Quality Control
For CT Scanners
SAJID EJAZ
DEPARTMENT OF
RADIOLOGY
Quality Control in CT
 A good idea? Yes
 Required for accreditation? Sometimes
 Improves image quality? Sometimes
 Depends on starting point
 Reduces downtime? Sometimes
 Depends on starting point
 Possible recognition of problems before
 clinical manifestations
 total failure
 may reduce unscheduled downtime
High Contrast Resolution
 Baseline established when scanner in good operati
ng condition
 Subsequent test compared to baseline
CT Quality Control
 Standard protocol required
 same kVp, mAs, scan diameter, slice width, algorith
m, focal spot, filter, etc
 When performed
 At regular intervals
 After major service
 tube change
CT # of Water
 What do we check?
 Average (ideally 0 by definition)
 Standard Deviation (ideally 0)
 Flatness
 ROI’s at various locations in phantom
 Results depend upon protocol
 kVp, mA, scan time
 slice width
 phantom size
 ROI position
 Problem causes
 Mis-calibration
 Noise / drift in detector system
CT # of Water
 Can repeat for various
 Slice thicknesses
 Reconstruction algorithms
 Centering positions
 Phantom diameters
 Fields of view
CT Number Accuracy
 Phantom with several materials of known CT #’s
 Do ROI’s at each and check CT # accuracy
High Contrast (or Limiting) Reso
lution
 Phantom
 rows of equally spaced high contrast objects
 holes get smaller from row to row
 Bars of decreasing size & separation
 Determine smallest row of holes where all
holes can be clearly seen
 smaller holes = better resolution
 Visual scoring
Low Contrast Resolution Phanto
m
 holes get smaller from column to column
 Holes have less contrast from row to row
 Visual scoring
Low Contrast Resolution
 Noise limited
 less noise = more rows visualized
 Standardize
 mAs
 more mAs  less noise  better low co
ntrast resolution
 reconstruction algorithm / filter
 smoothing reduces noise
 improves low contrast resolution
Low Contrast Resolution
 Possible sources of failure: anything that can increa
se noise
 decreased dose to detectors
 reduced tube output
 electronic noise
 detectors
 amplifiers
 A - D convertor
Distance-Measuring Accuracy
 Phantom object with prec
isely known object distan
ces
 Use distance-measuring s
oftware
 Check vertical, horizontal
& diagonal
 Check monitor & hard co
py
Monitor Performance / Hard Copy Output
 Standard gray scale ima
ge needed
 computer generated (S
MPTE pattern)
 Check monitor & hard c
opy
 5% patches
 Contrast patches
 Resolution bars
 Hard copy
 Measure optical densit
y of steps with densito
meter
Slice Localization / Thickness Accuracy
 Phantom with
 slanted wire
 Measure length
 Regularly spaced (in z direction) obj
ects
 count
 Align phantom slice with laser
 Make single scan with known slice t
hickness
Slice Localization / Thickness Accuracy
Table Indexing
 X-ray cassette on table
 Load table with weight
 Select
 multiple scans
 small beam width
 large mm table increment
Table Indexing
 Image should show series of exposed b
ands
 Measure distance between bands
 Possible causes of failure
 slippage of table drive mechanism
 table mis-calibration
Table Backlash
 Does table return to same position
from both directions?
 Test
 Apply masking tape to table
 Load table with weight
 Note numerical position
 Mark tape at starting position usin
g laser position
 Move table away from starting posit
ion & then back to same numerical
location
 Check laser vs. mark on tape
Laser Accuracy
 Align phantom on table m
atching outer laser to exter
nal marks on phantom
 “Zero” table
 Scan phantom
 “0” z-location should sho
w correct phantom position
on image
 Repeat for internal laser
CT Noise Characteristics
 Water phantom
 Multiple scans changing only mAs
 Measure standard deviation of CT #’s using ident
ical ROI’s
 Noise proportional to standard deviation
CT Noise Characteristics
 Excessive noise can be caused by
 detector problems
 electronic noise in detector amplifier circuits
 reduced output per mAs
Ct quality control

More Related Content

PPTX
Quality Assurance Programme in Computed Tomography
PPT
QC of CT using Catphan 600
PPTX
Ct image artifacts
PPT
CTDI (Computed Tomography Dose Index
PPTX
Ct image quality artifacts and it remedy
PPT
Quality Management for Diagnostic Imaging
PPTX
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
PPTX
Image Reconstruction in Computed Tomography
Quality Assurance Programme in Computed Tomography
QC of CT using Catphan 600
Ct image artifacts
CTDI (Computed Tomography Dose Index
Ct image quality artifacts and it remedy
Quality Management for Diagnostic Imaging
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
Image Reconstruction in Computed Tomography

What's hot (20)

PPTX
Computed Tomography Dose Index
PDF
PET-MRI BASIC PRINCIPAL
PPTX
Encoding and image formation
PPTX
Post processing of computed tomography
PPT
Radiation protection in nuclear medicine shafiee
PPTX
Slip ring
PPTX
CT Procedure OF Abdomen & Pelvis
PPTX
Dose reduction technique in ct scan
PPTX
CT scanner
PPTX
Tools in interventional radiology
PPTX
MDCT (2)
PPTX
CT Guided Biopsy.pptx
PPTX
Basic principle of gamma camera
PPTX
Mammography physics and technique
PPTX
MDCT Principles and Applications- Avinesh Shrestha
PPTX
TOMOSYNTHESIS.pptx
PDF
Chapter4 radiation monitoring instruments
PPT
Operative Cholangiogram
PPT
radiation protection in ct
Computed Tomography Dose Index
PET-MRI BASIC PRINCIPAL
Encoding and image formation
Post processing of computed tomography
Radiation protection in nuclear medicine shafiee
Slip ring
CT Procedure OF Abdomen & Pelvis
Dose reduction technique in ct scan
CT scanner
Tools in interventional radiology
MDCT (2)
CT Guided Biopsy.pptx
Basic principle of gamma camera
Mammography physics and technique
MDCT Principles and Applications- Avinesh Shrestha
TOMOSYNTHESIS.pptx
Chapter4 radiation monitoring instruments
Operative Cholangiogram
radiation protection in ct
Ad

Similar to Ct quality control (20)

PPTX
Chapter 6 image quality in ct
PPTX
CT based Image Guided Radiotherapy - Physics & QA
PPTX
06.CT6.pptx the best lecturer in CT technical
PPT
ct 6 chapter compmd tomography chapter 6
PPT
QC of gamma camera using different phantoms and their specification
PPT
Qa of ct and mamography
PDF
computedtomography-170122041110.pdf
PPTX
Computed tomography
PDF
Quality Control in MRI- Avinesh Shrestha
PPTX
PPT
Anjali (qa qc on cr dr)
PPTX
Image Quality And Artifacts in Computed Tomography.pptx
PPT
Quality assurance in ultrasound machines
PPTX
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
PPT
Physics of Multidetector CT Scan
PPTX
Computed tomography
PPT
QA QC CR Pak Bagus.ppt
PPTX
Ct ppt 2
PPT
14184864.ppt
PPTX
Physics_of_CT , CT machine and it’s parts, ct generations
Chapter 6 image quality in ct
CT based Image Guided Radiotherapy - Physics & QA
06.CT6.pptx the best lecturer in CT technical
ct 6 chapter compmd tomography chapter 6
QC of gamma camera using different phantoms and their specification
Qa of ct and mamography
computedtomography-170122041110.pdf
Computed tomography
Quality Control in MRI- Avinesh Shrestha
Anjali (qa qc on cr dr)
Image Quality And Artifacts in Computed Tomography.pptx
Quality assurance in ultrasound machines
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
Physics of Multidetector CT Scan
Computed tomography
QA QC CR Pak Bagus.ppt
Ct ppt 2
14184864.ppt
Physics_of_CT , CT machine and it’s parts, ct generations
Ad

Recently uploaded (20)

PDF
SEMEN PREPARATION TECHNIGUES FOR INTRAUTERINE INSEMINATION.pdf
PPTX
Neonate anatomy and physiology presentation
PPTX
preoerative assessment in anesthesia and critical care medicine
PPTX
Radiation Dose Management for Patients in Medical Imaging- Avinesh Shrestha
PDF
focused on the development and application of glycoHILIC, pepHILIC, and comm...
PPTX
Effects of lipid metabolism 22 asfelagi.pptx
PPTX
NUCLEAR-MEDICINE-Copy.pptxbabaabahahahaahha
PDF
Lecture on Anesthesia for ENT surgery 2025pptx.pdf
PPTX
Electrolyte Disturbance in Paediatric - Nitthi.pptx
PDF
The Digestive System Science Educational Presentation in Dark Orange, Blue, a...
PDF
Calcified coronary lesions management tips and tricks
PPTX
09. Diabetes in Pregnancy/ gestational.pptx
PPTX
HYPERSENSITIVITY REACTIONS - Pathophysiology Notes for Second Year Pharm D St...
PPTX
Post Op complications in general surgery
PPTX
y4d nutrition and diet in pregnancy and postpartum
PPTX
Acute Coronary Syndrome for Cardiology Conference
PPT
nephrology MRCP - Member of Royal College of Physicians ppt
PPTX
IMAGING EQUIPMENiiiiìiiiiiTpptxeiuueueur
PDF
The_EHRA_Book_of_Interventional Electrophysiology.pdf
PPTX
Manage HIV exposed child and a child with HIV infection.pptx
SEMEN PREPARATION TECHNIGUES FOR INTRAUTERINE INSEMINATION.pdf
Neonate anatomy and physiology presentation
preoerative assessment in anesthesia and critical care medicine
Radiation Dose Management for Patients in Medical Imaging- Avinesh Shrestha
focused on the development and application of glycoHILIC, pepHILIC, and comm...
Effects of lipid metabolism 22 asfelagi.pptx
NUCLEAR-MEDICINE-Copy.pptxbabaabahahahaahha
Lecture on Anesthesia for ENT surgery 2025pptx.pdf
Electrolyte Disturbance in Paediatric - Nitthi.pptx
The Digestive System Science Educational Presentation in Dark Orange, Blue, a...
Calcified coronary lesions management tips and tricks
09. Diabetes in Pregnancy/ gestational.pptx
HYPERSENSITIVITY REACTIONS - Pathophysiology Notes for Second Year Pharm D St...
Post Op complications in general surgery
y4d nutrition and diet in pregnancy and postpartum
Acute Coronary Syndrome for Cardiology Conference
nephrology MRCP - Member of Royal College of Physicians ppt
IMAGING EQUIPMENiiiiìiiiiiTpptxeiuueueur
The_EHRA_Book_of_Interventional Electrophysiology.pdf
Manage HIV exposed child and a child with HIV infection.pptx

Ct quality control

  • 1. Quality Control For CT Scanners SAJID EJAZ DEPARTMENT OF RADIOLOGY
  • 2. Quality Control in CT  A good idea? Yes  Required for accreditation? Sometimes  Improves image quality? Sometimes  Depends on starting point  Reduces downtime? Sometimes  Depends on starting point  Possible recognition of problems before  clinical manifestations  total failure  may reduce unscheduled downtime
  • 3. High Contrast Resolution  Baseline established when scanner in good operati ng condition  Subsequent test compared to baseline
  • 4. CT Quality Control  Standard protocol required  same kVp, mAs, scan diameter, slice width, algorith m, focal spot, filter, etc  When performed  At regular intervals  After major service  tube change
  • 5. CT # of Water  What do we check?  Average (ideally 0 by definition)  Standard Deviation (ideally 0)  Flatness  ROI’s at various locations in phantom  Results depend upon protocol  kVp, mA, scan time  slice width  phantom size  ROI position  Problem causes  Mis-calibration  Noise / drift in detector system
  • 6. CT # of Water  Can repeat for various  Slice thicknesses  Reconstruction algorithms  Centering positions  Phantom diameters  Fields of view
  • 7. CT Number Accuracy  Phantom with several materials of known CT #’s  Do ROI’s at each and check CT # accuracy
  • 8. High Contrast (or Limiting) Reso lution  Phantom  rows of equally spaced high contrast objects  holes get smaller from row to row  Bars of decreasing size & separation  Determine smallest row of holes where all holes can be clearly seen  smaller holes = better resolution  Visual scoring
  • 9. Low Contrast Resolution Phanto m  holes get smaller from column to column  Holes have less contrast from row to row  Visual scoring
  • 10. Low Contrast Resolution  Noise limited  less noise = more rows visualized  Standardize  mAs  more mAs  less noise  better low co ntrast resolution  reconstruction algorithm / filter  smoothing reduces noise  improves low contrast resolution
  • 11. Low Contrast Resolution  Possible sources of failure: anything that can increa se noise  decreased dose to detectors  reduced tube output  electronic noise  detectors  amplifiers  A - D convertor
  • 12. Distance-Measuring Accuracy  Phantom object with prec isely known object distan ces  Use distance-measuring s oftware  Check vertical, horizontal & diagonal  Check monitor & hard co py
  • 13. Monitor Performance / Hard Copy Output  Standard gray scale ima ge needed  computer generated (S MPTE pattern)  Check monitor & hard c opy  5% patches  Contrast patches  Resolution bars  Hard copy  Measure optical densit y of steps with densito meter
  • 14. Slice Localization / Thickness Accuracy  Phantom with  slanted wire  Measure length  Regularly spaced (in z direction) obj ects  count  Align phantom slice with laser  Make single scan with known slice t hickness
  • 15. Slice Localization / Thickness Accuracy
  • 16. Table Indexing  X-ray cassette on table  Load table with weight  Select  multiple scans  small beam width  large mm table increment
  • 17. Table Indexing  Image should show series of exposed b ands  Measure distance between bands  Possible causes of failure  slippage of table drive mechanism  table mis-calibration
  • 18. Table Backlash  Does table return to same position from both directions?  Test  Apply masking tape to table  Load table with weight  Note numerical position  Mark tape at starting position usin g laser position  Move table away from starting posit ion & then back to same numerical location  Check laser vs. mark on tape
  • 19. Laser Accuracy  Align phantom on table m atching outer laser to exter nal marks on phantom  “Zero” table  Scan phantom  “0” z-location should sho w correct phantom position on image  Repeat for internal laser
  • 20. CT Noise Characteristics  Water phantom  Multiple scans changing only mAs  Measure standard deviation of CT #’s using ident ical ROI’s  Noise proportional to standard deviation
  • 21. CT Noise Characteristics  Excessive noise can be caused by  detector problems  electronic noise in detector amplifier circuits  reduced output per mAs