HISTORICAL PERSPECTIVE
oG.N Hounsfield (an engineer with
EMI) is credited with the invention of
CT
oAllan MacLeod Cormack of united
state developed solution to the
mathematical problem in CT
oIn 1972, the first head CT scanner
was invented by Goldfrey Hounsfield
AVI00719
GN Hounsfield
2
oin 1972, Hounsfield shared the
Nobel prize in medicine and
physiology with Allan Cormack
AVI00719
Allan Cormack
3
COMPONENTS OF CT
• CT has three major systems
1) Imaging system
a. Gantry b) table
2)Computer system
3)Operating system
IMAGING SYSTEM
A)GANTRY
• A mounted framework that surrounds the patient in a vertical plane
• The gantry includes :-a)slip ring
b) x-ray tube
c)high voltage generator
d)collimator
e)detector
d) data aquisition system
EXTERNAL APPEARANCE OF GANTRY
1.GANTRY APPERTURE
2.MICROPHONE
3.SAGITTAL LASER LIGHT
ALIGNMENT
4.PATIENT GUIDE TABLE
5.X-RAY EXPOSURE INDICATOR
LIGHT
6.EMERGENCY STOP BUTTOM
7.GANTRY CONTROL PANNEL
8.EXTERNAL LASER
ALIGNMENT LIGHT
9.PATIENT COUCH
10.ECG GATING MONITOR
6AVI00719
INTERNAL APPEARANCE OF GANTRY
1.X-RAY TUBE
2.FILTER,COLLIMATOR,
3.INTERNAL PROJECTOR
4.OIL COOLER
5.HIGH VOLTAGE GENERATOR(0-
75kv)
6DIRECT DRIVE GANTRY MOTOR
7.ROTATION CONTROL UNIT
8.DAS
9.DETECTOR
10.SLIP RING
11.DETECTOR COTROL UNIT
12.HVG (75-150 KV)
13.POWER UNIT
14.LINE NOISE FILTER
7AVI00719
Cont……
• Two important features of gantry :-
oGantry aperture :-50cm -90cm
oTilting range :- 30 degree towards or away from the patient table
• Three intense white or low power laser light
SLIP RING TECHNOLOGY
• Introduced in 1990
• Electromechanical device
• Consist of circular electrical conductive rings and brushes
• Slip ring transmit electrical energy across a rotating interface
• It allows continuous gantry rotation
DESIGN OF SLIP RING
• TWO DESIGN
1. DISC DESIGN 2. CYLINDRICAL DESIGN
CYLINDRICAL DESIGN
CT X-RAY TUBE
• Located in the heart of the gantry
• Provide radiation source for CT
• First and second generation scanner used fixed anode , oil cooled x-
ray tube
• Borosilicate used as glass envelope in early CT
• Nowadays metal ceramic x-ray tube is used
Cont…..
• In modern CT x-ray tube
- heat capacity : above 5MHU
- Has large anode with graphite backing
- Utilize a bigger filament
- Anode angle is 7-10 degree
AQUILLION X-RAY TUBE
• High capacity multi slice CT tube
• Heat storage capacity – 7.5 MHU
• Cooling rate – 1.7 MHU/min
• Anode grounded
• Focal spot – 1.4mm x1.4mm
• Air cooled
STRATON X –RAY TUBE
• Zero heat storage capacity
• Cooling rate – 4.7 MHU /min
• Cooled down within 20 sec
• Tube current – 500 mA
• Based on rotating envelope tube
• focal spot – tungsten and rhenium
GENERATOR
• conventional generator for CT scanner were based on the 60 Hz
voltage frequency
• CT scanner now use high frequency generator
• HVG circuit is referred to as high frequency inverter circuit
• Mounted on the rotating frame with the x-ray tube
• Have very low ripple <1 %
FILTRATION IN CT
• Absorbs low energy x-rays
• Reduce scatter and patient dose
• There are mainly two types of filtration utilized in CT
i. Mathematical filter
ii. Inherent filter :-a)flat filter
b)bowtie filter
COLLIMATOR
• TYPES OF COLLIMATOR IN CT SCAN
I. Pre-patient collimator
II. Post- patient collimator
“1D” anti scatter collimator “ 2D” anti scatter collimator
CT DETECTORS
• TYPES OF DETECTOR :-
1. Gas ionization detector
convert x-ray energy directly into electrical signal
2. Scintillation detector
convert x-ray energy into light
1. GAS IONIZATION DETECTOR
• Based on the principal of ionization
• Use high pressure
• Used non-radioactive xenon gas
• Consist of series of individual gas chamber
• QDE :60-90%
AVI00719 20
Charges
SCINTILLATION DETECTOR
• Use a photomultiplier tube
• NaI and CsI were used in early generation of scanners
• NaI was replaced by Bismuth germanate & cadmium tungstate
• Helical & MDCT scanner use combined Cadmium tungstate ,yttrium
or gadolinium ceramic scintillator with a photodetector
• Now a days solid state photodiode multiplier scintillation crystal
detector are used
DETECTOR CHARACTERISTIC
• Efficiency
• Stability
• Response time
• After glow
• Dynamic range
DATA AQUISTISTION SYSTEM
• DETECTOR - ELETRICAL SIGNAL – DAS – COMPUTER - DIGITAL SIGNAL-
ARRAY PROCESSOR (IMAGE GENERATOR OR RECONSTOCTOR
GENERATOR)
PATIENT COUCH/TABLE
• Should be strong and rigid
• Either curved or flat
• Capable of moving up and down
• Constructed of low atomic no. of carbon graphite fiber
COMPUTER SYSTEM
• Uses minicomputer
• Unique components of CT
• Must have sufficient speed and memory
• Calculate the attenuation of the individual voxels using algorithm
• Must be very fast
OPERATOR CONSOLE
• Permits control of all scan parameter
• Commands computer to reconstruct and transfer of image data for
storage in data file
• pre-programed with kvp & mA
• Has three console
SUMMARY
AVI00719 27
Components of ct

More Related Content

PPTX
CT SCAN equipment.pptx
PPTX
Computed Tomography Instrumentation and Detector Configuration
PPTX
CT scanner
PPTX
Computed tomography by jay&jay
PDF
Image reconstrsuction in ct pdf
PPTX
ct image quality
PPT
CT History
PPTX
5lab components of ct scanner
CT SCAN equipment.pptx
Computed Tomography Instrumentation and Detector Configuration
CT scanner
Computed tomography by jay&jay
Image reconstrsuction in ct pdf
ct image quality
CT History
5lab components of ct scanner

What's hot (20)

PPTX
Ct instrument
PPTX
Post processing of computed tomography
PPTX
Advances in ct technology
PPTX
Generations and advancement of CT
PPTX
Computer tomography components
PPTX
Helical and Multislice CT
PPTX
Fluroscopy
PPTX
CT detectors
PPTX
Generations of CT scanner
PPTX
MDCT Principles and Applications- Avinesh Shrestha
PPTX
Ultrasound Transducer Constriction And It’s Physics.pptx
PPTX
Gamma Camera
PDF
4-PLANNING-OF-DIAGNOSTIC-X-RAY-INSTALLATIONS.pdf
PPTX
mr angiography.pptx
PPTX
CT image quality and image resolution.pptx
PPTX
Components of CT Scan Machine
PPTX
Mri gradient coils
PPTX
Ct quality control
PPTX
The control of scattered radiation
Ct instrument
Post processing of computed tomography
Advances in ct technology
Generations and advancement of CT
Computer tomography components
Helical and Multislice CT
Fluroscopy
CT detectors
Generations of CT scanner
MDCT Principles and Applications- Avinesh Shrestha
Ultrasound Transducer Constriction And It’s Physics.pptx
Gamma Camera
4-PLANNING-OF-DIAGNOSTIC-X-RAY-INSTALLATIONS.pdf
mr angiography.pptx
CT image quality and image resolution.pptx
Components of CT Scan Machine
Mri gradient coils
Ct quality control
The control of scattered radiation
Ad

Similar to Components of ct (20)

PPT
Xexh4 wzswh6ucn7f5hzc6ytgnibrrc
PPTX
INTRODUCTION OF CT , VARIOUS GENERATIONS .pptx
PPTX
Physical principle of Computed Tomography (Scanning principle & Data acquisit...
PPTX
Ct scan final (2)
PPTX
CT INSTRUMENTATIONSsnsjsjsjwjwjnwjajaj.pptx
PPTX
Multi detector CT AND ITS ADVANCEMENTS by joginder.pptx
PPTX
Ct instrumentation and types of detector configuration
PPTX
CT Generations and Artefacts
PPTX
Computed Tomography.pptx
PPT
Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOU...
PDF
BASICS OF WORKING OF COMPUTER TOMOGRAPHY
PPTX
Computer Tomography
PPTX
ct physivvbbbghdhdjrjrnnrntbjfndncs.pptx
PPTX
Basics of CT & MRI
PPTX
Advanced technique & Instrumentation of.pptx
PPTX
Computed tomography
PPT
4 CT instrumentation Radiation CT scan Lecture
PPT
Computed tomography - Physics and Clinical
PPTX
Ct Scan my sb dbdhdb dd. dd d d s s d e e .pptx
PPTX
CT scanner
Xexh4 wzswh6ucn7f5hzc6ytgnibrrc
INTRODUCTION OF CT , VARIOUS GENERATIONS .pptx
Physical principle of Computed Tomography (Scanning principle & Data acquisit...
Ct scan final (2)
CT INSTRUMENTATIONSsnsjsjsjwjwjnwjajaj.pptx
Multi detector CT AND ITS ADVANCEMENTS by joginder.pptx
Ct instrumentation and types of detector configuration
CT Generations and Artefacts
Computed Tomography.pptx
Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOU...
BASICS OF WORKING OF COMPUTER TOMOGRAPHY
Computer Tomography
ct physivvbbbghdhdjrjrnnrntbjfndncs.pptx
Basics of CT & MRI
Advanced technique & Instrumentation of.pptx
Computed tomography
4 CT instrumentation Radiation CT scan Lecture
Computed tomography - Physics and Clinical
Ct Scan my sb dbdhdb dd. dd d d s s d e e .pptx
CT scanner
Ad

Recently uploaded (20)

PDF
Saundersa Comprehensive Review for the NCLEX-RN Examination.pdf
PDF
Anesthesia in Laparoscopic Surgery in India
PDF
Computing-Curriculum for Schools in Ghana
PDF
Basic Mud Logging Guide for educational purpose
PPTX
Final Presentation General Medicine 03-08-2024.pptx
PPTX
Renaissance Architecture: A Journey from Faith to Humanism
PPTX
IMMUNITY IMMUNITY refers to protection against infection, and the immune syst...
PDF
RMMM.pdf make it easy to upload and study
PDF
Physiotherapy_for_Respiratory_and_Cardiac_Problems WEBBER.pdf
PPTX
PPH.pptx obstetrics and gynecology in nursing
PPTX
master seminar digital applications in india
PPTX
GDM (1) (1).pptx small presentation for students
PPTX
Microbial diseases, their pathogenesis and prophylaxis
PDF
Module 4: Burden of Disease Tutorial Slides S2 2025
PDF
O7-L3 Supply Chain Operations - ICLT Program
PPTX
Lesson notes of climatology university.
PPTX
Institutional Correction lecture only . . .
PDF
Microbial disease of the cardiovascular and lymphatic systems
PPTX
1st Inaugural Professorial Lecture held on 19th February 2020 (Governance and...
PDF
2.FourierTransform-ShortQuestionswithAnswers.pdf
Saundersa Comprehensive Review for the NCLEX-RN Examination.pdf
Anesthesia in Laparoscopic Surgery in India
Computing-Curriculum for Schools in Ghana
Basic Mud Logging Guide for educational purpose
Final Presentation General Medicine 03-08-2024.pptx
Renaissance Architecture: A Journey from Faith to Humanism
IMMUNITY IMMUNITY refers to protection against infection, and the immune syst...
RMMM.pdf make it easy to upload and study
Physiotherapy_for_Respiratory_and_Cardiac_Problems WEBBER.pdf
PPH.pptx obstetrics and gynecology in nursing
master seminar digital applications in india
GDM (1) (1).pptx small presentation for students
Microbial diseases, their pathogenesis and prophylaxis
Module 4: Burden of Disease Tutorial Slides S2 2025
O7-L3 Supply Chain Operations - ICLT Program
Lesson notes of climatology university.
Institutional Correction lecture only . . .
Microbial disease of the cardiovascular and lymphatic systems
1st Inaugural Professorial Lecture held on 19th February 2020 (Governance and...
2.FourierTransform-ShortQuestionswithAnswers.pdf

Components of ct

  • 1. BY MUNA BUDHATHOKI B.SC. MIT 2ND YEAR CMCTH,CHITWAN
  • 2. HISTORICAL PERSPECTIVE oG.N Hounsfield (an engineer with EMI) is credited with the invention of CT oAllan MacLeod Cormack of united state developed solution to the mathematical problem in CT oIn 1972, the first head CT scanner was invented by Goldfrey Hounsfield AVI00719 GN Hounsfield 2
  • 3. oin 1972, Hounsfield shared the Nobel prize in medicine and physiology with Allan Cormack AVI00719 Allan Cormack 3
  • 4. COMPONENTS OF CT • CT has three major systems 1) Imaging system a. Gantry b) table 2)Computer system 3)Operating system
  • 5. IMAGING SYSTEM A)GANTRY • A mounted framework that surrounds the patient in a vertical plane • The gantry includes :-a)slip ring b) x-ray tube c)high voltage generator d)collimator e)detector d) data aquisition system
  • 6. EXTERNAL APPEARANCE OF GANTRY 1.GANTRY APPERTURE 2.MICROPHONE 3.SAGITTAL LASER LIGHT ALIGNMENT 4.PATIENT GUIDE TABLE 5.X-RAY EXPOSURE INDICATOR LIGHT 6.EMERGENCY STOP BUTTOM 7.GANTRY CONTROL PANNEL 8.EXTERNAL LASER ALIGNMENT LIGHT 9.PATIENT COUCH 10.ECG GATING MONITOR 6AVI00719
  • 7. INTERNAL APPEARANCE OF GANTRY 1.X-RAY TUBE 2.FILTER,COLLIMATOR, 3.INTERNAL PROJECTOR 4.OIL COOLER 5.HIGH VOLTAGE GENERATOR(0- 75kv) 6DIRECT DRIVE GANTRY MOTOR 7.ROTATION CONTROL UNIT 8.DAS 9.DETECTOR 10.SLIP RING 11.DETECTOR COTROL UNIT 12.HVG (75-150 KV) 13.POWER UNIT 14.LINE NOISE FILTER 7AVI00719
  • 8. Cont…… • Two important features of gantry :- oGantry aperture :-50cm -90cm oTilting range :- 30 degree towards or away from the patient table • Three intense white or low power laser light
  • 9. SLIP RING TECHNOLOGY • Introduced in 1990 • Electromechanical device • Consist of circular electrical conductive rings and brushes • Slip ring transmit electrical energy across a rotating interface • It allows continuous gantry rotation
  • 10. DESIGN OF SLIP RING • TWO DESIGN 1. DISC DESIGN 2. CYLINDRICAL DESIGN CYLINDRICAL DESIGN
  • 11. CT X-RAY TUBE • Located in the heart of the gantry • Provide radiation source for CT • First and second generation scanner used fixed anode , oil cooled x- ray tube • Borosilicate used as glass envelope in early CT • Nowadays metal ceramic x-ray tube is used
  • 12. Cont….. • In modern CT x-ray tube - heat capacity : above 5MHU - Has large anode with graphite backing - Utilize a bigger filament - Anode angle is 7-10 degree
  • 13. AQUILLION X-RAY TUBE • High capacity multi slice CT tube • Heat storage capacity – 7.5 MHU • Cooling rate – 1.7 MHU/min • Anode grounded • Focal spot – 1.4mm x1.4mm • Air cooled
  • 14. STRATON X –RAY TUBE • Zero heat storage capacity • Cooling rate – 4.7 MHU /min • Cooled down within 20 sec • Tube current – 500 mA • Based on rotating envelope tube • focal spot – tungsten and rhenium
  • 15. GENERATOR • conventional generator for CT scanner were based on the 60 Hz voltage frequency • CT scanner now use high frequency generator • HVG circuit is referred to as high frequency inverter circuit • Mounted on the rotating frame with the x-ray tube • Have very low ripple <1 %
  • 16. FILTRATION IN CT • Absorbs low energy x-rays • Reduce scatter and patient dose • There are mainly two types of filtration utilized in CT i. Mathematical filter ii. Inherent filter :-a)flat filter b)bowtie filter
  • 17. COLLIMATOR • TYPES OF COLLIMATOR IN CT SCAN I. Pre-patient collimator II. Post- patient collimator “1D” anti scatter collimator “ 2D” anti scatter collimator
  • 18. CT DETECTORS • TYPES OF DETECTOR :- 1. Gas ionization detector convert x-ray energy directly into electrical signal 2. Scintillation detector convert x-ray energy into light
  • 19. 1. GAS IONIZATION DETECTOR • Based on the principal of ionization • Use high pressure • Used non-radioactive xenon gas • Consist of series of individual gas chamber • QDE :60-90%
  • 21. SCINTILLATION DETECTOR • Use a photomultiplier tube • NaI and CsI were used in early generation of scanners • NaI was replaced by Bismuth germanate & cadmium tungstate • Helical & MDCT scanner use combined Cadmium tungstate ,yttrium or gadolinium ceramic scintillator with a photodetector • Now a days solid state photodiode multiplier scintillation crystal detector are used
  • 22. DETECTOR CHARACTERISTIC • Efficiency • Stability • Response time • After glow • Dynamic range
  • 23. DATA AQUISTISTION SYSTEM • DETECTOR - ELETRICAL SIGNAL – DAS – COMPUTER - DIGITAL SIGNAL- ARRAY PROCESSOR (IMAGE GENERATOR OR RECONSTOCTOR GENERATOR)
  • 24. PATIENT COUCH/TABLE • Should be strong and rigid • Either curved or flat • Capable of moving up and down • Constructed of low atomic no. of carbon graphite fiber
  • 25. COMPUTER SYSTEM • Uses minicomputer • Unique components of CT • Must have sufficient speed and memory • Calculate the attenuation of the individual voxels using algorithm • Must be very fast
  • 26. OPERATOR CONSOLE • Permits control of all scan parameter • Commands computer to reconstruct and transfer of image data for storage in data file • pre-programed with kvp & mA • Has three console

Editor's Notes