HPLC: Principle and Maintenance with Application
Yogesh Kumar
Research Scholar,
Delhi Pharmaceutical Sciences
ABSTRACT
High performance liquid chromatography (HPLC) is a
significant qualitative and quantitative technique,
usually used for the estimation of pharmaceutical and
biological samples. The chromatography term is
derived from the Greek words namely chroma
(colour) and graphein (to write). The chromatography
is very accepted technique and it is mostly used
analytically. It is the most resourceful, safest, reliable
and fastest chromatographic technique for the quality
control of drug components. This technique involves
2 phases’ stationary and mobile phases. There are
different types of chromatographic techniques. The
separation of constituents is based on the variation
between the partition coefficients of the two phases.
This article is primed with an aim to review different
aspects of HPLC, such as principle, types,
instrumentation and application with maintenance.
KEYWORD: HPLC; Instrumentation; maintenance;
application; chromatography
I. INTRODUCTION
Analytical chemistry is widely used in determining
the qualitative and quantitative composition of typical
mixtures. These two methods are the main aspects to
understand the sample materials. Generally, analytical
chemistry divides into two branches i.e. q
and quantitative analysis. High performance liquid
chromatography also is known as High Pressure
Liquid Chromatography. HPLC is an analysis
technique used for separation, identification and
quantification of typical mixtures such as organic,
inorganic, biological, ionic and polymeric materials.
HPLC is a type of column chromatography in which
solvent flows with high pressure, so that sample can
be separated into different constituents divided into
different types based on modes of separation, the
@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018
ISSN No: 2456 - 6470 | www.ijtsrd.com | Volume
International Journal of Trend in Scientific
Research and Development (IJTSRD)
International Open Access Journal
HPLC: Principle and Maintenance with Application
Yogesh Kumar, Sayed Md Mumtaz, Mustaq Ahmad
search Scholar, Pharmacology Department,
Delhi Pharmaceutical Sciences and Research University Delhi, Delhi, India
High performance liquid chromatography (HPLC) is a
significant qualitative and quantitative technique,
usually used for the estimation of pharmaceutical and
chromatography term is
derived from the Greek words namely chroma
(colour) and graphein (to write). The chromatography
is very accepted technique and it is mostly used
analytically. It is the most resourceful, safest, reliable
echnique for the quality
control of drug components. This technique involves
2 phases’ stationary and mobile phases. There are
different types of chromatographic techniques. The
separation of constituents is based on the variation
fficients of the two phases.
This article is primed with an aim to review different
aspects of HPLC, such as principle, types,
instrumentation and application with maintenance.
HPLC; Instrumentation; maintenance;
Analytical chemistry is widely used in determining
the qualitative and quantitative composition of typical
mixtures. These two methods are the main aspects to
understand the sample materials. Generally, analytical
chemistry divides into two branches i.e. qualitative
and quantitative analysis. High performance liquid
chromatography also is known as High Pressure
Liquid Chromatography. HPLC is an analysis
technique used for separation, identification and
quantification of typical mixtures such as organic,
rganic, biological, ionic and polymeric materials.
HPLC is a type of column chromatography in which
solvent flows with high pressure, so that sample can
be separated into different constituents divided into
different types based on modes of separation, the
principle of separation, elution technique, scale of
separation and based on the type of analysis
Detector is the heart of this instrument and the
efficiency of this system depends upon detecting
techniques. Many types of HPLC detectors such as
UV-Vs, refractive index detector, photodiode
detector, fluorescence, electrical conductivity
detector.
A. Chromatography:
Chromatography is an analytical method in which
compounds are physically separated prior to
measurement. The main purpose of chromatography
is to separate and quantify the target sample in the
matrix. The first developer of
Tswett.
B. Scope of HPLC:
HPLC is used in various fields like in
pharmaceuticals, biochemicals, food products,
industrial chemicals, forensic chemistry,
environmental field, clinical medicine for analysis for
typical mixtures like antibiotics, amino acids, fatty
acids, drugs, poisons, inorganic ions, urine extracts,
oestrogens [2, 3].
C. Need of HPLC
The use of HPLC are increased day by day across the
world due to its unique properties like high
resolutions, high sensitivity (ppm
repeatability, small sample size, moderate analysis
condition, no need to vaporise the sample as in the gas
chromatography, easy to fractionate the sample and
purify [4, 5].
Aug 2018 Page: 1618
6470 | www.ijtsrd.com | Volume - 2 | Issue – 5
Scientific
(IJTSRD)
International Open Access Journal
HPLC: Principle and Maintenance with Application
, India
principle of separation, elution technique, scale of
separation and based on the type of analysis [1]. The
Detector is the heart of this instrument and the
efficiency of this system depends upon detecting
techniques. Many types of HPLC detectors such as
Vs, refractive index detector, photodiode
detector, fluorescence, electrical conductivity
Chromatography is an analytical method in which
compounds are physically separated prior to
measurement. The main purpose of chromatography
is to separate and quantify the target sample in the
matrix. The first developer of chromatography was M.
HPLC is used in various fields like in
pharmaceuticals, biochemicals, food products,
industrial chemicals, forensic chemistry,
environmental field, clinical medicine for analysis for
ntibiotics, amino acids, fatty
acids, drugs, poisons, inorganic ions, urine extracts,
The use of HPLC are increased day by day across the
world due to its unique properties like high
resolutions, high sensitivity (ppm-ppb), good
repeatability, small sample size, moderate analysis
condition, no need to vaporise the sample as in the gas
atography, easy to fractionate the sample and
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1619
D. Instrumentation:
The HPLC instrumentation involves pump, injector,
column, detector, and integrator and display system.
 Solvent reservoir: In HPLC the mobile phase or
solvent is a mixture of polar and non-polar liquid
components. Depending on the composition of the
sample, the polar and non-polar solvents will be
varied [6].
 Pump: The pump suctions the mobile phase from
the solvent reservoir and forces it to the column
and then passes to the detector. 42000 KPa is the
operating pressure of the pump. This operating
pressure depends on column dimensions, particle
size, flow rate and composition of mobile phase
[7].
 Sample Injector: An injector for an HPLC
framework should give an infusion of the fluid
specimen inside the scope of 0.1 mL to 100 mL of
volume with high reproducibility and under high
pressure (up to 4000 psi).
 Columns: Columns are typically made of cleaned
stainless steel, are somewhere around 50 mm and
300 mm long and have an inward distance across
of somewhere around 2 and 5 mm. They are
generally loaded with a stationary phase with a
molecule size of 3 μm to 10 μm [8].
 Detector: The HPLC detector, situated toward the
end of the column distinguishes the analytes as
they elute from the chromatographic column.
Regularly utilized detectors are UV-spectroscopy,
fluorescence, mass spectrometric and
electrochemical identifiers [9].
 Data Collection Devices or Integrator: Signals
from the detector might be gathered on graph
recorders or electronic integrators that fluctuate in
many-sided quality and in their capacity to
process, store and reprocess chromatographic
information. The PC coordinates the reaction of
the indicator to every part and places it into a
chromatograph that is anything but difficult to
interpret [10].
Fig.1. Flow diagram of HPLC
II. TYPES OF HPLC-
A. BASED ON MODE OF SEPARATION:
 Normal phase Chromatography: In this method,
separation is based on polarity. The first developer
used CaCo3 as separation column and Petroleum
ether as developing solvent. The combination of
stationary phase and mobile phase called normal
phase chromatography. The stationary phase is
polar, like silica gel type, cyanotype, amino type
and mobile phase is non-polar example organic
solvents, hexane, benzene [11].
 Reversed phase Chromatography: It is the
reverse of the normal phase chromatography in
which stationary phase is non- polar and mobile
phase is polar. The example of the mobile phase
are organic solvents (methanol, acetonitrile),
buffer (phosphate buffer)[10, 12].
B. Based on the principle of Separation:
 Absorption Chromatography-
In the absorption chromatography solute molecules
bond directly to the surface of the stationary phase
[13]. The component which has more affinity towards
mobile phase elutes first and the component which
has less affinity towards stationary phase elutes later.
No two components have the same affinity towards
mobile phase and stationary phase.
 Ion- exchange chromatography-
Ion exchange chromatography is a process that allows
the separation of ions and polar molecules based on
their charge. It can be used for almost any kind of
charged molecules including large proteins, small
nucleotides and amino acids. Retention is based on
the attraction between solute ions and charged sites
bound to the stationary phase. Ions of the same
charged are excluded. The use of resin (the stationary
solid phase) is used to covalently attach anions or
cations onto it. Solute ions of the opposite charge in
the mobile liquid phase are attracted to the resin by
electrostatic forces [10].
C. Ion- Pair chromatography-
It is a form of chromatography in which ions in
solution can be “paired” or neutralized and separated
as an ion pair on a reversed- phase column. Ion-
pairing agents are usually ionic compounds that
contain a hydrocarbon chain that imparts a certain
hydrophobicity so that the ion pair can be retained on
a reversed-phase column [14].
 Gel permeation chromatography-
This type of chromatography lacks an attractive
interaction between the stationary phase and solute.
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1620
The liquid or gaseous phase passes through a porous
gel which separates the molecules according to its
size. The pores are normally small and exclude the
larger solute molecules, but allow smaller molecules
to enter the gel, causing them to flow through a larger
volume. This causes the larger molecules to pass
through the column at a faster rate than the smaller
ones [15].
 Affinity chromatography-
This is the most selective type of chromatography
employed. It utilizes the specific interaction between
one kind of solute molecule and a second molecule
that is immobilized on a stationary phase. For
example, the immobilized molecule may be an
antibody to some specific protein. When a solute
containing a mixture of proteins are passed by this
molecule, only the specific protein reacted to this
antibody, binding it to the stationary phase. This
protein is later extracted by changing the ionic
strength or pH [16].
 Chiral chromatography-
It involves the separation of stereoisomers. In the case
of enantiomers, these have no chemical or physical
differences apart from being three- dimensional
mirror images. Conventional chromatography or other
separation processes are incapable of separating them.
To enable chiral separations to take place, either the
mobile phase or the stationary phase must themselves
be made chiral, giving differing affinities between the
analytes [17].
D. Based on elution technique-
 Isocratic elution:
A separation in which the mobile phase composition
remains constant throughout the procedure is termed
isocratic elution. In isocratic elution, peak width
increases with retention time linearly with the number
of theoretical plates. This leads to the disadvantage
that late-eluting peaks get very flat and broad. Best for
simple separations. Often used in quality control
application those supports and are in close proximity
to a manufacturing process [18].
 Gradient elution-
A separation in which the mobile phase composition
is changed during the separation process is described
as a gradient elution Gradient elution decreases the
retention of the later-eluting components so that they
elute faster, giving narrower peaks. This also
improves the peak shape and the peak height. Best for
the analysis of complex samples. Often used in
method development for unknown mixtures. Linear
gradients are most popular [19].
E. Based on a scale of operation-
 Analytical HPLC: No recovery of individual
components of a substance.
 Preparative HPLC: Individual components of
substance can be removed.
F. Based on the type of Analysis-
 Qualitative analysis: Analysis of a substance in
order to ascertain the nature of its chemical
constituents and we can separate individual
components but cannot assess the quantity in this
analysis.
 Quantitative analysis: Determining the amounts
and proportions of its chemical constituents and
the quantity of the impurity and individual
components can be assessed.
Separation Mechanism: Compound are separated
because the molecules are moves at different rates in
the column.
Fig.2. Separation mechanism
Due to different interactions between stationary phase
and different sample, the molecules move to different
rate, therefore separation can be done [20].
Fig.3. Separation mechanism
III. DIFFERENT TYPES OF HPLC
DETECTORS:
Detectors in HPLC is placed at the end of the
analytical column. The function of the detector is to
examine the solution which is eluting from the
column. An electronic signal is proportional to the
concentration of individual components of the
analyte.
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1621
HPLC detector has particular characteristics such as
excellent linear response as a function of the good
sensitivity works approximately in the range of 0.01-
100 µg of the compound in elutes.
A. UV-Visible- Ultraviolet Visible detector: It is the
most commonly used detector in the HPLC. Most
of the organic compounds absorb light in the
region of UV (190-400nm) and in the visible
region (400-750nm). It is based on Beer-Lambert
law; deuterium and high-pressure xenon lamp are
the sources of the UV. It has various advantages
and disadvantage [21].
 Advantages: It has high sensitivity, relative
robust to temperature, compatible with gradient
elution.
 Disadvantages: only compounds with UV or
visible absorption could be detected.
B. Refractive index detectors: Refractive index is a
bulk property of column eluent. In this detector,
detection depends on solute modifying the overall
refractive index of the mobile phase. Bulk
property detectors have an inherently limited
sensitivity. For the detection of the non-ionic
compounds refractive index can be very useful
and non-ionic compounds are not absorbed in the
UV region and in fluoresce [22]. Various types of
RI detectors are the following [23].
 Christiansen effect detector
 Interferometer detector
 Thermal lens detector
 Dielectric constant detector
 Advantages:
1. Responds to nearly all solutes
2. Unaffected by the flow rate
 Disadvantages:
1. Not as sensitive as most other types of detectors
2. Could not be used with a gradient elution
C. Photodiode Array detector (PAD), Diode Array
detector: Photodiode arrays (semiconductor
devices) are used in the detection unit. A DAD
detects the absorption in UV to VIS region. While
a UV-VIS detector has only one sample-side light-
receiving section, a DAD has multiple (1024 for
L-2455/2455U) photodiode arrays to obtain
information over a wide range of wavelengths at
one time, which is a merit of the DAD [24].
 Advantages:
1. .PDA Detector could analyze a sample
simultaneously at many different wavelengths
2. UV visible spectra are useful for compound
identification, checking peak purity, as well as
finding the optimum absorbance for the
compounds.
3. UV visible spectra of many compounds could be
stored in the spectrum libraries, which are used for
compound identification.
4. Relatively robust to temperature and flow rate
fluctuations
5. Compatible with gradient elution.
 Disadvantages:
1. Slightly less sensitive than the UV-Visible
detector.
D. Fluorescence Detector: It is the most sensitive,
specific detector among all existing HPLC
detector. It is possible to detect the presence of
single analyte molecule in the flow cell. The
sensitivity of this detector is 10-1000 times higher
than UV detector [25]. Various types of
Fluorescence detector are:
 The single wavelength excitation Fluorescence
detector
 Multi wavelength Fluorescence detector
 Laser-induced Fluorescence detector
Advantages:
1. Sensitivity is higher than UV-Vis detector
2. Selectivity is high because of relatively few-
compounds fluorescences.
3. Compatible with gradient elution
Disadvantages:
1. Difficult to predict fluorescence
2. Greatly affected by environment, solvent, pH,
temperature, viscosity, ionic strength, dissolved
gas.
E. Electrical Conductivity Detector: It provides
universal, reproducible, high-sensitivity detection
of all charged species such as anions, cations,
metals, organic acids. These detectors measure the
conductivity of the total mobile phase hence
categorized in bulk density detectors. Electrodes
of this detector are usually made up of platinum,
stainless steel or some other noble metal [26].
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1622
 Advantages:
1. Respond to ionic compounds and suitable for ion
chromatography.
2. High sensitivity for low concentration range.
 Disadvantages:
1. Sensitive to the fluctuations in the solvent flow
and mobile phase composition
2. Not compatible with gradient elution.
IV. SELECTION OF DETECTOR:
Table1. Selection of Detector [27]
S.NO DETECTORS Type of compounds can be
detected
1. UV-Vis
&PDA
Compounds with
chromophores, such as
aromatic rings or multiple
alternating double bonds.
2. RF Fluorescent compounds,
usually with fused rings or
highly conjugated planar
system.
3. CDD Charged compounds, such as
inorganic ions and organic
acids.
4. ECD For easily oxidized
compounds like quinones or
amines.
V. BASIC MAINTENANCE OF HPLC:
A. Daily maintenance (pre-analysis check):
 Filter mobile phase
 Wait for mobile phase to reach room temperature
 Purge all the flow lines with mobile phase
 Replace mobile phase as necessary
 If the solvent tubing is not completely filled with
mobile phase, purge the flow line
 After periods of disuse or changes in ambient
temperature, purge the flow line.
 If the buffer is used as mobile phase, wash the
back of plunger seal.
 If auto sampler is used, verify the auto sampler
rinse the bottle fully.
 Purge the autosampler rinse solution
 Checks for leaks
 Checks the pump pressure
 Perform a baseline check.
B. Daily maintenance (post-analysis checks):
 Wash the column
 If the column is not going to be used for a long
time, take off it from the system after washing,
and cap both ends of the column, store it in a cool
and dark place.
 Clean the flow line or the rest of the instrument, as
necessary
 Store the solvent reservoir filters in rinse solvent
to prevent them from drying out
 Solvent Reservoir Maintenance: The material of
containers is glass which is generally used and
plastic which has high sensitivity inorganic
analysis.
1. Always cover the containers
2. Prevent the evaporation of solvents
3. Prevent the dust enter the mobile phase
4. Reduce the vapours in the room
5. Do not make containers as this may lead to the
creation of partial vacuum and pump system.
6. Do not expose the bottle to direct sunlight/wind
7. Always keep reservoir above solvent delivery
system, ensures good siphon feed to pump and
prevents starving pumping system.
8. Always use a suction filter, prevent the
particulates from reaching pumps and clean the
suction filter by sonicating in isopropanol or 1N
nitric acid and water.
9. Good practice in the lab to label all bottles with
composition inside as well as the date of
prescription.
 Mobile Phase: If the solvent has background
absorption, the linear range will be smaller, the
noise will be bigger and baseline stability,
especially during gradient, runs will be poor.
1. The use of pure water systems is ideal as both
inorganic and organic contaminants are removed.
2. Modern pure water systems give TOC values as
indicated of total organic content in water.
3. Ghost peak will appear in gradient elution if water
is bad quality.
VI. PRECAUTIONS IN HANDLING SOME
HPLC GRADE ORGANIC SOLVENTS.
1. Analytical grade THF contains butylated
hydroxytoluene (BHT) as an oxidation inhibitor.
As BHT has strong UV absorption, it is not added
for HPLC grade THF.
2. As such, HPLC grade THF must be stored in cool
and dark places to reduce oxidation to form
peroxides. Accumulation of a large amount of
organic peroxides may lead to explosions.
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1623
A. Use of buffer solutions:
 Buffer solutions must be filtered before use.
 Buffer solutions must not be left in the system to
avoid crystallization.
 Effect on the pump- damage plunger and seal
 Effect on the column- the creation of column
voids
 Effect on flow line- corrosion of stainless steel
lines
 Possible bacterial growth, especially phosphate
buffers good medium for bacterial and fungus
growth.
 Ideally, solutions should be prepared fresh every
day [28].
B. Premixing of Solvents: For isocratic systems,
different solvents are premixed.
Example: to prepare a 1000ml of 50:50 mixture of
water/methanol.
 Measure out 500 ml of water and 500ml of
methanol separately, and then mix them together.
 Measure 500ml of water and top up to the 1000ml
mark of measuring cylinder/ volumetric flask with
methanol.
 Measure 500ml of methanol and top up to the
1000ml mark of measuring cylinder/ volumetric
flask with water.
All three methods will give different compositions,
whichever method selected must be used consistently
for the repeatable result.
C. Pump maintenance:
 Clean behind plunger seal when using buffer
solution-
 Automatic rinse kit for LC-10Avp and LC_20A
series.
 Useful spares to be maintained- plunger seals,
plunger, check valves, filter frit for line filter.
 Most’s tools required for maintenance work come
with the pump accessories kit- do not lose them.
D. Sample treatment: Always filtered sample with
0.2µ m or 0.45µ m filters before injection. If the
sample contains a lot of solid, centrifuge it, and
the filter the solution.
 Select suitable solvent to dissolve the sample.
Normally dissolve sample in the mobile phase.
But do not select a high soluble solvent as a
sample solvent.
E. Injector maintenance:
Manual injector:
Precautions:
 Do not use pointed or beveled needle tip. Use only
flat and end type.
 If pH higher than 10 is to be used, Delrin rotter
seal is required.
 When both manual and auto-injector are present in
the system, auto sampler is always connected first,
followed by manual.
 Manual injector always set to load position when
auto sampler is used and to inject position
otherwise.
F. Auto-Injector: General Maintenance
 Purge flow line at least once a day.
 Change purge liquid frequently.
 Never used the salt solution as a purge liquid.
 Used septums for sample vial recommended by
the manufacturer only.
 Septums should not be reused.
 Do not fill vials to the brim.
G. Detector maintenance:
 UV-Vs detector: Check D2 lamp energy- Check
the energy of reference cell. Change D2 lamp if
the energy is lower than 800.
(Note- do not touch the surface of D2 lamp by hand.)
 Waste bottle maintenance:
 Remember to empty waste bottle regularly.
 Do not mix concentrated nitric acid with an
organic solvent.
 Do not immerse the drain tubing in the waste
liquid as this may create back pressure on the flow
cells [30].
 Advantages of HPLC-
 High Resolution
 Good repeatability
 High Sensitivity (ppm-ppb)
 Small sample size
 Moderate analyses condition
 No need to vaporize the sample like GC
 Easy to fractionate the sample and purify
 Non-Destructive
 Separations fast and efficient (High Resolution
Power)
 Continuous monitoring of the column effluent
 It can be applied to the separation and analysis of
very complex mixtures.
 Accurate quantitative measurements
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1624
 Repetitive and reproducible analysis using the
same column
 Adsorption, partition, ion exchange and exclusion
column separations are excellently made
 HPLC is more versatile than GLC in some
respects because it has the advantage of not being
restricted to volatile and thermally stable solute
and the choice of mobile and stationary phases is
much wider in HPLC.
 Both aqueous and non- aqueous samples can be
analyzed with little or no sample pre-treatment
 A variety of solvents and column packing are
available, providing a high degree of selectivity
for specific analysis.
 It provides a means for determination of multiple
components in a single analysis.
 Application-
HPLC is one of the most widely applied analytical
separation techniques-[29]
 Pharmaceutical:
 Tablet dissolution of pharmaceutical dosages.
 Shelf life determination of pharmaceutical
products.
 Identification of counterfeit drug products.
 Pharmaceutical quality control.
 Environmental:
 Phenols in drinking water
 Identification of diphenhydramine in sediment
samples.
 Bio monitoring of PAH pollution in high-altitude
mountain lakes through the analysis of fish bile
 Estrogens in coastal waters- The sewage source
 Toxicity of tetracyclines and tetracycline
degradation products to environmentally relevant
bacteria
 Assessment of TNT toxicity in sediment.
 Forensics:
 A mobile HPLC apparatus at dance parties- on-
site identification and quantification of the drug
Ecstasy.
 Identification of anabolic steroids in serum, urine,
sweat and hair
 Forensic analysis of textile dyes
 Simultaneous quantification of psychotherapeutic
drugs in human plasma
 Clinical:
 Quantification of DEET in Human Urine.
 Analysis of antibiotics
 Increased urinary excretion of aquaporin 2 in
patients with liver cirrhosis
 Detection of endogenous neuropeptides in brain
extracellular fluids.
 Food & Flavour:
 Ensuring soft drink consistency and quality
 Analysis of vicinal dike tones in beer
 Sugar analysis in fruit juices
 Polycyclic aromatic hydrocarbons in Brazilian
vegetables and fruits.
 Trace analysis of military high explosives in
agricultural crops
 Stability of aspartame in the presence of glucose
and vanillin [2, 10].
VII. CONCLUSION
In this review, we include all fundamental aspects of
HPLC technique. We describe the general technique
of HPLC method and maintenance of the instrument.
The general approach for these all point makes the
precise quality of techniques. This method is widely
accepted in the laboratory as well as in the clinical
and forensic science lab due to the accuracy and
precession and specificity with the HPLC. The only
disadvantage of HPLC is high cost and complex
methodology.
VIII.REFERNCES:
1. Olander, D.P., Instrumental Methods of Analysis,
Sixth Edition (Willard, Hobart H.; Merritt, Lynne
L.; Dean, John A.; Settle, Frank A., Jr.). Journal of
Chemical Education, 1984. 61(8): p. A222.
2. Lipka, E. and C. Vaccher, Quantitative analysis of
drugs in biological matrices by HPLC hyphenated
to fluorescence detection. Bioanalysis, 2015. 7(6):
p. 743-762.
3. Choudhury, H., et al., Development and validation
of RP-HPLC method: scope of application in the
determination of oil solubility of paclitaxel. J
Chromatogr Sci, 2014. 52(1): p. 68-74.
4. Md Sabir, A., M. Moloy, and P. S Bhasin, HPLC
METHOD DEVELOPMENT AND
VALIDATION: A REVIEW. Vol. 4. 2015. 39-46.
5. Malherbe, C.J., D. de Beer, and E. Joubert,
Development of on-line high performance liquid
chromatography (HPLC)-biochemical detection
methods as tools in the identification of
bioactives. Int J Mol Sci, 2012. 13(3): p. 3101-33.
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1625
6. Hopmann, E., W. Arlt, and M. Minceva, Solvent
system selection in counter-current
chromatography using conductor-like screening
model for real solvents. J Chromatogr A, 2011.
1218(2): p. 242-50.
7. Xiang, Y., Y. Liu, and M.L. Lee, Ultrahigh
pressure liquid chromatography using elevated
temperature. J Chromatogr A, 2006. 1104(1-2): p.
198-202.
8. Martin, M. and G. Guiochon, Effects of high
pressure in liquid chromatography. J Chromatogr
A, 2005. 1090(1-2): p. 16-38.
9. Seger, C., S. Sturm, and H. Stuppner, Mass
spectrometry and NMR spectroscopy: modern
high-end detectors for high resolution separation
techniques--state of the art in natural product
HPLC-MS, HPLC-NMR, and CE-MS
hyphenations. Nat Prod Rep, 2013. 30(7): p. 970-
87.
10. Malviya, R., et al., High performance liquid
chromatography: A short review. Vol. 2. 2010.
22-26.
11. Lababidi, S. and W. Schrader, Online normal-
phase high-performance liquid
chromatography/Fourier transform ion cyclotron
resonance mass spectrometry: effects of different
ionization methods on the characterization of
highly complex crude oil mixtures. Rapid
Commun Mass Spectrom, 2014. 28(12): p. 1345-
52.
12. Fueangfung, S., Y. Yuan, and S. Fang, Denaturing
reversed-phase HPLC using a mobile phase
containing urea for oligodeoxynucleotide analysis.
Nucleosides Nucleotides Nucleic Acids, 2014.
33(7): p. 481-8.
13. Shibasaki, H., et al., Separation and quantitative
determination of 6alpha-hydroxycortisol and
6beta-hydroxycortisol in human urine by high-
performance liquid chromatography with
ultraviolet absorption detection. Anal Bioanal
Chem, 2012. 402(9): p. 2945-52.
14. Jerz, G., et al., Separation of amaranthine-type
betacyanins by ion-pair high-speed countercurrent
chromatography. J Chromatogr A, 2014. 1344: p.
42-50.
15. O'Fagain, C., P.M. Cummins, and B.F. O'Connor,
Gel-filtration chromatography. Methods Mol Biol,
2011. 681: p. 25-33.
16. Pfaunmiller, E.L., et al., Affinity monolith
chromatography: a review of principles and recent
analytical applications. Anal Bioanal Chem, 2013.
405(7): p. 2133-45.
17. Leon-Gonzalez, M.E., et al., Two-dimensional
liquid chromatography for direct chiral
separations: a review. Biomed Chromatogr, 2014.
28(1): p. 59-83.
18. Xu, M., et al., [Determination of chlorogenic acid
and baicalin in Yinhuang film-coated tablets by
HPLC isocratic elution]. Zhongguo Zhong Yao Za
Zhi, 2013. 38(5): p. 691-3.
19. D'Archivio, A.A., M.A. Maggi, and F. Ruggieri,
Prediction of the retention of s-triazines in
reversed-phase high-performance liquid
chromatography under linear gradient-elution
conditions. J Sep Sci, 2014. 37(15): p. 1930-6.
20. Marchetti, N., et al., Determination of adsorption
isotherms by means of HPLC: adsorption
mechanism elucidation and separation
optimization. J Sep Sci, 2009. 32(5-6): p. 727-41.
21. Kraiczek, K.G., et al., Highly flexible UV-vis
radiation sources and novel detection schemes for
spectrophotometric HPLC detection. Anal Chem,
2014. 86(2): p. 1146-52.
22. Ping, B.T.Y., H.A. Aziz, and Z. Idris, Comparison
of Peak-area Ratios and Percentage Peak Area
Derived from HPLC-evaporative Light Scattering
and Refractive Index Detectors for Palm Oil and
its Fractions. J Oleo Sci, 2018. 67(3): p. 265-272.
23. Kupina, S. and M. Roman, Determination of total
carbohydrates in wine and wine-like beverages by
HPLC with a refractive index detector: First
Action 2013.12. J AOAC Int, 2014. 97(2): p. 498-
505.
24. Pragst, F., M. Herzler, and B.T. Erxleben,
Systematic toxicological analysis by high-
performance liquid chromatography with diode
array detection (HPLC-DAD). Clin Chem Lab
Med, 2004. 42(11): p. 1325-40.
25. Raut, P.P. and S.Y. Charde, Simultaneous
estimation of levodopa and carbidopa by RP-
HPLC using a fluorescence detector: its
application to a pharmaceutical dosage form.
Luminescence, 2014. 29(7): p. 762-71.
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1626
26. Zhang, M., et al., Monitoring gradient profile on-
line in micro- and nano-high performance liquid
chromatography using conductivity detection. J
Chromatogr A, 2016. 1460: p. 68-73.
27. Stubbs, C. and I. Kanfer, A stability-indicating
high-performance liquid chromatographic assay of
erythromycin estolate in pharmaceutical dosage
forms. International Journal of Pharmaceutics,
1990. 63(2): p. 113-119.
28. Bakshi, M. and S. Singh, Development of
validated stability-indicating assay methods--
critical review. J Pharm Biomed Anal, 2002.
28(6): p. 1011-40.
29. Michalski, R., Application of ion chromatography
in clinical studies and pharmaceutical industry.
Mini Rev Med Chem, 2014. 14(10): p. 862-72.

More Related Content

PPT
Kaumil Hplc
PPTX
Columns in HPLC
PDF
PPTX
RATE TH,PLATE SYS SUTIABILITY.pptx
PPTX
PPTX
Partition column-chromatography
Kaumil Hplc
Columns in HPLC
RATE TH,PLATE SYS SUTIABILITY.pptx
Partition column-chromatography

What's hot (20)

PPTX
HPLC (HIGH PERFORMANCE LIQUID CHROMATOGRAPHY)
PPTX
Prep hplc 1
PPTX
sample preparation in hplc
PPT
high performance thin layer chromatography [HPTLC]
PPTX
HPLC method development
PPTX
High performance liquid chromatography
PPTX
HPLC AND ITS APPLICATIONS
PPTX
counter current chromatography
PPTX
PPTX
HPLC
PPTX
Tlc (Thin layer chromatography)
PPTX
Ultra performance liquid chromatography
PPTX
PRINCIPLE , INSTRUMENTATION & APPLICATION OF SUPER CRITICAL FLUID CHROMATOGRAPHY
PPTX
Preperative HPLC.pptx
PPTX
High performance liquid chromatography
PPTX
Ultra Performance Liquid Chromatography
PPTX
gas chromatography
DOCX
gas chromatography
PDF
Capillary electrophoresis
HPLC (HIGH PERFORMANCE LIQUID CHROMATOGRAPHY)
Prep hplc 1
sample preparation in hplc
high performance thin layer chromatography [HPTLC]
HPLC method development
High performance liquid chromatography
HPLC AND ITS APPLICATIONS
counter current chromatography
HPLC
Tlc (Thin layer chromatography)
Ultra performance liquid chromatography
PRINCIPLE , INSTRUMENTATION & APPLICATION OF SUPER CRITICAL FLUID CHROMATOGRAPHY
Preperative HPLC.pptx
High performance liquid chromatography
Ultra Performance Liquid Chromatography
gas chromatography
gas chromatography
Capillary electrophoresis
Ad

Similar to HPLC: Principle and Maintenance with Application (20)

PPTX
Simultaneous estimation of uv spectroscopy and method development and validat...
DOCX
High performance liquid chromatography.docx
PPTX
Difference between HPLC and HPTLC and Applications.
PPTX
Sagar kanade m pharm indrustial project
PDF
Development and estimation of carvedilol in oral dosage form using HPLC method
PPTX
high performance liquid chromatography in chem
PPTX
Royal ppt hplc
PPTX
Nwoke_stephen[1].pptxlllllllklkkkllkllkkok
PPTX
Nwoke_stephen[1].pptxlllllllklkkkllkllkkok
PDF
A Review HPLC Method Development and Validation
PPTX
A brief review on development and validation of hplc method.
PPTX
High Performance Liquid Chromatography by Dharmendra Kumar Parte
PDF
Development and validation of a stability indicating RP-HPLC method for estim...
PPTX
Chromatography Part-III
PPTX
High-Performance Liquid Chromatography(HPLC)
PPTX
FINAL PPT OF MPHARM PROJECT2024 - Copy.pptx
PDF
PPTX
High performance liquid chromato graphy
PPTX
Hplc by Bachaspati jana
Simultaneous estimation of uv spectroscopy and method development and validat...
High performance liquid chromatography.docx
Difference between HPLC and HPTLC and Applications.
Sagar kanade m pharm indrustial project
Development and estimation of carvedilol in oral dosage form using HPLC method
high performance liquid chromatography in chem
Royal ppt hplc
Nwoke_stephen[1].pptxlllllllklkkkllkllkkok
Nwoke_stephen[1].pptxlllllllklkkkllkllkkok
A Review HPLC Method Development and Validation
A brief review on development and validation of hplc method.
High Performance Liquid Chromatography by Dharmendra Kumar Parte
Development and validation of a stability indicating RP-HPLC method for estim...
Chromatography Part-III
High-Performance Liquid Chromatography(HPLC)
FINAL PPT OF MPHARM PROJECT2024 - Copy.pptx
High performance liquid chromato graphy
Hplc by Bachaspati jana
Ad

More from ijtsrd (20)

PDF
A Study of School Dropout in Rural Districts of Darjeeling and Its Causes
PDF
Pre extension Demonstration and Evaluation of Soybean Technologies in Fedis D...
PDF
Pre extension Demonstration and Evaluation of Potato Technologies in Selected...
PDF
Pre extension Demonstration and Evaluation of Animal Drawn Potato Digger in S...
PDF
Pre extension Demonstration and Evaluation of Drought Tolerant and Early Matu...
PDF
Pre extension Demonstration and Evaluation of Double Cropping Practice Legume...
PDF
Pre extension Demonstration and Evaluation of Common Bean Technology in Low L...
PDF
Enhancing Image Quality in Compression and Fading Channels A Wavelet Based Ap...
PDF
Manpower Training and Employee Performance in Mellienium Ltdawka, Anambra State
PDF
A Statistical Analysis on the Growth Rate of Selected Sectors of Nigerian Eco...
PDF
Automatic Accident Detection and Emergency Alert System using IoT
PDF
Corporate Social Responsibility Dimensions and Corporate Image of Selected Up...
PDF
The Role of Media in Tribal Health and Educational Progress of Odisha
PDF
Advancements and Future Trends in Advanced Quantum Algorithms A Prompt Scienc...
PDF
A Study on Seismic Analysis of High Rise Building with Mass Irregularities, T...
PDF
Descriptive Study to Assess the Knowledge of B.Sc. Interns Regarding Biomedic...
PDF
Performance of Grid Connected Solar PV Power Plant at Clear Sky Day
PDF
Vitiligo Treated Homoeopathically A Case Report
PDF
Vitiligo Treated Homoeopathically A Case Report
PDF
Uterine Fibroids Homoeopathic Perspectives
A Study of School Dropout in Rural Districts of Darjeeling and Its Causes
Pre extension Demonstration and Evaluation of Soybean Technologies in Fedis D...
Pre extension Demonstration and Evaluation of Potato Technologies in Selected...
Pre extension Demonstration and Evaluation of Animal Drawn Potato Digger in S...
Pre extension Demonstration and Evaluation of Drought Tolerant and Early Matu...
Pre extension Demonstration and Evaluation of Double Cropping Practice Legume...
Pre extension Demonstration and Evaluation of Common Bean Technology in Low L...
Enhancing Image Quality in Compression and Fading Channels A Wavelet Based Ap...
Manpower Training and Employee Performance in Mellienium Ltdawka, Anambra State
A Statistical Analysis on the Growth Rate of Selected Sectors of Nigerian Eco...
Automatic Accident Detection and Emergency Alert System using IoT
Corporate Social Responsibility Dimensions and Corporate Image of Selected Up...
The Role of Media in Tribal Health and Educational Progress of Odisha
Advancements and Future Trends in Advanced Quantum Algorithms A Prompt Scienc...
A Study on Seismic Analysis of High Rise Building with Mass Irregularities, T...
Descriptive Study to Assess the Knowledge of B.Sc. Interns Regarding Biomedic...
Performance of Grid Connected Solar PV Power Plant at Clear Sky Day
Vitiligo Treated Homoeopathically A Case Report
Vitiligo Treated Homoeopathically A Case Report
Uterine Fibroids Homoeopathic Perspectives

Recently uploaded (20)

PDF
O7-L3 Supply Chain Operations - ICLT Program
PDF
VCE English Exam - Section C Student Revision Booklet
PDF
Module 4: Burden of Disease Tutorial Slides S2 2025
PDF
Microbial disease of the cardiovascular and lymphatic systems
PPTX
Pharmacology of Heart Failure /Pharmacotherapy of CHF
PDF
grade 11-chemistry_fetena_net_5883.pdf teacher guide for all student
PPTX
Presentation on HIE in infants and its manifestations
PPTX
1st Inaugural Professorial Lecture held on 19th February 2020 (Governance and...
PPTX
Lesson notes of climatology university.
PDF
Black Hat USA 2025 - Micro ICS Summit - ICS/OT Threat Landscape
PPTX
IMMUNITY IMMUNITY refers to protection against infection, and the immune syst...
PDF
01-Introduction-to-Information-Management.pdf
PDF
2.FourierTransform-ShortQuestionswithAnswers.pdf
PDF
FourierSeries-QuestionsWithAnswers(Part-A).pdf
PDF
102 student loan defaulters named and shamed – Is someone you know on the list?
PDF
Classroom Observation Tools for Teachers
PDF
Supply Chain Operations Speaking Notes -ICLT Program
PPTX
human mycosis Human fungal infections are called human mycosis..pptx
PDF
Computing-Curriculum for Schools in Ghana
PDF
Abdominal Access Techniques with Prof. Dr. R K Mishra
O7-L3 Supply Chain Operations - ICLT Program
VCE English Exam - Section C Student Revision Booklet
Module 4: Burden of Disease Tutorial Slides S2 2025
Microbial disease of the cardiovascular and lymphatic systems
Pharmacology of Heart Failure /Pharmacotherapy of CHF
grade 11-chemistry_fetena_net_5883.pdf teacher guide for all student
Presentation on HIE in infants and its manifestations
1st Inaugural Professorial Lecture held on 19th February 2020 (Governance and...
Lesson notes of climatology university.
Black Hat USA 2025 - Micro ICS Summit - ICS/OT Threat Landscape
IMMUNITY IMMUNITY refers to protection against infection, and the immune syst...
01-Introduction-to-Information-Management.pdf
2.FourierTransform-ShortQuestionswithAnswers.pdf
FourierSeries-QuestionsWithAnswers(Part-A).pdf
102 student loan defaulters named and shamed – Is someone you know on the list?
Classroom Observation Tools for Teachers
Supply Chain Operations Speaking Notes -ICLT Program
human mycosis Human fungal infections are called human mycosis..pptx
Computing-Curriculum for Schools in Ghana
Abdominal Access Techniques with Prof. Dr. R K Mishra

HPLC: Principle and Maintenance with Application

  • 1. @ IJTSRD | Available Online @ www.ijtsrd.com ISSN No: 2456 International Research HPLC: Principle and Maintenance with Application Yogesh Kumar Research Scholar, Delhi Pharmaceutical Sciences ABSTRACT High performance liquid chromatography (HPLC) is a significant qualitative and quantitative technique, usually used for the estimation of pharmaceutical and biological samples. The chromatography term is derived from the Greek words namely chroma (colour) and graphein (to write). The chromatography is very accepted technique and it is mostly used analytically. It is the most resourceful, safest, reliable and fastest chromatographic technique for the quality control of drug components. This technique involves 2 phases’ stationary and mobile phases. There are different types of chromatographic techniques. The separation of constituents is based on the variation between the partition coefficients of the two phases. This article is primed with an aim to review different aspects of HPLC, such as principle, types, instrumentation and application with maintenance. KEYWORD: HPLC; Instrumentation; maintenance; application; chromatography I. INTRODUCTION Analytical chemistry is widely used in determining the qualitative and quantitative composition of typical mixtures. These two methods are the main aspects to understand the sample materials. Generally, analytical chemistry divides into two branches i.e. q and quantitative analysis. High performance liquid chromatography also is known as High Pressure Liquid Chromatography. HPLC is an analysis technique used for separation, identification and quantification of typical mixtures such as organic, inorganic, biological, ionic and polymeric materials. HPLC is a type of column chromatography in which solvent flows with high pressure, so that sample can be separated into different constituents divided into different types based on modes of separation, the @ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 ISSN No: 2456 - 6470 | www.ijtsrd.com | Volume International Journal of Trend in Scientific Research and Development (IJTSRD) International Open Access Journal HPLC: Principle and Maintenance with Application Yogesh Kumar, Sayed Md Mumtaz, Mustaq Ahmad search Scholar, Pharmacology Department, Delhi Pharmaceutical Sciences and Research University Delhi, Delhi, India High performance liquid chromatography (HPLC) is a significant qualitative and quantitative technique, usually used for the estimation of pharmaceutical and chromatography term is derived from the Greek words namely chroma (colour) and graphein (to write). The chromatography is very accepted technique and it is mostly used analytically. It is the most resourceful, safest, reliable echnique for the quality control of drug components. This technique involves 2 phases’ stationary and mobile phases. There are different types of chromatographic techniques. The separation of constituents is based on the variation fficients of the two phases. This article is primed with an aim to review different aspects of HPLC, such as principle, types, instrumentation and application with maintenance. HPLC; Instrumentation; maintenance; Analytical chemistry is widely used in determining the qualitative and quantitative composition of typical mixtures. These two methods are the main aspects to understand the sample materials. Generally, analytical chemistry divides into two branches i.e. qualitative and quantitative analysis. High performance liquid chromatography also is known as High Pressure Liquid Chromatography. HPLC is an analysis technique used for separation, identification and quantification of typical mixtures such as organic, rganic, biological, ionic and polymeric materials. HPLC is a type of column chromatography in which solvent flows with high pressure, so that sample can be separated into different constituents divided into different types based on modes of separation, the principle of separation, elution technique, scale of separation and based on the type of analysis Detector is the heart of this instrument and the efficiency of this system depends upon detecting techniques. Many types of HPLC detectors such as UV-Vs, refractive index detector, photodiode detector, fluorescence, electrical conductivity detector. A. Chromatography: Chromatography is an analytical method in which compounds are physically separated prior to measurement. The main purpose of chromatography is to separate and quantify the target sample in the matrix. The first developer of Tswett. B. Scope of HPLC: HPLC is used in various fields like in pharmaceuticals, biochemicals, food products, industrial chemicals, forensic chemistry, environmental field, clinical medicine for analysis for typical mixtures like antibiotics, amino acids, fatty acids, drugs, poisons, inorganic ions, urine extracts, oestrogens [2, 3]. C. Need of HPLC The use of HPLC are increased day by day across the world due to its unique properties like high resolutions, high sensitivity (ppm repeatability, small sample size, moderate analysis condition, no need to vaporise the sample as in the gas chromatography, easy to fractionate the sample and purify [4, 5]. Aug 2018 Page: 1618 6470 | www.ijtsrd.com | Volume - 2 | Issue – 5 Scientific (IJTSRD) International Open Access Journal HPLC: Principle and Maintenance with Application , India principle of separation, elution technique, scale of separation and based on the type of analysis [1]. The Detector is the heart of this instrument and the efficiency of this system depends upon detecting techniques. Many types of HPLC detectors such as Vs, refractive index detector, photodiode detector, fluorescence, electrical conductivity Chromatography is an analytical method in which compounds are physically separated prior to measurement. The main purpose of chromatography is to separate and quantify the target sample in the matrix. The first developer of chromatography was M. HPLC is used in various fields like in pharmaceuticals, biochemicals, food products, industrial chemicals, forensic chemistry, environmental field, clinical medicine for analysis for ntibiotics, amino acids, fatty acids, drugs, poisons, inorganic ions, urine extracts, The use of HPLC are increased day by day across the world due to its unique properties like high resolutions, high sensitivity (ppm-ppb), good repeatability, small sample size, moderate analysis condition, no need to vaporise the sample as in the gas atography, easy to fractionate the sample and
  • 2. International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470 @ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1619 D. Instrumentation: The HPLC instrumentation involves pump, injector, column, detector, and integrator and display system.  Solvent reservoir: In HPLC the mobile phase or solvent is a mixture of polar and non-polar liquid components. Depending on the composition of the sample, the polar and non-polar solvents will be varied [6].  Pump: The pump suctions the mobile phase from the solvent reservoir and forces it to the column and then passes to the detector. 42000 KPa is the operating pressure of the pump. This operating pressure depends on column dimensions, particle size, flow rate and composition of mobile phase [7].  Sample Injector: An injector for an HPLC framework should give an infusion of the fluid specimen inside the scope of 0.1 mL to 100 mL of volume with high reproducibility and under high pressure (up to 4000 psi).  Columns: Columns are typically made of cleaned stainless steel, are somewhere around 50 mm and 300 mm long and have an inward distance across of somewhere around 2 and 5 mm. They are generally loaded with a stationary phase with a molecule size of 3 μm to 10 μm [8].  Detector: The HPLC detector, situated toward the end of the column distinguishes the analytes as they elute from the chromatographic column. Regularly utilized detectors are UV-spectroscopy, fluorescence, mass spectrometric and electrochemical identifiers [9].  Data Collection Devices or Integrator: Signals from the detector might be gathered on graph recorders or electronic integrators that fluctuate in many-sided quality and in their capacity to process, store and reprocess chromatographic information. The PC coordinates the reaction of the indicator to every part and places it into a chromatograph that is anything but difficult to interpret [10]. Fig.1. Flow diagram of HPLC II. TYPES OF HPLC- A. BASED ON MODE OF SEPARATION:  Normal phase Chromatography: In this method, separation is based on polarity. The first developer used CaCo3 as separation column and Petroleum ether as developing solvent. The combination of stationary phase and mobile phase called normal phase chromatography. The stationary phase is polar, like silica gel type, cyanotype, amino type and mobile phase is non-polar example organic solvents, hexane, benzene [11].  Reversed phase Chromatography: It is the reverse of the normal phase chromatography in which stationary phase is non- polar and mobile phase is polar. The example of the mobile phase are organic solvents (methanol, acetonitrile), buffer (phosphate buffer)[10, 12]. B. Based on the principle of Separation:  Absorption Chromatography- In the absorption chromatography solute molecules bond directly to the surface of the stationary phase [13]. The component which has more affinity towards mobile phase elutes first and the component which has less affinity towards stationary phase elutes later. No two components have the same affinity towards mobile phase and stationary phase.  Ion- exchange chromatography- Ion exchange chromatography is a process that allows the separation of ions and polar molecules based on their charge. It can be used for almost any kind of charged molecules including large proteins, small nucleotides and amino acids. Retention is based on the attraction between solute ions and charged sites bound to the stationary phase. Ions of the same charged are excluded. The use of resin (the stationary solid phase) is used to covalently attach anions or cations onto it. Solute ions of the opposite charge in the mobile liquid phase are attracted to the resin by electrostatic forces [10]. C. Ion- Pair chromatography- It is a form of chromatography in which ions in solution can be “paired” or neutralized and separated as an ion pair on a reversed- phase column. Ion- pairing agents are usually ionic compounds that contain a hydrocarbon chain that imparts a certain hydrophobicity so that the ion pair can be retained on a reversed-phase column [14].  Gel permeation chromatography- This type of chromatography lacks an attractive interaction between the stationary phase and solute.
  • 3. International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470 @ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1620 The liquid or gaseous phase passes through a porous gel which separates the molecules according to its size. The pores are normally small and exclude the larger solute molecules, but allow smaller molecules to enter the gel, causing them to flow through a larger volume. This causes the larger molecules to pass through the column at a faster rate than the smaller ones [15].  Affinity chromatography- This is the most selective type of chromatography employed. It utilizes the specific interaction between one kind of solute molecule and a second molecule that is immobilized on a stationary phase. For example, the immobilized molecule may be an antibody to some specific protein. When a solute containing a mixture of proteins are passed by this molecule, only the specific protein reacted to this antibody, binding it to the stationary phase. This protein is later extracted by changing the ionic strength or pH [16].  Chiral chromatography- It involves the separation of stereoisomers. In the case of enantiomers, these have no chemical or physical differences apart from being three- dimensional mirror images. Conventional chromatography or other separation processes are incapable of separating them. To enable chiral separations to take place, either the mobile phase or the stationary phase must themselves be made chiral, giving differing affinities between the analytes [17]. D. Based on elution technique-  Isocratic elution: A separation in which the mobile phase composition remains constant throughout the procedure is termed isocratic elution. In isocratic elution, peak width increases with retention time linearly with the number of theoretical plates. This leads to the disadvantage that late-eluting peaks get very flat and broad. Best for simple separations. Often used in quality control application those supports and are in close proximity to a manufacturing process [18].  Gradient elution- A separation in which the mobile phase composition is changed during the separation process is described as a gradient elution Gradient elution decreases the retention of the later-eluting components so that they elute faster, giving narrower peaks. This also improves the peak shape and the peak height. Best for the analysis of complex samples. Often used in method development for unknown mixtures. Linear gradients are most popular [19]. E. Based on a scale of operation-  Analytical HPLC: No recovery of individual components of a substance.  Preparative HPLC: Individual components of substance can be removed. F. Based on the type of Analysis-  Qualitative analysis: Analysis of a substance in order to ascertain the nature of its chemical constituents and we can separate individual components but cannot assess the quantity in this analysis.  Quantitative analysis: Determining the amounts and proportions of its chemical constituents and the quantity of the impurity and individual components can be assessed. Separation Mechanism: Compound are separated because the molecules are moves at different rates in the column. Fig.2. Separation mechanism Due to different interactions between stationary phase and different sample, the molecules move to different rate, therefore separation can be done [20]. Fig.3. Separation mechanism III. DIFFERENT TYPES OF HPLC DETECTORS: Detectors in HPLC is placed at the end of the analytical column. The function of the detector is to examine the solution which is eluting from the column. An electronic signal is proportional to the concentration of individual components of the analyte.
  • 4. International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470 @ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1621 HPLC detector has particular characteristics such as excellent linear response as a function of the good sensitivity works approximately in the range of 0.01- 100 µg of the compound in elutes. A. UV-Visible- Ultraviolet Visible detector: It is the most commonly used detector in the HPLC. Most of the organic compounds absorb light in the region of UV (190-400nm) and in the visible region (400-750nm). It is based on Beer-Lambert law; deuterium and high-pressure xenon lamp are the sources of the UV. It has various advantages and disadvantage [21].  Advantages: It has high sensitivity, relative robust to temperature, compatible with gradient elution.  Disadvantages: only compounds with UV or visible absorption could be detected. B. Refractive index detectors: Refractive index is a bulk property of column eluent. In this detector, detection depends on solute modifying the overall refractive index of the mobile phase. Bulk property detectors have an inherently limited sensitivity. For the detection of the non-ionic compounds refractive index can be very useful and non-ionic compounds are not absorbed in the UV region and in fluoresce [22]. Various types of RI detectors are the following [23].  Christiansen effect detector  Interferometer detector  Thermal lens detector  Dielectric constant detector  Advantages: 1. Responds to nearly all solutes 2. Unaffected by the flow rate  Disadvantages: 1. Not as sensitive as most other types of detectors 2. Could not be used with a gradient elution C. Photodiode Array detector (PAD), Diode Array detector: Photodiode arrays (semiconductor devices) are used in the detection unit. A DAD detects the absorption in UV to VIS region. While a UV-VIS detector has only one sample-side light- receiving section, a DAD has multiple (1024 for L-2455/2455U) photodiode arrays to obtain information over a wide range of wavelengths at one time, which is a merit of the DAD [24].  Advantages: 1. .PDA Detector could analyze a sample simultaneously at many different wavelengths 2. UV visible spectra are useful for compound identification, checking peak purity, as well as finding the optimum absorbance for the compounds. 3. UV visible spectra of many compounds could be stored in the spectrum libraries, which are used for compound identification. 4. Relatively robust to temperature and flow rate fluctuations 5. Compatible with gradient elution.  Disadvantages: 1. Slightly less sensitive than the UV-Visible detector. D. Fluorescence Detector: It is the most sensitive, specific detector among all existing HPLC detector. It is possible to detect the presence of single analyte molecule in the flow cell. The sensitivity of this detector is 10-1000 times higher than UV detector [25]. Various types of Fluorescence detector are:  The single wavelength excitation Fluorescence detector  Multi wavelength Fluorescence detector  Laser-induced Fluorescence detector Advantages: 1. Sensitivity is higher than UV-Vis detector 2. Selectivity is high because of relatively few- compounds fluorescences. 3. Compatible with gradient elution Disadvantages: 1. Difficult to predict fluorescence 2. Greatly affected by environment, solvent, pH, temperature, viscosity, ionic strength, dissolved gas. E. Electrical Conductivity Detector: It provides universal, reproducible, high-sensitivity detection of all charged species such as anions, cations, metals, organic acids. These detectors measure the conductivity of the total mobile phase hence categorized in bulk density detectors. Electrodes of this detector are usually made up of platinum, stainless steel or some other noble metal [26].
  • 5. International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470 @ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1622  Advantages: 1. Respond to ionic compounds and suitable for ion chromatography. 2. High sensitivity for low concentration range.  Disadvantages: 1. Sensitive to the fluctuations in the solvent flow and mobile phase composition 2. Not compatible with gradient elution. IV. SELECTION OF DETECTOR: Table1. Selection of Detector [27] S.NO DETECTORS Type of compounds can be detected 1. UV-Vis &PDA Compounds with chromophores, such as aromatic rings or multiple alternating double bonds. 2. RF Fluorescent compounds, usually with fused rings or highly conjugated planar system. 3. CDD Charged compounds, such as inorganic ions and organic acids. 4. ECD For easily oxidized compounds like quinones or amines. V. BASIC MAINTENANCE OF HPLC: A. Daily maintenance (pre-analysis check):  Filter mobile phase  Wait for mobile phase to reach room temperature  Purge all the flow lines with mobile phase  Replace mobile phase as necessary  If the solvent tubing is not completely filled with mobile phase, purge the flow line  After periods of disuse or changes in ambient temperature, purge the flow line.  If the buffer is used as mobile phase, wash the back of plunger seal.  If auto sampler is used, verify the auto sampler rinse the bottle fully.  Purge the autosampler rinse solution  Checks for leaks  Checks the pump pressure  Perform a baseline check. B. Daily maintenance (post-analysis checks):  Wash the column  If the column is not going to be used for a long time, take off it from the system after washing, and cap both ends of the column, store it in a cool and dark place.  Clean the flow line or the rest of the instrument, as necessary  Store the solvent reservoir filters in rinse solvent to prevent them from drying out  Solvent Reservoir Maintenance: The material of containers is glass which is generally used and plastic which has high sensitivity inorganic analysis. 1. Always cover the containers 2. Prevent the evaporation of solvents 3. Prevent the dust enter the mobile phase 4. Reduce the vapours in the room 5. Do not make containers as this may lead to the creation of partial vacuum and pump system. 6. Do not expose the bottle to direct sunlight/wind 7. Always keep reservoir above solvent delivery system, ensures good siphon feed to pump and prevents starving pumping system. 8. Always use a suction filter, prevent the particulates from reaching pumps and clean the suction filter by sonicating in isopropanol or 1N nitric acid and water. 9. Good practice in the lab to label all bottles with composition inside as well as the date of prescription.  Mobile Phase: If the solvent has background absorption, the linear range will be smaller, the noise will be bigger and baseline stability, especially during gradient, runs will be poor. 1. The use of pure water systems is ideal as both inorganic and organic contaminants are removed. 2. Modern pure water systems give TOC values as indicated of total organic content in water. 3. Ghost peak will appear in gradient elution if water is bad quality. VI. PRECAUTIONS IN HANDLING SOME HPLC GRADE ORGANIC SOLVENTS. 1. Analytical grade THF contains butylated hydroxytoluene (BHT) as an oxidation inhibitor. As BHT has strong UV absorption, it is not added for HPLC grade THF. 2. As such, HPLC grade THF must be stored in cool and dark places to reduce oxidation to form peroxides. Accumulation of a large amount of organic peroxides may lead to explosions.
  • 6. International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470 @ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1623 A. Use of buffer solutions:  Buffer solutions must be filtered before use.  Buffer solutions must not be left in the system to avoid crystallization.  Effect on the pump- damage plunger and seal  Effect on the column- the creation of column voids  Effect on flow line- corrosion of stainless steel lines  Possible bacterial growth, especially phosphate buffers good medium for bacterial and fungus growth.  Ideally, solutions should be prepared fresh every day [28]. B. Premixing of Solvents: For isocratic systems, different solvents are premixed. Example: to prepare a 1000ml of 50:50 mixture of water/methanol.  Measure out 500 ml of water and 500ml of methanol separately, and then mix them together.  Measure 500ml of water and top up to the 1000ml mark of measuring cylinder/ volumetric flask with methanol.  Measure 500ml of methanol and top up to the 1000ml mark of measuring cylinder/ volumetric flask with water. All three methods will give different compositions, whichever method selected must be used consistently for the repeatable result. C. Pump maintenance:  Clean behind plunger seal when using buffer solution-  Automatic rinse kit for LC-10Avp and LC_20A series.  Useful spares to be maintained- plunger seals, plunger, check valves, filter frit for line filter.  Most’s tools required for maintenance work come with the pump accessories kit- do not lose them. D. Sample treatment: Always filtered sample with 0.2µ m or 0.45µ m filters before injection. If the sample contains a lot of solid, centrifuge it, and the filter the solution.  Select suitable solvent to dissolve the sample. Normally dissolve sample in the mobile phase. But do not select a high soluble solvent as a sample solvent. E. Injector maintenance: Manual injector: Precautions:  Do not use pointed or beveled needle tip. Use only flat and end type.  If pH higher than 10 is to be used, Delrin rotter seal is required.  When both manual and auto-injector are present in the system, auto sampler is always connected first, followed by manual.  Manual injector always set to load position when auto sampler is used and to inject position otherwise. F. Auto-Injector: General Maintenance  Purge flow line at least once a day.  Change purge liquid frequently.  Never used the salt solution as a purge liquid.  Used septums for sample vial recommended by the manufacturer only.  Septums should not be reused.  Do not fill vials to the brim. G. Detector maintenance:  UV-Vs detector: Check D2 lamp energy- Check the energy of reference cell. Change D2 lamp if the energy is lower than 800. (Note- do not touch the surface of D2 lamp by hand.)  Waste bottle maintenance:  Remember to empty waste bottle regularly.  Do not mix concentrated nitric acid with an organic solvent.  Do not immerse the drain tubing in the waste liquid as this may create back pressure on the flow cells [30].  Advantages of HPLC-  High Resolution  Good repeatability  High Sensitivity (ppm-ppb)  Small sample size  Moderate analyses condition  No need to vaporize the sample like GC  Easy to fractionate the sample and purify  Non-Destructive  Separations fast and efficient (High Resolution Power)  Continuous monitoring of the column effluent  It can be applied to the separation and analysis of very complex mixtures.  Accurate quantitative measurements
  • 7. International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470 @ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1624  Repetitive and reproducible analysis using the same column  Adsorption, partition, ion exchange and exclusion column separations are excellently made  HPLC is more versatile than GLC in some respects because it has the advantage of not being restricted to volatile and thermally stable solute and the choice of mobile and stationary phases is much wider in HPLC.  Both aqueous and non- aqueous samples can be analyzed with little or no sample pre-treatment  A variety of solvents and column packing are available, providing a high degree of selectivity for specific analysis.  It provides a means for determination of multiple components in a single analysis.  Application- HPLC is one of the most widely applied analytical separation techniques-[29]  Pharmaceutical:  Tablet dissolution of pharmaceutical dosages.  Shelf life determination of pharmaceutical products.  Identification of counterfeit drug products.  Pharmaceutical quality control.  Environmental:  Phenols in drinking water  Identification of diphenhydramine in sediment samples.  Bio monitoring of PAH pollution in high-altitude mountain lakes through the analysis of fish bile  Estrogens in coastal waters- The sewage source  Toxicity of tetracyclines and tetracycline degradation products to environmentally relevant bacteria  Assessment of TNT toxicity in sediment.  Forensics:  A mobile HPLC apparatus at dance parties- on- site identification and quantification of the drug Ecstasy.  Identification of anabolic steroids in serum, urine, sweat and hair  Forensic analysis of textile dyes  Simultaneous quantification of psychotherapeutic drugs in human plasma  Clinical:  Quantification of DEET in Human Urine.  Analysis of antibiotics  Increased urinary excretion of aquaporin 2 in patients with liver cirrhosis  Detection of endogenous neuropeptides in brain extracellular fluids.  Food & Flavour:  Ensuring soft drink consistency and quality  Analysis of vicinal dike tones in beer  Sugar analysis in fruit juices  Polycyclic aromatic hydrocarbons in Brazilian vegetables and fruits.  Trace analysis of military high explosives in agricultural crops  Stability of aspartame in the presence of glucose and vanillin [2, 10]. VII. CONCLUSION In this review, we include all fundamental aspects of HPLC technique. We describe the general technique of HPLC method and maintenance of the instrument. The general approach for these all point makes the precise quality of techniques. This method is widely accepted in the laboratory as well as in the clinical and forensic science lab due to the accuracy and precession and specificity with the HPLC. The only disadvantage of HPLC is high cost and complex methodology. VIII.REFERNCES: 1. Olander, D.P., Instrumental Methods of Analysis, Sixth Edition (Willard, Hobart H.; Merritt, Lynne L.; Dean, John A.; Settle, Frank A., Jr.). Journal of Chemical Education, 1984. 61(8): p. A222. 2. Lipka, E. and C. Vaccher, Quantitative analysis of drugs in biological matrices by HPLC hyphenated to fluorescence detection. Bioanalysis, 2015. 7(6): p. 743-762. 3. Choudhury, H., et al., Development and validation of RP-HPLC method: scope of application in the determination of oil solubility of paclitaxel. J Chromatogr Sci, 2014. 52(1): p. 68-74. 4. Md Sabir, A., M. Moloy, and P. S Bhasin, HPLC METHOD DEVELOPMENT AND VALIDATION: A REVIEW. Vol. 4. 2015. 39-46. 5. Malherbe, C.J., D. de Beer, and E. Joubert, Development of on-line high performance liquid chromatography (HPLC)-biochemical detection methods as tools in the identification of bioactives. Int J Mol Sci, 2012. 13(3): p. 3101-33.
  • 8. International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470 @ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1625 6. Hopmann, E., W. Arlt, and M. Minceva, Solvent system selection in counter-current chromatography using conductor-like screening model for real solvents. J Chromatogr A, 2011. 1218(2): p. 242-50. 7. Xiang, Y., Y. Liu, and M.L. Lee, Ultrahigh pressure liquid chromatography using elevated temperature. J Chromatogr A, 2006. 1104(1-2): p. 198-202. 8. Martin, M. and G. Guiochon, Effects of high pressure in liquid chromatography. J Chromatogr A, 2005. 1090(1-2): p. 16-38. 9. Seger, C., S. Sturm, and H. Stuppner, Mass spectrometry and NMR spectroscopy: modern high-end detectors for high resolution separation techniques--state of the art in natural product HPLC-MS, HPLC-NMR, and CE-MS hyphenations. Nat Prod Rep, 2013. 30(7): p. 970- 87. 10. Malviya, R., et al., High performance liquid chromatography: A short review. Vol. 2. 2010. 22-26. 11. Lababidi, S. and W. Schrader, Online normal- phase high-performance liquid chromatography/Fourier transform ion cyclotron resonance mass spectrometry: effects of different ionization methods on the characterization of highly complex crude oil mixtures. Rapid Commun Mass Spectrom, 2014. 28(12): p. 1345- 52. 12. Fueangfung, S., Y. Yuan, and S. Fang, Denaturing reversed-phase HPLC using a mobile phase containing urea for oligodeoxynucleotide analysis. Nucleosides Nucleotides Nucleic Acids, 2014. 33(7): p. 481-8. 13. Shibasaki, H., et al., Separation and quantitative determination of 6alpha-hydroxycortisol and 6beta-hydroxycortisol in human urine by high- performance liquid chromatography with ultraviolet absorption detection. Anal Bioanal Chem, 2012. 402(9): p. 2945-52. 14. Jerz, G., et al., Separation of amaranthine-type betacyanins by ion-pair high-speed countercurrent chromatography. J Chromatogr A, 2014. 1344: p. 42-50. 15. O'Fagain, C., P.M. Cummins, and B.F. O'Connor, Gel-filtration chromatography. Methods Mol Biol, 2011. 681: p. 25-33. 16. Pfaunmiller, E.L., et al., Affinity monolith chromatography: a review of principles and recent analytical applications. Anal Bioanal Chem, 2013. 405(7): p. 2133-45. 17. Leon-Gonzalez, M.E., et al., Two-dimensional liquid chromatography for direct chiral separations: a review. Biomed Chromatogr, 2014. 28(1): p. 59-83. 18. Xu, M., et al., [Determination of chlorogenic acid and baicalin in Yinhuang film-coated tablets by HPLC isocratic elution]. Zhongguo Zhong Yao Za Zhi, 2013. 38(5): p. 691-3. 19. D'Archivio, A.A., M.A. Maggi, and F. Ruggieri, Prediction of the retention of s-triazines in reversed-phase high-performance liquid chromatography under linear gradient-elution conditions. J Sep Sci, 2014. 37(15): p. 1930-6. 20. Marchetti, N., et al., Determination of adsorption isotherms by means of HPLC: adsorption mechanism elucidation and separation optimization. J Sep Sci, 2009. 32(5-6): p. 727-41. 21. Kraiczek, K.G., et al., Highly flexible UV-vis radiation sources and novel detection schemes for spectrophotometric HPLC detection. Anal Chem, 2014. 86(2): p. 1146-52. 22. Ping, B.T.Y., H.A. Aziz, and Z. Idris, Comparison of Peak-area Ratios and Percentage Peak Area Derived from HPLC-evaporative Light Scattering and Refractive Index Detectors for Palm Oil and its Fractions. J Oleo Sci, 2018. 67(3): p. 265-272. 23. Kupina, S. and M. Roman, Determination of total carbohydrates in wine and wine-like beverages by HPLC with a refractive index detector: First Action 2013.12. J AOAC Int, 2014. 97(2): p. 498- 505. 24. Pragst, F., M. Herzler, and B.T. Erxleben, Systematic toxicological analysis by high- performance liquid chromatography with diode array detection (HPLC-DAD). Clin Chem Lab Med, 2004. 42(11): p. 1325-40. 25. Raut, P.P. and S.Y. Charde, Simultaneous estimation of levodopa and carbidopa by RP- HPLC using a fluorescence detector: its application to a pharmaceutical dosage form. Luminescence, 2014. 29(7): p. 762-71.
  • 9. International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470 @ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 2 | Issue – 5 | Jul-Aug 2018 Page: 1626 26. Zhang, M., et al., Monitoring gradient profile on- line in micro- and nano-high performance liquid chromatography using conductivity detection. J Chromatogr A, 2016. 1460: p. 68-73. 27. Stubbs, C. and I. Kanfer, A stability-indicating high-performance liquid chromatographic assay of erythromycin estolate in pharmaceutical dosage forms. International Journal of Pharmaceutics, 1990. 63(2): p. 113-119. 28. Bakshi, M. and S. Singh, Development of validated stability-indicating assay methods-- critical review. J Pharm Biomed Anal, 2002. 28(6): p. 1011-40. 29. Michalski, R., Application of ion chromatography in clinical studies and pharmaceutical industry. Mini Rev Med Chem, 2014. 14(10): p. 862-72.