Quality Assurance Programme in Computed Tomography
Outline
 Introduction to Computed Tomography
 Component of CT System
 CT Data Acquisition System
 Quality Assurance in CT
 QA/QC Test in CT
 Conclusion
 References
Computed Tomography
Is a medical imaging procedure of creating
a cross-sectional tomographic plane of any
part of the body .
CT scanners create cross-sectional images of high radiographic contrast
The dose to the patient may be significantly higher than with
alternative imaging modalities
Computed Tomography
Component of CT System
Computer (Operating Console)
Gantry and Table
X-ray Tube
Filter
Detector
Computed Tomography
Computer
Provides the link between the CT technologist and the other
components of the system.
Control of data acquisition,
Image reconstruction,
Storage of image data, and
Image display
Computed Tomography
Gantry and Table
A circular device that house the;
 The X-ray tube
The Detector Array
Data Acquisition System
The Generator
Computed Tomography
Gantry and Table continue…
The table is a flat or slightly curved platform that the patient lies on
during the scan.
The table is responsible for positioning the patient with the gantry
The patient couch on the table must be constructed of low Z-material,
such as carbon fiber to not interfere with x-ray beam transmission
and patient imaging.
Computed Tomography
X-ray Tube
They have an anode heat storage
capacity of 4-5 MHU (million heat
unit) or more.
CT x-ray tube use a rotating anode to
increase heat dissipation
COMPUTED TOMOGRAPHY
Detector
The detectors in CT function as image receptors.
It measures the amount of radiation transmitted
through the body and converts the measurement into
an electrical signal.
The two basic detector types used in CT are scintillation
(solid state) and ionization (xenon gas) detectors
Detector
Rings
Computed Tomography
Data Acquisition System
X-ray photons come on the detector.
The detector detects the intensity in form of current.
The current is converted into voltage.
The analog signal is converted into digital form.
This signal can now be processed and reconstructed in the computer
Detector
Current to
Voltage
Convertor
Pre
Amplifier
Analog to
Digital
Convertor
Computer
Quality Assurance
Quality assurance (QA) is a program used by management to maintain
optimal diagnostic image quality with minimum hazard and distress to
patients.
The program includes;
Periodic quality control tests,
Administrative Procedure
Corrective action and
Training
QA/QC Test In CT
Field Uniformity Test
Data Analysis
Daily Test Objective
Equipment Warm-up Test To ensure equipment is working properly
CT Number for Water To ensure consistency for CT calibrated number for water
Field Uniformity Test To evaluate the scan plane uniformity
QA/QC Test In CT
Monthly Test Objective
CT Light Alignment Test To ensure that the internal and external alignment lights are properly aligned
with the tomographic plane and the couch
SPR Accuracy To ensure that the SPR image accurately indicates the patient position
CT Number Accuracy, Image
Noise, Uniformity And
Image Artifacts
To ensure a) CT number accuracy; b) Noise levels of CT images being within the
tolerances with respect to their reference values; c) Uniformity of CT number
and image noise across the FOV; d) Minimal image artefacts
Image Display and Printing To confirm that the image displays are reproducing all of the grey scale
information in the image accurately and that the printed images are a true
replica of the image display
Low Contrast Resolution
Test
To determines the capability of the scanner to discriminate low contrast objects
High Contrast Spatial
Resolution
To determines the high contrast spatial frequency limits of the CT scanner under
various conditions
Scan Increment Accuracy To determines the deviation of indicated versus actual scan increment
QA/QC Test In CT
Annual Test Objective
Visual Inspection and
Programme Review
1. To ensure that the CT scanner and adjacent areas are safe and have the necessary
radiation protection supplies
2. To ensure that CT QC test are being conducted at appropriate time periods and
corrective actions are taken when necessary
X-ray Generator If the patient dose is excessively high, then kV and HVL should be investigated to
determine the cause of high dose
Radiation Dose 1. To Compare the determined CVOL (CTDVOL) with the manufacturer’s specification.
2. To determine the CVOL (CTDIVOL) and CT (DLP) values for selected protocols
representing the main examinations clinically performed at the facility
3. To Review and update the CVOL (CTDIVOL) and CT (DLP) values for selected
protocols representing the main examinations clinically performed at the facility
X-ray Beam Width To determine the accuracy of collimator settings
Spatial Resolution To ensure that the spatial resolution of a reconstructed image complies with
manufacture standards
QA/QC Test In CT
Software
Spatial/geometry accuracy tests
oCorrect data transfer
Image Registration
Image reconstruction
oTests should verify that the software accurately reconstructs known
phantoms
QA/QC CT Test Phantom
Line Pair High Contrast Phantom Siemens Phantom
CT number regions in RMI
ACR accreditation phantom
RMI Model 464
Head and body dosimetry
phantoms
Conclusion
The QA program is designed to improve accuracy of patient
treatments and efficiency of the treatment planning process
while keeping dose ALARA
QA program should evolve and adapt as the device used for
CT simulation process change
References
Quality Assurance Programme for Computed Tomography; Diagnostic and
Therapy Application, International Atomic Energy Agency IAEA, Human Health
Series No. 19, 2012, ISBN 978–92–0–128910–0
Quality assurance workbook for radiographers & radiological technologists, World
Health Organization, 2001, ISBN 9241546425
American Association of Physicists in Medicine. Quality Assurance for Computed-
Tomography Simulators and the Computed Tomography- Simulation Process:
Report of the AAPM Radiation Therapy Committee Task Group No. 66.,2003.
Quality Control Recommendations for Diagnostic Radiography Conference of
Radiation Control Programme Directors Inc. CRCPD, Volume 2, 2001
References
American College of Radiology, Computed Tomography Quality Control Manual,
2012
Smith Steven W., The Scientist and Engineer’s Guide to Digital Signal Processing,
1997-1998
Phillip W. Ballinger, Eugene D. Frank, volume 3: Merrill's Atlas of Radiographic
Positions and Radiologic Procedures, Edition 10, ISBN (Set) 0-323-01604-9
Jerrold T. Bushberg, J. Anthony Seibert, Edwin M. Leidholdt Jr., John N. Boone,
The Essential Physic of Medical Imaging, second edition, ISBN 0-683-30118-7
Kak A.C., Slaney M., Principles of Computed Tomographic Imaging. SIAM, 1987.
Flohr T.G., Schaller S., Stierstorfer K., et al., "Multi-Detector Row CT Systems and
Image-Reconstruction Techniques," Radiology, June 2005.
The End ???..... I Don’t Think So!!!

More Related Content

PPTX
Quality control
PPTX
Quality assurance
PPTX
Ct quality control
PPTX
Qc in xray
PPTX
Quality assurance
PPTX
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
PPT
Quality Management for Diagnostic Imaging
PPTX
QUALITY CONTROL PDF.pptx
Quality control
Quality assurance
Ct quality control
Qc in xray
Quality assurance
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
Quality Management for Diagnostic Imaging
QUALITY CONTROL PDF.pptx

What's hot (20)

PPTX
Ct protocols of thorax
PPT
Flouroscopic imging
PPTX
MDCT (2)
PPTX
CT artifact
PDF
Quality Control in MRI- Avinesh Shrestha
PPTX
Mri artifacts
PPTX
Generations of CT scanner
PPTX
MACRORADIOGRAPHY.pptx
PPTX
Magnification(macro and micro radiography), distortion
PPTX
DIGITAL FLUOROSCOPY DMR.pptx
PPT
MAMMOGRAPHY VIEWS .ppt
PPT
Basics in ct
PPTX
ICRP-International commission on Radiation Protection
PPTX
Post processing of computed tomography
PPTX
mr angiography.pptx
PPTX
Fluroscopy
PPT
CT History
PPTX
MRI Coil and Gradient power-point slide pk
PPTX
Computer tomography components
PPT
Anjali (qa qc on cr dr)
Ct protocols of thorax
Flouroscopic imging
MDCT (2)
CT artifact
Quality Control in MRI- Avinesh Shrestha
Mri artifacts
Generations of CT scanner
MACRORADIOGRAPHY.pptx
Magnification(macro and micro radiography), distortion
DIGITAL FLUOROSCOPY DMR.pptx
MAMMOGRAPHY VIEWS .ppt
Basics in ct
ICRP-International commission on Radiation Protection
Post processing of computed tomography
mr angiography.pptx
Fluroscopy
CT History
MRI Coil and Gradient power-point slide pk
Computer tomography components
Anjali (qa qc on cr dr)
Ad

Similar to Quality Assurance Programme in Computed Tomography (20)

PPTX
06.CT6.pptx the best lecturer in CT technical
PPTX
COMPUTED TOMOGRPAHY SCAN LESSONS PPT FOR RADIOLOGIC TECHONOLOGY
PPTX
QUALITY ASSUARENCE AND CONTROL RADIOLOGY.pptx
PPT
Qa of ct and mamography
PPTX
DR-DDR(QUALITY CONTROL TESTDigital Radiography).pptx
PPTX
medical equipment lec 6 CT scan
PPTX
5lab components of ct scanner
PPTX
CT image quality and image resolution.pptx
PPT
Principles of ct
PPT
7 Physical principles of CT.pptasfbmhjugr
PPTX
CT COMPONENTS
PPTX
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
PPTX
Ct Scan my sb dbdhdb dd. dd d d s s d e e .pptx
PPT
Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOU...
PPT
Computed tomography - Physics and Clinical
PPTX
PPT
Basic Computed Tomography for Medical Imaging
PPTX
Computed Tomography Equipments.pptx
PPTX
COMPUTED TOMOGRAPHY SCAN
PPTX
INTRODUCTION OF CT , VARIOUS GENERATIONS .pptx
06.CT6.pptx the best lecturer in CT technical
COMPUTED TOMOGRPAHY SCAN LESSONS PPT FOR RADIOLOGIC TECHONOLOGY
QUALITY ASSUARENCE AND CONTROL RADIOLOGY.pptx
Qa of ct and mamography
DR-DDR(QUALITY CONTROL TESTDigital Radiography).pptx
medical equipment lec 6 CT scan
5lab components of ct scanner
CT image quality and image resolution.pptx
Principles of ct
7 Physical principles of CT.pptasfbmhjugr
CT COMPONENTS
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
Ct Scan my sb dbdhdb dd. dd d d s s d e e .pptx
Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOU...
Computed tomography - Physics and Clinical
Basic Computed Tomography for Medical Imaging
Computed Tomography Equipments.pptx
COMPUTED TOMOGRAPHY SCAN
INTRODUCTION OF CT , VARIOUS GENERATIONS .pptx
Ad

More from Ramzee Small (11)

PDF
Patient Dose Audit in Computed Tomography at Cancer Institute of Guyana
PPTX
Pacs configuration
PPTX
Positron Emission Tomography
PPTX
Side lobes and Harmonic Imaging
PPTX
Splenomegaly
PPTX
Spleen ultrasound
PDF
Ultrasonography in Abdominal Aortic Aneurysm Diagnosis (A Literature Review)
PPTX
Mammalian cycle
PPTX
Doppler effect
PPTX
Emphysema
PPTX
Cardiophrenic angle
Patient Dose Audit in Computed Tomography at Cancer Institute of Guyana
Pacs configuration
Positron Emission Tomography
Side lobes and Harmonic Imaging
Splenomegaly
Spleen ultrasound
Ultrasonography in Abdominal Aortic Aneurysm Diagnosis (A Literature Review)
Mammalian cycle
Doppler effect
Emphysema
Cardiophrenic angle

Recently uploaded (20)

DOCX
Copies if quanti.docxsegdfhfkhjhlkjlj,klkj
PPT
Pyramid Points Lab Values Power Point(11).ppt
PPTX
Nancy Caroline Emergency Paramedic Chapter 16
PPTX
Arthritis Types, Signs & Treatment with physiotherapy management
PPTX
Nancy Caroline Emergency Paramedic Chapter 17
PPTX
FUNCTIONS OF BLOOD PART I AND PART 2 WHOLE
PPTX
Nancy Caroline Emergency Paramedic Chapter 13
PPTX
unit1-introduction of nursing education..
PPTX
guidance--unit 1 semester-5 bsc nursing.
PPTX
DeployedMedicineMedical EquipmentTCCC.pptx
PPT
12.08.2025 Dr. Amrita Ghosh_Stocks Standards_ Smart_Inventory Management_GCLP...
PPTX
Care Facilities Alcatel lucenst Presales
PPTX
Nancy Caroline Emergency Paramedic Chapter 15
PDF
Introduction to Clinical Psychology, 4th Edition by John Hunsley Test Bank.pdf
PPTX
HIGHLIGHTS of NDCT 2019 WITH IMPACT ON CLINICAL RESEARCH.pptx
PPTX
POSTURE.pptx......,............. .........
PPTX
Nancy Caroline Emergency Paramedic Chapter 4
PPTX
Diabetes_Pathology_Colourful_With_Diagrams.pptx
PPTX
PEDIATRIC OSCE, MBBS, by Dr. Sangit Chhantyal(IOM)..pptx
PDF
01. Histology New Classification of histo is clear calssification
Copies if quanti.docxsegdfhfkhjhlkjlj,klkj
Pyramid Points Lab Values Power Point(11).ppt
Nancy Caroline Emergency Paramedic Chapter 16
Arthritis Types, Signs & Treatment with physiotherapy management
Nancy Caroline Emergency Paramedic Chapter 17
FUNCTIONS OF BLOOD PART I AND PART 2 WHOLE
Nancy Caroline Emergency Paramedic Chapter 13
unit1-introduction of nursing education..
guidance--unit 1 semester-5 bsc nursing.
DeployedMedicineMedical EquipmentTCCC.pptx
12.08.2025 Dr. Amrita Ghosh_Stocks Standards_ Smart_Inventory Management_GCLP...
Care Facilities Alcatel lucenst Presales
Nancy Caroline Emergency Paramedic Chapter 15
Introduction to Clinical Psychology, 4th Edition by John Hunsley Test Bank.pdf
HIGHLIGHTS of NDCT 2019 WITH IMPACT ON CLINICAL RESEARCH.pptx
POSTURE.pptx......,............. .........
Nancy Caroline Emergency Paramedic Chapter 4
Diabetes_Pathology_Colourful_With_Diagrams.pptx
PEDIATRIC OSCE, MBBS, by Dr. Sangit Chhantyal(IOM)..pptx
01. Histology New Classification of histo is clear calssification

Quality Assurance Programme in Computed Tomography

  • 2. Outline  Introduction to Computed Tomography  Component of CT System  CT Data Acquisition System  Quality Assurance in CT  QA/QC Test in CT  Conclusion  References
  • 3. Computed Tomography Is a medical imaging procedure of creating a cross-sectional tomographic plane of any part of the body . CT scanners create cross-sectional images of high radiographic contrast The dose to the patient may be significantly higher than with alternative imaging modalities
  • 4. Computed Tomography Component of CT System Computer (Operating Console) Gantry and Table X-ray Tube Filter Detector
  • 5. Computed Tomography Computer Provides the link between the CT technologist and the other components of the system. Control of data acquisition, Image reconstruction, Storage of image data, and Image display
  • 6. Computed Tomography Gantry and Table A circular device that house the;  The X-ray tube The Detector Array Data Acquisition System The Generator
  • 7. Computed Tomography Gantry and Table continue… The table is a flat or slightly curved platform that the patient lies on during the scan. The table is responsible for positioning the patient with the gantry The patient couch on the table must be constructed of low Z-material, such as carbon fiber to not interfere with x-ray beam transmission and patient imaging.
  • 8. Computed Tomography X-ray Tube They have an anode heat storage capacity of 4-5 MHU (million heat unit) or more. CT x-ray tube use a rotating anode to increase heat dissipation
  • 9. COMPUTED TOMOGRAPHY Detector The detectors in CT function as image receptors. It measures the amount of radiation transmitted through the body and converts the measurement into an electrical signal. The two basic detector types used in CT are scintillation (solid state) and ionization (xenon gas) detectors Detector Rings
  • 10. Computed Tomography Data Acquisition System X-ray photons come on the detector. The detector detects the intensity in form of current. The current is converted into voltage. The analog signal is converted into digital form. This signal can now be processed and reconstructed in the computer Detector Current to Voltage Convertor Pre Amplifier Analog to Digital Convertor Computer
  • 11. Quality Assurance Quality assurance (QA) is a program used by management to maintain optimal diagnostic image quality with minimum hazard and distress to patients. The program includes; Periodic quality control tests, Administrative Procedure Corrective action and Training
  • 12. QA/QC Test In CT Field Uniformity Test Data Analysis Daily Test Objective Equipment Warm-up Test To ensure equipment is working properly CT Number for Water To ensure consistency for CT calibrated number for water Field Uniformity Test To evaluate the scan plane uniformity
  • 13. QA/QC Test In CT Monthly Test Objective CT Light Alignment Test To ensure that the internal and external alignment lights are properly aligned with the tomographic plane and the couch SPR Accuracy To ensure that the SPR image accurately indicates the patient position CT Number Accuracy, Image Noise, Uniformity And Image Artifacts To ensure a) CT number accuracy; b) Noise levels of CT images being within the tolerances with respect to their reference values; c) Uniformity of CT number and image noise across the FOV; d) Minimal image artefacts Image Display and Printing To confirm that the image displays are reproducing all of the grey scale information in the image accurately and that the printed images are a true replica of the image display Low Contrast Resolution Test To determines the capability of the scanner to discriminate low contrast objects High Contrast Spatial Resolution To determines the high contrast spatial frequency limits of the CT scanner under various conditions Scan Increment Accuracy To determines the deviation of indicated versus actual scan increment
  • 14. QA/QC Test In CT Annual Test Objective Visual Inspection and Programme Review 1. To ensure that the CT scanner and adjacent areas are safe and have the necessary radiation protection supplies 2. To ensure that CT QC test are being conducted at appropriate time periods and corrective actions are taken when necessary X-ray Generator If the patient dose is excessively high, then kV and HVL should be investigated to determine the cause of high dose Radiation Dose 1. To Compare the determined CVOL (CTDVOL) with the manufacturer’s specification. 2. To determine the CVOL (CTDIVOL) and CT (DLP) values for selected protocols representing the main examinations clinically performed at the facility 3. To Review and update the CVOL (CTDIVOL) and CT (DLP) values for selected protocols representing the main examinations clinically performed at the facility X-ray Beam Width To determine the accuracy of collimator settings Spatial Resolution To ensure that the spatial resolution of a reconstructed image complies with manufacture standards
  • 15. QA/QC Test In CT Software Spatial/geometry accuracy tests oCorrect data transfer Image Registration Image reconstruction oTests should verify that the software accurately reconstructs known phantoms
  • 16. QA/QC CT Test Phantom Line Pair High Contrast Phantom Siemens Phantom CT number regions in RMI ACR accreditation phantom RMI Model 464 Head and body dosimetry phantoms
  • 17. Conclusion The QA program is designed to improve accuracy of patient treatments and efficiency of the treatment planning process while keeping dose ALARA QA program should evolve and adapt as the device used for CT simulation process change
  • 18. References Quality Assurance Programme for Computed Tomography; Diagnostic and Therapy Application, International Atomic Energy Agency IAEA, Human Health Series No. 19, 2012, ISBN 978–92–0–128910–0 Quality assurance workbook for radiographers & radiological technologists, World Health Organization, 2001, ISBN 9241546425 American Association of Physicists in Medicine. Quality Assurance for Computed- Tomography Simulators and the Computed Tomography- Simulation Process: Report of the AAPM Radiation Therapy Committee Task Group No. 66.,2003. Quality Control Recommendations for Diagnostic Radiography Conference of Radiation Control Programme Directors Inc. CRCPD, Volume 2, 2001
  • 19. References American College of Radiology, Computed Tomography Quality Control Manual, 2012 Smith Steven W., The Scientist and Engineer’s Guide to Digital Signal Processing, 1997-1998 Phillip W. Ballinger, Eugene D. Frank, volume 3: Merrill's Atlas of Radiographic Positions and Radiologic Procedures, Edition 10, ISBN (Set) 0-323-01604-9 Jerrold T. Bushberg, J. Anthony Seibert, Edwin M. Leidholdt Jr., John N. Boone, The Essential Physic of Medical Imaging, second edition, ISBN 0-683-30118-7 Kak A.C., Slaney M., Principles of Computed Tomographic Imaging. SIAM, 1987. Flohr T.G., Schaller S., Stierstorfer K., et al., "Multi-Detector Row CT Systems and Image-Reconstruction Techniques," Radiology, June 2005.
  • 20. The End ???..... I Don’t Think So!!!