Axial Length Measurement
( Biometry )
Mohammad Reza ZARRIN (optometrist- MsC)
Tehran University of Medical Sciences
IOL Power Calculation
1. Keratometry
2. A-Scan Biometry
3. IOL Formula
A-Scan Biometry
Measurement of Axial Eye Length
by Ultrasound
Average Axial Length of
Normal Eye
23.06 mm
Majority 22.0 to 24.5 mm
Accuracy of AL measurement
using A-scan ultrasound is
+ 0.1 mm
Difference in AL measurement
Between both eyes
+ 0.3 mm
Biometry
Instrumentation
Biometry
Examination Procedure
1. History Taking
2. Patient Preparation
3. Biometry Technique
Biometry Technique
 Contact
- Applanation Method
- Hand-Held Method
 Immersion
Biometry
Biometry
Values are 0.14 to 0.36 mm
longer with immersion technique
than with contact method
Biometry
Biometry
Biometry
Biometry
Biometry
Biometry
Biometry
Potential Sources of Error
with Contact Method
1.Corneal Compression
2. Fluid Excess
3. Misalignment of Sound Beam
4. Inappropriate Eye type
Biometry
Error caused by
1 mm Corneal Compression
Average eye 2.5 D
Long eye 1.75 D
Short eye 3.75 D
Biometry
Potential Sources of Error
with Immersion Method
1. Air bubbles within fluid
2. Inappropriate eye type
Instrument Setting
1. Measurement Mode
2. Gates
3. Gain
4. Eye Type
Measurement Mode
 Automatic
 Semiautomatic
 Manual
Gates
Gates are electronic markers on the
screen that provide measurement of
distance between 2 or more anatomic
interfaces .
Biometry
Biometry
Gain Setting
Initially high gain setting should be
used to assess the overall appearance
of the echogram , then gain should
be reduced to a medium level to
improve resolution of spikes .
Biometry
Error can occur when the gain
is set too high or too low .
Very high gain short reading
Very low gain long reading
Eye Type
( Sound Velocity )
1. Phakic
2. Aphakic
3. Pseudophakic
Biometry
Biometry
Biometry
Biometry
Biometry
Use of average sound velocity ,although
sufficient in normal phakic eye , may
result in slight error when the lens is
inordinately thin or thick or when the
eye is very short or very long .
Biometry
The use of individual sound velocity
may provide more consistent and
accurate AL reading .
Biometry
Aphakia & Pseudophakia
Manual measurement mode is
better to help ensure alignment
of sound beam .
If an incorrect eye type is used
an erroneous measurement will
occur .
For determination of correct value
Velocity Conversion Equation
should be used .
Velocity Conversion Equation
True AL = V c /Vm x Apparent AL
Biometry
in
Special Cases
1.Inadequate Patient Fixation
 Low Vision
 Nystagmus
 Blepharospasm
 Strabismus
2. Posterior Staphyloma
Posterior staphylomas often causes
an irregular shape of the ocular wall
resulting in an inability to display a
distinct , high retinal spike , leading
to a significant error in A-scan
measurement .
Deepest portion of the staphyloma
may be located eccentric to macula
thus te measurement may be longer
than true AL along the visual axis .
B-scan can be used to demonstrate
the shape of posterior ocular wall
and the relationship of macula to
the staphyloma .
Biometry
Biometry
Biometry
Probes with fixation light
are preferable
3. High Hyperopia
Immersion technique
is preferable .
 Edema
 DMS
 Tumor
4. Macular Lesions
 RD
The presence of an elevated macular
lesion may prevent the display of a
distinct retinal spike and often causes
a shortened AL measurement .
Biometry
Biometry
5. Vitreous Lesions
 Asteroid Hyalosis
 Vitreous Hemorrhage
 Gas Bubble
Biometry
6. Dense Cataract
Strong sound attenuation produced
by a very dense cataract can
significantly impair the ability to
display spikes from the various
interfaces along the visual axis .
Maximum gain setting may be
required to obtain spikes of
sufficient height from the
posterior lens capsule and retina .
Semiautomatic mode should be
used in eyes with dense cataract
Biometry
7. Silicone Oil
Sound velocity in silicone oil
1040 m/s 5000 cs
980 m/s 1000cs
This low sound velocity can result
in pronounced sound attenuation
and difficulty in identifying the
retinal spikes .
Biometry
If proper sound velocity are not
used , erroneously long AL
measurement will be obtained .
For accurate AL measurement ,
various ocular components should
be measured separately with
appropriate sound velocity .
If biometer provides only preset
sound velocity , AL measurement
can be obtained using velocity
conversion equation .
The least preferred method is
use of average sound velocity
Average sound velocity in eyes with
average length (23.5 mm)
1,139 m/s phakic eye
1,052 m/s aphakic eye
Due to strong sound attenuation
AL measurement often can not be
obtained from an eye containing
emulsified silicone oil .
IOL Master
Biometry
Biometry
Zeiss IOL Master
 Axial Length
 ACD
 Corneal Power
 IOL Power Calculation
Hoffer-Q , SRK/T ,Holladay 1, Haigis
Biometry
Biometry
Keratometry
A second person should confirm measurements prior to A-scan
ultrasonography if: The corneal power is less than 40.0 diopters, or
greater than 47.0 diopters.
If there has been prior keratorefractive surgery. In this case the corneal
power will need to be estimated by either the historical, or the contact lens
method.
The average corneal power difference between the two eyes is greater
than 1.00 diopter.
The patient cannot fixate, as seen with a mature cataract, or macular hole.
The amount of corneal astigmatism by keratometry, or
topography, correlates poorly with the amount of astigmatism on the most
recent manifest refraction.
The corneal diameter is less than 11.00 mm.
There is any problem with patient cooperation, or understanding.
Immersion A-scan Ultrasonography
A second person should re-measure both eyes if: The axial length is less
than 22.00 mm, or greater than 25.00 mm in either eye.
The axial length is greater than 26.0 mm, and there is a poor retinal
spike, or wide variability in the readings.
There is a difference in axial length between the two eyes of greater than
0.33 mm that cannot be correlated with the patient's oldest refraction.
Axial length measurements do not correlate with the patient's refractive
error. In general, myopes should have eyes longer than 24.0 mm and
hyperopes should have eyes shorter than 24.0 mm. Exceptions to this rule
involve steep, or flat corneas. Be sure to use the oldest refractive data.
There is difficulty obtaining correctly positioned, high, steeply rising
echoes, or wide variability in individual axial length readings for either
eye.
There is a difference in axial length between the two eyes of greater
than 0.33 mm that cannot be correlated with the patient's oldest refraction.
Axial length measurements do not correlate with the patient's refractive error.
In general, myopes should have eyes longer than 24.0 mm and hyperopes
should have eyes shorter than 24.0 mm. Exceptions to this rule involve steep,
or flat corneas. Be sure to use the oldest refractive data.
There is difficulty obtaining correctly positioned, high, steeply rising echoes,
or wide variability in individual axial length readings for either eye.
Intraocular Lens Power
A second person should repeat the axial length measurements, keratometry
readings and re-run the IOL power calculations for both eyes if: The IOL
power for emmetropia is greater than 3.00 diopters different than
anticipated.
There is a difference in IOL power of greater than 1.00 diopter between the
two eyes.
If the patient has had prior keratorefractive surgery and the calculated IOL
power for standard phacoemulsification is less than +20.0 D or greater than
+23.0 D.
Formula
for
IOL Power Calculation
IOL Power Formula
 Theoretical
 Regression
 Refractive
Theoretical Formulas
These are derived from
geometrical optics
Regression Formulas
Actual postop refractive results
of many lens implantations are
used to predict IOL power
Theoretical Formula
These formulas contain many
assumptions including values of
postop ACD , refractive index of
cornea and ocular humors , retinal
thickness
Theoretical Formula
These formulas are reliable for
average AL , but overestimates
in short eyes and underestimates
in long eyes
Refractive Formulas
IOL power calculation without
determination of axial length
Biometry
Biometry
Biometry
Biometry
Biometry
SRK I
(Sanders,Retzlaff,Kraff)
P = A – 2.5L – 0.9K
It generally undercorrects short eyes
and overcorrects long eyes
SRK II
A1 = A + 3 AL < 20mm
A1 = A + 2 AL 20-21
A1 = A + 1 AL 21-22
A1 = A AL 22-24.5
A1 = A – 0.5 AL >24.5
SRK/T
It is a nonlinear theoretical optical
formula empirically optimized for
postop ACD , retinal thickness ,
corneal refractive index .
It combines advantages of theoretical
and regression formulas .
Generations of IOL Formulas
1st Generation
Fyodorov , Colenbrander ,Hoffer , SRK I
2nd Generation
Binkhorst , SRK II
3rd Generation
Holladay 1 , Hoffer-Q , SRK/T
4th Generation
Holladay 2 , Haigis
Biometry
Biometry
Biometry
There are currently three IOL constants in use: The
SRK/T formula uses an "A-constant."
The Holladay 1 formula uses a "Surgeon Factor."
The Holladay 2 formula, and the Hoffer Q formula,
both use an "Anterior Chamber Depth." aka: ACD.
d = the effective lens position, where ...
d = a0 + (a1 * ACD) + (a2 * AL)
Haigis Formula
* The a0 constant basically moves the curve up,
or down, in much the same way that the A-
constant, Surgeon Factor, or ACD does for the
Holladay 1, Holladay 2, Hoffer Q and SRK/T
formulas.
* The a1 constant is tied to the measured
anterior chamber depth.
* The a2 constant is tied to the measured axial
length. The way the a0, a1 and a2 constants are
derived is by generating a set of surgeon, and
IOL-specific
Formula Choice
AL < 19 mm (<0.1%)
Holladay 2
AL 19-22 mm (8%)
Holladay 2 , Hoffer-Q
AL 22-24.5 mm (72%)
SRK II , Hoffer-Q ,Holladay 1
AL 24.5-26 mm (15%)
Holladay 1 , Hoffer-Q
AL > 26 mm ( 15%)
SRK/T
Biometry
Axial Length in mm Haigis
unoptimized
Hoffer Q Holladay 1 Holladay 2 SRK/T
20.00 to 21.99 0.25 D 0.25 D 0.25 - 0.50 D 0.25 D 0.51 - 1.0 D
22.00 to 24.49 0.25 D 0.25 D 0.25 D 0.25 D 0.25 D
24.50 to 25.99 0.25 D 0.25 D 0.25 D 0.25 D 0.25 D
26.00 to 28.00 0.25 - 0.50 D 0.25 - 0.50 D 0.25 D 0.25 D 0.25 D
28.00 to 30.00 0.25 - 0.50 D 0.25 - 0.50 D 0.25 D 0.25 D 0.25 - 0.50 D
Minus power IOLs 0.51 - 1.0 D 0.51 - 1.0 D 0.25 - 0.50 D 0.25 D 0.25 - 0.50 D
Haigis formula
may be appropriate for all
ranges of axial lengths
Biometry
Biometry
Biometry

More Related Content

PPTX
Biometry
PPTX
IOL power calculation special situations
PPTX
Biometry & Iol calculations
PPTX
PPTX
IOL Master
PPTX
Extraocular muscles
PPTX
A scan biometry
PPTX
Maddox rod n wing
Biometry
IOL power calculation special situations
Biometry & Iol calculations
IOL Master
Extraocular muscles
A scan biometry
Maddox rod n wing

What's hot (20)

PPTX
Amblyopia : classification & Occlusion therapy
PDF
Examination protocol for binocular vision
PPTX
Direct & indirect ophthalmoscopy
PPT
Corneal topography
PPTX
Anterior segment OCT & UBM
PPTX
Log mar chart
PPTX
Introduction to binocular single vision (BSV)
PPTX
Accommodation/ Accommodation of Eye, Measurement of Accommodation of Eye (hea...
PPTX
Diplopia charting
PPTX
Hirschberg and krimsky test.pptx
PPTX
Convergence insufficiency
PPTX
Colour vision test
PPTX
Pediatric refraction
PPTX
Visiontherapy
PPTX
Synaptophore
PPTX
Maddox Rod test
PPTX
Optical Coherence Tomography (OCT)
PPTX
Optics of contact lens
PDF
Tear function tests
Amblyopia : classification & Occlusion therapy
Examination protocol for binocular vision
Direct & indirect ophthalmoscopy
Corneal topography
Anterior segment OCT & UBM
Log mar chart
Introduction to binocular single vision (BSV)
Accommodation/ Accommodation of Eye, Measurement of Accommodation of Eye (hea...
Diplopia charting
Hirschberg and krimsky test.pptx
Convergence insufficiency
Colour vision test
Pediatric refraction
Visiontherapy
Synaptophore
Maddox Rod test
Optical Coherence Tomography (OCT)
Optics of contact lens
Tear function tests
Ad

Viewers also liked (20)

PPT
Biometry for Cataract
PPTX
IOL power calculation formulae
PPT
Biometry: Iol calculation
PPTX
Post lasik iol Surgery
PPT
Biometry
PPTX
Multifocal iols
PPT
IOL Power Calculation in Normal Eyes
PPTX
Ocular viscosurgical devices
PPTX
Update in intraocular lenses
PPTX
PPTX
Ocular viscoelastic devices(OVD)
PPTX
RECENT ADVANCES IN INTRAOCULAR LENS
PPT
Goldman applanation tonometry
PPT
Marketing an optometric practice
PPTX
Blepharitis
PPT
Spectacle prescription fulfillment in medical optometry cope approved
PPTX
Intermittent exotropia
PPT
Primetry
PPTX
Presbyopia
Biometry for Cataract
IOL power calculation formulae
Biometry: Iol calculation
Post lasik iol Surgery
Biometry
Multifocal iols
IOL Power Calculation in Normal Eyes
Ocular viscosurgical devices
Update in intraocular lenses
Ocular viscoelastic devices(OVD)
RECENT ADVANCES IN INTRAOCULAR LENS
Goldman applanation tonometry
Marketing an optometric practice
Blepharitis
Spectacle prescription fulfillment in medical optometry cope approved
Intermittent exotropia
Primetry
Presbyopia
Ad

Similar to Biometry (20)

PPTX
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
PPTX
Biometery
PPT
Biometry Yonas.res.ppt
PPT
Biometry Yonas.res.ppt
PPTX
Iol power calculation normal and post lasik eyes
PPTX
Optical Biometry Measurements For Future Iol’S
DOCX
A project on
PPTX
PPTX
Biometry
PPTX
A Scan- Basics and Update
PPT
High Precision Biometry
PPT
biometry for ON.ppt
PPTX
Biometry-1, rifat.pptx
PDF
A scan biometry | How to Use A-scan? Types of A-Scan Biometry?
PPT
Pre And Postoperative Care Of The Modern Cataract Patient
PPTX
OCULAR BIOMETRY AND IOL.pptx
PDF
Biometry made easy
PPTX
Achieving accurate iol power calculations
PPTX
Ocular Biometry- IOL calculation methods
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
Biometery
Biometry Yonas.res.ppt
Biometry Yonas.res.ppt
Iol power calculation normal and post lasik eyes
Optical Biometry Measurements For Future Iol’S
A project on
Biometry
A Scan- Basics and Update
High Precision Biometry
biometry for ON.ppt
Biometry-1, rifat.pptx
A scan biometry | How to Use A-scan? Types of A-Scan Biometry?
Pre And Postoperative Care Of The Modern Cataract Patient
OCULAR BIOMETRY AND IOL.pptx
Biometry made easy
Achieving accurate iol power calculations
Ocular Biometry- IOL calculation methods

More from Hossein Mirzaie (20)

PDF
Mba monograph13 0412_marketinganoptometricpractice
PPT
PPT
keratometry
PPTX
Funduscopy
PPT
Optometry, a focus on vision
PPT
Astigmatism 2
PPT
Non surgical management of strabismus .ppt
PPTX
Contact Lens-Induced Acute Red Eye(CLARE)
PPT
Amblyopia treatment dr mirzajani
PPSX
optometric prescription
PPT
Peripheral refraction in myopia
PDF
Auto perimetry
PPT
Toric rgp fit
PPT
Soft contact lens changes through history (in persian)
PPSX
charles bonnet syndrome
PPTX
3D vision syndrome
PPT
Staining grid lecture
PPTX
Prism.ppt
PPT
Eyeglass frames
PPT
Chapter 27-ophthalmic-care
Mba monograph13 0412_marketinganoptometricpractice
keratometry
Funduscopy
Optometry, a focus on vision
Astigmatism 2
Non surgical management of strabismus .ppt
Contact Lens-Induced Acute Red Eye(CLARE)
Amblyopia treatment dr mirzajani
optometric prescription
Peripheral refraction in myopia
Auto perimetry
Toric rgp fit
Soft contact lens changes through history (in persian)
charles bonnet syndrome
3D vision syndrome
Staining grid lecture
Prism.ppt
Eyeglass frames
Chapter 27-ophthalmic-care

Recently uploaded (20)

PPTX
Pathophysiology And Clinical Features Of Peripheral Nervous System .pptx
PPTX
surgery guide for USMLE step 2-part 1.pptx
PPTX
Electromyography (EMG) in Physiotherapy: Principles, Procedure & Clinical App...
PPTX
neonatal infection(7392992y282939y5.pptx
DOCX
RUHS II MBBS Microbiology Paper-II with Answer Key | 6th August 2025 (New Sch...
PPT
Management of Acute Kidney Injury at LAUTECH
PPTX
POLYCYSTIC OVARIAN SYNDROME.pptx by Dr( med) Charles Amoateng
PPT
ASRH Presentation for students and teachers 2770633.ppt
DOC
Adobe Premiere Pro CC Crack With Serial Key Full Free Download 2025
PPTX
SKIN Anatomy and physiology and associated diseases
PPTX
Note on Abortion.pptx for the student note
PDF
NEET PG 2025 | 200 High-Yield Recall Topics Across All Subjects
PDF
Handout_ NURS 220 Topic 10-Abnormal Pregnancy.pdf
PPTX
15.MENINGITIS AND ENCEPHALITIS-elias.pptx
PPT
Obstructive sleep apnea in orthodontics treatment
PDF
Khadir.pdf Acacia catechu drug Ayurvedic medicine
PPTX
1 General Principles of Radiotherapy.pptx
PDF
Human Health And Disease hggyutgghg .pdf
PPTX
Chapter-1-The-Human-Body-Orientation-Edited-55-slides.pptx
PPTX
Slider: TOC sampling methods for cleaning validation
Pathophysiology And Clinical Features Of Peripheral Nervous System .pptx
surgery guide for USMLE step 2-part 1.pptx
Electromyography (EMG) in Physiotherapy: Principles, Procedure & Clinical App...
neonatal infection(7392992y282939y5.pptx
RUHS II MBBS Microbiology Paper-II with Answer Key | 6th August 2025 (New Sch...
Management of Acute Kidney Injury at LAUTECH
POLYCYSTIC OVARIAN SYNDROME.pptx by Dr( med) Charles Amoateng
ASRH Presentation for students and teachers 2770633.ppt
Adobe Premiere Pro CC Crack With Serial Key Full Free Download 2025
SKIN Anatomy and physiology and associated diseases
Note on Abortion.pptx for the student note
NEET PG 2025 | 200 High-Yield Recall Topics Across All Subjects
Handout_ NURS 220 Topic 10-Abnormal Pregnancy.pdf
15.MENINGITIS AND ENCEPHALITIS-elias.pptx
Obstructive sleep apnea in orthodontics treatment
Khadir.pdf Acacia catechu drug Ayurvedic medicine
1 General Principles of Radiotherapy.pptx
Human Health And Disease hggyutgghg .pdf
Chapter-1-The-Human-Body-Orientation-Edited-55-slides.pptx
Slider: TOC sampling methods for cleaning validation

Biometry

  • 1. Axial Length Measurement ( Biometry ) Mohammad Reza ZARRIN (optometrist- MsC) Tehran University of Medical Sciences
  • 2. IOL Power Calculation 1. Keratometry 2. A-Scan Biometry 3. IOL Formula
  • 3. A-Scan Biometry Measurement of Axial Eye Length by Ultrasound
  • 4. Average Axial Length of Normal Eye 23.06 mm Majority 22.0 to 24.5 mm
  • 5. Accuracy of AL measurement using A-scan ultrasound is + 0.1 mm
  • 6. Difference in AL measurement Between both eyes + 0.3 mm
  • 10. Examination Procedure 1. History Taking 2. Patient Preparation 3. Biometry Technique
  • 11. Biometry Technique  Contact - Applanation Method - Hand-Held Method  Immersion
  • 14. Values are 0.14 to 0.36 mm longer with immersion technique than with contact method
  • 22. Potential Sources of Error with Contact Method 1.Corneal Compression 2. Fluid Excess 3. Misalignment of Sound Beam 4. Inappropriate Eye type
  • 24. Error caused by 1 mm Corneal Compression Average eye 2.5 D Long eye 1.75 D Short eye 3.75 D
  • 26. Potential Sources of Error with Immersion Method 1. Air bubbles within fluid 2. Inappropriate eye type
  • 27. Instrument Setting 1. Measurement Mode 2. Gates 3. Gain 4. Eye Type
  • 28. Measurement Mode  Automatic  Semiautomatic  Manual
  • 29. Gates Gates are electronic markers on the screen that provide measurement of distance between 2 or more anatomic interfaces .
  • 32. Gain Setting Initially high gain setting should be used to assess the overall appearance of the echogram , then gain should be reduced to a medium level to improve resolution of spikes .
  • 34. Error can occur when the gain is set too high or too low . Very high gain short reading Very low gain long reading
  • 35. Eye Type ( Sound Velocity ) 1. Phakic 2. Aphakic 3. Pseudophakic
  • 41. Use of average sound velocity ,although sufficient in normal phakic eye , may result in slight error when the lens is inordinately thin or thick or when the eye is very short or very long .
  • 43. The use of individual sound velocity may provide more consistent and accurate AL reading .
  • 45. Aphakia & Pseudophakia Manual measurement mode is better to help ensure alignment of sound beam .
  • 46. If an incorrect eye type is used an erroneous measurement will occur . For determination of correct value Velocity Conversion Equation should be used .
  • 47. Velocity Conversion Equation True AL = V c /Vm x Apparent AL
  • 49. 1.Inadequate Patient Fixation  Low Vision  Nystagmus  Blepharospasm  Strabismus
  • 50. 2. Posterior Staphyloma Posterior staphylomas often causes an irregular shape of the ocular wall resulting in an inability to display a distinct , high retinal spike , leading to a significant error in A-scan measurement .
  • 51. Deepest portion of the staphyloma may be located eccentric to macula thus te measurement may be longer than true AL along the visual axis .
  • 52. B-scan can be used to demonstrate the shape of posterior ocular wall and the relationship of macula to the staphyloma .
  • 56. Probes with fixation light are preferable
  • 57. 3. High Hyperopia Immersion technique is preferable .
  • 58.  Edema  DMS  Tumor 4. Macular Lesions  RD
  • 59. The presence of an elevated macular lesion may prevent the display of a distinct retinal spike and often causes a shortened AL measurement .
  • 62. 5. Vitreous Lesions  Asteroid Hyalosis  Vitreous Hemorrhage  Gas Bubble
  • 64. 6. Dense Cataract Strong sound attenuation produced by a very dense cataract can significantly impair the ability to display spikes from the various interfaces along the visual axis .
  • 65. Maximum gain setting may be required to obtain spikes of sufficient height from the posterior lens capsule and retina .
  • 66. Semiautomatic mode should be used in eyes with dense cataract
  • 68. 7. Silicone Oil Sound velocity in silicone oil 1040 m/s 5000 cs 980 m/s 1000cs
  • 69. This low sound velocity can result in pronounced sound attenuation and difficulty in identifying the retinal spikes .
  • 71. If proper sound velocity are not used , erroneously long AL measurement will be obtained .
  • 72. For accurate AL measurement , various ocular components should be measured separately with appropriate sound velocity .
  • 73. If biometer provides only preset sound velocity , AL measurement can be obtained using velocity conversion equation .
  • 74. The least preferred method is use of average sound velocity Average sound velocity in eyes with average length (23.5 mm) 1,139 m/s phakic eye 1,052 m/s aphakic eye
  • 75. Due to strong sound attenuation AL measurement often can not be obtained from an eye containing emulsified silicone oil .
  • 79. Zeiss IOL Master  Axial Length  ACD  Corneal Power  IOL Power Calculation Hoffer-Q , SRK/T ,Holladay 1, Haigis
  • 82. Keratometry A second person should confirm measurements prior to A-scan ultrasonography if: The corneal power is less than 40.0 diopters, or greater than 47.0 diopters. If there has been prior keratorefractive surgery. In this case the corneal power will need to be estimated by either the historical, or the contact lens method. The average corneal power difference between the two eyes is greater than 1.00 diopter. The patient cannot fixate, as seen with a mature cataract, or macular hole. The amount of corneal astigmatism by keratometry, or topography, correlates poorly with the amount of astigmatism on the most recent manifest refraction. The corneal diameter is less than 11.00 mm. There is any problem with patient cooperation, or understanding.
  • 83. Immersion A-scan Ultrasonography A second person should re-measure both eyes if: The axial length is less than 22.00 mm, or greater than 25.00 mm in either eye. The axial length is greater than 26.0 mm, and there is a poor retinal spike, or wide variability in the readings. There is a difference in axial length between the two eyes of greater than 0.33 mm that cannot be correlated with the patient's oldest refraction. Axial length measurements do not correlate with the patient's refractive error. In general, myopes should have eyes longer than 24.0 mm and hyperopes should have eyes shorter than 24.0 mm. Exceptions to this rule involve steep, or flat corneas. Be sure to use the oldest refractive data. There is difficulty obtaining correctly positioned, high, steeply rising echoes, or wide variability in individual axial length readings for either eye.
  • 84. There is a difference in axial length between the two eyes of greater than 0.33 mm that cannot be correlated with the patient's oldest refraction. Axial length measurements do not correlate with the patient's refractive error. In general, myopes should have eyes longer than 24.0 mm and hyperopes should have eyes shorter than 24.0 mm. Exceptions to this rule involve steep, or flat corneas. Be sure to use the oldest refractive data. There is difficulty obtaining correctly positioned, high, steeply rising echoes, or wide variability in individual axial length readings for either eye.
  • 85. Intraocular Lens Power A second person should repeat the axial length measurements, keratometry readings and re-run the IOL power calculations for both eyes if: The IOL power for emmetropia is greater than 3.00 diopters different than anticipated. There is a difference in IOL power of greater than 1.00 diopter between the two eyes. If the patient has had prior keratorefractive surgery and the calculated IOL power for standard phacoemulsification is less than +20.0 D or greater than +23.0 D.
  • 87. IOL Power Formula  Theoretical  Regression  Refractive
  • 88. Theoretical Formulas These are derived from geometrical optics
  • 89. Regression Formulas Actual postop refractive results of many lens implantations are used to predict IOL power
  • 90. Theoretical Formula These formulas contain many assumptions including values of postop ACD , refractive index of cornea and ocular humors , retinal thickness
  • 91. Theoretical Formula These formulas are reliable for average AL , but overestimates in short eyes and underestimates in long eyes
  • 92. Refractive Formulas IOL power calculation without determination of axial length
  • 98. SRK I (Sanders,Retzlaff,Kraff) P = A – 2.5L – 0.9K It generally undercorrects short eyes and overcorrects long eyes
  • 99. SRK II A1 = A + 3 AL < 20mm A1 = A + 2 AL 20-21 A1 = A + 1 AL 21-22 A1 = A AL 22-24.5 A1 = A – 0.5 AL >24.5
  • 100. SRK/T It is a nonlinear theoretical optical formula empirically optimized for postop ACD , retinal thickness , corneal refractive index . It combines advantages of theoretical and regression formulas .
  • 101. Generations of IOL Formulas 1st Generation Fyodorov , Colenbrander ,Hoffer , SRK I 2nd Generation Binkhorst , SRK II 3rd Generation Holladay 1 , Hoffer-Q , SRK/T 4th Generation Holladay 2 , Haigis
  • 105. There are currently three IOL constants in use: The SRK/T formula uses an "A-constant." The Holladay 1 formula uses a "Surgeon Factor." The Holladay 2 formula, and the Hoffer Q formula, both use an "Anterior Chamber Depth." aka: ACD.
  • 106. d = the effective lens position, where ... d = a0 + (a1 * ACD) + (a2 * AL) Haigis Formula
  • 107. * The a0 constant basically moves the curve up, or down, in much the same way that the A- constant, Surgeon Factor, or ACD does for the Holladay 1, Holladay 2, Hoffer Q and SRK/T formulas. * The a1 constant is tied to the measured anterior chamber depth. * The a2 constant is tied to the measured axial length. The way the a0, a1 and a2 constants are derived is by generating a set of surgeon, and IOL-specific
  • 109. AL < 19 mm (<0.1%) Holladay 2 AL 19-22 mm (8%) Holladay 2 , Hoffer-Q AL 22-24.5 mm (72%) SRK II , Hoffer-Q ,Holladay 1 AL 24.5-26 mm (15%) Holladay 1 , Hoffer-Q AL > 26 mm ( 15%) SRK/T
  • 111. Axial Length in mm Haigis unoptimized Hoffer Q Holladay 1 Holladay 2 SRK/T 20.00 to 21.99 0.25 D 0.25 D 0.25 - 0.50 D 0.25 D 0.51 - 1.0 D 22.00 to 24.49 0.25 D 0.25 D 0.25 D 0.25 D 0.25 D 24.50 to 25.99 0.25 D 0.25 D 0.25 D 0.25 D 0.25 D 26.00 to 28.00 0.25 - 0.50 D 0.25 - 0.50 D 0.25 D 0.25 D 0.25 D 28.00 to 30.00 0.25 - 0.50 D 0.25 - 0.50 D 0.25 D 0.25 D 0.25 - 0.50 D Minus power IOLs 0.51 - 1.0 D 0.51 - 1.0 D 0.25 - 0.50 D 0.25 D 0.25 - 0.50 D
  • 112. Haigis formula may be appropriate for all ranges of axial lengths