• The radiographic grid consists of a series of Lead foil strips
separated by X-ray transparent spacers .
• Invented by Dr.Gustave Bucky in 1913
• Most effective way of removing scatter radiation from large
radiographic fields .
• The interspaces of grids are filled either with aluminium or some
organic compound .
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
GRID RATIO : Ratio between the height of the Lead strips and
the distance between them .
• GRID PATTERN : Orientation of the Lead strips in their
longitudinal axis .
• Pattern of the grid that we see from as a view from the top
I. LINEAR GRID
II. CROSSED GRID
III. PARALLEL GRID
IV. FOCUSSED GRID
LINEAR GRID
The Lead strips are parallel to each other in their longitudinal axis
Major advantage is that they allow us to angle the X-ray tube
along the length of the Grid without loss of primary radiation
from grid ‘cutoff ‘
CROSSED GRID
Made of two superimposed linear grids that have the same
focusing distance .
Grid ratio of two crossed grids is equal to the sum of the ratio of
the two linear grids .
Cannot be used with oblique techniques requiring angulation of
the Xray tube .
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
FOCUSSED GRID
Grid made up of lead strips that are slightly angled so that they
focus in space .
May be either linear or crossed .
Line focused grids converge at a line in space called CONVERGENT
LINE .
Crossed Grids converge in a point in space called CONVERGENT
POINT .
FOCAL DISTANCE : Perpendicular distance between the grid and
the convergent line or point .
PARALLEL GRID
Lead strips are parallel when viewed in cross section .
Focused at infinity so they do not have a convergent line .
Can only be used effectively with either very small Xray fields
or long target-grid distances .
LINES PER INCH : Number of Lead strips per inch of the grid .
Calculated by adding the thickness of the Lead strips and
interspaces and dividing this sum into 1 .
• Three main methods :
I. PRIMARY TRANSMISSION (Tp)
II. BUCKY FACTOR (B)
III. CONTRAST IMPROVEMENT FACTOR (K)
• Measurement of the percentage of primary radiation
transmitted through a grid .
1. Made with the grid in place to determine the intensity of the
radiation through the grid .
2. Made after removal of the grid to determine the intensity of
the radiation directed at the grid .
Tp = Ip/Ip’ x 100
• Ratio of the incident radiation on the grid to the transmitted
radiation passing through the grid .
• Indicates the amount of exposure factors that should be
increased when the change from grid to a non grid technique is
made .
• Measure of the Grid’s ability to absorb scatter radiation
• Unlike primary transmission , Bucky factor indicates the
absorption of both primary and secondary radiation .
B= INCIDENT RADIATION / TRANSMITTED RADIATION
• Ratio of the contrast with a grid to the contrast without a grid .
• Measure of the grid’s ability to improve contrast which is its
primary function .
• It depends on :
1. kVp
2. Field size
3. Phantom thickness
• More closely related to the Lead content of the grid than any
other factor . (g/cm²)
• Loss of primary radiation that occurs when the images of the
Lead strips are projected wider than they would be with
ordinary magnification .
• Result of a poor geometeric relationship between the primary
beam and the Lead foil strips of the grid .
• Cut off is complete and no primary radiation reaches the film
when the projected images of the Lead strips are thicker than
the width of the interspaces.
• Amount of cut off is always greatest with high ratio grids and
short grid focus distances .
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
• 4 SITUATIONS THAT PRODUCE GRID CUT OFF :
1. FOCUSED GRIDS USED UPSIDE DOWN
2. LATERAL DECENTERING (GRID ANGULATION )
3. FOCUS GRID DISTANCE DECENTERING
4. COMBINED LATERAL AND FOCUS-GRID DISTANCE
DECENTERING .
• When a focused grid is used upside down, there is severe
peripheral cut off with a dark band of exposure in the center
of the film with no exposure at the periphery .
• The higher the grid ratio,narrower the exposed area .
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
• When the X-ray tube is positioned lateral to the convergent line
but at the correct focal distance.
• Uniform loss of radiation over the entire surface of the grid,
producing a uniformly light radiograph .
• 3 factors affect the magnitude of cut off from lateral
decentering :
-Grid ratio
-Focal distance
-Amount of decentering
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
OFF LEVEL GRIDS
When a linear grid is tilted , there is uniform loss
of primary radiation across the entire surface of
the grid .
Effect on the film being same as that of lateral
decentering .
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
• The target of the X-ray tube is correctly centered to the grid ,
but it is positioned above or below the convergent line .
• Cut off is greater with near than far focus –grid distance
decentering
• The central portion of the film isnt affected but the periphery is
light .
• The loss of primary radiation is directly proportional to the grid
ratio and the distance from the center line.
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
• Causes an uneven exposure resulting in a film that is light on
one side and dark on the other side.
• Directly proportional to the grid ratio and decentering distance
and inversely proportional to the focal distance of the grid .
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
• Invented by Dr.Hollis.E.Potter in 1920 .
• Also called a Bucky grid
• Moved to blur out the shadows cast by the lead strips .
• Continuously moves 1 to 3 cm back and forth throughout the
exposure .
• Precautions :
1. The grid must move fast enough to blue the lead strips
2. The transverse motion of the grid should be synchronous with
the pulses of the Xray generator .
DISADVANTAGES :
• Costly
• Subject to failure
• May vibrate the Xray table
• Places a limit on the minimum exposure time because they move
slowly .
• INCREASES THE PATIENT RADIATION DOSE.
• The price of increased “cleanup “ with high ratio grids is that
patient exposure is considerably increased and that Xray tube
centering becomes critical .
• 8:1 grids will give adequate results below 90 kVp
• Above 90 kvp,12:1 grids are preferred
• In cases of Biplane cerebral angiography, crossed grids are
preferred as there is a great deal of scatter radiation.
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
• Scatter radiation arising from the patient from Compton
reactions disperses in all directions , so the patient acts like a
large light bulb .
• The closer the patient is to the film, the greater the
concentration of scatter per unit area .
• With an air gap, the concentration decreases due to more
photons missing the film in the gap .
• Used in Magnification radiography and Chest radiography .
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
1. Thicker the part,the more advantageous a larger air gap
2. The first inch of any air gap improves contrast more than any
subsequent inch
3. Image sharpness deteriorates with increasing gap width unless
the focal film distance is increased to compensate for the
greater magnification.
4. If the gap is widened by moving the patient away from the
film with a fixed focal film distance the patient is closer to the
X-ray tube and his exposure increases.
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT

More Related Content

PDF
Grids
PPTX
8. GRIDS slide gkjhgkjhvukgvjhbkuyvkbliuh
PPT
PPTX
Grids and Types Of Grids and Airgap gap technique
PPTX
Radiographic grid uses and its types .pptx
PPTX
grids (4).pptxdfnirni3nijfnrijfncvrijfvifn
PPTX
XRAY GRIDS (1).pptx
PPTX
Filters grids and collimators
Grids
8. GRIDS slide gkjhgkjhvukgvjhbkuyvkbliuh
Grids and Types Of Grids and Airgap gap technique
Radiographic grid uses and its types .pptx
grids (4).pptxdfnirni3nijfnrijfncvrijfvifn
XRAY GRIDS (1).pptx
Filters grids and collimators

Similar to GRIDS IS PRESENTATION IN RADIO PHYSICS MUST IMPORTANT (20)

PPTX
Radiographic grids
PPTX
grid.pptx
PPT
Grids JP .ppt
PPTX
PPTX
Filters grids and collimators
PPTX
Grid.pptx radio physics presentation 123
PPT
Rad 206 p12 Fundamentals of Imaging - Control of Scatter Radiation
PPTX
Filters used in radiology.ppt.radiology.
PPTX
grids,films.pptx
PPTX
Radiographic grid Swastik
PPTX
Neelotpal Mitra
PPTX
GRIDS_TYPES.pptx
PPTX
6 Lecture 6 Control of scatter radiation.pptx
PPT
GRIDS.ppt
PPTX
Beam restriction devices in x rays by sushil pattar
PPTX
X-RAY GRID BUCKY
PPTX
X-ray_Grids.pptx how grid is effective to reduce scattered radiation?
PPT
PPT
Xray filters beam restrictors and Grids Radiodiagnosis.ppt
PPTX
Radiographic grids
grid.pptx
Grids JP .ppt
Filters grids and collimators
Grid.pptx radio physics presentation 123
Rad 206 p12 Fundamentals of Imaging - Control of Scatter Radiation
Filters used in radiology.ppt.radiology.
grids,films.pptx
Radiographic grid Swastik
Neelotpal Mitra
GRIDS_TYPES.pptx
6 Lecture 6 Control of scatter radiation.pptx
GRIDS.ppt
Beam restriction devices in x rays by sushil pattar
X-RAY GRID BUCKY
X-ray_Grids.pptx how grid is effective to reduce scattered radiation?
Xray filters beam restrictors and Grids Radiodiagnosis.ppt
Ad

Recently uploaded (20)

PPTX
Anatomy and physiology of the digestive system
PDF
Intl J Gynecology Obste - 2021 - Melamed - FIGO International Federation o...
PPT
Obstructive sleep apnea in orthodontics treatment
PPTX
ONCOLOGY Principles of Radiotherapy.pptx
PPTX
surgery guide for USMLE step 2-part 1.pptx
PDF
Oral Aspect of Metabolic Disease_20250717_192438_0000.pdf
PPTX
NASO ALVEOLAR MOULDNIG IN CLEFT LIP AND PALATE PATIENT
PPT
neurology Member of Royal College of Physicians (MRCP).ppt
PPT
Copy-Histopathology Practical by CMDA ESUTH CHAPTER(0) - Copy.ppt
PPTX
PRESENTACION DE TRAUMA CRANEAL, CAUSAS, CONSEC, ETC.
PPTX
Morphology of Bacterial Cell for bsc sud
PPTX
1. Basic chemist of Biomolecule (1).pptx
PPTX
the psycho-oncology for psychiatrists pptx
PPT
Infections Member of Royal College of Physicians.ppt
PPT
STD NOTES INTRODUCTION TO COMMUNITY HEALT STRATEGY.ppt
PDF
Copy of OB - Exam #2 Study Guide. pdf
PPTX
Post Op complications in general surgery
PPTX
Human Reproduction: Anatomy, Physiology & Clinical Insights.pptx
PPTX
ANATOMY OF MEDULLA OBLANGATA AND SYNDROMES.pptx
PPTX
Clinical approach and Radiotherapy principles.pptx
Anatomy and physiology of the digestive system
Intl J Gynecology Obste - 2021 - Melamed - FIGO International Federation o...
Obstructive sleep apnea in orthodontics treatment
ONCOLOGY Principles of Radiotherapy.pptx
surgery guide for USMLE step 2-part 1.pptx
Oral Aspect of Metabolic Disease_20250717_192438_0000.pdf
NASO ALVEOLAR MOULDNIG IN CLEFT LIP AND PALATE PATIENT
neurology Member of Royal College of Physicians (MRCP).ppt
Copy-Histopathology Practical by CMDA ESUTH CHAPTER(0) - Copy.ppt
PRESENTACION DE TRAUMA CRANEAL, CAUSAS, CONSEC, ETC.
Morphology of Bacterial Cell for bsc sud
1. Basic chemist of Biomolecule (1).pptx
the psycho-oncology for psychiatrists pptx
Infections Member of Royal College of Physicians.ppt
STD NOTES INTRODUCTION TO COMMUNITY HEALT STRATEGY.ppt
Copy of OB - Exam #2 Study Guide. pdf
Post Op complications in general surgery
Human Reproduction: Anatomy, Physiology & Clinical Insights.pptx
ANATOMY OF MEDULLA OBLANGATA AND SYNDROMES.pptx
Clinical approach and Radiotherapy principles.pptx
Ad

GRIDS IS PRESENTATION IN RADIO PHYSICS MUST IMPORTANT

  • 2. • The radiographic grid consists of a series of Lead foil strips separated by X-ray transparent spacers . • Invented by Dr.Gustave Bucky in 1913 • Most effective way of removing scatter radiation from large radiographic fields . • The interspaces of grids are filled either with aluminium or some organic compound .
  • 4. GRID RATIO : Ratio between the height of the Lead strips and the distance between them .
  • 5. • GRID PATTERN : Orientation of the Lead strips in their longitudinal axis . • Pattern of the grid that we see from as a view from the top I. LINEAR GRID II. CROSSED GRID III. PARALLEL GRID IV. FOCUSSED GRID
  • 6. LINEAR GRID The Lead strips are parallel to each other in their longitudinal axis Major advantage is that they allow us to angle the X-ray tube along the length of the Grid without loss of primary radiation from grid ‘cutoff ‘ CROSSED GRID Made of two superimposed linear grids that have the same focusing distance . Grid ratio of two crossed grids is equal to the sum of the ratio of the two linear grids . Cannot be used with oblique techniques requiring angulation of the Xray tube .
  • 9. FOCUSSED GRID Grid made up of lead strips that are slightly angled so that they focus in space . May be either linear or crossed . Line focused grids converge at a line in space called CONVERGENT LINE . Crossed Grids converge in a point in space called CONVERGENT POINT . FOCAL DISTANCE : Perpendicular distance between the grid and the convergent line or point .
  • 10. PARALLEL GRID Lead strips are parallel when viewed in cross section . Focused at infinity so they do not have a convergent line . Can only be used effectively with either very small Xray fields or long target-grid distances .
  • 11. LINES PER INCH : Number of Lead strips per inch of the grid . Calculated by adding the thickness of the Lead strips and interspaces and dividing this sum into 1 .
  • 12. • Three main methods : I. PRIMARY TRANSMISSION (Tp) II. BUCKY FACTOR (B) III. CONTRAST IMPROVEMENT FACTOR (K)
  • 13. • Measurement of the percentage of primary radiation transmitted through a grid . 1. Made with the grid in place to determine the intensity of the radiation through the grid . 2. Made after removal of the grid to determine the intensity of the radiation directed at the grid . Tp = Ip/Ip’ x 100
  • 14. • Ratio of the incident radiation on the grid to the transmitted radiation passing through the grid . • Indicates the amount of exposure factors that should be increased when the change from grid to a non grid technique is made . • Measure of the Grid’s ability to absorb scatter radiation • Unlike primary transmission , Bucky factor indicates the absorption of both primary and secondary radiation . B= INCIDENT RADIATION / TRANSMITTED RADIATION
  • 15. • Ratio of the contrast with a grid to the contrast without a grid . • Measure of the grid’s ability to improve contrast which is its primary function . • It depends on : 1. kVp 2. Field size 3. Phantom thickness • More closely related to the Lead content of the grid than any other factor . (g/cm²)
  • 16. • Loss of primary radiation that occurs when the images of the Lead strips are projected wider than they would be with ordinary magnification . • Result of a poor geometeric relationship between the primary beam and the Lead foil strips of the grid . • Cut off is complete and no primary radiation reaches the film when the projected images of the Lead strips are thicker than the width of the interspaces. • Amount of cut off is always greatest with high ratio grids and short grid focus distances .
  • 18. • 4 SITUATIONS THAT PRODUCE GRID CUT OFF : 1. FOCUSED GRIDS USED UPSIDE DOWN 2. LATERAL DECENTERING (GRID ANGULATION ) 3. FOCUS GRID DISTANCE DECENTERING 4. COMBINED LATERAL AND FOCUS-GRID DISTANCE DECENTERING .
  • 19. • When a focused grid is used upside down, there is severe peripheral cut off with a dark band of exposure in the center of the film with no exposure at the periphery . • The higher the grid ratio,narrower the exposed area .
  • 21. • When the X-ray tube is positioned lateral to the convergent line but at the correct focal distance. • Uniform loss of radiation over the entire surface of the grid, producing a uniformly light radiograph . • 3 factors affect the magnitude of cut off from lateral decentering : -Grid ratio -Focal distance -Amount of decentering
  • 23. OFF LEVEL GRIDS When a linear grid is tilted , there is uniform loss of primary radiation across the entire surface of the grid . Effect on the film being same as that of lateral decentering .
  • 25. • The target of the X-ray tube is correctly centered to the grid , but it is positioned above or below the convergent line . • Cut off is greater with near than far focus –grid distance decentering • The central portion of the film isnt affected but the periphery is light . • The loss of primary radiation is directly proportional to the grid ratio and the distance from the center line.
  • 28. • Causes an uneven exposure resulting in a film that is light on one side and dark on the other side. • Directly proportional to the grid ratio and decentering distance and inversely proportional to the focal distance of the grid .
  • 30. • Invented by Dr.Hollis.E.Potter in 1920 . • Also called a Bucky grid • Moved to blur out the shadows cast by the lead strips . • Continuously moves 1 to 3 cm back and forth throughout the exposure . • Precautions : 1. The grid must move fast enough to blue the lead strips 2. The transverse motion of the grid should be synchronous with the pulses of the Xray generator .
  • 31. DISADVANTAGES : • Costly • Subject to failure • May vibrate the Xray table • Places a limit on the minimum exposure time because they move slowly . • INCREASES THE PATIENT RADIATION DOSE.
  • 32. • The price of increased “cleanup “ with high ratio grids is that patient exposure is considerably increased and that Xray tube centering becomes critical . • 8:1 grids will give adequate results below 90 kVp • Above 90 kvp,12:1 grids are preferred • In cases of Biplane cerebral angiography, crossed grids are preferred as there is a great deal of scatter radiation.
  • 34. • Scatter radiation arising from the patient from Compton reactions disperses in all directions , so the patient acts like a large light bulb . • The closer the patient is to the film, the greater the concentration of scatter per unit area . • With an air gap, the concentration decreases due to more photons missing the film in the gap . • Used in Magnification radiography and Chest radiography .
  • 36. 1. Thicker the part,the more advantageous a larger air gap 2. The first inch of any air gap improves contrast more than any subsequent inch 3. Image sharpness deteriorates with increasing gap width unless the focal film distance is increased to compensate for the greater magnification. 4. If the gap is widened by moving the patient away from the film with a fixed focal film distance the patient is closer to the X-ray tube and his exposure increases.