Get 1cc Autogenous Bone within Seconds
ACMAuto Chip Maker
Contents
• Indication
• Specification and Type
• Precautions
• Caution
• Cleaning
• Sterilization
• Storage and Management
• Clinical case
• Clinical Movie
Indication
ACM(Autochip Maker) is an instrument to collect autograft bone from mandibular molars or incisor for GBR in
defected area
Precautions
1. Instrument is susceptible to damage and wear and should be inspected before use.
2. Check the latch lock shank for wear to ensure that the connection is not damaged by reused instrument.
3. Instrument must be cleaned and sterilized prior to use
4. Stopper is ready to use. Do not sterilize plastic stopper in autoclave prior to use.
Specification and Type
Diameter Length(mm) Code
4.0
10mm ACM40ISETES
14mm ACM40ISETS
4.5
10mm ACM45ISETES
14mm ACM45ISETS
5.0
10mm ACM50ISETES
14mm ACM50ISETS
6.0
10mm ACM60ISETES
14mm ACM60ISETS
1 box : 1drill and 7 stoppers
1. Mount stopper to ACM and connect dental implant motor.
2. Locate the drill on the surface of operation area by pressing it slightly with saline irrigation.
✓ Recommended drilling speed is 50~300rpm and maximum torque over 50Ncm.
(50rpm without irrigation, Over 100rpm with irrigation)
✓ The stopper does not go deeper than 4mm.
✓ Maximum 4mm depth autograft bone, drill until 3~4mm depth and move to different area to collect the
same quality bone chip.
✓ Drill with shaking to left and right for effective cutting and cooling.
3. When autograft bone is fully filled in the space, take out the bone chips from ACM to the medical bowl after
removing the stopper.
4. Repeat procedure 1,2,3 until taking the enough autograft bone.
5. GBR with the collected autograft bone in the defected area.
If the thickness of cortical bone is below 4mm, drill can be stopped by over torque due to touching
the cancellous bone.
To prevent drill stop, drill below 4mm to collect cortical bone only.
Otherwise shake toward left and right with maximum torque.
!
Procedure
1. Must irrigate with saline while drilling to prevent bone heating.
2. Recommendation speed is 50~300rpm and maximum torque over 50Ncm.
3. Recommended using times are below 7 to prevent bone necrosis caused by drill abrasion.
4. Clean the instrument immediately as the blood, tissue remains, bone debris or secretion could be dried on it.
!
1. Disassemble multi-piece components. Rinse with cool to lukewarm water for two and one half minutes.
2. Place all parts in an ultrasonic cleaner with an enzymatic detergent diluted with tap water per the manufacture’s
guidelines. Sonicate for ten minutes. Rinse with tap water for three minutes.
Cleaning
Caution
Stopper is disposable. When sterilizing plastic stopper in autoclave, it will melt down.
Remove all protective packaging and separate stopper from ACM before sterilization to ensure maximum efficacy.
!
* Note : To ensure autoclave is performing effectively, periodic use of biologic indicators should be considered. Dry
heat and Chemclave sterilization are NOT recommended.
1 Minimum validated steam sterilization time and temperature required to achieve a 10-6 sterility assurance level(SAL)
2 Local or national specifications should be followed where steam sterilization requirements are stricter or more conservative
than those listed in this table.
3 Sea level
Sterilization
Cycle Type Temperature Pressure 3 Exposure Time Dry Time
Instrument,
Wrapped
Pre-Vacuum 1,2 132℃
270℉
2 bars
28.5 psi
3 minutes 30 minutes
Instrument,
Wrapped
Gravity 1 121℃
250℉
1 bars
14.5 psi
40 minutes 30 minutes
Recommended Steam Sterilization Parameters
1. Store in a dry area at room temperature.
2. Used product must be cleaned, sterilized and dried befor storage.
3. Discard the instrument which used 7 times.
4. Reuse instrument may affect cutting efficiency and damage of latch lock shank. If inspection reveals signs of
wear or damage, replace the instruments.
Storage and Management
Clinical Case
Clinical Movie
ACMAuto Chip Maker

More Related Content

PDF
20150827 user guide (sinus all)
PDF
Clinical case SR Kit with Movie
PDF
Sr 2 kit version 01
PDF
Sca user guide
PDF
FR Kit user guide
PDF
User guide universal prosthetic kit -151207
PDF
IS II active Catalog
PDF
Sla user guide
20150827 user guide (sinus all)
Clinical case SR Kit with Movie
Sr 2 kit version 01
Sca user guide
FR Kit user guide
User guide universal prosthetic kit -151207
IS II active Catalog
Sla user guide

What's hot (20)

PDF
Osung catalogue 2013
PDF
SCA Kit case(Neobiotech)
PDF
2 brochure dhsp-dcsp
PDF
Neo Biotech Implant System Catalog
PPT
Abutments design & selection /orthodontics training courses
DOC
HOT Screen Spare Parts & Maintenance
PDF
Orbit valves-brochure
PPT
2013 implant restorative overview
PPTX
IOLCatridge-Injector Systems
PDF
5840 Brochure
PPT
Basic implant surgeryraju/prosthodontic courses
PDF
T308 epoxy-summary-brochure-1299505
PDF
How to use thread plug gages
PDF
Split Rotary Seals Brochure Small
PDF
PPTX
Gingival retraction techniques for implants versus teeth/ cosmetic dentistry ...
PDF
How to use thread ring gages
PPT
Fluid control and Soft tissue management
PPTX
Debonding of orhtodontic brackets
Osung catalogue 2013
SCA Kit case(Neobiotech)
2 brochure dhsp-dcsp
Neo Biotech Implant System Catalog
Abutments design & selection /orthodontics training courses
HOT Screen Spare Parts & Maintenance
Orbit valves-brochure
2013 implant restorative overview
IOLCatridge-Injector Systems
5840 Brochure
Basic implant surgeryraju/prosthodontic courses
T308 epoxy-summary-brochure-1299505
How to use thread plug gages
Split Rotary Seals Brochure Small
Gingival retraction techniques for implants versus teeth/ cosmetic dentistry ...
How to use thread ring gages
Fluid control and Soft tissue management
Debonding of orhtodontic brackets
Ad

Viewers also liked (7)

PDF
Clinical case (ibrush)
PDF
Peri implant treatment Step
PDF
2015 Company & Product Introduction
PDF
GBR Kit catalog
PDF
CTi mem new version 01
PDF
Prosthetic guide - 151210 update
PDF
Neo surgical accessory kit
Clinical case (ibrush)
Peri implant treatment Step
2015 Company & Product Introduction
GBR Kit catalog
CTi mem new version 01
Prosthetic guide - 151210 update
Neo surgical accessory kit
Ad

More from Kei Lim (17)

PDF
201601001 Guide seminar
PDF
User guide NEO GUIDE SYSTEM - 20160621
PDF
20160222 Neobiotech article collection.
PDF
20160626 symposium leaflet v04
PDF
Any time loading proof leaflet 151217
PDF
Predictable and Accurate 3D Guided implant placement for a 3-D world
PDF
20151021 CAMELEON 5AXIS
PDF
2015 NEOBIOTECH CADCAM SYSTEM
PDF
ACM poster from Michigan University
PDF
Live surgery Sketch
PDF
20150925 jin y kim basic
PDF
CAMELEON CS (Chairside milling)
PDF
Neo guide system_201505_ver04
PDF
Prosthetic guide leaflet
PDF
CADCAM system
PDF
20150625 tijuana Implant live surgery
PDF
Smiler ITS Live Surgery Course
201601001 Guide seminar
User guide NEO GUIDE SYSTEM - 20160621
20160222 Neobiotech article collection.
20160626 symposium leaflet v04
Any time loading proof leaflet 151217
Predictable and Accurate 3D Guided implant placement for a 3-D world
20151021 CAMELEON 5AXIS
2015 NEOBIOTECH CADCAM SYSTEM
ACM poster from Michigan University
Live surgery Sketch
20150925 jin y kim basic
CAMELEON CS (Chairside milling)
Neo guide system_201505_ver04
Prosthetic guide leaflet
CADCAM system
20150625 tijuana Implant live surgery
Smiler ITS Live Surgery Course

Recently uploaded (20)

PDF
Glaucoma Definition, Introduction, Etiology, Epidemiology, Clinical Presentat...
PDF
04 dr. Rahajeng - dr.rahajeng-KOGI XIX 2025-ed1.pdf
PPTX
thio and propofol mechanism and uses.pptx
PPT
nephrology MRCP - Member of Royal College of Physicians ppt
PPTX
Neoplasia III.pptxjhghgjhfj fjfhgfgdfdfsrbvhv
PDF
The_EHRA_Book_of_Interventional Electrophysiology.pdf
PDF
Forensic Psychology and Its Impact on the Legal System.pdf
PDF
OSCE SERIES - Set 7 ( Questions & Answers ).pdf
PPTX
Reading between the Rings: Imaging in Brain Infections
PPTX
y4d nutrition and diet in pregnancy and postpartum
PDF
B C German Homoeopathy Medicineby Dr Brij Mohan Prasad
PDF
OSCE SERIES ( Questions & Answers ) - Set 5.pdf
PPTX
Impression Materials in dental materials.pptx
PPTX
HYPERSENSITIVITY REACTIONS - Pathophysiology Notes for Second Year Pharm D St...
PDF
Lecture 8- Cornea and Sclera .pdf 5tg year
PPTX
@K. CLINICAL TRIAL(NEW DRUG DISCOVERY)- KIRTI BHALALA.pptx
PPTX
CARDIOVASCULAR AND RENAL DRUGS.pptx for health study
PDF
AGE(Acute Gastroenteritis)pdf. Specific.
PPTX
Antepartum_Haemorrhage_Guidelines_2024.pptx
PDF
Nursing manual for conscious sedation.pdf
Glaucoma Definition, Introduction, Etiology, Epidemiology, Clinical Presentat...
04 dr. Rahajeng - dr.rahajeng-KOGI XIX 2025-ed1.pdf
thio and propofol mechanism and uses.pptx
nephrology MRCP - Member of Royal College of Physicians ppt
Neoplasia III.pptxjhghgjhfj fjfhgfgdfdfsrbvhv
The_EHRA_Book_of_Interventional Electrophysiology.pdf
Forensic Psychology and Its Impact on the Legal System.pdf
OSCE SERIES - Set 7 ( Questions & Answers ).pdf
Reading between the Rings: Imaging in Brain Infections
y4d nutrition and diet in pregnancy and postpartum
B C German Homoeopathy Medicineby Dr Brij Mohan Prasad
OSCE SERIES ( Questions & Answers ) - Set 5.pdf
Impression Materials in dental materials.pptx
HYPERSENSITIVITY REACTIONS - Pathophysiology Notes for Second Year Pharm D St...
Lecture 8- Cornea and Sclera .pdf 5tg year
@K. CLINICAL TRIAL(NEW DRUG DISCOVERY)- KIRTI BHALALA.pptx
CARDIOVASCULAR AND RENAL DRUGS.pptx for health study
AGE(Acute Gastroenteritis)pdf. Specific.
Antepartum_Haemorrhage_Guidelines_2024.pptx
Nursing manual for conscious sedation.pdf

ACM user guide

  • 1. Get 1cc Autogenous Bone within Seconds ACMAuto Chip Maker
  • 2. Contents • Indication • Specification and Type • Precautions • Caution • Cleaning • Sterilization • Storage and Management • Clinical case • Clinical Movie
  • 3. Indication ACM(Autochip Maker) is an instrument to collect autograft bone from mandibular molars or incisor for GBR in defected area Precautions 1. Instrument is susceptible to damage and wear and should be inspected before use. 2. Check the latch lock shank for wear to ensure that the connection is not damaged by reused instrument. 3. Instrument must be cleaned and sterilized prior to use 4. Stopper is ready to use. Do not sterilize plastic stopper in autoclave prior to use. Specification and Type Diameter Length(mm) Code 4.0 10mm ACM40ISETES 14mm ACM40ISETS 4.5 10mm ACM45ISETES 14mm ACM45ISETS 5.0 10mm ACM50ISETES 14mm ACM50ISETS 6.0 10mm ACM60ISETES 14mm ACM60ISETS 1 box : 1drill and 7 stoppers
  • 4. 1. Mount stopper to ACM and connect dental implant motor. 2. Locate the drill on the surface of operation area by pressing it slightly with saline irrigation. ✓ Recommended drilling speed is 50~300rpm and maximum torque over 50Ncm. (50rpm without irrigation, Over 100rpm with irrigation) ✓ The stopper does not go deeper than 4mm. ✓ Maximum 4mm depth autograft bone, drill until 3~4mm depth and move to different area to collect the same quality bone chip. ✓ Drill with shaking to left and right for effective cutting and cooling. 3. When autograft bone is fully filled in the space, take out the bone chips from ACM to the medical bowl after removing the stopper. 4. Repeat procedure 1,2,3 until taking the enough autograft bone. 5. GBR with the collected autograft bone in the defected area. If the thickness of cortical bone is below 4mm, drill can be stopped by over torque due to touching the cancellous bone. To prevent drill stop, drill below 4mm to collect cortical bone only. Otherwise shake toward left and right with maximum torque. ! Procedure
  • 5. 1. Must irrigate with saline while drilling to prevent bone heating. 2. Recommendation speed is 50~300rpm and maximum torque over 50Ncm. 3. Recommended using times are below 7 to prevent bone necrosis caused by drill abrasion. 4. Clean the instrument immediately as the blood, tissue remains, bone debris or secretion could be dried on it. ! 1. Disassemble multi-piece components. Rinse with cool to lukewarm water for two and one half minutes. 2. Place all parts in an ultrasonic cleaner with an enzymatic detergent diluted with tap water per the manufacture’s guidelines. Sonicate for ten minutes. Rinse with tap water for three minutes. Cleaning Caution
  • 6. Stopper is disposable. When sterilizing plastic stopper in autoclave, it will melt down. Remove all protective packaging and separate stopper from ACM before sterilization to ensure maximum efficacy. ! * Note : To ensure autoclave is performing effectively, periodic use of biologic indicators should be considered. Dry heat and Chemclave sterilization are NOT recommended. 1 Minimum validated steam sterilization time and temperature required to achieve a 10-6 sterility assurance level(SAL) 2 Local or national specifications should be followed where steam sterilization requirements are stricter or more conservative than those listed in this table. 3 Sea level Sterilization Cycle Type Temperature Pressure 3 Exposure Time Dry Time Instrument, Wrapped Pre-Vacuum 1,2 132℃ 270℉ 2 bars 28.5 psi 3 minutes 30 minutes Instrument, Wrapped Gravity 1 121℃ 250℉ 1 bars 14.5 psi 40 minutes 30 minutes Recommended Steam Sterilization Parameters
  • 7. 1. Store in a dry area at room temperature. 2. Used product must be cleaned, sterilized and dried befor storage. 3. Discard the instrument which used 7 times. 4. Reuse instrument may affect cutting efficiency and damage of latch lock shank. If inspection reveals signs of wear or damage, replace the instruments. Storage and Management