Pumps, Oxygenators and Priming
Solutions
Pumps in Cardiopulmonary Bypass
Ideal pump
• Move large volumes against significant pressure
(7 l/min against 500 mmHg)
• Minimize flow velocity- limit damage to blood
• Inert pump components – no activation of coagulation and inflammation
• Minimal dead space – avoid stagnation and turbulence
• Calibration - easy, reliable, and reproducible
• Manual operations possible
Positive displacement pump
• Periodic volumetric change of a working space
• Low volume but high pressure flow
Centrifugal pump
• Energy transfer due to velocity deflection
• High volume low pressure flow
Pumps - Classification
Roller pump – Working principle
 Contain a length of tubing inside a curved
raceway placed at the travel perimeter of
rollers mounted on the ends of rotating
arms
 “Rolls” blood through piece of tubing.
 Generates both negative and positive
pressures.
 Independent of resistance (afterload)
hydrostatic pressure head (preload)
Stroke volume and blood flow
Roller pump – Working principle
TUBING DIAMETER(IN) STROKE VOLUME(ML) BLOOD FLOW(L/MIN) AT 150
RPM
3/16 7 1050
1/4 13 1950
3/8 27 4050
1/2 54 8100
Advantages
Simple to use.
Low cost.
Preload and Afterload independent
Disadvantages
Spallation
Tube material fatigue
Blood cell damage
Pump air
Cavitation
Potential pressurisation of arterial line.
Non occlusive roller pumps
 Rhone – Poulenc in France.
 MC 3 Pump.
 Passive filling peristaltic pump.
 Combines advantages of centrifugal and roller pumps.
 Two sheets of flat polyurethane tubing bonded at edges which are
stretched under tension over three rollers.
 Completely flat pump chamber.
 No backing plate against which the tubing can be compressed.
 Priming volume 120 ml.
Non occlusive roller pumps – Working Principle
Non occlusive roller pumps – Working Principle
METAPLUS PUMP
•Pump position is fixed in relation to the hard shell venous reservoir.
•Large bore, semi-rigid, U-shaped tubing connects the outlet of venous reservoir to inlet of pump
PUMP ROTOR AND MOTOR
ASSEMBLY
VENOUS RESERVOIR AND
MEMBRANE OXYGENATOR
Advantages
Preload dependent – cannot suck air.
No retrograde flow when pump stopped.
Blood damage and microbubble generation is reduced as no negative
pressure generated.
Non occlusive nature – tubing wear reduced.
Centrifugal pump
• In early 1970s, research related to the development of an
artificial heart was basis of the development of centrifugal
pumps for CPB.
• Boimedicus 600 - 1973.
•In the United States, the centrifugal pump is extensively used.
Centrifugal Pump – Working Principle
 Creating pressure gradient between inlet and oulet of pump.
 This pressure gradient results from the creation of a vortex by the
rotation of the pump head.
 The vortex can be created by using cones that impart motion to the
blood by viscous shear or by rotating impellers.
 The rotating motion creates an area of low pressure in the center
and an area of high pressure on the sides.
Centrifugal Pump – Working Principle
Centrifugal Pump – Working Principle
• Resultant blood flow-
•The resistance at the outlet is a function of two components: the CPB circuit
and SVR.
•Centrifugal pumps are afterload dependent and flow is influenced by
changes in resistance in both the circuit and the patient
•Flow meter necessary.
Centrifugal force
Centrifugal Pump – Working Principle
Centrifugal Pump – Working Principle
•HEAT GENERATION: All centrifugal pumps will generate heat
depending on the amount of energy that is impaired into the
blood.
•Combination with the low flow in the center of the pump head -
may create blood clots and blood cell activation in the pump.
SPECIFIC CLINICALLY AVAILABLE CENTRIFUGAL PUMPS
•BIOMEDICUS PUMP
• In 1976, the first centrifugal pump was used for CPB.
• The pump head is acrylic, with inlet and outlet ports oriented at right angles to each other, and its
priming volume is 80 ml.
• Cones driven by magnetic coupling to external console.
CENTRIFUGAL PUMPS - Capiox pump
Rotor with unique straight path
design to reduce pump
rotational speed without
decreasing hydraulic efficiency
Small priming volume – 46 ml
reduces stagnant flow within
the rotor
CENTRIFUGAL PUMPS – Nikkiso Pump
•Smallest commercially available pump
•Priming volume of 25 ml.
•Made of polycarbonate, with a V shaped ring seal that
separates the pump housing and the actuator chamber.
•Seal – made of fluororubber, suppresses heat generation and
prevents blood leakage.
•Six washout holes are incorporated into impeller to generate
blood flow from the back to front surface of the impeller.
•These holes prevent thrombus formation in areas behind the
impeller and around the sealing part.
Problems associated with centrifugal pumps
Flow rate affected by preload and after load.
Retrograde flows down the arterial line.
Potential air entrapment if inadequate aferload.
For forward flow pressure in the pump head (PP) must be greater than the
combined patient pressure (PPT) and the pressure head (PH) {hydrostatic
pressure}.
--PP > PPT + PH forward flow
Centrifugal Pump vs Roller Pump
Expensive Inexpensive
Pump flow function of SVR Flow predictable based on pump speed
Cannot pump large amount of air Can pump large amount of air
Pump stalls on occlusion without generating
high suction or outlet pressure.
Potential to overpressurize circuit if
inadvertently clamped.
Retrograde flow when pump slows / stops No retrograde flow.
Does not require continous monitoring Continuous strict monitoring of blood level
Hemolysis and damage to formed blood
elements is less
More hemolysis and damage to formed blood
elements.
Less wear and tear of pump. No spallation More wear and tear of tubing in pump head.
Spallation.
Oxygenator
 Blood gas exchange device.
Oxygenator
Oxygenate venous blood.
Remove CO2
Represent the largest surface area to which circulating blood is exposed.
Components -
Membrane module
Heat exchanger
Reservoir
AN IDEAL OXYGENATOR
Oxygenation of venous blood: device must have sufficient capacity to
provide oxygenation over a wide range of venous flow rate.
Carbon dioxide elimination to avoid hypercarbia or hypocarbia.
Minimum trauma to the blood
Small priming volume - to limit the deleterious effects of hemodilution
Safety
CLASSIFICATION
Bubble oxygenator:
The earliest oxygenators.
Exchange gases through direct interaction of gas and blood.
These devices were used during early advent of CPB.
Membrane oxygenator
Semi-permeable barrier that separates fluid from gas.
Diffusive qualities of the membrane material determine the
transfer of oxygen and carbon dioxide between phases.
BUBBLE OXYGENATORS
• First widely available commercial oxygenators
• Structure : 3 sections of operation
Bubble column
Defoaming area
Arterial reservoir
• Desaturated blood passively enters mixing chamber, where 100%
oxygen flows across a disparager plate into the stream of blood,
which forms small bubbles
BUBBLE OXYGENATORS
Blood becomes oxygenated and carbon dioxide is reduced
as stream of gas percolates through blood.
Blood is defoamed by the presence of silicone antifoam-A,
which consist of the liquid polymer dimethylpolysiloxane
(96%) and particulate silica (4%), which destabilizes the
bubbles, causing them to implode.
BUBBLE OXYGENATORS
The arterialized blood is collected in an arterial reservoir that is
then actively pumped.
The simple design of bubble oxygenators relies on the
hydrostatic pressure head from the patient to the mixing
chamber connected by the venous line.
The pressure drop through bubble oxygenator is <30 cm of
water, in contrast to the 100 cm of water pressure drop
typically found in membrane oxygenators.
BUBBLE OXYGENATORS
Bubble size is critical to adequate gas
transfer.
The bubble size selected must be a
compromise between optimal surface
area for oxygenation and volume for
carbon dioxide transfer.
Decreasing size of bubbles increases total
surface area of blood gas interface with
better oxygenation but limiting total CO2
transfer.
Bubble sizes of 3 to 7 mm are used to
optimize both O2 and CO2 transfer.
MEMBRANE OXYGENATORS
• Complete barrier between the gas and blood phases and diffusion is through membrane
material
• Costly to manufacture and require large priming volume
• Most membrane lungs used for CPB have micropores
Willem J. Kolff
During dialysis in 1943 noticed that the
blue blood in the rotating-drum
artificial kidney became red
Clowes and Neville(1958) Poineers in
using membrane oxygenators (teflon
flat membranes)
MEMBRANE OXYGENATORS – Materials
Historically – Cellulose.
Polytetrafluoroethylene
Polyethylene
Currently -
Silicon rubber (homogeneous, nonporous membrane).
Polypropylene (heterogeneous, microporous, hydrophobic
membrane).
Silicone Vs Polypropylene Membrane
Silicone membrane
Long term support
Without a diminution in gas transfer capacity
Avoid plasma leakage and membrane wet out
Microporous polypropylene membrane
Cheap
Good for short periods
New generation membranes that incorporates benefits of
silicone with polypropylene have been developed.
MEMBRANE OXYGENATORS – Designs
Membrane materials are organized in three configurations:
Scrolled envelope
Parallel plate
Hollow fiber
Types of Membrane Oxygenators
Plaque oxygenators
-microporous expanded polypropylene
-folded Z shape
-blood & gas flow opposite direction
-Cobel Excel, Cobe VPCML,Shirley M2000
Spiral oxygenators
-silicon membranes
-rolled around central axis
-Kolobow oxygenator.
Types of Membrane Oxygenators
Hollow fibre oxygenator
1970 Benlips introduced
Capillary fibers of microporous polypropylene
Spiral Membrane Oxygenators
Kolobow
Silicon membrane in shape of an envelope that is coiled on
itself.
Used primarily in ECMO
Ability to maintain stable CO2 and O2 for long periods
(weeks).
Available in gas exchange surface area sizes from 0.5 to 4.5
m2.
Hollow Membrane Oxygenators
Blood flow inside the capillaries
Gas flow inside the capillaries
Blood flow through the fiber was abandoned -
High trans membrane pressure
Activation of platelets
Increased haemolysis
Blood flow either perpendicular or in the direction of fiber bundle
In latter case, blood will flow in a counter current direction to the gas
flow - Optimized gas gradients during the dwell time
Membrane oxygenators &
lung
MEMB OXYGN LUNG
SURFACE AREA(M2) 0.5-4 70
Blood path width (µm) 200 8
Blood path length (µm) 250000 200
Memrane thickness (µm) 150 0.5
Max O2 transfer(ml/min) 400-600 2000
ADVANCES IN OXYGENATORS
Biocompatibility
Heparin coating – 1980's
Surface Modifying Additive—
Polydimethylsiloxane polycaprolactone oligomer
Phosphorylcholine
Bioline
X Coating
Trillium Biopassive Surface—polyethylene oxide.
Cost-effectiveness'????
Priming Solutions
Priming solutions for CPB Circuit
 Need of prime – to achieve adequate flow rates on initiation of
CPB without air embolism.
 Ideal prime -
Similar electrolyte content, osmolarity and pH as
that of plasma.
On mixing with blood maintains oxygen delivery, CO2 removal and
physiological homeostasis.
Historical perspective of prime solutions
Homologous blood syndrome
 Blood borne infection
 Severe pulmonary insufficiency
 Impaired immunity – wound infection, sepsis
 Impaired resistance to Malignant cell transformation
 Graft versus host disease
Impetus for nonhemic prime
Severe strain on hospital blood bank
Increased access to the emergency surgery
Increased exposure to O2 in polycythemics
Refusal of hemic prime from Jehovah’s witness faith
Experimental success of hemodilution in CPB
Glucose in priming solution
1962, Cooley – 5% dextrose in addition to blood - improves outcome
Solution with glucose as major component – isotonic but after
metabolisation of glucose become severely hypotonic
Fluid shift from Extracellular to Intracellular compartment -
Red blood cell lysis
Pulmonary edema
Cerebral edema
Hyperglycemia – poor neurological outcome
Priming solution – Normotonic, near physiologic sodium concentration.
Colloidal Priming solution
 Hemodilution – decreased colloid oncotic pressure – fluid
shift into intracellular compartment – cellular edema and
dysfunction.
 Colloid solution - counteract reduction in colloid oncotic
pressure – prevent fluid shift.
 CPB – systemic inflammatory response – tight junctions at
endothelial lining "permeable" to high molecular weight
proteins – high molecular weight protein trapped in ECF –
paradoxical increase in cellular edema.
Common priming solutions
Additives to priming solution
Experimental prime solutions
Perfluoro carbons
- 0.118 microns, half the viscocity of blood
-O2 release even at low po2 environments
-O2 relase is not related to pH/ temp
- can perfuse distal capillareis
- still at experimental stage
Stroma free Hb
-natural O2 carrying capacity &osmotic activity
-lower viscocity than blood
-do not cause immunosuppression
-still in preclinical testing
Priming Solution (AIIMS)
Ringer lactate – 20 ml/kg.
Hydroxy Ethyl Starch – 10 ml/kg.
Mannitol – 5 ml/kg.
Soda bicarbonate – 1 ml/kg.
Heparin – on the basis of circuit used.
If blood is added to prime – additional soda bicarbonate 10ml/300ml blood
is added.
Mannitol
 Potent osmotic diuretic.
 Maintain urine output during CPB and in immediate
post bypass period.
 Preserves renal function.
 Free radical scavenger.
Oxygenator-heat exchanger
Heat exchanger
Membrane oxygenator
Pump tubings
Internal diameter
(inch)
Volume
(ml/feet)
Used as
1/4 8.6 Suction / vent
<6kg – art./venous line
3/8 21.6 <10 kg – venous line.
>11 kg – arterial line
1/2 37.0 > 20 kgVenous line
CPB circuit according to age
Weight
(kg)
Oxygenator Prime vol
(ml)
Art. line Venous line Circuit
Priming
vol. (ml)
Heparin
(mg)
Max. Flow
(lt/min)
< 6 Baby Rx 35 1/4 1/4 400 - 500 25 1.2
6 - 10 Minimax 109 1/4 1/4 or 3/8 400 –600 25 - 30 2.3
11 - 20 Sx 10 135 3/8 3/8 1000 - 1100 50 3.5
>20 Affinity NT 290 3/8 1/2 1500 - 1700 75 7
Priming volume
 Volume required to fill the arterial and venous limbs, adequate
voulme in reservoir to prevent air entering the arterial line on
initiation of CPB.
 Acceptable hemodilution??
Calculation of Blood Volume to be added
Pts estimated blood volume x Hct
Predicted Hct= ____________________________
Pts estimated blood vol +CPB prime+
pre CPB iv fluid volume (TCV)
RBC Vol. To be added = TCV (Hct Desired – Hct Predicted)
Bank Blood volume to be added = RBC Vol to be added / 0.7
Blood preservation techniques
Pre CPB – Retrograde circulation
Adult
Hct > 32%
CVP > 8 mm hg.
SBP > 80 mm hg
On CPB – Hemofilter.
Cell Saver.
Post CPB – Chase prime with crystalloid.
Completely reverse and pack
Thank you

More Related Content

PPTX
Basics of cpb
PPTX
Oxygenators
PPTX
CARDIOPULMONARY BYPASS
PPTX
Conduction system of the heart
PPTX
Blood pump perfusion
PPT
Oxygenators
PPTX
History of cardiopulmonary bypass
PPTX
CPB(CARDIO PULMONARY BYPASS)
Basics of cpb
Oxygenators
CARDIOPULMONARY BYPASS
Conduction system of the heart
Blood pump perfusion
Oxygenators
History of cardiopulmonary bypass
CPB(CARDIO PULMONARY BYPASS)

What's hot (20)

PPSX
Ultrafiltration during cardiopulmonary_bypass
PPTX
Cpb oxygenators DR NIKUNJ R SHEKHADA (MBBS,MS GEN SURG ,DNB CTS SR)
PPT
cannulation techniques during cpb
PPTX
Roller pump
PPTX
Heat exchanger
PPTX
CARDIO PULMONARYBYPASS BLOOD PUMPS DR NIKUNJ R SHEKHADA (MBBS,MSGEN SURG ,DN...
PDF
Emergencies in cpb
PPT
10.heat transfer during cpb
PPTX
Blood pumps
PPTX
ALPHA STAT & PH STAT.pptx
PPTX
Cardiopulmonary bypass
PDF
Hypothermia, DHCA, RCP, ACP,Oxygen consumption,Cooling, Rewarming.
PPTX
Basic principles of myocardial proctection
PPTX
Cannulation and cardiopulmonary bypass.
PPTX
DHCA
PPT
7 Adequacy Of Perfusion During Cardiopulmonary Bypass
PPT
Routine cpb weaning
PPT
Ecmo presentation final
PPT
Cardiopulmonary bypass
PPTX
Anticoagulation and haemostasis during cardiopulmonary bypass
Ultrafiltration during cardiopulmonary_bypass
Cpb oxygenators DR NIKUNJ R SHEKHADA (MBBS,MS GEN SURG ,DNB CTS SR)
cannulation techniques during cpb
Roller pump
Heat exchanger
CARDIO PULMONARYBYPASS BLOOD PUMPS DR NIKUNJ R SHEKHADA (MBBS,MSGEN SURG ,DN...
Emergencies in cpb
10.heat transfer during cpb
Blood pumps
ALPHA STAT & PH STAT.pptx
Cardiopulmonary bypass
Hypothermia, DHCA, RCP, ACP,Oxygen consumption,Cooling, Rewarming.
Basic principles of myocardial proctection
Cannulation and cardiopulmonary bypass.
DHCA
7 Adequacy Of Perfusion During Cardiopulmonary Bypass
Routine cpb weaning
Ecmo presentation final
Cardiopulmonary bypass
Anticoagulation and haemostasis during cardiopulmonary bypass
Ad

Similar to Pumps, oxygenators and priming solution (20)

PPTX
Sinz cpb
PDF
oxygenators-copy-200308141605.pdf
PDF
Unit 1.pdf
PPTX
Cardio Pulmonary Bypass Machine Hardware and Primes.pptx
PPTX
PRESENTATION.pptx
PPTX
Pumps and its operation characteristic and performance
PDF
Heart lung machine.pdf
PDF
Vacuum system jn thermal power system to use
PPTX
radial piston pump
PDF
heartlungmachine-141115052523-conversion-gate01.pdf
PDF
heartlungmachine-141115052523-conversion-gate01.pdf
PPSX
Heart–lung machine
PDF
ilide.info-centrifugal-pump-pr_1aa9ede45af0412df3436cf39846eaa3.pdf
PDF
8. Reciprocating compressors.pdf
PDF
Babatunde Mobisola - Accumulator Systems
PPTX
Babatunde Mobisola - Accumulator Systems
PPTX
Physiology of Cardiopulmonary Bypass
PPTX
PPT
compressors_pumps course
PDF
Energy Savings in Industrial Water Pumping Systems
Sinz cpb
oxygenators-copy-200308141605.pdf
Unit 1.pdf
Cardio Pulmonary Bypass Machine Hardware and Primes.pptx
PRESENTATION.pptx
Pumps and its operation characteristic and performance
Heart lung machine.pdf
Vacuum system jn thermal power system to use
radial piston pump
heartlungmachine-141115052523-conversion-gate01.pdf
heartlungmachine-141115052523-conversion-gate01.pdf
Heart–lung machine
ilide.info-centrifugal-pump-pr_1aa9ede45af0412df3436cf39846eaa3.pdf
8. Reciprocating compressors.pdf
Babatunde Mobisola - Accumulator Systems
Babatunde Mobisola - Accumulator Systems
Physiology of Cardiopulmonary Bypass
compressors_pumps course
Energy Savings in Industrial Water Pumping Systems
Ad

More from India CTVS (20)

PPTX
Weaning from MECHANICAL VENTILATION
PPTX
Tracheostomy
PPTX
Protocol for ventilator settings
PPTX
Infective endocardiitis
PPTX
Management of cc tga
PPTX
Iabp presentation
PPTX
Coagulation monitoring and teg
PPT
Coronary artery anomalies in chd
PPTX
Adult ecmo
PPTX
Evolution of management stratergy for TGA
PPTX
Heart transplant guidelines
PPTX
HYPOPLASTIC LEFT HEART SYNDROME
PPTX
Blood presentation
PPTX
Tga management
PPTX
Pediatric ecmo
PPTX
Management of tapvc
PPTX
Constrictive pericarditis pathophysiology
PPTX
Mechanical heart valve substitutes
PPTX
Bioprostheic heart valve prosthesis
PPTX
PATHOPHYSIOLOGY ,NATURAL HISTORY OF VSD
Weaning from MECHANICAL VENTILATION
Tracheostomy
Protocol for ventilator settings
Infective endocardiitis
Management of cc tga
Iabp presentation
Coagulation monitoring and teg
Coronary artery anomalies in chd
Adult ecmo
Evolution of management stratergy for TGA
Heart transplant guidelines
HYPOPLASTIC LEFT HEART SYNDROME
Blood presentation
Tga management
Pediatric ecmo
Management of tapvc
Constrictive pericarditis pathophysiology
Mechanical heart valve substitutes
Bioprostheic heart valve prosthesis
PATHOPHYSIOLOGY ,NATURAL HISTORY OF VSD

Recently uploaded (20)

PPTX
Electromyography (EMG) in Physiotherapy: Principles, Procedure & Clinical App...
PPTX
POLYCYSTIC OVARIAN SYNDROME.pptx by Dr( med) Charles Amoateng
PPTX
CME 2 Acute Chest Pain preentation for education
PPT
MENTAL HEALTH - NOTES.ppt for nursing students
PPTX
Uterus anatomy embryology, and clinical aspects
PPT
Breast Cancer management for medicsl student.ppt
PPTX
Acid Base Disorders educational power point.pptx
PPTX
post stroke aphasia rehabilitation physician
PDF
Deadly Stampede at Yaounde’s Olembe Stadium Forensic.pdf
PPTX
1 General Principles of Radiotherapy.pptx
PPTX
ACID BASE management, base deficit correction
PPTX
CEREBROVASCULAR DISORDER.POWERPOINT PRESENTATIONx
PPT
ASRH Presentation for students and teachers 2770633.ppt
PDF
NEET PG 2025 | 200 High-Yield Recall Topics Across All Subjects
PPTX
Note on Abortion.pptx for the student note
PPTX
History and examination of abdomen, & pelvis .pptx
PPTX
Respiratory drugs, drugs acting on the respi system
PPTX
neonatal infection(7392992y282939y5.pptx
PDF
Intl J Gynecology Obste - 2021 - Melamed - FIGO International Federation o...
DOCX
RUHS II MBBS Microbiology Paper-II with Answer Key | 6th August 2025 (New Sch...
Electromyography (EMG) in Physiotherapy: Principles, Procedure & Clinical App...
POLYCYSTIC OVARIAN SYNDROME.pptx by Dr( med) Charles Amoateng
CME 2 Acute Chest Pain preentation for education
MENTAL HEALTH - NOTES.ppt for nursing students
Uterus anatomy embryology, and clinical aspects
Breast Cancer management for medicsl student.ppt
Acid Base Disorders educational power point.pptx
post stroke aphasia rehabilitation physician
Deadly Stampede at Yaounde’s Olembe Stadium Forensic.pdf
1 General Principles of Radiotherapy.pptx
ACID BASE management, base deficit correction
CEREBROVASCULAR DISORDER.POWERPOINT PRESENTATIONx
ASRH Presentation for students and teachers 2770633.ppt
NEET PG 2025 | 200 High-Yield Recall Topics Across All Subjects
Note on Abortion.pptx for the student note
History and examination of abdomen, & pelvis .pptx
Respiratory drugs, drugs acting on the respi system
neonatal infection(7392992y282939y5.pptx
Intl J Gynecology Obste - 2021 - Melamed - FIGO International Federation o...
RUHS II MBBS Microbiology Paper-II with Answer Key | 6th August 2025 (New Sch...

Pumps, oxygenators and priming solution

  • 1. Pumps, Oxygenators and Priming Solutions
  • 3. Ideal pump • Move large volumes against significant pressure (7 l/min against 500 mmHg) • Minimize flow velocity- limit damage to blood • Inert pump components – no activation of coagulation and inflammation • Minimal dead space – avoid stagnation and turbulence • Calibration - easy, reliable, and reproducible • Manual operations possible
  • 4. Positive displacement pump • Periodic volumetric change of a working space • Low volume but high pressure flow Centrifugal pump • Energy transfer due to velocity deflection • High volume low pressure flow Pumps - Classification
  • 5. Roller pump – Working principle  Contain a length of tubing inside a curved raceway placed at the travel perimeter of rollers mounted on the ends of rotating arms  “Rolls” blood through piece of tubing.  Generates both negative and positive pressures.  Independent of resistance (afterload) hydrostatic pressure head (preload)
  • 6. Stroke volume and blood flow Roller pump – Working principle TUBING DIAMETER(IN) STROKE VOLUME(ML) BLOOD FLOW(L/MIN) AT 150 RPM 3/16 7 1050 1/4 13 1950 3/8 27 4050 1/2 54 8100
  • 7. Advantages Simple to use. Low cost. Preload and Afterload independent
  • 8. Disadvantages Spallation Tube material fatigue Blood cell damage Pump air Cavitation Potential pressurisation of arterial line.
  • 9. Non occlusive roller pumps  Rhone – Poulenc in France.  MC 3 Pump.  Passive filling peristaltic pump.  Combines advantages of centrifugal and roller pumps.  Two sheets of flat polyurethane tubing bonded at edges which are stretched under tension over three rollers.  Completely flat pump chamber.  No backing plate against which the tubing can be compressed.  Priming volume 120 ml.
  • 10. Non occlusive roller pumps – Working Principle
  • 11. Non occlusive roller pumps – Working Principle
  • 12. METAPLUS PUMP •Pump position is fixed in relation to the hard shell venous reservoir. •Large bore, semi-rigid, U-shaped tubing connects the outlet of venous reservoir to inlet of pump PUMP ROTOR AND MOTOR ASSEMBLY VENOUS RESERVOIR AND MEMBRANE OXYGENATOR
  • 13. Advantages Preload dependent – cannot suck air. No retrograde flow when pump stopped. Blood damage and microbubble generation is reduced as no negative pressure generated. Non occlusive nature – tubing wear reduced.
  • 14. Centrifugal pump • In early 1970s, research related to the development of an artificial heart was basis of the development of centrifugal pumps for CPB. • Boimedicus 600 - 1973. •In the United States, the centrifugal pump is extensively used.
  • 15. Centrifugal Pump – Working Principle  Creating pressure gradient between inlet and oulet of pump.  This pressure gradient results from the creation of a vortex by the rotation of the pump head.  The vortex can be created by using cones that impart motion to the blood by viscous shear or by rotating impellers.  The rotating motion creates an area of low pressure in the center and an area of high pressure on the sides.
  • 16. Centrifugal Pump – Working Principle
  • 17. Centrifugal Pump – Working Principle • Resultant blood flow- •The resistance at the outlet is a function of two components: the CPB circuit and SVR. •Centrifugal pumps are afterload dependent and flow is influenced by changes in resistance in both the circuit and the patient •Flow meter necessary.
  • 18. Centrifugal force Centrifugal Pump – Working Principle
  • 19. Centrifugal Pump – Working Principle •HEAT GENERATION: All centrifugal pumps will generate heat depending on the amount of energy that is impaired into the blood. •Combination with the low flow in the center of the pump head - may create blood clots and blood cell activation in the pump.
  • 20. SPECIFIC CLINICALLY AVAILABLE CENTRIFUGAL PUMPS •BIOMEDICUS PUMP • In 1976, the first centrifugal pump was used for CPB. • The pump head is acrylic, with inlet and outlet ports oriented at right angles to each other, and its priming volume is 80 ml. • Cones driven by magnetic coupling to external console.
  • 21. CENTRIFUGAL PUMPS - Capiox pump Rotor with unique straight path design to reduce pump rotational speed without decreasing hydraulic efficiency Small priming volume – 46 ml reduces stagnant flow within the rotor
  • 22. CENTRIFUGAL PUMPS – Nikkiso Pump •Smallest commercially available pump •Priming volume of 25 ml. •Made of polycarbonate, with a V shaped ring seal that separates the pump housing and the actuator chamber. •Seal – made of fluororubber, suppresses heat generation and prevents blood leakage. •Six washout holes are incorporated into impeller to generate blood flow from the back to front surface of the impeller. •These holes prevent thrombus formation in areas behind the impeller and around the sealing part.
  • 23. Problems associated with centrifugal pumps Flow rate affected by preload and after load. Retrograde flows down the arterial line. Potential air entrapment if inadequate aferload. For forward flow pressure in the pump head (PP) must be greater than the combined patient pressure (PPT) and the pressure head (PH) {hydrostatic pressure}. --PP > PPT + PH forward flow
  • 24. Centrifugal Pump vs Roller Pump Expensive Inexpensive Pump flow function of SVR Flow predictable based on pump speed Cannot pump large amount of air Can pump large amount of air Pump stalls on occlusion without generating high suction or outlet pressure. Potential to overpressurize circuit if inadvertently clamped. Retrograde flow when pump slows / stops No retrograde flow. Does not require continous monitoring Continuous strict monitoring of blood level Hemolysis and damage to formed blood elements is less More hemolysis and damage to formed blood elements. Less wear and tear of pump. No spallation More wear and tear of tubing in pump head. Spallation.
  • 25. Oxygenator  Blood gas exchange device.
  • 26. Oxygenator Oxygenate venous blood. Remove CO2 Represent the largest surface area to which circulating blood is exposed. Components - Membrane module Heat exchanger Reservoir
  • 27. AN IDEAL OXYGENATOR Oxygenation of venous blood: device must have sufficient capacity to provide oxygenation over a wide range of venous flow rate. Carbon dioxide elimination to avoid hypercarbia or hypocarbia. Minimum trauma to the blood Small priming volume - to limit the deleterious effects of hemodilution Safety
  • 28. CLASSIFICATION Bubble oxygenator: The earliest oxygenators. Exchange gases through direct interaction of gas and blood. These devices were used during early advent of CPB. Membrane oxygenator Semi-permeable barrier that separates fluid from gas. Diffusive qualities of the membrane material determine the transfer of oxygen and carbon dioxide between phases.
  • 29. BUBBLE OXYGENATORS • First widely available commercial oxygenators • Structure : 3 sections of operation Bubble column Defoaming area Arterial reservoir • Desaturated blood passively enters mixing chamber, where 100% oxygen flows across a disparager plate into the stream of blood, which forms small bubbles
  • 30. BUBBLE OXYGENATORS Blood becomes oxygenated and carbon dioxide is reduced as stream of gas percolates through blood. Blood is defoamed by the presence of silicone antifoam-A, which consist of the liquid polymer dimethylpolysiloxane (96%) and particulate silica (4%), which destabilizes the bubbles, causing them to implode.
  • 31. BUBBLE OXYGENATORS The arterialized blood is collected in an arterial reservoir that is then actively pumped. The simple design of bubble oxygenators relies on the hydrostatic pressure head from the patient to the mixing chamber connected by the venous line. The pressure drop through bubble oxygenator is <30 cm of water, in contrast to the 100 cm of water pressure drop typically found in membrane oxygenators.
  • 32. BUBBLE OXYGENATORS Bubble size is critical to adequate gas transfer. The bubble size selected must be a compromise between optimal surface area for oxygenation and volume for carbon dioxide transfer. Decreasing size of bubbles increases total surface area of blood gas interface with better oxygenation but limiting total CO2 transfer. Bubble sizes of 3 to 7 mm are used to optimize both O2 and CO2 transfer.
  • 33. MEMBRANE OXYGENATORS • Complete barrier between the gas and blood phases and diffusion is through membrane material • Costly to manufacture and require large priming volume • Most membrane lungs used for CPB have micropores
  • 34. Willem J. Kolff During dialysis in 1943 noticed that the blue blood in the rotating-drum artificial kidney became red Clowes and Neville(1958) Poineers in using membrane oxygenators (teflon flat membranes)
  • 35. MEMBRANE OXYGENATORS – Materials Historically – Cellulose. Polytetrafluoroethylene Polyethylene Currently - Silicon rubber (homogeneous, nonporous membrane). Polypropylene (heterogeneous, microporous, hydrophobic membrane).
  • 36. Silicone Vs Polypropylene Membrane Silicone membrane Long term support Without a diminution in gas transfer capacity Avoid plasma leakage and membrane wet out Microporous polypropylene membrane Cheap Good for short periods New generation membranes that incorporates benefits of silicone with polypropylene have been developed.
  • 37. MEMBRANE OXYGENATORS – Designs Membrane materials are organized in three configurations: Scrolled envelope Parallel plate Hollow fiber
  • 38. Types of Membrane Oxygenators Plaque oxygenators -microporous expanded polypropylene -folded Z shape -blood & gas flow opposite direction -Cobel Excel, Cobe VPCML,Shirley M2000 Spiral oxygenators -silicon membranes -rolled around central axis -Kolobow oxygenator.
  • 39. Types of Membrane Oxygenators Hollow fibre oxygenator 1970 Benlips introduced Capillary fibers of microporous polypropylene
  • 40. Spiral Membrane Oxygenators Kolobow Silicon membrane in shape of an envelope that is coiled on itself. Used primarily in ECMO Ability to maintain stable CO2 and O2 for long periods (weeks). Available in gas exchange surface area sizes from 0.5 to 4.5 m2.
  • 41. Hollow Membrane Oxygenators Blood flow inside the capillaries Gas flow inside the capillaries Blood flow through the fiber was abandoned - High trans membrane pressure Activation of platelets Increased haemolysis Blood flow either perpendicular or in the direction of fiber bundle In latter case, blood will flow in a counter current direction to the gas flow - Optimized gas gradients during the dwell time
  • 42. Membrane oxygenators & lung MEMB OXYGN LUNG SURFACE AREA(M2) 0.5-4 70 Blood path width (µm) 200 8 Blood path length (µm) 250000 200 Memrane thickness (µm) 150 0.5 Max O2 transfer(ml/min) 400-600 2000
  • 43. ADVANCES IN OXYGENATORS Biocompatibility Heparin coating – 1980's Surface Modifying Additive— Polydimethylsiloxane polycaprolactone oligomer Phosphorylcholine Bioline X Coating Trillium Biopassive Surface—polyethylene oxide. Cost-effectiveness'????
  • 45. Priming solutions for CPB Circuit  Need of prime – to achieve adequate flow rates on initiation of CPB without air embolism.  Ideal prime - Similar electrolyte content, osmolarity and pH as that of plasma. On mixing with blood maintains oxygen delivery, CO2 removal and physiological homeostasis.
  • 46. Historical perspective of prime solutions
  • 47. Homologous blood syndrome  Blood borne infection  Severe pulmonary insufficiency  Impaired immunity – wound infection, sepsis  Impaired resistance to Malignant cell transformation  Graft versus host disease
  • 48. Impetus for nonhemic prime Severe strain on hospital blood bank Increased access to the emergency surgery Increased exposure to O2 in polycythemics Refusal of hemic prime from Jehovah’s witness faith Experimental success of hemodilution in CPB
  • 49. Glucose in priming solution 1962, Cooley – 5% dextrose in addition to blood - improves outcome Solution with glucose as major component – isotonic but after metabolisation of glucose become severely hypotonic Fluid shift from Extracellular to Intracellular compartment - Red blood cell lysis Pulmonary edema Cerebral edema Hyperglycemia – poor neurological outcome Priming solution – Normotonic, near physiologic sodium concentration.
  • 50. Colloidal Priming solution  Hemodilution – decreased colloid oncotic pressure – fluid shift into intracellular compartment – cellular edema and dysfunction.  Colloid solution - counteract reduction in colloid oncotic pressure – prevent fluid shift.  CPB – systemic inflammatory response – tight junctions at endothelial lining "permeable" to high molecular weight proteins – high molecular weight protein trapped in ECF – paradoxical increase in cellular edema.
  • 53. Experimental prime solutions Perfluoro carbons - 0.118 microns, half the viscocity of blood -O2 release even at low po2 environments -O2 relase is not related to pH/ temp - can perfuse distal capillareis - still at experimental stage Stroma free Hb -natural O2 carrying capacity &osmotic activity -lower viscocity than blood -do not cause immunosuppression -still in preclinical testing
  • 54. Priming Solution (AIIMS) Ringer lactate – 20 ml/kg. Hydroxy Ethyl Starch – 10 ml/kg. Mannitol – 5 ml/kg. Soda bicarbonate – 1 ml/kg. Heparin – on the basis of circuit used. If blood is added to prime – additional soda bicarbonate 10ml/300ml blood is added.
  • 55. Mannitol  Potent osmotic diuretic.  Maintain urine output during CPB and in immediate post bypass period.  Preserves renal function.  Free radical scavenger.
  • 57. Pump tubings Internal diameter (inch) Volume (ml/feet) Used as 1/4 8.6 Suction / vent <6kg – art./venous line 3/8 21.6 <10 kg – venous line. >11 kg – arterial line 1/2 37.0 > 20 kgVenous line
  • 58. CPB circuit according to age Weight (kg) Oxygenator Prime vol (ml) Art. line Venous line Circuit Priming vol. (ml) Heparin (mg) Max. Flow (lt/min) < 6 Baby Rx 35 1/4 1/4 400 - 500 25 1.2 6 - 10 Minimax 109 1/4 1/4 or 3/8 400 –600 25 - 30 2.3 11 - 20 Sx 10 135 3/8 3/8 1000 - 1100 50 3.5 >20 Affinity NT 290 3/8 1/2 1500 - 1700 75 7
  • 59. Priming volume  Volume required to fill the arterial and venous limbs, adequate voulme in reservoir to prevent air entering the arterial line on initiation of CPB.  Acceptable hemodilution??
  • 60. Calculation of Blood Volume to be added Pts estimated blood volume x Hct Predicted Hct= ____________________________ Pts estimated blood vol +CPB prime+ pre CPB iv fluid volume (TCV) RBC Vol. To be added = TCV (Hct Desired – Hct Predicted) Bank Blood volume to be added = RBC Vol to be added / 0.7
  • 61. Blood preservation techniques Pre CPB – Retrograde circulation Adult Hct > 32% CVP > 8 mm hg. SBP > 80 mm hg On CPB – Hemofilter. Cell Saver. Post CPB – Chase prime with crystalloid. Completely reverse and pack