SlideShare a Scribd company logo
LINEAR ACCELERATOR
BY- PALLAVI JAIN
MODERATOR- ANKUR MOURYA SIR
HISTORY
• The first linear accelerator principle was invented
by Rolf Wideroe in 1928 to accelerate heavy ions
• In 1952, the first linear accelerator designed for
radiotherapy was installed at Hammersmith
hospital, London
• In 1956, the first patient was treated at Stanford
University in USA
A model of the 1953 8 MeV
linac installation at
Hammersmith Hospital.
LINAC
• Device that uses high-frequency electromagnetic waves to accelerate
charged particles such as electrons to high energies through a linear
tube using accelerator wave guide. It can be made to strike a target to
produce x-rays.
MAJOR COMPONENTS OF LINAC
• Power Supply
• Modulator
• Magnetron
• Electron Gun
• Wave Guide system
• Accelerator Tube
• Treatment Head
• Treatment Couch
• Control Console
A BLOCK DIAGRAM OF TYPICAL MEDICAL LINEAR
ACCELERATOR
TYPE OF SYSTEMS COMPONENTS
ELECTRON EJECTION SYSTEMS ELECTRON GUN
MICROWAVE SYSTEMS MAGNETRON OR KLYSTRON
POWER SUPPLY SYSTEMS MODULATOR CABINET
BEAM TRANSPORT SYSTEMS ACCELERATING WAVE GUIDE
AUXILLARY SYSTEMS VACCUM PUMP
,CIRCULATING COOLING WATER,
RF FREQUENCY TUBE, PRESSURIZED DIELECTRIC
GAS (for RF transmission), RF ISOLATOR OR
THYRATRON
SAFETY INTERLOCK SYSTEMS Both HARDWARE and SOFTWARE
systems
COMPUTER CONTROLLED FEEDBACK SYSTEM MONITOR CHAMBER, hardware position
, limiting micro-switchers
BEAM COLLIMETER/APPLICATOR SYSTEMS JAW COLLIMETERS, MLCS
DRIVE STAND
• It is large rectangular cabinet that is firmly
secured to the floor of the room
• Gantry rotates on horizontal axis in the
drive stand.
• Major components of the drive stand is:
1. Power supply
2. Modulator
3. Cooling system/ Water tank
4. Magnetron or klystron
DRIVE
STAND
POWER SUPPLY
• Provides DC current power to the
modulator which includes pulse
forming network
• High voltage pulses from modulator
section are flat topped DC pulses of
few microseconds in duration.
MODULATOR CABINET
• Contains three major subcomponents-
1. Fan control ( cooling the power distribution systems)
2. Primary power distribution systems (gives power to
machine)
3. Auxiliary power distribution systems. (emergency off
button to shut off power to treatment unit)
• Supplies pulsed power to magnetron and electron
gun when triggered by a timing pulse from control
console.
COOLING SYSTEM/ WATER TANK
• Locted in drive stand
• Provide thermal stability to the system
• Helps in maintaining a constant temperatures so that the components in drive
stand and gantry functions properly.
• Demineralized water is used for cooling.
• Normal water temperature is 20-30 C
• SF6 is used which acts as dielectric and thus prevent sparking.
MAGNETRON
• Device that produces microwaves
• Functions as a high-power oscillator, generating microwave
pulses of 3000MHz
• High voltage short duration pulses in synchrony with
electron gun is supplied to inner circular cathode which
result in thermo-ionic electron emission to outer anode in
presence of perpendicular magnetic field .This produces
microwaves.
Cross-sectional
diagram showing
principle of
magnetron
operation
KLYSTRON
• Klystron is not a generator of microwaves but rather a microwave amplifier
• It needs to be driven by a low-power microwave oscillator
Cross-sectional drawing of a two-cavity klystron
GANTRY
• It can rotate by 360
degree
• It is isocentrically
mounted
• C- Type
• It consist of :
• Electron gun
• Wave guide systems
• Treatment head
• The linac injection system is
the source of electrons,
referred to as the electron
gun
• Produced thermionically
• Injected onto the central
axis of the waveguide.
ELECTRON GUN
ELECTRON GUN
• It consist of:
• Cathode – negatively charged ,1 cm3 of barium
aluminate, produce electron by thermo-ionic
emission
• Gride / gate- switch like mechanism , consist of
copper screen
• Anode – positively charged , attracts electron,
perforated so that some electrons are passed into
accelerating wave guide
Brief description of Linear accelerator machine
WAVE GUIDE SYSTEM
• Accelerating waveguide is obtained from a
cylindrical uniform waveguide by adding a series of
disks (irises) with circular holes at the centre, placed
at equal distances along the tube to form a series of
cavities
• The accelerating waveguide is evacuated to allow
free propagation of electrons.
• Often pressurized with dielectric freon or SF6 to
reduce possibility of electric breakdown and
increase power handling capacity.
TYPES OF ACCELERATING WAVEGUIDE
• Travelling wave structure : structures require a terminating, or
“dummy,” load to absorb the residual power at the end of the
structure, thus preventing a backward reflected wave.
• Standing wave structure: provide maximum reflection of the
waves at both ends of the structure so that the combination of
forward and reverse traveling waves will give rise to stationary
waves, more efficient than the traveling wave designs since axial,
beam transport cavities, and the side cavities can be
independently optimized
TREATMENT HEAD
• Bending magnet
• X-Ray Target – tungsten
• Primary Collimator
• Flattening Filter –to produce a “flat” beam
• Monitor Chambers – ion chamber
• Asymmetrical jaws- X and Y
• MLCs
• Shielding against leakage - tungsten, or lead–tungsten
alloy
• Light localizer system
BENDING MAGNET
• Projects the electron beam onto the
target.
• Produce different beam path for different
energies.
• Present in high energies linacs
• Two common bending magnets afe:
• 90 degree- chromatic deflection
• 270 degree- achromatic deflection
• 112.5 degree- slalom
X RAY TARGET
• Bremsstrahlung x-rays are produced when the electrons are incident on a target of a
high-Z material such as tungsten
• target is water cooled
• target is thick enough to absorb the incident electrons
• the electron energy is converted into a spectrum of x-ray energies
• maximum energy equal to the incident electron energy
• average photon energy of the beam is approximately one-third of the maximum
energy
THE ELECTRON BEAM
• It exits the window of accelerator tube , a narrow pencil of about
2mm in diameter
• This beam instead of striking the target ,is made to strike the on
electron scattering foil to spread the beam as well get a uniform
electron fluence across the treatment field.
Components of treatment head. A: X-ray therapy mode. B: Electron therapy mode
COLLIMATION SYSTEMS
• Limiting the radiation beam
• Beam modifying device
• Primary/fixed collimation
• Secondary/movable collimator
• Tertiary collimator
PRIMARY/FIXED COLLIMATION
• defines a maximum circular field
• one end of conical opening of the
collimator projecting onto edges of the
target and the other end to the flattening
filter
• Depleted uranium/Tungsten
• Transmission <0.2%
FLATTENING FILTER
• Conical metallic filter
• Situated between target and ion chamber
• Produce a uniform intensity distribution
• Usually made up of lead , steel or copper or
tungsten ,aluminum or uranium combinations
• Modifies the narrow non uniform photon beam
at the isocenter into clinically useful flattened
beam with uniformity and symmetry.
IONIZATION CHAMBER
• Transmission ionization
chambers consist of two
separately sealed ionization
chambers with completely
independent biasing power
supplies and readout
electrometers for increased
patient safety
• monitor dose rate, integrated
dose, and field symmetry
• Scattering foil –
• Thin , high Z foil(tungsten or lead or aluminium)
• In electron mode of linac
• Electron beam of 3 mm diameter instead of striking the target ,strike
this foil to produce uniform electron fluence across the treatment
• Carrousel-
• Helps in movement of flattening filters of different energies as well as
scattering foils
SECONDARY/MOVABLE COLLIMATOR
• After passing through the ion chambers, the beam is further collimated by a
continuously movable x-ray collimator
• consists of two pairs of lead or tungsten blocks (jaws) mounted on either side of
the central axis
• provide a rectangular opening from 0 × 0 to the maximum field size (40 × 40 cm2
or a little less) projected at a standard distance such as 100 cm from the x-ray
source
• Transmission <0.5%
MULTILEAF COLLIMATORS
• consists of a large number of collimating blocks or leaves
• driven automatically, independent of each other, to generate a field of
any shape
• consist of 60 to 80 pairs, independently driven
• leaves are made of tungsten alloy (density = 17.0 to 18.5 g/cm3)
• thickness of leaves along the beam direction - 6 cm to 7.5 cm
• primary x-ray transmission through the leaves - < 2%
• interleaf (between sides) transmission - > 3%
• thickness on inner leaves is 0.5 cm - 40 pairs, while its 1cm for the outer 20
pairs
TYPES OF MLC
Siemens - retain upper but
replaces lower Jaw
Elekta - retain lower jaw +
backup collimators, replaces
upper jaw
Varian - retain all jaws + MLC
BEAM LOCALIZATION
• The field size definition is provided by light localizing system in treatment head
• Consists of mirror & light source
• Located in space between chambers & jaws
• Provides an intense light field, duplicating in size & shape the radiation field
• Facilitates positioning of patient’s for treatment
AUXILIARY SERVICE
• Vacuum pumping system producing high vacuum in the accelerating waveguide
• Water cooling system for cooling the accelerating waveguide, target, circulator
and RF generator
• Shielding against leakage radiation produced by target, beam transport system
and RF generator
COUCH/ TREATMENT TABLE
• It is provided to support
the patient to be treated
• It is isocentrically mounted
as the Gantry & Collimator
• Made up of carbon fiber
Dynamic
Floor
In Room
Display
Lasers
Green laser- Silicon diode
2 sagittal , 1 coronal
Tolerance of laser position-
2mm
COUCH MOVEMENTS Can move horizontal,
vertical, lateral and
rotational directions.
HAND PENDANTS
• Contains all the control
switches which can used to
access the movement of
gantry , couch, collimeters ,
jaw, etc
CONTROL CONSOLE
• 4D integral treatment console
• Treatment console/service monitor : Provides status
information on treatment modality, accessories in use,
prescribed dose & dose delivered, interlock status,
emergency off
• ARIA information system : appointment scheduling
• ARIA is a comprehensive information system that
addresses the clinical, administrative, and imaging
needs of the oncology department. It provides a single
user interface that integrates record and verifies the
multileaf collimator, and PortalVision
Service
Monitor
ARIA Information
System
4D ITC
ELECTRONIC PORTAL IMAGING DEVICE(EPID)
• On board imaging
• MV Xray
• Real time imaging
• IGRT
• Flat panel amorphous silicon
• Detectors mounted on retractable arm
• 1024 * 768 resolutions
RADIATION SAFETY /INTERLOCK SYSTEMS
• Last man out(LMOS) switch
• Door interlock
• Beam on-off switch
• Emergency switches
• Safety lights( red, yellow, green)
EMERGENCY BUTTONS
• Total 7 in number
• Completely turn off the entire unit with single switch during emergency
situations.
AUDIO VISUAL MONITORS
TREATMENT PLANNING SYSTEM
After completion of the installation, a formal radiation protection
survey is carried out –
• measurement of head leakage
• area survey
• tests of interlocks
• warning lights
• emergency switches
• There are daily ,monthly and annual quality assessment carried
out for proper functioning of the LINAC.
THANK YOU

More Related Content

PPTX
LINEAR ACCELARATOR
PPTX
Linear accelerator
PPTX
LINAC- LINEAR ACCELERATOR
PPTX
Cobalt & linac
PPTX
Linear Accelerators
PDF
Tomotherapy
PPTX
Radiotherapy machines upto cobalt 60
PPTX
Linac presentation
LINEAR ACCELARATOR
Linear accelerator
LINAC- LINEAR ACCELERATOR
Cobalt & linac
Linear Accelerators
Tomotherapy
Radiotherapy machines upto cobalt 60
Linac presentation

What's hot (20)

PDF
Medical linear accelerator
PPT
Treatment plannings i kiran
PPTX
Basic of medical linac
PDF
Teletherapy Cobalt-60 Machines
PPTX
Linear accelator
PPTX
Physics and Operation of Linear Accelerator
PPTX
Teletherapy equipments and
PPTX
Brachytherapy.ar
PPTX
COMPONENTS OF CO-60 UNIT.pptx
PDF
Linear accelerator amal al-yasiri
PPTX
Percentage Depth Dose In Radiotherapy.pptx
PPT
Clinical radiation generators
PPTX
Linear accelerator
PPT
Beam modification-in-radiotherapy-
PDF
Commissioning of Truebeam LINAC
PPTX
Brachytherapy dosimetry
PDF
Ionization Chambers
PPTX
TRS 398 (Technical Report Series)
Medical linear accelerator
Treatment plannings i kiran
Basic of medical linac
Teletherapy Cobalt-60 Machines
Linear accelator
Physics and Operation of Linear Accelerator
Teletherapy equipments and
Brachytherapy.ar
COMPONENTS OF CO-60 UNIT.pptx
Linear accelerator amal al-yasiri
Percentage Depth Dose In Radiotherapy.pptx
Clinical radiation generators
Linear accelerator
Beam modification-in-radiotherapy-
Commissioning of Truebeam LINAC
Brachytherapy dosimetry
Ionization Chambers
TRS 398 (Technical Report Series)
Ad

Similar to Brief description of Linear accelerator machine (20)

PPTX
Linear Accelerator
PPTX
Linear Accelerators.pptx
PPTX
Linear Accelerators: Principles, Components, Mechanism of Action, and Their V...
PPTX
M LINAC JJ.pptx
PPTX
Linear accelerator.pptx
PPTX
LINEAR ACCELERATOR PRINCIPAL AND WORKING
PDF
Madical liner accelartaer. Morden radiotherapy
PPTX
Clinical Radiation Generators Radiation Physics Dr Syed Aman
PPTX
Radiation Generating equipments of LINAC
PPTX
Linac- A Practical Approach
PPTX
Linear accelerator vinay
PPTX
CLINICAL RADIATION GENERATORS
PPT
Linear Accelerator working and construction
PPTX
Linac
PPTX
Clinical Generators in Radiotherapy by Dr.Avilash.pptx
PPTX
Lec.12 Seminar2023 LINEAR ACCELERATOR.pptx
PDF
[equipment iv] linacc
PPTX
LINAC by Dr. Musaib Mushtaq.pptx
PPTX
linear accelerator
PPTX
Medical Equipment lec radiotherapy
Linear Accelerator
Linear Accelerators.pptx
Linear Accelerators: Principles, Components, Mechanism of Action, and Their V...
M LINAC JJ.pptx
Linear accelerator.pptx
LINEAR ACCELERATOR PRINCIPAL AND WORKING
Madical liner accelartaer. Morden radiotherapy
Clinical Radiation Generators Radiation Physics Dr Syed Aman
Radiation Generating equipments of LINAC
Linac- A Practical Approach
Linear accelerator vinay
CLINICAL RADIATION GENERATORS
Linear Accelerator working and construction
Linac
Clinical Generators in Radiotherapy by Dr.Avilash.pptx
Lec.12 Seminar2023 LINEAR ACCELERATOR.pptx
[equipment iv] linacc
LINAC by Dr. Musaib Mushtaq.pptx
linear accelerator
Medical Equipment lec radiotherapy
Ad

More from Dr. Pallavi Jain (10)

PPTX
External Beam RT IN CArcinoma PROSTATE.pptx
PPTX
LINAC based SRS treatment - Dr Pallavi.pptx
PPTX
role of bevacizumab in radiation induced necrosis _PALLAVI.pptx
PPTX
STUPP TRIAL - treatment of high grade glioma
PPTX
stockholm trial
PPTX
Cyberknife- machine and its clinical applications
PPTX
Pre treatment work-up of carcinoma endometrium
PDF
Lymphatic system
PPTX
Pre management salivary gland by dr pallavi jain
PPTX
Breast screening pallavi
External Beam RT IN CArcinoma PROSTATE.pptx
LINAC based SRS treatment - Dr Pallavi.pptx
role of bevacizumab in radiation induced necrosis _PALLAVI.pptx
STUPP TRIAL - treatment of high grade glioma
stockholm trial
Cyberknife- machine and its clinical applications
Pre treatment work-up of carcinoma endometrium
Lymphatic system
Pre management salivary gland by dr pallavi jain
Breast screening pallavi

Recently uploaded (20)

PPTX
Lesson notes of climatology university.
PDF
Anesthesia in Laparoscopic Surgery in India
PDF
The Lost Whites of Pakistan by Jahanzaib Mughal.pdf
PPTX
PPT- ENG7_QUARTER1_LESSON1_WEEK1. IMAGERY -DESCRIPTIONS pptx.pptx
PDF
Module 4: Burden of Disease Tutorial Slides S2 2025
PDF
Computing-Curriculum for Schools in Ghana
PDF
ANTIBIOTICS.pptx.pdf………………… xxxxxxxxxxxxx
PDF
GENETICS IN BIOLOGY IN SECONDARY LEVEL FORM 3
PPTX
Introduction-to-Literarature-and-Literary-Studies-week-Prelim-coverage.pptx
PPTX
IMMUNITY IMMUNITY refers to protection against infection, and the immune syst...
PDF
Abdominal Access Techniques with Prof. Dr. R K Mishra
PDF
Saundersa Comprehensive Review for the NCLEX-RN Examination.pdf
PPTX
202450812 BayCHI UCSC-SV 20250812 v17.pptx
PPTX
Presentation on HIE in infants and its manifestations
PPTX
1st Inaugural Professorial Lecture held on 19th February 2020 (Governance and...
PPTX
school management -TNTEU- B.Ed., Semester II Unit 1.pptx
PDF
O5-L3 Freight Transport Ops (International) V1.pdf
PDF
Supply Chain Operations Speaking Notes -ICLT Program
PPTX
Cell Types and Its function , kingdom of life
PDF
Chinmaya Tiranga quiz Grand Finale.pdf
Lesson notes of climatology university.
Anesthesia in Laparoscopic Surgery in India
The Lost Whites of Pakistan by Jahanzaib Mughal.pdf
PPT- ENG7_QUARTER1_LESSON1_WEEK1. IMAGERY -DESCRIPTIONS pptx.pptx
Module 4: Burden of Disease Tutorial Slides S2 2025
Computing-Curriculum for Schools in Ghana
ANTIBIOTICS.pptx.pdf………………… xxxxxxxxxxxxx
GENETICS IN BIOLOGY IN SECONDARY LEVEL FORM 3
Introduction-to-Literarature-and-Literary-Studies-week-Prelim-coverage.pptx
IMMUNITY IMMUNITY refers to protection against infection, and the immune syst...
Abdominal Access Techniques with Prof. Dr. R K Mishra
Saundersa Comprehensive Review for the NCLEX-RN Examination.pdf
202450812 BayCHI UCSC-SV 20250812 v17.pptx
Presentation on HIE in infants and its manifestations
1st Inaugural Professorial Lecture held on 19th February 2020 (Governance and...
school management -TNTEU- B.Ed., Semester II Unit 1.pptx
O5-L3 Freight Transport Ops (International) V1.pdf
Supply Chain Operations Speaking Notes -ICLT Program
Cell Types and Its function , kingdom of life
Chinmaya Tiranga quiz Grand Finale.pdf

Brief description of Linear accelerator machine

  • 1. LINEAR ACCELERATOR BY- PALLAVI JAIN MODERATOR- ANKUR MOURYA SIR
  • 2. HISTORY • The first linear accelerator principle was invented by Rolf Wideroe in 1928 to accelerate heavy ions • In 1952, the first linear accelerator designed for radiotherapy was installed at Hammersmith hospital, London • In 1956, the first patient was treated at Stanford University in USA A model of the 1953 8 MeV linac installation at Hammersmith Hospital.
  • 3. LINAC • Device that uses high-frequency electromagnetic waves to accelerate charged particles such as electrons to high energies through a linear tube using accelerator wave guide. It can be made to strike a target to produce x-rays.
  • 4. MAJOR COMPONENTS OF LINAC • Power Supply • Modulator • Magnetron • Electron Gun • Wave Guide system • Accelerator Tube • Treatment Head • Treatment Couch • Control Console
  • 5. A BLOCK DIAGRAM OF TYPICAL MEDICAL LINEAR ACCELERATOR
  • 6. TYPE OF SYSTEMS COMPONENTS ELECTRON EJECTION SYSTEMS ELECTRON GUN MICROWAVE SYSTEMS MAGNETRON OR KLYSTRON POWER SUPPLY SYSTEMS MODULATOR CABINET BEAM TRANSPORT SYSTEMS ACCELERATING WAVE GUIDE AUXILLARY SYSTEMS VACCUM PUMP ,CIRCULATING COOLING WATER, RF FREQUENCY TUBE, PRESSURIZED DIELECTRIC GAS (for RF transmission), RF ISOLATOR OR THYRATRON SAFETY INTERLOCK SYSTEMS Both HARDWARE and SOFTWARE systems COMPUTER CONTROLLED FEEDBACK SYSTEM MONITOR CHAMBER, hardware position , limiting micro-switchers BEAM COLLIMETER/APPLICATOR SYSTEMS JAW COLLIMETERS, MLCS
  • 7. DRIVE STAND • It is large rectangular cabinet that is firmly secured to the floor of the room • Gantry rotates on horizontal axis in the drive stand. • Major components of the drive stand is: 1. Power supply 2. Modulator 3. Cooling system/ Water tank 4. Magnetron or klystron DRIVE STAND
  • 8. POWER SUPPLY • Provides DC current power to the modulator which includes pulse forming network • High voltage pulses from modulator section are flat topped DC pulses of few microseconds in duration.
  • 9. MODULATOR CABINET • Contains three major subcomponents- 1. Fan control ( cooling the power distribution systems) 2. Primary power distribution systems (gives power to machine) 3. Auxiliary power distribution systems. (emergency off button to shut off power to treatment unit) • Supplies pulsed power to magnetron and electron gun when triggered by a timing pulse from control console.
  • 10. COOLING SYSTEM/ WATER TANK • Locted in drive stand • Provide thermal stability to the system • Helps in maintaining a constant temperatures so that the components in drive stand and gantry functions properly. • Demineralized water is used for cooling. • Normal water temperature is 20-30 C • SF6 is used which acts as dielectric and thus prevent sparking.
  • 11. MAGNETRON • Device that produces microwaves • Functions as a high-power oscillator, generating microwave pulses of 3000MHz • High voltage short duration pulses in synchrony with electron gun is supplied to inner circular cathode which result in thermo-ionic electron emission to outer anode in presence of perpendicular magnetic field .This produces microwaves. Cross-sectional diagram showing principle of magnetron operation
  • 12. KLYSTRON • Klystron is not a generator of microwaves but rather a microwave amplifier • It needs to be driven by a low-power microwave oscillator Cross-sectional drawing of a two-cavity klystron
  • 13. GANTRY • It can rotate by 360 degree • It is isocentrically mounted • C- Type • It consist of : • Electron gun • Wave guide systems • Treatment head
  • 14. • The linac injection system is the source of electrons, referred to as the electron gun • Produced thermionically • Injected onto the central axis of the waveguide. ELECTRON GUN
  • 15. ELECTRON GUN • It consist of: • Cathode – negatively charged ,1 cm3 of barium aluminate, produce electron by thermo-ionic emission • Gride / gate- switch like mechanism , consist of copper screen • Anode – positively charged , attracts electron, perforated so that some electrons are passed into accelerating wave guide
  • 17. WAVE GUIDE SYSTEM • Accelerating waveguide is obtained from a cylindrical uniform waveguide by adding a series of disks (irises) with circular holes at the centre, placed at equal distances along the tube to form a series of cavities • The accelerating waveguide is evacuated to allow free propagation of electrons. • Often pressurized with dielectric freon or SF6 to reduce possibility of electric breakdown and increase power handling capacity.
  • 18. TYPES OF ACCELERATING WAVEGUIDE • Travelling wave structure : structures require a terminating, or “dummy,” load to absorb the residual power at the end of the structure, thus preventing a backward reflected wave. • Standing wave structure: provide maximum reflection of the waves at both ends of the structure so that the combination of forward and reverse traveling waves will give rise to stationary waves, more efficient than the traveling wave designs since axial, beam transport cavities, and the side cavities can be independently optimized
  • 19. TREATMENT HEAD • Bending magnet • X-Ray Target – tungsten • Primary Collimator • Flattening Filter –to produce a “flat” beam • Monitor Chambers – ion chamber • Asymmetrical jaws- X and Y • MLCs • Shielding against leakage - tungsten, or lead–tungsten alloy • Light localizer system
  • 20. BENDING MAGNET • Projects the electron beam onto the target. • Produce different beam path for different energies. • Present in high energies linacs • Two common bending magnets afe: • 90 degree- chromatic deflection • 270 degree- achromatic deflection • 112.5 degree- slalom
  • 21. X RAY TARGET • Bremsstrahlung x-rays are produced when the electrons are incident on a target of a high-Z material such as tungsten • target is water cooled • target is thick enough to absorb the incident electrons • the electron energy is converted into a spectrum of x-ray energies • maximum energy equal to the incident electron energy • average photon energy of the beam is approximately one-third of the maximum energy
  • 22. THE ELECTRON BEAM • It exits the window of accelerator tube , a narrow pencil of about 2mm in diameter • This beam instead of striking the target ,is made to strike the on electron scattering foil to spread the beam as well get a uniform electron fluence across the treatment field.
  • 23. Components of treatment head. A: X-ray therapy mode. B: Electron therapy mode
  • 24. COLLIMATION SYSTEMS • Limiting the radiation beam • Beam modifying device • Primary/fixed collimation • Secondary/movable collimator • Tertiary collimator
  • 25. PRIMARY/FIXED COLLIMATION • defines a maximum circular field • one end of conical opening of the collimator projecting onto edges of the target and the other end to the flattening filter • Depleted uranium/Tungsten • Transmission <0.2%
  • 26. FLATTENING FILTER • Conical metallic filter • Situated between target and ion chamber • Produce a uniform intensity distribution • Usually made up of lead , steel or copper or tungsten ,aluminum or uranium combinations • Modifies the narrow non uniform photon beam at the isocenter into clinically useful flattened beam with uniformity and symmetry.
  • 27. IONIZATION CHAMBER • Transmission ionization chambers consist of two separately sealed ionization chambers with completely independent biasing power supplies and readout electrometers for increased patient safety • monitor dose rate, integrated dose, and field symmetry
  • 28. • Scattering foil – • Thin , high Z foil(tungsten or lead or aluminium) • In electron mode of linac • Electron beam of 3 mm diameter instead of striking the target ,strike this foil to produce uniform electron fluence across the treatment • Carrousel- • Helps in movement of flattening filters of different energies as well as scattering foils
  • 29. SECONDARY/MOVABLE COLLIMATOR • After passing through the ion chambers, the beam is further collimated by a continuously movable x-ray collimator • consists of two pairs of lead or tungsten blocks (jaws) mounted on either side of the central axis • provide a rectangular opening from 0 × 0 to the maximum field size (40 × 40 cm2 or a little less) projected at a standard distance such as 100 cm from the x-ray source • Transmission <0.5%
  • 30. MULTILEAF COLLIMATORS • consists of a large number of collimating blocks or leaves • driven automatically, independent of each other, to generate a field of any shape • consist of 60 to 80 pairs, independently driven • leaves are made of tungsten alloy (density = 17.0 to 18.5 g/cm3) • thickness of leaves along the beam direction - 6 cm to 7.5 cm • primary x-ray transmission through the leaves - < 2% • interleaf (between sides) transmission - > 3% • thickness on inner leaves is 0.5 cm - 40 pairs, while its 1cm for the outer 20 pairs
  • 31. TYPES OF MLC Siemens - retain upper but replaces lower Jaw Elekta - retain lower jaw + backup collimators, replaces upper jaw Varian - retain all jaws + MLC
  • 32. BEAM LOCALIZATION • The field size definition is provided by light localizing system in treatment head • Consists of mirror & light source • Located in space between chambers & jaws • Provides an intense light field, duplicating in size & shape the radiation field • Facilitates positioning of patient’s for treatment
  • 33. AUXILIARY SERVICE • Vacuum pumping system producing high vacuum in the accelerating waveguide • Water cooling system for cooling the accelerating waveguide, target, circulator and RF generator • Shielding against leakage radiation produced by target, beam transport system and RF generator
  • 34. COUCH/ TREATMENT TABLE • It is provided to support the patient to be treated • It is isocentrically mounted as the Gantry & Collimator • Made up of carbon fiber
  • 35. Dynamic Floor In Room Display Lasers Green laser- Silicon diode 2 sagittal , 1 coronal Tolerance of laser position- 2mm
  • 36. COUCH MOVEMENTS Can move horizontal, vertical, lateral and rotational directions.
  • 37. HAND PENDANTS • Contains all the control switches which can used to access the movement of gantry , couch, collimeters , jaw, etc
  • 38. CONTROL CONSOLE • 4D integral treatment console • Treatment console/service monitor : Provides status information on treatment modality, accessories in use, prescribed dose & dose delivered, interlock status, emergency off • ARIA information system : appointment scheduling • ARIA is a comprehensive information system that addresses the clinical, administrative, and imaging needs of the oncology department. It provides a single user interface that integrates record and verifies the multileaf collimator, and PortalVision Service Monitor ARIA Information System 4D ITC
  • 39. ELECTRONIC PORTAL IMAGING DEVICE(EPID) • On board imaging • MV Xray • Real time imaging • IGRT • Flat panel amorphous silicon • Detectors mounted on retractable arm • 1024 * 768 resolutions
  • 40. RADIATION SAFETY /INTERLOCK SYSTEMS • Last man out(LMOS) switch • Door interlock • Beam on-off switch • Emergency switches • Safety lights( red, yellow, green)
  • 41. EMERGENCY BUTTONS • Total 7 in number • Completely turn off the entire unit with single switch during emergency situations.
  • 44. After completion of the installation, a formal radiation protection survey is carried out – • measurement of head leakage • area survey • tests of interlocks • warning lights • emergency switches • There are daily ,monthly and annual quality assessment carried out for proper functioning of the LINAC.