Introduction
Alternatives to animal testing and the development of methods that help
reducing the number of animals in preclinical studies are much sought
after. In capillary micro sampling, a low volume of blood (for example 30
µL) is collected in a capillary, centrifuged and an exact volume of plasma
(for example 10 µL) transferred into another capillary. For analysis, the
plasma is washed or blown out into a sample tube and diluted to a
volume, that can be handled reliably and reproducibly. In contrast to
DBS techniques, plasma is processed and analyzed, thus avoiding
investigation of critical parameters like effect of hematrocit and offering
standard procedures for addition of stabilizers or internal standards
during analysis. Even though necessary, dilution should be minimized as
it directly affects the achievable limits of quantification. In this poster we
describe an automated SPE procedure using small, single use
cartridges in a modified CTC liquid handling auto sampler. Results from
this approach are compared to results obtained from identical samples
processed manually by protein precipitation.
Evaluation of an automated SPE method for samples
derived from capillary microsampling
Paul-Gerhard Lassahn1, Petra Lombardi1, Winfried Wagner-Redeker1, Marie-Claude
Roubaudi-Fraschini2, Reto Bolliger3, Guenter Boehm3
1 Swiss BioAnalytics AG, Basel, Switzerland
2 Debiopharm International S.A., Lausanne, Switzerland
3 CTC Analytics AG, Zwingen, Switzerland
Experimental
General: Samples were processed by dilution of spiked plasma with water containing the internal standard
followed by online sample-preparation described below.
Working-solutions and sample preparation: The analyte (MW 581) was dissolved in DMSO and diluted in
ACN/DMSO (50:50, v/v) to obtain working-solutions at concentrations twenty times higher than those of the
final CAL and QC samples. Aliquots of the working-solutions were spiked into blank rat plasma and diluted with
water containing the internal standard in a ratio of 1:10 (v/v). The samples were vortexed, centrifuged and
transferred to the ITSP sample-preparation plate.
Online-sample-preparation: ITSP (ITSP Solutions Inc.) cartridges were conditioned with 80µL MeOH and 80µL
0.2% aqueous formic acid before loading 10µL sample. After drying the cartridge with 80µL air, the ITSP
cartridges were washed twice with 80µl 0.2% aqueous formic acid and dried again. The compounds were
eluted using 80µL EtOH and 10µL of the eluate was injected onto the analytical column. Sample-preparation of
each plasma sample was performed directly before the injection (see Figure 1 for the workflow and the results
of the optimization process).
Chromatography: 10 µL of the extracted samples were injected onto a Thermo Hypersil Gold (2.1 x 100 mm, 3
µm) analytical column. Eluent A consisted of 0.5% aqueous formic acid, eluent B of 0.5% formic acid in ACN
and LC flow rate was 0.5 mL/min. The eluent composition was changed linearly from 70% to 5% A in 2.25
minutes. Eluent composition was kept at 5% B for one minute and then went back ballistically to starting
conditions.
Calibration range: 50.0 - 25’000 ng/mL (high-range, HR) and 5.00 - 2’500 ng/mL (low-range, LR).
Instrumentation and MS-method: TSQ Quantum Ultra (Thermo Scientific), Rheos 2200 quarternary pump (Flux
Instruments), HTX PAL autosampler (CTC Analytics).
Data-Processing: Raw data were acquired with Xcalibur 2.0 software and processed with LCQuan 2.5 (Thermo
Scientific).
Validation parameters: Intra- and inter-assay accuracy and precision, linearity, and carry-over.
Table 1. MS-parameter (SRM-assay).
www.swissbioanalytics.com
Results
Blank rat plasma was spiked with the analyte and diluted with water
containing the internal standard in a ratio of 1:10 (v/v). CALs and QCs
were prepared to cover concentrations of 50.0 - 25’000 ng/mL in the
high-calibration range and 5.00 - 2’500 ng/mL in low calibration range.
The diluted plasma samples were transferred to the ITSP sample-
preparation plate and automatically extracted directly before analysis.
Sample preparation and LC-MS/MS measurements were performed in
parallel using the “Prep Ahead Function” of the CTC autosampler
(Figure 4). Chromatograms of the lowest calibration standard Cal1 are
presented in Figure 2 and compared with the chromatograms of the
same calibration standard prepared by protein-precipitation.
Three independent precision/accuracy runs were conducted for both
calibration ranges and the intra- and inter-run variabilities are
presented in Table 2. An example of a typical calibration curve for
each range is presented in Figure 3. Both assays gave very similar
results and the intra- and inter-run performance is equivalent to the
data obtained using a standard protein-precipitation method.
The carry-over in the analytical assay does not meet the acceptance
criteria of the EMA guideline but can be significantly reduced by using
Analyte Parent ion mass
[Da]
Product ion mass
[Da]
Collision energy
[V]
Analyte 582.3 255.1 35
Internal Standard 588.3 487.2 30
ESI positive, spray voltage 3.5 kV, capillary temperature 350 °C, collision gas pressure
(Ar) 1.5 mTorr, scan-time 100 msec
Figure 6. Linearity of the loading-volume and capacity of cartridges.
Loading
Eluting
Area Analyte Area IS Area RatioArea Analyte Area IS Area Ratio
5µL 10µL Injection volume 40µL
Conclusions
•The automated workflow described above gave equivalent
results (LLOQ, precision, accuracy) compared to samples
processed manually by protein precipitation. The overall
time/sample is slightly higher for the automated protocol.
However, since only minimal intervention is required, process
safety is superior and human resources are freed up for
other tasks.
•The automated protocol may be further optimized in several
aspects. As shown in Figure 7, a volume of 20 µL is sufficient
for complete elution of the analyte from the cartridge. Figure
6 indicates that an loading volume of 2-5 µL diluted plasma is
sufficient allowing the use of low-volume samples. Both
aspects are important for further optimizing this protocol for
analysis of samples originating from studies using
microsampling techniques.
•Methods for the automated optimization of the SPE
conditions exist and will be applied in future studies.
Reference: Nilsson LB, Ahnoff M, Jonsson O; Bioanalysis
(2013) 5(6), 731–738
Low-Range
Y = 0.000781592+0.00128746*X+1.0801e-008*X^2
R^2 = 0.9990 W: 1/X
0 1000 2000
ng/mL
1
2
3
AreaRatio
High-Range
Y = 0.00191642+0.000195293*X-1.32222e-009*X^2
R^2 = 0.9993 W: 1/X
0 10000 20000
ng/mL
1
2
3
4
Figure 3. Calibration-curves (low-range assay left, high-range assay right).
Figure 1. Workflow and results of the optimization process.
0.E+00
5.E+06
1.E+07
2.E+07
2.E+07 Wash 1
Wash 2
Elution 1
Elution 2
0.E+00
1.E+07
2.E+07
4.E+07
5.E+07
6.E+07
7.E+07
8.E+07
1.E+08
1.E+08
1.E+08
Elution 1
Elution 2
Elution pre-testing:
(a) 10mM NH4Fo pH9
(b) 10mM NH4Fo pH9/MeOH (50:50)
(c) 10mM NH4Fo pH9/ACN (50:50)
(d) ACN
(e) MeOH
(f) ACN/MeOH (50:50)
(g) ACN/MeOH (75:25)
(h) ACN/MeOH (75:25)
(i) EtOH
selected
0%
20%
40%
60%
80%
100%
0.1% HFo 0.2% HFo 0.5% HFo
*
* *
*
Elution 1
Elution 2
Wash 1
Wash 2
Elution 2
Elution 1
0.1% HFo 0.2% HFo 0.5% HFo
(e) (f) (i)
*
(a) (b) (c) (d) (e) (f) (g) (h) (i)
Washing variable Washing variable
Figure 2. Comparison of protein-precipitation (PP) and ITSP at LLOQ.
J:Projekte...Cal1_LR_Precipitation_02 8/23/2013 1:40:16 PM 9 3 13020 CAL1 1 1
RT: 0.0 - 4.0
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5
Time (min)
0
10
20
30
40
50
60
70
80
90
100
RelativeAbundance
0
10
20
30
40
50
60
70
80
90
100
RelativeAbundance
RT: 2.6
RT: 2.7
NL: 8.83E3
TIC F: + c ESI SRM
ms2
582.300@cid35.00
[255.050-255.150]
MS
Cal1_LR_Precipitatio
n_02
NL: 8.32E5
TIC F: + c ESI SRM
ms2
588.300@cid30.00
[487.150-487.250]
MS
Cal1_LR_Precipitatio
n_02
RT: 0.0 - 4.0
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0
Time (min)
0
10
20
30
40
50
60
70
80
90
100
RelativeAbundance
0
10
20
30
40
50
60
70
80
90
100
RelativeAbundance
RT: 2.5
RT: 2.5
NL: 2.57E4
TIC F: + c ESI
SRM ms2
582.300@cid35.0
0
[255.050-
255.150] MS
04112013_05
NL: 4.25E6
TIC F: + c ESI
SRM ms2
588.300@cid30.0
0
[487.150-
487.250] MS
04112013_05
J:Projekte...Cal1_HR_Precipitation 7/30/2013 12:10:33 PM 2 3 13020 CAL1 1 1
RT: 0.0 - 4.0 SM: 7G
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0
Time (min)
0
10
20
30
40
50
60
70
80
90
100
RelativeAbundance
0
10
20
30
40
50
60
70
80
90
100
RelativeAbundance
RT: 2.5
RT: 2.6
NL: 1.82E5
TIC F: + c ESI SRM
ms2
582.300@cid35.00
[255.050-255.150]
MS
Cal1_HR_Precipitati
on
NL: 1.20E7
TIC F: + c ESI SRM
ms2
588.300@cid30.00
[487.150-487.250]
MS
Cal1_HR_Precipitati
on
RT: 0.0 - 4.0 SM: 7G
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5
Time (min)
0
10
20
30
40
50
60
70
80
90
100
RelativeAbundance
0
10
20
30
40
50
60
70
80
90
100
RelativeAbundance
RT: 2.4
RT: 2.5
NL: 7.44E4
TIC F: + c ESI
SRM ms2
582.300@cid35.0
0
[255.050-
255.150] MS
24102013_5
NL: 8.40E6
TIC F: + c ESI
SRM ms2
588.300@cid30.0
0
[487.150-
487.250] MS
24102013_5
Low-Range
Analyte
Internal
Standard
PP ITSP
High-Range
PP ITSP
0.0 100.0 200.0 300.0 400.0 500.0 600.0
Zentrifugieren
Verdünnen
Transfer ITSP
Fällung IS
MS
0.0 200.0 400.0 600.0 800.0 1000.0 1200.0 1400.0
PP
PP
ITSP
PP
ITSP
Gradient
5min
7min
10min
Run-Time [min]
ITSP
# of Samples: 100
centrifuge
dilute
Transfer ITSP-Plate
Protein precipitation
MS-measurement
#ofSamples
Run-Time [min]
100
200
400
Sample-Preparation:
Figure 5. Comparison of protein-precipitation (PP) and ITSP (#
Samples: variable; gradient-time: variable; ITSP: sample-preparation
7min in parallel; PP: pre-injection-time 1.0min).
PP
ITSP
PP
ITSP
PP
ITSP
at LLOQ (LR) at LLOQ (HR) > LLOQ (LR) > LLOQ (HR)
PrecIntra 9.7-10.5% 5.0-6.0% 1.5-10.5% 1.4-8.3%
AccIntra 93.2-107.4% 95.7-113.0% 101.0-105.2% 95.5-113.5%
PrecInter 11.1% 8.9% 4.4-9.5% 2.5-8.1%
AccInter 100.6% 105.1% 101.0-105.2% 104.8-108.6%
Table 2. Intra- and inter-run precision and accuracy (3 runs, [%]).
Calibration-Range Carry-over*
[% of CAL8]
Carry-over*
[% of Cal1]
Low-Range [1st Blank after Cal8] 0.6 ** 229.7 **
Low-Range [2nd Blank after Cal8] 0.2 ** 80.3 **
High-Range [1st Blank after Cal8] 0.7 278.6
High-Range [2nd Blank after Cal8] 0.3 108.5
* Standard-Wash-solution; Wash 1: H2O/EtOH (50:50, v/v); Wash 2
ACN/MeOH/i-Prop (1:1:1, v/v/v)
** Carry-Over can be significantly reduced using 0.3% Tween in
Wash-solution 1 (below 30% of Cal1 for the 2nd Blank after Cal8)
Table 3. Carry-over.
Figure 4. Prep Ahead Function of the CTC autosampler.
Area Analyte
20µL10µL 80µL
Area Analyte
40µL
Figure 7. Volume of the elution solvent EtOH (area of analyte
normalized to concentration).
0.3% Tween as an additive in the aqueous wash-solution
(see Table 3).
The influence of the amount of (diluted) plasma loaded onto
the cartridge and the volume of the extraction solvent is
presented in Figures 6 and 7. Elution from the ITSP
cartridge is complete with as little as 20 µL of extraction
solvent (Figure 7) resulting in a factor 4 higher
concentration of the analyte in the eluate compared to the
standard-elution method. The capacity of the cartridges has
been tested between 2 to 40 µL diluted plasma (Figure 6).
The area ratio of analyte to internal standard is constant
across this range and demonstrate excellent performance
of the method across a wide range of sample volumes.
The sample-preparation time and the overall run-time of
the automated and manual method are compared in Figure
5. Since only minimal manual intervention is required for the
automated approach, this workflow offers significant
advantages by high sample-workload and longer LC-
MS/MS run-times.

More Related Content

PDF
Immunosuppressants in Blood at UWMC Poster-MicroLiter
PDF
Immunosuppressants in Blood at OpAns Poster-MicroLiter
PPT
Blood amps poster_090210
PDF
Automated SPE-LC/MS/MS method development using ITSP
PDF
dry blood spotting technique,DBS
PDF
Symposium presentation: Development of a Dried Blood Spot Method for Leptin
PDF
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
PDF
Automated hplc screening of newborns for sickle cell anemia
Immunosuppressants in Blood at UWMC Poster-MicroLiter
Immunosuppressants in Blood at OpAns Poster-MicroLiter
Blood amps poster_090210
Automated SPE-LC/MS/MS method development using ITSP
dry blood spotting technique,DBS
Symposium presentation: Development of a Dried Blood Spot Method for Leptin
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Automated hplc screening of newborns for sickle cell anemia

What's hot (19)

PDF
K based-chromatography: future and potential of Countercurrent Chromatography
PDF
Optimization parameters in Countercurrent Chromatography
PDF
Analysis of the Monoclonal Antibody Adalimumab in Human Blood Collected via V...
PDF
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
PDF
Comparison of CCS Instruments
PDF
Quantification of a Novel Peptide, CPT31 in Rat and Monkey Plasma by LC-MS
PDF
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
PPTX
PattersonEWRI2014 (1)
PPTX
Automated Online Extraction and Chromatography with Supercritical Fluids
PDF
Analysis of “The Big Four” Heavy Metals in Hops by Graphite Furnace Atomic Ab...
PDF
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
PPTX
"Productivity and Simplicity" - Streamlining Cumbersome Sample Preparation i...
PDF
VALIDATED LIQUID CHROMATOGRAPHY/TANDEM MASS SPECTROMETRY METHOD FOR DETERMINA...
PDF
solvent system formulation
PDF
Solvent System Selection
PPT
Colitag 16 - 48 Hour Presentation
DOCX
Final Project Report
PDF
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
PDF
Elution methods in Countercurrent Chromatography
K based-chromatography: future and potential of Countercurrent Chromatography
Optimization parameters in Countercurrent Chromatography
Analysis of the Monoclonal Antibody Adalimumab in Human Blood Collected via V...
Are You Sticking with Drinking Bottled Water? Assessment of PFAS Content in C...
Comparison of CCS Instruments
Quantification of a Novel Peptide, CPT31 in Rat and Monkey Plasma by LC-MS
Determination of Ethanol and Isopropanol Content in Hand Sanitizers Using Nit...
PattersonEWRI2014 (1)
Automated Online Extraction and Chromatography with Supercritical Fluids
Analysis of “The Big Four” Heavy Metals in Hops by Graphite Furnace Atomic Ab...
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
"Productivity and Simplicity" - Streamlining Cumbersome Sample Preparation i...
VALIDATED LIQUID CHROMATOGRAPHY/TANDEM MASS SPECTROMETRY METHOD FOR DETERMINA...
solvent system formulation
Solvent System Selection
Colitag 16 - 48 Hour Presentation
Final Project Report
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
Elution methods in Countercurrent Chromatography
Ad

Viewers also liked (12)

PDF
DOCX
Bogotá 06 julio del 2012
DOCX
Conclusión del proyecto semestral
DOCX
DOCX
Bogotá 06 julio del 2012
DOCX
1 to 9 point to point express
TXT
제주렌트카할인 필리핀골프gsa
DOCX
1 to 9 point to point express
DOC
Disquera
PDF
Zaintza diagnostikoaren aurkezpena
PDF
Industeel duplex
Bogotá 06 julio del 2012
Conclusión del proyecto semestral
Bogotá 06 julio del 2012
1 to 9 point to point express
제주렌트카할인 필리핀골프gsa
1 to 9 point to point express
Disquera
Zaintza diagnostikoaren aurkezpena
Industeel duplex
Ad

Similar to Automated SPE for Capillary Microsampling Poster (20)

PDF
Bio outsource hcp
PDF
Automated sample preparation using the GERSTEL MPS Series and MAESTRO softwar...
PDF
Analytical methods for therapeutic antibody characterization, comparability, ...
PDF
Western Blotting Protocol & Troubleshooting
PPTX
Western blotting protocol & troubleshooting
PDF
Pba2003poster(Bock Etal)
PDF
From Screening to QC: Development Considerations for Octet Methods
PPT
Oligonucleotide Capabilities at Pine Lake Laboratories
PPT
Lab talk 070510 inducing r_rel1 for establishing fret assay_ic50 studies with...
PPTX
extraction of drug from biological matrix.pptx
PDF
High Throughput Purification and Characterization of a Protein Variant Library
PPTX
What is Nephelometry,and a fully automated Nephlometer analyzer for protein a...
PDF
Biolyzer 300 technical specifications.pdf
PPSX
Nova Analytica
PPTX
AAPS 2015_M1028_U-PLEX Feasibility Poster_Russell
PDF
High-throughput and Automated Process Development for Accelerated Biotherapeu...
PDF
Immunosuppressants in Blood Article - MicroLiter
PPTX
Sample preparation for Chromatography
PDF
LC-HRMS Method FDA NITROSAMINES REVIEWS A
PDF
Two-Tailed PCR - New Ultrasensitive and Ultraspecific Technique for the Quant...
Bio outsource hcp
Automated sample preparation using the GERSTEL MPS Series and MAESTRO softwar...
Analytical methods for therapeutic antibody characterization, comparability, ...
Western Blotting Protocol & Troubleshooting
Western blotting protocol & troubleshooting
Pba2003poster(Bock Etal)
From Screening to QC: Development Considerations for Octet Methods
Oligonucleotide Capabilities at Pine Lake Laboratories
Lab talk 070510 inducing r_rel1 for establishing fret assay_ic50 studies with...
extraction of drug from biological matrix.pptx
High Throughput Purification and Characterization of a Protein Variant Library
What is Nephelometry,and a fully automated Nephlometer analyzer for protein a...
Biolyzer 300 technical specifications.pdf
Nova Analytica
AAPS 2015_M1028_U-PLEX Feasibility Poster_Russell
High-throughput and Automated Process Development for Accelerated Biotherapeu...
Immunosuppressants in Blood Article - MicroLiter
Sample preparation for Chromatography
LC-HRMS Method FDA NITROSAMINES REVIEWS A
Two-Tailed PCR - New Ultrasensitive and Ultraspecific Technique for the Quant...

More from Rick Youngblood (20)

PDF
Automated Chromatographic Solid-Phase Extraction Using an Autosampler _ Ameri...
PPSX
ITSP PAL Training - Intermediate
PPSX
ITSP PAL Training - Advanced
PPSX
ITSP PAL Training - Basic
PPSX
ITSP Dual Rail Project
PPSX
ITSP New Stuff
PPSX
Optimization of Automated Online SPE-LC-MS
PPSX
Unified-Drug-Analysis-PittCon-2016
PPSX
ITSP Setup on PAL-xt
PPSX
ITSP Introduction
PDF
Barbiturates and THCA in Urine at OpAns
PDF
DUI-D in Urine Poster
PDF
Androstenedione and Testosterone in Serum
PDF
Drugs of Abuse in Post Mortem Blood Poster
PDF
Manual Single Cartridge Holder Flyer
PDF
Streptomycin in Honey Poster
PDF
PDF
QuEChERS Pittcon 2014 Poster
PDF
Drugs_of_Abuse_Application_Note
PDF
Go Green Brochure-MicroLiter
Automated Chromatographic Solid-Phase Extraction Using an Autosampler _ Ameri...
ITSP PAL Training - Intermediate
ITSP PAL Training - Advanced
ITSP PAL Training - Basic
ITSP Dual Rail Project
ITSP New Stuff
Optimization of Automated Online SPE-LC-MS
Unified-Drug-Analysis-PittCon-2016
ITSP Setup on PAL-xt
ITSP Introduction
Barbiturates and THCA in Urine at OpAns
DUI-D in Urine Poster
Androstenedione and Testosterone in Serum
Drugs of Abuse in Post Mortem Blood Poster
Manual Single Cartridge Holder Flyer
Streptomycin in Honey Poster
QuEChERS Pittcon 2014 Poster
Drugs_of_Abuse_Application_Note
Go Green Brochure-MicroLiter

Automated SPE for Capillary Microsampling Poster

  • 1. Introduction Alternatives to animal testing and the development of methods that help reducing the number of animals in preclinical studies are much sought after. In capillary micro sampling, a low volume of blood (for example 30 µL) is collected in a capillary, centrifuged and an exact volume of plasma (for example 10 µL) transferred into another capillary. For analysis, the plasma is washed or blown out into a sample tube and diluted to a volume, that can be handled reliably and reproducibly. In contrast to DBS techniques, plasma is processed and analyzed, thus avoiding investigation of critical parameters like effect of hematrocit and offering standard procedures for addition of stabilizers or internal standards during analysis. Even though necessary, dilution should be minimized as it directly affects the achievable limits of quantification. In this poster we describe an automated SPE procedure using small, single use cartridges in a modified CTC liquid handling auto sampler. Results from this approach are compared to results obtained from identical samples processed manually by protein precipitation. Evaluation of an automated SPE method for samples derived from capillary microsampling Paul-Gerhard Lassahn1, Petra Lombardi1, Winfried Wagner-Redeker1, Marie-Claude Roubaudi-Fraschini2, Reto Bolliger3, Guenter Boehm3 1 Swiss BioAnalytics AG, Basel, Switzerland 2 Debiopharm International S.A., Lausanne, Switzerland 3 CTC Analytics AG, Zwingen, Switzerland Experimental General: Samples were processed by dilution of spiked plasma with water containing the internal standard followed by online sample-preparation described below. Working-solutions and sample preparation: The analyte (MW 581) was dissolved in DMSO and diluted in ACN/DMSO (50:50, v/v) to obtain working-solutions at concentrations twenty times higher than those of the final CAL and QC samples. Aliquots of the working-solutions were spiked into blank rat plasma and diluted with water containing the internal standard in a ratio of 1:10 (v/v). The samples were vortexed, centrifuged and transferred to the ITSP sample-preparation plate. Online-sample-preparation: ITSP (ITSP Solutions Inc.) cartridges were conditioned with 80µL MeOH and 80µL 0.2% aqueous formic acid before loading 10µL sample. After drying the cartridge with 80µL air, the ITSP cartridges were washed twice with 80µl 0.2% aqueous formic acid and dried again. The compounds were eluted using 80µL EtOH and 10µL of the eluate was injected onto the analytical column. Sample-preparation of each plasma sample was performed directly before the injection (see Figure 1 for the workflow and the results of the optimization process). Chromatography: 10 µL of the extracted samples were injected onto a Thermo Hypersil Gold (2.1 x 100 mm, 3 µm) analytical column. Eluent A consisted of 0.5% aqueous formic acid, eluent B of 0.5% formic acid in ACN and LC flow rate was 0.5 mL/min. The eluent composition was changed linearly from 70% to 5% A in 2.25 minutes. Eluent composition was kept at 5% B for one minute and then went back ballistically to starting conditions. Calibration range: 50.0 - 25’000 ng/mL (high-range, HR) and 5.00 - 2’500 ng/mL (low-range, LR). Instrumentation and MS-method: TSQ Quantum Ultra (Thermo Scientific), Rheos 2200 quarternary pump (Flux Instruments), HTX PAL autosampler (CTC Analytics). Data-Processing: Raw data were acquired with Xcalibur 2.0 software and processed with LCQuan 2.5 (Thermo Scientific). Validation parameters: Intra- and inter-assay accuracy and precision, linearity, and carry-over. Table 1. MS-parameter (SRM-assay). www.swissbioanalytics.com Results Blank rat plasma was spiked with the analyte and diluted with water containing the internal standard in a ratio of 1:10 (v/v). CALs and QCs were prepared to cover concentrations of 50.0 - 25’000 ng/mL in the high-calibration range and 5.00 - 2’500 ng/mL in low calibration range. The diluted plasma samples were transferred to the ITSP sample- preparation plate and automatically extracted directly before analysis. Sample preparation and LC-MS/MS measurements were performed in parallel using the “Prep Ahead Function” of the CTC autosampler (Figure 4). Chromatograms of the lowest calibration standard Cal1 are presented in Figure 2 and compared with the chromatograms of the same calibration standard prepared by protein-precipitation. Three independent precision/accuracy runs were conducted for both calibration ranges and the intra- and inter-run variabilities are presented in Table 2. An example of a typical calibration curve for each range is presented in Figure 3. Both assays gave very similar results and the intra- and inter-run performance is equivalent to the data obtained using a standard protein-precipitation method. The carry-over in the analytical assay does not meet the acceptance criteria of the EMA guideline but can be significantly reduced by using Analyte Parent ion mass [Da] Product ion mass [Da] Collision energy [V] Analyte 582.3 255.1 35 Internal Standard 588.3 487.2 30 ESI positive, spray voltage 3.5 kV, capillary temperature 350 °C, collision gas pressure (Ar) 1.5 mTorr, scan-time 100 msec Figure 6. Linearity of the loading-volume and capacity of cartridges. Loading Eluting Area Analyte Area IS Area RatioArea Analyte Area IS Area Ratio 5µL 10µL Injection volume 40µL Conclusions •The automated workflow described above gave equivalent results (LLOQ, precision, accuracy) compared to samples processed manually by protein precipitation. The overall time/sample is slightly higher for the automated protocol. However, since only minimal intervention is required, process safety is superior and human resources are freed up for other tasks. •The automated protocol may be further optimized in several aspects. As shown in Figure 7, a volume of 20 µL is sufficient for complete elution of the analyte from the cartridge. Figure 6 indicates that an loading volume of 2-5 µL diluted plasma is sufficient allowing the use of low-volume samples. Both aspects are important for further optimizing this protocol for analysis of samples originating from studies using microsampling techniques. •Methods for the automated optimization of the SPE conditions exist and will be applied in future studies. Reference: Nilsson LB, Ahnoff M, Jonsson O; Bioanalysis (2013) 5(6), 731–738 Low-Range Y = 0.000781592+0.00128746*X+1.0801e-008*X^2 R^2 = 0.9990 W: 1/X 0 1000 2000 ng/mL 1 2 3 AreaRatio High-Range Y = 0.00191642+0.000195293*X-1.32222e-009*X^2 R^2 = 0.9993 W: 1/X 0 10000 20000 ng/mL 1 2 3 4 Figure 3. Calibration-curves (low-range assay left, high-range assay right). Figure 1. Workflow and results of the optimization process. 0.E+00 5.E+06 1.E+07 2.E+07 2.E+07 Wash 1 Wash 2 Elution 1 Elution 2 0.E+00 1.E+07 2.E+07 4.E+07 5.E+07 6.E+07 7.E+07 8.E+07 1.E+08 1.E+08 1.E+08 Elution 1 Elution 2 Elution pre-testing: (a) 10mM NH4Fo pH9 (b) 10mM NH4Fo pH9/MeOH (50:50) (c) 10mM NH4Fo pH9/ACN (50:50) (d) ACN (e) MeOH (f) ACN/MeOH (50:50) (g) ACN/MeOH (75:25) (h) ACN/MeOH (75:25) (i) EtOH selected 0% 20% 40% 60% 80% 100% 0.1% HFo 0.2% HFo 0.5% HFo * * * * Elution 1 Elution 2 Wash 1 Wash 2 Elution 2 Elution 1 0.1% HFo 0.2% HFo 0.5% HFo (e) (f) (i) * (a) (b) (c) (d) (e) (f) (g) (h) (i) Washing variable Washing variable Figure 2. Comparison of protein-precipitation (PP) and ITSP at LLOQ. J:Projekte...Cal1_LR_Precipitation_02 8/23/2013 1:40:16 PM 9 3 13020 CAL1 1 1 RT: 0.0 - 4.0 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 Time (min) 0 10 20 30 40 50 60 70 80 90 100 RelativeAbundance 0 10 20 30 40 50 60 70 80 90 100 RelativeAbundance RT: 2.6 RT: 2.7 NL: 8.83E3 TIC F: + c ESI SRM ms2 582.300@cid35.00 [255.050-255.150] MS Cal1_LR_Precipitatio n_02 NL: 8.32E5 TIC F: + c ESI SRM ms2 588.300@cid30.00 [487.150-487.250] MS Cal1_LR_Precipitatio n_02 RT: 0.0 - 4.0 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 Time (min) 0 10 20 30 40 50 60 70 80 90 100 RelativeAbundance 0 10 20 30 40 50 60 70 80 90 100 RelativeAbundance RT: 2.5 RT: 2.5 NL: 2.57E4 TIC F: + c ESI SRM ms2 582.300@cid35.0 0 [255.050- 255.150] MS 04112013_05 NL: 4.25E6 TIC F: + c ESI SRM ms2 588.300@cid30.0 0 [487.150- 487.250] MS 04112013_05 J:Projekte...Cal1_HR_Precipitation 7/30/2013 12:10:33 PM 2 3 13020 CAL1 1 1 RT: 0.0 - 4.0 SM: 7G 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 Time (min) 0 10 20 30 40 50 60 70 80 90 100 RelativeAbundance 0 10 20 30 40 50 60 70 80 90 100 RelativeAbundance RT: 2.5 RT: 2.6 NL: 1.82E5 TIC F: + c ESI SRM ms2 582.300@cid35.00 [255.050-255.150] MS Cal1_HR_Precipitati on NL: 1.20E7 TIC F: + c ESI SRM ms2 588.300@cid30.00 [487.150-487.250] MS Cal1_HR_Precipitati on RT: 0.0 - 4.0 SM: 7G 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 Time (min) 0 10 20 30 40 50 60 70 80 90 100 RelativeAbundance 0 10 20 30 40 50 60 70 80 90 100 RelativeAbundance RT: 2.4 RT: 2.5 NL: 7.44E4 TIC F: + c ESI SRM ms2 582.300@cid35.0 0 [255.050- 255.150] MS 24102013_5 NL: 8.40E6 TIC F: + c ESI SRM ms2 588.300@cid30.0 0 [487.150- 487.250] MS 24102013_5 Low-Range Analyte Internal Standard PP ITSP High-Range PP ITSP 0.0 100.0 200.0 300.0 400.0 500.0 600.0 Zentrifugieren Verdünnen Transfer ITSP Fällung IS MS 0.0 200.0 400.0 600.0 800.0 1000.0 1200.0 1400.0 PP PP ITSP PP ITSP Gradient 5min 7min 10min Run-Time [min] ITSP # of Samples: 100 centrifuge dilute Transfer ITSP-Plate Protein precipitation MS-measurement #ofSamples Run-Time [min] 100 200 400 Sample-Preparation: Figure 5. Comparison of protein-precipitation (PP) and ITSP (# Samples: variable; gradient-time: variable; ITSP: sample-preparation 7min in parallel; PP: pre-injection-time 1.0min). PP ITSP PP ITSP PP ITSP at LLOQ (LR) at LLOQ (HR) > LLOQ (LR) > LLOQ (HR) PrecIntra 9.7-10.5% 5.0-6.0% 1.5-10.5% 1.4-8.3% AccIntra 93.2-107.4% 95.7-113.0% 101.0-105.2% 95.5-113.5% PrecInter 11.1% 8.9% 4.4-9.5% 2.5-8.1% AccInter 100.6% 105.1% 101.0-105.2% 104.8-108.6% Table 2. Intra- and inter-run precision and accuracy (3 runs, [%]). Calibration-Range Carry-over* [% of CAL8] Carry-over* [% of Cal1] Low-Range [1st Blank after Cal8] 0.6 ** 229.7 ** Low-Range [2nd Blank after Cal8] 0.2 ** 80.3 ** High-Range [1st Blank after Cal8] 0.7 278.6 High-Range [2nd Blank after Cal8] 0.3 108.5 * Standard-Wash-solution; Wash 1: H2O/EtOH (50:50, v/v); Wash 2 ACN/MeOH/i-Prop (1:1:1, v/v/v) ** Carry-Over can be significantly reduced using 0.3% Tween in Wash-solution 1 (below 30% of Cal1 for the 2nd Blank after Cal8) Table 3. Carry-over. Figure 4. Prep Ahead Function of the CTC autosampler. Area Analyte 20µL10µL 80µL Area Analyte 40µL Figure 7. Volume of the elution solvent EtOH (area of analyte normalized to concentration). 0.3% Tween as an additive in the aqueous wash-solution (see Table 3). The influence of the amount of (diluted) plasma loaded onto the cartridge and the volume of the extraction solvent is presented in Figures 6 and 7. Elution from the ITSP cartridge is complete with as little as 20 µL of extraction solvent (Figure 7) resulting in a factor 4 higher concentration of the analyte in the eluate compared to the standard-elution method. The capacity of the cartridges has been tested between 2 to 40 µL diluted plasma (Figure 6). The area ratio of analyte to internal standard is constant across this range and demonstrate excellent performance of the method across a wide range of sample volumes. The sample-preparation time and the overall run-time of the automated and manual method are compared in Figure 5. Since only minimal manual intervention is required for the automated approach, this workflow offers significant advantages by high sample-workload and longer LC- MS/MS run-times.