Implementing an end-to-end SGRT
workflow for breath-hold SABR
Presenters: Chelsea Carnall
Senior Regional Therapeutic Radiographer (Central)
Ellen Dear
Senior Regional Therapeutic Radiographer (East)
©GenesisCare 2022
Objective
©GenesisCare 2022 2
Discuss the implementation, benefits and challenges for SGRT to plan and
treat breath-hold SABR.
Introduction to GenesisCare
©GenesisCare 2022 3
● Global provider of cancer care - in
Australia, UK, Spain and USA
● We have 14 sites across the UK
SGRT at GenesisCare
©GenesisCare 2022 4
● At GenesisCare, we have
incorporated SGRT since January
2018, using VisionRT
● Rolled out throughout all our 14 sites
across the UK March - April 2018
● We are completely tattooless, using
SGRT for all treatment sites
Case Study Outline
©GenesisCare 2022 5
● 77yr old gentleman with castration-
resistant prostate cancer (diagnosed
2004), manubrial bony met since 2015,
decision to treat due to progression
● Internal specialist SABR MDT approved
for treatment plan of 30Gy/3#
● Treated on Varian Edge Linac with
6DoF couch (treated similar cases on
our Elekta VersaHD’s)
● Patient treated in breath-hold
VisionRT - SimRT & Breath-hold Coaching
©GenesisCare 2022 6
Overview
• SimRT used at planning
CT, which is used routinely
for DIBH and 4DCT
• Ceiling mounted camera
with a stable tracking point
on the patient’s anatomy
• Able to assess patient’s
ability to hold their breath,
can then offer coaching if
needed
Breath-hold coaching
• Coaching the patient into a
suitable breath-hold level
is vital
• We can assess the
patient’s suitability before
delivering any radiation
• In room coaching
○ Guide into breath–
hold
○ Check for back
arching and chest
breaths
Using SimRT for coaching
• SimRT allows us to accurately
assess the breath-hold level
• Asked the patient to hold for 25-
30 seconds (average treatment
beam delivery time)
• We can visualise any air ‘leakage’
©GenesisCare 2022 7
• 1x free-breathing 3DCT & 1x breath-hold 3DCT as per our
standard protocol for DIBH patients
• Free-breathing 3DCT to use for on initial set-up on treatment
• Breath-hold 3DCT for accurate assessment of breath-hold
level and treatment
• Can use SimRT to see depth of breath-hold
○ We were able to confidently complete scanning
and treatment without the need for a 4DCT
(ALARP)
○ SimRT gives us confidence that planning datasets
are accurate and will be reproducible on treatment
VisionRT - SimRT for this case
Dosimetric comparison: Free breathing vs Breath hold
©GenesisCare 2022 8
● Compared the body surface between the free-
breathing and breath-hold scans
● Showed a difference between 1-2mm
● The planned (BH) CTV was expanded by 2mm to
derive a theoretical and conservative free-
breathing CTV volume
● The PTV was then expanded by a further 3mm
as per SABR protocol
Planned volume on
BH (cc)
Volume on FB (cc)* % Difference
CTV 66.5 85.9 29.20%
PTV (3mm margin) 110.2 134.7 22.25%
* Estimated by expanding BH margins by 2 mm Ant-Post, equivalent to FB surface shifts
Green = FB
Red = BH
This case
Treatment Set Up - AlignRT
©GenesisCare 2022 9
● Wingboard & indexed knee rest
● Used postural mode to aid with arm
position
In-room set up
• Two surface structures - FB and BH, created
from the two 3DCT scans
• Initial set-up in free-breathing using AlignRT to
get patient to the isocentre position and correct
for rotations (after #1 can also use saved bed
position)
• Patient holds breath and use BH surface
structure to adjust lateral and longitudinal
position - vertical controlled by level of breath-
hold
• Patient can use real-time-coach (RTC) as a
visual aid.
(SimRT-
VisionRT)
This case
Treatment Delivery - AlignRT
©GenesisCare 2022 10
• CBCT used to confirm internal position -
performed in breath-hold
• Treatment delivery is gated, AlignRT
gates 0.2cm movement either side of CT
breath-hold level
(SimRT-
VisionRT)
Overview
• 2 x 6MV FFF VMAT beams
• Patient in control at all times and able to
comfortably hold breath throughout
• When not in correct position, AlignRT
gates the beam
Daily CBCT
workflow by
SABR Technique
Simple Complex Spine
Pre-CBCT Y Y Y
Mid-CBCT At
operator
discretion
At operator
discretion
Y
Post-CBCT Y Y Y
Case Study #1 - AlignRT & CBCT workflow
Initial CBCT showed a max
translational shift of 1.8mm
and rotation of 1.2 degrees.
Translations and rotations
applied using the 6DOF
couch.
During treatment AlignRT
picked up 3mm longitudinal
difference.
Treatment was paused and
a CBCT taken.
CBCT confirmed 3mm
longitudinal shift, applied
and treatment resumed.
Post CBCT showed
intrafractional
motion/discrepancy of
<1mm.
©GenesisCare 2022 11
Reflection
Case Study #2&3 - AlignRT & CBCT workflow
©GenesisCare 2022 12
● AlignRT allowed for precise set-up,
reducing need for additional CBCTs and
minimising intrafractional motion.
● AlignRT allowed for positional monitoring
during treatment and accurate tracking of
a consistent breath hold.
● AlignRT detected movement of the
patient. Radiographers were able to
pause treatment, re-image and assess
this before any further treatment was
delivered inaccurately. The shift was
corroborated with CBCT, was applied,
and treatment continued to be delivered
with complete accuracy.
#2 & #3 imaging
● #2 and #3 patient stability improved -
● Able to accurately set up and monitor
patient’s breath-hold with AlignRT
● All CBCTs, including the initial set up
CBCT, showed <1mm shifts and <1
degree rotations.
Daily CBCT tolerances for simple SABR
Translations ≤0.5cm
Shift & treat
>0.5cm and ≤1cm
Shift & re-image
>1cm re-
set up
Rotations 3 degrees
Limitations and troubleshooting with SGRT
©GenesisCare 2022 13
Tracking point location in
SimRT
Tracking point can be lost as CT bed
moves through the scanner.
Troubleshooted by choosing tracking
point at end of topogram when patient
inside the scanner.
Patients with small breath-hold
With a smaller breath-hold, it is harder
to determine if a patients is in breath-
hold or free breathing.
Troubleshooted by coaching patients,
assessing suitability for breath hold, or
sometimes sending away to practice.
Initial struggle with skin tones
Initially we found AlignRT would
“flicker” or “reflect” and lose the skin
surface. Spoke to VisionRT who
advised to use the skin tone setting to
choose appropriately which has helped.
CBCT not gated - relies on
radiographer observation to
interrupt
Radiographers always watching when
patient in breath-hold to interrupt if
patient releases breath.
To Conclude
©GenesisCare 2022 14
● Treating in breath-hold allowed us to limit dose to the patient as we didn’t have to perform a
4DCT (ALARP)
● Treating in breath-hold allowed for reduced irradiated volumes due to minimal target-motion.
● At CT, SimRT allows us to assess level and consistency of breath-hold.
● AlignRT allows patient to be in control of breath-hold during gated treatment
● AlignRT detected patient movement during treatment - able to re-CBCT and apply moves to
ensure accurate treatment
● SGRT allows us to deliver treatment confidently and accurately without need for permanent
tattoo marks.
Studies
©GenesisCare 2022 15
● Hamming, V.C., Visser, C., Batin, E., McDermott, L.N., Busz, D.M., Both, S., Langendijk,
J.A. and Sijtsema, N.M. (2019). Evaluation of a 3D surface imaging system for deep
inspiration breath-hold patient positioning and intra-fraction monitoring. Radiation Oncology,
14(1). doi:10.1186/s13014-019-1329-6.
● Heinzerling, J.H., Biester, E.C., Robinson, M., Moeller, B.J., Prabhu, R.S., Ward, M.C.,
Corso, C.D., Gregory, J., Hampton, C.J., Burri, S.H. and Foster, R.D. (2021). Prospective
Study of Surface Guided Radiation Therapy (SGRT) for Breath Hold SBRT Treatments of
the Lung: Analysis of Reliability of Surface Guidance Alone for Internal Tumor Position
During Breath Hold. International Journal of Radiation Oncology*Biology*Physics, 111(3),
p.e539. doi:10.1016/j.ijrobp.2021.07.1468.
● Nankali, S., Hansen, R., Worm, E., Yates, E.S., Thomsen, M.S., Offersen, B. and Poulsen,
P.R. (2022). Accuracy and potential improvements of surface-guided breast cancer
radiotherapy in deep inspiration breath-hold with daily image-guidance. Physics in Medicine
and Biology, [online] 67(19). doi:10.1088/1361-6560/ac9109.
References
©GenesisCare 2022 16
● Carnall, C. (2022). Varian Edge Linac. [Photo].
● Dear, E. (2022). PTV outline. [Photo].
● Dear, E. (2022). VisionRT SimRT. [Photo].
● Vision RT. (n.d.). Breast And Deep Inspiration Breath Hold (DIBH). [online] Available at:
https://www.visionrt.com/applications/dibh/ [Accessed 7 Nov. 2022].
● Vision RT. (n.d.). SimRT. [online] Available at: https://www.visionrt.com/simrt/ [Accessed 7 Nov.
2022].
● Vision RT. (n.d.). Virtual Learning. [online] Available at: https://www.visionrt.com/virtual/
[Accessed 7 Nov. 2022].
● www.genesiscare.com. (n.d.). UK Cancer Treatment Centres & Diagnostic Clinics |
GenesisCare UK. [online] Available at: https://www.genesiscare.com/uk/our-centres [Accessed
7 Nov. 2022].
● Vision RT. (n.d) Real Time Coach Display. [online] Available at:
https://www.visionrt.com/products/the-real-time-coach-display/ [Accessed 9 Nov. 2022]
Further questions?
©GenesisCare 2022 17
If you have any further questions about our experience please feel free to email us:
chelsea.carnall@genesiscare.co.uk
ellen.dear@genesiscare.co.uk

More Related Content

PPTX
Starting out with DIBH
PDF
Surface Guided Radiotherapy for Accuracy, Volume Reduction, Real time Trackin...
PPTX
Adaptive radiotherapy in head and neck cancer
ODP
IMRT and 3D CRT in cervical Cancers
PPTX
Final ICRU 62 ( International commission on radiation units and measurements)
PPTX
Icru reports in external beam radiotherapy
PPTX
Icru – 83 dr. upasna
PPTX
SBRT Liver when and how.pptx
Starting out with DIBH
Surface Guided Radiotherapy for Accuracy, Volume Reduction, Real time Trackin...
Adaptive radiotherapy in head and neck cancer
IMRT and 3D CRT in cervical Cancers
Final ICRU 62 ( International commission on radiation units and measurements)
Icru reports in external beam radiotherapy
Icru – 83 dr. upasna
SBRT Liver when and how.pptx

What's hot (20)

PDF
SGRT for DIBH: from zero to hero 
PPTX
Three dimensional conformal radiation therapy
PPTX
Smart radiotherapy
PPTX
Prophylactic cranial irradiation
PPTX
Multileaf collimator basic concepts.pptx
PDF
Image guided adaptive radiotherapy
PPTX
TARGET DELINEATION OF SOFT TISSUE SARCOMA
PPTX
Brachytherapy in breast cancer
PPTX
Patient positional correction stategies in radiotherapy
PPTX
Hippocampal sparing whole brain radiation therapy- Making a case!
PPT
Intra Operative Radiotherapy
PPTX
Plan evaluation in RADIOTHERAPY
PPTX
HOLISTIC APPROACH IN WHOLE BRAIN RADIATION IN BRAIN METS
PDF
SRS & SBRT - Unflattened Beam
PDF
Motion Management in Radiation Therapy
PPTX
RT breast apbi
PPTX
Esophagus Contouring.pptx
PPTX
DECISION MAKING IN HEAD AND NECK CANCER RE-IRRADIATION
ODP
Concurrent Chemoradiation in Postoperative Setting In LAHNC. A comparision of...
PPTX
Icru seminar 38
SGRT for DIBH: from zero to hero 
Three dimensional conformal radiation therapy
Smart radiotherapy
Prophylactic cranial irradiation
Multileaf collimator basic concepts.pptx
Image guided adaptive radiotherapy
TARGET DELINEATION OF SOFT TISSUE SARCOMA
Brachytherapy in breast cancer
Patient positional correction stategies in radiotherapy
Hippocampal sparing whole brain radiation therapy- Making a case!
Intra Operative Radiotherapy
Plan evaluation in RADIOTHERAPY
HOLISTIC APPROACH IN WHOLE BRAIN RADIATION IN BRAIN METS
SRS & SBRT - Unflattened Beam
Motion Management in Radiation Therapy
RT breast apbi
Esophagus Contouring.pptx
DECISION MAKING IN HEAD AND NECK CANCER RE-IRRADIATION
Concurrent Chemoradiation in Postoperative Setting In LAHNC. A comparision of...
Icru seminar 38
Ad

Similar to Implementing an End-to-End SGRT Workflow for Breath-Hold SABR (20)

PDF
Breath-hold for hearing-impaired patients
PDF
Feasibility Study of Deep Inspiration Breath-Hold Technique with AlignRT InBo...
PPTX
SBRT: The Importance of Continuous Motion Management​
PPTX
Management dell’organ motion nei trattamenti stereo-RT e radiochirurgici:
PDF
Advancing Motion Control: An Evaluation of VRT In-Bore Guided Breath-Hold Usi...
PDF
Increased Efficiency through SGRT Usage for Prostate Treatments on a Bore-bas...
PPTX
Lung sbrt ppt
PPTX
AlignRT as a Respiratory Motion Management Tool for SBRT
PPTX
Motion management in Radiation Oncology - 2020
PDF
Feasibility and Tolerability of Surface Guided Radiotherapy in Breath-Hold Li...
PDF
AlignRT’s Respiratory Module: Results of Pre-clinical Testing and Initial Exp...
PDF
Surface Guided Simulation for Improved Accuracy in Breast Treatments
PPTX
SBRT in lung cancer
PDF
Breath Hold Insurance: How SGRT Gives Therapists the Tools for an Accurate Br...
PPTX
MOTION MANAGEMANT IN LUNG SBRT BY DR KANHU CHARAN PATRO
PPT
Igrt And Resp Gating Final Version
PPTX
Managing Motion from End to End
PDF
A Surface Guided Method to Improved Pancreatic SBRT Patient Experience
PDF
We Can See Clearly Now: Implementation of Surface Guided Radiation Therapy
Breath-hold for hearing-impaired patients
Feasibility Study of Deep Inspiration Breath-Hold Technique with AlignRT InBo...
SBRT: The Importance of Continuous Motion Management​
Management dell’organ motion nei trattamenti stereo-RT e radiochirurgici:
Advancing Motion Control: An Evaluation of VRT In-Bore Guided Breath-Hold Usi...
Increased Efficiency through SGRT Usage for Prostate Treatments on a Bore-bas...
Lung sbrt ppt
AlignRT as a Respiratory Motion Management Tool for SBRT
Motion management in Radiation Oncology - 2020
Feasibility and Tolerability of Surface Guided Radiotherapy in Breath-Hold Li...
AlignRT’s Respiratory Module: Results of Pre-clinical Testing and Initial Exp...
Surface Guided Simulation for Improved Accuracy in Breast Treatments
SBRT in lung cancer
Breath Hold Insurance: How SGRT Gives Therapists the Tools for an Accurate Br...
MOTION MANAGEMANT IN LUNG SBRT BY DR KANHU CHARAN PATRO
Igrt And Resp Gating Final Version
Managing Motion from End to End
A Surface Guided Method to Improved Pancreatic SBRT Patient Experience
We Can See Clearly Now: Implementation of Surface Guided Radiation Therapy
Ad

More from SGRT Community (20)

PDF
Efficiency in the Workplace, Natasha Calle, RT
PDF
Maskless Head & Neck Treatments, Danae McCarthy, RT(T)
PDF
Broad Implementation of SGRT: Faster, More Accurate Treatments
PDF
Clinical Experiences with InBore Surface-Guided Radiation Therapy
PDF
Implementing SGRT on All Electron Set-ups
PDF
Designing an Optimal Immobilization Device for SGRT Enhanced Treatment Delivery
PDF
Using SGRT to Improve Prone Breast Setups
PDF
Tattoos and Marks: How SGRT Makes Them Unnecessary for Treatments
PDF
The Role of SGRT for Online Adaptive SBRT Treatments Using AlignRT InBore on ...
PDF
Implementing a Markless Workflow in a Community Hospital
PDF
Cross-Vendor Comparison of Beam Hold Duration (Latency)
PDF
Reimbursement of SGRT in Free-Standing Clinics
PDF
Pushing the Boundaries in SGRT, Mike Tallhamer, DABR
PDF
Using Cherenkov Imaging and Scintillation Dosimetry to Quantify Contralateral...
PDF
Improving Plan Quality & Safety Using Surface Guided Planning and Dose Visual...
PDF
Clinical Experience with Noncoplanar Treatments
PDF
AlignRT for Proton Therapy Breast Treatments
PDF
AlignRT Benefits for Protons – Our Experience
PDF
Innovations and Challenges: Implementing SGRT in Proton Radiation Therapy
PDF
Future of SGRT: Beyond Traditional Motion Management
Efficiency in the Workplace, Natasha Calle, RT
Maskless Head & Neck Treatments, Danae McCarthy, RT(T)
Broad Implementation of SGRT: Faster, More Accurate Treatments
Clinical Experiences with InBore Surface-Guided Radiation Therapy
Implementing SGRT on All Electron Set-ups
Designing an Optimal Immobilization Device for SGRT Enhanced Treatment Delivery
Using SGRT to Improve Prone Breast Setups
Tattoos and Marks: How SGRT Makes Them Unnecessary for Treatments
The Role of SGRT for Online Adaptive SBRT Treatments Using AlignRT InBore on ...
Implementing a Markless Workflow in a Community Hospital
Cross-Vendor Comparison of Beam Hold Duration (Latency)
Reimbursement of SGRT in Free-Standing Clinics
Pushing the Boundaries in SGRT, Mike Tallhamer, DABR
Using Cherenkov Imaging and Scintillation Dosimetry to Quantify Contralateral...
Improving Plan Quality & Safety Using Surface Guided Planning and Dose Visual...
Clinical Experience with Noncoplanar Treatments
AlignRT for Proton Therapy Breast Treatments
AlignRT Benefits for Protons – Our Experience
Innovations and Challenges: Implementing SGRT in Proton Radiation Therapy
Future of SGRT: Beyond Traditional Motion Management

Recently uploaded (20)

PPTX
Nancy Caroline Emergency Paramedic Chapter 11
PPTX
Arthritis Types, Signs & Treatment with physiotherapy management
PPTX
unit1-introduction of nursing education..
PPTX
POSTURE.pptx......,............. .........
PPTX
Rheumatic heart diseases with Type 2 Diabetes Mellitus
PPTX
Newer Technologies in medical field.pptx
PDF
health promotion and maintenance of elderly
DOCX
ch 9 botes for OB aka Pregnant women eww
PPTX
FUNCTIONS OF BLOOD PART I AND PART 2 WHOLE
PPTX
Nancy Caroline Emergency Paramedic Chapter 4
DOCX
Copies if quanti.docxsegdfhfkhjhlkjlj,klkj
PPTX
Acute renal failure.pptx for BNs 2nd year
PPTX
Nancy Caroline Emergency Paramedic Chapter 16
PPTX
Public Health. Disasater mgt group 1.pptx
PPTX
Full Slide Deck - SY CF Talk Adelaide 10June.pptx
PPTX
Nepal health service act.pptx by Sunil Sharma
PPTX
First Aid and Basic Life Support Training.pptx
PPTX
Understanding The Self : 1Sexual health
PDF
Essentials of Hysteroscopy at World Laparoscopy Hospital
PPT
12.08.2025 Dr. Amrita Ghosh_Stocks Standards_ Smart_Inventory Management_GCLP...
Nancy Caroline Emergency Paramedic Chapter 11
Arthritis Types, Signs & Treatment with physiotherapy management
unit1-introduction of nursing education..
POSTURE.pptx......,............. .........
Rheumatic heart diseases with Type 2 Diabetes Mellitus
Newer Technologies in medical field.pptx
health promotion and maintenance of elderly
ch 9 botes for OB aka Pregnant women eww
FUNCTIONS OF BLOOD PART I AND PART 2 WHOLE
Nancy Caroline Emergency Paramedic Chapter 4
Copies if quanti.docxsegdfhfkhjhlkjlj,klkj
Acute renal failure.pptx for BNs 2nd year
Nancy Caroline Emergency Paramedic Chapter 16
Public Health. Disasater mgt group 1.pptx
Full Slide Deck - SY CF Talk Adelaide 10June.pptx
Nepal health service act.pptx by Sunil Sharma
First Aid and Basic Life Support Training.pptx
Understanding The Self : 1Sexual health
Essentials of Hysteroscopy at World Laparoscopy Hospital
12.08.2025 Dr. Amrita Ghosh_Stocks Standards_ Smart_Inventory Management_GCLP...

Implementing an End-to-End SGRT Workflow for Breath-Hold SABR

  • 1. // Month Year Implementing an end-to-end SGRT workflow for breath-hold SABR Presenters: Chelsea Carnall Senior Regional Therapeutic Radiographer (Central) Ellen Dear Senior Regional Therapeutic Radiographer (East) ©GenesisCare 2022
  • 2. Objective ©GenesisCare 2022 2 Discuss the implementation, benefits and challenges for SGRT to plan and treat breath-hold SABR.
  • 3. Introduction to GenesisCare ©GenesisCare 2022 3 ● Global provider of cancer care - in Australia, UK, Spain and USA ● We have 14 sites across the UK
  • 4. SGRT at GenesisCare ©GenesisCare 2022 4 ● At GenesisCare, we have incorporated SGRT since January 2018, using VisionRT ● Rolled out throughout all our 14 sites across the UK March - April 2018 ● We are completely tattooless, using SGRT for all treatment sites
  • 5. Case Study Outline ©GenesisCare 2022 5 ● 77yr old gentleman with castration- resistant prostate cancer (diagnosed 2004), manubrial bony met since 2015, decision to treat due to progression ● Internal specialist SABR MDT approved for treatment plan of 30Gy/3# ● Treated on Varian Edge Linac with 6DoF couch (treated similar cases on our Elekta VersaHD’s) ● Patient treated in breath-hold
  • 6. VisionRT - SimRT & Breath-hold Coaching ©GenesisCare 2022 6 Overview • SimRT used at planning CT, which is used routinely for DIBH and 4DCT • Ceiling mounted camera with a stable tracking point on the patient’s anatomy • Able to assess patient’s ability to hold their breath, can then offer coaching if needed Breath-hold coaching • Coaching the patient into a suitable breath-hold level is vital • We can assess the patient’s suitability before delivering any radiation • In room coaching ○ Guide into breath– hold ○ Check for back arching and chest breaths Using SimRT for coaching • SimRT allows us to accurately assess the breath-hold level • Asked the patient to hold for 25- 30 seconds (average treatment beam delivery time) • We can visualise any air ‘leakage’
  • 7. ©GenesisCare 2022 7 • 1x free-breathing 3DCT & 1x breath-hold 3DCT as per our standard protocol for DIBH patients • Free-breathing 3DCT to use for on initial set-up on treatment • Breath-hold 3DCT for accurate assessment of breath-hold level and treatment • Can use SimRT to see depth of breath-hold ○ We were able to confidently complete scanning and treatment without the need for a 4DCT (ALARP) ○ SimRT gives us confidence that planning datasets are accurate and will be reproducible on treatment VisionRT - SimRT for this case
  • 8. Dosimetric comparison: Free breathing vs Breath hold ©GenesisCare 2022 8 ● Compared the body surface between the free- breathing and breath-hold scans ● Showed a difference between 1-2mm ● The planned (BH) CTV was expanded by 2mm to derive a theoretical and conservative free- breathing CTV volume ● The PTV was then expanded by a further 3mm as per SABR protocol Planned volume on BH (cc) Volume on FB (cc)* % Difference CTV 66.5 85.9 29.20% PTV (3mm margin) 110.2 134.7 22.25% * Estimated by expanding BH margins by 2 mm Ant-Post, equivalent to FB surface shifts Green = FB Red = BH
  • 9. This case Treatment Set Up - AlignRT ©GenesisCare 2022 9 ● Wingboard & indexed knee rest ● Used postural mode to aid with arm position In-room set up • Two surface structures - FB and BH, created from the two 3DCT scans • Initial set-up in free-breathing using AlignRT to get patient to the isocentre position and correct for rotations (after #1 can also use saved bed position) • Patient holds breath and use BH surface structure to adjust lateral and longitudinal position - vertical controlled by level of breath- hold • Patient can use real-time-coach (RTC) as a visual aid. (SimRT- VisionRT)
  • 10. This case Treatment Delivery - AlignRT ©GenesisCare 2022 10 • CBCT used to confirm internal position - performed in breath-hold • Treatment delivery is gated, AlignRT gates 0.2cm movement either side of CT breath-hold level (SimRT- VisionRT) Overview • 2 x 6MV FFF VMAT beams • Patient in control at all times and able to comfortably hold breath throughout • When not in correct position, AlignRT gates the beam Daily CBCT workflow by SABR Technique Simple Complex Spine Pre-CBCT Y Y Y Mid-CBCT At operator discretion At operator discretion Y Post-CBCT Y Y Y
  • 11. Case Study #1 - AlignRT & CBCT workflow Initial CBCT showed a max translational shift of 1.8mm and rotation of 1.2 degrees. Translations and rotations applied using the 6DOF couch. During treatment AlignRT picked up 3mm longitudinal difference. Treatment was paused and a CBCT taken. CBCT confirmed 3mm longitudinal shift, applied and treatment resumed. Post CBCT showed intrafractional motion/discrepancy of <1mm. ©GenesisCare 2022 11
  • 12. Reflection Case Study #2&3 - AlignRT & CBCT workflow ©GenesisCare 2022 12 ● AlignRT allowed for precise set-up, reducing need for additional CBCTs and minimising intrafractional motion. ● AlignRT allowed for positional monitoring during treatment and accurate tracking of a consistent breath hold. ● AlignRT detected movement of the patient. Radiographers were able to pause treatment, re-image and assess this before any further treatment was delivered inaccurately. The shift was corroborated with CBCT, was applied, and treatment continued to be delivered with complete accuracy. #2 & #3 imaging ● #2 and #3 patient stability improved - ● Able to accurately set up and monitor patient’s breath-hold with AlignRT ● All CBCTs, including the initial set up CBCT, showed <1mm shifts and <1 degree rotations. Daily CBCT tolerances for simple SABR Translations ≤0.5cm Shift & treat >0.5cm and ≤1cm Shift & re-image >1cm re- set up Rotations 3 degrees
  • 13. Limitations and troubleshooting with SGRT ©GenesisCare 2022 13 Tracking point location in SimRT Tracking point can be lost as CT bed moves through the scanner. Troubleshooted by choosing tracking point at end of topogram when patient inside the scanner. Patients with small breath-hold With a smaller breath-hold, it is harder to determine if a patients is in breath- hold or free breathing. Troubleshooted by coaching patients, assessing suitability for breath hold, or sometimes sending away to practice. Initial struggle with skin tones Initially we found AlignRT would “flicker” or “reflect” and lose the skin surface. Spoke to VisionRT who advised to use the skin tone setting to choose appropriately which has helped. CBCT not gated - relies on radiographer observation to interrupt Radiographers always watching when patient in breath-hold to interrupt if patient releases breath.
  • 14. To Conclude ©GenesisCare 2022 14 ● Treating in breath-hold allowed us to limit dose to the patient as we didn’t have to perform a 4DCT (ALARP) ● Treating in breath-hold allowed for reduced irradiated volumes due to minimal target-motion. ● At CT, SimRT allows us to assess level and consistency of breath-hold. ● AlignRT allows patient to be in control of breath-hold during gated treatment ● AlignRT detected patient movement during treatment - able to re-CBCT and apply moves to ensure accurate treatment ● SGRT allows us to deliver treatment confidently and accurately without need for permanent tattoo marks.
  • 15. Studies ©GenesisCare 2022 15 ● Hamming, V.C., Visser, C., Batin, E., McDermott, L.N., Busz, D.M., Both, S., Langendijk, J.A. and Sijtsema, N.M. (2019). Evaluation of a 3D surface imaging system for deep inspiration breath-hold patient positioning and intra-fraction monitoring. Radiation Oncology, 14(1). doi:10.1186/s13014-019-1329-6. ● Heinzerling, J.H., Biester, E.C., Robinson, M., Moeller, B.J., Prabhu, R.S., Ward, M.C., Corso, C.D., Gregory, J., Hampton, C.J., Burri, S.H. and Foster, R.D. (2021). Prospective Study of Surface Guided Radiation Therapy (SGRT) for Breath Hold SBRT Treatments of the Lung: Analysis of Reliability of Surface Guidance Alone for Internal Tumor Position During Breath Hold. International Journal of Radiation Oncology*Biology*Physics, 111(3), p.e539. doi:10.1016/j.ijrobp.2021.07.1468. ● Nankali, S., Hansen, R., Worm, E., Yates, E.S., Thomsen, M.S., Offersen, B. and Poulsen, P.R. (2022). Accuracy and potential improvements of surface-guided breast cancer radiotherapy in deep inspiration breath-hold with daily image-guidance. Physics in Medicine and Biology, [online] 67(19). doi:10.1088/1361-6560/ac9109.
  • 16. References ©GenesisCare 2022 16 ● Carnall, C. (2022). Varian Edge Linac. [Photo]. ● Dear, E. (2022). PTV outline. [Photo]. ● Dear, E. (2022). VisionRT SimRT. [Photo]. ● Vision RT. (n.d.). Breast And Deep Inspiration Breath Hold (DIBH). [online] Available at: https://www.visionrt.com/applications/dibh/ [Accessed 7 Nov. 2022]. ● Vision RT. (n.d.). SimRT. [online] Available at: https://www.visionrt.com/simrt/ [Accessed 7 Nov. 2022]. ● Vision RT. (n.d.). Virtual Learning. [online] Available at: https://www.visionrt.com/virtual/ [Accessed 7 Nov. 2022]. ● www.genesiscare.com. (n.d.). UK Cancer Treatment Centres & Diagnostic Clinics | GenesisCare UK. [online] Available at: https://www.genesiscare.com/uk/our-centres [Accessed 7 Nov. 2022]. ● Vision RT. (n.d) Real Time Coach Display. [online] Available at: https://www.visionrt.com/products/the-real-time-coach-display/ [Accessed 9 Nov. 2022]
  • 17. Further questions? ©GenesisCare 2022 17 If you have any further questions about our experience please feel free to email us: chelsea.carnall@genesiscare.co.uk ellen.dear@genesiscare.co.uk