70th TRAINING PROGRAM OF NITUB
Use, Maintenance and
Troubleshooting of Small Lab-
Instruments
Dr. Md Shahinul Haque Khan
Professor (cc)
Department of Chemistry
Bangladesh University of Health Sciences
125/1, Darus salam, Mirpur 1
Emai: khanshahinul@yahoo.com
Name of the common Instruments
used in a Laboratory
1. Ultrasonic cleaner
2. Homogenizer
3. Drying oven
4. Desiccators
5. pH meter
6. Balance, Weight box
7. Rotary Vacuum Evaporator
8. Freeze dryer
9. Centrifuge machine
10. Water bath
11. Melting point apparatus
12. Hot plate with Magnetic stirrer
13. Heating Mantle
14. UV Chamber
15. Vortex Mixer
16. Filtration Unit
Ultrasonic Cleaner
Ultrasonic cleaning is a process that
uses ultrasound (usually from 20–400 kHz) and an
appropriate cleaning solvent to clean items.
 Ultrasonic cleaning uses cavitation bubbles
induced by high frequency pressure (sound)
waves to agitate a liquid. The agitation produces
high forces on contaminants adhering to
substrates like metals, plastics, glass, rubber, and
ceramics.
 used to clean many different types of objects,
including jewelry, lenses and other optical
parts, watches, dental and surgical
instruments, tools, car fuel injectors, musical
instruments, industrial parts and electronic
equipment. electronic repair workshops.
Homogenizer
Used for
 the homogenization of various types of
material, such as tissue, plant, food, soil, and
many others.
 sample preparation step prior to the analysis
of nucleic acids, proteins, Cell, pathogens, and
many other targets.
Drying Oven
Desiccators
Desiccators are sealable or sealable with vacuum
enclosures containing desiccants.
It is used to
 preserving moisture-sensitive items,
 protect chemicals which are hygroscopic ,
 prevent reaction with water from humidity,
 keep materials in drying condition,
 reduce temperature during moisture determination
70th training program of nitub shahin
Vacuum Desiccators
UV Blocking Vacuum Desiccator
Sizes of Desiccators
1. Desiccator, Glass, Clear, I.D. 110 mm
2. Desiccator, Glass, Clear, I.D. 180 mm
3. Desiccator, Glass, Clear, I.D. 240 mm
4. Vacuum Desiccator, Glass, Amber, I.D. 240 mm
5. Space Saver Vacuum Desiccator, 19.7cm ID
6. Desiccator, Glass, Clear, I.D. 150 mm
7. Cabinet type
Desiccant
A desiccant is a hygroscopic substance that induces or sustains a state of
dryness in its vicinity.
Characteristics : chemically inert, non-toxic, water insoluble
Performance efficiency
# ratio or percentage (water storable : mass of desiccant)
# relative humidity of the air or other substance being dried
70th training program of nitub shahin
pH meter
It is a scientific instrument used to measure
 the hydrogen-ion activity in water-based solutions,
 the acidity or alkalinity,
 the difference in electrical potential between a pH
electrode and a reference electrode.
Electrode:
# pH electrode and
# reference electrode design
pH-sensitive glass membrane, 0.4 mm thickness
Construction of an electrode
1. A sensing part, a bulb made from a specific glass
2. Internal electrode, silver chloride electrode or calomel electrode
3. Internal solution, usually a pH=7 buffered solution of 0.1 mol/L
KCl for pH electrodes.
4. When using the silver chloride electrode, a small amount of
AgCl can precipitate inside the glass electrode
5. Reference electrode,
6. Reference internal solution, usually 0.1 mol/L KCl
7. Junction with studied solution, usually made from ceramics or
capillary with asbestos or quartz fiber.
8. Body of electrode, made from non-conductive glass or plastics.
70th training program of nitub shahin
Selection of pH electrode
The types of pH electrodes are many and various. The epoxy
bodied combination electrode is suitable for the majority of
tests carried out on aqueous solutions.
For other applications, such as low ionic strength, high
temperatures and strongly acidic solutions a more suitable
pH/reference electrode pair may be required
Storing the pH electrode
1. After Use - Rinse thoroughly with distilled water.
Short Term Storage - Immerse in storage solution
Long Term Storage - Fit wetting cap filled with storage solution
and replace in original packaging.
2. Electrodes should be stored
a) away from direct sunlight
b) in a vertical position
c) within their specified temperature range
3. Always ensure the electrode is used within its specified
temperature range. Ageing of electrodes used above their
specified temperature is rapid and irreversible.
Precautions
1. DO NOT touch the sensitive glass pH membrane or reference
junction. Excess droplets of solution may be removed by gently
blotting with filter paper or tissue.
2. DO NOT wipe the electrode as this may induce an electrostatic
charge
CLEANING OF GLASS ELECTRODES
Deposit Cleaning agents
General deposits Mild detergent solution
Inorganic coatings Commercial glass cleaning solution (not strongly acidic)
Metal compounds Acid solution, not stronger than 1M
Oil/Grease Complexing agent (EDTA) or suitable solvent
Resins/Lignins Acetone, alcohol or detergent (not strongly alkaline).
Proteins (blood, etc) Enzyme solutions e.g. Pepsin in 0.1M HCl
Stubborn deposits Weak hydrogen peroxide solution, Sodium Hypochlorite
solution or domestic bleach.
If Electrodes is dried out, (dry deposit of KCl crystals on the electrode body), by
soaking overnight in warm deionised water.
Note: The epoxy bodied electrode should not be cleaned with aggressive solvents.
ROUBLESHOOTING GUIDE
FAULT PROBABLE CAUSE ACTION
power a. Battery failure
b. Battery polarity incorrect
a. Replace batteries
b. Refit batteries
Cal error 1, 2 or 3 a. Incorrect value buffer used
b. Contaminated buffer solution
a. Use correct value
b. Use fresh buffer solutions
Cal error No 1 or 2 a. pH electrode defective
b. Same buffer used for 2 point cal
a. Replace pH electrode
b. Recalibrate by 2 point buffer solution
Unstable display pH electrode defective Replace pH electrode
Display shows 1 or -
1 only
a. Intermittant or no connection
b. pH electrode defective
c. Contaminated solutions
a. Check electrode connection to unit
b. Replace pH electrode
c. Replace solutions
Intermittant display Probes not fitted correctly Check connections
I/O switch not
working
a. I/O key not held down >2 secs
b. Batteries not fitted
a. Retry holding key for 2 seconds
b. Fit batteries
CHECKING the pH METER FUNCTION
 Adjust temperature to 25°C using the keys.
 Check buffer solution
 pH 7 and the display should read 7 pH.
 pH 4 and the display should read 4 pH.
 pH 10 the display should show 10 pH.
 the run unknown sample to read its pH value.
Weighing Scales
These are devices to measure weight or calculate
mass.
a. Spring balances or spring scales: measure
force (weight) of gravity against the force on a
spring. Example: electronic balance, digital scale.
b. A balance or pair of scales: compares masses
by balancing the weight due to the mass of an
object against the weight of a known mass.
Example: Pan balance
Installation of a balance
# On a solid, vibration-free surface, away from direct sunlight and at an
constant temperature.
# precisely levelled, using the built-in spirit level:
# Protect the balance from drafts of air. Air moving across the pans will
cause inaccurate readings.
# Always keep the balance and weights clean and dry to protect them
from corrosion.
# Check the weight 10 mg, 50 mg and 100 mg and keep note
Other items
Weighing paper;
Set of calibrating weights;
Dust brush.
Silica blue desiccant
Precautions
# Never put materials directly on the pan of the balance.
# Use the weighing container or paper.
# Keep container or the weighing paper in the middle of the pan.
# Clean the balance with a soft, clean brush after use.
# Pans and the working area can be disinfected with 70% ethanol
Operating procedure
# Check the power cord
# Keep the balance ‘ON’ for 15 min before use
# Check whether the sensitivity
# Zero the balance with weighing paper keeping door closed
# Put material on weighing paper
# Read the weight of the substance directly from the screen
Note: When the balance is not in use, keep it under an airtight plastic
cover for protection against dust. Place a dish of with blue silica under the
cover to remove humidity from the air. (When the silica desiccant has
turned red, it must be regenerated by heating.)
Quality Control
Daily
Using an appropriate milligram/gram weight, check daily that the
weight registered by the balance agrees with the calibrating weights.
Record this check in the logbook.
Annually
Calibration of the balance should be carried out annually or after any
repair or re-siting by a qualified service technician and registered in
the logbook.
Repair of the balance should be carried out by a qualified service
technician.
Rotary Vacuum Evaporator
It’s a Laboratory Instruments used to reduce solvent by
reducing pressure at an elevated temperature (less than the
boiling point of the solvent to be evaporated). This
promotes the rapid removal of excess solvent from less
volatile samples.
Parts of the
1. Water bath
2. Evaporating tube coupled with a rotor
3. Condenser
4. Solvent trap
5. Chiller / cooling water
6. Vacuum pump
What to do before start rotary evaporator?
Pour the required volume of water to the water bath
Connect the tubing between vacuum pump and condenser
Connect cooling water inlet and outlet tubing
Clamp the solvent trap flask
Connect all the electric cord to the mains
 Keep close the air releasing bulb
How to start a Rotary Evaporator?
Turn the mains ON position
Turn the switch of the water bath to ON
Set the desired temperature i.e. 400C
Wait until the bath temperature reached to set temperature
Connect the round bottomed flask with sample to the evaporating tube
Close the air release bulb
Turn on the vacuum pump
Turn on the rotor
Quickly evaporating solvents
1. n-Hexane
2. DCM
3. Chloroform
4. Ethyl acetate
5. Methanol
6. Ethanol
Slowly evaporating solvent
1. n- Butanol
2. Water
How to turn off the rotary evaporator?
Stop the rotation
Release the vacuum
Take out the flask with semidried sample
Turn off the switch
Close the cooling water tap.
Maintenance and care
Distilled water should be used in the heating bath to minimize the
scale build up in the bath. Clean the water bath regularly
To remove algae gunk from the inside of a coiled water condenser, the
condenser has to be removed from the rotavap and the coil is soaked
in a dilute nitric acid solution for a few hours.
The ground glass joint holding the flask does not need to be greased,
If a mechanical pump is used instead of an aspirator to produce a
vacuum, a secondary trap has to be used to prevent that the solvent
destroys the membrane or is absorbed in the oil.
Freeze Dryer
Freeze-drying or lyophilisation, lyophilization, or cryodesiccation—
It’s a dehydration process
Solid phase to gas phase (sublime method) i.e. Ice to vapor
It is used to
# preserve a perishable material or
# make the material more convenient for transport.
# dry frozen material under reduced pressure.
1 atm = 101.325 kPa = 760 torr
70th training program of nitub shahin
Operating procedure
Turing ON the freeze dryer
 Turn on the mains
 Turn on the refrigerator
 Turn on pump after <- 400C has reached
 until the vacuum has reached < 400 mTorr
 Connect sample and open vacuum bulb
Turning OFF the freeze dryer
Release the vacuum
Turning off the pump
Defrost the chamber
Turning off the mains
Maintenance and troubleshooting
If the freeze dryer does not obtain a satisfactory vacuum, perform the
following maintenance tests
Gaskets, Tubing, Connections, Sample Valves
A. Ensure the gasket is clean. Apply vacuum grease to the gasket.
B. Check the drain line and ensure the drain plug.
C. Check all rubber vacuum tubing for signs of deterioration or cracking.
D. Check all connections and make sure they are secure and leak tight.
E. Check all sample valves on the drying chamber
F. When checking the sample valves,
g. check the vacuum pump
Centrifuge machine
It is a piece of equipment that puts an object in rotation around a fixed
axis (spins it in a circle), applying a potentially strong force perpendicular
to the axis of spin (outward).
The centrifuge works using the sedimentation principle, where the
centripetal acceleration causes denser substances and particles to move
outward in the radial direction.
Objects that are less dense are displaced and move to the center.
Denser particles to settle to the bottom of the tube, while low-density
substances rise to the top.
70th training program of nitub shahin
Maintenance and care
Avoiding Rotor Failures
Tube Breakage
Chemical Resistance : If uncommon solvents are used then precautions
should be taken
Aerosol Generation: If any liquid is spilled on a rotor, it will be dispersed
as a particulate mist when the centrifuge is run. Rotor should be
cleaned immediately.
Handling Human Samples: Human blood or blood components can
transmit an infectious disease or virus if the patient or donor carries
these.
Lubricating
Disinfection
Water bath
Used to
 incubate samples
warming of reagents,
 melting of substrates
 incubation of cell cultures
 enable certain chemical reactions
 heating flammable chemicals instead of an
open flame to prevent ignition
Precaution
 Do not use moisture sensitive substances
 Use distilled water only
 It should be use inside the fume- hood
 Raise temperature up to 900 C for
decontamination,
Melting point apparatus
Used to
 determine the melting point of a
substance.
Hot Plate with magnetic stirrer
Used for
 chemical reactions
 making homogeneous solution
Heating mantle
Used to
 apply heat to containers
 extract oil using Soxhlet extractor
 prevent excessive temperature
gradients
 distillate solvent
 heat oil bath
UV Chamber
Used to
 Set wavelength at 254 nm and
366 nm used to detect the spot or
band sample on the tlc plate
Vortex Mixer
The variable rotational speed settings
ranging from 100 to 3,200 rpm
Used to
 properly mix solution in a small vials
 mix the reagents of an assay
 mix an experimental sample and
a dilutant.
Filtration unit
Used to
 filter mobile phase
 filter solvent or mixutre
solvents
Thank you
for
your kind patience

More Related Content

PPTX
Distillation and Evaporation
PPTX
SPIRITS. And how is prepared for laboratory use
PPTX
Phenanthrene by Mr. pradeep swarnkar
PPTX
SIZE SEPARATION.pptx
PPTX
Physical pharmacy processes 1
PDF
HISTOTECHNIQUES MICROTOMY.ppt · version 1.pdf
PPTX
PPT
The animal kingdom
Distillation and Evaporation
SPIRITS. And how is prepared for laboratory use
Phenanthrene by Mr. pradeep swarnkar
SIZE SEPARATION.pptx
Physical pharmacy processes 1
HISTOTECHNIQUES MICROTOMY.ppt · version 1.pdf
The animal kingdom

What's hot (20)

PPTX
Animal kingdom Chordates
PPTX
Distillation1
PPTX
Exp. no. 01. study of apparatus used in experimental microbiology
PPTX
Phenols
PPTX
FACTORS INFLUENCING (AFFECTING) ON SOLUBILITY OF DRUGS IN SOLVENT
PPTX
Presentation on Buffer
PDF
Simple and fractional distillation
PDF
Steam-Distillation.pdf
PDF
LIPIDS ( FATS, OILS & WAXES)
PPTX
Stability of colloids
PPTX
Conductometry titration
PDF
Use of Soxhalet and rotary evaporator for extraction
PPTX
pH determination (electrometric method)
PPT
Handling of pipette ,buret,separatory funnnel, graduated cylinder
PDF
Osmosis, osmotic pressure, laws of osmotic pressure and theories of osmosis p...
PPT
POLYNUCLEAR AROMATIC HYDROCARBON- PHARMACEUTICAL ORGANIC CHEMISTRY
PPTX
Distillation by ankita yagnik
PPT
Physical processes
Animal kingdom Chordates
Distillation1
Exp. no. 01. study of apparatus used in experimental microbiology
Phenols
FACTORS INFLUENCING (AFFECTING) ON SOLUBILITY OF DRUGS IN SOLVENT
Presentation on Buffer
Simple and fractional distillation
Steam-Distillation.pdf
LIPIDS ( FATS, OILS & WAXES)
Stability of colloids
Conductometry titration
Use of Soxhalet and rotary evaporator for extraction
pH determination (electrometric method)
Handling of pipette ,buret,separatory funnnel, graduated cylinder
Osmosis, osmotic pressure, laws of osmotic pressure and theories of osmosis p...
POLYNUCLEAR AROMATIC HYDROCARBON- PHARMACEUTICAL ORGANIC CHEMISTRY
Distillation by ankita yagnik
Physical processes
Ad

Similar to 70th training program of nitub shahin (20)

PPTX
Analytical instruments in clinical chemistry, uses and maintenance
PDF
B2 df9d01
PPT
pH measurement.ppt
PDF
Manualwqc 24
PPTX
How to Maintain Your Temperature Calibration Equipment Webinar
PDF
Instruments for notes for BSc MLT, DMLT .pdf
PPT
RAC-050 Engineer Training V1.0e-202202.ppt
PDF
pHmetro OAKTAN 1100.pdf
PPTX
Qualification of instruments
PPTX
Qualification of analytical instruments
PPTX
Understand the importance of each step to minimise Laboratory errors
PPTX
Details of some Lab equipment's.
DOC
Validation and calibration of equipment
PPTX
Presentation1
PDF
pH kit by Metrohm
PPTX
CALIBRATION OF pH METER
PDF
Common Laboratory Equipment with their Working Principles
DOCX
Maa sharda pharmacy college ayodhya SOP
PDF
BTU1500-Operating-Manual.pdf
Analytical instruments in clinical chemistry, uses and maintenance
B2 df9d01
pH measurement.ppt
Manualwqc 24
How to Maintain Your Temperature Calibration Equipment Webinar
Instruments for notes for BSc MLT, DMLT .pdf
RAC-050 Engineer Training V1.0e-202202.ppt
pHmetro OAKTAN 1100.pdf
Qualification of instruments
Qualification of analytical instruments
Understand the importance of each step to minimise Laboratory errors
Details of some Lab equipment's.
Validation and calibration of equipment
Presentation1
pH kit by Metrohm
CALIBRATION OF pH METER
Common Laboratory Equipment with their Working Principles
Maa sharda pharmacy college ayodhya SOP
BTU1500-Operating-Manual.pdf
Ad

More from Muhammad Tareque Hasan (13)

PPTX
Analytical microbiology
PPT
PPTX
Blood culture
PPTX
Nitub workshop july 2018
PPT
PPTX
Nitub 2018 mou
PPTX
Microscope 14.07.2018
PPT
Israt ara hossain nst 17-18
PPT
Elisa ppt nitub-02
PPTX
Automated blood cell counter by prof. dr. ma mohsin
PPT
2. biomolecule
PPT
Spectrophotometry july 2018 nitub
Analytical microbiology
Blood culture
Nitub workshop july 2018
Nitub 2018 mou
Microscope 14.07.2018
Israt ara hossain nst 17-18
Elisa ppt nitub-02
Automated blood cell counter by prof. dr. ma mohsin
2. biomolecule
Spectrophotometry july 2018 nitub

Recently uploaded (20)

PPTX
Post Op complications in general surgery
PPTX
thio and propofol mechanism and uses.pptx
PPT
Dermatology for member of royalcollege.ppt
PPTX
IMAGING EQUIPMENiiiiìiiiiiTpptxeiuueueur
PDF
Transcultural that can help you someday.
DOCX
PEADIATRICS NOTES.docx lecture notes for medical students
PDF
Plant-Based Antimicrobials: A New Hope for Treating Diarrhea in HIV Patients...
PPTX
Manage HIV exposed child and a child with HIV infection.pptx
PPTX
Approach to chest pain, SOB, palpitation and prolonged fever
PPT
Rheumatology Member of Royal College of Physicians.ppt
PPTX
Vaccines and immunization including cold chain , Open vial policy.pptx
PDF
The_EHRA_Book_of_Interventional Electrophysiology.pdf
PPT
neurology Member of Royal College of Physicians (MRCP).ppt
PDF
Lecture on Anesthesia for ENT surgery 2025pptx.pdf
PPTX
09. Diabetes in Pregnancy/ gestational.pptx
PDF
Lecture 8- Cornea and Sclera .pdf 5tg year
PDF
B C German Homoeopathy Medicineby Dr Brij Mohan Prasad
PPTX
HYPERSENSITIVITY REACTIONS - Pathophysiology Notes for Second Year Pharm D St...
PPT
nephrology MRCP - Member of Royal College of Physicians ppt
PPTX
Introduction to Medical Microbiology for 400L Medical Students
Post Op complications in general surgery
thio and propofol mechanism and uses.pptx
Dermatology for member of royalcollege.ppt
IMAGING EQUIPMENiiiiìiiiiiTpptxeiuueueur
Transcultural that can help you someday.
PEADIATRICS NOTES.docx lecture notes for medical students
Plant-Based Antimicrobials: A New Hope for Treating Diarrhea in HIV Patients...
Manage HIV exposed child and a child with HIV infection.pptx
Approach to chest pain, SOB, palpitation and prolonged fever
Rheumatology Member of Royal College of Physicians.ppt
Vaccines and immunization including cold chain , Open vial policy.pptx
The_EHRA_Book_of_Interventional Electrophysiology.pdf
neurology Member of Royal College of Physicians (MRCP).ppt
Lecture on Anesthesia for ENT surgery 2025pptx.pdf
09. Diabetes in Pregnancy/ gestational.pptx
Lecture 8- Cornea and Sclera .pdf 5tg year
B C German Homoeopathy Medicineby Dr Brij Mohan Prasad
HYPERSENSITIVITY REACTIONS - Pathophysiology Notes for Second Year Pharm D St...
nephrology MRCP - Member of Royal College of Physicians ppt
Introduction to Medical Microbiology for 400L Medical Students

70th training program of nitub shahin

  • 2. Use, Maintenance and Troubleshooting of Small Lab- Instruments Dr. Md Shahinul Haque Khan Professor (cc) Department of Chemistry Bangladesh University of Health Sciences 125/1, Darus salam, Mirpur 1 Emai: khanshahinul@yahoo.com
  • 3. Name of the common Instruments used in a Laboratory 1. Ultrasonic cleaner 2. Homogenizer 3. Drying oven 4. Desiccators 5. pH meter 6. Balance, Weight box 7. Rotary Vacuum Evaporator 8. Freeze dryer 9. Centrifuge machine 10. Water bath 11. Melting point apparatus 12. Hot plate with Magnetic stirrer 13. Heating Mantle 14. UV Chamber 15. Vortex Mixer 16. Filtration Unit
  • 4. Ultrasonic Cleaner Ultrasonic cleaning is a process that uses ultrasound (usually from 20–400 kHz) and an appropriate cleaning solvent to clean items.  Ultrasonic cleaning uses cavitation bubbles induced by high frequency pressure (sound) waves to agitate a liquid. The agitation produces high forces on contaminants adhering to substrates like metals, plastics, glass, rubber, and ceramics.  used to clean many different types of objects, including jewelry, lenses and other optical parts, watches, dental and surgical instruments, tools, car fuel injectors, musical instruments, industrial parts and electronic equipment. electronic repair workshops.
  • 5. Homogenizer Used for  the homogenization of various types of material, such as tissue, plant, food, soil, and many others.  sample preparation step prior to the analysis of nucleic acids, proteins, Cell, pathogens, and many other targets.
  • 7. Desiccators Desiccators are sealable or sealable with vacuum enclosures containing desiccants. It is used to  preserving moisture-sensitive items,  protect chemicals which are hygroscopic ,  prevent reaction with water from humidity,  keep materials in drying condition,  reduce temperature during moisture determination
  • 10. UV Blocking Vacuum Desiccator
  • 11. Sizes of Desiccators 1. Desiccator, Glass, Clear, I.D. 110 mm 2. Desiccator, Glass, Clear, I.D. 180 mm 3. Desiccator, Glass, Clear, I.D. 240 mm 4. Vacuum Desiccator, Glass, Amber, I.D. 240 mm 5. Space Saver Vacuum Desiccator, 19.7cm ID 6. Desiccator, Glass, Clear, I.D. 150 mm 7. Cabinet type
  • 12. Desiccant A desiccant is a hygroscopic substance that induces or sustains a state of dryness in its vicinity. Characteristics : chemically inert, non-toxic, water insoluble Performance efficiency # ratio or percentage (water storable : mass of desiccant) # relative humidity of the air or other substance being dried
  • 14. pH meter It is a scientific instrument used to measure  the hydrogen-ion activity in water-based solutions,  the acidity or alkalinity,  the difference in electrical potential between a pH electrode and a reference electrode.
  • 15. Electrode: # pH electrode and # reference electrode design
  • 16. pH-sensitive glass membrane, 0.4 mm thickness
  • 17. Construction of an electrode 1. A sensing part, a bulb made from a specific glass 2. Internal electrode, silver chloride electrode or calomel electrode 3. Internal solution, usually a pH=7 buffered solution of 0.1 mol/L KCl for pH electrodes. 4. When using the silver chloride electrode, a small amount of AgCl can precipitate inside the glass electrode 5. Reference electrode, 6. Reference internal solution, usually 0.1 mol/L KCl 7. Junction with studied solution, usually made from ceramics or capillary with asbestos or quartz fiber. 8. Body of electrode, made from non-conductive glass or plastics.
  • 19. Selection of pH electrode The types of pH electrodes are many and various. The epoxy bodied combination electrode is suitable for the majority of tests carried out on aqueous solutions. For other applications, such as low ionic strength, high temperatures and strongly acidic solutions a more suitable pH/reference electrode pair may be required
  • 20. Storing the pH electrode 1. After Use - Rinse thoroughly with distilled water. Short Term Storage - Immerse in storage solution Long Term Storage - Fit wetting cap filled with storage solution and replace in original packaging. 2. Electrodes should be stored a) away from direct sunlight b) in a vertical position c) within their specified temperature range 3. Always ensure the electrode is used within its specified temperature range. Ageing of electrodes used above their specified temperature is rapid and irreversible.
  • 21. Precautions 1. DO NOT touch the sensitive glass pH membrane or reference junction. Excess droplets of solution may be removed by gently blotting with filter paper or tissue. 2. DO NOT wipe the electrode as this may induce an electrostatic charge
  • 22. CLEANING OF GLASS ELECTRODES Deposit Cleaning agents General deposits Mild detergent solution Inorganic coatings Commercial glass cleaning solution (not strongly acidic) Metal compounds Acid solution, not stronger than 1M Oil/Grease Complexing agent (EDTA) or suitable solvent Resins/Lignins Acetone, alcohol or detergent (not strongly alkaline). Proteins (blood, etc) Enzyme solutions e.g. Pepsin in 0.1M HCl Stubborn deposits Weak hydrogen peroxide solution, Sodium Hypochlorite solution or domestic bleach. If Electrodes is dried out, (dry deposit of KCl crystals on the electrode body), by soaking overnight in warm deionised water. Note: The epoxy bodied electrode should not be cleaned with aggressive solvents.
  • 23. ROUBLESHOOTING GUIDE FAULT PROBABLE CAUSE ACTION power a. Battery failure b. Battery polarity incorrect a. Replace batteries b. Refit batteries Cal error 1, 2 or 3 a. Incorrect value buffer used b. Contaminated buffer solution a. Use correct value b. Use fresh buffer solutions Cal error No 1 or 2 a. pH electrode defective b. Same buffer used for 2 point cal a. Replace pH electrode b. Recalibrate by 2 point buffer solution Unstable display pH electrode defective Replace pH electrode Display shows 1 or - 1 only a. Intermittant or no connection b. pH electrode defective c. Contaminated solutions a. Check electrode connection to unit b. Replace pH electrode c. Replace solutions Intermittant display Probes not fitted correctly Check connections I/O switch not working a. I/O key not held down >2 secs b. Batteries not fitted a. Retry holding key for 2 seconds b. Fit batteries
  • 24. CHECKING the pH METER FUNCTION  Adjust temperature to 25°C using the keys.  Check buffer solution  pH 7 and the display should read 7 pH.  pH 4 and the display should read 4 pH.  pH 10 the display should show 10 pH.  the run unknown sample to read its pH value.
  • 25. Weighing Scales These are devices to measure weight or calculate mass. a. Spring balances or spring scales: measure force (weight) of gravity against the force on a spring. Example: electronic balance, digital scale. b. A balance or pair of scales: compares masses by balancing the weight due to the mass of an object against the weight of a known mass. Example: Pan balance
  • 26. Installation of a balance # On a solid, vibration-free surface, away from direct sunlight and at an constant temperature. # precisely levelled, using the built-in spirit level: # Protect the balance from drafts of air. Air moving across the pans will cause inaccurate readings. # Always keep the balance and weights clean and dry to protect them from corrosion. # Check the weight 10 mg, 50 mg and 100 mg and keep note Other items Weighing paper; Set of calibrating weights; Dust brush. Silica blue desiccant
  • 27. Precautions # Never put materials directly on the pan of the balance. # Use the weighing container or paper. # Keep container or the weighing paper in the middle of the pan. # Clean the balance with a soft, clean brush after use. # Pans and the working area can be disinfected with 70% ethanol
  • 28. Operating procedure # Check the power cord # Keep the balance ‘ON’ for 15 min before use # Check whether the sensitivity # Zero the balance with weighing paper keeping door closed # Put material on weighing paper # Read the weight of the substance directly from the screen Note: When the balance is not in use, keep it under an airtight plastic cover for protection against dust. Place a dish of with blue silica under the cover to remove humidity from the air. (When the silica desiccant has turned red, it must be regenerated by heating.)
  • 29. Quality Control Daily Using an appropriate milligram/gram weight, check daily that the weight registered by the balance agrees with the calibrating weights. Record this check in the logbook. Annually Calibration of the balance should be carried out annually or after any repair or re-siting by a qualified service technician and registered in the logbook. Repair of the balance should be carried out by a qualified service technician.
  • 30. Rotary Vacuum Evaporator It’s a Laboratory Instruments used to reduce solvent by reducing pressure at an elevated temperature (less than the boiling point of the solvent to be evaporated). This promotes the rapid removal of excess solvent from less volatile samples. Parts of the 1. Water bath 2. Evaporating tube coupled with a rotor 3. Condenser 4. Solvent trap 5. Chiller / cooling water 6. Vacuum pump
  • 31. What to do before start rotary evaporator? Pour the required volume of water to the water bath Connect the tubing between vacuum pump and condenser Connect cooling water inlet and outlet tubing Clamp the solvent trap flask Connect all the electric cord to the mains  Keep close the air releasing bulb
  • 32. How to start a Rotary Evaporator? Turn the mains ON position Turn the switch of the water bath to ON Set the desired temperature i.e. 400C Wait until the bath temperature reached to set temperature Connect the round bottomed flask with sample to the evaporating tube Close the air release bulb Turn on the vacuum pump Turn on the rotor
  • 33. Quickly evaporating solvents 1. n-Hexane 2. DCM 3. Chloroform 4. Ethyl acetate 5. Methanol 6. Ethanol Slowly evaporating solvent 1. n- Butanol 2. Water
  • 34. How to turn off the rotary evaporator? Stop the rotation Release the vacuum Take out the flask with semidried sample Turn off the switch Close the cooling water tap.
  • 35. Maintenance and care Distilled water should be used in the heating bath to minimize the scale build up in the bath. Clean the water bath regularly To remove algae gunk from the inside of a coiled water condenser, the condenser has to be removed from the rotavap and the coil is soaked in a dilute nitric acid solution for a few hours. The ground glass joint holding the flask does not need to be greased, If a mechanical pump is used instead of an aspirator to produce a vacuum, a secondary trap has to be used to prevent that the solvent destroys the membrane or is absorbed in the oil.
  • 36. Freeze Dryer Freeze-drying or lyophilisation, lyophilization, or cryodesiccation— It’s a dehydration process Solid phase to gas phase (sublime method) i.e. Ice to vapor It is used to # preserve a perishable material or # make the material more convenient for transport. # dry frozen material under reduced pressure.
  • 37. 1 atm = 101.325 kPa = 760 torr
  • 39. Operating procedure Turing ON the freeze dryer  Turn on the mains  Turn on the refrigerator  Turn on pump after <- 400C has reached  until the vacuum has reached < 400 mTorr  Connect sample and open vacuum bulb
  • 40. Turning OFF the freeze dryer Release the vacuum Turning off the pump Defrost the chamber Turning off the mains
  • 41. Maintenance and troubleshooting If the freeze dryer does not obtain a satisfactory vacuum, perform the following maintenance tests Gaskets, Tubing, Connections, Sample Valves A. Ensure the gasket is clean. Apply vacuum grease to the gasket. B. Check the drain line and ensure the drain plug. C. Check all rubber vacuum tubing for signs of deterioration or cracking. D. Check all connections and make sure they are secure and leak tight. E. Check all sample valves on the drying chamber F. When checking the sample valves, g. check the vacuum pump
  • 42. Centrifuge machine It is a piece of equipment that puts an object in rotation around a fixed axis (spins it in a circle), applying a potentially strong force perpendicular to the axis of spin (outward). The centrifuge works using the sedimentation principle, where the centripetal acceleration causes denser substances and particles to move outward in the radial direction. Objects that are less dense are displaced and move to the center. Denser particles to settle to the bottom of the tube, while low-density substances rise to the top.
  • 44. Maintenance and care Avoiding Rotor Failures Tube Breakage Chemical Resistance : If uncommon solvents are used then precautions should be taken Aerosol Generation: If any liquid is spilled on a rotor, it will be dispersed as a particulate mist when the centrifuge is run. Rotor should be cleaned immediately. Handling Human Samples: Human blood or blood components can transmit an infectious disease or virus if the patient or donor carries these. Lubricating Disinfection
  • 45. Water bath Used to  incubate samples warming of reagents,  melting of substrates  incubation of cell cultures  enable certain chemical reactions  heating flammable chemicals instead of an open flame to prevent ignition Precaution  Do not use moisture sensitive substances  Use distilled water only  It should be use inside the fume- hood  Raise temperature up to 900 C for decontamination,
  • 46. Melting point apparatus Used to  determine the melting point of a substance.
  • 47. Hot Plate with magnetic stirrer Used for  chemical reactions  making homogeneous solution
  • 48. Heating mantle Used to  apply heat to containers  extract oil using Soxhlet extractor  prevent excessive temperature gradients  distillate solvent  heat oil bath
  • 49. UV Chamber Used to  Set wavelength at 254 nm and 366 nm used to detect the spot or band sample on the tlc plate
  • 50. Vortex Mixer The variable rotational speed settings ranging from 100 to 3,200 rpm Used to  properly mix solution in a small vials  mix the reagents of an assay  mix an experimental sample and a dilutant.
  • 51. Filtration unit Used to  filter mobile phase  filter solvent or mixutre solvents