IGRT and Respiratory Gating: Can We See It and Can We Hit It? Joe L. Meadows, M.S. Medical Physicist The Lacks Cancer Center Saint Mary’s Health Care Grand Rapids, MI
Outline Basic Principles of Image Guided Radiation Therapy (IGRT) and Respiratory Gating (RG) Workflow: IGRT Image acquisition Positioning accuracy and image quality Potential use for Image Guided Radiation Therapy through selected clinical examples
Outline Different types of CBCT – Vendor specific approaches Workflow:  Respiratory Gating Establishing good sinus pattern acquisition Establishing the treatment “Gates” Monitoring patient respiratory rhythm during treatment
Objectives The participants will understand the current practice of IGRT and DGRT as it related to CBCT The participants will be aware of future trends of CBCT Understanding of the importance of Respiratory Gating and it’s relationship with IGRT Understanding the basic principles of respiratory pattern and treatment “gates”
IGRT- Working Definition Treatment systems with integrated imaging capability enabling a highly accurate patient set-up, compensating for any interfraction motion “ Old Days” IMRT, SBRT,.. “ Now-A Days”
OPTICAL TRACKING TECHNOLOGY Over the past 20-30 years our ability to “visualize” what we are treating (beyond skin marks) has dramatically changed! IGRT- Working Definition Prostate US H&N EPID http://biomedcentral.inist kV cone beam CT (CBCT)   University of Heidelberg radonc.ucsf.edu Medical Dosimetry, Vol. 32, No. 2, pp. 111-120, 2007 Calypso employs radiofrequency technology comparable to the global positioning system to continuously and precisely monitor biologically inert transponders implanted in the prostate. When prostate displacement exceeds predetermined thresholds, (i.e., 5 mm for more than 10 seconds) treatment can be halted temporarily to readjust the patient's position. Calypso® System Tracking and Target Localization during Radiotherapy for Prostate Cancer https://www.virginiamason.org Odd couple:  A prototype device combines a magnetic resonance imager with a linear accelerator. The blue cylinders facing each other are the imaging magnets.  The metal circle visible to the left at the back is a magnetic and radiation shield that protects the accelerator’s waveguide.  Credit: University of Alberta Cross Cancer Institute   Acquire five two-dimensional MRI images per second using  Compressive Sensing , acquire only the most important 10 percent of the image data in the first place-  FAST!
Choice Of IGRT Helical MVCT   6   YES   NO   Pretty good “ The data showed a sharper peak and a narrower spread of values for AlignRT than for OBI”   CBCT(kV)   6   YES   NO   YES   Optical Tracking  6  NO  YES  NO CBCT(MV)     6   YES   NO   Not great RF Tracking   6?   Surrogate   YES   NO “ Choose Your Poison” Technique   DOF   Internal Anatomy   RealTime   Soft Tissue Port Film   3   NO   NO   NO Novel-MRI   6   YES   YES?   YES The best choice might be a “hybrid” or combination of imaging technologies. ASTRO 2007 (abstract 43).
IGRT- Is It Really Necessary? We are not dealing with a static image!
Cone Beam CT Conventional CT ‘ Fan’ beam 1D detector 1 rotation = 1 slice Cone Beam CT ‘ Cone’ beam 2D detector 1 rotation = volume (many slices) CBCT is based on the same principle as conventional fan beam CT, except that an entire volumetric dataset can be obtained through a single rotation of the source and detector.  This is made possible by the use of a 2-D detector, as opposed to the one-dimensional (1-D) detectors used in conventional CT.
Cone Beam CT Available Cone Beam Systems Elekta Synergy™ Siemens Artiste™ Varian Trilogy™ Bench Top C-Arms
Cone Beam CT CBCT versus Fan Beam CT
Cone Beam CT Moving Gas
Cone Beam CT Motion Artifact
IGRT- Why Do We Need It?
IGRT- Why Do We Need It?
IGRT- What Does It Take? Patient set-up- Nothing new here Acquire CBCT Image 3D Image Reconstruction- 200 projections Image Fusion- CBCT and Planning CT Visual validation of match- What can go wrong? UCSF CRAZY Shifts! Re-position patient using Table shifts & Tx!
IGRT- High Z Artifacts UCSF kV MV
IGRT- Respiratory Motion The perfect transition since we now know the importance of IGRT, let’s now consider Respiratory Gating and 4D!
Step 1- 4D Acquisition Impact of Breathing-Related Motion on CT AAPM TG76 Report (Keall et al)
Comparison of ungated and gated case
4D Acquisition Siemens Uses Anzai Belt With Transducers Belt is placed around patient The transducers is held in place by belt The transducer signal is interfaced to CT scanner to provide the Respiratory Cycle This  correspondence allows  each slice to be “tagged” with the breathing cyle. (“4th” –D)
Data “Mining” Respiration is monitored by system. Patient does not need to maintain a constant breathing pattern.  Software categorize the breathing cycle into “BINS” which are decided by user. For example 0%-25%-50%-75%-100% IN and EX Or 0%-20%-40%-60%-80%-100% IN and EX The operator decides the number of  BINS for reconstruction
4D Reconstruction All of the determined “BINS” of data are reconstructed to produce a 4 Dimensional image of the Breathing patient
4D Visualization The 4 Dimensional patient is viewed by the Physician, Physicist/Dosimetrist prior to export to the RTP system.
 
 
4D Treatment Planning Disclaimer: I am by no means an “expert” on this topic. I will share with you what I know today This step involves using the 1,000 or more images to characterize the motion of the target and normal anatomy in designing the treatment fields
4D Treatment Planning Now to contour  on ALL the BIN’s of data (>1,000 slices!) “ Houstin, we need more time!”
 
4D Treatment Planning- “Poor Man’”
 
4D Treatment Planning Some of these new tools can “propagate” what is contoured on ONE data set (BIN) to all the remaining 900+ slices! The result is the creation of an “ITV” or Internal Target Volume which encompasses the GTV on ALL the 4D data-sets. However, this approach comes with the “cost” of treating more normal lung.  To spare more normal lung you could propagate the GTV only to the BIN’s or portions of the Respiratory Cycle that will be used to TREAT  the patient = GATED DELIVERY!
 
Respiratory Gated Delivery This is the final step from 4DCT Simulation   4D Treatment Planning   GATED Delivery The patient is set-up with the same Respiratory monitoring equipment used during the 4D Simulation Next, the Respiratory signal is optimized and the treatment “GATES” are set to coincide with the 4D Planning data The patient has the “routine” IGRT session followed by the GATED Treatment Delivery
 
Pretty good example
Pretty “bad” example
Moral To The Story Not ALL patients have a consistent breathing pattern. Some are good, others NOT This doesn’t necessarily mean you  cannot  use Respiratory Gating for the treatment delivery However you should ask yourself “is it worth it?” Using Gating for delivery will increase the total delivery time by a factor of  FOUR If you are performing SBRT, that is a very long treatment delivery!
IGRT and Respiratory Gating We have see the benefits of IGRT on set-up accuracy We have “seen” the effect of breathing motion and we can no longer afford to use static CT data, especially for SBRT! If we can image the patient (IGRT) AND Gate the delivery (Resp. Gating), this will enable us to: Dose Escalate and Hypo fractionate
Thank You for your time !

More Related Content

PPT
Respiratory Gating with IMRT
PPTX
Concept of bed in radiobiology
PPTX
Evolution of Intra-cavitary Brachytherapy
PPTX
Topic -Analysis of Patients specific Quality Assurance of IMRT/VMAT
Respiratory Gating with IMRT
Concept of bed in radiobiology
Evolution of Intra-cavitary Brachytherapy
Topic -Analysis of Patients specific Quality Assurance of IMRT/VMAT

What's hot (20)

PDF
Motion Management in Radiation Therapy
PDF
history of VMAT
PPTX
Respiratory gating
PPTX
Brachytherapy dosimetry systems .R
PPTX
The vmat vs other recent radiotherapy techniques
PPTX
Co60 vs Ir192
PPTX
Beam modifying devices
PPTX
Principals of brachytherapy shk
PPTX
Three dimensional conformal radiation therapy
PPTX
Electron Beam Therapy
PPTX
IMRT and 3DCRT
PPTX
Craniospinal irradiation
PPTX
ROSE CASE OF SRS BRAIN METASTASIS
PPTX
Breast Cancer Radiation Therapy: RT Plan evaluation & Recent Advances - 4DCT ...
PPT
Treatment Planning Ii Patient Data, Corrections, And Set Up
PPTX
Patient positional correction stategies in radiotherapy
PPTX
Plan evaluation in RADIOTHERAPY
PPTX
Image Guided Radiotherapy
PDF
Image Guided Radiation Therapy (IGRT)
PPTX
Total skin electron irradiation
Motion Management in Radiation Therapy
history of VMAT
Respiratory gating
Brachytherapy dosimetry systems .R
The vmat vs other recent radiotherapy techniques
Co60 vs Ir192
Beam modifying devices
Principals of brachytherapy shk
Three dimensional conformal radiation therapy
Electron Beam Therapy
IMRT and 3DCRT
Craniospinal irradiation
ROSE CASE OF SRS BRAIN METASTASIS
Breast Cancer Radiation Therapy: RT Plan evaluation & Recent Advances - 4DCT ...
Treatment Planning Ii Patient Data, Corrections, And Set Up
Patient positional correction stategies in radiotherapy
Plan evaluation in RADIOTHERAPY
Image Guided Radiotherapy
Image Guided Radiation Therapy (IGRT)
Total skin electron irradiation
Ad

Viewers also liked (20)

PPT
Image Guided Radiotherapy
PDF
Igrt Perspectives
PDF
Igrt Srt For Lung Cancer
PPTX
Image guided radiation therapy (IGRT)
PDF
Enhancement of clinical outcome using OBI and Cone Beam CT in Radiotherapy
PDF
TPS modeling and Beam commissioning at UIHC Radiation Oncology
PDF
IGRT: MVCBCT Calibration and Acceptance testing Procedure
PPTX
AlignRT as a Respiratory Motion Management Tool for SBRT
PPTX
Jason Morton - SOP Linac Commissioning
PPTX
Commissioning AlignRT with Minimal Disruption
PPT
Thastro08 igrt1
PPT
Radioterapia En El Nuevo Milenio Santa Rosafinal
PDF
Motion Management in Lung Cancer Radiotherapy
PDF
Igrt For Prostate Cancer
PPTX
Intracranial Calcification in Cone Beam CT & Medical CT
PPT
IMPROVISED RADIOTHERAPY TECHNIQUES IN TELE COBALT WITHOUT MLC
PDF
4 D Igrt
PPTX
Stereotactic Radiosurgery/ Radiotherapy
PPTX
Computer Tomography
PPT
Dental conebeamct
Image Guided Radiotherapy
Igrt Perspectives
Igrt Srt For Lung Cancer
Image guided radiation therapy (IGRT)
Enhancement of clinical outcome using OBI and Cone Beam CT in Radiotherapy
TPS modeling and Beam commissioning at UIHC Radiation Oncology
IGRT: MVCBCT Calibration and Acceptance testing Procedure
AlignRT as a Respiratory Motion Management Tool for SBRT
Jason Morton - SOP Linac Commissioning
Commissioning AlignRT with Minimal Disruption
Thastro08 igrt1
Radioterapia En El Nuevo Milenio Santa Rosafinal
Motion Management in Lung Cancer Radiotherapy
Igrt For Prostate Cancer
Intracranial Calcification in Cone Beam CT & Medical CT
IMPROVISED RADIOTHERAPY TECHNIQUES IN TELE COBALT WITHOUT MLC
4 D Igrt
Stereotactic Radiosurgery/ Radiotherapy
Computer Tomography
Dental conebeamct
Ad

Similar to Igrt And Resp Gating Final Version (20)

PDF
Surface Guided Radiotherapy for Accuracy, Volume Reduction, Real time Trackin...
PPT
4D-CBCT (Symmetry) - a useful tool to verify and treat traditional ITV withou...
PDF
IMAGE GUDIED RADIOTHERAPY IN OUR COUNTRY INDIA
PDF
Session 8 - IGRT_ Guidelines for Matching _ Alignment, Methods & Workflows.pdf
PPT
Fueling Start-Up Medical Companies through 3D Modeling
PPTX
How Centers Can​ Thrive in the Modern Era
PPTX
How Centers Can​ Thrive in the Modern Era
PPTX
behgal cancer institute
PDF
Implementing an End-to-End SGRT Workflow for Breath-Hold SABR
PPTX
Intensity modulated radiation therapy and Image guided radiation therapy
ODP
New Techniques in Radiotherapy
PPTX
Image guided radiation therapy
ODP
IMRT and 3D CRT in cervical Cancers
PPTX
Cyberknife At Saint Raphael’s Campus_revB
PPT
Dominion Vi 3D
PDF
Cyberknife An expertise state of art Technology to execute Stereotactic Robot...
PPTX
Brachytherapy IGABT
PDF
PPTX
Cone Beam Computed Tomography.pptx
Surface Guided Radiotherapy for Accuracy, Volume Reduction, Real time Trackin...
4D-CBCT (Symmetry) - a useful tool to verify and treat traditional ITV withou...
IMAGE GUDIED RADIOTHERAPY IN OUR COUNTRY INDIA
Session 8 - IGRT_ Guidelines for Matching _ Alignment, Methods & Workflows.pdf
Fueling Start-Up Medical Companies through 3D Modeling
How Centers Can​ Thrive in the Modern Era
How Centers Can​ Thrive in the Modern Era
behgal cancer institute
Implementing an End-to-End SGRT Workflow for Breath-Hold SABR
Intensity modulated radiation therapy and Image guided radiation therapy
New Techniques in Radiotherapy
Image guided radiation therapy
IMRT and 3D CRT in cervical Cancers
Cyberknife At Saint Raphael’s Campus_revB
Dominion Vi 3D
Cyberknife An expertise state of art Technology to execute Stereotactic Robot...
Brachytherapy IGABT
Cone Beam Computed Tomography.pptx

Igrt And Resp Gating Final Version

  • 1. IGRT and Respiratory Gating: Can We See It and Can We Hit It? Joe L. Meadows, M.S. Medical Physicist The Lacks Cancer Center Saint Mary’s Health Care Grand Rapids, MI
  • 2. Outline Basic Principles of Image Guided Radiation Therapy (IGRT) and Respiratory Gating (RG) Workflow: IGRT Image acquisition Positioning accuracy and image quality Potential use for Image Guided Radiation Therapy through selected clinical examples
  • 3. Outline Different types of CBCT – Vendor specific approaches Workflow: Respiratory Gating Establishing good sinus pattern acquisition Establishing the treatment “Gates” Monitoring patient respiratory rhythm during treatment
  • 4. Objectives The participants will understand the current practice of IGRT and DGRT as it related to CBCT The participants will be aware of future trends of CBCT Understanding of the importance of Respiratory Gating and it’s relationship with IGRT Understanding the basic principles of respiratory pattern and treatment “gates”
  • 5. IGRT- Working Definition Treatment systems with integrated imaging capability enabling a highly accurate patient set-up, compensating for any interfraction motion “ Old Days” IMRT, SBRT,.. “ Now-A Days”
  • 6. OPTICAL TRACKING TECHNOLOGY Over the past 20-30 years our ability to “visualize” what we are treating (beyond skin marks) has dramatically changed! IGRT- Working Definition Prostate US H&N EPID http://biomedcentral.inist kV cone beam CT (CBCT) University of Heidelberg radonc.ucsf.edu Medical Dosimetry, Vol. 32, No. 2, pp. 111-120, 2007 Calypso employs radiofrequency technology comparable to the global positioning system to continuously and precisely monitor biologically inert transponders implanted in the prostate. When prostate displacement exceeds predetermined thresholds, (i.e., 5 mm for more than 10 seconds) treatment can be halted temporarily to readjust the patient's position. Calypso® System Tracking and Target Localization during Radiotherapy for Prostate Cancer https://www.virginiamason.org Odd couple: A prototype device combines a magnetic resonance imager with a linear accelerator. The blue cylinders facing each other are the imaging magnets. The metal circle visible to the left at the back is a magnetic and radiation shield that protects the accelerator’s waveguide. Credit: University of Alberta Cross Cancer Institute Acquire five two-dimensional MRI images per second using Compressive Sensing , acquire only the most important 10 percent of the image data in the first place- FAST!
  • 7. Choice Of IGRT Helical MVCT 6 YES NO Pretty good “ The data showed a sharper peak and a narrower spread of values for AlignRT than for OBI” CBCT(kV) 6 YES NO YES Optical Tracking 6 NO YES NO CBCT(MV) 6 YES NO Not great RF Tracking 6? Surrogate YES NO “ Choose Your Poison” Technique DOF Internal Anatomy RealTime Soft Tissue Port Film 3 NO NO NO Novel-MRI 6 YES YES? YES The best choice might be a “hybrid” or combination of imaging technologies. ASTRO 2007 (abstract 43).
  • 8. IGRT- Is It Really Necessary? We are not dealing with a static image!
  • 9. Cone Beam CT Conventional CT ‘ Fan’ beam 1D detector 1 rotation = 1 slice Cone Beam CT ‘ Cone’ beam 2D detector 1 rotation = volume (many slices) CBCT is based on the same principle as conventional fan beam CT, except that an entire volumetric dataset can be obtained through a single rotation of the source and detector. This is made possible by the use of a 2-D detector, as opposed to the one-dimensional (1-D) detectors used in conventional CT.
  • 10. Cone Beam CT Available Cone Beam Systems Elekta Synergy™ Siemens Artiste™ Varian Trilogy™ Bench Top C-Arms
  • 11. Cone Beam CT CBCT versus Fan Beam CT
  • 12. Cone Beam CT Moving Gas
  • 13. Cone Beam CT Motion Artifact
  • 14. IGRT- Why Do We Need It?
  • 15. IGRT- Why Do We Need It?
  • 16. IGRT- What Does It Take? Patient set-up- Nothing new here Acquire CBCT Image 3D Image Reconstruction- 200 projections Image Fusion- CBCT and Planning CT Visual validation of match- What can go wrong? UCSF CRAZY Shifts! Re-position patient using Table shifts & Tx!
  • 17. IGRT- High Z Artifacts UCSF kV MV
  • 18. IGRT- Respiratory Motion The perfect transition since we now know the importance of IGRT, let’s now consider Respiratory Gating and 4D!
  • 19. Step 1- 4D Acquisition Impact of Breathing-Related Motion on CT AAPM TG76 Report (Keall et al)
  • 20. Comparison of ungated and gated case
  • 21. 4D Acquisition Siemens Uses Anzai Belt With Transducers Belt is placed around patient The transducers is held in place by belt The transducer signal is interfaced to CT scanner to provide the Respiratory Cycle This correspondence allows each slice to be “tagged” with the breathing cyle. (“4th” –D)
  • 22. Data “Mining” Respiration is monitored by system. Patient does not need to maintain a constant breathing pattern. Software categorize the breathing cycle into “BINS” which are decided by user. For example 0%-25%-50%-75%-100% IN and EX Or 0%-20%-40%-60%-80%-100% IN and EX The operator decides the number of BINS for reconstruction
  • 23. 4D Reconstruction All of the determined “BINS” of data are reconstructed to produce a 4 Dimensional image of the Breathing patient
  • 24. 4D Visualization The 4 Dimensional patient is viewed by the Physician, Physicist/Dosimetrist prior to export to the RTP system.
  • 25.  
  • 26.  
  • 27. 4D Treatment Planning Disclaimer: I am by no means an “expert” on this topic. I will share with you what I know today This step involves using the 1,000 or more images to characterize the motion of the target and normal anatomy in designing the treatment fields
  • 28. 4D Treatment Planning Now to contour on ALL the BIN’s of data (>1,000 slices!) “ Houstin, we need more time!”
  • 29.  
  • 30. 4D Treatment Planning- “Poor Man’”
  • 31.  
  • 32. 4D Treatment Planning Some of these new tools can “propagate” what is contoured on ONE data set (BIN) to all the remaining 900+ slices! The result is the creation of an “ITV” or Internal Target Volume which encompasses the GTV on ALL the 4D data-sets. However, this approach comes with the “cost” of treating more normal lung. To spare more normal lung you could propagate the GTV only to the BIN’s or portions of the Respiratory Cycle that will be used to TREAT the patient = GATED DELIVERY!
  • 33.  
  • 34. Respiratory Gated Delivery This is the final step from 4DCT Simulation  4D Treatment Planning  GATED Delivery The patient is set-up with the same Respiratory monitoring equipment used during the 4D Simulation Next, the Respiratory signal is optimized and the treatment “GATES” are set to coincide with the 4D Planning data The patient has the “routine” IGRT session followed by the GATED Treatment Delivery
  • 35.  
  • 38. Moral To The Story Not ALL patients have a consistent breathing pattern. Some are good, others NOT This doesn’t necessarily mean you cannot use Respiratory Gating for the treatment delivery However you should ask yourself “is it worth it?” Using Gating for delivery will increase the total delivery time by a factor of FOUR If you are performing SBRT, that is a very long treatment delivery!
  • 39. IGRT and Respiratory Gating We have see the benefits of IGRT on set-up accuracy We have “seen” the effect of breathing motion and we can no longer afford to use static CT data, especially for SBRT! If we can image the patient (IGRT) AND Gate the delivery (Resp. Gating), this will enable us to: Dose Escalate and Hypo fractionate
  • 40. Thank You for your time !