognize, that one signal disappears faster , than the other. This has two different causes : may be one bus drives faster than the other (loss of signal would thus be due to external influences, the external magnetic field inhomogeneities.). Or the difference in signal intensities, the difference in signal, may be due to difference of inherent properties of the two groups (internal inhomogeneities); may be in one bus, there are only the “party animals”, who do not become tired fast, as the people in the other crowd.. To figure out what actually is the reason for the signal disappearing , you can make the buses turn around after a certain time TE/2, and have them drive back with the same speed also for the time TE/2. After time 2 x TE/2 = TE the buses will be back at the starting line. The signal intensity, that you record with your microphone then depends only on inherent properties, i.e. how tired the crowd have become. If you do not use a 180 0 pulse to neutralize constant external inhomogeneities , the protons will experience larger difference in magnetic field strength when the RF pulse is switched off. Due to this they will be out of phase faster, the transversal relaxation time will be shorter. To distinguish this shorter transversal relaxation time from the T 2 After the 180 0 pulse, it is called T* 2 (T2 star). The corresponding effects are named T* 2 -effects are important with the so called fast imaging sequences. The type of pulse sequence, that we used in our experiment, is called a spin echo sequence, consisting of 90 0 pulse and a 180 0 pulse (causing the echo). This pulse sequence is very important in MR imaging, as it is the workhorse among the pulse sequences, which can be used for many things. It is important to realize that with a spin echo sequence, we cannot only produce T2-, but also T1- and proton density- weighted pictures.
  • #39: First we sent in a 90 0 pulse we have a maximum transversal magnetization. However , this transversal magnetization disappears, due to T 2 -effect. How fast transversal magnetization disappears can be seen from a T2-curve. Tissue A having a short T2(Brain), tissue B having a long T2 (Water). Both curve start at 0, which is the time immediately after the 90 0 pulse is switched off. When we wait for a certain time TE/2 to send the 180 0 pulse, transversal magnetization will have become smaller. After waiting another time TE/2 (that is TE after the 90 0 pulse is switched off ), we will receive a signal , the spin echo . The intensity of this echo is given by the T2 curve at the time TE. This time TE between the 90 0 pulse and spin echo is called TE = time to echo. The time TE can be chosen by the operator. And as we can see from the T 2 curve , TE influences the resulting signal and thus also the image. The shorter the TE , the stronger the signal that we get from a tissue. To get the best, strong signal, it may seem reasonable to use a short TE, because with longer TEs signal intensity decreases. With a short TE, however, there will be a problem. If we wait a very short TE, the difference in signal intensity between tissue A and tissue B is very small, both tissue may hardly be distinguished as there is hardly any contrast (which is the difference in signal intensity of tissue). Consequence; with a short TE, difference in T2 do not influence tissue contrast very much. As both T2 curves diverge, with a longer TE the difference in T2- curves and thus the difference in signal intensity = contrast, is more pronounced in our example. So it might be reasonable to wait a very long TE; the resulting picture should be very heavily T2-weighed. But if wait longer, the total signal intensity becomes smaller and smaller. The signal to noise ratio becomes smaller, the picture appears grainy
  • #40: A short TR is one that is about as short as the smallest/shortest T1 that we are interested in. a long TR is 3 times as long as a short TR. A TR of less then 500msec is considered to be short, a TR of more than 1500msec to be long. A short TE is one that is as short as possible, a long TE is about 3 times as long. A TE of less than 30 mesc is considered to be short, a TE greater than 80 msec to be long.