Atomic Force Microscope (AFM)
Introduction
The atomic force microscope (AFM) was invented
in1986 by Binnig, Quate and Gerber.
The AFM raster scans a sharp probe over the surface
of a sample and measures the changes in force
between the probe tip and the sample.
2
Working Concept
The physical parameter probed is a force resulting from different
interactions.
Thus, an AFM image is generated by recording the force changes as
the probe (or sample) is scanned in the x and y directions.
The sample is mounted on a piezoelectric scanner, which ensures
three-dimensional positioning with high resolution.
The force is monitored by attaching the probe to a pliable cantilever,
which acts as a spring, and measuring the bending or "deflection" of
the cantilever.
3
Principle
• The AFM consists of a cantilever with a sharp tip (probe) at its end that is
used to scan the specimen surface.
• The cantilever is typically silicon or silicon nitride with a tip radius of
curvature on the order of nanometers.
• When the tip is brought into proximity of a sample surface, forces between
the tip and the sample lead to a deflection of the cantilever according
to Hooke's law.
• Depending on the situation, forces that are measured in AFM include
mechanical contact force, van der Waals forces, capillary forces, chemical
bonding, electrostatic forces.
4
5
Working concept of AFM
Basic set-up of an AFM
The ability of an AFM to achieve near atomic scale
resolution depends on the three essential
components:
(1) a cantilever with a sharp tip,
(2) a scanner that controls the x-y-z position, and
(3) the feedback control and loop.
6
• Cantiliever with a sharp tip. The stiffness of the cantilever
needs to be less the effective spring constant holding atoms
together, which is on the order of 1 - 10 nm.
• The tip should have a radius of curvature less than 20-50 nm
(smaller is better) a cone angle between 10-20 degrees.
• 2. Scanner. The movement of the tip in the sample in the x, y,
and z-directions is controlled by a piezo-electric tube scanner.
• For typical AFM scanners, the maximum ranges for are 80 mm
x 80 mm in the x-y plane and 5 mm for the z-direction.
7
3. Feedback control. The forces that are exerted
between the tip and the sample are measured by the
amount of bending (or deflection) of the cantilever.
By calculating the difference signal in the photodiode
quadrants, the amount of deflection can be
correlated with a height .
Because the cantilever obeys Hooke's Law for small
displacements, the interaction force between the tip
and the sample can be determined.
8
Imaging Modes
• Contact Mode
• Tapping Mode
• Non contact Mode
9
Contact mode
• The tip is "dragged" across the surface of the sample and the contours of
the surface are measured either using the deflection of the cantilever
directly or, more commonly, using the feedback signal required to keep the
cantilever at a constant position.
• Because the measurement of a static signal is prone to noise and drift, low
stiffness cantilevers are used to achieve a large enough deflection signal
while keeping the interaction force low.
• Thus, contact mode AFM is almost always done at a depth where the
overall force is repulsive, that is, in firm "contact" with the solid surface.
10
Tapping mode
• The cantilever is driven to oscillate up and down at or near its resonance
frequency.
• The amplitude of this oscillation usually varies from several nm to 200 nm.
• The interaction of forces acting on the cantilever when the tip comes close
to the surface, Van der Waals forces, dipole-dipole
interactions, electrostatic forces, etc.
• This causes the amplitude of the cantilever's oscillation to change (usually
decrease) as the tip gets closer to the sample.
• This amplitude is used as the parameter that goes into the electronic
servo that controls the height of the cantilever above the sample.
• A tapping AFM image is therefore produced by imaging the force of the
intermittent contacts of the tip with the sample surface.
11
Non contact method
• In non-contact mode, the tip of the cantilever does not contact the sample
surface.
• The cantilever is instead oscillated at either its resonant frequency or just
above (amplitude modulation) where the amplitude of oscillation is
typically a few nanometers (<10 nm) down to a few picometers.
• The van der Waals forces, which are strongest from 1 nm to 10 nm above
the surface acts to decrease the resonance frequency of the cantilever.
• Measuring the tip-to-sample distance at each (x,y) data point allows the
scanning software to construct a topographic image of the sample surface.
• Non-contact mode AFM does not suffer from tip or sample degradation
effects that are sometimes observed after taking numerous scans with
contact AFM. This makes non-contact AFM preferable to contact AFM for
measuring soft samples, e.g. biological samples and organic thin film.
12
Applications
The AFM is useful for obtaining three-dimensional topographic
information of insulating and conducting structures with lateral
resolution down to 1.5 nm and vertical resolution down to 0.05 nm.
These samples include clusters of atoms and molecules, individual
macromolecules, and biologic al species (cells, DNA, proteins).
Minimal sample preparation involved for AFM imaging.
The AFM can operate in gas, ambient, and fluid environments and
can measure physical properties including elasticity, adhesion,
hardness, friction and chemical functionality.
13
Advantages
The AFM has several advantages over the scanning electron
microscope (SEM).
Unlike the electron microscope which provides a two-
dimensional projection or a two-dimensional image of a
sample, the AFM provides a true three-dimensional surface
profile.
Additionally, samples viewed by AFM do not require any
special treatments (such as metal/carbon coatings) that would
irreversibly change or damage the sample.
14
While an electron microscope needs an expensive
vacuum environment for proper operation, most AFM
modes can work perfectly well in ambient air or even a
liquid environment.
This makes it possible to study biological
macromolecules and even living organisms.
In principle, AFM can provide higher resolution than
SEM.
15
Disadvantages
A disadvantage of AFM compared with the scanning electron
microscope (SEM) is the image size.
The SEM can image an area on the order of millimetres by millimetres
with a depth of field on the order of millimetres.
The AFM can only image a maximum height on the order of
micrometres and a maximum scanning area of around 150 by 150
micrometres.
Another inconvenience is that at high resolution, the quality of an
image is limited by the radius of curvature of the probe tip, and an
incorrect choice of tip for the required resolution can lead to image
artifacts.
16

More Related Content

PPT
Atomic force microscopy principle and application
PPT
Atomic force microscope : principle and application
PDF
AFM technique aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
PDF
Lecture_10_Atomic force microscopy(1).pdf
PPTX
AFM(Atomic Force microscope ) Nanotechnology
PPT
Microscopy
PPTX
Atomic force microscope (AFM) full .pptx
PPTX
AFM and STM (Scanning probe microscopy)
Atomic force microscopy principle and application
Atomic force microscope : principle and application
AFM technique aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
Lecture_10_Atomic force microscopy(1).pdf
AFM(Atomic Force microscope ) Nanotechnology
Microscopy
Atomic force microscope (AFM) full .pptx
AFM and STM (Scanning probe microscopy)

Similar to Atomic force microscopy presentation. Ppt (20)

PPTX
Non contact mode (AFM)
PPTX
Scanning Tunneling Microscope and Atomic Tunneling Microscope
PPTX
Atomic Force Microscope: Fundamental Principles
PPTX
Charecterization techniques
PPT
Atomic force microscopy
PPTX
Atomic force microscopy
PPTX
Atomic Force Microscopy (AFM)
PPTX
Atomic force microscope
PPT
Microscopic examination using Atomic force microscopy and Confocal scanning ...
PPT
PPTX
Atomic Force Microscopy
PPTX
PPTX
Introduction and applications of Atomic force microscopy
PDF
AFM.pdf
PPT
cantilever array sensors
PPT
Atomic Force Microscope and its potential use in biology
PPTX
Atomic Foce Microscopy by Syed Ahsan Bukhari.pptx
PPT
ATOMIC FORCE MICROSCOPE MITHILESH CHOUDHARY
PPT
Afm modified
PDF
afm by anam rana
Non contact mode (AFM)
Scanning Tunneling Microscope and Atomic Tunneling Microscope
Atomic Force Microscope: Fundamental Principles
Charecterization techniques
Atomic force microscopy
Atomic force microscopy
Atomic Force Microscopy (AFM)
Atomic force microscope
Microscopic examination using Atomic force microscopy and Confocal scanning ...
Atomic Force Microscopy
Introduction and applications of Atomic force microscopy
AFM.pdf
cantilever array sensors
Atomic Force Microscope and its potential use in biology
Atomic Foce Microscopy by Syed Ahsan Bukhari.pptx
ATOMIC FORCE MICROSCOPE MITHILESH CHOUDHARY
Afm modified
afm by anam rana
Ad

Recently uploaded (20)

PDF
OBE - B.A.(HON'S) IN INTERIOR ARCHITECTURE -Ar.MOHIUDDIN.pdf
PDF
O7-L3 Supply Chain Operations - ICLT Program
PPTX
Microbial diseases, their pathogenesis and prophylaxis
PDF
STATICS OF THE RIGID BODIES Hibbelers.pdf
PDF
01-Introduction-to-Information-Management.pdf
PDF
RMMM.pdf make it easy to upload and study
PDF
Microbial disease of the cardiovascular and lymphatic systems
PPTX
Introduction-to-Literarature-and-Literary-Studies-week-Prelim-coverage.pptx
DOC
Soft-furnishing-By-Architect-A.F.M.Mohiuddin-Akhand.doc
PPTX
Tissue processing ( HISTOPATHOLOGICAL TECHNIQUE
PDF
Module 4: Burden of Disease Tutorial Slides S2 2025
PDF
Supply Chain Operations Speaking Notes -ICLT Program
PDF
FourierSeries-QuestionsWithAnswers(Part-A).pdf
PPTX
Cell Types and Its function , kingdom of life
PDF
RTP_AR_KS1_Tutor's Guide_English [FOR REPRODUCTION].pdf
PDF
Trump Administration's workforce development strategy
PDF
O5-L3 Freight Transport Ops (International) V1.pdf
PPTX
Final Presentation General Medicine 03-08-2024.pptx
PDF
Complications of Minimal Access Surgery at WLH
PPTX
school management -TNTEU- B.Ed., Semester II Unit 1.pptx
OBE - B.A.(HON'S) IN INTERIOR ARCHITECTURE -Ar.MOHIUDDIN.pdf
O7-L3 Supply Chain Operations - ICLT Program
Microbial diseases, their pathogenesis and prophylaxis
STATICS OF THE RIGID BODIES Hibbelers.pdf
01-Introduction-to-Information-Management.pdf
RMMM.pdf make it easy to upload and study
Microbial disease of the cardiovascular and lymphatic systems
Introduction-to-Literarature-and-Literary-Studies-week-Prelim-coverage.pptx
Soft-furnishing-By-Architect-A.F.M.Mohiuddin-Akhand.doc
Tissue processing ( HISTOPATHOLOGICAL TECHNIQUE
Module 4: Burden of Disease Tutorial Slides S2 2025
Supply Chain Operations Speaking Notes -ICLT Program
FourierSeries-QuestionsWithAnswers(Part-A).pdf
Cell Types and Its function , kingdom of life
RTP_AR_KS1_Tutor's Guide_English [FOR REPRODUCTION].pdf
Trump Administration's workforce development strategy
O5-L3 Freight Transport Ops (International) V1.pdf
Final Presentation General Medicine 03-08-2024.pptx
Complications of Minimal Access Surgery at WLH
school management -TNTEU- B.Ed., Semester II Unit 1.pptx
Ad

Atomic force microscopy presentation. Ppt

  • 2. Atomic Force Microscope (AFM) Introduction The atomic force microscope (AFM) was invented in1986 by Binnig, Quate and Gerber. The AFM raster scans a sharp probe over the surface of a sample and measures the changes in force between the probe tip and the sample. 2
  • 3. Working Concept The physical parameter probed is a force resulting from different interactions. Thus, an AFM image is generated by recording the force changes as the probe (or sample) is scanned in the x and y directions. The sample is mounted on a piezoelectric scanner, which ensures three-dimensional positioning with high resolution. The force is monitored by attaching the probe to a pliable cantilever, which acts as a spring, and measuring the bending or "deflection" of the cantilever. 3
  • 4. Principle • The AFM consists of a cantilever with a sharp tip (probe) at its end that is used to scan the specimen surface. • The cantilever is typically silicon or silicon nitride with a tip radius of curvature on the order of nanometers. • When the tip is brought into proximity of a sample surface, forces between the tip and the sample lead to a deflection of the cantilever according to Hooke's law. • Depending on the situation, forces that are measured in AFM include mechanical contact force, van der Waals forces, capillary forces, chemical bonding, electrostatic forces. 4
  • 6. Basic set-up of an AFM The ability of an AFM to achieve near atomic scale resolution depends on the three essential components: (1) a cantilever with a sharp tip, (2) a scanner that controls the x-y-z position, and (3) the feedback control and loop. 6
  • 7. • Cantiliever with a sharp tip. The stiffness of the cantilever needs to be less the effective spring constant holding atoms together, which is on the order of 1 - 10 nm. • The tip should have a radius of curvature less than 20-50 nm (smaller is better) a cone angle between 10-20 degrees. • 2. Scanner. The movement of the tip in the sample in the x, y, and z-directions is controlled by a piezo-electric tube scanner. • For typical AFM scanners, the maximum ranges for are 80 mm x 80 mm in the x-y plane and 5 mm for the z-direction. 7
  • 8. 3. Feedback control. The forces that are exerted between the tip and the sample are measured by the amount of bending (or deflection) of the cantilever. By calculating the difference signal in the photodiode quadrants, the amount of deflection can be correlated with a height . Because the cantilever obeys Hooke's Law for small displacements, the interaction force between the tip and the sample can be determined. 8
  • 9. Imaging Modes • Contact Mode • Tapping Mode • Non contact Mode 9
  • 10. Contact mode • The tip is "dragged" across the surface of the sample and the contours of the surface are measured either using the deflection of the cantilever directly or, more commonly, using the feedback signal required to keep the cantilever at a constant position. • Because the measurement of a static signal is prone to noise and drift, low stiffness cantilevers are used to achieve a large enough deflection signal while keeping the interaction force low. • Thus, contact mode AFM is almost always done at a depth where the overall force is repulsive, that is, in firm "contact" with the solid surface. 10
  • 11. Tapping mode • The cantilever is driven to oscillate up and down at or near its resonance frequency. • The amplitude of this oscillation usually varies from several nm to 200 nm. • The interaction of forces acting on the cantilever when the tip comes close to the surface, Van der Waals forces, dipole-dipole interactions, electrostatic forces, etc. • This causes the amplitude of the cantilever's oscillation to change (usually decrease) as the tip gets closer to the sample. • This amplitude is used as the parameter that goes into the electronic servo that controls the height of the cantilever above the sample. • A tapping AFM image is therefore produced by imaging the force of the intermittent contacts of the tip with the sample surface. 11
  • 12. Non contact method • In non-contact mode, the tip of the cantilever does not contact the sample surface. • The cantilever is instead oscillated at either its resonant frequency or just above (amplitude modulation) where the amplitude of oscillation is typically a few nanometers (<10 nm) down to a few picometers. • The van der Waals forces, which are strongest from 1 nm to 10 nm above the surface acts to decrease the resonance frequency of the cantilever. • Measuring the tip-to-sample distance at each (x,y) data point allows the scanning software to construct a topographic image of the sample surface. • Non-contact mode AFM does not suffer from tip or sample degradation effects that are sometimes observed after taking numerous scans with contact AFM. This makes non-contact AFM preferable to contact AFM for measuring soft samples, e.g. biological samples and organic thin film. 12
  • 13. Applications The AFM is useful for obtaining three-dimensional topographic information of insulating and conducting structures with lateral resolution down to 1.5 nm and vertical resolution down to 0.05 nm. These samples include clusters of atoms and molecules, individual macromolecules, and biologic al species (cells, DNA, proteins). Minimal sample preparation involved for AFM imaging. The AFM can operate in gas, ambient, and fluid environments and can measure physical properties including elasticity, adhesion, hardness, friction and chemical functionality. 13
  • 14. Advantages The AFM has several advantages over the scanning electron microscope (SEM). Unlike the electron microscope which provides a two- dimensional projection or a two-dimensional image of a sample, the AFM provides a true three-dimensional surface profile. Additionally, samples viewed by AFM do not require any special treatments (such as metal/carbon coatings) that would irreversibly change or damage the sample. 14
  • 15. While an electron microscope needs an expensive vacuum environment for proper operation, most AFM modes can work perfectly well in ambient air or even a liquid environment. This makes it possible to study biological macromolecules and even living organisms. In principle, AFM can provide higher resolution than SEM. 15
  • 16. Disadvantages A disadvantage of AFM compared with the scanning electron microscope (SEM) is the image size. The SEM can image an area on the order of millimetres by millimetres with a depth of field on the order of millimetres. The AFM can only image a maximum height on the order of micrometres and a maximum scanning area of around 150 by 150 micrometres. Another inconvenience is that at high resolution, the quality of an image is limited by the radius of curvature of the probe tip, and an incorrect choice of tip for the required resolution can lead to image artifacts. 16