National Ocular Biometry
Course 2015
An echoing presentation
Prepared by:
Anis Suzanna Binti Mohamad
Rohaila Binti Ariffin
Pegawai Optometri U41
Hospital Sultanah Bahiyah, Alor Setar
COURSE OBJECTIVES:
 Understand the basic principal of biometry
 Perform, define the discuss the pathology that would require the
biometry test
 Interpret the result of biometry test
 Measure axial length
 Name four ocular structures that reflect ultrasound echoes
 Able to perform biometry test
 Able to perform topography, discuss corneal contours of normal,
pathological and post-surgical cornea
 Take pre-operative testing and intraocular lens power calculation
method to the next level of accuracy and efficiency
 Utilize standardize diagnostic ultrasonography as a valuable
adjuvant to current clinical evaluation.
Cataract surgery
• What is cataract surgery?
– Precise biometry is essential for accurate outcomes in cataract and
refractive surgeries.
– The measurement of axial length by ultrasound was the gold
standard for many years.
– With the introduction of optical biometry in the US in the year
2000, this technology has become more & more popular & is now
the most common method for the measurement of axial length.
– Optical biometry uses a partially coherent wave that has approx. 9x
the resolution of a 10 MHz sound wave, making the measurement
of axial length very precise.
– It helps to avoid operator variations in measurement.
– Also, to increase the accuracy as contact with cornea is not needed
(eliminates compression to cornea).
– Since Optical biometry measures to the centre of macula, it gives
the refractive AXL vs the anatomical AXL achieved with ulrasound.
Pre-op assessment
1. Keratometry
2. A-Scan Biometry
3. IOL Formula
1. K-reading
Keratometry
Manual K Topography
Placido disc
Scheimflug
topography
Auto-K
Optical
biometry
Manual Keratometry Topography (Placido Disc)
K-reading in our center
Auto-K Optical Biometry (Lenstar)
K-reading in our center
Why we choose Lenstar in our center?
2. A-Scan Biometry
• Measurement of Axial Eye Length by Ultrasound
• Average Axial Length of Normal Eye 23.06 mm (Majority :
22.0 to 24.5 mm)
• Accuracy of AL measurement using A-scan ultrasound is +
0.1 mm
• Difference in AL measurement between both eyes + 0.3 mm
• Values are 0.14 to 0.36 mm longer with immersion
technique than with contact method
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
Examination Procedure
1. History Taking
2. Patient Preparation
3. Biometry Technique
Biometry
Technique
Ultrasound
Applanation Immersion
PCI
IOL Master Lenstar
Ultrasound biometry
Ultrasound is
produced in the
ultrasonographic
probe by the
oscillation of a
pizoelectric
crystal
Converting
electrical energy
into mechanical
energy
The probe emits
and receives
pulsed signal
Reflectivity vs
time is displayed
for the single
direction in which
the probe is
pointing
This value then
be converted to
milimitres
Contact A-scan
Immersion A-scan
Applanation (NIDEK)
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
Immersion
Scleral shell
(Prager)
Scleral shell
(Prager) (Ossoining) (Kohn)
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
Potential Sources of Error
with Contact Method
1. Corneal Compression
2. Fluid Excess
3. Misalignment of Sound Beam
4. Inappropriate Eye type
Error caused by 1 mm Corneal Compression:
Average eye 2.5 D
Long eye 1.75 D
Short eye 3.75 D
Corneal compression
Fluid excess
Potential Sources of Error
with Immersion Method
1. Air bubbles within fluid
2. Inappropriate eye type
2. Inappropriate eye type
Instrument Setting
1. Eye to measure (OD/OS)
2. Eye type
3. Measurement mode
4. Technique
5. Gain
Eye Type
( Sound Velocity )
1. Phakic
2. Aphakic
3. Pseudophakic
C A P + R S
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
Measurement Mode
 Automatic
 Semiautomatic
 Manual
Gain Setting
• Initially high gain setting should be used to assess the overall
appearance of the echogram , then gain should be reduced to a
medium level to improve resolution of spikes .
• Error can occur when the gain is set too high or too low .
Very high gain : short reading
Very low gain : long reading
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
Use of average sound velocity ,although
sufficient in normal phakic eye , may
result in slight error when the lens is
inordinately thin or thick or when the
eye is very short or very long .
• The use of individual sound velocity may provide
more consistent and accurate AL reading .
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
• Manual measurement mode is better to help
ensure alignment of sound beam .
Aphakia & Pseudophakia
Posterior Staphyloma
• Posterior staphylomas often causes an irregular shape of the ocular wall
resulting in an inability to display a distinct , high retinal spike , leading to a
significant error in A-scan measurement .
• Deepest portion of the staphyloma may be located eccentric to macula thus the
measurement may be longer than true AL along the visual axis .
• B-scan can be used to demonstrate the shape of posterior ocular wall and the
relationship of macula to the staphyloma .
• Probes with fixation light are preferable
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
High Hyperopia
• Immersion technique is preferable .
 Edema
 DMS
 Tumor
Macular Lesions
 RD
The presence of an elevated macular lesion may
prevent the display of a distinct retinal spike and
often causes a shortened AL measurement .
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
Vitreous Lesions
 Asteroid Hyalosis
 Vitreous Hemorrhage
 Gas Bubble
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
Dense Cataract
• Strong sound attenuation produced by a very dense cataract can
significantly impair the ability to display spikes from the various
interfaces along the visual axis .
• Maximum gain setting may be required to obtain spikes of
sufficient height from the posterior lens capsule and retina .
• Semiautomatic mode should be used in eyes with dense cataract.
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
Silicone Oil
• Sound velocity in silicone oil
1040 m/s 5000 cs
980 m/s 1000cs
• This low sound velocity can result in
pronounced sound attenuation and difficulty
in identifying the retinal spikes .
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
• If proper sound velocity are not used , erroneously long AL
measurement will be obtained .
• For accurate AL measurement ,various ocular components
should be measured separately with appropriate sound
velocity .
• The least preferred method is use of average sound velocity.
Average sound velocity in eyes with average length (23.5 mm):
1,139 m/s phakic eye
1,052 m/s aphakic eye
• Due to strong sound attenuation AL measurement often can
not be obtained from an eye containing emulsified silicone oil .
IOL Master/Lenstar
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
Zeiss IOL Master
 Axial Length
 ACD
 Corneal Power
 IOL Power Calculation
Hoffer-Q , SRK/T ,Holladay 1, Haigis
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
Keratometry
A second person should confirm measurements prior to A-scan
ultrasonography if: The corneal power is less than 40.0 diopters, or
greater than 47.0 diopters.
If there has been prior keratorefractive surgery. In this case the corneal
power will need to be estimated by either the historical, or the contact lens
method.
The average corneal power difference between the two eyes is greater
than 1.00 diopter.
The patient cannot fixate, as seen with a mature cataract, or macular hole.
The amount of corneal astigmatism by keratometry, or topography,
correlates poorly with the amount of astigmatism on the most recent
manifest refraction.
The corneal diameter is less than 11.00 mm.
There is any problem with patient cooperation, or understanding.
Immersion A-scan Ultrasonography
A second person should re-measure both eyes if: The axial length is less
than 22.00 mm, or greater than 25.00 mm in either eye.
The axial length is greater than 26.0 mm, and there is a poor retinal spike,
or wide variability in the readings.
There is a difference in axial length between the two eyes of greater
than 0.33 mm that cannot be correlated with the patient's oldest
refraction.
Axial length measurements do not correlate with the patient's refractive
error. In general, myopes should have eyes longer than 24.0 mm and
hyperopes should have eyes shorter than 24.0 mm. Exceptions to this
rule involve steep, or flat corneas. Be sure to use the oldest refractive
data.
There is difficulty obtaining correctly positioned, high, steeply rising
echoes, or wide variability in individual axial length readings for either
eye.
Intraocular Lens Power
A second person should repeat the axial length measurements,
keratometry readings and re-run the IOL power calculations for both eyes
if: The IOL power for emmetropia is greater than 3.00 diopters different
than anticipated.
There is a difference in IOL power of greater than 1.00 diopter between the
two eyes.
If the patient has had prior keratorefractive surgery and the calculated IOL
power for standard phacoemulsification is less than +20.0 D or greater than
+23.0 D.
Formula
for
IOL Power Calculation
IOL Power Formula
 Theoretical
 Regression
 Refractive
Theoretical Formulas
These are derived from
geometrical optics
Regression Formulas
Actual postop refractive results
of many lens implantations are
used to predict IOL power
Theoretical Formula
These formulas contain many
assumptions including values of
postop ACD , refractive index of
cornea and ocular humors , retinal
thickness
Theoretical Formula
These formulas are reliable for
average AL , but overestimates
in short eyes and underestimates
in long eyes
Refractive Formulas
IOL power calculation without
determination of axial length
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
SRK I
(Sanders,Retzlaff,Kraff)
P = A – 2.5L – 0.9K
It generally undercorrects short eyes
and overcorrects long eyes
SRK II
A1 = A + 3 AL < 20mm
A1 = A + 2 AL 20-21
A1 = A + 1 AL 21-22
A1 = A AL 22-24.5
A1 = A – 0.5 AL >24.5
SRK/T
• It is a nonlinear theoretical optical formula
empirically optimized for postop ACD, retinal
thickness , corneal refractive index . It
combines advantages of theoretical and
regression formulas .
Generations of IOL Formulas
• 1st Generation
Fyodorov , Colenbrander ,Hoffer , SRK I
• 2nd Generation
Binkhorst , SRK II
• 3rd Generation
Holladay 1 , Hoffer-Q , SRK/T
• 4th Generation
Holladay 2 , Haigis, Hoffer-H
• 5th Generation (2014)
Hoffer H-5
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
There are currently three IOL constants in use: The
SRK/T formula uses an "A-constant."
The Holladay 1 formula uses a "Surgeon Factor."
The Holladay 2 formula, and the Hoffer Q formula,
both use an "Anterior Chamber Depth." aka: ACD.
d = the effective lens position, where ...
d = a0 + (a1 * ACD) + (a2 * AL)
Haigis Formula
* The a0 constant basically moves the curve up,
or down, in much the same way that the A-
constant, Surgeon Factor, or ACD does for the
Holladay 1, Holladay 2, Hoffer Q and SRK/T
formulas.
* The a1 constant is tied to the measured
anterior chamber depth.
* The a2 constant is tied to the measured axial
length. The way the a0, a1 and a2 constants are
derived is by generating a set of surgeon, and
IOL-specific
AL < 19 mm (<0.1%)
Holladay 2
AL 19-22 mm (8%)
Holladay 2 , Hoffer-Q
AL 22-24.5 mm (72%)
SRK II , Hoffer-Q ,Holladay 1
AL 24.5-26 mm (15%)
Holladay 1 , Hoffer-Q
AL > 26 mm ( 15%)
SRK/T
Formula Choice
Haigis formula may be appropriate for all ranges of axial lengths
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation

More Related Content

PPTX
Biometry
PPT
Iol master
PPTX
Keratometer and keratometry
PPTX
IOL POWER CALCULATION IN DIFFICULT SITUATIONS
PPTX
Phakic IOL
PPTX
IOL Master
PPT
Scleral lenses presentation final (1)
PPTX
Biometry.pptx
Biometry
Iol master
Keratometer and keratometry
IOL POWER CALCULATION IN DIFFICULT SITUATIONS
Phakic IOL
IOL Master
Scleral lenses presentation final (1)
Biometry.pptx

What's hot (20)

PPTX
Biometery
PPTX
IOL power calculation special situations
PPTX
RGP Complications
PPTX
Biometry & Iol calculations
PPT
Biometry for Cataract
PPTX
Worth 4 dot test
PPTX
Bandage Contact Lens
PPTX
Progressive addition lenses
PPTX
Ultrasonography in ophthalmology
PDF
Lenstar ls 900[1]
PPTX
Iol power calculation in pediatric patients
PPTX
pentacam
PPTX
Low vision aids magnification
PPTX
BIOMETRY AND IOL POWER CALCULATION
PPTX
PPTX
Optical and Non-optical Methods of Measuring Axial Length of Eye
PPTX
ABBEROMETRY.pptx
PPTX
Diplopia charting
PPTX
Introduction to binocular single vision (BSV)
PPTX
Scheimpflug imaging in ophthalmology
Biometery
IOL power calculation special situations
RGP Complications
Biometry & Iol calculations
Biometry for Cataract
Worth 4 dot test
Bandage Contact Lens
Progressive addition lenses
Ultrasonography in ophthalmology
Lenstar ls 900[1]
Iol power calculation in pediatric patients
pentacam
Low vision aids magnification
BIOMETRY AND IOL POWER CALCULATION
Optical and Non-optical Methods of Measuring Axial Length of Eye
ABBEROMETRY.pptx
Diplopia charting
Introduction to binocular single vision (BSV)
Scheimpflug imaging in ophthalmology
Ad

Viewers also liked (20)

PPTX
Biometry
PPTX
Biometry
PPT
High Precision Biometry
PPT
Biometry: Iol calculation
PPTX
Optical Biometry Measurements For Future Iol’S
PPTX
Brief history biometry iol calculation formula &amp; a constant optimization
PPTX
Biometry instruments & equipment
PPT
Biometry
PPT
IOL Power Calculation in Normal Eyes
PPTX
A scan ultrasonography
PPTX
IOL power calculation formulae
PPT
Measurement of corneal curvature
PPT
Keratometer Slides
PPT
High Performance IOL Power Calculation…. at the push of a button
PPTX
Biometry
PPT
Cosmetic Correction Of Anterior Staphyloma
PDF
Dioptric Distress
PPTX
Temporal vision
PPT
Holladay2
PPTX
Retinoscopy on human eye
Biometry
Biometry
High Precision Biometry
Biometry: Iol calculation
Optical Biometry Measurements For Future Iol’S
Brief history biometry iol calculation formula &amp; a constant optimization
Biometry instruments & equipment
Biometry
IOL Power Calculation in Normal Eyes
A scan ultrasonography
IOL power calculation formulae
Measurement of corneal curvature
Keratometer Slides
High Performance IOL Power Calculation…. at the push of a button
Biometry
Cosmetic Correction Of Anterior Staphyloma
Dioptric Distress
Temporal vision
Holladay2
Retinoscopy on human eye
Ad

Similar to National Ocular Biometry Course (NOBC) 2015 An echoslide presentation (20)

PPTX
OCULAR BIOMETRY AND IOL.pptx
PPTX
22. Biometry ophthalmology by Dr Rahul.pptx
PPTX
introduction into Biometry and it's formulas.pptx
PPTX
DOCX
A project on
PPTX
Biometry-1, rifat.pptx
PDF
A scan biometry | How to Use A-scan? Types of A-Scan Biometry?
PPTX
KERATOMETRY AND IOL POWER CALCULATION-1.pptx
PPT
biometry for ON.ppt
PPTX
IOL CALCULATIONS BY MRS ANTHONIA.pptx
PPT
Biometry Yonas.res.ppt
PPT
Biometry Yonas.res.ppt
PPTX
Biometry mistakes and how to avoid
PPTX
biometry and iol calculation in children 1.pptx
PPTX
Ocular Biometry- IOL calculation methods
PDF
biometry-160212135415 (1).pdf
PPTX
A scan biometry
PPTX
A-Scan Biometry.pptx
PPTX
Biometry explanation and the used formulas.pptx
PDF
Biometry made easy
OCULAR BIOMETRY AND IOL.pptx
22. Biometry ophthalmology by Dr Rahul.pptx
introduction into Biometry and it's formulas.pptx
A project on
Biometry-1, rifat.pptx
A scan biometry | How to Use A-scan? Types of A-Scan Biometry?
KERATOMETRY AND IOL POWER CALCULATION-1.pptx
biometry for ON.ppt
IOL CALCULATIONS BY MRS ANTHONIA.pptx
Biometry Yonas.res.ppt
Biometry Yonas.res.ppt
Biometry mistakes and how to avoid
biometry and iol calculation in children 1.pptx
Ocular Biometry- IOL calculation methods
biometry-160212135415 (1).pdf
A scan biometry
A-Scan Biometry.pptx
Biometry explanation and the used formulas.pptx
Biometry made easy

More from Anis Suzanna Mohamad (20)

PDF
Chapter 9. CVI Skills Inventory & Strategies Worksheet.pdf
PPTX
NED CSR OUTCOME REPORT NOV 2024 - JAN 2025
PDF
APA ITU MYOPIA? BERTINDAK SEBELUM TERLAMBAT !
PDF
OPTOMETRIC ASSESSMENT SO WHAT THEY DO REALLY DO?
PDF
CEREBRAL VISUAL IMPAIRMENT SERVICE IN HPP.pdf
PDF
Autism Awareness for Optometrist Going Down The Rabbit Hole.pdf
PPTX
Retinopathy_of_Prematurity_Presentation.pptx
PPTX
EYE SAFETY AT WORK IN CONJUCTION WITH WORLD SIGHT DAY 2023
PPTX
BV and Pediatrics_Make it interesting.pptx
PDF
A quick revision on Corneal topography
PPTX
Cataract management in children from optometrist perspective
PPTX
Prosedur Operasi Standard Perkhidmatan Penjagaan pesakit Anomali Penglihatan ...
PPTX
Revision about Hess Chart
PPTX
Case presentation: Consecutive esotropia
PPTX
Diagnosis & management of accomodative esotropia
PPTX
Binocular vision patient....what should I do?
PPTX
The vascular coat of the eye
PPTX
Saringan penglihatan kanak kanak
PPT
Ocular disease in peadiatric
PPT
Eso deviation
Chapter 9. CVI Skills Inventory & Strategies Worksheet.pdf
NED CSR OUTCOME REPORT NOV 2024 - JAN 2025
APA ITU MYOPIA? BERTINDAK SEBELUM TERLAMBAT !
OPTOMETRIC ASSESSMENT SO WHAT THEY DO REALLY DO?
CEREBRAL VISUAL IMPAIRMENT SERVICE IN HPP.pdf
Autism Awareness for Optometrist Going Down The Rabbit Hole.pdf
Retinopathy_of_Prematurity_Presentation.pptx
EYE SAFETY AT WORK IN CONJUCTION WITH WORLD SIGHT DAY 2023
BV and Pediatrics_Make it interesting.pptx
A quick revision on Corneal topography
Cataract management in children from optometrist perspective
Prosedur Operasi Standard Perkhidmatan Penjagaan pesakit Anomali Penglihatan ...
Revision about Hess Chart
Case presentation: Consecutive esotropia
Diagnosis & management of accomodative esotropia
Binocular vision patient....what should I do?
The vascular coat of the eye
Saringan penglihatan kanak kanak
Ocular disease in peadiatric
Eso deviation

Recently uploaded (20)

PPTX
neonatal infection(7392992y282939y5.pptx
PPTX
Imaging of parasitic D. Case Discussions.pptx
PPT
Copy-Histopathology Practical by CMDA ESUTH CHAPTER(0) - Copy.ppt
PPTX
CEREBROVASCULAR DISORDER.POWERPOINT PRESENTATIONx
PPTX
Pathophysiology And Clinical Features Of Peripheral Nervous System .pptx
PDF
CT Anatomy for Radiotherapy.pdf eryuioooop
PDF
Deadly Stampede at Yaounde’s Olembe Stadium Forensic.pdf
PPTX
JUVENILE NASOPHARYNGEAL ANGIOFIBROMA.pptx
DOC
Adobe Premiere Pro CC Crack With Serial Key Full Free Download 2025
PPTX
SKIN Anatomy and physiology and associated diseases
PPTX
Fundamentals of human energy transfer .pptx
DOCX
RUHS II MBBS Microbiology Paper-II with Answer Key | 6th August 2025 (New Sch...
PPTX
NEET PG 2025 Pharmacology Recall | Real Exam Questions from 3rd August with D...
PPT
MENTAL HEALTH - NOTES.ppt for nursing students
PPTX
Gastroschisis- Clinical Overview 18112311
PPT
genitourinary-cancers_1.ppt Nursing care of clients with GU cancer
PPT
CHAPTER FIVE. '' Association in epidemiological studies and potential errors
PDF
Khadir.pdf Acacia catechu drug Ayurvedic medicine
PPT
ASRH Presentation for students and teachers 2770633.ppt
PPTX
Electromyography (EMG) in Physiotherapy: Principles, Procedure & Clinical App...
neonatal infection(7392992y282939y5.pptx
Imaging of parasitic D. Case Discussions.pptx
Copy-Histopathology Practical by CMDA ESUTH CHAPTER(0) - Copy.ppt
CEREBROVASCULAR DISORDER.POWERPOINT PRESENTATIONx
Pathophysiology And Clinical Features Of Peripheral Nervous System .pptx
CT Anatomy for Radiotherapy.pdf eryuioooop
Deadly Stampede at Yaounde’s Olembe Stadium Forensic.pdf
JUVENILE NASOPHARYNGEAL ANGIOFIBROMA.pptx
Adobe Premiere Pro CC Crack With Serial Key Full Free Download 2025
SKIN Anatomy and physiology and associated diseases
Fundamentals of human energy transfer .pptx
RUHS II MBBS Microbiology Paper-II with Answer Key | 6th August 2025 (New Sch...
NEET PG 2025 Pharmacology Recall | Real Exam Questions from 3rd August with D...
MENTAL HEALTH - NOTES.ppt for nursing students
Gastroschisis- Clinical Overview 18112311
genitourinary-cancers_1.ppt Nursing care of clients with GU cancer
CHAPTER FIVE. '' Association in epidemiological studies and potential errors
Khadir.pdf Acacia catechu drug Ayurvedic medicine
ASRH Presentation for students and teachers 2770633.ppt
Electromyography (EMG) in Physiotherapy: Principles, Procedure & Clinical App...

National Ocular Biometry Course (NOBC) 2015 An echoslide presentation

  • 1. National Ocular Biometry Course 2015 An echoing presentation Prepared by: Anis Suzanna Binti Mohamad Rohaila Binti Ariffin Pegawai Optometri U41 Hospital Sultanah Bahiyah, Alor Setar
  • 2. COURSE OBJECTIVES:  Understand the basic principal of biometry  Perform, define the discuss the pathology that would require the biometry test  Interpret the result of biometry test  Measure axial length  Name four ocular structures that reflect ultrasound echoes  Able to perform biometry test  Able to perform topography, discuss corneal contours of normal, pathological and post-surgical cornea  Take pre-operative testing and intraocular lens power calculation method to the next level of accuracy and efficiency  Utilize standardize diagnostic ultrasonography as a valuable adjuvant to current clinical evaluation.
  • 3. Cataract surgery • What is cataract surgery? – Precise biometry is essential for accurate outcomes in cataract and refractive surgeries. – The measurement of axial length by ultrasound was the gold standard for many years. – With the introduction of optical biometry in the US in the year 2000, this technology has become more & more popular & is now the most common method for the measurement of axial length. – Optical biometry uses a partially coherent wave that has approx. 9x the resolution of a 10 MHz sound wave, making the measurement of axial length very precise. – It helps to avoid operator variations in measurement. – Also, to increase the accuracy as contact with cornea is not needed (eliminates compression to cornea). – Since Optical biometry measures to the centre of macula, it gives the refractive AXL vs the anatomical AXL achieved with ulrasound.
  • 4. Pre-op assessment 1. Keratometry 2. A-Scan Biometry 3. IOL Formula
  • 5. 1. K-reading Keratometry Manual K Topography Placido disc Scheimflug topography Auto-K Optical biometry
  • 6. Manual Keratometry Topography (Placido Disc) K-reading in our center
  • 7. Auto-K Optical Biometry (Lenstar) K-reading in our center
  • 8. Why we choose Lenstar in our center?
  • 9. 2. A-Scan Biometry • Measurement of Axial Eye Length by Ultrasound • Average Axial Length of Normal Eye 23.06 mm (Majority : 22.0 to 24.5 mm) • Accuracy of AL measurement using A-scan ultrasound is + 0.1 mm • Difference in AL measurement between both eyes + 0.3 mm • Values are 0.14 to 0.36 mm longer with immersion technique than with contact method
  • 12. Examination Procedure 1. History Taking 2. Patient Preparation 3. Biometry Technique
  • 14. Ultrasound biometry Ultrasound is produced in the ultrasonographic probe by the oscillation of a pizoelectric crystal Converting electrical energy into mechanical energy The probe emits and receives pulsed signal Reflectivity vs time is displayed for the single direction in which the probe is pointing This value then be converted to milimitres
  • 22. Potential Sources of Error with Contact Method 1. Corneal Compression 2. Fluid Excess 3. Misalignment of Sound Beam 4. Inappropriate Eye type Error caused by 1 mm Corneal Compression: Average eye 2.5 D Long eye 1.75 D Short eye 3.75 D
  • 25. Potential Sources of Error with Immersion Method 1. Air bubbles within fluid 2. Inappropriate eye type
  • 27. Instrument Setting 1. Eye to measure (OD/OS) 2. Eye type 3. Measurement mode 4. Technique 5. Gain
  • 28. Eye Type ( Sound Velocity ) 1. Phakic 2. Aphakic 3. Pseudophakic C A P + R S
  • 30. Measurement Mode  Automatic  Semiautomatic  Manual
  • 31. Gain Setting • Initially high gain setting should be used to assess the overall appearance of the echogram , then gain should be reduced to a medium level to improve resolution of spikes . • Error can occur when the gain is set too high or too low . Very high gain : short reading Very low gain : long reading
  • 33. Use of average sound velocity ,although sufficient in normal phakic eye , may result in slight error when the lens is inordinately thin or thick or when the eye is very short or very long .
  • 34. • The use of individual sound velocity may provide more consistent and accurate AL reading .
  • 36. • Manual measurement mode is better to help ensure alignment of sound beam . Aphakia & Pseudophakia
  • 37. Posterior Staphyloma • Posterior staphylomas often causes an irregular shape of the ocular wall resulting in an inability to display a distinct , high retinal spike , leading to a significant error in A-scan measurement . • Deepest portion of the staphyloma may be located eccentric to macula thus the measurement may be longer than true AL along the visual axis . • B-scan can be used to demonstrate the shape of posterior ocular wall and the relationship of macula to the staphyloma . • Probes with fixation light are preferable
  • 41. High Hyperopia • Immersion technique is preferable .
  • 42.  Edema  DMS  Tumor Macular Lesions  RD The presence of an elevated macular lesion may prevent the display of a distinct retinal spike and often causes a shortened AL measurement .
  • 45. Vitreous Lesions  Asteroid Hyalosis  Vitreous Hemorrhage  Gas Bubble
  • 47. Dense Cataract • Strong sound attenuation produced by a very dense cataract can significantly impair the ability to display spikes from the various interfaces along the visual axis . • Maximum gain setting may be required to obtain spikes of sufficient height from the posterior lens capsule and retina . • Semiautomatic mode should be used in eyes with dense cataract.
  • 49. Silicone Oil • Sound velocity in silicone oil 1040 m/s 5000 cs 980 m/s 1000cs • This low sound velocity can result in pronounced sound attenuation and difficulty in identifying the retinal spikes .
  • 51. • If proper sound velocity are not used , erroneously long AL measurement will be obtained . • For accurate AL measurement ,various ocular components should be measured separately with appropriate sound velocity . • The least preferred method is use of average sound velocity. Average sound velocity in eyes with average length (23.5 mm): 1,139 m/s phakic eye 1,052 m/s aphakic eye • Due to strong sound attenuation AL measurement often can not be obtained from an eye containing emulsified silicone oil .
  • 55. Zeiss IOL Master  Axial Length  ACD  Corneal Power  IOL Power Calculation Hoffer-Q , SRK/T ,Holladay 1, Haigis
  • 58. Keratometry A second person should confirm measurements prior to A-scan ultrasonography if: The corneal power is less than 40.0 diopters, or greater than 47.0 diopters. If there has been prior keratorefractive surgery. In this case the corneal power will need to be estimated by either the historical, or the contact lens method. The average corneal power difference between the two eyes is greater than 1.00 diopter. The patient cannot fixate, as seen with a mature cataract, or macular hole. The amount of corneal astigmatism by keratometry, or topography, correlates poorly with the amount of astigmatism on the most recent manifest refraction. The corneal diameter is less than 11.00 mm. There is any problem with patient cooperation, or understanding.
  • 59. Immersion A-scan Ultrasonography A second person should re-measure both eyes if: The axial length is less than 22.00 mm, or greater than 25.00 mm in either eye. The axial length is greater than 26.0 mm, and there is a poor retinal spike, or wide variability in the readings. There is a difference in axial length between the two eyes of greater than 0.33 mm that cannot be correlated with the patient's oldest refraction. Axial length measurements do not correlate with the patient's refractive error. In general, myopes should have eyes longer than 24.0 mm and hyperopes should have eyes shorter than 24.0 mm. Exceptions to this rule involve steep, or flat corneas. Be sure to use the oldest refractive data. There is difficulty obtaining correctly positioned, high, steeply rising echoes, or wide variability in individual axial length readings for either eye.
  • 60. Intraocular Lens Power A second person should repeat the axial length measurements, keratometry readings and re-run the IOL power calculations for both eyes if: The IOL power for emmetropia is greater than 3.00 diopters different than anticipated. There is a difference in IOL power of greater than 1.00 diopter between the two eyes. If the patient has had prior keratorefractive surgery and the calculated IOL power for standard phacoemulsification is less than +20.0 D or greater than +23.0 D.
  • 62. IOL Power Formula  Theoretical  Regression  Refractive
  • 63. Theoretical Formulas These are derived from geometrical optics
  • 64. Regression Formulas Actual postop refractive results of many lens implantations are used to predict IOL power
  • 65. Theoretical Formula These formulas contain many assumptions including values of postop ACD , refractive index of cornea and ocular humors , retinal thickness
  • 66. Theoretical Formula These formulas are reliable for average AL , but overestimates in short eyes and underestimates in long eyes
  • 67. Refractive Formulas IOL power calculation without determination of axial length
  • 72. SRK I (Sanders,Retzlaff,Kraff) P = A – 2.5L – 0.9K It generally undercorrects short eyes and overcorrects long eyes
  • 73. SRK II A1 = A + 3 AL < 20mm A1 = A + 2 AL 20-21 A1 = A + 1 AL 21-22 A1 = A AL 22-24.5 A1 = A – 0.5 AL >24.5
  • 74. SRK/T • It is a nonlinear theoretical optical formula empirically optimized for postop ACD, retinal thickness , corneal refractive index . It combines advantages of theoretical and regression formulas .
  • 75. Generations of IOL Formulas • 1st Generation Fyodorov , Colenbrander ,Hoffer , SRK I • 2nd Generation Binkhorst , SRK II • 3rd Generation Holladay 1 , Hoffer-Q , SRK/T • 4th Generation Holladay 2 , Haigis, Hoffer-H • 5th Generation (2014) Hoffer H-5
  • 79. There are currently three IOL constants in use: The SRK/T formula uses an "A-constant." The Holladay 1 formula uses a "Surgeon Factor." The Holladay 2 formula, and the Hoffer Q formula, both use an "Anterior Chamber Depth." aka: ACD.
  • 80. d = the effective lens position, where ... d = a0 + (a1 * ACD) + (a2 * AL) Haigis Formula
  • 81. * The a0 constant basically moves the curve up, or down, in much the same way that the A- constant, Surgeon Factor, or ACD does for the Holladay 1, Holladay 2, Hoffer Q and SRK/T formulas. * The a1 constant is tied to the measured anterior chamber depth. * The a2 constant is tied to the measured axial length. The way the a0, a1 and a2 constants are derived is by generating a set of surgeon, and IOL-specific
  • 82. AL < 19 mm (<0.1%) Holladay 2 AL 19-22 mm (8%) Holladay 2 , Hoffer-Q AL 22-24.5 mm (72%) SRK II , Hoffer-Q ,Holladay 1 AL 24.5-26 mm (15%) Holladay 1 , Hoffer-Q AL > 26 mm ( 15%) SRK/T Formula Choice
  • 83. Haigis formula may be appropriate for all ranges of axial lengths