UNIT OUTCOME 3
• Need of defibrillator
• DC Defibrillator
• AED (advisory external defibrillator )
• Implantable Defibrillator
• Defibrillator electrodes
Conducting system of heart
VENTRICULAR FIBRILLATION
Need of defibrillator
Life-threatening cardiac arrhythmias specifically ventricular fibrillation (V-Fib) and
non-perfusing ventricular tachycardia (V-Tach).
Arrhythmias, also known as cardiac arrhythmias, heart arrhythmias, or
dysrhythmias, are irregularities in the heartbeat, including when it is too fast or too
slow. A resting heart rate that is too fast – above 100 beats per minute in adults – is
called tachycardia, and a resting heart rate that is too slow – below 60 beats per
minute – is called bradycardia.
Ventricular fibrillation (V-fib or VF) is an abnormal heart rhythm in which the
ventricles of the heart quiver. It is due to disorganized electrical activity. Ventricular
fibrillation results in cardiac arrest with loss of consciousness and no pulse. This is
followed by sudden cardiac death in the absence of treatment. Ventricular fibrillation
is initially found in about 10% of people with cardiac arrest.
Ventricular tachycardia (V-tach or VT) is a fast heart rate arising from the lower
chambers of the heart.Although a few seconds of VT may not result in permanent
problems, longer periods are dangerous; and multiple episodes over a short period of
time are referred to as an electrical storm. lightheadedness, palpitations, or chest pain.
Ventricular tachycardia may result in ventricular fibrillation (VF) and turn into
cardiac arrest. It is found initially in about 7% of people in cardiac arrest.
NEED OF DEFIBRILLATOR
Ventricular fibrillation is serious cardiac emergency resulting from
asynchronous contraction of heart muscle. This uncoordinated movement
of ventricle walls of the heart may result from electric shock or from
abnormalities of body chemistry.
Main problem of fibrillation is continuously stimulation of adjacent cells of
heart muscle fibers. So there is no synchronized succession of events that
follow heart action.
This Fibrillation leads to loss of cardiac output and irreversible brain damage
or death if not reversed within 5 minutes of onset
Ventricular fibrillation can be converted into more efficient rhythm by applying
a high energy shock to the heart. This sudden surge across the heart
cause all muscle fibers to contract simultaneously ,possibly the fibers
then respond to normal physiological intrinsic pulse.
What is Cardioversion
Defibrillator is a device that deliver a
therapeutic dose of electrical energy (electric
shock) to the affected heart (fibrillated heart
or other shockable rhythm) to force the heart
to produce normal cardiac rhythm.
It is any process that aims to convert an
arrhythmia back to sinus rhythm by
therapeutic dose of electrical energy
Unlike defibrillation, which is used in cardiac
arrest patients, synchronized cardioversion is
performed on patients that still have a pulse
What is defibrillator
Defibrillator
CARDIO
VERSION
DEFIBRILLATION VS. CARDIO
VERSION
WHAT IS
DEFIBRILLATOR
MANUAL
DEFIBRILLATOR
HOW DEFIBRILLATOR IS
USED
DC
DEFIBRILLATOR
Variable Auto transformer T1:
 Forms primary of high voltage of transformer T2.
 T1 variable autotransformer is usually designed with a significant number of
primary windings to produce a secondary voltage for T2 which can be
adjusted from a few volts
 Auto transformers ensure a consistent linear regulation of AC voltage to
provide high output and negligible waveform distortion.
Step-up transformer T2:
 It forms secondary of variable auto transformer T2.
 A transformer transfers power from the primary coil to the secondary coil.
 Since the power must stay the same, if the voltage increases, the current
must decrease. (P=VI)
Diode Rectifier:
The output voltage of the transformer is rectified and is connected
to
vacuum type high voltage change-over switch.
It converts an alternating current into a direct one by allowing a current to
flow through it in one direction only.
DC
DEFIBRILLATOR
Switch:
 It is a vacuum type high voltage change over switch.
 Its one end is connected to Diode rectifier at 1 and another to one end of an oil-filled
16 micro-farad capacitor.
 In position A the capacitor charges to a voltage set by the positioning of the auto
transformer.
 When the high voltage switch changes over the position 2, the capacitor is discharged
across the heart through the electrodes.
Capacitor:
 Oil filled 16 micro farad capacitor is used.
 During transthoracic defibrillation an energy storage capacitor is charged at relatively
slow rate in order of seconds by means of step-up transformer and then delivered
rapid rate in order of milliseconds to the chest of the patient.
Inductor:
 The rapid discharge rate impairs(weaken/damage) the contractibility of ventricles.
To overcome it, a current limiting inductor is inserted in series with the patient circuit.
The inductor slows down the discharge from the capacitor by the induced counter
voltage.
 This gives the output pulse a physiologically favorable shape.
DC
DEFIBRILLATOR
T2
T1
Block Diagram of DC defibrillator
POWER
SUPPLY
CHARGING
CIRCUIT
ENERGY
STORAGE
WAVEFORM
SHAPING
CHARGE
CONTROL
DISCHARGE
CONTROL
AC
POWER
SOURCE
USER INPUT
TO PATIENT
PADDLE
Energy: E=1/2 CV2
C= 16 microfarad
E= 400 Ws
V= 7000V with load resistance 50-200 ohms
For internal defibrillation E= 100Ws with load
resistance 25-50 ohms
Most of the defibrillator will deliver between
60% and 80% of their stored energy to a 50
ohms load.
• The shape of the waveform
that appears across electrodes
will depend
on the value of capacitor and
the
inductor used in the circuit (100mH).
• The discharge resistance which patient
represents for the defibrillating pulse
may be regarded as purely 50 to 100
ohms with electrode size 80cm2.
• The most common waveform utilized
in the RLC circuit employs an under
damped response with damping factor
less than unity.
• This particular waveform is called as
Lown waveform.
• This waveform is more or less of an
oscillatory character, with both
positive and negative portions.
LOWN
WAVEFORM
MONOPHASIC VS BIPHASIC DEFIBRILLATION
Terms
Monophasic Defibrillators Biphasic Defibrillators
Definition
The type waveform formed
during this process where the
shock gets delivered to the
heart with the help of just one
vector and hence gets the
name.
The type waveform formed
during this process where the
shock gets delivered to the
heart with the aid of two
vectors and hence gets the
name.
Usage
Different life-threatening cardiac
dysrhythmias and ventricular
fibrillation.
Various life-threatening
cardiac dysrhythmias and
ventricular fibrillation.
Devices Implantable Defibrillators
detects the irregular heart
rate.
External Defibrillators moves
the high-level values without
any problems
Nature Most common used process.
Most common available
process.
Energy On a monophasic defibrillator, this is
On a biphasic defibrillator, this
is usually between 120 joules
Defibrillator.pptx.                       Jei3j
DC
DEFIBRILLATOR
IMPLANTABLE
DEFIBRILLATOR
IMPLANTABLE
DEFIBRILLATOR
• The use of automatic implantable defibrillator (AID) is recommended for
patients who are at high risk for ventricular fibrillation.
• It used integrated circuits to reduce the device size.
• It has enhancing functionality of the device.
• It has ability to store recording of the patient’s heart
signals and collects extensive therapy history and
diagnostic data files.
• This device is hermetically(airtight) sealed, biocompatible,
and able to survive a temperature range of -30 degree Celsius
to 60 degree Celsius.
• This device allows the physician to non-invasively program it.
• It has sophisticated algorithms for rhythm classification and
storage of patient’s heart signal.
IMPLANTABLE
DEFIBRILLATOR
• An implantable defibrillator is continuously monitors a patient’s heart rhythm.
• If the device detects fibrillation, the capacitor within the device are charged up to 750
V.
• The capacitor then discharged into the heart which mostly represents a resistive load of
50 ohm and to bring the heart into normal rhythm.
• The shock duration is approximately 4-8 ms which results in the delivery of
approximately 30-35 J at 750 volts.
Implantable defibrillator system have main system component:
1. AID
2. The Lead system
3. The programmer recorder/monitor (PRM)
4. Pulse generator
5. Battery
IMPLANTABLE DEFIBRILLATOR SYSTEM
ARCHITECTURE
AID LEAD
SYSTEM
PULSE GENERATOR
BATTERY
IMPLANTABLE
COMPONENTS
AID
 It houses the power source , sensing, defibrillation,
pacing and telemetric communication system.
 The lead system provides physical and electrical
connection between the defibrillator and the heart tissue.
 The PRM communicates with the AID and allows the
physician to view status information and modify the
function of the device as needed.
Leads
 The defibrillating high energy pulse is delivered to the
heart via 6cm*9cm titanium mesh patch with
electrodes placed directly on the surface of heart.
 Electrode is placed invasively through leads screwed in
the heart.
 Leads are used for defibrillating and sensing.
IMPLANTABLE
COMPONENTS
Programmer Recorder/Monitor (PRM):
 The PRM is an external device that provides a bidirectional communications link to
an implanted AID.
 A telemetry link is established between PRM and AID.
 Telemetry channel is used to retrieve real time and stored intracardiac ECG, therapy
history, battery status and other information.
 It allows the programmer to change the algorithm as per the require for patients
abnormality.
Pulse generator:
 It has a microprocessor which controls overall system functions.
 ROM provides non-volatile memory for system start-up task and some program space
 RAM is required for storage of operating parameters and storage
of
electrocardiogram data.
 System Control supports the circuitry of microprocessor like a telemetry
interface,
typically implemented with a UART-like (universal
asynchronous receiver/transmitter) interface and general purpose timers.
Defibrillator.pptx.                       Jei3j
IMPLANTABLE
COMPONENTS
Battery:
 Two connected in series to form approximate 6 V Lithium Silver Vanadium
oxide (LiSVO)
 Separate voltage supplies are generally used for pacing(5V) i.e. low volt
supply and control of the charging circuit (10-15 V) i.e. high volt supply.
 High power circuits convert 3-6V battery voltage to the 750V necessary
for defibrillation.
 DC to DC convertor coverts 6 V battery to 750V, the output storage
capacitor and high output switch are used for defibrillation.
Sense Amplifier
 Proper sensing of heart requires precise sensing and discrimination
of each
of the component to differentiate tachycardia and ventricular
fibrillation.
 It should be able to sense 30 to 360 bpm.
 It should be able to quickly respond to widely varying signals
present during arrhythmia.
 Heat signals lies in the range of 100micoV to 20microV as a result
needs to
amplify to be processed further.
AID/
ICD
AI
D
AUTOMATIC OR ADVISORY EXTERNAL
DEFIBRILLATORS
 Smart automatic or advisory external defibrillator (AED) are capable of accurately
analyzing the ECG and making reliable shock decisions.
 The are designed to detect ventricular fibrillation with sensitivity and specificity
comparable to that of well trained paramedics then deliver(automatic) or recommend
(advisory) an appropriate high energy defibrillating shock.
 AED require self adhesive electrodes instead of hand held paddles.
 ECG signal are acquired are faster and more accurate and with less noise and higher
quality.
 Hands-off defibrillation is a safer procedure for the operator especially when operator
has no training.
 Multifunction self adhesive electrodes are commonly used with defibrillator-monitor-
pacer instruments.
 AED has the ability of the device to accurately assess the patient heart and make
therapy decision.
 The defibrillator performs this evaluation by sensing electrical signals from the
patient heart via electrodes and using a computerized algorithm to interpret the
electrical signals.
Defibrillator.pptx.                       Jei3j
DC Defibrillator with synchronization
AE
D
DEFIBRILLATOR
ELECTRODES
 Types of Defibrillator electrodes:-
 a) Spoon shaped electrode Applied directly to the heart.
 b)Paddle type electrode Applied against the chest wall
 c)Pad type electrode Applied directly on chest wall
 The electrodes for external defibrillation are metal discs about 3-5 cm in
diameter (or rectangular flat paddles 5x10 cm ) and attached to highly insulated
handle.
 The size of electrodes plays an important part in determining the chest wall
impedance which influence the efficiency of defibrillation.
 The capacitor is discharged only when the electrodes make a good and firm
contact with the chest of the patient.
 For internal defibrillation when the chest is open, large spoon- shaped
electrodes are used.
 Self adhesive pre-gelled electrodes :Self-adhering, solid gel construction ensures
easy use with no extra gel needed. Flexible design allows electrodes to better
conform to body contours.
Defibrillator.pptx.                       Jei3j
METAL DISC AND SPOON
ELECTRODE
DEFIBRILLATOR
ELECTRODE

More Related Content

PPTX
Animal Communication- Auditory and Visual.pptx
DOCX
Echolocation
PDF
Ants - Social organisation and types of nests
PPTX
Cardiac arrhythmia and anti arrhythmic drugs
PDF
Innate Behavior (Animal Behavior) Taxis, Fixed Action Pattern, reflex.
PPTX
Animal behaviour
PPTX
Transient response analysis
PPT
Power system contingencies
Animal Communication- Auditory and Visual.pptx
Echolocation
Ants - Social organisation and types of nests
Cardiac arrhythmia and anti arrhythmic drugs
Innate Behavior (Animal Behavior) Taxis, Fixed Action Pattern, reflex.
Animal behaviour
Transient response analysis
Power system contingencies

What's hot (12)

PDF
Ants
PPT
Animal Behavior
PPT
Magnetic field lines
PPTX
animal behaviours and its types with examples
DOCX
Animal behavior
PPTX
Radial vs Axial ventilation
DOCX
Biotelemetry
PPTX
Principles of electrocardiography ppt
PPT
Galvanometer
PPTX
Bee dance
PPTX
Territoiality in arthropoda
PPTX
Thermoreception in Insects
Ants
Animal Behavior
Magnetic field lines
animal behaviours and its types with examples
Animal behavior
Radial vs Axial ventilation
Biotelemetry
Principles of electrocardiography ppt
Galvanometer
Bee dance
Territoiality in arthropoda
Thermoreception in Insects
Ad

Similar to Defibrillator.pptx. Jei3j (20)

PPTX
Defibrillators
PPTX
Defibrillators
PPTX
Defibrillatorppt 131028115457-phpapp01
PPTX
Defibrillator (ppt)
PPTX
defibrillator.pptx
PPTX
Defibrillator
PPTX
Defibrillator and defibrillation presentation
PPTX
Bio-Medical Therapeutic of in Pacemaker& Respiratory
PPTX
DEFIBRILLATOR.akash pptttttttttttttttttt
PPT
Cardiovertor
PPTX
U5 - AVB, DEFIB, HOLTER,AVB,DEFIB,HOLTER.pptx
PDF
Medical Electronics - Hints for Slow Learner
PPTX
Defibrillator ppt.pptx
PPT
Defibrillation
PPTX
Defibrillater
PPT
Defibrillation.ppt3333333333333333333333
PDF
D & C 2012
PPTX
Defibrillator prakash
PPTX
Pacemaker_Presentation .pptx
PDF
Dc shock
Defibrillators
Defibrillators
Defibrillatorppt 131028115457-phpapp01
Defibrillator (ppt)
defibrillator.pptx
Defibrillator
Defibrillator and defibrillation presentation
Bio-Medical Therapeutic of in Pacemaker& Respiratory
DEFIBRILLATOR.akash pptttttttttttttttttt
Cardiovertor
U5 - AVB, DEFIB, HOLTER,AVB,DEFIB,HOLTER.pptx
Medical Electronics - Hints for Slow Learner
Defibrillator ppt.pptx
Defibrillation
Defibrillater
Defibrillation.ppt3333333333333333333333
D & C 2012
Defibrillator prakash
Pacemaker_Presentation .pptx
Dc shock
Ad

Recently uploaded (20)

PPTX
neonatal infection(7392992y282939y5.pptx
PDF
Medical Evidence in the Criminal Justice Delivery System in.pdf
PPTX
NEET PG 2025 Pharmacology Recall | Real Exam Questions from 3rd August with D...
PPTX
Imaging of parasitic D. Case Discussions.pptx
PPT
1b - INTRODUCTION TO EPIDEMIOLOGY (comm med).ppt
PPTX
CEREBROVASCULAR DISORDER.POWERPOINT PRESENTATIONx
PPTX
ACID BASE management, base deficit correction
DOC
Adobe Premiere Pro CC Crack With Serial Key Full Free Download 2025
PPTX
Neuropathic pain.ppt treatment managment
PDF
NEET PG 2025 | 200 High-Yield Recall Topics Across All Subjects
PPTX
Fundamentals of human energy transfer .pptx
PDF
Human Health And Disease hggyutgghg .pdf
PPT
CHAPTER FIVE. '' Association in epidemiological studies and potential errors
PPTX
JUVENILE NASOPHARYNGEAL ANGIOFIBROMA.pptx
PPTX
CME 2 Acute Chest Pain preentation for education
PPTX
Important Obstetric Emergency that must be recognised
DOCX
RUHS II MBBS Microbiology Paper-II with Answer Key | 6th August 2025 (New Sch...
PDF
Therapeutic Potential of Citrus Flavonoids in Metabolic Inflammation and Ins...
PPTX
1 General Principles of Radiotherapy.pptx
PDF
Handout_ NURS 220 Topic 10-Abnormal Pregnancy.pdf
neonatal infection(7392992y282939y5.pptx
Medical Evidence in the Criminal Justice Delivery System in.pdf
NEET PG 2025 Pharmacology Recall | Real Exam Questions from 3rd August with D...
Imaging of parasitic D. Case Discussions.pptx
1b - INTRODUCTION TO EPIDEMIOLOGY (comm med).ppt
CEREBROVASCULAR DISORDER.POWERPOINT PRESENTATIONx
ACID BASE management, base deficit correction
Adobe Premiere Pro CC Crack With Serial Key Full Free Download 2025
Neuropathic pain.ppt treatment managment
NEET PG 2025 | 200 High-Yield Recall Topics Across All Subjects
Fundamentals of human energy transfer .pptx
Human Health And Disease hggyutgghg .pdf
CHAPTER FIVE. '' Association in epidemiological studies and potential errors
JUVENILE NASOPHARYNGEAL ANGIOFIBROMA.pptx
CME 2 Acute Chest Pain preentation for education
Important Obstetric Emergency that must be recognised
RUHS II MBBS Microbiology Paper-II with Answer Key | 6th August 2025 (New Sch...
Therapeutic Potential of Citrus Flavonoids in Metabolic Inflammation and Ins...
1 General Principles of Radiotherapy.pptx
Handout_ NURS 220 Topic 10-Abnormal Pregnancy.pdf

Defibrillator.pptx. Jei3j

  • 2. UNIT OUTCOME 3 • Need of defibrillator • DC Defibrillator • AED (advisory external defibrillator ) • Implantable Defibrillator • Defibrillator electrodes
  • 5. Need of defibrillator Life-threatening cardiac arrhythmias specifically ventricular fibrillation (V-Fib) and non-perfusing ventricular tachycardia (V-Tach). Arrhythmias, also known as cardiac arrhythmias, heart arrhythmias, or dysrhythmias, are irregularities in the heartbeat, including when it is too fast or too slow. A resting heart rate that is too fast – above 100 beats per minute in adults – is called tachycardia, and a resting heart rate that is too slow – below 60 beats per minute – is called bradycardia. Ventricular fibrillation (V-fib or VF) is an abnormal heart rhythm in which the ventricles of the heart quiver. It is due to disorganized electrical activity. Ventricular fibrillation results in cardiac arrest with loss of consciousness and no pulse. This is followed by sudden cardiac death in the absence of treatment. Ventricular fibrillation is initially found in about 10% of people with cardiac arrest. Ventricular tachycardia (V-tach or VT) is a fast heart rate arising from the lower chambers of the heart.Although a few seconds of VT may not result in permanent problems, longer periods are dangerous; and multiple episodes over a short period of time are referred to as an electrical storm. lightheadedness, palpitations, or chest pain. Ventricular tachycardia may result in ventricular fibrillation (VF) and turn into cardiac arrest. It is found initially in about 7% of people in cardiac arrest.
  • 6. NEED OF DEFIBRILLATOR Ventricular fibrillation is serious cardiac emergency resulting from asynchronous contraction of heart muscle. This uncoordinated movement of ventricle walls of the heart may result from electric shock or from abnormalities of body chemistry. Main problem of fibrillation is continuously stimulation of adjacent cells of heart muscle fibers. So there is no synchronized succession of events that follow heart action. This Fibrillation leads to loss of cardiac output and irreversible brain damage or death if not reversed within 5 minutes of onset Ventricular fibrillation can be converted into more efficient rhythm by applying a high energy shock to the heart. This sudden surge across the heart cause all muscle fibers to contract simultaneously ,possibly the fibers then respond to normal physiological intrinsic pulse.
  • 7. What is Cardioversion Defibrillator is a device that deliver a therapeutic dose of electrical energy (electric shock) to the affected heart (fibrillated heart or other shockable rhythm) to force the heart to produce normal cardiac rhythm. It is any process that aims to convert an arrhythmia back to sinus rhythm by therapeutic dose of electrical energy Unlike defibrillation, which is used in cardiac arrest patients, synchronized cardioversion is performed on patients that still have a pulse What is defibrillator
  • 14. DC DEFIBRILLATOR Variable Auto transformer T1:  Forms primary of high voltage of transformer T2.  T1 variable autotransformer is usually designed with a significant number of primary windings to produce a secondary voltage for T2 which can be adjusted from a few volts  Auto transformers ensure a consistent linear regulation of AC voltage to provide high output and negligible waveform distortion. Step-up transformer T2:  It forms secondary of variable auto transformer T2.  A transformer transfers power from the primary coil to the secondary coil.  Since the power must stay the same, if the voltage increases, the current must decrease. (P=VI) Diode Rectifier: The output voltage of the transformer is rectified and is connected to vacuum type high voltage change-over switch. It converts an alternating current into a direct one by allowing a current to flow through it in one direction only.
  • 15. DC DEFIBRILLATOR Switch:  It is a vacuum type high voltage change over switch.  Its one end is connected to Diode rectifier at 1 and another to one end of an oil-filled 16 micro-farad capacitor.  In position A the capacitor charges to a voltage set by the positioning of the auto transformer.  When the high voltage switch changes over the position 2, the capacitor is discharged across the heart through the electrodes. Capacitor:  Oil filled 16 micro farad capacitor is used.  During transthoracic defibrillation an energy storage capacitor is charged at relatively slow rate in order of seconds by means of step-up transformer and then delivered rapid rate in order of milliseconds to the chest of the patient. Inductor:  The rapid discharge rate impairs(weaken/damage) the contractibility of ventricles. To overcome it, a current limiting inductor is inserted in series with the patient circuit. The inductor slows down the discharge from the capacitor by the induced counter voltage.  This gives the output pulse a physiologically favorable shape.
  • 17. Block Diagram of DC defibrillator POWER SUPPLY CHARGING CIRCUIT ENERGY STORAGE WAVEFORM SHAPING CHARGE CONTROL DISCHARGE CONTROL AC POWER SOURCE USER INPUT TO PATIENT PADDLE
  • 18. Energy: E=1/2 CV2 C= 16 microfarad E= 400 Ws V= 7000V with load resistance 50-200 ohms For internal defibrillation E= 100Ws with load resistance 25-50 ohms Most of the defibrillator will deliver between 60% and 80% of their stored energy to a 50 ohms load.
  • 19. • The shape of the waveform that appears across electrodes will depend on the value of capacitor and the inductor used in the circuit (100mH). • The discharge resistance which patient represents for the defibrillating pulse may be regarded as purely 50 to 100 ohms with electrode size 80cm2. • The most common waveform utilized in the RLC circuit employs an under damped response with damping factor less than unity. • This particular waveform is called as Lown waveform. • This waveform is more or less of an oscillatory character, with both positive and negative portions. LOWN WAVEFORM
  • 20. MONOPHASIC VS BIPHASIC DEFIBRILLATION Terms Monophasic Defibrillators Biphasic Defibrillators Definition The type waveform formed during this process where the shock gets delivered to the heart with the help of just one vector and hence gets the name. The type waveform formed during this process where the shock gets delivered to the heart with the aid of two vectors and hence gets the name. Usage Different life-threatening cardiac dysrhythmias and ventricular fibrillation. Various life-threatening cardiac dysrhythmias and ventricular fibrillation. Devices Implantable Defibrillators detects the irregular heart rate. External Defibrillators moves the high-level values without any problems Nature Most common used process. Most common available process. Energy On a monophasic defibrillator, this is On a biphasic defibrillator, this is usually between 120 joules
  • 24. IMPLANTABLE DEFIBRILLATOR • The use of automatic implantable defibrillator (AID) is recommended for patients who are at high risk for ventricular fibrillation. • It used integrated circuits to reduce the device size. • It has enhancing functionality of the device. • It has ability to store recording of the patient’s heart signals and collects extensive therapy history and diagnostic data files. • This device is hermetically(airtight) sealed, biocompatible, and able to survive a temperature range of -30 degree Celsius to 60 degree Celsius. • This device allows the physician to non-invasively program it. • It has sophisticated algorithms for rhythm classification and storage of patient’s heart signal.
  • 25. IMPLANTABLE DEFIBRILLATOR • An implantable defibrillator is continuously monitors a patient’s heart rhythm. • If the device detects fibrillation, the capacitor within the device are charged up to 750 V. • The capacitor then discharged into the heart which mostly represents a resistive load of 50 ohm and to bring the heart into normal rhythm. • The shock duration is approximately 4-8 ms which results in the delivery of approximately 30-35 J at 750 volts. Implantable defibrillator system have main system component: 1. AID 2. The Lead system 3. The programmer recorder/monitor (PRM) 4. Pulse generator 5. Battery
  • 26. IMPLANTABLE DEFIBRILLATOR SYSTEM ARCHITECTURE AID LEAD SYSTEM PULSE GENERATOR BATTERY
  • 27. IMPLANTABLE COMPONENTS AID  It houses the power source , sensing, defibrillation, pacing and telemetric communication system.  The lead system provides physical and electrical connection between the defibrillator and the heart tissue.  The PRM communicates with the AID and allows the physician to view status information and modify the function of the device as needed. Leads  The defibrillating high energy pulse is delivered to the heart via 6cm*9cm titanium mesh patch with electrodes placed directly on the surface of heart.  Electrode is placed invasively through leads screwed in the heart.  Leads are used for defibrillating and sensing.
  • 28. IMPLANTABLE COMPONENTS Programmer Recorder/Monitor (PRM):  The PRM is an external device that provides a bidirectional communications link to an implanted AID.  A telemetry link is established between PRM and AID.  Telemetry channel is used to retrieve real time and stored intracardiac ECG, therapy history, battery status and other information.  It allows the programmer to change the algorithm as per the require for patients abnormality. Pulse generator:  It has a microprocessor which controls overall system functions.  ROM provides non-volatile memory for system start-up task and some program space  RAM is required for storage of operating parameters and storage of electrocardiogram data.  System Control supports the circuitry of microprocessor like a telemetry interface, typically implemented with a UART-like (universal asynchronous receiver/transmitter) interface and general purpose timers.
  • 30. IMPLANTABLE COMPONENTS Battery:  Two connected in series to form approximate 6 V Lithium Silver Vanadium oxide (LiSVO)  Separate voltage supplies are generally used for pacing(5V) i.e. low volt supply and control of the charging circuit (10-15 V) i.e. high volt supply.  High power circuits convert 3-6V battery voltage to the 750V necessary for defibrillation.  DC to DC convertor coverts 6 V battery to 750V, the output storage capacitor and high output switch are used for defibrillation. Sense Amplifier  Proper sensing of heart requires precise sensing and discrimination of each of the component to differentiate tachycardia and ventricular fibrillation.  It should be able to sense 30 to 360 bpm.  It should be able to quickly respond to widely varying signals present during arrhythmia.  Heat signals lies in the range of 100micoV to 20microV as a result needs to amplify to be processed further.
  • 32. AI D
  • 33. AUTOMATIC OR ADVISORY EXTERNAL DEFIBRILLATORS  Smart automatic or advisory external defibrillator (AED) are capable of accurately analyzing the ECG and making reliable shock decisions.  The are designed to detect ventricular fibrillation with sensitivity and specificity comparable to that of well trained paramedics then deliver(automatic) or recommend (advisory) an appropriate high energy defibrillating shock.  AED require self adhesive electrodes instead of hand held paddles.  ECG signal are acquired are faster and more accurate and with less noise and higher quality.  Hands-off defibrillation is a safer procedure for the operator especially when operator has no training.  Multifunction self adhesive electrodes are commonly used with defibrillator-monitor- pacer instruments.  AED has the ability of the device to accurately assess the patient heart and make therapy decision.  The defibrillator performs this evaluation by sensing electrical signals from the patient heart via electrodes and using a computerized algorithm to interpret the electrical signals.
  • 35. DC Defibrillator with synchronization
  • 36. AE D
  • 37. DEFIBRILLATOR ELECTRODES  Types of Defibrillator electrodes:-  a) Spoon shaped electrode Applied directly to the heart.  b)Paddle type electrode Applied against the chest wall  c)Pad type electrode Applied directly on chest wall  The electrodes for external defibrillation are metal discs about 3-5 cm in diameter (or rectangular flat paddles 5x10 cm ) and attached to highly insulated handle.  The size of electrodes plays an important part in determining the chest wall impedance which influence the efficiency of defibrillation.  The capacitor is discharged only when the electrodes make a good and firm contact with the chest of the patient.  For internal defibrillation when the chest is open, large spoon- shaped electrodes are used.  Self adhesive pre-gelled electrodes :Self-adhering, solid gel construction ensures easy use with no extra gel needed. Flexible design allows electrodes to better conform to body contours.
  • 39. METAL DISC AND SPOON ELECTRODE

Editor's Notes