SlideShare a Scribd company logo
Contact temperature sensor calibration
Contents
• What is Calibration
• When Calibration is required
• Method of Calibration
• Primary Calibration
• Fixed Point calibration
• Maintenance Apparatus
• Secondary Calibration
• Selection of Source
• Stable Temperature source
• Master Sensors
• Cold Junction Compensation
• Reference Junction
• Reference Junction Units
• Uncertainty Calculations
• Documentation Requirement
• Importance of accreditation
• Calibration Report
• Calibration is the process of determining or adjusting the deviation of a
instrument (which is under calibration) with reference to a said standard.
• This is done by comparing the device or the output of an instrument to a
standard having known characteristics.
• To improve the quality of the calibration and have the results accepted by
outside organization it is desirable for the calibration and subsequent
measurement to be “traceable” to the internationally defined measurement
units.
What is calibration?
• With a new instrument.
• When a specified time period is elapsed.
• When a specified usage (operating hours) has elapsed.
• When an instrument has had a shock or vibration which potentially may have
put it out of calibration.
• Whenever observations appear questionable.
When Calibration is required
Method of Calibration
Generally contact type temperature sensor calibration can be done in two ways:
1. Primary calibration - Primary calibration is done by taking fixed points, whose
temperature remain same. E.g. Of fixed points are melting and freezing points
of water.
2. Secondary calibration - Secondary calibration is done by comparison method.
i.e. by comparing master and test sensor.
Primary Calibration
Certain systems in nature always seem to have the same temperature, for
example
• Melting Ice
• Boiling Water at sea level
• The interior of healthy human body
• Phase equilibrium points of pure material
These “landmarks” of temperature can serve as thermometric fixed points which
can be used as an apparatus for calibrating temperature sensors.
Fixed point method
• This method consists in setting up a thermometer in a fixed-
point cell that provides an isothermal environment.
• The fixed point cell is a flask filled with pure material and
protected by a shell. The cell is placed in an apparatus that
must provide good temperature control and sufficient cell
immersion to generate a uniform temperature in the
measurement zone.
• The apparatus must provide for fixed-point cooling as well as
controlled heat.
• Few fixed points are chosen, consistent with the need to
establish a good interpolating formula.
• This method gives a highly accurate calibration and is used
only for highest quality thermometry.
Maintenance apparatus
for fixed point cells
Secondary Calibration
• Secondary calibration is done by comparison method. In
comparison method we need:
a.) Stable temperature source
b.) Calibrated master sensor
c.) Calibrated meter
• large numbers of thermometers can be calibrated
Simultaneously by comparison method.
• Suitable for most of the industrial calibrations.
• Comparison calibrations can be performed in a
laboratory or in the field. High accuracy can be obtained
with careful selection of equipment
Stable Temperature Source
Dry Block Calibrator
• The unit consists of a special designed heating block which has located internal holes
for the probes. Although the block temperature is accurately controlled.
CALsys 1200L….300° to 1200°C
Stirred Liquid Bath
• Provide superior thermal environment for probe immersion as no air gap exist between
the probe and the medium. Thermal coupling is therefore much better. Methanol is
used for temperature below 0°C, water from 0 to 80°C and silicon oil for up to 250°C.
CALsys 650….50° to 650° C
Selection of Sources
Below are important specifications to consider when deciding on a particular temperature
bath
Stability: dynamic variability around a static measurement. Utopia would be when heat
input minus losses equals zero for an indefinite amount of time. Always expressed as +/- of
a value.
Radial Uniformity: horizontal or side to side, one point to another variation. Always
expressed as a single value, no + or -.
Axial Uniformity: vertical or lengthwise, one point to another variation. Always expressed
as a single value, no + or -.
Heat up and cool down time: how fast will the bath heat up to its maximum temperature
and then the amount of time it takes to cool down to a safe level.
Accuracy: true or known proximity to desired setting as verified by a traceable standard of
much higher accuracy. Always expressed as +/- of a value.
SSPRT : Pt100/Pt25, Temperature range 0 to 661°C
RTD : Pt100
Accuracy : 1/10, 1/5, 1/3, 1/2 DIN, Class A
Sheath Material : SS316/ Inconel/Quartz
THERMOCOUPLE : K/N/R/S
Accuracy : Special, Class 1, with option cold junction compensation
Sheath Material : Inconel/Ceramic (KER 710-C 799)
Master Sensors
Master Sensors
• To calibrate industrial sensors the laboratory must compare them to standard
thermometers whose characteristics have been defined and are traceable to National
Standards. Such thermometers that are traceable to national standard are known as
'Master Sensors'.
• Master sensor is used as a reference sensor in comparison type calibration process.
• Master sensor is put in the highly stable temperature zone along with sensor under
calibration, their reading is compared and error is calculated.
• Master should have high accuracy, and repeatability and should be recalibrated after
certain period of time (generally time period suggested by calibration Laboratories).
Cold Junction Compensation
• By connecting any thermocouple to measurement device three dissimilar metal
junction are created in the circuit: the thermocouple junction itself, or hot junction,
and the junction between each lead and the measurement device, or cold junctions.
• These cold junction provide their own thermoelectric voltages that are proportional
to the temperature at the device terminals.
• A technology known as cold junction compensation is therefore used to remove this
unwanted effect.
Reference Junction
• Thermocouple output is dependent of temperature of hot junction and
temperature of cold junction.
• The Cold junction temperature needs to be known to get the hot junction
temperature.
• To reduce the uncertainty of cold junction measurement Reference junction boxes
are used.
• A reference junction box terminates the thermocouple cold end at a known and
constant temperature. This can be 0° or 60° C, as per user requirement.
• The EMF of the reference junction is fixed and is added on to the output of the
thermocouple.
Reference Junction Units
Spec / Model CalREF 0 CalREF 60
Reference Junction
0°C(or Ambient -25° to
90°)
60°C(or Ambient +10° to
90°)
Accuracy +/- 0.5°C,errors can be compensated
Stability +/-0.05°C
Stabilization time 10 Min
Input J/K/T/N/S/R type thermocouple 6-24 channel
Control Peltier cells + RTD + PID Heater + RTD + PID
Resolution 0.1 °C
Dimension
400 H x 500 W x 200 D
mm
Weight 13 kg
Mounting Wall mounting, 19" rack or standalone
Output
Plug/Jack or Terminals; optional temperature
feedback RTD, Alram
Power supply 230 VAC/ 24 V DC
Uncertainty calculation
Uncertainty Factors
Type B-Evaluation (tabular form)
source Distribution DOF Calculation Value
stability of calibration bath Rectangular infinity 0.500/1.732 0.2887
uncertainty of master sensor Normal infinity 1.000/2.000 0.5
uncertainty of precision thermometer Normal infinity 0.170/2.000 0.085
least count of precision thermometer Rectangular infinity 0.010/1.732 0.0058
Type A -Evaluation
Readings Xi Average X Xi-X Std. Dev
Ref UUC Ref UUC Ref UUC Ref UUC
993.67 994.1
993.652 994.098
0.018 0.002
0.040249 0.020494
993.67 994.08 0.018 -0.018
993.58 994.08 -0.072 -0.018
993.67 994.1 0.018 0.002
993.67 994.13 0.018 0.032
Degree of freedom : 4
The standard uncertainty type A assuming Normal Probability Distribution is given b
uA = 0.0205/sqrt5
= 0.0092ºC
Degree of freedom = 4
Formula
uA = standard deviation /sqrt(n)
Uncertainty Budget
Source of
Uncertainty
Limits(°C) Probability Distribution Standard
Uncertainty(°C)
Sensitivity
Coefficient
Uncertainty
Contribution
Degree
of freedom
X1 ±Xi Type A or Type B U (Xi) Ci °C
u1 0.5000 Rectangular 1.732 B 0.5000/1.732 1 0.2887 Infinitive
u2 1.0000 Normal 1.96 B 1.0000/2.000 1 0.5 Infinitive
u3 0.1700 Normal 1.96 B 0.1700/2.000 1 0.085 Infinitive
u4 0.0100 Rectangular 1.732 B 0.0100/1.732 1 0.0058 Infinitive
uA 0.0402 Normal Sqrt (5) A 0.0205/2.236 1 0.0092 4
Result of Uncertainty Calculation
• Combined Standard Uncertainty, Uc(°C) : 0.5837
• Effective degree of freedom, Veff : 65,795,465.03
• Coverage factor, K : 2.000
• Expanded uncertainty, Uex (°C) : 1.4000
Formulas:
1. Uc = sqrt (uA2 +u12 +u22+u32+u42)
2. Veff = (n-1)*(Uc/uA) ^4; (n=no of observations)
3. If Veff>30 then DOF=infinitive, so K=2
4. Uex = K*Uc
Documentation Requirement
Documentation Requirement
Documentation Requirement
Documentation Requirement
Importance of Accreditation
Accreditation is important in following ways:
A Recognition of Testing Competence
Laboratory accreditation is a means of determining the technical competence of
laboratories to perform specific types of testing, measurement and calibration.
A BENCHMARK FOR PERFORMANCE
Laboratory accreditation benefits laboratories by allowing them to determine whether
they are performing their work correctly and to appropriate standards, and provides
them with a benchmark for maintaining that competence
A Marketing Advantage
Accreditation is an effective marketing tool for testing, calibration and measurement
organizations, and a passport to submit tenders to contractors that require
independently verified laboratories.
International Recognition for Laboratory
Many countries around the world have one or more organizations responsible for the
accreditation of their nation’s laboratories
Calibration Report
Quality
Measured/Instruments
Temperature Range Calibration &
Measurement Capability
Contact Type RTD,
Thermocouples,
Thermometers
-196° C 0.05° C
-80° C to -38° C 0.05° C
-38° C to 0° C 0.03° C
> 0°C to 140°C 0.03° C
> 140°C to 250°C 0.04° C
> 250°C to 650°C 0.12° C
> 650°C to 1000°C 1.26° C
> 1000°C to 1200°C 1.26° C
> 1200°C to 1600°C 2.64° C
Non Contact type
Pyrometer
0°C to 100°C 1.5° C
>100°C to 500°C
>500°C to 1500°C
>1500°C to 1700°C
2.4° C
2.72° C
3.27° C
>1700°C to 2700°C 5.3° C
In House Calibration Facility:
Best Measurement Capabilities and
temperature range is as shown in table
Quality
Measured/Instruments
Temperature Range Calibration &
Measurement Capability
Calibration of SPRT/PRTS/
THERMOCOUPLES at
Triple Point of Water (0.01°C) 0.0035°C
Melting Point of Gallium (29.7646°C) 0.0070°C
Freezing Point of Tin (231.928°C) 0.0070°C
Freezing Point of Zinc (419.527°C) 0.0075°C
Freezing Point of Aluminum (660.323 °C) 0.0075°C
Fixed Point Calibration Facility
We are first private company to provide fixed point temperature calibration for Triple
point of water (Tpw), Gallium (Ga), Tin (Sn), Zinc (Zn) & Aluminum (Al) Cells.
Tempsens calibration Centres-
Udaipur, Vadodara ,Bangalore
Udaipur: Tempsens Calibration Centre
B-188A, Road No.5, M.I.A., Madri,
Udaipur – 313003 – Rajasthan – INDIA
Phone:91-294-3057724
Email: lab@tempsens.com
Baroda: Tempsens Calibration Center
TF- 304, Florance classic, 10, Ashapuri Society,
Near Unnati School Akota,
Baroda - 390020 (Gujarat)
Mob. 09327157887
Email: gujarat@tempsens.com
Bangalore: Tempsens Calibration Center
No.7, Ist Main Coconut Garden
Nagarabhavil Main Road
Bangalore-560072 (K.N)
Mob. 09343183607
Email: bangalore@tempsens.com
• Tempsens calibration centers functions as per ISO17025/NABL standards.
• Calibration of contact type sensors can be made in temperature range of -196°C to
1600°C and Calibration of non contact type sensors can be made in temperature
range 0°C to 2700°C
THANK YOU!
Tempsens Instruments (I) Pvt. Ltd.
B-188 A, Road No. 5, MIA, Madri
Udaipur - 313 003 (Rajasthan) INDIA
Phone : +91-294-3057700
Fax : +91-294-3057750
E-mail : info@tempsens.com
Website : www.tempsens.com
Temperature Sensing, Heating & Cables

More Related Content

PPTX
Temperature Measurement
PPTX
Unit 5 temperature measurement
PPT
Industrial Temperature Measurement
PDF
Temperature measurement part II
PPTX
Hygrometers
PDF
Sensors and Transducers (1).pdf
PPTX
Temperature Measurement
PPTX
Characteristics of Measuring instruments
Temperature Measurement
Unit 5 temperature measurement
Industrial Temperature Measurement
Temperature measurement part II
Hygrometers
Sensors and Transducers (1).pdf
Temperature Measurement
Characteristics of Measuring instruments

What's hot (20)

PPTX
RTD-RESISTANCE TEMPERATURE DETECTOR
PPTX
Temperature sensor
PPTX
“Temperature Sensors” Thermocouple | Thermistor | Resister Temperature Detectors
PPTX
Temperature measurement ppt
PPTX
Rtd and thermocouples
PDF
Unit – 4 Transducers and sensors:Definition and types of transducers
PPT
Calibration
PPTX
Introduction to Industrial Instrumentation
PPTX
PPTX
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015
PPT
Level Measurement
PPTX
Temperature Sensor Thermocouple and RTD
PPSX
Pressure measurement gauges
PDF
Industrial Instrumentation An Introduction
PPTX
Introduction to measurements systems
PPSX
Instrumentation measurement principles
PPT
Data Acquisition System and Data loggers
PPTX
Rtd (resistance temperature detector)
RTD-RESISTANCE TEMPERATURE DETECTOR
Temperature sensor
“Temperature Sensors” Thermocouple | Thermistor | Resister Temperature Detectors
Temperature measurement ppt
Rtd and thermocouples
Unit – 4 Transducers and sensors:Definition and types of transducers
Calibration
Introduction to Industrial Instrumentation
How to Calibrate an RTD Using a Dry Block Calibrator Webinar_061015
Level Measurement
Temperature Sensor Thermocouple and RTD
Pressure measurement gauges
Industrial Instrumentation An Introduction
Introduction to measurements systems
Instrumentation measurement principles
Data Acquisition System and Data loggers
Rtd (resistance temperature detector)
Ad

Similar to Contact temperature sensor calibration (20)

PDF
PDF
Calibration times march 2011
PDF
Calibration times march 2011
PPTX
Understanding AMS2750E (intermediate) v2
PDF
4.0 technical reference for monitoring equipment
PPT
Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...
DOCX
1.)(485 pages) C&I Measurements and Conversions.docx
PPTX
Thermal analysis 4
PDF
Temperature measurement
PPTX
thermometry
PDF
Comparative analysis of different temperature analysis
DOC
Project report on design and fabrication of temperature measurement setup
DOCX
PARAMETERS AND THEIR APPROXIMATE MEASUREMENT POINTS IN A THERMAL POWER PLANT
DOCX
Last Rev. August 2014 Calibration and Temperature Measurement.docx
PDF
Fluke Temperature Calibration Catalog
PPT
TemperatureMeasurement TemperatureMeasurement
PPT
Temperature Measurement
PPT
Presentation (Temperature )
PPTX
Tempreture measurement
PDF
Temperature Sensors used in Pharmaceutical Industry and its application
Calibration times march 2011
Calibration times march 2011
Understanding AMS2750E (intermediate) v2
4.0 technical reference for monitoring equipment
Introduction to Temperature Measurement and Calibration Presented by Fluke Ca...
1.)(485 pages) C&I Measurements and Conversions.docx
Thermal analysis 4
Temperature measurement
thermometry
Comparative analysis of different temperature analysis
Project report on design and fabrication of temperature measurement setup
PARAMETERS AND THEIR APPROXIMATE MEASUREMENT POINTS IN A THERMAL POWER PLANT
Last Rev. August 2014 Calibration and Temperature Measurement.docx
Fluke Temperature Calibration Catalog
TemperatureMeasurement TemperatureMeasurement
Temperature Measurement
Presentation (Temperature )
Tempreture measurement
Temperature Sensors used in Pharmaceutical Industry and its application
Ad

More from Tempsens Instruments (13)

PPTX
Thermowells & Accessories
PPTX
Tempsens- cable and wires
PPTX
Mineral insulated (mi) cables
PPT
Temperature Calibration equipment for Thermcouple, RTDs, and Pyrometers
PPTX
Laboratory furnace
PPTX
Process , Component and Industrial heaters
PPTX
Ultra high temperature thermocouple
PPTX
Self Powered Neutron Detectors - SPND
PPTX
Furnace calibration - temperature uniformity survey
PPTX
Calibration of NON Contact Pyrometers
PPTX
Basics of Wires and Cables
PPTX
Basics of Thermocouples and RTDs
PPTX
Basics Of Pyrometers
Thermowells & Accessories
Tempsens- cable and wires
Mineral insulated (mi) cables
Temperature Calibration equipment for Thermcouple, RTDs, and Pyrometers
Laboratory furnace
Process , Component and Industrial heaters
Ultra high temperature thermocouple
Self Powered Neutron Detectors - SPND
Furnace calibration - temperature uniformity survey
Calibration of NON Contact Pyrometers
Basics of Wires and Cables
Basics of Thermocouples and RTDs
Basics Of Pyrometers

Recently uploaded (20)

PDF
Well-logging-methods_new................
PDF
The CXO Playbook 2025 – Future-Ready Strategies for C-Suite Leaders Cerebrai...
PPTX
Fundamentals of safety and accident prevention -final (1).pptx
PPTX
bas. eng. economics group 4 presentation 1.pptx
PDF
composite construction of structures.pdf
PDF
Embodied AI: Ushering in the Next Era of Intelligent Systems
PDF
Operating System & Kernel Study Guide-1 - converted.pdf
PPTX
Foundation to blockchain - A guide to Blockchain Tech
PDF
R24 SURVEYING LAB MANUAL for civil enggi
PDF
Enhancing Cyber Defense Against Zero-Day Attacks using Ensemble Neural Networks
PPTX
Construction Project Organization Group 2.pptx
PPTX
CH1 Production IntroductoryConcepts.pptx
PDF
Mitigating Risks through Effective Management for Enhancing Organizational Pe...
PPTX
Internet of Things (IOT) - A guide to understanding
PPT
Project quality management in manufacturing
PDF
PPT on Performance Review to get promotions
PPTX
Geodesy 1.pptx...............................................
PPTX
Engineering Ethics, Safety and Environment [Autosaved] (1).pptx
PDF
737-MAX_SRG.pdf student reference guides
PPTX
FINAL REVIEW FOR COPD DIANOSIS FOR PULMONARY DISEASE.pptx
Well-logging-methods_new................
The CXO Playbook 2025 – Future-Ready Strategies for C-Suite Leaders Cerebrai...
Fundamentals of safety and accident prevention -final (1).pptx
bas. eng. economics group 4 presentation 1.pptx
composite construction of structures.pdf
Embodied AI: Ushering in the Next Era of Intelligent Systems
Operating System & Kernel Study Guide-1 - converted.pdf
Foundation to blockchain - A guide to Blockchain Tech
R24 SURVEYING LAB MANUAL for civil enggi
Enhancing Cyber Defense Against Zero-Day Attacks using Ensemble Neural Networks
Construction Project Organization Group 2.pptx
CH1 Production IntroductoryConcepts.pptx
Mitigating Risks through Effective Management for Enhancing Organizational Pe...
Internet of Things (IOT) - A guide to understanding
Project quality management in manufacturing
PPT on Performance Review to get promotions
Geodesy 1.pptx...............................................
Engineering Ethics, Safety and Environment [Autosaved] (1).pptx
737-MAX_SRG.pdf student reference guides
FINAL REVIEW FOR COPD DIANOSIS FOR PULMONARY DISEASE.pptx

Contact temperature sensor calibration

  • 2. Contents • What is Calibration • When Calibration is required • Method of Calibration • Primary Calibration • Fixed Point calibration • Maintenance Apparatus • Secondary Calibration • Selection of Source • Stable Temperature source • Master Sensors • Cold Junction Compensation • Reference Junction • Reference Junction Units • Uncertainty Calculations • Documentation Requirement • Importance of accreditation • Calibration Report
  • 3. • Calibration is the process of determining or adjusting the deviation of a instrument (which is under calibration) with reference to a said standard. • This is done by comparing the device or the output of an instrument to a standard having known characteristics. • To improve the quality of the calibration and have the results accepted by outside organization it is desirable for the calibration and subsequent measurement to be “traceable” to the internationally defined measurement units. What is calibration?
  • 4. • With a new instrument. • When a specified time period is elapsed. • When a specified usage (operating hours) has elapsed. • When an instrument has had a shock or vibration which potentially may have put it out of calibration. • Whenever observations appear questionable. When Calibration is required
  • 5. Method of Calibration Generally contact type temperature sensor calibration can be done in two ways: 1. Primary calibration - Primary calibration is done by taking fixed points, whose temperature remain same. E.g. Of fixed points are melting and freezing points of water. 2. Secondary calibration - Secondary calibration is done by comparison method. i.e. by comparing master and test sensor.
  • 6. Primary Calibration Certain systems in nature always seem to have the same temperature, for example • Melting Ice • Boiling Water at sea level • The interior of healthy human body • Phase equilibrium points of pure material These “landmarks” of temperature can serve as thermometric fixed points which can be used as an apparatus for calibrating temperature sensors.
  • 7. Fixed point method • This method consists in setting up a thermometer in a fixed- point cell that provides an isothermal environment. • The fixed point cell is a flask filled with pure material and protected by a shell. The cell is placed in an apparatus that must provide good temperature control and sufficient cell immersion to generate a uniform temperature in the measurement zone. • The apparatus must provide for fixed-point cooling as well as controlled heat. • Few fixed points are chosen, consistent with the need to establish a good interpolating formula. • This method gives a highly accurate calibration and is used only for highest quality thermometry.
  • 9. Secondary Calibration • Secondary calibration is done by comparison method. In comparison method we need: a.) Stable temperature source b.) Calibrated master sensor c.) Calibrated meter • large numbers of thermometers can be calibrated Simultaneously by comparison method. • Suitable for most of the industrial calibrations. • Comparison calibrations can be performed in a laboratory or in the field. High accuracy can be obtained with careful selection of equipment
  • 10. Stable Temperature Source Dry Block Calibrator • The unit consists of a special designed heating block which has located internal holes for the probes. Although the block temperature is accurately controlled. CALsys 1200L….300° to 1200°C
  • 11. Stirred Liquid Bath • Provide superior thermal environment for probe immersion as no air gap exist between the probe and the medium. Thermal coupling is therefore much better. Methanol is used for temperature below 0°C, water from 0 to 80°C and silicon oil for up to 250°C. CALsys 650….50° to 650° C
  • 12. Selection of Sources Below are important specifications to consider when deciding on a particular temperature bath Stability: dynamic variability around a static measurement. Utopia would be when heat input minus losses equals zero for an indefinite amount of time. Always expressed as +/- of a value. Radial Uniformity: horizontal or side to side, one point to another variation. Always expressed as a single value, no + or -. Axial Uniformity: vertical or lengthwise, one point to another variation. Always expressed as a single value, no + or -. Heat up and cool down time: how fast will the bath heat up to its maximum temperature and then the amount of time it takes to cool down to a safe level. Accuracy: true or known proximity to desired setting as verified by a traceable standard of much higher accuracy. Always expressed as +/- of a value.
  • 13. SSPRT : Pt100/Pt25, Temperature range 0 to 661°C RTD : Pt100 Accuracy : 1/10, 1/5, 1/3, 1/2 DIN, Class A Sheath Material : SS316/ Inconel/Quartz THERMOCOUPLE : K/N/R/S Accuracy : Special, Class 1, with option cold junction compensation Sheath Material : Inconel/Ceramic (KER 710-C 799) Master Sensors
  • 14. Master Sensors • To calibrate industrial sensors the laboratory must compare them to standard thermometers whose characteristics have been defined and are traceable to National Standards. Such thermometers that are traceable to national standard are known as 'Master Sensors'. • Master sensor is used as a reference sensor in comparison type calibration process. • Master sensor is put in the highly stable temperature zone along with sensor under calibration, their reading is compared and error is calculated. • Master should have high accuracy, and repeatability and should be recalibrated after certain period of time (generally time period suggested by calibration Laboratories).
  • 15. Cold Junction Compensation • By connecting any thermocouple to measurement device three dissimilar metal junction are created in the circuit: the thermocouple junction itself, or hot junction, and the junction between each lead and the measurement device, or cold junctions. • These cold junction provide their own thermoelectric voltages that are proportional to the temperature at the device terminals. • A technology known as cold junction compensation is therefore used to remove this unwanted effect.
  • 16. Reference Junction • Thermocouple output is dependent of temperature of hot junction and temperature of cold junction. • The Cold junction temperature needs to be known to get the hot junction temperature. • To reduce the uncertainty of cold junction measurement Reference junction boxes are used. • A reference junction box terminates the thermocouple cold end at a known and constant temperature. This can be 0° or 60° C, as per user requirement. • The EMF of the reference junction is fixed and is added on to the output of the thermocouple.
  • 17. Reference Junction Units Spec / Model CalREF 0 CalREF 60 Reference Junction 0°C(or Ambient -25° to 90°) 60°C(or Ambient +10° to 90°) Accuracy +/- 0.5°C,errors can be compensated Stability +/-0.05°C Stabilization time 10 Min Input J/K/T/N/S/R type thermocouple 6-24 channel Control Peltier cells + RTD + PID Heater + RTD + PID Resolution 0.1 °C Dimension 400 H x 500 W x 200 D mm Weight 13 kg Mounting Wall mounting, 19" rack or standalone Output Plug/Jack or Terminals; optional temperature feedback RTD, Alram Power supply 230 VAC/ 24 V DC
  • 18. Uncertainty calculation Uncertainty Factors Type B-Evaluation (tabular form) source Distribution DOF Calculation Value stability of calibration bath Rectangular infinity 0.500/1.732 0.2887 uncertainty of master sensor Normal infinity 1.000/2.000 0.5 uncertainty of precision thermometer Normal infinity 0.170/2.000 0.085 least count of precision thermometer Rectangular infinity 0.010/1.732 0.0058
  • 19. Type A -Evaluation Readings Xi Average X Xi-X Std. Dev Ref UUC Ref UUC Ref UUC Ref UUC 993.67 994.1 993.652 994.098 0.018 0.002 0.040249 0.020494 993.67 994.08 0.018 -0.018 993.58 994.08 -0.072 -0.018 993.67 994.1 0.018 0.002 993.67 994.13 0.018 0.032 Degree of freedom : 4 The standard uncertainty type A assuming Normal Probability Distribution is given b uA = 0.0205/sqrt5 = 0.0092ºC Degree of freedom = 4 Formula uA = standard deviation /sqrt(n)
  • 20. Uncertainty Budget Source of Uncertainty Limits(°C) Probability Distribution Standard Uncertainty(°C) Sensitivity Coefficient Uncertainty Contribution Degree of freedom X1 ±Xi Type A or Type B U (Xi) Ci °C u1 0.5000 Rectangular 1.732 B 0.5000/1.732 1 0.2887 Infinitive u2 1.0000 Normal 1.96 B 1.0000/2.000 1 0.5 Infinitive u3 0.1700 Normal 1.96 B 0.1700/2.000 1 0.085 Infinitive u4 0.0100 Rectangular 1.732 B 0.0100/1.732 1 0.0058 Infinitive uA 0.0402 Normal Sqrt (5) A 0.0205/2.236 1 0.0092 4
  • 21. Result of Uncertainty Calculation • Combined Standard Uncertainty, Uc(°C) : 0.5837 • Effective degree of freedom, Veff : 65,795,465.03 • Coverage factor, K : 2.000 • Expanded uncertainty, Uex (°C) : 1.4000 Formulas: 1. Uc = sqrt (uA2 +u12 +u22+u32+u42) 2. Veff = (n-1)*(Uc/uA) ^4; (n=no of observations) 3. If Veff>30 then DOF=infinitive, so K=2 4. Uex = K*Uc
  • 26. Importance of Accreditation Accreditation is important in following ways: A Recognition of Testing Competence Laboratory accreditation is a means of determining the technical competence of laboratories to perform specific types of testing, measurement and calibration. A BENCHMARK FOR PERFORMANCE Laboratory accreditation benefits laboratories by allowing them to determine whether they are performing their work correctly and to appropriate standards, and provides them with a benchmark for maintaining that competence A Marketing Advantage Accreditation is an effective marketing tool for testing, calibration and measurement organizations, and a passport to submit tenders to contractors that require independently verified laboratories. International Recognition for Laboratory Many countries around the world have one or more organizations responsible for the accreditation of their nation’s laboratories
  • 28. Quality Measured/Instruments Temperature Range Calibration & Measurement Capability Contact Type RTD, Thermocouples, Thermometers -196° C 0.05° C -80° C to -38° C 0.05° C -38° C to 0° C 0.03° C > 0°C to 140°C 0.03° C > 140°C to 250°C 0.04° C > 250°C to 650°C 0.12° C > 650°C to 1000°C 1.26° C > 1000°C to 1200°C 1.26° C > 1200°C to 1600°C 2.64° C Non Contact type Pyrometer 0°C to 100°C 1.5° C >100°C to 500°C >500°C to 1500°C >1500°C to 1700°C 2.4° C 2.72° C 3.27° C >1700°C to 2700°C 5.3° C In House Calibration Facility: Best Measurement Capabilities and temperature range is as shown in table
  • 29. Quality Measured/Instruments Temperature Range Calibration & Measurement Capability Calibration of SPRT/PRTS/ THERMOCOUPLES at Triple Point of Water (0.01°C) 0.0035°C Melting Point of Gallium (29.7646°C) 0.0070°C Freezing Point of Tin (231.928°C) 0.0070°C Freezing Point of Zinc (419.527°C) 0.0075°C Freezing Point of Aluminum (660.323 °C) 0.0075°C Fixed Point Calibration Facility We are first private company to provide fixed point temperature calibration for Triple point of water (Tpw), Gallium (Ga), Tin (Sn), Zinc (Zn) & Aluminum (Al) Cells.
  • 30. Tempsens calibration Centres- Udaipur, Vadodara ,Bangalore Udaipur: Tempsens Calibration Centre B-188A, Road No.5, M.I.A., Madri, Udaipur – 313003 – Rajasthan – INDIA Phone:91-294-3057724 Email: lab@tempsens.com Baroda: Tempsens Calibration Center TF- 304, Florance classic, 10, Ashapuri Society, Near Unnati School Akota, Baroda - 390020 (Gujarat) Mob. 09327157887 Email: gujarat@tempsens.com Bangalore: Tempsens Calibration Center No.7, Ist Main Coconut Garden Nagarabhavil Main Road Bangalore-560072 (K.N) Mob. 09343183607 Email: bangalore@tempsens.com • Tempsens calibration centers functions as per ISO17025/NABL standards. • Calibration of contact type sensors can be made in temperature range of -196°C to 1600°C and Calibration of non contact type sensors can be made in temperature range 0°C to 2700°C
  • 31. THANK YOU! Tempsens Instruments (I) Pvt. Ltd. B-188 A, Road No. 5, MIA, Madri Udaipur - 313 003 (Rajasthan) INDIA Phone : +91-294-3057700 Fax : +91-294-3057750 E-mail : info@tempsens.com Website : www.tempsens.com Temperature Sensing, Heating & Cables