Monitoring of Inorganic Ions in Ambient Air
Mr. S. D. Joshi
Scientist
Air Pollution Control Division, NEERI, Nagpur - 440020
Introduction
The air besides containing gaseous pollutants like S02, NOx, CO, HC also
contains particulate matter which is made up of complex organic compounds,
inorganic elements and their oxides and secondary particulates viz, sulphate,
nitrate etc. These secondary particulates are formed from their precursors in
presence of sunlight humidity and metal ion species by oxidation or reduction
processes. The acid precursors viz. sulfate and nitrate, and chloride require
more concern as these have deleterious effects on aquatic and forest eco-
systems and are potentially harmful for a variety of building materials, accelerate
corrosion of metals, damage sculptures and cause chronic respiratory problems
in human beings. In view of possible large social, economic, ecological and
aesthetic value, it has become important to analyse these pollutants in the
ambient air.
These particulates are emitted in the atmosphere by different
anthropogenic and natural sources. Sulphate (SO4-) is formed in the
atmosphere by reduction or oxidation in presence of hydroxy! ion (OH"). The
main sources of sulfur oxides are power houses, sulphuric acid, petroleum
industries, oil refineries, coal burning and domestic use of fuels. Acid
manufacturing, automobile exhaust, explosive industry, fuel burning etc. are the
main sources of nitrogen oxides. Chloride occurs predominantly in coastal cities.
Other sources of chloride are coal burning, waste incineration, automobile
exhaust, burning of chlorinated hydrocarbons. Fluoride is emitted in the
atmosphere by fertiliser industry, aluminium industry, steel making and acid
making industries.
Analytical Techniques
Different methods are available for the analysis of anions. The
conventional methods, viz., titrimetric, colorimetric, spectrophotometric,
gravimetric and ion selective electrode methods though commonly used, are
ion specific, time consuming and less sensitive. Ion Chromatography (IC) is the
only instrumental technique which can be used for rapid, sequential analysis of
different anions simultaneously.
Ion Chromatography
Chromatography encompasses wide range of techniques of separation of
the components of a mixture as a result of differential migration of the component
molecules between stationary phase and mobile phase. It was Hamish, Small of
DOW chemical corporation whose pioneering work in mid seventies in the
analysis of multiple -ion mixture developed into, 'Ion Chromatography" Ion
chromatograph provides a single instrumental technique for rapid sequential
analysis of different anions present in a sample. Ion chromatograph eliminates
the need to use hazardous reagents and it effectively distinguishes among the
halides (Cr,F~) and the oxides (NO3", SO4"). For ambient air monitoring, IC is
the most suitable technique though it is expensive. The anions SO4", NO3", CI",
F" are being monitored in ambient air using IC 2000i/SP and DX-100 models
under NAQM project, NEERI.
Several approaches have been initiated to solve the problem of separation
and detection of ions either independently or in integrated fashion. Three major
approaches are:
i) suppressed ion chromatography
ii) non-suppressed ion chromatography and
iii) indirect detection ion chromatography
Suppressed Ion chromatography
In the group of separation techniques employing ionic interactions ion
exchange is the oldest one. The separation is carried out with the packing that
possess charge bearing functional groups. The most common retention
mechanism is simple ion exchange of sample ions X and mobile phase ion Y
with the charged groups R of the stationary phase.
X" + R+
Y_
Y" + R+
X" anion exchange
X+
+ R"Y+
Y+
+ R"X+
cation exchange
For anion exchange separation the sample ions X" are in competition with
the mobile phase ions Y" for the ionic sites R+
of the ion exchanger. Sample ions
that interact weakly with the ion exchanger, in the presence of competing mobile
phase ion will be retained on the column for a shorter duration whereas sample
ion interacting strongly to with the ion exchanger will be retained for larger.
Conductivity being universal property of ionic species in solution, showing
simple dependence on the concentration of ions, has been considered to solved
detection problem since long. However, in ion exchange separation conductivity
response is masked by the mobile phase electrolyte. This problem was tackled
by employing a novel combination of ion exchange columns to remove the
background electrolyte leaving only the ionic solute of interest as conducting
species in the column effluent.
The second column called the stripper or the suppresser complicates the
use of ion chromatography. The stripper column needs periodic regeneration
and furthermore ion exclusion effects and some band broadening in the
suppresser column deteriorates the analysis.
To optimise the regeneration and chromatographic efficiency of
conventional suppressor column, the relative volumes and specific ion exchange
capacities of separator and stripper columns are adjusted. For good
chromatographic efficiency volume ratio of stripper and separator column is kept
iii
low unity being desirable but values upto 10 acceptable. Regeneration
requirements are minimised by using low capacity pellicular particles or surface
modified resins in separating columns, low ionic strength mobile phase, small
sample size and conventional high capacity porous ion exchange resin in the
stripper. Typical low capacity separator column can be used alongwith
conventional high capacity suppresser column for 8 to 10 hours without requiring
regeneration. Automatic - regeneration feature in commercial instruments
further simplify the analysis. However suppresor should not react with the
sample ions in a way that would remove them from the eluant or reduce their
conductivities or retain them permanently undergoing irreversible changes.
Non Suppressed Ion Chromatography
The ion chromatographic systems with suppresser columns are generally
dedicated and relatively costly. Analysis by non-suppressed systems depends
on existence of a significant measurable difference between sample ions and the
prevailing eiuent ions. For improving sensitivity low capacity exchanges and
proper displacing ions are required. Low capacity exchanger match the low ionic
strength eiuent employed, which enable detection of small amounts of samples
and proper displacing ions can display a useful difference in equivalent
conductance in comparison with common inorganic ions. Low capacity micro
porous anion exchange resin also have been developed for the separation of
inorganic anions. The resins used macro reticular cross-linked polystyrene
beads as substrate. The conductivity detector used for this method requires a
large electronic offset range for nulling the background conductivity of the eiuent
and small cell currents to minimise heat dissipation in the cell and resulting
baseline drift and noise. This detector given best results when used with low
background conductivity of the eiuent and fine temperature stability.
Indirect Detection Ion Chromatography
Many inorganic ions display strong absorbance in UV region but at
wavelengths that were previously inaccessible to liquid chromatography
photometers. With the development of UV detectors that reach down to 190 nm.
These ions are easily amenable to sensitive monitoring and determination. Direct
iv
UV absorbance has been coupled with several separation modes for variety of
samples. However, the major drawback of these methods is the low sensitivity
even after employing high purity solutes to minimise background absorbance.
Recently the method of indirect detection is gaining applications.
Generally in ion exchange or ion pair separation ultraviolet detector is used in
conjunction with mobile phases with very low absorbance at the monitoring
wavelength.
The sample contains the chromospheres and when they elute and pass
through the detector, absorption of light takes place and positive peak is
recorded.
Generally low capacity ion exchange columns are employed as separators
and an aqueous solutions of potassium hydrogen phthalate (10~4
to 10"3
M) as
mobile phase. This salt gives suitable absorption at the wavelength 265 mm.
The choice of the mobile phase and its strength is decided by the anions of
interest and analysis time.
A large number of organic and and inorganic ions have been studied by
this technique. Analogous schemes for cation analysis have also been reported.
Simultaneous analysis of anions and cations in a single chromatogram also can
be done by this system.
Refractive index detector also has been used in this indirect mode. Both
UV & Rl methods are more sensitive than conductivity detection and they give
less baseline noise.
Interference : Any substance that has retention time coinciding with that
of any anion to be determined will interference. For example relatively high
concentrations of low molecular weight organic acids interfere with the
determination of chloride and fluoride. Sample dilution over comes many
interferences. Spurious peaks may result from contaminants in reagent water,
glassware or sample processing apparatus.
Minimum Detectable Limit
The minimum detectable concentration of an anion is a function of sample
size and conductivity scale used. Generally, minimum detectable concentrations
are near 0.1 mg/l for Br, CI", NO3-, N02", P04
3
" and S04
2
" with a 25 pi loop
and a 10 pS/cm full scale setting on the conductivity detector.
Air Monitoring/Sampling Collection
For ambient air anion sampling and analytical technique being used for
these pollutants require a large volume of air to be sampled in order to reach
needed detections limits. This has been accomplished utilising the high volume
(1400 Ipm) air sampler for the collection of total suspended particulate matter.
While for respirable suspended particulate matter, respirable dust sampler
is used. This sampler can be adapted with an optional PM-10 aerodynamic
aerosol inlet cut-point design which is insensitive to small variations in sampling
flow rate.
Preparation of Sample Filter
The filter papers are dried in desiccators for hours before, use. To ensure
acceptable filters, they are extracted with water and extracts are analysed by ion
chromatography. A filter blank of less than 0.1 pg per filter is considered
acceptable for field use.
Calibration of Sampling System
Each sampler is to be calibrated i) when new ii) after major repairs or
maintenance iii) whenever any audit point deviates from the calibration curve by
more than 7% iv) when a different sample collection media, other than that which
the sampler was originally calibrate to will be used for sampling, v) before and
after each test series.
vi
Construct a best fit curve for the points generated and use this relationship
for future work employing the flow sensor device.
Sample Retrieval
At the end of the desired sampling period the power is turned off.
Carefully remove the sampling head containing the filter. Remove filter from the
upper chamber using clean, teflon tipped forceps. Fold the filter in half twice
(Sampled side inward) and place it in an labeled envelope. These filters should
stored in dessicator containing dry silica gel. If the time span between sample
collection and laboratory analysis is to exceed 24 hour samples must be kept
refrigerated at 4°C. At least on field filter should be returned to the laboratory
with each group of samples. A field blank is treated exactly as a sample except
that no air is drawn throughout the filter.
Analysis
Reagents : Deionised or distilled water free from interferences at the
minimum detection limit of each constituent.
Eluant Solution : Eluant solution is a mixture of 1.7 mm sodium
bicarbonate and 1.8 mm Sodium Carbonate.
Regenerant Solution
Regenerant solution is all glass double distilled water for continous
regeneration.
Standard Anion Solution
Prepare a series of standard anion solutions by weighing the inducted
amount of salt, dried to a constant weight at 105°C to 1000 ml. Store in plastic
bottles in a refrigerator. These solutions are stable for at least one month. Verify
stability.
vii
Combine Working Standard Solution
Prepare a combined working standard solution by mixing appropriate
quantity of stock solution and store in plastic bottle protected from light. Prepare
fresh daily.
Ultrasonic Extraction Method
The extraction of anions is carried out by ultrasonic compact cleaner
model SW 45 (Toshniwal). The tank volume of extractor is 4.5 lits. It has a
sonifer cell disrupter .40 KHz power ultrasonic generator capable of dialing
150/130 w accurately with 127 cm horn distrupter sonabox.
From the exposed filter paper 18 circles of 1.5 cm diameter each were
taken by punching with steel punch (area 31.8 cm2
) in a clean 100 ml beaker. To
these circles 20 ml deionised water was added and extracted for 5 min in
ultrasonicator. The extract was vacuum filtered with the help of G4 sintered glass
crucible in a clean plastic bottle. Again 20 ml water was added and extraction
was done for 10 min. The filtrate was collected in same bottle. The extraction
procedure was repeated third time for 15 minutes. The extracted samples were
stored in fridge till analysis.
Calibration Curve
Prepare standards of different concentration by mixing known volume of
different ions. Inject standards containing single anion or a mixture and
determine approximate retention time. Inject atleast three different concentrations
for each anion to be measured and construct a calibration curve by plotting peak
height or area against concentration on linear graph paper. Recalibrate
whenever detector setting is changed. Record the peak height or area and
retention time for calculation of the calibration factor CF.
Calibration Factor CF =
Total area of peak
Mass injected (in microgram)
if the percent relative standard deviation (% RSD) of the calibration factor is less
than 20% over the working range, linearity through the origin can be assumed,
and average calibration factor can be used in place of a calibration curve.
The working calibration curve or calibration factor must be verified on each
working day by the injection of one or more calibration standards. If the response
factor for any analyte varies from the predicted response by more than + 20% a
new calibration curve must be prepared for that analyte. Calculate the percent
variance by the following equation-
Percent variance = [(R2 -Ri)/R<|] X 100
Before analysis can be performed the retention time windows must be
established for each analyte. Make three injections of the standard containing all
compounds for retention time window determination. The retention time window
is defined as the plus or minus three times the standard deviation of the absolute
retention times for each standard.
System Equillibration
Turn on ion chromatograph and adjust eluant flow rate. Adjust detector to
desired setting and let system come to equilibrium (20 to 30 min). A stable
baseline indicates equilibrium conditions. Adjust detector offset to zero out eluant
conductivity; with fibre or membrane suppressor adjust the regeneration flow rate
to maintain stability usually 2 to 3 mL/min.
Ion Chromatography with Conductivity Detector
Under the calibration procedures (external)the % RSD of the calibration
factor should be <20% over the linear working range of a five point calibration
curve.
Under the calibration procedures (external) daily working calibration curve
for each analyte should not vary from the predicted response by more than +
20%.
ix
For each analyte the retention time window must be established, verified
on a daily basis and established for each analyte throughout the course of a 72
hour period.
For each analyte the mid level standard must fall within the retention time
window on a daily basis as a qualitative performance evaluation of the IC system.
The surrogate standard recovery must not deviate by more than 20%.
Calibration
The stock standard solutions were prepared by adding exactly weighed
known amount of compound to a one litre volumetric flask and voiume was made
up by deionised water. A series of different concentrations of standards were
prepared by diluting stock standard solution.
Retention Time
The retention time is the time required for complete elution of a component
from the point of injection of sample. The retention time windows for each analyst
was determined. The retention time window is defined as plus or minus three
times the standard deviation of the absolute retention times for each standard.
Range and Sensitivity
The minimum detectable limit for chloride ion at sensitivity 100 is 2.44 pg.
The minimum detectable concentration of chloride ion for particulates collected
on one glass fiber filter of approximately 400 cm2
is 0.064 pg if 480 m3
of air are
sampled in the ambient atmosphere. The minimum detectable limit for SO4 is
3 pg for fluoride 0.4 pg and for nitrate 0.5 pg at sensitivity 100 on 2000 i/SP
Dionex ion chromatograph.
Precision and Accuracy
The blank fiber filter paper was spiked by known concentration of anions
and extracted ultrasonically. The relative standard deviation (RSD) for 3
ultrasonic extract for SO4 was ±1.35. For nitrate 5.54% for chloride 4.23% and for
fluoride 4.5%. The percentage recovery for SO4 is 98.02 for chloride 97.4, for
nitrate 98.0 and for fluoride 95.0.
Similarly the exposed filter papers were also spiked by known
concentration of anions. The RSD for 2 ultrasonic extract for sulphate was 1.44
for nitrate 5.8 for chloride 5.2 & for fluoride 5.5. The percentage recovery for SO4
is 97.01 for chloride 95.2 for nitrate 95 and for fluoride 90.58%.
Calculations
The concentration of different ions is calculated in ug/m3
concentration = peak height concentration Volumeof Total area of 1
(ug/m3
) of sample (cm) X of standard (ug) X sample X filter paper(cm2
) X
peak height of Area taken for volume of
standard (cm) extraction(cm2
) air (m3
)
xi
F
SAMPLE
Anions Separation in Rain Water Sample
by Ion-chromatograph
SAMPLE
Cations Separation in Rain Water Sample
by lon-'chromasograph

More Related Content

PDF
Biological Nutrient Removal Applications for Monitoring ORP | YSI
PDF
Chromatography: Anions and Metals Analysis for Waters Impacted by Hydraulic F...
PDF
DEVELOPMENT OF A NITRATE CONCENTRATON PREDICTION MODEL USING UV
PDF
Ad044203208
PDF
Ion chromatography copy
PDF
Improve Analysis Precision for ICP-OES and ICP-MS for Environmental and Geolo...
PPT
Ce icp-ms
PPTX
Synthesis and Characterization of Silver Nanoparticle and Liquid Crystal Nano...
Biological Nutrient Removal Applications for Monitoring ORP | YSI
Chromatography: Anions and Metals Analysis for Waters Impacted by Hydraulic F...
DEVELOPMENT OF A NITRATE CONCENTRATON PREDICTION MODEL USING UV
Ad044203208
Ion chromatography copy
Improve Analysis Precision for ICP-OES and ICP-MS for Environmental and Geolo...
Ce icp-ms
Synthesis and Characterization of Silver Nanoparticle and Liquid Crystal Nano...

What's hot (20)

PDF
PDF
Elemental Analysis of Plant Material
PPTX
Elemental Analysis of Plant Material
PDF
sci comm research-vivek
PPTX
The Role of IrO2 in Mediating Hole Transfer at the TiO2 Interface
PPTX
Applications of uv with hplc
PDF
Bacterial cable
PDF
Synthesis & Characterization of Fluorescent Silver Nanoparticles stabilized b...
PPTX
Introduction to nanoscience and nanotechnology
PPTX
Gas chromatography
PDF
Kumar Vanga-Poster_IEEE Nano
PPTX
Valente - Plasma Etching of Niobium SRF Cavities
DOCX
ESTIMATION OF SODIUM IN TAP WATER SAMPLE BY ATOMIC ABSORPTION SPECTROPHOTOMETER
PPTX
Synthesis of Spinel based Catalysts by Wet chemical methods for Colour Remova...
PPTX
Roll of nanomaterials in water treatment as photocatalysts copy
PPTX
Iec seminar
PDF
Q045068996
PDF
Low-cost Fiber Optic Chemical Sensor Development for Fishpond Application
PDF
Chemo sensor ppt
Elemental Analysis of Plant Material
Elemental Analysis of Plant Material
sci comm research-vivek
The Role of IrO2 in Mediating Hole Transfer at the TiO2 Interface
Applications of uv with hplc
Bacterial cable
Synthesis & Characterization of Fluorescent Silver Nanoparticles stabilized b...
Introduction to nanoscience and nanotechnology
Gas chromatography
Kumar Vanga-Poster_IEEE Nano
Valente - Plasma Etching of Niobium SRF Cavities
ESTIMATION OF SODIUM IN TAP WATER SAMPLE BY ATOMIC ABSORPTION SPECTROPHOTOMETER
Synthesis of Spinel based Catalysts by Wet chemical methods for Colour Remova...
Roll of nanomaterials in water treatment as photocatalysts copy
Iec seminar
Q045068996
Low-cost Fiber Optic Chemical Sensor Development for Fishpond Application
Chemo sensor ppt
Ad

Viewers also liked (7)

PDF
Preventive measures for occupational
PDF
Some critical facets in air pollution
PPT
Dieu Hien - Bai 17
PPT
Autoservicio
PDF
2015 tgcsa best bagger manual
PDF
BIM Portfolio
PPTX
Franz Zhang et al Weevil Workshop 2016 Neotropical Entiminae Systematics evol...
Preventive measures for occupational
Some critical facets in air pollution
Dieu Hien - Bai 17
Autoservicio
2015 tgcsa best bagger manual
BIM Portfolio
Franz Zhang et al Weevil Workshop 2016 Neotropical Entiminae Systematics evol...
Ad

Similar to Monitoring of inorganic ions (20)

DOCX
ion exclusion chromatogrpahy by sumith.docx
PPTX
Ion-Chromatography-IC.pptx ion chromatography ppt
PPTX
ion exclusion chromatography.pptx
PPTX
High-effective IONIC CHROMATOGRAPHY
DOCX
Ion Chromatography Paper
PPTX
Ion exchange chromatoghraphy
PDF
Nigel Poster final final
PPTX
ION EXCLUSION CHROMATOGRAPHY
PPTX
Ion selective electrode (hw)85
PPTX
Atomic absorption meth spectroscopy.pptx
PPT
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
PPTX
Nanocatalysts and nanozymes in heavy metal rmoval
PDF
Trace and Heavy Metals in Wastewater
PPTX
Mass Spectrometry Ionization Techniques
PPTX
Atomic absorption spectroscopy
PPT
Affinity chromatography 1
PPT
Uvspectroscopyup 140119003106-phpapp01
PPTX
IR spectroscopy
PPTX
Atomic absorption spectrophotometry
ion exclusion chromatogrpahy by sumith.docx
Ion-Chromatography-IC.pptx ion chromatography ppt
ion exclusion chromatography.pptx
High-effective IONIC CHROMATOGRAPHY
Ion Chromatography Paper
Ion exchange chromatoghraphy
Nigel Poster final final
ION EXCLUSION CHROMATOGRAPHY
Ion selective electrode (hw)85
Atomic absorption meth spectroscopy.pptx
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
Nanocatalysts and nanozymes in heavy metal rmoval
Trace and Heavy Metals in Wastewater
Mass Spectrometry Ionization Techniques
Atomic absorption spectroscopy
Affinity chromatography 1
Uvspectroscopyup 140119003106-phpapp01
IR spectroscopy
Atomic absorption spectrophotometry

More from ECRD2015 (20)

PDF
Pollution control in small and medium scale industry
PDF
Indoor air quality
PDF
Air pollution control
PDF
Vehicular Pollution Management
PPT
Strategy of control of urban air pollution
PDF
Spectrophotometer meter
PPT
Slm_1
PDF
Shop floor monitoring
PDF
Quality assurance and quality control
PPT
Pm10 measmnt
PPT
Ozone
PDF
Nitrogen oxide monitoring
PDF
Minutes of brain storming session on air pollution load carrying capacity of ...
PDF
Measurement of respirable suspended pm10
PDF
Measurement of hydrogen ion concentration (p h)
PPT
Maintenance
PPT
Maintenance of high volume
PDF
Industrial emission control techniques
DOC
Indophenol method
PDF
Industrial ventilation
Pollution control in small and medium scale industry
Indoor air quality
Air pollution control
Vehicular Pollution Management
Strategy of control of urban air pollution
Spectrophotometer meter
Slm_1
Shop floor monitoring
Quality assurance and quality control
Pm10 measmnt
Ozone
Nitrogen oxide monitoring
Minutes of brain storming session on air pollution load carrying capacity of ...
Measurement of respirable suspended pm10
Measurement of hydrogen ion concentration (p h)
Maintenance
Maintenance of high volume
Industrial emission control techniques
Indophenol method
Industrial ventilation

Recently uploaded (20)

PDF
Ornithology-Basic-Concepts.pdf..........
PDF
1748933543SJA_41_2_826-834 SJA Ihsan ullha.pdf
PDF
Tree Biomechanics, a concise presentation
PPT
PPTPresentation3 jhsvdasvdjhavsdhsvjcksjbc.jasb..ppt
PDF
Bai bao Minh chứng sk2-DBTrong-003757.pdf
PDF
Effects of rice-husk biochar and aluminum sulfate application on rice grain q...
PPTX
Plant_Cell_Presentation.pptx.com learning purpose
PDF
Effective factors on adoption of intercropping and it’s role on development o...
PDF
Urban Hub 50: Spirits of Place - & the Souls' of Places
PPTX
Corporate Social Responsibility & Governance
PDF
Insitu conservation seminar , national park ,enthobotanical significance
PPTX
Conformity-and-Deviance module 7 ucsp grade 12
PPTX
Unit 1 - Environmental management, politics and.pptx
PDF
Effect of salinity on biochimical and anatomical characteristics of sweet pep...
PPTX
Concept of Safe and Wholesome Water.pptx
PPTX
Biodiversity of nature in environmental studies.pptx
PPTX
Green Modern Sustainable Living Nature Presentation_20250226_230231_0000.pptx
PPTX
STL Academy - Highlights & Impact 2020-21-v2 (1).pptx
PPTX
"One Earth Celebrating World Environment Day"
DOCX
Epoxy Coated Steel Bolted Tanks for Beverage Wastewater Storage Manages Liqui...
Ornithology-Basic-Concepts.pdf..........
1748933543SJA_41_2_826-834 SJA Ihsan ullha.pdf
Tree Biomechanics, a concise presentation
PPTPresentation3 jhsvdasvdjhavsdhsvjcksjbc.jasb..ppt
Bai bao Minh chứng sk2-DBTrong-003757.pdf
Effects of rice-husk biochar and aluminum sulfate application on rice grain q...
Plant_Cell_Presentation.pptx.com learning purpose
Effective factors on adoption of intercropping and it’s role on development o...
Urban Hub 50: Spirits of Place - & the Souls' of Places
Corporate Social Responsibility & Governance
Insitu conservation seminar , national park ,enthobotanical significance
Conformity-and-Deviance module 7 ucsp grade 12
Unit 1 - Environmental management, politics and.pptx
Effect of salinity on biochimical and anatomical characteristics of sweet pep...
Concept of Safe and Wholesome Water.pptx
Biodiversity of nature in environmental studies.pptx
Green Modern Sustainable Living Nature Presentation_20250226_230231_0000.pptx
STL Academy - Highlights & Impact 2020-21-v2 (1).pptx
"One Earth Celebrating World Environment Day"
Epoxy Coated Steel Bolted Tanks for Beverage Wastewater Storage Manages Liqui...

Monitoring of inorganic ions

  • 1. Monitoring of Inorganic Ions in Ambient Air Mr. S. D. Joshi Scientist Air Pollution Control Division, NEERI, Nagpur - 440020 Introduction The air besides containing gaseous pollutants like S02, NOx, CO, HC also contains particulate matter which is made up of complex organic compounds, inorganic elements and their oxides and secondary particulates viz, sulphate, nitrate etc. These secondary particulates are formed from their precursors in presence of sunlight humidity and metal ion species by oxidation or reduction processes. The acid precursors viz. sulfate and nitrate, and chloride require more concern as these have deleterious effects on aquatic and forest eco- systems and are potentially harmful for a variety of building materials, accelerate corrosion of metals, damage sculptures and cause chronic respiratory problems in human beings. In view of possible large social, economic, ecological and aesthetic value, it has become important to analyse these pollutants in the ambient air. These particulates are emitted in the atmosphere by different anthropogenic and natural sources. Sulphate (SO4-) is formed in the atmosphere by reduction or oxidation in presence of hydroxy! ion (OH"). The main sources of sulfur oxides are power houses, sulphuric acid, petroleum industries, oil refineries, coal burning and domestic use of fuels. Acid manufacturing, automobile exhaust, explosive industry, fuel burning etc. are the main sources of nitrogen oxides. Chloride occurs predominantly in coastal cities. Other sources of chloride are coal burning, waste incineration, automobile exhaust, burning of chlorinated hydrocarbons. Fluoride is emitted in the atmosphere by fertiliser industry, aluminium industry, steel making and acid making industries.
  • 2. Analytical Techniques Different methods are available for the analysis of anions. The conventional methods, viz., titrimetric, colorimetric, spectrophotometric, gravimetric and ion selective electrode methods though commonly used, are ion specific, time consuming and less sensitive. Ion Chromatography (IC) is the only instrumental technique which can be used for rapid, sequential analysis of different anions simultaneously. Ion Chromatography Chromatography encompasses wide range of techniques of separation of the components of a mixture as a result of differential migration of the component molecules between stationary phase and mobile phase. It was Hamish, Small of DOW chemical corporation whose pioneering work in mid seventies in the analysis of multiple -ion mixture developed into, 'Ion Chromatography" Ion chromatograph provides a single instrumental technique for rapid sequential analysis of different anions present in a sample. Ion chromatograph eliminates the need to use hazardous reagents and it effectively distinguishes among the halides (Cr,F~) and the oxides (NO3", SO4"). For ambient air monitoring, IC is the most suitable technique though it is expensive. The anions SO4", NO3", CI", F" are being monitored in ambient air using IC 2000i/SP and DX-100 models under NAQM project, NEERI. Several approaches have been initiated to solve the problem of separation and detection of ions either independently or in integrated fashion. Three major approaches are: i) suppressed ion chromatography ii) non-suppressed ion chromatography and iii) indirect detection ion chromatography
  • 3. Suppressed Ion chromatography In the group of separation techniques employing ionic interactions ion exchange is the oldest one. The separation is carried out with the packing that possess charge bearing functional groups. The most common retention mechanism is simple ion exchange of sample ions X and mobile phase ion Y with the charged groups R of the stationary phase. X" + R+ Y_ Y" + R+ X" anion exchange X+ + R"Y+ Y+ + R"X+ cation exchange For anion exchange separation the sample ions X" are in competition with the mobile phase ions Y" for the ionic sites R+ of the ion exchanger. Sample ions that interact weakly with the ion exchanger, in the presence of competing mobile phase ion will be retained on the column for a shorter duration whereas sample ion interacting strongly to with the ion exchanger will be retained for larger. Conductivity being universal property of ionic species in solution, showing simple dependence on the concentration of ions, has been considered to solved detection problem since long. However, in ion exchange separation conductivity response is masked by the mobile phase electrolyte. This problem was tackled by employing a novel combination of ion exchange columns to remove the background electrolyte leaving only the ionic solute of interest as conducting species in the column effluent. The second column called the stripper or the suppresser complicates the use of ion chromatography. The stripper column needs periodic regeneration and furthermore ion exclusion effects and some band broadening in the suppresser column deteriorates the analysis. To optimise the regeneration and chromatographic efficiency of conventional suppressor column, the relative volumes and specific ion exchange capacities of separator and stripper columns are adjusted. For good chromatographic efficiency volume ratio of stripper and separator column is kept iii
  • 4. low unity being desirable but values upto 10 acceptable. Regeneration requirements are minimised by using low capacity pellicular particles or surface modified resins in separating columns, low ionic strength mobile phase, small sample size and conventional high capacity porous ion exchange resin in the stripper. Typical low capacity separator column can be used alongwith conventional high capacity suppresser column for 8 to 10 hours without requiring regeneration. Automatic - regeneration feature in commercial instruments further simplify the analysis. However suppresor should not react with the sample ions in a way that would remove them from the eluant or reduce their conductivities or retain them permanently undergoing irreversible changes. Non Suppressed Ion Chromatography The ion chromatographic systems with suppresser columns are generally dedicated and relatively costly. Analysis by non-suppressed systems depends on existence of a significant measurable difference between sample ions and the prevailing eiuent ions. For improving sensitivity low capacity exchanges and proper displacing ions are required. Low capacity exchanger match the low ionic strength eiuent employed, which enable detection of small amounts of samples and proper displacing ions can display a useful difference in equivalent conductance in comparison with common inorganic ions. Low capacity micro porous anion exchange resin also have been developed for the separation of inorganic anions. The resins used macro reticular cross-linked polystyrene beads as substrate. The conductivity detector used for this method requires a large electronic offset range for nulling the background conductivity of the eiuent and small cell currents to minimise heat dissipation in the cell and resulting baseline drift and noise. This detector given best results when used with low background conductivity of the eiuent and fine temperature stability. Indirect Detection Ion Chromatography Many inorganic ions display strong absorbance in UV region but at wavelengths that were previously inaccessible to liquid chromatography photometers. With the development of UV detectors that reach down to 190 nm. These ions are easily amenable to sensitive monitoring and determination. Direct iv
  • 5. UV absorbance has been coupled with several separation modes for variety of samples. However, the major drawback of these methods is the low sensitivity even after employing high purity solutes to minimise background absorbance. Recently the method of indirect detection is gaining applications. Generally in ion exchange or ion pair separation ultraviolet detector is used in conjunction with mobile phases with very low absorbance at the monitoring wavelength. The sample contains the chromospheres and when they elute and pass through the detector, absorption of light takes place and positive peak is recorded. Generally low capacity ion exchange columns are employed as separators and an aqueous solutions of potassium hydrogen phthalate (10~4 to 10"3 M) as mobile phase. This salt gives suitable absorption at the wavelength 265 mm. The choice of the mobile phase and its strength is decided by the anions of interest and analysis time. A large number of organic and and inorganic ions have been studied by this technique. Analogous schemes for cation analysis have also been reported. Simultaneous analysis of anions and cations in a single chromatogram also can be done by this system. Refractive index detector also has been used in this indirect mode. Both UV & Rl methods are more sensitive than conductivity detection and they give less baseline noise. Interference : Any substance that has retention time coinciding with that of any anion to be determined will interference. For example relatively high concentrations of low molecular weight organic acids interfere with the determination of chloride and fluoride. Sample dilution over comes many
  • 6. interferences. Spurious peaks may result from contaminants in reagent water, glassware or sample processing apparatus. Minimum Detectable Limit The minimum detectable concentration of an anion is a function of sample size and conductivity scale used. Generally, minimum detectable concentrations are near 0.1 mg/l for Br, CI", NO3-, N02", P04 3 " and S04 2 " with a 25 pi loop and a 10 pS/cm full scale setting on the conductivity detector. Air Monitoring/Sampling Collection For ambient air anion sampling and analytical technique being used for these pollutants require a large volume of air to be sampled in order to reach needed detections limits. This has been accomplished utilising the high volume (1400 Ipm) air sampler for the collection of total suspended particulate matter. While for respirable suspended particulate matter, respirable dust sampler is used. This sampler can be adapted with an optional PM-10 aerodynamic aerosol inlet cut-point design which is insensitive to small variations in sampling flow rate. Preparation of Sample Filter The filter papers are dried in desiccators for hours before, use. To ensure acceptable filters, they are extracted with water and extracts are analysed by ion chromatography. A filter blank of less than 0.1 pg per filter is considered acceptable for field use. Calibration of Sampling System Each sampler is to be calibrated i) when new ii) after major repairs or maintenance iii) whenever any audit point deviates from the calibration curve by more than 7% iv) when a different sample collection media, other than that which the sampler was originally calibrate to will be used for sampling, v) before and after each test series. vi
  • 7. Construct a best fit curve for the points generated and use this relationship for future work employing the flow sensor device. Sample Retrieval At the end of the desired sampling period the power is turned off. Carefully remove the sampling head containing the filter. Remove filter from the upper chamber using clean, teflon tipped forceps. Fold the filter in half twice (Sampled side inward) and place it in an labeled envelope. These filters should stored in dessicator containing dry silica gel. If the time span between sample collection and laboratory analysis is to exceed 24 hour samples must be kept refrigerated at 4°C. At least on field filter should be returned to the laboratory with each group of samples. A field blank is treated exactly as a sample except that no air is drawn throughout the filter. Analysis Reagents : Deionised or distilled water free from interferences at the minimum detection limit of each constituent. Eluant Solution : Eluant solution is a mixture of 1.7 mm sodium bicarbonate and 1.8 mm Sodium Carbonate. Regenerant Solution Regenerant solution is all glass double distilled water for continous regeneration. Standard Anion Solution Prepare a series of standard anion solutions by weighing the inducted amount of salt, dried to a constant weight at 105°C to 1000 ml. Store in plastic bottles in a refrigerator. These solutions are stable for at least one month. Verify stability. vii
  • 8. Combine Working Standard Solution Prepare a combined working standard solution by mixing appropriate quantity of stock solution and store in plastic bottle protected from light. Prepare fresh daily. Ultrasonic Extraction Method The extraction of anions is carried out by ultrasonic compact cleaner model SW 45 (Toshniwal). The tank volume of extractor is 4.5 lits. It has a sonifer cell disrupter .40 KHz power ultrasonic generator capable of dialing 150/130 w accurately with 127 cm horn distrupter sonabox. From the exposed filter paper 18 circles of 1.5 cm diameter each were taken by punching with steel punch (area 31.8 cm2 ) in a clean 100 ml beaker. To these circles 20 ml deionised water was added and extracted for 5 min in ultrasonicator. The extract was vacuum filtered with the help of G4 sintered glass crucible in a clean plastic bottle. Again 20 ml water was added and extraction was done for 10 min. The filtrate was collected in same bottle. The extraction procedure was repeated third time for 15 minutes. The extracted samples were stored in fridge till analysis. Calibration Curve Prepare standards of different concentration by mixing known volume of different ions. Inject standards containing single anion or a mixture and determine approximate retention time. Inject atleast three different concentrations for each anion to be measured and construct a calibration curve by plotting peak height or area against concentration on linear graph paper. Recalibrate whenever detector setting is changed. Record the peak height or area and retention time for calculation of the calibration factor CF. Calibration Factor CF = Total area of peak Mass injected (in microgram)
  • 9. if the percent relative standard deviation (% RSD) of the calibration factor is less than 20% over the working range, linearity through the origin can be assumed, and average calibration factor can be used in place of a calibration curve. The working calibration curve or calibration factor must be verified on each working day by the injection of one or more calibration standards. If the response factor for any analyte varies from the predicted response by more than + 20% a new calibration curve must be prepared for that analyte. Calculate the percent variance by the following equation- Percent variance = [(R2 -Ri)/R<|] X 100 Before analysis can be performed the retention time windows must be established for each analyte. Make three injections of the standard containing all compounds for retention time window determination. The retention time window is defined as the plus or minus three times the standard deviation of the absolute retention times for each standard. System Equillibration Turn on ion chromatograph and adjust eluant flow rate. Adjust detector to desired setting and let system come to equilibrium (20 to 30 min). A stable baseline indicates equilibrium conditions. Adjust detector offset to zero out eluant conductivity; with fibre or membrane suppressor adjust the regeneration flow rate to maintain stability usually 2 to 3 mL/min. Ion Chromatography with Conductivity Detector Under the calibration procedures (external)the % RSD of the calibration factor should be <20% over the linear working range of a five point calibration curve. Under the calibration procedures (external) daily working calibration curve for each analyte should not vary from the predicted response by more than + 20%. ix
  • 10. For each analyte the retention time window must be established, verified on a daily basis and established for each analyte throughout the course of a 72 hour period. For each analyte the mid level standard must fall within the retention time window on a daily basis as a qualitative performance evaluation of the IC system. The surrogate standard recovery must not deviate by more than 20%. Calibration The stock standard solutions were prepared by adding exactly weighed known amount of compound to a one litre volumetric flask and voiume was made up by deionised water. A series of different concentrations of standards were prepared by diluting stock standard solution. Retention Time The retention time is the time required for complete elution of a component from the point of injection of sample. The retention time windows for each analyst was determined. The retention time window is defined as plus or minus three times the standard deviation of the absolute retention times for each standard. Range and Sensitivity The minimum detectable limit for chloride ion at sensitivity 100 is 2.44 pg. The minimum detectable concentration of chloride ion for particulates collected on one glass fiber filter of approximately 400 cm2 is 0.064 pg if 480 m3 of air are sampled in the ambient atmosphere. The minimum detectable limit for SO4 is 3 pg for fluoride 0.4 pg and for nitrate 0.5 pg at sensitivity 100 on 2000 i/SP Dionex ion chromatograph. Precision and Accuracy The blank fiber filter paper was spiked by known concentration of anions and extracted ultrasonically. The relative standard deviation (RSD) for 3 ultrasonic extract for SO4 was ±1.35. For nitrate 5.54% for chloride 4.23% and for
  • 11. fluoride 4.5%. The percentage recovery for SO4 is 98.02 for chloride 97.4, for nitrate 98.0 and for fluoride 95.0. Similarly the exposed filter papers were also spiked by known concentration of anions. The RSD for 2 ultrasonic extract for sulphate was 1.44 for nitrate 5.8 for chloride 5.2 & for fluoride 5.5. The percentage recovery for SO4 is 97.01 for chloride 95.2 for nitrate 95 and for fluoride 90.58%. Calculations The concentration of different ions is calculated in ug/m3 concentration = peak height concentration Volumeof Total area of 1 (ug/m3 ) of sample (cm) X of standard (ug) X sample X filter paper(cm2 ) X peak height of Area taken for volume of standard (cm) extraction(cm2 ) air (m3 ) xi
  • 12. F SAMPLE Anions Separation in Rain Water Sample by Ion-chromatograph
  • 13. SAMPLE Cations Separation in Rain Water Sample by lon-'chromasograph