Introduction
Majority of the congenital lesions are asso. with
intracardiac shunts.
Detection, localisation and quantification of shunts –
integral part of hemodynamic evaluation of these
patients.
Quantification of shunts by:
Invasive
Noninvasive methods.
Non invasive methods
Clinical evaluation
CXR and ECG
Echo and Doppler
MRI
CT
Nuclear scanning
Invasive
Oximetry
Indicator dilution
Shunt quantification
2 –D Echocardiography:
direct visualization of any communication and their
shunting
excellent for localization of shunt
limited in the ability to quantify shunt
Indirect estimations
ASD with large shunting
 RA, RV dilated
 diastolic septal flattening
VSD with large shunting
 LA, LV dilatation
 Systolic and diastolic septal flattening
Contrast Echo
Agitated saline
 Detection of small R  L shunts
Large left to right shunts may cause a negative
contrast jet in the Rt. atrium
Doppler echocardiography
Most of the studies were done in ASD & VSD.
Based on the measurement of flow across the valves
or the defect itself.
Based on the formula that flow can be calculated
from CSA & VTI.
Volumetric flow measurement requires
- calculation of temporal mean flow velocity and
cross sectional area
- velocity profile is uniform across the vessel
- single sampling of the flow velocity in the center
of the vessel is recorded
To calculate 
computerised calculation of the area under the
doppler curve/ flow period
For systemic flow
Aortic velocity is recorded from the apical view.
Vessel diameter is measured from the parasternal long
axis view.
Pulmonary blood flow:
Pulm. mean velocity is recorded from the
parasternal short axis view
In case of large shunt and a shunt located close
to PA, the doppler tracing shows spectral
broadening or signs of disturbed flow, in such
cases uniform velocity cannot be assumed
Doppler tracing cannot be used if any PS is
present.
Measurement of the cross sectional area
- at the valve annulus (as it is the flow limiting
point of the vessel)
- measurement is made from the inner edge to
the inner edge.
- diameter is measured at the early systole –
time when vessel is at peak systolic dimension.
- flow velocity should be measured at same
location
Methods:
1.Qp/Qs calculation from flow velocity (velocity
time integral) and cross sectional area
Qp/Qs = Area of Pa x VTI
Area of Aorta x VTI
Sanders et al used pulmonary flow with good
correlation , r=0.85.
Barron et al used Mitral flow with better
correlation , r= 0.9 vs 0.69.
2. Simplified method - the square of the pulmonary
to aortic diameter ratio was substituted for the
area ratio and the flow peak velocity ratio for the
velocity time integral.
 Qp/Qs = Vel. Pk PA x r2
Vel. Pk Ao x r2
r= radius / diameter
square of the ratio of the pulmonary to aortic
luminal diameter (or radius)is used instead of
vessel areas.
3. Qp/Qs = (pulm + mitral flow)/2
(tricuspid + aortic flow)/2
 Assumption that tricuspid and aortic flow
represent systemic flow and pulmonary &
mitral flows represent pulmonary flow.
 Maximal diameter of mitral and tricuspid
valve annulus is measured during diastole in
an apical 4 –chamber view & area calculated
assuming the annulus to be circular.
 Laborious & time taking.
 Combines measurements from all the 4 valves
but was not shown to be better than the
previous 2.
4. Qp/Qs = Shunt flow + aortic flow
aortic flow
 Shunt (VSD) flow  as the product of the VTI
and color-derived cross-sectional area of VSD jet
measured at its narrowest point.
 This method does not use the pulmonary flow &
outflow diameter which are more variable.
Difficulties : It is assumed that
1. The defect is circular.
2. The size does not vary during systole.
3. No angle correction is needed.
4 . Flow velocity is uniform across the orifice.
Most studies show moderate correlation of
doppler techniques with oximetry data.
Sabry et al, studied above 4 methods with
cardiac cath calculated shunts.
First 2 methods showed moderate correlation
( r=0.54 & 0.56 ). Correlation improved
when large VSDs were excluded (r=0.62 &
0.66 ).
3rd
method showed no better correlation & it
was more time taking (r =0.57).
4th
method showed best correlation (r =0.82).
When large VSDs were excluded it was r = 0.9.
Sources of error
Mainly from the assumptions made
1) assumption that velocity profile is uniform
through out the cross sectional area
2) alignment of the doppler beam with the
direction of the flow ( angle < 200
)
3) turbulent flow in the vessels mainly in the
PA.
4) inaccuracy in measurement of cross
sectional area of vessels
- PA dimensions vary significantly during
cardiac cycle
5) errors in calculation because of presence of
additional defects
e.g PDA present downstream from the PA
6) semilunar valve regurgitation results in
overestimation of flow because of failure to measure
the regurgitant volume
7) All these methods overestimate the shunt in large
VSDs.
8) Quantification of PDA is not accurate due to the
turbulent flow across the defect & difficulty in
measuring the flow distal to the shunt.
Quantification of ASD
TTE estimation of ASD diameters & shunt
quantification by cath showed only fair correlation
(r=0.56).
• The size of the defect by transesophageal Doppler
color flow mapping correlated fairly well with the
size estimated at surgery (r = 0.73 ).
Other measurements by TTE for ASD are –
- RV/LV diameter (r=0.64)
- Area of PA (r=0.62)
- RV volume (r=o.71)
- PA/Ao (r=0.89)
Doppler colour flow mapping by
TEE
• Area of the ASD- calculated by assuming it to be
circular and taking the maximal Doppler color flow jet
width at the defect site as its diameter.
• The pulsed Doppler sample volume is to be placed
parallel to the shunt flow direction at the defect site to
obtain the mean velocity and flow duration.
• From these values, the shunt volume can
calculated as a product of the defect area, mean
velocity, flow duration and heart rate.
• The calculated shunt flow volume obtained by
transesophageal study showed a good correlation
with shunt flow volume (r = 0.91) and pulmonary to
systemic blood flow ratio (r = 0.84) obtained at
cardiac catheterization.
R  L shunt quantification
Aortic flow + VSD flow = Input to RV
Pulmonary flow + R  L shunt = output from RV.
Hence R  L shunt = Aortic flow + VSD flow –
Pulmonary flow.
Only fair correlation found with Cath data (r = 0.61 –
0.77 )
Automated cardiac flow
measurement method
ACM method – using spatial and temporal
integration of colour Doppler profiles .
Method  velocity profile in a region of interest
set on the flow tract is detected in each frame ,
imaged & recorded on the image memory
Flow volume rate is calculated by integration of
the velocity profile
Stroke volume is measured by temporal
integration, throughout the systolic period.
ACM method:
Advantages:
Quick and requires only two manual procedures.-
selection of systolic period in the stored image
memory, positioning at the area of interest
Calculations are done without tracing the
Doppler wave form to measure VTI
No need to measure area of outflow tract, because
the edge of the color profile is detected as the width
of the flow tract.
ACM method
ACM uses velocity profile across the flow tract
diameter, while in conventional pulsed Doppler
method the spectral velocity from a centrally
placed sample point is measured
Flow rate at each frame is temporally integrated
for the stroke volume calculation
Thus this method requires fewer assumptions
than pulsed Doppler method.
Martin et al found strong correlations between
ACM and invasive QP/QS ratio and the agreement
with invasive data was better using ACM than
using conventional echocardiographic method ( r
= 0.91 )
The only restriction remains to select manually
the systolic period and to carefully choose the
region of interest.
Gain must be optimised, allowing to see only one
colour throughout the outflow tract
Avoids the potential error of conventional PWD
method for the calculation of pulmonary output,
linked to the difficulty of measuring the exact
pulmonary diameter.
This method was applied to ASD & VSD with better
correlation .
Shunt quantification
Principle: oxygenated blood shunted from left
side of the heart to the rt cause an abnormal
increase (step – up) in the oxygen content or
saturation of blood in the chamber into which
shunting occurs
Dexter et al, max. increase in oxygen content
from,
RV to PA - 0.5 ml/dl
RA to RV - 1.0 ml/dl
SVC to RA - 2.0 ml/dl
Barratt -Boyes and Wood suggested that multiple
blood samples to be obtained from each Rt. heart
chamber.
Data to be averaged before applying Dexter
Criteria
Advantages
- easy to perform, results available immediately
- can ascertain the site of shunt
- magnitude of the shunt can be calculated
Detection of left to right shunt
Level of
shunt
Mean O2 sat % in
distal chambers
(Highest value in
proximal
chambers)
Mean O2
vol. % in
proximal
chambers
(Highest
value in
distal
chambers)
Min Qp/Qs
reqd. for
detection at
3L/min/m2
Possible causes
of step up
Atrial
(SVC/IVC
to RA)
≥7 (≥11) ≥1.3 (≥2.0) 1.5 – 1.9 ASD, RSOV,
VSD with TR,
coronary fistula
to RA,
anomalous PV
drainage
Ventricular
(RA to RV)
≥5 (≥10) ≥1.0 (≥1.7) 1.3 – 1.5 VSD, PDA with
PR, Primum
ASD, coronary
fistula to RV
Detection of left to right shunt
Level of
shunt
Mean O2 sat % in
distal chambers
(Highest value in
proximal
chambers)
Mean O2
vol. % in
proximal
chambers
(Highest
value in
distal
chambers)
Min Qp/Qs
reqd. for
detection at
3L/min/m2
Possible causes
of step up
Great
vessel
(RV to PA)
≥5 (≥5) ≥1.0 (≥1.0) ≥1.3 PDA, Aorta-
pulmonic
window,
Aberrant
coronary artery
origin
Any level
(SVC to
PA)
≥7 (≥8) ≥1.3 (≥1.5) ≥1.5 All of the above
Detection of left to right shunt
Chambers
sampled
Single
Sample
Multiple
samples
SVC – RA 7% 5%
RA – RV 5% 3%
RV – PA 4% 3%
PV/LA –
LV/SA
-3% -2%
Oximetry – limitations:
It lacks sensitivity – i.e. does not allow detection of
small left to right shunts
Requirement of steady state during collection of blood
samples
Calculation of Mixed venous saturation: Normal
variability of blood oxygen saturation in the Rt heart
chambers is strongly influenced by the magnitude of
systemic blood flow.
Higher levels of systemic blood flow  higher mixed
venous saturation  blunting of inter-chamber
variability.
Step up in oxygen content of the receiving chamber
depends also on the oxygen carrying capacity of blood
i.e. Hb conc.
Depends on the associated lesions also – like TR, PR.
Precautions
Samples at multiple sites to be obtained rapidly ,
should take < 7 min.
O2 saturation data rather than O2 content are
preferable,
Comparision of mean values is preferable to
highest values.
When the systemic blood flow at rest is low,
exercise shoud be used.
At higher FiO2 the dissolved O2 should also be
considered.
Shunt calculation
Qp = VO2 / (PVO2- PAO2) (Hb) (1.36) (10)
Qs = VO2 / (AoO2 – SVCO2) (Hb) (1.36) (10)
Qp eff. = VO2 / (PVO2 – SVC O2) (Hb) (1.36) (10)
Qp / Qs = (AoO2 – SVC O2) / (PVO2 – PAO2)
Q L  R shunt = (Qp – Qp eff)
 Q R  L = (Qs – Qs eff)
Carter formula
Determines the left to right shunting.
Best used when there is a smooth downstroke
They can detect a minimum shunt of 25 – 30%.
Simplified formula of Victoria & Gessner is used when
there is no smooth transition.
Calculation of Rt to Lt shunt
BT1
T2
%Qs = BT1 x P1x 100
BT1xP1 + 0.44xT2xP2
%Qs = percentage of systemic blood flow due to Rt to
Lt shunt.
PC1 = height of first peak.
BT1 = bulid up time from appearance to first peak.
T2 = time from injection to P2.
Shunt quantification

More Related Content

PPTX
Shunt quantification
PPTX
Shunt quantification and reversibility
PPTX
Assessment of operability of left to right shunts
PPTX
Cath hemodynamics vir
PPT
Intracardiac shunts
PPT
Shunt Detection And Quantification
PPTX
PRESSURE MEASUREMENT by Cardiac catheterisation_Dr Amol Patil.pptx
PPTX
CATH MEET PDA
Shunt quantification
Shunt quantification and reversibility
Assessment of operability of left to right shunts
Cath hemodynamics vir
Intracardiac shunts
Shunt Detection And Quantification
PRESSURE MEASUREMENT by Cardiac catheterisation_Dr Amol Patil.pptx
CATH MEET PDA

What's hot (20)

PPTX
BMV balloons- FINAL.pptx
PPTX
Right heart catheters
PDF
Pressure, Damping and Ventricularization_Crimson Publishers
PPTX
ECHOCARDIOGRAPHY IN CARDIAC TAMPONADE
PPTX
Lv systolic function
PPTX
Non infarction Q waves
PPTX
FRACTIONAL FLOW RESERVE
PPTX
VSD devices
PPTX
Diastolic dysfunction
PPT
Evaluation of prosthetic valve function and clinical utility.
PPTX
CORONARY ANOMALY
PPTX
Basics of Electrophysiologic study, part 1 (2020)
PPTX
Echo assessment of Aortic Regurgitation
PPTX
Assessment of prosthetic valve function
PDF
Echo assessment of aortic stenosis
PPTX
Cath hemodynamics vir
PDF
Echo in cardiomyopathies part 1
PPSX
Right heart catheterization
PPTX
Coronary angiogram
BMV balloons- FINAL.pptx
Right heart catheters
Pressure, Damping and Ventricularization_Crimson Publishers
ECHOCARDIOGRAPHY IN CARDIAC TAMPONADE
Lv systolic function
Non infarction Q waves
FRACTIONAL FLOW RESERVE
VSD devices
Diastolic dysfunction
Evaluation of prosthetic valve function and clinical utility.
CORONARY ANOMALY
Basics of Electrophysiologic study, part 1 (2020)
Echo assessment of Aortic Regurgitation
Assessment of prosthetic valve function
Echo assessment of aortic stenosis
Cath hemodynamics vir
Echo in cardiomyopathies part 1
Right heart catheterization
Coronary angiogram
Ad

Viewers also liked (20)

PPT
Assessment of shunt by cardiac catheterization
PPTX
Pulmonary hypertension with cardiac shunt determination
PPTX
Surgical shunts
PPTX
Calculation of bidirectional shunts
PDF
Portal hypertension
PPT
Coronary Calcium and other CVD Risk Biomarkers: From Epidemiology to Comparat...
PPTX
HCM – Presentation, Hemodynamics and Intervention
PPTX
Framingham
PPT
All About Framingham Heart Study
PPTX
Jupiter Trial
PPTX
Takayasu's arteritis
PPTX
Natural history of right to left shunts
PPTX
Portal hypertension surgical management
PPTX
Treatment of portal hypertension
PPTX
Constrictive pericarditis
PDF
Euroecho2010 intracardiac-shunts
PPTX
Echo Differentiation of Restrictive Cardiomyopathy and Constrictive Pericarditis
PPTX
Intracoronary optical coherence tomography
PPTX
Aortic stenosis - Echocardiography
Assessment of shunt by cardiac catheterization
Pulmonary hypertension with cardiac shunt determination
Surgical shunts
Calculation of bidirectional shunts
Portal hypertension
Coronary Calcium and other CVD Risk Biomarkers: From Epidemiology to Comparat...
HCM – Presentation, Hemodynamics and Intervention
Framingham
All About Framingham Heart Study
Jupiter Trial
Takayasu's arteritis
Natural history of right to left shunts
Portal hypertension surgical management
Treatment of portal hypertension
Constrictive pericarditis
Euroecho2010 intracardiac-shunts
Echo Differentiation of Restrictive Cardiomyopathy and Constrictive Pericarditis
Intracoronary optical coherence tomography
Aortic stenosis - Echocardiography
Ad

Similar to Shunt quantification (20)

PPTX
Hemodynamics in echo lab by Dr. Ranjeet S.Palkar
PPTX
Basics in echocardiography - an initiative in evaluation of valvular heart di...
PPT
Hemodynamic assessment in cardiology
PPTX
ECHO views and measurements-Dr. Razu.pptx
PPTX
pediatric echo measurements technique z score
PPTX
Echocardiography
PPTX
Functional ECHO.pptx
PPTX
Shunt quantification and reversibility
PPT
Echocardiographic evaluation of lv function
PDF
Invasive Hemodynamics: Assessment and interpretation
PDF
Doppler echocardiography
PDF
Haemodynamics, Gradients and calculations - Inga Voges.pdf
PPTX
ASD ECHO ASSESSMENT renewed slideshow.pptx
PPTX
Use of echocardiography in icu
PPTX
siêu âm tim trong thông liên nhĩ.pptx
PPTX
Cardiac catheteriztion, Oximetery study in a patient with VSD
PPTX
Aortic valve assessment
PDF
Hemodynamics assessment in the Catheterization lab
PPTX
Echocardiogram
Hemodynamics in echo lab by Dr. Ranjeet S.Palkar
Basics in echocardiography - an initiative in evaluation of valvular heart di...
Hemodynamic assessment in cardiology
ECHO views and measurements-Dr. Razu.pptx
pediatric echo measurements technique z score
Echocardiography
Functional ECHO.pptx
Shunt quantification and reversibility
Echocardiographic evaluation of lv function
Invasive Hemodynamics: Assessment and interpretation
Doppler echocardiography
Haemodynamics, Gradients and calculations - Inga Voges.pdf
ASD ECHO ASSESSMENT renewed slideshow.pptx
Use of echocardiography in icu
siêu âm tim trong thông liên nhĩ.pptx
Cardiac catheteriztion, Oximetery study in a patient with VSD
Aortic valve assessment
Hemodynamics assessment in the Catheterization lab
Echocardiogram

Recently uploaded (20)

PPTX
Slider: TOC sampling methods for cleaning validation
PPTX
Gastroschisis- Clinical Overview 18112311
PPTX
Fundamentals of human energy transfer .pptx
PPT
genitourinary-cancers_1.ppt Nursing care of clients with GU cancer
PPTX
Chapter-1-The-Human-Body-Orientation-Edited-55-slides.pptx
PPT
CHAPTER FIVE. '' Association in epidemiological studies and potential errors
PPT
Obstructive sleep apnea in orthodontics treatment
PPT
Copy-Histopathology Practical by CMDA ESUTH CHAPTER(0) - Copy.ppt
PPT
1b - INTRODUCTION TO EPIDEMIOLOGY (comm med).ppt
PPT
MENTAL HEALTH - NOTES.ppt for nursing students
PPTX
15.MENINGITIS AND ENCEPHALITIS-elias.pptx
PPTX
NEET PG 2025 Pharmacology Recall | Real Exam Questions from 3rd August with D...
PPTX
SKIN Anatomy and physiology and associated diseases
PDF
Human Health And Disease hggyutgghg .pdf
PPTX
Pathophysiology And Clinical Features Of Peripheral Nervous System .pptx
PPTX
surgery guide for USMLE step 2-part 1.pptx
PDF
Medical Evidence in the Criminal Justice Delivery System in.pdf
PPTX
Important Obstetric Emergency that must be recognised
PPTX
ca esophagus molecula biology detailaed molecular biology of tumors of esophagus
PDF
Khadir.pdf Acacia catechu drug Ayurvedic medicine
Slider: TOC sampling methods for cleaning validation
Gastroschisis- Clinical Overview 18112311
Fundamentals of human energy transfer .pptx
genitourinary-cancers_1.ppt Nursing care of clients with GU cancer
Chapter-1-The-Human-Body-Orientation-Edited-55-slides.pptx
CHAPTER FIVE. '' Association in epidemiological studies and potential errors
Obstructive sleep apnea in orthodontics treatment
Copy-Histopathology Practical by CMDA ESUTH CHAPTER(0) - Copy.ppt
1b - INTRODUCTION TO EPIDEMIOLOGY (comm med).ppt
MENTAL HEALTH - NOTES.ppt for nursing students
15.MENINGITIS AND ENCEPHALITIS-elias.pptx
NEET PG 2025 Pharmacology Recall | Real Exam Questions from 3rd August with D...
SKIN Anatomy and physiology and associated diseases
Human Health And Disease hggyutgghg .pdf
Pathophysiology And Clinical Features Of Peripheral Nervous System .pptx
surgery guide for USMLE step 2-part 1.pptx
Medical Evidence in the Criminal Justice Delivery System in.pdf
Important Obstetric Emergency that must be recognised
ca esophagus molecula biology detailaed molecular biology of tumors of esophagus
Khadir.pdf Acacia catechu drug Ayurvedic medicine

Shunt quantification

  • 1. By: Dr. Ankur Gupta Resident, DM Cardiolgy
  • 2. Introduction Majority of the congenital lesions are asso. with intracardiac shunts. Detection, localisation and quantification of shunts – integral part of hemodynamic evaluation of these patients. Quantification of shunts by: Invasive Noninvasive methods.
  • 3. Non invasive methods Clinical evaluation CXR and ECG Echo and Doppler MRI CT Nuclear scanning
  • 6. 2 –D Echocardiography: direct visualization of any communication and their shunting excellent for localization of shunt limited in the ability to quantify shunt Indirect estimations ASD with large shunting  RA, RV dilated  diastolic septal flattening VSD with large shunting  LA, LV dilatation  Systolic and diastolic septal flattening
  • 7. Contrast Echo Agitated saline  Detection of small R  L shunts Large left to right shunts may cause a negative contrast jet in the Rt. atrium
  • 8. Doppler echocardiography Most of the studies were done in ASD & VSD. Based on the measurement of flow across the valves or the defect itself. Based on the formula that flow can be calculated from CSA & VTI.
  • 9. Volumetric flow measurement requires - calculation of temporal mean flow velocity and cross sectional area - velocity profile is uniform across the vessel - single sampling of the flow velocity in the center of the vessel is recorded To calculate  computerised calculation of the area under the doppler curve/ flow period
  • 10. For systemic flow Aortic velocity is recorded from the apical view. Vessel diameter is measured from the parasternal long axis view.
  • 11. Pulmonary blood flow: Pulm. mean velocity is recorded from the parasternal short axis view In case of large shunt and a shunt located close to PA, the doppler tracing shows spectral broadening or signs of disturbed flow, in such cases uniform velocity cannot be assumed Doppler tracing cannot be used if any PS is present.
  • 12. Measurement of the cross sectional area - at the valve annulus (as it is the flow limiting point of the vessel) - measurement is made from the inner edge to the inner edge. - diameter is measured at the early systole – time when vessel is at peak systolic dimension. - flow velocity should be measured at same location
  • 13. Methods: 1.Qp/Qs calculation from flow velocity (velocity time integral) and cross sectional area Qp/Qs = Area of Pa x VTI Area of Aorta x VTI Sanders et al used pulmonary flow with good correlation , r=0.85. Barron et al used Mitral flow with better correlation , r= 0.9 vs 0.69.
  • 14. 2. Simplified method - the square of the pulmonary to aortic diameter ratio was substituted for the area ratio and the flow peak velocity ratio for the velocity time integral.  Qp/Qs = Vel. Pk PA x r2 Vel. Pk Ao x r2 r= radius / diameter square of the ratio of the pulmonary to aortic luminal diameter (or radius)is used instead of vessel areas.
  • 15. 3. Qp/Qs = (pulm + mitral flow)/2 (tricuspid + aortic flow)/2  Assumption that tricuspid and aortic flow represent systemic flow and pulmonary & mitral flows represent pulmonary flow.  Maximal diameter of mitral and tricuspid valve annulus is measured during diastole in an apical 4 –chamber view & area calculated assuming the annulus to be circular.  Laborious & time taking.  Combines measurements from all the 4 valves but was not shown to be better than the previous 2.
  • 16. 4. Qp/Qs = Shunt flow + aortic flow aortic flow  Shunt (VSD) flow  as the product of the VTI and color-derived cross-sectional area of VSD jet measured at its narrowest point.  This method does not use the pulmonary flow & outflow diameter which are more variable.
  • 17. Difficulties : It is assumed that 1. The defect is circular. 2. The size does not vary during systole. 3. No angle correction is needed. 4 . Flow velocity is uniform across the orifice.
  • 18. Most studies show moderate correlation of doppler techniques with oximetry data. Sabry et al, studied above 4 methods with cardiac cath calculated shunts. First 2 methods showed moderate correlation ( r=0.54 & 0.56 ). Correlation improved when large VSDs were excluded (r=0.62 & 0.66 ). 3rd method showed no better correlation & it was more time taking (r =0.57). 4th method showed best correlation (r =0.82). When large VSDs were excluded it was r = 0.9.
  • 19. Sources of error Mainly from the assumptions made 1) assumption that velocity profile is uniform through out the cross sectional area 2) alignment of the doppler beam with the direction of the flow ( angle < 200 ) 3) turbulent flow in the vessels mainly in the PA. 4) inaccuracy in measurement of cross sectional area of vessels - PA dimensions vary significantly during cardiac cycle
  • 20. 5) errors in calculation because of presence of additional defects e.g PDA present downstream from the PA 6) semilunar valve regurgitation results in overestimation of flow because of failure to measure the regurgitant volume 7) All these methods overestimate the shunt in large VSDs. 8) Quantification of PDA is not accurate due to the turbulent flow across the defect & difficulty in measuring the flow distal to the shunt.
  • 21. Quantification of ASD TTE estimation of ASD diameters & shunt quantification by cath showed only fair correlation (r=0.56). • The size of the defect by transesophageal Doppler color flow mapping correlated fairly well with the size estimated at surgery (r = 0.73 ). Other measurements by TTE for ASD are – - RV/LV diameter (r=0.64) - Area of PA (r=0.62) - RV volume (r=o.71) - PA/Ao (r=0.89)
  • 22. Doppler colour flow mapping by TEE • Area of the ASD- calculated by assuming it to be circular and taking the maximal Doppler color flow jet width at the defect site as its diameter. • The pulsed Doppler sample volume is to be placed parallel to the shunt flow direction at the defect site to obtain the mean velocity and flow duration.
  • 23. • From these values, the shunt volume can calculated as a product of the defect area, mean velocity, flow duration and heart rate. • The calculated shunt flow volume obtained by transesophageal study showed a good correlation with shunt flow volume (r = 0.91) and pulmonary to systemic blood flow ratio (r = 0.84) obtained at cardiac catheterization.
  • 24. R  L shunt quantification Aortic flow + VSD flow = Input to RV Pulmonary flow + R  L shunt = output from RV. Hence R  L shunt = Aortic flow + VSD flow – Pulmonary flow. Only fair correlation found with Cath data (r = 0.61 – 0.77 )
  • 25. Automated cardiac flow measurement method ACM method – using spatial and temporal integration of colour Doppler profiles . Method  velocity profile in a region of interest set on the flow tract is detected in each frame , imaged & recorded on the image memory Flow volume rate is calculated by integration of the velocity profile Stroke volume is measured by temporal integration, throughout the systolic period.
  • 26. ACM method: Advantages: Quick and requires only two manual procedures.- selection of systolic period in the stored image memory, positioning at the area of interest Calculations are done without tracing the Doppler wave form to measure VTI No need to measure area of outflow tract, because the edge of the color profile is detected as the width of the flow tract.
  • 27. ACM method ACM uses velocity profile across the flow tract diameter, while in conventional pulsed Doppler method the spectral velocity from a centrally placed sample point is measured Flow rate at each frame is temporally integrated for the stroke volume calculation Thus this method requires fewer assumptions than pulsed Doppler method.
  • 28. Martin et al found strong correlations between ACM and invasive QP/QS ratio and the agreement with invasive data was better using ACM than using conventional echocardiographic method ( r = 0.91 ) The only restriction remains to select manually the systolic period and to carefully choose the region of interest. Gain must be optimised, allowing to see only one colour throughout the outflow tract
  • 29. Avoids the potential error of conventional PWD method for the calculation of pulmonary output, linked to the difficulty of measuring the exact pulmonary diameter. This method was applied to ASD & VSD with better correlation .
  • 31. Principle: oxygenated blood shunted from left side of the heart to the rt cause an abnormal increase (step – up) in the oxygen content or saturation of blood in the chamber into which shunting occurs Dexter et al, max. increase in oxygen content from, RV to PA - 0.5 ml/dl RA to RV - 1.0 ml/dl SVC to RA - 2.0 ml/dl
  • 32. Barratt -Boyes and Wood suggested that multiple blood samples to be obtained from each Rt. heart chamber. Data to be averaged before applying Dexter Criteria Advantages - easy to perform, results available immediately - can ascertain the site of shunt - magnitude of the shunt can be calculated
  • 33. Detection of left to right shunt Level of shunt Mean O2 sat % in distal chambers (Highest value in proximal chambers) Mean O2 vol. % in proximal chambers (Highest value in distal chambers) Min Qp/Qs reqd. for detection at 3L/min/m2 Possible causes of step up Atrial (SVC/IVC to RA) ≥7 (≥11) ≥1.3 (≥2.0) 1.5 – 1.9 ASD, RSOV, VSD with TR, coronary fistula to RA, anomalous PV drainage Ventricular (RA to RV) ≥5 (≥10) ≥1.0 (≥1.7) 1.3 – 1.5 VSD, PDA with PR, Primum ASD, coronary fistula to RV
  • 34. Detection of left to right shunt Level of shunt Mean O2 sat % in distal chambers (Highest value in proximal chambers) Mean O2 vol. % in proximal chambers (Highest value in distal chambers) Min Qp/Qs reqd. for detection at 3L/min/m2 Possible causes of step up Great vessel (RV to PA) ≥5 (≥5) ≥1.0 (≥1.0) ≥1.3 PDA, Aorta- pulmonic window, Aberrant coronary artery origin Any level (SVC to PA) ≥7 (≥8) ≥1.3 (≥1.5) ≥1.5 All of the above
  • 35. Detection of left to right shunt Chambers sampled Single Sample Multiple samples SVC – RA 7% 5% RA – RV 5% 3% RV – PA 4% 3% PV/LA – LV/SA -3% -2%
  • 36. Oximetry – limitations: It lacks sensitivity – i.e. does not allow detection of small left to right shunts Requirement of steady state during collection of blood samples Calculation of Mixed venous saturation: Normal variability of blood oxygen saturation in the Rt heart chambers is strongly influenced by the magnitude of systemic blood flow. Higher levels of systemic blood flow  higher mixed venous saturation  blunting of inter-chamber variability.
  • 37. Step up in oxygen content of the receiving chamber depends also on the oxygen carrying capacity of blood i.e. Hb conc. Depends on the associated lesions also – like TR, PR.
  • 38. Precautions Samples at multiple sites to be obtained rapidly , should take < 7 min. O2 saturation data rather than O2 content are preferable, Comparision of mean values is preferable to highest values. When the systemic blood flow at rest is low, exercise shoud be used. At higher FiO2 the dissolved O2 should also be considered.
  • 39. Shunt calculation Qp = VO2 / (PVO2- PAO2) (Hb) (1.36) (10) Qs = VO2 / (AoO2 – SVCO2) (Hb) (1.36) (10) Qp eff. = VO2 / (PVO2 – SVC O2) (Hb) (1.36) (10) Qp / Qs = (AoO2 – SVC O2) / (PVO2 – PAO2) Q L  R shunt = (Qp – Qp eff)  Q R  L = (Qs – Qs eff)
  • 40. Carter formula Determines the left to right shunting. Best used when there is a smooth downstroke They can detect a minimum shunt of 25 – 30%. Simplified formula of Victoria & Gessner is used when there is no smooth transition.
  • 41. Calculation of Rt to Lt shunt BT1 T2
  • 42. %Qs = BT1 x P1x 100 BT1xP1 + 0.44xT2xP2 %Qs = percentage of systemic blood flow due to Rt to Lt shunt. PC1 = height of first peak. BT1 = bulid up time from appearance to first peak. T2 = time from injection to P2.