Orion Star™ and
Star Plus Meter
User Guide




             • English

             • Español

             • Français

             • Deutsch

             • Italiano
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc.

AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure
Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are
registered trademarks of Thermo Fisher.

1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER,
Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoE Z, Cable-Free, CERTI-CAL, CISA, DataCOLLECT,
DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value,
FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum
Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results,
Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK,
rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21,
Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are
trademarks of Thermo Fisher.

Guaranteed Success and The Technical Edge are service marks of Thermo Fisher.

PerpHecT meters are protected by U.S. patent 6,168,707.

PerpHecT ROSS are protected by U.S. patent 6,168,707.

ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753.

ionplus electrodes and Optimum Results solutions are protected by US Patent 5,830,338.

ROSS Ultra electrodes are protected by US patents 6,793,787.

Orion ORP Standard is protected by US Patent 6,350,367.

Orion NoCal electrodes are protected by US Patent 7,276,142.


© 2008 Thermo Fisher Scientific Inc. All rights reserved. RDO is a registered trademark of In-Situ Inc.,
Fort Collins, CO USA. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.

The specifications, descriptions, drawings, ordering information and part numbers within this document are
subject to change without notice.

This publication supersedes all previous publications on this subject.
EN                                                                                                                                                   ENGLISH SECTION




     Table of Contents
     Chapter I Introduction
     Meter Features  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-2
     Chapter II Display
     General Description .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-3
     Chapter III Keypad
     General Description .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-5
     Key Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .EN-6
     Chapter IV Preparation
     Installing the Power Adapter .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-7
     Installing the Batteries .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-8
     Connecting the Electrodes  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-9
     Turning on the Instrument .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-10
     Meter Maintenance .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-10
     Chapter V Meter Setup
     Setup Menu . .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .   EN-11
     	 Setup Menu Table  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .              EN-12
     General Menu Settings .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                  EN-14
     Time and Date Settings .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                 EN-15
     AUTO-READ™, Continuous or Timed Measurement Settings  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                                                                         EN-16
     Selecting the Measurement Parameter  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                                          EN-17
     Method Setup .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .      EN-18
     Chapter VI pH Technique
     pH Setup Menu .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .        EN-19
     pH Calibration  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .   EN-20
     pH Measurement  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .            EN-21
     pH Temperature Display and Calibration  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                                       EN-22
     Chapter VII mV, Relative mV and ORP Technique
     Relative mV and ORP Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .EN-23
     mV, Relative mV and ORP Measurement .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-24
     Chapter VIII Dissolved Oxygen Technique
     Dissolved Oxygen Setup Menu .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                              EN-25
     Dissolved Oxygen Calibration  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                         EN-26
     Dissolved Oxygen Measurement  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                                  EN-28
     Dissolved Oxygen Temperature Display and Calibration  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                                                             EN-29


     Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                                                                                                         EN-I
EN



     Chapter IX RDO® Optical Dissolved Oxygen Technique
     RDO Cap Overview .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-31
     RDO Optical Dissolved Oxygen Setup Menu  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-32
     	 Setup Menu for RDO Optical Dissolved Oxygen Sensor . . . . . . . . . . . . . . . . .EN-33
     RDO Optical Dissolved Oxygen Calibration .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-35
     RDO Optical Dissolved Oxygen Measurement .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-37
     RDO Optical Dissolved Oxygen Temperature Display and Calibration .  .  .  .  .  .  .  .  . EN-38
     Chapter X Conductivity Technique
     Conductivity Setup Menu  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                     EN-39
     Conductivity Calibration . .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                EN-40
     Conductivity Measurement .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                        EN-41
     Conductivity Temperature Display and Calibration .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                                                   EN-42
     Chapter XI ISE Technique
     ISE Setup Menu  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .         EN-43
     ISE Calibration .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .   EN-44
     ISE Measurement .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .            EN-45
     ISE Temperature Display and Calibration .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                                       EN-46
     Chapter XII Data Archiving and Retrieval
     Datalog and Calibration Log . .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                      EN-47
     	 Automatic Datalog Feature  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                          EN-47
     	 Datalog Deletion Setting .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                      EN-48
     	 Viewing and Printing the Datalog and Calibration Log . .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                                                             EN-49
     Chapter XIII Declaration of Conformity
     Declaration of Conformity .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-51
     WEEE Compliance . .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-52
     Chapter XIV Troubleshooting
     Meter Self Test  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-53
     Meter Error Codes . .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-54
     General Troubleshooting  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-56
     Chapter XV Meter Specifications
     Meter Specifications .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-59
     Ordering Information .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  . EN-63
     Appendix A Meter Setup Menu Features
     pH Setup Menu Features  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                      EN-65
     Dissolved Oxygen Setup Menu Features  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                                            EN-65
     Conductivity Setup Menu Features .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                                  EN-66
     ISE Setup Menu Features .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .  .                      EN-68

     EN-II                                              Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Chapter I         Introduction
Congratulations! You have selected an industry-leading Thermo Scientific Orion Star
or Star Plus series meter that is designed for electrochemistry measurements in the
field or in the laboratory.

•	 2-Star meters provide the single parameter measurement of pH.

•	 3-Star Plus meters provide the single parameter measurement of pH,
   dissolved oxygen, RDO® optical dissolved oxygen or conductivity.

•	 4-Star Plus meters provide the dual parameter measurements of
   pH/dissolved oxygen, pH/conductivity, pH/ISE (ion selective electrode) or
   RDO optical dissolved oxygen/conductivity.

•	 5-Star Plus meters provide the multi-parameter measurements of
   pH/ISE/dissolved oxygen/conductivity, pH/dissolved oxygen/conductivity or
   pH/RDO optical dissolved oxygen/conductivity.

All meters include a temperature measurement function. All meters with pH
measurement capability include a mV/relative mV/ORP function.

Built to meet the demands of busy, multiple user laboratory or plant environments,
all Orion Star and Star Plus series meters are microprocessor controlled, which aids
in the delivery of accurate and precise measurements. The waterproof portable
meters can even withstand submersion for short periods of time.

The 3-Star Plus, 4-Star Plus and 5-Star Plus meters have been enhanced to include
an increased number of datalog points, improved temperature displays and a new
temperature calibration mode for each measurement parameter. Benchtop 3-Star
Plus, 4-Star Plus and 5-Star Plus meters include autosampler capability and are
compatible with the AutoTration™-500 autosampler. Refer to the AutoTration-500
user guide for information on operating the Star Plus meters with the autosampler.

Note: Please read this user guide thoroughly before using your benchtop or
portable meter. Any use outside of these instructions may invalidate your warranty
and cause permanent damage to the meter.

Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide             EN-1
Introduction


Meter Features
To better meet the needs of users in environmental protection and control, food and
beverage, pharmaceutical and consumer product laboratories, the Orion Star and
Star Plus series meters include these key features:

•	 Password Protected Methods – The meter will save up to ten custom
   measurements and calibrations for future reference. Password protection of
   each method eliminates any tampering with methods as multiple users access
   only the procedure most appropriate to their work.

•	 AUTO-READ™ – The meter takes a measurement and automatically prints or
   logs data when the reading becomes stable. The measurement is frozen on the
   display until the user prompts the meter to take a new measurement.

•	 Stirrer Control – Benchtop meters (3-Star and higher) have a stirrer control
   for the stirrer probe, Cat. No. 096019, and the AUTO-STIR™ BOD probe, which
   eliminates the need for additional stir plates and stir bars.

•	 SMART STABILITY™ and SMART AVERAGING™ – Remove the
   guesswork by automatically compensating for measurement conditions and
   optimizing the meter response time.

•	 Display Backlight – All 3-Star, 4-Star and 5-Star meters include a display
   backlight feature. When the meter is on, a quick press of       will turn the
   backlight on and off. When the meter is operating on battery power, the
   backlight will automatically turn off after two minutes to conserve power. When
   batteries are low, the backlight will no longer turn on.

•	 Automatic Shutoff – The meters will shut down after 20 minutes without
   a keypress. This maximizes battery power on portable meters and benchtop
   meters that are being run on battery power.

•	 Audible Signals – The meter will beep whenever a key is pressed, providing
   immediate verification that the user’s input was received.

An easy-to-use reference guide, attached to each meter, supports daily meter use.



EN-2                   Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Chapter II         Display
General Description
Throughout a given process, the display on an Orion Star or Star Plus series meter
provides temperature and calibration data. The temperature appears in the left, top
corner of the display. The     icon indicates that a calibration mode or calibration
setup menu is active. The man, 2, 4, 7, 9, 10, and 12 icons indicate which pH
buffers were saved after a pH calibration is performed. The setup icon only appears
when the meter is in setup mode. The          icon indicates an error condition and
when it is displayed with the    icon, a calibration alarm or sensor quality issue
exists. The      icon indicates that the AUTO-READ measurement mode is active
and is discussed in greater detail in the Meter Setup section.




Temperature
  Calibration
                                                                      pH/ISE/ORP
                                                                      Unit

                                                                      Conductivity
                                                                      Unit

                                                                      DO Unit


5 Star Meter
This is the display of the 5-Star meter capable of multi-parameter measurements.
The single and dual parameter meters will have fewer measurement lines,
depending on the meter capabilities.

Note: In the measurement mode, the three main lines of data on the meter
display correspond to what is being measured.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide             EN-3
Display


Measurement Unit Icons
In the measurement mode, the arrow icon on the left side of the display screen
indicates the active line. Press      to move the arrow icon to the desired line and
press       /     to scroll through the measurement unit icons associated with the
selected line. The measurement unit icons for the 5-Star multi-parameter meter are
shown below. The single and dual parameter meters will have fewer measurement
lines and icons, depending on the meter capabilities.




                               A             	   p
                                                  H, mV, relative mV, ISE,
                                                 or temperature


                               B             	   C
                                                  onductivity, TDS, salinity, resistivity,
                                                 or temperature


                               C         	       D
                                                  issolved oxygen as % saturation,
                                                 dissolved oxygen as concentration,
                                                 barometric pressure or temperature




The units of measurement, which are displayed on the right side of the screen, will
flash until the reading is stable.

Note: If a measurement line is not needed, press       to move the arrow
icon to the measurement line that is not needed and press     /     until the
measurement line is completely blank.




EN-4                   Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Chapter III            Keypad
General Description
The keypad layout is the same for all Orion Star and Star Plus series meters. The
portable and 2-Star benchtop meters have nine keys. The 3-Star, 4-Star and 5-Star
benchtop meters have 10 keys due to the addition of the stir key –      .



Portable Keypad

            Calibrate                                                 Measure
                                                                      Save/Print
                  Power                                               Backlight
                                                                      Scroll Up
                  Digits                                              Line Select
         Scroll Down

                  Setup                                               View Log



Benchtop Keypad
                                Power       Backlight
                                                                         Measure
      Calibrate                                                          Save/Print
                                                                         Stirrer On/Off,
     Scroll Up
                                                                         (not on 2-Star)
         Digits                                                          Line Select

         Setup                                                           View Log

                                                                         Scroll Down




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                        EN-5
Keypad


Key Definitions

 Key     Description                          Key     Description
         Turns the meter on, if the meter             Changes the measurement
         is off.                                      units of the selected line in the
         Toggles the backlight on and off,            measurement mode.
         if the meter is on (3-Star, 4-Star           Changes the value on the
         and 5-Star meters only).                     selected line in the setup,
         If the meter is on, hold down                methods and log view modes.
         the key for about three seconds              Edits the value of the flashing
         to turn off the meter.                       digit for setup, password entry
                                                      and calibration modes.
         Scrolls the arrow icon on left of            Selects the next digit to edit
         screen among the three display               and moves the decimal point
         lines, so the selected line can              for setup, password entry and
         be edited or calibrated.                     calibration modes.
         Starts the calibration for                   Prints and logs a measurement
         the selected line in the                     in the continuous or timed
         measurement mode.                            measurement modes.
         If the arrow icon points to the              Prints, logs and freezes the
         top line and the displayed units             display when the reading
         are pH, pressing the key will                becomes stable in the AUTO-
         start a pH calibration.                      READ measurement mode.
         Each time the key is pressed in              Exits the setup menu and returns
         the calibration mode, the meter              to measurement mode.
         will accept the calibration point            Accepts the calibration and
         and move to the next point                   returns to measurement mode.
         until the maximum number of
         calibration points are reached.
         Enters the setup menu, starting              Enters the log view and
         with selected line in the                    download menu.
         measurement mode.
         If the arrow icon points to the              Turns the stirrer on and off.
         top line and the displayed units
         are ISE, pressing the key will
         enter the ISE setup screen.


EN-6                     Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Chapter IV           Preparation
Installing the Power Adapter
The universal power adapter that is included with your benchtop meter is the only
power adapter recommended for use with this unit. The use of any other power
adapter will void your meter warranty. The external electrical power adapter is rated
to be operated at 100 to 240 VAC, 0.5 A, 50/60 Hz.

Based on your wall outlet, select one of the four plug plates provided (110 V, 220 V,
240V) and slide it into the grooves on the adapter. A click will be heard when the
plug is properly in place.




Connect the output plug of the power adapter to the power input on the benchtop
meter. Refer to the diagram in the Connecting the Electrodes section.

Batteries can be installed in the benchtop Orion Star or Star Plus series meters, so
the meter setup settings are protected if the meter is disconnected from the wall
outlet or a brief power outage occurs.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide               EN-7
Preparation


Installing the Batteries
Orion Star and Star Plus series meters use four AA alkaline batteries. Do not use
lithium or rechargeable batteries. Improper installation of non-alkaline batteries
could create a hazard.

Note: For benchtop meters, the installation of batteries is not required if the unit
will always be connected to a power source via the universal power supply. For
portable meters, the batteries are supplied from the factory. To access the battery
compartment in portable meters, loosen the two screws in the back of the meter.

1.	 Confirm that the meter is off and gently place the meter upside down on a
    clean, lint-free cloth to prevent scratching the LCD.

2.	 Remove the battery case cover.

3.	 Insert new batteries with the + side orientation as depicted in the battery
    compartment housing.

4.	 Replace the battery case cover.

5.	 Stored data, calibrations and methods will remain in the meter’s nonvolatile
    memory when the batteries are being replaced. However, the date and time
    may need to be reset when the batteries are changed.




EN-8                    Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Preparation


Connecting the Electrodes
Follow the diagrams below to correctly connect electrodes and probes to the meter.
The multi-parameter meter is depicted; single parameter and dual parameter meters
will have fewer connections, depending on the meter measurement capabilities.

Benchtop Meter Electrode Connections




                    Reference,       8 pin miniDIN,   RS232     Power
                     pH, ORP          Cond or ATC
  Stirrer jack,       or ISE
 stirrer or BOD                   BNC,          9 pin     Ground
AUTO-STIR probe                  pH, ORP       miniDIN,
                                  or ISE         DO

Portable Meter Electrode Connections




                8 pin miniDIN,     RS232
                 Cond or ATC

           BNC,            9 pin
          pH, ORP         miniDIN,
           or ISE           DO


Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide           EN-9
Preparation


Meter Connections with Multiple Functions
•	 Use the BNC input to connect pH, ISE and ORP electrodes with a BNC or
   waterproof BNC connector.

•	 Benchtop meters have a reference input that is used to connect a separate
   reference electrode. Reference electrodes require an separate, appropriate
   sensing electrode for measurements.

•	 The 970899WP dissolved oxygen probe can be used on the BNC input.

•	 Use the 8 pin miniDIN input for conductivity probes or for automatic temperature
   compensation (ATC) probes.

•	 The DO AUTO-STIR probe uses the 9 pin miniDIN input and the stirrer jack.

•	 Connect a printer or computer to the RS232 input using the appropriate cable.


Turning on the Instrument
With the batteries installed in the portable meters and the power adapter attached
or the batteries installed in the benchtop meter, press     to turn on the meter.

If using a 3-Star, 4-Star or 5-Star meter, press  when the meter is powered on
to toggle the backlight on and off. When the benchtop meter is drawing line power,
the backlight will stay on until     is pressed.

To turn off the meter, press and hold      for about three seconds.


Meter Maintenance
For routine meter maintenance, dust and wipe the meter with a damp cloth. If
necessary, a warm water or a mild water-based detergent can be used. Perform
meter maintenance on a daily, weekly or monthly basis, as required by the
environment in which the meter is operated.

Immediately remove any spilled substance from the meter using the proper cleaning
procedure for the type of spill.


EN-10                   Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Chapter V           Meter Setup
Setup Menu
To navigate the setup menu:

1.	 Press       to enter the setup menu.

2.	 Press       /       until the desired setup option is displayed on the top line.

3.	 Press       to move the arrow icon to the middle line.

4.	 Press       /       until the desired setup option is displayed on the middle line.

5.	 Press       to move the arrow icon to the bottom line.

6.	 To scroll through a list of options on the bottom line, press     /     until the
    desired option is displayed. To enter a numeric value for an option on the bottom
    line, press     /       to adjust each digit and     to move to the next digit.
    For example, to change the pH measurement resolution press            to scroll from
    0.01 to 0.001 on the bottom display line.

	                   	                 	




7 Press
 .	             to move the arrow icon to the top line.

8.	 Repeat steps 2 through 7 to program a new setup option or press             to exit
    the setup menu and return to the measurement mode.

Note: Refer to Appendix A for a description of the special setup menu features.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                    EN-11
Meter Setup


Setup Menu Table – The following table is for the complete line of Orion Star and
Star Plus meters. Meters may not include all of the options listed in this table.

 Top Line   Middle Line    Bottom Line                     Setup Menu Description (default setting, method specific)
 PH         rES            0.1, 0.01, 0.001                pH measurement resolution (0.01, yes)
 PH         bUF            USA, EUr0                       pH buffer set for automatic buffer recognition during calibration,
                                                           USA buffers are 1.68, 4.01, 7.00, 10.01, 12.46 and EUrO buffers are
                                                           1.68, 4.01, 6.86, 9.18 (USA, yes)
 ISE        rES            1, 2, 3                         ISE measurement resolution in significant figures (1, yes)
 ISE        UnIt           m, mgL, PEr, PPb, n0nE          ISE measurement units (PPb, yes)
 ISE        rAng           L0w, HIgH                       ISE concentration range for calibration stability criteria (HIgH, yes)
 ISE        nLIn           AUt0, 0FF                       ISE automatic blank correction for low-level calibration (AUt0, yes)
 C0nd       tC             0FF, LIn, nLF                   Conductivity temperature compensation type, LIn is for linear, nLF is
                                                           for non-linear pure water samples (LIn, yes)
 C0nd       C0EF           0.0 to 10.0                     Conductivity temperature compensation coefficient in % change in
                                                           conductivity per °C, appears if LIn was selected for tC (2.1, yes)
 C0nd       tdSF           0.00 to 10.0                    Conductivity TDS factor value (0.49, yes)
 C0nd       CELL           0.001 to 199.0                  Conductivity default cell constant value for automatic conductivity
                                                           calibration mode (0.475, yes)
 C0nd       trEF           5, 10, 15, 20, 25               Conductivity reference temperature (25, yes)
 C0nd       tyPE           Std, 1, 2, 3, 4, 5, 6, 7, USP   Conductivity cell type and selectable range (Std, yes)
 See the Setup Menu for RDO Optical Dissolved Oxygen Sensor section for details on the RDO sensor information menus.
 d0         rES            0.1, 1 % sat                    DO % saturation measurement resolution (0.1, yes)
 d0         rES            0.01, 0.1 mg/L                  DO mg/L measurement resolution (0.01, yes)
 d0         bAr            AUt0, mAn                       DO barometric pressure compensation type (AUt0, yes)
 d0         PrES           450.0 to 850.0                  DO manual barometric pressure compensation value, appears if
                                                           mAn was selected for bAr (760.0, yes)
 d0         SAL            AUt0, mAn                       DO salinity correction type (AUt0, yes)
 d0         SALF           0 to 45                         DO manual salinity correction value, appears if mAn was selected
                                                           for SAL or a DO meter without a conductivity mode is used (0, yes)
 d0         CALt           AIr, H20, mAn, SEt0             DO calibration type (AIr, yes)
 d0         LIFE           365 to 0                        RDO cap replacement countdown in days – 3, 4 and 5 Star Plus
                                                           RDO meters only (Set by RDO sensor, no)
 dUE        PH             0 to 9999                       pH calibration alarm value in hours, 0 is off (0, yes)
 dUE        0rP            0 to 9999                       ORP calibration alarm value in hours, 0 is off (0, yes)
 dUE        ISE            0 to 9999                       ISE calibration alarm value in hours, 0 is off (0, yes)
 dUE        C0nd           0 to 9999                       Conductivity calibration alarm value in hours, 0 is off (0, yes)
 dUE        d0             0 to 9999                       DO calibration alarm value in hours, 0 is off (0, yes)


EN-12                           Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Meter Setup




 Top Line   Middle Line   Bottom Line                Setup Menu Description (default setting, method specific)
 rEAd       tyPE          AUt0, tImE, C0nt,          Measurement read type as AUTO-READ, timed or continuous
                                                     (AUt0, yes)
 rEAd       tInE          00:05 to 99:59             Timed measurement value in minutes and seconds (01:00, yes)
 L0g        dEL           n0, YES                    Delete datalog after download option (n0, yes)
 L0g        AUt0          0FF, 0n                    Automatic datalog point saving option (0FF, yes)
 gEn        dEgC          -5.0 to 105                Manual temperature value (25.0, yes)
 gEn        StIr          0FF, 1, 2, 3, 4, 5, 6, 7   Stirrer speed – 3, 4 and 5 Star benchtop meters only (4, yes)
 gEn        PASS          0000 to 9999               Meter password entry (0000, yes)
 gEn        AUt0          0n, 0FF                    Automatic meter shutoff option (0n, no)
 gLP        SEt           0FF, 0n                    GLP option, GLP feature enables or disables methods (0FF, no)
 dAtE       H0Ur          HH00 to HH23               Hour setting (HH12, no)
 dAtE       tInE          mm00 to mm59               Minute setting (mm00, no)
 dAtE       tYPE          mdY, dmY                   Date format as month, day, year or day, month, year (mdY, no)
 dAtE       YEAr          2000 to 2099               Year setting (2004, no)
 dAtE       dAtE          mm01 to mm12               Month setting (mm01, no)
 dAtE       dAY           dd01 to dd31               Day of the month setting (dd01, no)
 r232       bAUd          1200, 2400, 4800, 9600     Baud rate setting (9600, no)
 r232       0UtF          Prnt, C0mP                 Output format for printer or computer, COmP format is comma
                                                     delimited (Prnt, no)
 AUt0       SAPL          0FF, 0n                    Autosampler – 3, 4 and 5 Star Plus benchtop meters only (OFF, no)
 AUt0       trAY          28, 48                     Beaker tray setting (28, no)
 AUt0       rInb          1, 2, 3, 4, 5              Number of rinse beakers (3, no)
 AUt0       rSEC          5 to 60                    Rinse time in each rinse beaker, seconds (10, no)
 AUt0       PH            0, 1, 2, 3                 pH calibration points (3, no)
 AUt0       0rP           n0, YES                    ORP calibration option, appears if 0 was selected for PH (n0, no)
 AUt0       ISE           0, 2, 3                    ISE calibration points, appears if 0 was selected for PH (2, no)
 1          AUt0          ISE1                       Concentration value of ISE standard 1, appears if 2 or 3 was
                                                     selected for the ISE calibration points (1, no)
 10         AUt0          ISE2                       Concentration value of ISE standard 1, appears if 2 or 3 was
                                                     selected for the ISE calibration points (10, no)
 100        AUt0          ISE3                       Concentration value of ISE standard 1, appears if 3 was selected
                                                     for the ISE calibration points (100, no)
 AUt0       C0nd          0, 1, 2, 3                 Conductivity calibration points (3, no)
 AUt0       n0SA          0 to 47                    Number of sample beakers (1, no)


Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                                                  EN-13
Meter Setup


General Menu Settings
                     •	   Manual Temperature controls temperature compensation
                          when no temperature sensor is attached to the meter.

                     •	   Stirrer Speed sets the stirrer speed from 1 (slowest)
                          through 7 (fastest) and off (3-Star, 4-Star and 5-Star
                          benchtop meters only).

                     •	   Password Protection protects setup menu options and
                          methods from being accidentally erased or tampered with
                          (3-Star, 4-Star and 5-Star meters only).

                     •	   Automatic Shutoff controls whether the instrument will
                          automatically turn off after 20 minutes without a keypress.

1.	 In the measurement mode, press            .

2.	 Press       /     to scroll through the setup menu until           is displayed on the
    top line.

3.	 Press       to accept the selection and move the arrow icon to the middle line.

4.	 Press      /      to scroll through           for the manual temperature setting,
           for the stirrer speed setting,           for password entry and         for
    the automatic shutoff setting.

5.	 Press       to accept the selection and move the arrow icon to the bottom line.

6.	 To scroll through a list of options on the bottom line, press     /     until the
    desired option is displayed. To enter a numeric value for an option on the bottom
    line, press     /       to adjust each digit and     to move to the next digit.

7.	 Press       to accept the selection and move the arrow icon to the top line.

8.	 Repeat steps 3 through 7 to change another general setting or press              to
    return to the measurement mode.



EN-14                     Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Meter Setup


Time and Date Settings
                     •	   The date and time settings are saved with the data and
                          calibration log points and are included with the data that is
                          sent to a computer or printer.

                     •	   The date format can be set to read month, day, year or
                          day, month, year according to the user’s preference.

1.	 In the measurement mode, press            .

2.	 Press       /     to scroll through the setup menu until          is displayed on
    the top line.

3.	 Press       to accept the selection and move the arrow icon to the middle line.

4.	 Press       /     to scroll through         for the current hour setting,
    for the current minute setting,          for the date format setting,        for
    the current month setting,         for the current day setting and        for the
    current year setting.

5.	 Press       to accept the selection and move the arrow icon to the bottom line.

6.	 To scroll through a list of options on the bottom line, press     /     until the
    desired option is displayed. To enter a numeric value for an option on the bottom
    line, press     /       to adjust each digit and     to move to the next digit.

7.	 Press       to accept the selection and move the arrow icon to the top line.

8.	 Repeat steps 3 through 7 to change another time and date setting or press
    to return to the measurement mode.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                EN-15
Meter Setup


AUTO-READ™, Continuous or Timed Measurement Settings
                      •	   In the AUTO-READ mode, the meter starts taking
                           a measurement when             is pressed. Once the
                           measurement is stable, the display freezes and the data is
                           logged and printed. The AUTO-READ mode also controls
                           the stirrer. The stirrer starts when       is pressed and stops
                           when the measurement becomes stable.

•	 In the continuous mode, the meter is constantly taking measurements and
   updating the display. Press    to log and print a measurement in this mode.

•	 In the timed mode, the meter is constantly taking measurements and updating
   the display. The meter logs and prints the measurement at the selected time
   interval. Timed dissolved oxygen measurements with the RDO sensor are taken
   only at the selected time interval, which conserves the meter battery power.

1.	 In the measurement mode, press              .

2.	 Press       /      to scroll through the setup menu until             is displayed on
    the top line.

3.	 Press        to accept the selection and move the arrow icon to the middle line.

4.	 Press         /     to scroll through           for the measurement read type and
              for the timed reading interval.

5.	 Press        to accept the selection and move the arrow icon to the bottom line.

6.	 To scroll through a list of options on the bottom line, press    /     until the
    desired option is displayed. To enter a numeric value for an option on the bottom
    line, press     /      to adjust each digit and    to move to the next digit.

7.	 Press        to accept the selection and move the arrow icon to the top line.

8.	 Repeat steps 3 through 7 to change another measurement setting or press
    to return to the measurement mode.

EN-16                      Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Meter Setup


Selecting the Measurement Parameter
In the measurement mode, the arrow icon on the left side of the display indicates
the active line. Press      to move the arrow icon to the desired measurement line
and press       /     to scroll through the measurement parameters associated with
the selected line.

The measurement lines and icons for the 5-Star multi-parameter meter are shown
below. The single and dual parameter meters will have fewer measurement lines
and icons, depending on the meter capabilities.

             pH
             mV
             RmV
             ISE
             No icon for temperature
             No icon and no measurement – the measurement line is turned off

             μS/cm or mS/cm for conductivity
             mg/L for TDS
             ppt for salinity
             MΩ-cm for resistivity
             No icon for temperature
             No icon and no measurement – the measurement line is turned off

             % sat for dissolved oxygen percent saturation
             mg/L for dissolved oxygen concentration
             No icon for barometric pressure
             No icon for sample temperature
             No icon for membrane temperature (polarographic DO readings only)
             No icon and no measurement – the measurement line is turned off

Note: If a measurement line is not needed, press       to move the arrow
icon to the measurement line that is not needed and press     /     until the
measurement line is completely blank.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide             EN-17
Meter Setup


Method Setup
The Orion 3-Star, 4-Star and 5-Star meters can save up to 10 methods when the
GLP function is enabled. When a method is selected, the meter will use the last
calibration performed in that method, so electrodes that share a common meter
connection can be more easily interchanged. When using multiple methods, a
calibration must be performed for each method that will be used.

1.	 To enable the GLP function:

	   a.	 In the measurement mode, press         .

	   b.	 Press     /     until      is displayed on top line.

	   c.	
       Press      to move the arrow icon to the middle line and press           /
       until     is displayed.

	   d.	
       Press       to move the arrow icon to the bottom line and press          /
       until    is displayed.

	   e.	 Press     to move the arrow icon to the top line.

	   f.	 Press     to exit the setup menu and return to the measurement mode.

2.	 To display and change the current method number:

	   a.	 the measurement mode, press
       In                                      . The current method number will
       be displayed.

	   b.	
       Press      /     to select a new method number.

	   c.	
       Press   to save the method number and press               to return to the
       measurement mode.




EN-18                  Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Chapter VI           pH Technique
pH Setup Menu
Note: Refer to the Setup Menu section for the Setup Menu Table, which contains
a complete list of meter setup options and descriptions. Refer to Appendix A for a
description of the special setup menu features.

1.	 In the measurement mode, press            .

2.	 Press       /     to scroll through the setup menu until          is displayed on the
    top line.

3.	 Press       to accept the selection and move the arrow icon to the middle line.

4.	 Press         /     to scroll through      for pH measurement resolution and
            for the automatic buffer recognition setting.

5.	 Press       to accept the selection and move the arrow icon to the bottom line.

6.	 To scroll through a list of options on the bottom line, press         /     until the
    desired option is displayed.

7 Press
 .	             to accept the selection and move the arrow icon to the top line.

8.	 Repeat steps 3 through 7 to change another pH setting or press             to return to
    the measurement mode.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                   EN-19
pH Technique


pH Calibration
1.	 Prepare the electrode according to the electrode user guide.

2.	 In the setup mode, select the buffer set (       or         ) that will be used for
    the automatic buffer recognition feature.

3.	 In the measurement mode, press        until the arrow icon points to the top line,
    press     until the pH icon is shown and press       to begin the calibration.

4.	 Rinse the electrode, and ATC probe if being used, with distilled water and place
    into the buffer.

5.	 Wait for the pH icon to stop flashing.

	      Automatic buffer recognition – When the pH icon stops flashing the meter
    a.	
       will display the temperature-corrected pH value for the buffer.
	      Manual calibration – When the pH icon stops flashing the meter will display
    b.	
       the actual pH value read by the electrode. Press          until the first digit
       to be changed is flashing, press        /      to change the value of the
       flashing digit and continue to change the digits until the meter displays the
       temperature-corrected pH value of the buffer. Once the pH buffer value is
       set, press       until the decimal point is in the correct location.

6.	 Press      to proceed to the next calibration point and repeat steps 4 and 5
    or press      to save and end the calibration.

7 The actual electrode slope, in percent, will be displayed in the main field and
 .	
       will be displayed in the lower field.

	   a.	 a one point calibration, press
       For                               and     /              to edit the slope and
       press    to return to the measurement mode.
	   b.	 or a two or more point calibration, the meter will automatically proceed to
        F
        the measurement mode after the slope is displayed.




EN-20                   Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
pH Technique


pH Measurement
Note: Turn on the automatic datalog feature to send measurements to the meter
datalog at the frequency specified in each measurement mode. Refer to the Data
Archiving and Retrieval section for details. If the automatic datalog feature is off,
connect the meter to a printer or computer to record the measurements.

1.	 Rinse the electrode with distilled or deionized water. Shake off any excess
    water and blot the electrode dry with lint-free tissue.

2.	 Place the electrode into the sample.

	   a. 	 the meter is in the continuous measurement mode, it will start reading
        If
        immediately and continuously update the display. The pH icon will flash
        until the reading is stable. Once the reading is stable, log and print the
        measurement by pressing           . If a benchtop meter is used and the stirrer
        is enabled, press       to start the stirrer. Press    again to turn off the
        stirrer before removing the electrode and stirrer from the sample.
	   b.	 the meter is in the AUTO-READ measurement mode, press
       If                                                                   to start
       the reading. The AR icon will flash until the reading is stable. Once the
       reading is stable, the meter will log and print the measurement and freeze
       the display. If a benchtop meter is used and the stirrer is enabled, the stirrer
       will turn on when       is pressed and turn off when the reading is stable.
	   c.	 the meter is in the timed measurement mode, it will start reading
       If
       immediately and continuously update the display. The meter will log and
       print the measurement at the frequency specified in the setup menu. If a
       benchtop meter is used and the stirrer is enabled, press        to start the
       stirrer. Press     again to turn off the stirrer before removing the electrode
       and stirrer from the sample.

3.	 Remove the electrode from the sample, rinse it with distilled or deionized water,
    blot it dry, place it in the next sample and repeat step 2.

4. 	 Once all of the samples have been measured, rinse the electrode with distilled
     or deionized water and blot it dry. Consult the electrode user guide for proper
     storage techniques.


Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                EN-21
pH Technique


pH Temperature Display and Calibration
pH Temperature Display
Star Plus meters allow the temperature to be viewed on individual measurement
lines in addition to the temperature display on the top, left of the screen.

To view the temperature for the pH measurement line:

1.	 In the measurement mode, press           to select the top display line. The arrow
    icon will point to the selected line.

2.	 Press      /     to change the value on the selected line. The top line can
    be changed to display pH (pH), millivolts (mV), relative millivolts (RmV),
    concentration (ISE), temperature (no icon) or a blank line.

pH Temperature Calibration
The temperature calibration mode of the Star Plus meter allows the temperature on
each measurement line to be manually adjusted.

To calibrate the temperature for the pH measurement line:

1.	 In the measurement mode, press       to choose the top measurement line and
    press     /    until the temperature is shown for the selected line.

2.	 Press      to begin the calibration.

3.	 When the reading stabilizes, the arrow icon and the first digit will flash. Enter
    the temperature by pressing       /     to adjust each digit and        to move to
    the next digit.

4.	 Press      to save and end the calibration.




EN-22                    Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
mV, Relative mV and
Chapter VII
ORP Technique
All meters with pH measurement capability include a mV, relative mV and ORP
function. Measure the raw millivolt (mV) values of an electrode in the mV mode.
Calibrate the relative millivolt (RmV) values of a redox electrode for oxidation-
reduction potential (ORP) measurements in the relative mV and ORP mode. Refer
to the pH Technique section for information on the temperature display and
temperature calibration of the pH/mV/RmV/ISE measurement line.

Note: The mV measurements are raw readings and cannot be calibrated. Use the
relative mV mode to calibrate mV measurements.


Relative mV and ORP Calibration
1.	 Prepare the electrode according to the electrode user guide.

2.	 In the measurement mode, press      until the arrow icon points to the top line,
    press     until the RmV icon is shown and press       to begin the calibration.

3.	 Rinse the electrode with distilled water and place it into the standard.

4.	 Wait for the RmV icon to stop flashing. When the RmV icon stops flashing, the
    meter will display 000.0 RmV. Press          until the first digit to be changed is
    flashing, press       /      to change the value of the flashing digit and continue
    to change the digits until the meter displays the millivolt value of the standard.
    To change the value to negative or positive number, press             until none of the
    digits are blinking and the arrow icon is blinking and then press           to change
    the sign of the millivolt value.

5.	 Press       to save and end the calibration.

6.	 The millivolt offset will be displayed and the meter will automatically proceed to
    the measurement mode.


Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                  EN-23
mV, Relative mV and ORP Technique


mV, Relative mV and ORP Measurement
Note: Turn on the automatic datalog feature to send measurements to the meter
datalog at the frequency specified in each measurement mode. Refer to the Data
Archiving and Retrieval section for details. If the automatic datalog feature is off,
connect the meter to a printer or computer to record the measurements.

1.	 Rinse the electrode with distilled or deionized water. Shake off any excess
    water and blot the electrode dry with lint-free tissue.

2.	 Place the electrode into the sample.

	   a. 	 the meter is in the continuous measurement mode, it will start reading
        If
        immediately and continuously update the display. The mV or RmV icon will
        flash until the reading is stable. Once the reading is stable, log and print the
        measurement by pressing           . If a benchtop meter is used and the stirrer
        is enabled, press       to start the stirrer. Press    again to turn off the
        stirrer before removing the electrode and stirrer from the sample.
	   b.	 the meter is in the AUTO-READ measurement mode, press
       If                                                                   to start
       the reading. The AR icon will flash until the reading is stable. Once the
       reading is stable, the meter will log and print the measurement and freeze
       the display. If a benchtop meter is used and the stirrer is enabled, the stirrer
       will turn on when       is pressed and turn off when the reading is stable.
	   c.	 the meter is in the timed measurement mode, it will start reading
       If
       immediately and continuously update the display. The meter will log and
       print the measurement at the frequency specified in the setup menu. If a
       benchtop meter is used and the stirrer is enabled, press        to start the
       stirrer. Press     again to turn off the stirrer before removing the electrode
       and stirrer from the sample.

3.	 Remove the electrode from the sample, rinse it with distilled or deionized water,
    blot it dry, place it in the next sample and repeat step 2.

4. 	 Once all of the samples have been measured, rinse the electrode with distilled
     or deionized water and blot it dry. Consult the electrode user guide for proper
     storage techniques.


EN-24                   Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Chapter VIII           Dissolved Oxygen Technique
Dissolved Oxygen Setup Menu
Note: Refer to the Setup Menu section for the Setup Menu Table, which contains
a complete list of meter setup options and descriptions. Refer to Appendix A for a
description of the special setup menu features.

1.	 In the measurement mode, press            .

2.	 Press       /     to scroll through the setup menu until          is displayed on the
    top line.

3.	 Press       to accept the selection and move the arrow icon to the middle line.

4.	 Press       /      to scroll through       for the % saturation resolution,      for
    the mg/L concentration resolution,          for the barometer type (automatic or
    manual),           for the manual barometric pressure compensation value,
    for the salinity compensation type (automatic or manual),            for the manual
    salinity correction value and         for the dissolved oxygen calibration type.

5.	 Press       to select the option and move the arrow icon to the bottom line.

6.	 To scroll through a list of options on the bottom line, press    /     until the
    desired option is displayed. To enter a numeric value for an option on the bottom
    line, press     /      to adjust each digit and    to move to the next digit.

7.	 Press       to accept the selection and move the arrow icon to the top line.

8.	 Repeat steps 3 through 7 to change another dissolved oxygen setting or press
        to return to the measurement mode.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                   EN-25
Dissolved Oxygen Technique


Dissolved Oxygen Calibration
•	 Prior to calibration, the dissolved oxygen probe must be prepared and polarized.
   The probe is continuously polarized when it is connected to the meter. When
   the probe is first connected or if the probe is disconnected for more than 60
   minutes, connect the probe to the meter, connect the meter to a power source
   and wait 30 to 60 minutes for the probe to polarize. Disconnecting the probe for
   less than one hour will require 5 to 25 minutes for polarization.

•	 The meters will supply a polarization current to the dissolved oxygen probe even
   when the meter power is off. To maximize the meter battery life, unplug the
   probe if it will not be used for an extended period.

1.	 Select one of the following calibration modes in the setup menu.

	   a.	     – An air calibration is performed in water saturated air using the
       calibration sleeve. This is the simplest and most accurate calibration. Due to
       the inherent differences between water saturated air and air saturated water,
       102.3% saturation will be displayed when the calibration reading is stable.

		      i.	
           The highest possible accuracy is reached when calibration temperature is
           the same as the measuring temperature.
		      ii.	 oisten the sponge or absorbent cloth in the calibration sleeve with
            M
            distilled water and insert the probe into the sleeve without touching the
            water saturated material. For BOD measurements, this calibration can be
            performed in a BOD bottle.

	   b.	     – A water calibration is performed using water that is 100% saturated
       with air. Bubble air into a water sample and gently stir the sample to prevent
       the buildup of air bubbles on the dissolved oxygen probe membrane.

	   c.	      – A manual calibration is performed using a water sample with a
       known concentration of dissolved oxygen. This method can be used to
       calibrate the dissolved oxygen probe to the value achieved by a
       Winkler titration.



EN-26                   Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Dissolved Oxygen Technique


		      i.	 manual calibration involves performing a Winkler titration and using
           A
           that sample as a calibration standard. The oxygen level result from the
           titration is entered in a manual calibration as the dissolved oxygen value.
           This correlates the meter input to the Winkler titration. This method is
           inherently less accurate, due to the possibility of titration errors.

	   d.	       – A zero point calibration is performed in an oxygen-free solution. A
       zero point calibration is not generally required unless measurements will be
       taken below 10% saturation or 1 mg/L. Zero the probe when using a new
       membrane, using fresh filling solution or when measuring dissolved oxygen
       levels below 1 mg/L. An air calibration should be performed prior to the zero
       point calibration.

2.	 Allow the probe and calibration standard (water saturated air, air saturated water,
    Winkler standard or oxygen-free solution) to reach equilibrium.

3.	 In the measurement mode, press         until the arrow icon points to the bottom
    line, press      until the % sat or mg/L icon is shown and press        to begin
    the calibration.

4.	 Wait for the dissolved oxygen reading to stabilize.

	   a.	 an air calibration is performed, the meter will display 102.3% and
       If
       automatically return to the measurement mode.
	   b.	 a water calibration is performed, the meter will display 100.0% and
       If
       automatically return to the measurement mode.
	   c.	 a manual calibration is performed, wait for the mg/L icon to stop flashing
       If
       and enter the dissolved oxygen value by pressing          until the first digit to
       be changed is flashing, press        /      to change the value of the flashing
       digit and continue to change the digits until the meter displays the correct
       dissolved oxygen value. Once the dissolved oxygen value is set, press
       until the decimal point is in the correct location.
	   d.	 a zero point calibration is performed, the meter will display 0.00 and
       If
       automatically return to the measurement mode.



Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                  EN-27
Dissolved Oxygen Technique


Dissolved Oxygen Measurement
Note: Turn on the automatic datalog feature to send measurements to the meter
datalog at the frequency specified in each measurement mode. Refer to the Data
Archiving and Retrieval section for details. If the automatic datalog feature is off,
connect the meter to a printer or computer to record the measurements.

1.	 Rinse the dissolved oxygen probe with distilled or deionized water. Shake off
    any excess water and blot the probe dry with lint-free tissue.

2.	 Place the dissolved oxygen probe into the sample.

	   a. 	 the meter is in the continuous measurement mode, it will start reading
        If
        immediately and continuously update the display. The mg/L or % sat icon
        will flash until the reading is stable. Once the reading is stable, log and print
        the measurement by pressing           . If a benchtop meter is used and the
        stirrer is enabled, press       to start the stirrer. Press    again to turn off
        the stirrer before removing the probe and stirrer from the sample.
	   b.	 the meter is in the AUTO-READ measurement mode, press
       If                                                                   to start
       the reading. The AR icon will flash until the reading is stable. Once the
       reading is stable, the meter will log and print the reading and freeze the
       display. If a benchtop meter is used and the stirrer is enabled, the stirrer will
       turn on when        is pressed and turn off when the reading is stable. If the
       BOD AUTO-STIR probe is used, press the button on the probe to start the
       AUTO-READ measurement.
		      N
         ote: Benchtop Star Plus dissolved oxygen meters allow the measurement
        and stirring functions to be controlled by the AUTO-STIR probe when the
        meter is in the AUTO-READ measurement mode. Press the button on the
        AUTO-STIR probe to start and stop measurement and stirring functions.
	   c.	 the meter is in the timed measurement mode, it will start reading
       If
       immediately and continuously update the display. The meter will log and
       print the measurement at the frequency specified in the setup menu. If a
       benchtop meter is used and the stirrer is enabled, press        to start the
       stirrer. Press     again to turn off the stirrer before removing the probe and
       stirrer from the sample.


EN-28                    Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Dissolved Oxygen Technique


3.	 Remove the dissolved oxygen probe from the sample, rinse it with distilled or
    deionized water, blot it dry, place it in the next sample and repeat step 2.

4. 	 Once all of the samples have been measured, rinse the dissolved oxygen probe
     with distilled or deionized water and blot it dry. Consult the dissolved oxygen
     probe user guide for proper storage techniques.


Dissolved Oxygen Temperature Display and Calibration
Dissolved Oxygen Temperature Display
Star Plus meters allow the temperature to be viewed on individual measurement
lines in addition to the temperature display on the top, left of the screen.

To view the temperature for the dissolved oxygen measurement line:

1.	 In the measurement mode, press           to choose the bottom display line. The
    arrow icon will point to the selected line.

2.	 Press      /    to change the value on the selected line. The bottom line
    can be changed to display dissolved oxygen (% saturation), dissolved oxygen
    (mg/L), barometric pressure (no icon), sample temperature (no icon), electrolyte
    solution/membrane temperature (no icon, after number) or a blank line.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                 EN-29
Dissolved Oxygen Technique


Dissolved Oxygen Temperature Calibration
The temperature calibration mode of the Star Plus meter allows the temperature
on each measurement line to be manually adjusted. The dissolved oxygen
measurement line displays the sample temperature and the electrolyte solution/
membrane temperature.

To calibrate the sample temperature for the dissolved oxygen measurement line:

1.	 In the measurement mode, press       to choose the bottom measurement line
    and press     /    until the sample temperature is displayed (i.e. ).

2.	 Press      to begin the calibration.

3.	 When the reading stabilizes, the arrow icon and the first digit will flash. Enter
    the temperature by pressing       /     to adjust each digit and        to move to
    the next digit.

4.	 Press      to save and end the calibration.

To calibrate the electrolyte solution/membrane temperature for the dissolved oxygen
measurement line:

1.	 In the measurement mode, press     to choose the bottom measurement line
    and press     /    until the membrane temperature is displayed (i.e.   ).

2.	 Press      to begin the calibration.

3.	 When the reading stabilizes, the arrow icon and the first digit will flash. Enter
    the temperature by pressing       /     to adjust each digit and        to move to
    the next digit.

4.	 Press      to save and end the calibration.




EN-30                   Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
RDO® Optical Dissolved
Chapter IX
Oxygen Technique
Note: The RDO optical dissolved oxygen meters have serial numbers that begin
with R (i.e. R12345). Only the RDO meters are compatible with the RDO optical
dissolved oxygen sensors.


RDO Cap Overview
The RDO sensor has an internal clock that counts down the 365 day lifespan of a
new RDO cap. The countdown begins when the RDO cap is installed on the RDO
sensor, the sensor is connected to the meter and the first measurement is taken.
The 365 day countdown cannot be reset or changed once the first measurement
is taken. Each RDO cap has a unique serial number that is recognized by the RDO
sensor, so reinstalling the cap will not reset the countdown.

Note: The power to the meter must be turned off when a new RDO cap is
installed on the RDO sensor. Once the RDO cap is installed, turn the power to the
meter on and the new cap information will be sent to the meter.

To print the RDO cap information:

1.	 Connect the meter to a printer or computer and verify the meter baud rate and
    output settings in the setup menu.

2.	 From the measurement mode, press             .

3.	 Press       /    to scroll through the setup menu until             is displayed on
    the top line and     is displayed on the middle line.

4.	 Press       two times to move the arrow icon to the bottom line.

5.	 Press       /     until        is displayed on the bottom line.

6.	 Press   to print the RDO cap information. Press               to return to the
    measurement mode.

Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                  EN-31
RDO® Optical Dissolved Oxygen Technique


RDO Optical Dissolved Oxygen Setup Menu
Note: Refer to the Setup Menu section for the Setup Menu Table, which contains
a complete list of meter setup options and descriptions.

1.	 From the measurement mode, press            .

2.	 Press       /     to scroll through the setup menu until        is displayed on the
    top line.

3.	 Press       to accept the selection and move the arrow icon to the middle line.

4.	 Press        /     to scroll through       for the % saturation resolution,      for
    the mg/L concentration resolution,          for the barometer type (automatic or
    manual),           for the manual barometric pressure compensation value,
    for the salinity compensation type (automatic or manual),            for the manual
    salinity correction value,         for the dissolved oxygen calibration type and
             for the RDO cap replacement countdown in days.

5.	 Press       to select the option and move the arrow icon to the bottom line.

6.	 To scroll through a list of options on the bottom line, press     /     until the
    desired option is displayed. To enter a numeric value for an option on the bottom
    line, press     /       to adjust each digit and     to move to the next digit.

7.	 Press       to accept the selection and move the arrow icon to the top line.

8.	 Repeat steps 3 through 7 to change another dissolved oxygen setting or press
        to return to the measurement mode.




EN-32                    Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
RDO® Optical Dissolved Oxygen Technique


Setup Menu for RDO Optical Dissolved Oxygen Sensor
The RDO optical dissolved oxygen meters have a special setup menu feature that
allows the user to print information about the RDO sensor.

 Top Line    Middle Line   Bottom Line   Setup Menu Description
 Pr0b        d0            tESt          The tESt option initiates a 30 second test of the communication link
                                         between the RDO sensor and meter. The temperature display will show a
                                         30 second countdown as the test progresses. The meter will send a report
                                         to the printer or computer when the test is done.
 Pr0b        d0            CAL           The CAL option prints the meter serial number, date, time and slope for the
                                         last five dissolved oxygen calibrations that were performed with the RDO
                                         sensor that is currently connected to the meter.
 Pr0b        d0            dFLt          The dFLt option prints the factory calibration information for the RDO
                                         sensor that is currently connected to the meter.
 Pr0b        d0            SLP           The SLP option prints a drift per minute value of the slope for the last five
                                         dissolved oxygen calibrations that were performed with the RDO sensor
                                         that is currently connected to the meter.
 Pr0b        d0            mEtH          The mEtH option prints the RDO specific method information of the last
                                         method used for the RDO sensor that is currently connected to the meter.
 Pr0b        d0            InF0          The InF0 option prints the RDO sensor information, including the sensor
                                         serial number and revision, sensor and RDO cap date of manufacturing,
                                         cap start date, cap life and internal real time clock for the sensor that is
                                         currently connected to the meter.


1.	 Connect the meter to a printer or computer and verify the meter baud rate and
    output settings in the setup menu.

2.	 From the measurement mode, press                      .

3.	 Press       /    to scroll through the setup menu until                                  is displayed on
    the top line and     is displayed on the middle line.

4.	 Press         two times to move the arrow icon to the bottom line.

5.	 Press       /     to scroll through         for the communication link test,
    for the RDO sensor calibration information,           for the RDO sensor factory
    calibration information,        for the RDO sensor slope information,         for
    the RDO method information and             for the general RDO sensor information.


Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                                              EN-33
RDO® Optical Dissolved Oxygen Technique


6.	 Press     to print the selected option.

7 Press
 .	             /     to select another option from the bottom line and press         to
    print the selected option or press      to return to the measurement mode.

Example RDO Sensor General Information Printout

 smartprobe info
 type		              1
 run_status	         255
 format_version	     1
 probe_SN		          6
 app_version	        9
 HW_version	         1

 cap_SN		            129614
 rtc		               1210257372   (05-08-2008   14:36:12)
 mfg_time		          1208371349   (04-16-2008   18:42:29)
 start_time	         1210178953   (05-07-2008   16:49:13)
 expiration	         1241736553   (05-07-2009   22:49:13)
 probe_mfg_time	     1208440800   (04-17-2008   14:00:00)
 write_count	        14

 pass_count	         12
 expire		            0x01
 days_left	364.3

 probe_reading	      149.529243
 temp_reading	       24.942651

 state		             21
 general_event	      0x00
 internal_event	     0x00
 internal_state	     3
 internal_retry	     0x00




EN-34                  Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
RDO® Optical Dissolved Oxygen Technique


RDO Optical Dissolved Oxygen Calibration
1.	 Select one of the following calibration modes in the setup menu.

	   a.	     – An air calibration is performed in water saturated air using the
       calibration sleeve. This is the simplest and most accurate calibration
       method.

		      i.	
           The highest possible accuracy is reached when calibration temperature is
           the same as the measuring temperature.
		      ii.	 oisten the sponge in the calibration sleeve with distilled water.
            M
            Insert the RDO sensor into the sleeve without touching the water
            saturated sponge.

	   b.	     – A water calibration is performed using water that is 100% saturated
       with air. Bubble air into a water sample and gently stir the sample to prevent
       the buildup of air bubbles on the RDO cap.

	   c.	      – A manual calibration is performed using a water sample with a
       known concentration of dissolved oxygen. This method can be used to
       calibrate the RDO sensor to the value achieved by a Winkler titration.

		      i.	 manual calibration involves performing a Winkler titration and using
           A
           that sample as a calibration standard. The oxygen level result from the
           titration is entered in a manual calibration as the dissolved oxygen value.
           This correlates the meter input to the Winkler titration. This method is
           inherently less accurate, due to the possibility of titration errors.

	   d.	       – A zero point calibration is performed in an oxygen-free solution. A
        zero point calibration is not generally required unless measurements will be
        taken below 10% saturation or 1 mg/L.

2.	 Allow the RDO sensor and calibration standard (water saturated air, air saturated
    water, Winkler standard or oxygen-free solution) to reach equilibrium.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                EN-35
RDO® Optical Dissolved Oxygen Technique



3.	 In the measurement mode, press         until the arrow icon points to the bottom
    line, press      until the % sat or mg/L icon is shown and press        to begin
    the calibration.

4.	 Wait for the dissolved oxygen reading to stabilize.

	   a.	 an air calibration is performed, the meter will display 100.0% and
       If
       automatically return to the measurement mode.

	   b.	 a water calibration is performed, the meter will display 100.0% and
       If
       automatically return to the measurement mode.

	   c.	 a manual calibration is performed, wait for the mg/L icon to stop flashing
       If
       and enter the dissolved oxygen value by pressing          until the first digit to
       be changed is flashing, press        /      to change the value of the flashing
       digit and continue to change the digits until the meter displays the correct
       dissolved oxygen value. Once the dissolved oxygen value is set, press
       until the decimal point is in the correct location.

	   d.	 a zero point calibration is performed, the meter will display 0.00 and
       If
       automatically return to the measurement mode.




EN-36                    Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
RDO® Optical Dissolved Oxygen Technique


RDO Optical Dissolved Oxygen Measurement
Note: Turn on the automatic datalog feature to send measurements to the meter
datalog at the frequency specified in each measurement mode. Refer to the Data
Archiving and Retrieval section for details. If the automatic datalog feature is off,
connect the meter to a printer or computer to record the measurements.

1.	 Rinse the RDO sensor with distilled or deionized water. Shake off any excess
    water and blot the sensor dry with lint-free tissue.

2.	 Place the RDO sensor into the sample.

	   a. 	 the meter is in the continuous measurement mode, it will start reading
        If
        immediately and continuously update the display. The mg/L or % sat icon
        will flash until the reading is stable. Once the reading is stable, log and print
        the measurement by pressing           . This mode will use a large amount of
        meter battery power.
	   b.	 the meter is in the AUTO-READ measurement mode, press
       If                                                                  to start
       the reading. The AR icon will flash until the reading is stable. Once the
       reading is stable, the meter will log and print the measurement and freeze
       the display. This mode uses various amounts of meter battery power,
       depending on how often        is pressed.
	   c.	 the meter is in the timed measurement mode, it will take a dissolved
       If
       oxygen reading at the frequency specified in the setup menu. If parameters
       other than dissolved oxygen are measured, the meter will continuously take
       the other readings and update the display. The meter will log and print the
       measurement at the predetermined time interval. This mode can be used
       to extend the meter battery life, since the power consumption of the RDO
       sensor is reduced, depending on the set time interval.

3.	 Remove the RDO sensor from the sample, rinse it with distilled or deionized
    water, blot it dry, place it in the next sample and repeat step 2.

4. 	 Once all of the samples have been measured, rinse the RDO sensor with
     distilled or deionized water and blot it dry. Consult the RDO sensor user guide
     for proper storage techniques.


Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                EN-37
RDO® Optical Dissolved Oxygen Technique


RDO Optical Dissolved Oxygen Temperature Display and
Calibration
RDO Optical Dissolved Oxygen Temperature Display
Star Plus meters allow the temperature to be viewed on individual measurement
lines in addition to the temperature display on the top, left of the screen.

To view the temperature for the RDO dissolved oxygen measurement line:

1.	 In the measurement mode, press           to choose the bottom display line. The
    arrow icon will point to the selected line.

2.	 Press     /     to change the value on the selected line. The bottom line can
    be changed to display dissolved oxygen (% saturation), dissolved oxygen (mg/L),
    barometric pressure (no icon), temperature (no icon) or a blank line.

RDO Optical Dissolved Oxygen Temperature Calibration
The temperature calibration mode of the Star Plus meter allows the temperature on
each measurement line to be manually adjusted.

To calibrate the temperature for the RDO dissolved oxygen measurement line:

1.	 In the measurement mode, press       to choose the bottom measurement line
    and press     /    until the temperature is displayed.

2.	 Press      to begin the calibration.

3.	 When the reading stabilizes, the arrow icon and the first digit will flash. Enter
    the temperature by pressing       /     to adjust each digit and        to move to
    the next digit.

4.	 Press      to save and end the calibration.




EN-38                   Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Chapter X           Conductivity Technique
Conductivity Setup Menu
Note: Refer to the Setup Menu section for the Setup Menu Table, which contains
a complete list of meter setup options and descriptions. Refer to Appendix A for a
description of the special setup menu features.

1.	 In the measurement mode, press            .

2.	 Press       /     to scroll through the setup menu until          is displayed on
    the top line.

3.	 Press       to accept the selection and move the arrow icon to the middle line.

4.	 Press       /     to scroll through       for the temperature compensation type,
            for the temperature coefficient value used for         temperature
    compensation,           for the TDS factor value used for total dissolved
    solids measurement,             for the nominal cell constant value of the
    conductivity probe,           for the reference temperature used for temperature
    compensation and             for the conductivity cell type.

5.	 Press       to accept the selection and move the arrow icon to the bottom line.

6.	 To scroll through a list of options on the bottom line, press     /     until the
    desired option is displayed. To enter a numeric value for an option on the bottom
    line, press     /       to adjust each digit and     to move to the next digit.

7.	 Press       to accept the selection and move the arrow icon to the top line.

8.	 Repeat steps 3 through 7 to change another conductivity setting or press            to
    return to the measurement mode.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                EN-39
Conductivity Technique


Conductivity Calibration
Note: For an automatic calibration, the nominal cell constant of the conductivity
probe must be entered in the setup menu before the calibration is performed.

1.	 In the measurement mode, press       until the arrow icon points to the middle
    line, press     until the µS/cm or mS/cm icon is shown and press         to
    begin the calibration.

2.	 Rinse the probe with deionized water and place it into the conductivity standard.

3.	 To perform a manual calibration – The manual calibration screen will display the
    cell constant on the top line, the conductivity value of the calibration standard
    on the middle line and            on the bottom line. To change the cell constant,
    press      until the first digit to be changed is flashing, press      /      to
    change the value of the flashing digit and continue to change the digits until the
    displayed conductivity value matches the value of the standard at the measured
    temperature. Once the value is set, press           until the decimal point is in the
    correct location. Press         to save and end the calibration.
	   Note: In the manual calibration screen, start changing the cell constant within
    five seconds or the meter will proceed to the automatic/direct calibration. If this
    occurs, press and hold     to abort the calibration and repeat the calibration.

4.	 To perform an automatic or direct calibration – Wait for the meter to go from
    the manual calibration screen to the automatic/direct calibration screen. The
    automatic/direct calibration screen will display the conductivity value of the
    calibration standard on the middle line and          on the bottom line.
	      Automatic calibration – When the µS/cm or mS/cm icon stops flashing, the
    a.	
        meter will display the temperature-corrected conductivity of the standard.
	      Direct calibration – When the µS/cm or mS/cm icon stops flashing, the
    b.	
        meter will display the actual conductivity value read by the probe. To
        change the conductivity value, press         until the first digit to be changed
        is flashing, press      /     to change the value of the flashing digit and
        continue to change the digits until the correct conductivity value of the
        standard at the measured temperature is displayed. Once the value is set,
        press       until the decimal point is in the correct location.

EN-40                    Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Conductivity Technique


5. 	 Press        to proceed to the next calibration point, rinse the conductivity probe
     with distilled or deionized water, place it into the next conductivity standard and
     repeat step 4a / 4b or press       to save and end the calibration.

6.	 The cell constant will be displayed in the main field and the meter will
    automatically advance to the measurement mode.


Conductivity Measurement
Note: Turn on the automatic datalog feature to send measurements to the meter
datalog at the frequency specified in each measurement mode. Refer to the Data
Archiving and Retrieval section for details. If the automatic datalog feature is off,
connect the meter to a printer or computer to record the measurements.

1.	 Rinse the conductivity probe with distilled or deionized water. Shake off any
    excess water and blot the probe dry with lint-free tissue.

2.	 Place the conductivity probe into the sample.

	   a. 	 the meter is in the continuous measurement mode, it will start reading
        If
        immediately and continuously update the display. The μS/cm or mS/cm
        icon will flash until the reading is stable. Once the reading is stable, log and
        print the measurement by pressing           . If a benchtop meter is used and
        the stirrer is enabled, press       to start the stirrer. Press    again to turn
        off the stirrer before removing the probe and stirrer from the sample.
	   b.	 the meter is in the AUTO-READ measurement mode, press
       If                                                                    to start
        the reading. The AR icon will flash until the reading is stable. Once the
        reading is stable, the meter will log and print the measurement and freeze
        the display. If a benchtop meter is used and the stirrer is enabled, the stirrer
        will turn on when       is pressed and turn off when the reading is stable.
	   c.	 the meter is in the timed measurement mode, it will start reading
       If
       immediately and continuously update the display. The meter will log and
       print the measurement at the frequency specified in the setup menu. If a
       benchtop meter is used and the stirrer is enabled, press        to start the
       stirrer. Press     again to turn off the stirrer before removing the probe and
       stirrer from the sample.

Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                   EN-41
Conductivity Technique


3.	 Remove the conductivity probe from the sample, rinse it with distilled or
    deionized water, blot it dry, place it in the next sample and repeat step 2.

4. 	 Once all of the samples have been measured, rinse the conductivity probe with
     distilled or deionized water and blot it dry. Consult the conductivity probe user
     guide for proper storage techniques.


Conductivity Temperature Display and Calibration
Conductivity Temperature Display
Star Plus meters allow the temperature to be viewed on individual measurement
lines in addition to the temperature display on the top, left of the screen.

1.	 In the measurement mode, press           to choose the middle display line. The
    arrow icon will point to the selected line.

2.	 Press      /     to change the value on the selected line. The middle line can
   be changed to display conductivity (µS/cm or mS/cm), total dissolved solids
   (mg/L), salinity (ppt), resistivity (MΩ-cm), temperature (no icon) or a blank line.

Conductivity Temperature Calibration
The temperature calibration mode of the Star Plus meter allows the temperature on
each measurement line to be manually adjusted.

1.	 In the measurement mode, press       to choose the middle measurement line
    and press     /    until the temperature is shown for the selected line.

2.	 Press      to begin the calibration.

3.	 When the reading stabilizes, the arrow icon and the first digit will flash. Enter
    the temperature by pressing       /     to adjust each digit and        to move to
    the next digit.

4.	 Press      to save and end the calibration.




EN-42                    Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Chapter XI            ISE Technique
ISE Setup Menu
Note: Refer to the Setup Menu section for the Setup Menu Table, which contains
a complete list of meter setup options and descriptions. Refer to Appendix A for a
description of the special setup menu features.

1.	 In the measurement mode, press             .

2.	 Press         /       to scroll through the setup menu until      is displayed on the
    top line.

3.	 Press         to accept the selection and move the arrow icon to the middle line.

4.	 Press       /     to scroll through       for the ISE measurement resolution,
            for the ISE measurement units,            for the ISE calibration range and
            for the non-linear blank correction feature.

5.	 Press         to accept the selection and move the arrow icon to the bottom line.

6.	 To scroll through a list of options on the bottom line, press    /     until the
    desired option is displayed. To enter a numeric value for an option on the bottom
    line, press       /       to adjust each digit and     to move to the next digit.

7.	 Press         to accept the selection and move the arrow icon to the top line.

8.	 Repeat steps 3 through 7 to change another ISE setting or press            to return
    to the measurement mode.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                     EN-43
ISE Technique


ISE Calibration
The calibration standards should be prepared in the same ISE units as the desired
sample results. Start the calibration with the lowest concentration calibration
standard and work up to the highest concentration calibration standard. Any
reagents, such as ionic strength adjustors, should be added to samples and
standards as specified in the electrode user guide.

1.	 Prepare the electrode, standards and any other required solutions for use
    according to the electrode user guide.

2.	 In the measurement mode, press        until the arrow icon points to the top line,
    press     until the ISE icon is shown and press       to begin the calibration.

3.	 Rinse the electrode with distilled or deionized water, shake any excess water off,
    blot it dry and place the electrode into the least concentrated standard.

4.	 Wait for ISE icon to stop flashing. Press      until the first digit to be changed
    is flashing, press       /   to change the value of the flashing digit and
    continue to change the digits until the meter displays the concentration value of
    the standard. Once the standard value is set, press         until the decimal point
    is in the correct location.

5.	 Press      to proceed to the next lowest calibration standard and repeat
    steps 3 and 4, working from the lowest concentration standard to the highest
    concentration standard, or press     to save and end the calibration.

6.	 The actual electrode slope, in mV per decade concentration, will be displayed in
    the main field and       will be displayed in the lower field.

	   a.	 a one point calibration, press
       For                                      and       /    to edit the slope. To
       change the sign of the slope to negative or positive, press        until none of
       the digits are blinking and the arrow icon is blinking and press       to change
       the sign of the slope. Press        to return to the measurement mode.
	   b.	 or a two or more point calibration, the meter will automatically proceed to
        F
        the measurement mode after the slope is displayed.



EN-44                   Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
ISE Technique


ISE Measurement
Note: Turn on the automatic datalog feature to send measurements to the meter
datalog at the frequency specified in each measurement mode. Refer to the Data
Archiving and Retrieval section for details. If the automatic datalog feature is off,
connect the meter to a printer or computer to record the measurements.

1.	 Rinse the electrode with distilled or deionized water. Shake off any excess
    water and blot the electrode dry with lint-free tissue.

2.	 Place the electrode into the sample.

	   a.	 the meter is in the continuous measurement mode, it will start reading
       If
       immediately and continuously update the display. The ISE icon will flash
       until the reading is stable. Once the reading is stable, log and print the
       measurement by pressing           . If a benchtop meter is used and the stirrer
       is enabled, press       to start the stirrer. Press    again to turn off the
       stirrer before removing the electrode and stirrer from the sample.
	   b.	 the meter is in the AUTO-READ measurement mode, press
       If                                                                   to start
       the reading. The AR icon will flash until the reading is stable. Once the
       reading is stable, the meter will log and print the measurement and freeze
       the display. If a benchtop meter is used and the stirrer is enabled, the stirrer
       will turn on when       is pressed and turn off when the reading is stable.
	   c.	 the meter is in the timed measurement mode, it will start reading
       If
       immediately and continuously update the display. The meter will log and
       print the measurement at the frequency specified in the setup menu. If a
       benchtop meter is used and the stirrer is enabled, press        to start the
       stirrer. Press     again to turn off the stirrer before removing the electrode
       and stirrer from the sample.

3.	 Remove the electrode from the sample, rinse it with distilled or deionized water,
    blot it dry, place it in the next sample and repeat step 2.

4. 	 Once all of the samples have been measured, rinse the electrode with distilled
     or deionized water and blot it dry. Consult the electrode user guide for proper
     storage techniques.


Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide               EN-45
ISE Technique


ISE Temperature Display and Calibration
ISE Temperature Display
Star Plus meters allow the temperature to be viewed on individual measurement
lines in addition to the temperature display on the top, left of the screen.

To view the temperature for the ISE measurement line:

1.	 Press and hold       until the meter displays the measurement mode.

2.	 Press       to choose the top display line. The arrow icon will point to the
    selected line.

3.	 Press      /     to change the value on the selected line. The top line can
    be changed to display pH (pH), millivolts (mV), relative millivolts (RmV),
    concentration (ISE), temperature (no icon) or a blank line.

ISE Temperature Calibration
The temperature calibration mode of the Star Plus meter allows the temperature on
each measurement line to be manually adjusted.

To calibrate the temperature for the ISE measurement line:

1.	 In the measurement mode, press       to choose the top measurement line and
    press     /    until the temperature is shown for the selected line.

2.	 Press       to begin the calibration.

3.	 When the reading stabilizes, the arrow icon and the first digit will flash. Enter
    the temperature by pressing       /     to adjust each digit and        to move to
    the next digit.

4.	 Press       to save and end the calibration.




EN-46                    Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Chapter XII           Data Archiving and Retrieval
Datalog and Calibration Log
All 3-Star Plus and the 4-Star Plus pH/ISE meters have a 1000 point datalog. The
4-Star Plus pH/conductivity and pH/dissolved oxygen meters have a 750 point
datalog. The 5-Star Plus meters have a 500 point datalog. The 2-Star meter has a
50 point datalog and all Orion Star meters have a 200 point datalog.

The Star Plus meter printouts have been enhanced to include additional information.
The pH and ISE calibration printouts now include the average slope, the slope
between points and the Eo per point. The polarographic dissolved oxygen printouts
now include slope, membrane temperature and solution temperature.

Automatic Datalog Feature
The 3-Star, 4-Star and 5-Star Plus meters have been enhanced to include an
automatic datalog feature that can be turned on or off. To enable the automatic
recording of data into the datalog:

1.	 In the measurement mode, press            .

2.	 Press       /     until       is displayed on top line.

3.	 Press     to accept the selection and move the arrow icon to the middle line
    and press     /     until        is displayed.

4.	 Press     to accept the selection and move the arrow icon to the bottom line
    and press     /     until    is displayed.

5.	 Press       to accept the selection and move the arrow icon to the top line.

6.	 Press       to save the setup option and return to measurement mode.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                EN-47
Data Archival and Retrieval


Datalog Deletion Setting
The datalog deletion setting determines if the meter will automatically delete
the datalog after it is downloaded to a printer or computer and if the meter will
overwrite the datalog points when the datalog is full. If the datalog deletion setting
is set to      , the meter will automatically delete the datalog after the datalog is
downloaded to a printer or computer. The meter will also display the
error message when all 200 datalog points are filled and the datalog must be
downloaded to a printer or computer to clear the error message. If the datalog
deletion setting is set to     , the meter will overwrite the oldest datalog point when
all 200 datalog points are filled and will not delete the datalog after the datalog is
downloaded to a printer or computer.

1.	 In the measurement mode, press           .

2.	 Press      /      until      is displayed on top line.

3.	 Press     to accept the selection and move the arrow icon to the middle line
    and press     /     until      is displayed.

4.	 Press     to accept the selection and move the arrow icon to the bottom line
    and press     /     until      or     is displayed.

5.	 Press      to accept the selection and move the arrow icon to the top line.

6.	 Press      to save the setup option and return to measurement mode.

Note: If the datalog is not required, set the datalog deletion setting to         to
prevent the error 038 (datalog full) message.




EN-48                    Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Viewing and Printing the Datalog and Calibration Log
The 3-Star, 4-Star and 5-Star Plus meters have been enhanced to include a
calibration log view feature in addition to the datalog view, datalog print and
calibration log print features. The Orion Star meters include the datalog view,
datalog print and calibration log print features.

To view the datalog or calibration log:

1.	 In the measurement mode, press            .

2.	 Press       /     to scroll through         to view the datalog or         to view
    the calibration log (Star Plus meters only).

3.	 Press      . The meter will display the date/time screen. The log number will be
    on the top of the screen and the time, date and year the log was recorded will
    be on the top, middle and bottom display lines respectively. Press     /      to
    scroll through the log.

4.	 Press     . The meter will display the data or calibration point associated with
    the selected date/time screen.

	   a.	 Press       to print the individual data point.

	   b.	 Press       /     to scroll through the log.

	   c.	 Press       to return to the date/time screen.

5.	 To exit the log view mode, press         until the meter displays the date/time
    screen and press      .

To send the datalog or calibration log to a printer or computer:

1.	 Connect the meter to a printer or computer and verify the meter baud rate and
    output settings in the setup menu.

2.	 In the measurement mode, press            .



Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                   EN-49
3.	 Press       /     to scroll through       to print the datalog or        to print
    the calibration log. The Orion Star meters will display         instead of        .

4.	 Press      to send the selected data to the printer or computer.

To interfacing the meter with a computer:

The Orion Star and Star Plus meters can send measurement and calibration data to
a computer in a comma delimited format that is easy to parse in computer programs
like Excel. Select the       ,       ,       output setting in the setup menu.

To send data from the meter to a computer using HyperTerminal:

1.	 Connect the meter to a computer port using the computer interface cable,
    Cat. No. 1010053.

2.	 Click on the start button on the lower left side of the computer screen. Select
    All Programs, Accessories, Communications and HyperTerminal.

3.	 When the HyperTerminal window opens, enter a file name, select an icon for the
    connection and click on the OK button.

4.	 When a new window opens, go to the Connect Using drop-down menu, select
    the COM port that the meter is connected to and click on the OK button.

5.	 A window will open with the COM port properties listed. Select the following
    settings from the drop-down menus and then click on the OK button.

	   Bits per second: 9600
	   Data bits: 8
	   Parity: None
	   Stop bits: 1
	   Flow control: Hardware

6.	 Send data from the meter to HyperTerminal.



EN-50                   Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Chapter XIII           Declaration of Conformity
Manufacturer:	             Thermo Fisher Scientific Inc.

Address:	                  1
                            66 Cummings Center
                           Beverly, MA 01915
                           USA

We declare that the following products described below conform to the Directive
and Standard listed below:

Product(s):  eters for measuring pH, conductivity, dissolved oxygen, and/or ISE
            M
            Benchtop models are rated 100 to 240 VAC, 50/60 Hz, 0.5 A
            Handheld models use four non-rechargeable AA batteries

Benchtop Meters                              Portable Meters
5-Star Plus pH/ISE/Conductivity/DO Meter     5-Star Plus pH/ISE/Conductivity/DO Meter
4-Star Plus pH/Conductivity Meter            5-Star Plus pH/Conductivity/DO Meter
4-Star Plus pH/DO                            4-Star Plus pH/Conductivity Meter
4-Star Plus pH/ISE Meter                     4-Star Plus pH/DO Meter
3-Star Plus Conductivity Meter               4-Star Plus pH/ISE Meter
3-Star Plus DO Meter                         3-Star Plus Conductivity Meter
3-Star Plus pH Meter                         3-Star Plus pH Meter
2-Star pH Meter                              3-Star Plus DO Meter
                                             5-Star Plus RDO® Optical DO/pH/Conductivity Meter
                                             4-Star Plus RDO Optical DO/Conductivity Meter
                                             3-Star Plus RDO Optical DO Meter

Equipment Class:  easurement, control and laboratory
                 M
                 Benchtop models are EMC Class A
                 Portable models are EMC Class D




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                     EN-51
Declaration of Conformity


Directive(s) and Standard(s):

•	 89/336/EEC – Electromagnetic Compatibility (EMC Directive)
	   •	 61326:1997 + A1:1998 + A2:2001 – Electrical equipment for
       EN
       measurement, control, and laboratory use – EMC requirements

•	 73/23/EEC – Low Voltage Directive (LVD)

	   •	 61010-1:2001 – Safety requirements for electrical equipment for
       EN
       measurement, control, and laboratory use – general requirements

Manufacturer’s Authorized Representative: 	                        Date:




Patrick Chiu	                                                      June 6, 2006
Senior Quality Assurance Engineering,
Regulatory Compliance




WEEE Compliance
	This product is required to comply with the European Union’s Waste
  Electrical  Electronic Equipment (WEEE) Directive 2002/96/EC.
  It is marked with the following symbol:

We have contracted with one or more recycling/disposal companies in each EU
Member State and this product should be disposed of or recycled through them.
Further information on compliance with these Directives, the recyclers in your
country, and information on Thermo Scientific Orion products which may assist the
detection of substances subject to the RoHS Directive are available at
www.thermo.com/WEEERoHS.




EN-52                       Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Chapter XIV            Troubleshooting
Meter Self Test
1.	 Disconnect all of the electrodes and probes from the meter and cover all of the
    meter inputs with the black caps.

2.	 Power on the meter, wait until the software revision is displayed and press         .

3.	 All the segments on the display will turn on. Visually inspect the display
    segments to verify that all of the segments are lit and press     .

4.	 All the segments on the display will turn off. Visually inspect the display
    segments to verify that all of the segments are not lit and press      .

5.	 The display will read      . Press every key on the keypad one at a time in
    any order. If the keys are not pressed within five seconds of one another, the
    display will read            , which indicates a key failure. Press   to clear
    the error 033 message and complete the self test. If all the keys are pressed
    and functioning, the meter will restart and proceed to the measurement mode.

Note: If the meter reads                during the self test, ensure that all of the
electrodes are disconnected from the meter, all of the meter inputs are covered
with the black caps and the BNC shorting cap is firmly attached to the BNC meter
input. This error code usually occurs if the BNC shorting cap is missing or not fully
connected to the BNC meter input during the meter self test.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide               EN-53
Troubleshooting


Meter Error Codes
•	 If the reading on the screen is flashing       , the value is out of range.
   Perform the meter self test, clean the electrode according to the electrode user
   guide and re-calibrate the electrode with new standards.

•	 If the     icon is lit and the reading is flashing, the sensor needs to be
   calibrated according to the user’s set calibration interval or the pH slope is
   outside the range of 85 % to 115%.

•	 Press       to clear an error code. Error codes show      on the middle line and
   a set of three alphanumeric characters on the bottom line. Some of these codes
   are errors, some are warnings and some are purely informational.

 Error Code	   Description              Troubleshooting
 002, 026,     Hardware or Memory       Press     to clear the error. If the error occurs again, contact
 E##, F##      Error                    Technical Support.
 005           Value Outside            Press     and re-enter the value. Check meter specifications for the
               Allowable Range          allowable range of values.
 033           Keypad Failure           Repeat the self test. When the meter reads        , press all the keys,
                                        including the power key, within five seconds of one another.
 034           BNC Input Failure        Disconnect all the electrodes from the meter, connect the BNC shorting cap
                                        to the meter and repeat the self test.
 038           Datalog Full             Download the datalog to a printer or computer, turn the automatic datalog
                                        feature off in the setup menu (Star Plus meters only) or change the datalog
                                        setting to       ,     ,    in the setup menu so the meter deletes the
                                        datalog points when the datalog is full.
 D##           Remote Control Error     Check the programming instructions to verify the correct commands, names
                                        and values.
 107           pH Calibration           The millivolts measured during calibration are the same for two buffers.
               Standard Error           Review the calibration procedure and verify that the electrode was placed
                                        in the buffers at the appropriate time. Clean the electrode according to the
                                        electrode user guide. Re-calibrate the electrode with fresh buffers.
 109           Bad pH Slope or          Clean the electrode according to the electrode user guide. Re-calibrate the
               Calibration Offset       electrode with new buffers.
 200           Autosampler              The meter is unable to send a signal to the autosampler. Make sure that
               Interface Error          the autosampler is properly connected to the meter.
 201           Autosampler              The autosampler is unable to receive a signal from the meter. Review the
               Signal Error             meter setup parameters and make sure that the baud rate of the meter is
                                        set to 1200.


EN-54                            Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Troubleshooting




 Error Code	   Description           Troubleshooting
 202           Autosampler is        Turn the autosampler off and wait 45 seconds before turning it back on.
               Jammed                The autosampler should return to the home position.
 203           Unstable Reading      The measurements taken using the autosampler are unstable. Check the
               from Autosampler      electrodes for proper function. Make sure that the electrode cables are
                                     properly connected.
 306           ISE Automatic         Disable the automatic blank feature in the setup menu and re-calibrate
               Blank Error           the meter without using a zero concentration standard.
 307           ISE Calibration       The millivolts measured during calibration are the same for two
               Standard Error        standards. Review the calibration procedure and verify that the electrode
                                     was placed in the standards at the appropriate times. Clean the electrode
                                     according to the electrode user guide. Re-calibrate the electrode with
                                     fresh standards.
 309           Bad ISE Slope         Clean the electrode according to the electrode user guide. Re-calibrate
                                     the electrode with freshly prepared standards.
 707           Conductivity          The conductivity value measured during calibration is the same for
               Calibration           two standards. Review the calibration procedure and verify that the
               Standard Error        conductivity probe was placed in the standards at the appropriate times.
                                     Clean the conductivity probe according to the probe user guide. Re-
                                     calibrate the probe with new standards.
 709           Conductivity Cell     The cell constant is not in the range of 0.001 to 199.0 cm-1. Clean the
               Constant Error        conductivity probe according to the probe user guide. Re-calibrate the
                                     probe with new standards.
 808           Bad Zero Point        An air calibration should be performed before the zero point calibration.
               DO Slope              Verify that a solution with zero oxygen is being used for the zero point
                                     calibration. A solution with 15 grams of Na2SO3 dissolved in 250 mL of
                                     distilled water is recommended.
 809           Bad DO Slope          For polarographic DO probes, connect the probe to the meter, power on
                                     the meter and let the probe to polarize for at least 30 minutes. For an
                                     air calibration, check that the sponge in the calibration sleeve is damp
                                     and there is no water on the probe membrane . For a water calibration,
                                     bubble air into the sample and stir to keep bubbles off the membrane.
                                     Clean the DO probe according to the probe user guide. Re-calibrate the
                                     DO probe.
 880           RDO® Optical DO       Verify that the RDO sensor is properly connected to the meter. Connect
               Sensor Not Attached   another, known working RDO sensor to the meter.
 881           RDO Optical DO        The RDO cap attached to the RDO sensor has expired. Install a new RDO
               Sensor Expired        cap according to the RDO sensor user guide.
 882           RDO Optical DO        Verify that the RDO cap on the RDO sensor was properly installed, not
               Sensor Failure        expired and not tampered with. Connect another, known working RDO
                                     sensor to the meter.


Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                                       EN-55
Troubleshooting


General Troubleshooting
 Problem:	    The display freezes and the measurement values will not change.
 Solution:	The meter is in the AUTO-READ measurement mode (the AR icon will
            appear in the top, right corner of the display). Press to start a new
            reading or select another measurement mode in the setup menu.

 Problem:	    When I press       the meter displays          .
 Solution:	The meter is printing and cannot enter the calibration mode until the
            printing is done. This should rarely occur if the meter is set to a 9600
            baud rate. If the meter is at a lower baud rate, the delay will be longer.
 Problem:	The meter did not accept the change I made in the setup menu.
 Solution:	After making a change in the setup menu, press         until the arrow
            icon points to the top line (confirms the change) and then press       to
            save the change and return to the measurement mode.
 Problem:	    How do I abort a calibration?
 Solution:	Press and hold  to abort any meter operation and return to the
            measurement mode.
 Problem:	    The printout is a string of numbers and units with commas.
 Solution:	The output format in the setup menu is set to the computer output or
            the printer baud rate is set incorrectly in the setup menu. Change the
            output format to the printer output in the setup menu. Change the
            baud rate to the correct value for the printer that is being used.
 Problem:	    When I press the stirrer button, the stirrer doesn’t work.
 Solution:	The current stirrer setting is off. Set the speed to 1 through 7 in the
            setup menu.
 Problem:	The timed measurement time entry screen does not appear in the
           setup menu.
 Solution:	The meter is in the AUTO-READ or continuous mode. When the meter
            is set to the timed mode, the next setup screen will be for time entry.
 Problem:	 cannot tell if I have the Star Plus meter or the Orion Star meter.
           I
 Solution:	   When the meter is powered on, the Star Plus meters with enhanced
              features will display                with the meter revision number
              (       or similar) and proceed to the measurement mode.


EN-56                    Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Troubleshooting



 pH Troubleshooting
 Problem:	    The meter does not recognize the pH buffer value during calibration.
 Solution:	   Verify that the correct buffer set was selected in the setup menu. The
              meter uses the raw mV reading of the electrode to recognize a buffer
              during calibration. As the electrode ages or becomes dirty, its mV
              readings will drift and you will need to manually enter the pH buffer value
              when calibrating.
 ISE Troubleshooting
 Problem:	    It takes several minutes for the readings to stabilize during a calibration.
 Solution:	   The concentration range in the setup menu is set to low. Change the
              concentration range to high.
              The ISE resolution is set to 3 digits in the setup menu. Change the ISE
              resolution to 2 digits for faster stabilization of the readings.
 Problem:	    When I use the automatic blank correction setting and calibrate an ISE,
              the meter gives a slope that is too low or cannot be manually checked.
 Solution:	   Turn the automatic blank correction setting off in the setup menu.
 Conductivity Troubleshooting
 Problem:	    The meter does not recognize the conductivity standard during
              calibration.
 Solution:	   Verify that the default cell constant was entered in the setup menu. The
              cell constant is usually printed on the conductivity probe cable. Verify
              that the conductivity standard is one that is programmed into the meter.
              Re-calibrate with a fresh standard.
 Problem:	The temperature coefficient value does not appear in the setup menu.
 Solution:	The current temperature compensation setting is nonlinear or off.
            Change the temperature compensation to linear and the next screen will
            be the temperature coefficient value entry screen.
 Problem:	    The measurement is out of range when it should be in range.
 Solution:	   Check that the conductivity probe is fully immersed in the solution.
              Verify that the cell constant is correct for the conductivity probe that is
              connected to the meter. Verify that the cell type selected in the setup
              menu is set to Std.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                   EN-57
Troubleshooting



 Dissolved Oxygen Troubleshooting
 Problem:	
          The manual barometric pressure entry screen does not appear in the
          setup menu.
 Solution:	
           The barometric pressure compensation is set to automatic in the setup
           menu. Change the barometric pressure compensation to manual and
           the next screen will be the manual pressure entry screen.
 Problem:	
          The manual salinity factor entry screen does not appear in the
          setup menu.
 Solution:	
           The salinity correction is set to automatic in the setup menu. Change
           the salinity correction to manual and the next screen will be the salinity
           factor entry screen.
 Problem:	
          The AUTO-STIR BOD probe does not turn on when the button on the
          probe is pressed.
 Solution:	
           The read type must be set to AUTO-READ in the setup menu and the
           stirrer speed must be set from 1 to 7 to initiate a measurement and
           start stirring by pressing the button on the AUTO-STIR BOD probe.
 RDO® Optical Dissolved Oxygen Troubleshooting
 Problem:	
          The meter displays an error 881 message and will not take a dissolved
          oxygen measurement.
 Solution:	
           Turn the meter off, replace the old RDO cap with a new cap and turn
           the meter on. This should clear the error message.

Assistance
After troubleshooting all components of your measurement system, contact
Technical Support. Within the United States call 1.800.225.1480 and outside the
United States call 978.232.6000 or fax 978.232.6031. In Europe, the Middle East
and Africa, contact your local authorized dealer. For the most current contact
information, visit www.thermo.com/contactwater.

For the latest application and technical resources for Thermo Scientific Orion
products, visit www.thermo.com/waterapps.

Warranty
For the most current warranty information, visit www.thermo.com/water.


EN-58                   Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Chapter XV                 Meter Specifications
Meter Specifications
Environmental Operating Conditions

 Portable and Benchtop Meter Environmental Operating Conditions

 Operating Ambient Temperature                           5 to 45 °C
 Operating Relative Humidity                             5 to 85 %, non-condensing
 Storage Temperature                                     -20 to +60 °C
 Storage Relative Humidity                               5 to 85 %, non-condensing
 Pollution                                               Degree 2
 Overvoltage                                             Category II
 Altitude                                                Up to 2000 meters
 Weight                                                  Portable: 0.45 kg
                                                         Benchtop: 0.91 kg
 Size                                                    Portable: 4.8 cm (H) x 9.7 cm (W) x 21.3 cm (D)
                                                         Benchtop: 9.4 cm (H) x 17.0 cm (W) x 22.4 cm (D)	
 AC Powered Meters                                       Indoor use only
 Battery Operated Meters                                 Indoor or outdoor use
 Regulatory and Safety                                   CE, CSA, TÜV, UL, FCC Class limits*
 Case Material                                           ABS
 Shock and Vibration                                     Vibration, shipping/handling per ISTA #1A
                                                         Shock, drop test in packaging per ISTA #1A
 Enclosure (designed to meet)                            IP67 (portable meter)
                                                         IP54 (benchtop meter)

* TÜV and UL certifications are pending for all Star Plus RDO® optical dissolved oxygen meters.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                                   EN-59
Meter Specifications



 Universal Power Adapter Environmental Operating Conditions

 Operating Ambient Temperature                        0 to 50 °C
 Operating Relative Humidity                          0 to 90 %, non-condensing
 Storage Temperature                                  -20 to +75 °C
 Storage Relative Humidity                            0 to 90 %, non-condensing
 Pollution                                            Degree 2
 Overvoltage                                          Category II
 Operating Altitude                                   Up to 2000 meters
 Benchtop Meters                                      Indoor use only


Meter Parameter Specifications
The following meter parameter specifications are for the complete line of Orion Star
and Star Plus series meters. Single parameter, dual parameter and some multi-
parameter meters will not include all of the parameters listed in this section.

 pH
 Range                             -2.000 to 19.999
 Resolution                        0.1, 0.01, 0.001
 Relative Accuracy                 ± 0.002
 Calibration Points                1 to 5
 pH (2-Star pH Meter Only)
 Range                             0.000 to 14.999
 Resolution                        0.1, 0.01, 0.001
 Relative Accuracy                 ± 0.002
 Calibration Points                1 to 3
 Millivolts, Relative Millivolts, ORP
 Range                             ± 1999.9 mV
 Resolution                        0.1 mV
 Relative Accuracy                 ± 0.2 mV or 0.05 % of reading, whichever is greater



EN-60                          Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Meter Specifications


 ISE
 Range                         0 to 19999
 Resolution                    1 to 3 significant figures
 Relative Accuracy             ± 0.2 mV or 0.05 %, whichever is greater
 Displayed Units               M, mg/L, %, ppb or no units
 Calibration Features          Linear point to point, selectable non-linear automatic blank
                               correction and low concentration range stability
 Dissolved Oxygen (Polarographic)
 Range                         0.00 to 90.0 mg/L
                               0.0 to 600 %
 Resolution                    0.1, 0.01 mg/L
                               0.1, 1 %
 Relative Accuracy             ± 0.2 mg/L
                               ±2%
 Salinity Factor               0 to 45 ppt
 Barometric Pressure           450 to 850 mm Hg
 Calibration Types             Water saturated air, air saturated water, manual (Winkler), zero point
 Probe Type                    Polarographic
 RDO® Optical Dissolved Oxygen
 Range                         0.00 to 20.0 mg/L
                               0.0 to 200 %
 Resolution                    0.1, 0.01 mg/L
                               0.1, 1 %
 Relative Accuracy             ± 0.1 mg/L up to 8 mg/L; ± 0.2 mg/L from 8 mg/L to 20 mg/L
                               ±2%
 Salinity Factor               0 to 45 ppt
 Barometric Pressure           450 to 850 mm Hg
 Calibration Types             Water saturated air, air saturated water, manual (Winkler), zero point
 Probe Type                    RDO optical




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                            EN-61
Meter Specifications



 Conductivity
 Range                            0.000 to 3000 mS/cm, auto-resolution with cell constant dependence
 Resolution                       4 significant figures down to 0.001 µS/cm, cell constant dependant
 Relative Accuracy                0.5 % ± 1 digit or 0.01 µS/cm, whichever is greater
 Cell Constant                    0.001 to 199.9 cm-1
 Reference Temperature            5 °C, 10 °C, 15 °C, 20 °C or 25 °C
 Resistivity Range                0.0001 to 100 Megohm
 Resistivity Resolution           Automatic
 Resistivity Relative Accuracy    0.5 % ± 1 digit
 Salinity Range                   0.1 to 80.0 ppt NaCl equivalent, 0.1 to 42 ppt practical salinity
 Salinity Resolution              0.1 ppt
 Salinity Relative Accuracy       0.1 ± 1 digit
 TDS Range                        0 to 19999 mg/L
 TDS Resolution                   1 mg/L
 TDS Relative Accuracy            0.5 % ± 1 digit
 Temperature
 Range                            -5 to 105 °C
 Resolution                       0.1 up to 99.9 °C, 1.0 over 99.9 °C
 Relative Accuracy                ± 0.1 °C
 Temperature* (RDO® Optical Dissolved Oxygen Meter Only)
 Range                            0 to 50 °C
 Resolution                       0.1 °C
 Relative Accuracy                ± 0.3 °C

*  his temperature specification is only for the Star Plus RDO optical dissolved
  T
  oxygen meter when it is used with the RDO optical dissolved oxygen sensor.
  The 4-Star and 5-Star RDO meters will have the standard temperature
  specifications when used with an electrode other than the RDO sensor.




Note: Specifications are subject to change without notice.



EN-62                         Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Ordering Information
 Cat. No.      Description
 1111000      2-Star pH benchtop meter with universal power adapter and user guide
 1112000      3-Star Plus pH benchtop meter with universal power adapter and user guide
 1212000      3-Star Plus pH portable meter with batteries and user guide
 1113000      3-Star Plus DO benchtop meter with universal power adapter and user guide
 1213000      3-Star Plus DO portable meter with batteries and user guide
 1114000      3-Star Plus conductivity benchtop meter with universal power adapter and user guide
 1214000      3-Star Plus conductivity portable meter with batteries and user guide
 1115000      4-Star Plus pH/ISE benchtop meter with universal power adapter and user guide
 1215000      4-Star Plus pH/ISE portable meter with batteries and user guide
 1116000      4-Star Plus pH/DO benchtop meter with universal power adapter and user guide
 1216000      4-Star Plus pH/DO portable meter with batteries and user guide
 1117000      4-Star Plus pH/conductivity benchtop meter with universal power adapter and
              user guide
 1217000      4-Star Plus pH/conductivity portable meter with batteries and user guide
 1218000      5-Star Plus pH/DO/conductivity portable meter with batteries and user guide
 1119000      5-Star Plus pH/ISE/DO/conductivity benchtop meter with universal power adapter and
              user guide
 1219000      5-Star Plus pH/ISE/DO/conductivity portable meter with batteries and user guide
 1213300      3-Star Plus RDO® optical DO portable meter with batteries and user guide
 1213310      4-Star Plus RDO optical DO/conductivity portable meter with batteries and
              user guide
 1213320      5-Star Plus RDO optical DO/pH/conductivity portable meter with batteries and
              user guide




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                             EN-63
Cat. No.          Description
 090043            Swing arm electrode stand
 1010003           Universal power adapter
 1010006           Star series printer with RS232 printer interface cable (Cat. No. 250302-001)
 1010053           RS232 computer interface cable
 096019            Stirrer probe with paddle, for 3-Star, 4-Star and 5-Star benchtop meters
 8102BNUWP         ROSS Ultra combination pH electrode with glass body
 8165BNWP          ROSS Sure-Flow combination pH electrode with epoxy body
 8172BNWP          ROSS Sure-Flow combination pH electrode with glass body
 9107BNMD          Gel-filled pH/ATC Triode with epoxy body
 9157BNMD          Refillable pH/ATC Triode with epoxy body
 9165BNWP          Sure-Flow combination pH electrode with epoxy body
 9172BNWP          Sure-Flow combination pH electrode with glass body
 927005MD          ATC probe with epoxy body
 927007MD          ATC probe with stainless steel body
 9512HPBNWP        High performance ammonia combination ion selective electrode
 9609BNWP          Fluoride combination ion selective electrode
 9707BNWP          Nitrate combination ion selective electrode
 8611BNWP          ROSS sodium combination ion selective electrode
 083005MD          Polarographic DO probe with calibration sleeve and 1.5 meter cable
 083010MD          Polarographic DO probe with calibration sleeve and 3 meter cable
 086030MD          Polarographic BOD AUTO-STIR DO probe with calibration sleeve
 087010MD          RDO® optical DO sensor with stainless steel guard, RDO cap and 3 meter cable
 087020MD          RDO optical DO sensor with stainless steel guard, RDO cap and 6 meter cable
 011050MD          Conductivity probe with 1 µS/cm to 20 mS/cm range and 1.5 meter cable
 013005MD          DuraProbe conductivity probe with 1 µS/cm to 200 mS/cm range and 1.5 meter cable
 013010MD          DuraProbe conductivity probe with 1 µS/cm to 200 mS/cm range and 3 meter cable
 013016MD          Conductivity probe with 0.01 µS/cm to 300 µS/cm range and 1.5 meter cable
Visit www.thermo.com/water for additional meter kits, accessories, electrodes and solutions.

EN-64                        Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Appendix A             Meter Setup Menu Features
pH Setup Menu Features
Automatic buffer recognition
The Orion Star and Star Plus pH meters are capable of automatically recognizing pH
1.68, 4.01, 6.86, 7.00, 9.18, 10.01 and 12.46 buffers during a pH calibration. During
a calibration, the meter uses the selected buffer set and the raw mV reading of the
pH electrode in the buffer to recognize and display the buffer value at the measured
temperature. The raw mV value must be about ± 30 mV from the theoretical mV
reading of the buffer in order for the meter to automatically recognize the buffer.

 Buffer       mV Range          Buffer       mV Range          Buffer   mV Range
 1.68         +285 to +345      7.00         - 30 to + 30      10.01    -207 to -147
 4.01         +207 to +147      9.18         -99 to -159       12.46    -293 to -353
 6.86         +38 to -22


Dissolved Oxygen Setup Menu Features
Barometric Pressure Compensation
The Orion Star and Star Plus dissolved oxygen meters have an internal barometer
that is used for pressure compensated dissolved oxygen readings. The meter can
also use manual barometric pressure compensation if dissolved oxygen is measured
with a submerged probe or in a pressurized vessel. The pressure must be entered
as mm Hg. 1 mm Hg = 0.03937 inch Hg = 1.3332 hPa (mBar) = 0.01934 PSI.

Salinity Correction
Automatic salinity correction for dissolved oxygen readings is available on Orion
Star and Star Plus dissolved oxygen meters that have a conductivity measurement
mode. The meter uses the conductivity value read by the conductivity probe to
calculate the salinity correction factor and applies the factor to dissolved oxygen
readings reported as mg/L.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                    EN-65
Meter Setup Menu Features


The meter can also use manual salinity correction for dissolved oxygen readings
reported as mg/L. The manual salinity correction factor must be entered as ppt
(parts per thousand).

   Conductivity     Salinity            Conductivity      Salinity            Conductivity     Salinity
   at 20 °C         Correction          at 20 °C          Correction          at 20 °C         Correction
   (mS/cm)          Value (ppt)         (mS/cm)           Value (ppt)         (mS/cm)          Value (ppt)
   5                3                   20                13                  35               25
   6                4                   21                14                  36               25
   7                4                   22                15                  37               26
   8                5                   23                15                  38               27
   9                6                   24                16                  39               28
   10               6                   25                17                  40               29
   11               7                   26                18                  42               30
   12               8                   27                18                  44               32
   13               8                   28                19                  46               33
   14               9                   29                20                  48               35
   15               10                  30                21                  50               37
   16               10                  31                22                  52               38
   17               11                  32                22                  54               40
   18               12                  33                23                  56               42
   19               13                  34                24
This table was calculated from the International Oceanographic Tables, Vol. 1, National Institute of Oceanography
of Great Britain, Womley, Godaming, Surrey, England and Unesco, Paris 1971.



Conductivity Setup Menu Features
Temperature Compensation and Reference Temperature
The Orion Star and Star Plus conductivity meters have the ability to use a
temperature compensation feature that calculates and displays the conductivity
measurements at a reference temperature of 5 °C, 10 °C, 15 °C, 20 °C or 25 °C
(Orion Star meters have a reference temperature of 15 °C, 20 °C or 25 °C only).
The temperature compensation can be set as linear for most aqueous samples,
non-linear for ultra pure and low ionic strength samples or off for non-temperature
compensated conductivity measurements.


EN-66                          Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Meter Setup Menu Features


The closer the sample temperature is to the selected reference temperature, the
more accurate the conductivity measurement will be, especially if the temperature
compensation coefficient is estimated or inaccurate.

The conductivity of a solution with a specific electrolyte concentration changes with
temperature and this relationship is described by the temperature coefficient of the
solution. The meter has a default temperature coefficient of 2.1 percent change in
conductivity per ˚C, which is representative of many aqueous samples.

 Solution (25 °C to 50 °C)                     Temperature Coefficient (% / °C)
 Ultra Pure Water                              4.55
 Salt (NaCl)                                   2.12
 5% NaOH                                       1.72
 Dilute Ammonia                                1.88
 10% HCl                                       1.32
 5% Sulfuric Acid                              0.96
 98% Sulfuric Acid                             2.84
 Sugar Syrup                                   5.64

Total Dissolved Solids (TDS)
The Orion Star and Star Plus conductivity meters measure TDS as the total amount
of dissolved inorganics in a solution. The dissolved inorganics carry a current that is
measured by the conductivity probe. Since there is a direct relationship between
conductivity and TDS, conductivity readings are used to estimate the presence of
inorganics. The user must enter a TDS factor between 0.01 and 10 mg/L in the
setup menu.

The standard method of determining TDS involves evaporating a sample to dryness
at 180 ˚C and weighing the residue. The TDS factor is calculated by taking the
residue weight and dividing it by the sample conductivity. Subsequent conductivity
readings are multiplied by the TDS factor to determine the TDS value of the sample.




Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide                 EN-67
Meter Setup Menu Features


Automatic Calibration
The Orion Star and Star Plus conductivity meters are capable of automatically
recognizing 100 µS/cm, 1413 µS/cm and 12.9 mS/cm conductivity standards when
the nominal cell constant of the conductivity probe is entered in the setup menu.
For the meter to recognize the conductivity standard, the entered cell constant must
be accurate within a factor of 3. For example, if the actual cell constant is 1.0 cm-1,
entering a nominal cell constant in the range of 0.3 cm-1 to 3.0 cm-1 would allow the
meter to identify the conductivity standard and perform the automatic calibration.


ISE Setup Menu Features
Concentration Range
The Orion Star and Star Plus ISE meters can be set for a high or low ISE
concentration range that is used to determine the calibration stability criteria. If a
high ISE concentration range is selected, the meter will perform a normal calibration
with no delay in displaying the calibration standard value. If a low ISE concentration
range is selected, the meter will wait about three to five minutes before displaying
a stable reading for the calibration standard values. The delay will depend on the
species being measured and the concentration of the calibration standards. The low
ISE concentration range is designed to improve the accuracy of low concentration
measurements by allowing the electrode to have a longer amount of time to
stabilize in the calibration standards.

Automatic Blank Correction
The Orion Star and Star Plus ISE meters have an ISE automatic blank correction
feature that uses an algorithm to compensate for the non-linearity of the ion
selective electrode in low-level standards and samples. Since the automatic blank
correction feature requires the use of a set of non-linear equations that can only
be calculated numerically, the user cannot analytically verify the calibration and the
average slope value that is displayed on the meter may be outside of the slope
range that is specified in the electrode user guide. In applications were analytical
verification is required, the automatic blank correction feature should be turned off.




EN-68                   Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
Environmental Instruments
Water Analysis Instruments
North America
166 Cummings Center
Beverly, MA 01915 USA
Toll Free: 1-800-225-1480
Tel: 1-978-232-6000
Dom. Fax: 1-978-232-6015
Int’l Fax: 978-232-6031

Europe
P.O. Box 254, 3860 AG Nijkerk
Wallerstraat 125K, 3862 BN Nijkerk,
The Netherlands
Tel: (31) 033-2463887
Fax: (31) 033-2460832

Asia Pacific
Blk 55, Ayer Rajah Crescent
#04-16/24, Singapore 139949
Tel: 65-6778-6876
Fax: 65-6773-0836

www.thermo.com/water
© 2008 Thermo Fisher Scientific Inc.
All rights reserved.




257250-001 Rev A 09-08

More Related Content

PPTX
C H N S ANALYSIS.pptx
PPTX
HPLC method development and data analysis
PPTX
pH measurement
PDF
Sop for ph meter
PDF
Sulfur Analysis by GC-SCD using Shimadzu’s SCD-2030 for ASTM D5504, D5623, an...
PDF
Hplc column
PDF
Sop water analysis-2
PPTX
Conductivity measurement
C H N S ANALYSIS.pptx
HPLC method development and data analysis
pH measurement
Sop for ph meter
Sulfur Analysis by GC-SCD using Shimadzu’s SCD-2030 for ASTM D5504, D5623, an...
Hplc column
Sop water analysis-2
Conductivity measurement

What's hot (18)

PPTX
TESTING OF NATURAL PRODUCTS AND DRUGS- DHANASHREE GIRIDHAR KOLHEKAR.pptx
PPT
Complexomety in analytical study
PPTX
Analytical balance
PDF
Gas chromatography, an introduction.pdf
PPTX
Conventional methods of quantitative analysis
PPTX
Karl Fischer Titration
PDF
Expiry date , calculation of yields, production record review, change control
PPTX
IVT Presentation Batch vs Continuous - 45min_REV3
PPTX
Basic of titration
PPT
Chapter 1
PPTX
Sana 13 nmr
PPTX
LC NMR HYPHENATED TECHNIQUES
PDF
Coulometry Manik
PPTX
Analytical procedures life cycle management
PDF
Ft nmr spectrometer 2
DOCX
Volumetric estimation of nitrite in the given solution of sodium nitrite usin...
PPT
Intro to titrations
TESTING OF NATURAL PRODUCTS AND DRUGS- DHANASHREE GIRIDHAR KOLHEKAR.pptx
Complexomety in analytical study
Analytical balance
Gas chromatography, an introduction.pdf
Conventional methods of quantitative analysis
Karl Fischer Titration
Expiry date , calculation of yields, production record review, change control
IVT Presentation Batch vs Continuous - 45min_REV3
Basic of titration
Chapter 1
Sana 13 nmr
LC NMR HYPHENATED TECHNIQUES
Coulometry Manik
Analytical procedures life cycle management
Ft nmr spectrometer 2
Volumetric estimation of nitrite in the given solution of sodium nitrite usin...
Intro to titrations
Ad

Similar to Orion star series_meter_users_guide (20)

PDF
[08 10-2014 10.14.03]orion-star_series_meter_users_guide
PDF
Jerome 431 manual
PDF
Coulter manual de usos
PDF
Oxymat 61 o2 analyzer
PDF
Vortex vn2000 manual badger meter flow meters hot tap insertion meter
PDF
pH.ORP Transmitter-Analog sensors-Digital sensors
PDF
Dynasonics dfx manual badger meter doppler ultrasonic flow meter
PDF
Dynasonics tfx 500w manual badger meter-clamp-on transit time ultrasonic flow...
PDF
GENESYS™ 10S UV-Vis (English)
PDF
Manual ohmart us
PDF
Guided Level Radar-Levelflex M FMP40
PDF
Promass80 flow meter
PDF
Waterpilot fmx21 endress+hauser datasheet-level transmitter
PDF
Hydrostatic Level Measurement-Deltapilot S DB50/50L/51/52/53
PDF
Hydrostatic level measurement-Deltapilot M FMB50/51/52/53
PDF
Waterpilot fmx167 endress+hauser datasheet-hydrostatic level measurement
PDF
Level limit switch compact design-Liquiphant T FTL20H
PDF
FLUKE 92B, 96B, 99B & 105B User.pdf
PDF
High accuracy liquid level measurement-Proservo NMS5
PDF
Drager Fabius GS Technical Service Manual Rev L (1).pdf
[08 10-2014 10.14.03]orion-star_series_meter_users_guide
Jerome 431 manual
Coulter manual de usos
Oxymat 61 o2 analyzer
Vortex vn2000 manual badger meter flow meters hot tap insertion meter
pH.ORP Transmitter-Analog sensors-Digital sensors
Dynasonics dfx manual badger meter doppler ultrasonic flow meter
Dynasonics tfx 500w manual badger meter-clamp-on transit time ultrasonic flow...
GENESYS™ 10S UV-Vis (English)
Manual ohmart us
Guided Level Radar-Levelflex M FMP40
Promass80 flow meter
Waterpilot fmx21 endress+hauser datasheet-level transmitter
Hydrostatic Level Measurement-Deltapilot S DB50/50L/51/52/53
Hydrostatic level measurement-Deltapilot M FMB50/51/52/53
Waterpilot fmx167 endress+hauser datasheet-hydrostatic level measurement
Level limit switch compact design-Liquiphant T FTL20H
FLUKE 92B, 96B, 99B & 105B User.pdf
High accuracy liquid level measurement-Proservo NMS5
Drager Fabius GS Technical Service Manual Rev L (1).pdf
Ad

Recently uploaded (20)

PDF
Network Security Unit 5.pdf for BCA BBA.
PDF
Getting Started with Data Integration: FME Form 101
PDF
Building Integrated photovoltaic BIPV_UPV.pdf
PDF
Video forgery: An extensive analysis of inter-and intra-frame manipulation al...
PDF
Build a system with the filesystem maintained by OSTree @ COSCUP 2025
PDF
gpt5_lecture_notes_comprehensive_20250812015547.pdf
PPTX
Spectroscopy.pptx food analysis technology
PPTX
Big Data Technologies - Introduction.pptx
PDF
7 ChatGPT Prompts to Help You Define Your Ideal Customer Profile.pdf
PPTX
SOPHOS-XG Firewall Administrator PPT.pptx
PDF
NewMind AI Weekly Chronicles - August'25-Week II
PDF
Per capita expenditure prediction using model stacking based on satellite ima...
PDF
Approach and Philosophy of On baking technology
PDF
MIND Revenue Release Quarter 2 2025 Press Release
PPTX
Tartificialntelligence_presentation.pptx
PPTX
KOM of Painting work and Equipment Insulation REV00 update 25-dec.pptx
PPTX
A Presentation on Artificial Intelligence
PDF
Dropbox Q2 2025 Financial Results & Investor Presentation
PDF
Reach Out and Touch Someone: Haptics and Empathic Computing
PDF
Profit Center Accounting in SAP S/4HANA, S4F28 Col11
Network Security Unit 5.pdf for BCA BBA.
Getting Started with Data Integration: FME Form 101
Building Integrated photovoltaic BIPV_UPV.pdf
Video forgery: An extensive analysis of inter-and intra-frame manipulation al...
Build a system with the filesystem maintained by OSTree @ COSCUP 2025
gpt5_lecture_notes_comprehensive_20250812015547.pdf
Spectroscopy.pptx food analysis technology
Big Data Technologies - Introduction.pptx
7 ChatGPT Prompts to Help You Define Your Ideal Customer Profile.pdf
SOPHOS-XG Firewall Administrator PPT.pptx
NewMind AI Weekly Chronicles - August'25-Week II
Per capita expenditure prediction using model stacking based on satellite ima...
Approach and Philosophy of On baking technology
MIND Revenue Release Quarter 2 2025 Press Release
Tartificialntelligence_presentation.pptx
KOM of Painting work and Equipment Insulation REV00 update 25-dec.pptx
A Presentation on Artificial Intelligence
Dropbox Q2 2025 Financial Results & Investor Presentation
Reach Out and Touch Someone: Haptics and Empathic Computing
Profit Center Accounting in SAP S/4HANA, S4F28 Col11

Orion star series_meter_users_guide

  • 1. Orion Star™ and Star Plus Meter User Guide • English • Español • Français • Deutsch • Italiano
  • 2. ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher. 1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoE Z, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of Thermo Fisher. Guaranteed Success and The Technical Edge are service marks of Thermo Fisher. PerpHecT meters are protected by U.S. patent 6,168,707. PerpHecT ROSS are protected by U.S. patent 6,168,707. ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results solutions are protected by US Patent 5,830,338. ROSS Ultra electrodes are protected by US patents 6,793,787. Orion ORP Standard is protected by US Patent 6,350,367. Orion NoCal electrodes are protected by US Patent 7,276,142. © 2008 Thermo Fisher Scientific Inc. All rights reserved. RDO is a registered trademark of In-Situ Inc., Fort Collins, CO USA. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. The specifications, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice. This publication supersedes all previous publications on this subject.
  • 3. EN ENGLISH SECTION Table of Contents Chapter I Introduction Meter Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-2 Chapter II Display General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-3 Chapter III Keypad General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-5 Key Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .EN-6 Chapter IV Preparation Installing the Power Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-7 Installing the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-8 Connecting the Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-9 Turning on the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-10 Meter Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-10 Chapter V Meter Setup Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-11 Setup Menu Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-12 General Menu Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-14 Time and Date Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-15 AUTO-READ™, Continuous or Timed Measurement Settings . . . . . . . . . . . . . . . EN-16 Selecting the Measurement Parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-17 Method Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-18 Chapter VI pH Technique pH Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-19 pH Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-20 pH Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-21 pH Temperature Display and Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-22 Chapter VII mV, Relative mV and ORP Technique Relative mV and ORP Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .EN-23 mV, Relative mV and ORP Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-24 Chapter VIII Dissolved Oxygen Technique Dissolved Oxygen Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-25 Dissolved Oxygen Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-26 Dissolved Oxygen Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-28 Dissolved Oxygen Temperature Display and Calibration . . . . . . . . . . . . . . . . . . . EN-29 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-I
  • 4. EN Chapter IX RDO® Optical Dissolved Oxygen Technique RDO Cap Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-31 RDO Optical Dissolved Oxygen Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-32 Setup Menu for RDO Optical Dissolved Oxygen Sensor . . . . . . . . . . . . . . . . .EN-33 RDO Optical Dissolved Oxygen Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-35 RDO Optical Dissolved Oxygen Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . EN-37 RDO Optical Dissolved Oxygen Temperature Display and Calibration . . . . . . . . . EN-38 Chapter X Conductivity Technique Conductivity Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-39 Conductivity Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-40 Conductivity Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-41 Conductivity Temperature Display and Calibration . . . . . . . . . . . . . . . . . . . . . . . . EN-42 Chapter XI ISE Technique ISE Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-43 ISE Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-44 ISE Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-45 ISE Temperature Display and Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-46 Chapter XII Data Archiving and Retrieval Datalog and Calibration Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-47 Automatic Datalog Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-47 Datalog Deletion Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-48 Viewing and Printing the Datalog and Calibration Log . . . . . . . . . . . . . . . . . . . EN-49 Chapter XIII Declaration of Conformity Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-51 WEEE Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-52 Chapter XIV Troubleshooting Meter Self Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-53 Meter Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-54 General Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-56 Chapter XV Meter Specifications Meter Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-59 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-63 Appendix A Meter Setup Menu Features pH Setup Menu Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-65 Dissolved Oxygen Setup Menu Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-65 Conductivity Setup Menu Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-66 ISE Setup Menu Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EN-68 EN-II Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 5. Chapter I Introduction Congratulations! You have selected an industry-leading Thermo Scientific Orion Star or Star Plus series meter that is designed for electrochemistry measurements in the field or in the laboratory. • 2-Star meters provide the single parameter measurement of pH. • 3-Star Plus meters provide the single parameter measurement of pH, dissolved oxygen, RDO® optical dissolved oxygen or conductivity. • 4-Star Plus meters provide the dual parameter measurements of pH/dissolved oxygen, pH/conductivity, pH/ISE (ion selective electrode) or RDO optical dissolved oxygen/conductivity. • 5-Star Plus meters provide the multi-parameter measurements of pH/ISE/dissolved oxygen/conductivity, pH/dissolved oxygen/conductivity or pH/RDO optical dissolved oxygen/conductivity. All meters include a temperature measurement function. All meters with pH measurement capability include a mV/relative mV/ORP function. Built to meet the demands of busy, multiple user laboratory or plant environments, all Orion Star and Star Plus series meters are microprocessor controlled, which aids in the delivery of accurate and precise measurements. The waterproof portable meters can even withstand submersion for short periods of time. The 3-Star Plus, 4-Star Plus and 5-Star Plus meters have been enhanced to include an increased number of datalog points, improved temperature displays and a new temperature calibration mode for each measurement parameter. Benchtop 3-Star Plus, 4-Star Plus and 5-Star Plus meters include autosampler capability and are compatible with the AutoTration™-500 autosampler. Refer to the AutoTration-500 user guide for information on operating the Star Plus meters with the autosampler. Note: Please read this user guide thoroughly before using your benchtop or portable meter. Any use outside of these instructions may invalidate your warranty and cause permanent damage to the meter. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-1
  • 6. Introduction Meter Features To better meet the needs of users in environmental protection and control, food and beverage, pharmaceutical and consumer product laboratories, the Orion Star and Star Plus series meters include these key features: • Password Protected Methods – The meter will save up to ten custom measurements and calibrations for future reference. Password protection of each method eliminates any tampering with methods as multiple users access only the procedure most appropriate to their work. • AUTO-READ™ – The meter takes a measurement and automatically prints or logs data when the reading becomes stable. The measurement is frozen on the display until the user prompts the meter to take a new measurement. • Stirrer Control – Benchtop meters (3-Star and higher) have a stirrer control for the stirrer probe, Cat. No. 096019, and the AUTO-STIR™ BOD probe, which eliminates the need for additional stir plates and stir bars. • SMART STABILITY™ and SMART AVERAGING™ – Remove the guesswork by automatically compensating for measurement conditions and optimizing the meter response time. • Display Backlight – All 3-Star, 4-Star and 5-Star meters include a display backlight feature. When the meter is on, a quick press of will turn the backlight on and off. When the meter is operating on battery power, the backlight will automatically turn off after two minutes to conserve power. When batteries are low, the backlight will no longer turn on. • Automatic Shutoff – The meters will shut down after 20 minutes without a keypress. This maximizes battery power on portable meters and benchtop meters that are being run on battery power. • Audible Signals – The meter will beep whenever a key is pressed, providing immediate verification that the user’s input was received. An easy-to-use reference guide, attached to each meter, supports daily meter use. EN-2 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 7. Chapter II Display General Description Throughout a given process, the display on an Orion Star or Star Plus series meter provides temperature and calibration data. The temperature appears in the left, top corner of the display. The icon indicates that a calibration mode or calibration setup menu is active. The man, 2, 4, 7, 9, 10, and 12 icons indicate which pH buffers were saved after a pH calibration is performed. The setup icon only appears when the meter is in setup mode. The icon indicates an error condition and when it is displayed with the icon, a calibration alarm or sensor quality issue exists. The icon indicates that the AUTO-READ measurement mode is active and is discussed in greater detail in the Meter Setup section. Temperature Calibration pH/ISE/ORP Unit Conductivity Unit DO Unit 5 Star Meter This is the display of the 5-Star meter capable of multi-parameter measurements. The single and dual parameter meters will have fewer measurement lines, depending on the meter capabilities. Note: In the measurement mode, the three main lines of data on the meter display correspond to what is being measured. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-3
  • 8. Display Measurement Unit Icons In the measurement mode, the arrow icon on the left side of the display screen indicates the active line. Press to move the arrow icon to the desired line and press / to scroll through the measurement unit icons associated with the selected line. The measurement unit icons for the 5-Star multi-parameter meter are shown below. The single and dual parameter meters will have fewer measurement lines and icons, depending on the meter capabilities. A p H, mV, relative mV, ISE, or temperature B C onductivity, TDS, salinity, resistivity, or temperature C D issolved oxygen as % saturation, dissolved oxygen as concentration, barometric pressure or temperature The units of measurement, which are displayed on the right side of the screen, will flash until the reading is stable. Note: If a measurement line is not needed, press to move the arrow icon to the measurement line that is not needed and press / until the measurement line is completely blank. EN-4 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 9. Chapter III Keypad General Description The keypad layout is the same for all Orion Star and Star Plus series meters. The portable and 2-Star benchtop meters have nine keys. The 3-Star, 4-Star and 5-Star benchtop meters have 10 keys due to the addition of the stir key – . Portable Keypad Calibrate Measure Save/Print Power Backlight Scroll Up Digits Line Select Scroll Down Setup View Log Benchtop Keypad Power Backlight Measure Calibrate Save/Print Stirrer On/Off, Scroll Up (not on 2-Star) Digits Line Select Setup View Log Scroll Down Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-5
  • 10. Keypad Key Definitions Key Description Key Description Turns the meter on, if the meter Changes the measurement is off. units of the selected line in the Toggles the backlight on and off, measurement mode. if the meter is on (3-Star, 4-Star Changes the value on the and 5-Star meters only). selected line in the setup, If the meter is on, hold down methods and log view modes. the key for about three seconds Edits the value of the flashing to turn off the meter. digit for setup, password entry and calibration modes. Scrolls the arrow icon on left of Selects the next digit to edit screen among the three display and moves the decimal point lines, so the selected line can for setup, password entry and be edited or calibrated. calibration modes. Starts the calibration for Prints and logs a measurement the selected line in the in the continuous or timed measurement mode. measurement modes. If the arrow icon points to the Prints, logs and freezes the top line and the displayed units display when the reading are pH, pressing the key will becomes stable in the AUTO- start a pH calibration. READ measurement mode. Each time the key is pressed in Exits the setup menu and returns the calibration mode, the meter to measurement mode. will accept the calibration point Accepts the calibration and and move to the next point returns to measurement mode. until the maximum number of calibration points are reached. Enters the setup menu, starting Enters the log view and with selected line in the download menu. measurement mode. If the arrow icon points to the Turns the stirrer on and off. top line and the displayed units are ISE, pressing the key will enter the ISE setup screen. EN-6 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 11. Chapter IV Preparation Installing the Power Adapter The universal power adapter that is included with your benchtop meter is the only power adapter recommended for use with this unit. The use of any other power adapter will void your meter warranty. The external electrical power adapter is rated to be operated at 100 to 240 VAC, 0.5 A, 50/60 Hz. Based on your wall outlet, select one of the four plug plates provided (110 V, 220 V, 240V) and slide it into the grooves on the adapter. A click will be heard when the plug is properly in place. Connect the output plug of the power adapter to the power input on the benchtop meter. Refer to the diagram in the Connecting the Electrodes section. Batteries can be installed in the benchtop Orion Star or Star Plus series meters, so the meter setup settings are protected if the meter is disconnected from the wall outlet or a brief power outage occurs. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-7
  • 12. Preparation Installing the Batteries Orion Star and Star Plus series meters use four AA alkaline batteries. Do not use lithium or rechargeable batteries. Improper installation of non-alkaline batteries could create a hazard. Note: For benchtop meters, the installation of batteries is not required if the unit will always be connected to a power source via the universal power supply. For portable meters, the batteries are supplied from the factory. To access the battery compartment in portable meters, loosen the two screws in the back of the meter. 1. Confirm that the meter is off and gently place the meter upside down on a clean, lint-free cloth to prevent scratching the LCD. 2. Remove the battery case cover. 3. Insert new batteries with the + side orientation as depicted in the battery compartment housing. 4. Replace the battery case cover. 5. Stored data, calibrations and methods will remain in the meter’s nonvolatile memory when the batteries are being replaced. However, the date and time may need to be reset when the batteries are changed. EN-8 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 13. Preparation Connecting the Electrodes Follow the diagrams below to correctly connect electrodes and probes to the meter. The multi-parameter meter is depicted; single parameter and dual parameter meters will have fewer connections, depending on the meter measurement capabilities. Benchtop Meter Electrode Connections Reference, 8 pin miniDIN, RS232 Power pH, ORP Cond or ATC Stirrer jack, or ISE stirrer or BOD BNC, 9 pin Ground AUTO-STIR probe pH, ORP miniDIN, or ISE DO Portable Meter Electrode Connections 8 pin miniDIN, RS232 Cond or ATC BNC, 9 pin pH, ORP miniDIN, or ISE DO Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-9
  • 14. Preparation Meter Connections with Multiple Functions • Use the BNC input to connect pH, ISE and ORP electrodes with a BNC or waterproof BNC connector. • Benchtop meters have a reference input that is used to connect a separate reference electrode. Reference electrodes require an separate, appropriate sensing electrode for measurements. • The 970899WP dissolved oxygen probe can be used on the BNC input. • Use the 8 pin miniDIN input for conductivity probes or for automatic temperature compensation (ATC) probes. • The DO AUTO-STIR probe uses the 9 pin miniDIN input and the stirrer jack. • Connect a printer or computer to the RS232 input using the appropriate cable. Turning on the Instrument With the batteries installed in the portable meters and the power adapter attached or the batteries installed in the benchtop meter, press to turn on the meter. If using a 3-Star, 4-Star or 5-Star meter, press when the meter is powered on to toggle the backlight on and off. When the benchtop meter is drawing line power, the backlight will stay on until is pressed. To turn off the meter, press and hold for about three seconds. Meter Maintenance For routine meter maintenance, dust and wipe the meter with a damp cloth. If necessary, a warm water or a mild water-based detergent can be used. Perform meter maintenance on a daily, weekly or monthly basis, as required by the environment in which the meter is operated. Immediately remove any spilled substance from the meter using the proper cleaning procedure for the type of spill. EN-10 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 15. Chapter V Meter Setup Setup Menu To navigate the setup menu: 1. Press to enter the setup menu. 2. Press / until the desired setup option is displayed on the top line. 3. Press to move the arrow icon to the middle line. 4. Press / until the desired setup option is displayed on the middle line. 5. Press to move the arrow icon to the bottom line. 6. To scroll through a list of options on the bottom line, press / until the desired option is displayed. To enter a numeric value for an option on the bottom line, press / to adjust each digit and to move to the next digit. For example, to change the pH measurement resolution press to scroll from 0.01 to 0.001 on the bottom display line. 7 Press . to move the arrow icon to the top line. 8. Repeat steps 2 through 7 to program a new setup option or press to exit the setup menu and return to the measurement mode. Note: Refer to Appendix A for a description of the special setup menu features. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-11
  • 16. Meter Setup Setup Menu Table – The following table is for the complete line of Orion Star and Star Plus meters. Meters may not include all of the options listed in this table. Top Line Middle Line Bottom Line Setup Menu Description (default setting, method specific) PH rES 0.1, 0.01, 0.001 pH measurement resolution (0.01, yes) PH bUF USA, EUr0 pH buffer set for automatic buffer recognition during calibration, USA buffers are 1.68, 4.01, 7.00, 10.01, 12.46 and EUrO buffers are 1.68, 4.01, 6.86, 9.18 (USA, yes) ISE rES 1, 2, 3 ISE measurement resolution in significant figures (1, yes) ISE UnIt m, mgL, PEr, PPb, n0nE ISE measurement units (PPb, yes) ISE rAng L0w, HIgH ISE concentration range for calibration stability criteria (HIgH, yes) ISE nLIn AUt0, 0FF ISE automatic blank correction for low-level calibration (AUt0, yes) C0nd tC 0FF, LIn, nLF Conductivity temperature compensation type, LIn is for linear, nLF is for non-linear pure water samples (LIn, yes) C0nd C0EF 0.0 to 10.0 Conductivity temperature compensation coefficient in % change in conductivity per °C, appears if LIn was selected for tC (2.1, yes) C0nd tdSF 0.00 to 10.0 Conductivity TDS factor value (0.49, yes) C0nd CELL 0.001 to 199.0 Conductivity default cell constant value for automatic conductivity calibration mode (0.475, yes) C0nd trEF 5, 10, 15, 20, 25 Conductivity reference temperature (25, yes) C0nd tyPE Std, 1, 2, 3, 4, 5, 6, 7, USP Conductivity cell type and selectable range (Std, yes) See the Setup Menu for RDO Optical Dissolved Oxygen Sensor section for details on the RDO sensor information menus. d0 rES 0.1, 1 % sat DO % saturation measurement resolution (0.1, yes) d0 rES 0.01, 0.1 mg/L DO mg/L measurement resolution (0.01, yes) d0 bAr AUt0, mAn DO barometric pressure compensation type (AUt0, yes) d0 PrES 450.0 to 850.0 DO manual barometric pressure compensation value, appears if mAn was selected for bAr (760.0, yes) d0 SAL AUt0, mAn DO salinity correction type (AUt0, yes) d0 SALF 0 to 45 DO manual salinity correction value, appears if mAn was selected for SAL or a DO meter without a conductivity mode is used (0, yes) d0 CALt AIr, H20, mAn, SEt0 DO calibration type (AIr, yes) d0 LIFE 365 to 0 RDO cap replacement countdown in days – 3, 4 and 5 Star Plus RDO meters only (Set by RDO sensor, no) dUE PH 0 to 9999 pH calibration alarm value in hours, 0 is off (0, yes) dUE 0rP 0 to 9999 ORP calibration alarm value in hours, 0 is off (0, yes) dUE ISE 0 to 9999 ISE calibration alarm value in hours, 0 is off (0, yes) dUE C0nd 0 to 9999 Conductivity calibration alarm value in hours, 0 is off (0, yes) dUE d0 0 to 9999 DO calibration alarm value in hours, 0 is off (0, yes) EN-12 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 17. Meter Setup Top Line Middle Line Bottom Line Setup Menu Description (default setting, method specific) rEAd tyPE AUt0, tImE, C0nt, Measurement read type as AUTO-READ, timed or continuous (AUt0, yes) rEAd tInE 00:05 to 99:59 Timed measurement value in minutes and seconds (01:00, yes) L0g dEL n0, YES Delete datalog after download option (n0, yes) L0g AUt0 0FF, 0n Automatic datalog point saving option (0FF, yes) gEn dEgC -5.0 to 105 Manual temperature value (25.0, yes) gEn StIr 0FF, 1, 2, 3, 4, 5, 6, 7 Stirrer speed – 3, 4 and 5 Star benchtop meters only (4, yes) gEn PASS 0000 to 9999 Meter password entry (0000, yes) gEn AUt0 0n, 0FF Automatic meter shutoff option (0n, no) gLP SEt 0FF, 0n GLP option, GLP feature enables or disables methods (0FF, no) dAtE H0Ur HH00 to HH23 Hour setting (HH12, no) dAtE tInE mm00 to mm59 Minute setting (mm00, no) dAtE tYPE mdY, dmY Date format as month, day, year or day, month, year (mdY, no) dAtE YEAr 2000 to 2099 Year setting (2004, no) dAtE dAtE mm01 to mm12 Month setting (mm01, no) dAtE dAY dd01 to dd31 Day of the month setting (dd01, no) r232 bAUd 1200, 2400, 4800, 9600 Baud rate setting (9600, no) r232 0UtF Prnt, C0mP Output format for printer or computer, COmP format is comma delimited (Prnt, no) AUt0 SAPL 0FF, 0n Autosampler – 3, 4 and 5 Star Plus benchtop meters only (OFF, no) AUt0 trAY 28, 48 Beaker tray setting (28, no) AUt0 rInb 1, 2, 3, 4, 5 Number of rinse beakers (3, no) AUt0 rSEC 5 to 60 Rinse time in each rinse beaker, seconds (10, no) AUt0 PH 0, 1, 2, 3 pH calibration points (3, no) AUt0 0rP n0, YES ORP calibration option, appears if 0 was selected for PH (n0, no) AUt0 ISE 0, 2, 3 ISE calibration points, appears if 0 was selected for PH (2, no) 1 AUt0 ISE1 Concentration value of ISE standard 1, appears if 2 or 3 was selected for the ISE calibration points (1, no) 10 AUt0 ISE2 Concentration value of ISE standard 1, appears if 2 or 3 was selected for the ISE calibration points (10, no) 100 AUt0 ISE3 Concentration value of ISE standard 1, appears if 3 was selected for the ISE calibration points (100, no) AUt0 C0nd 0, 1, 2, 3 Conductivity calibration points (3, no) AUt0 n0SA 0 to 47 Number of sample beakers (1, no) Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-13
  • 18. Meter Setup General Menu Settings • Manual Temperature controls temperature compensation when no temperature sensor is attached to the meter. • Stirrer Speed sets the stirrer speed from 1 (slowest) through 7 (fastest) and off (3-Star, 4-Star and 5-Star benchtop meters only). • Password Protection protects setup menu options and methods from being accidentally erased or tampered with (3-Star, 4-Star and 5-Star meters only). • Automatic Shutoff controls whether the instrument will automatically turn off after 20 minutes without a keypress. 1. In the measurement mode, press . 2. Press / to scroll through the setup menu until is displayed on the top line. 3. Press to accept the selection and move the arrow icon to the middle line. 4. Press / to scroll through for the manual temperature setting, for the stirrer speed setting, for password entry and for the automatic shutoff setting. 5. Press to accept the selection and move the arrow icon to the bottom line. 6. To scroll through a list of options on the bottom line, press / until the desired option is displayed. To enter a numeric value for an option on the bottom line, press / to adjust each digit and to move to the next digit. 7. Press to accept the selection and move the arrow icon to the top line. 8. Repeat steps 3 through 7 to change another general setting or press to return to the measurement mode. EN-14 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 19. Meter Setup Time and Date Settings • The date and time settings are saved with the data and calibration log points and are included with the data that is sent to a computer or printer. • The date format can be set to read month, day, year or day, month, year according to the user’s preference. 1. In the measurement mode, press . 2. Press / to scroll through the setup menu until is displayed on the top line. 3. Press to accept the selection and move the arrow icon to the middle line. 4. Press / to scroll through for the current hour setting, for the current minute setting, for the date format setting, for the current month setting, for the current day setting and for the current year setting. 5. Press to accept the selection and move the arrow icon to the bottom line. 6. To scroll through a list of options on the bottom line, press / until the desired option is displayed. To enter a numeric value for an option on the bottom line, press / to adjust each digit and to move to the next digit. 7. Press to accept the selection and move the arrow icon to the top line. 8. Repeat steps 3 through 7 to change another time and date setting or press to return to the measurement mode. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-15
  • 20. Meter Setup AUTO-READ™, Continuous or Timed Measurement Settings • In the AUTO-READ mode, the meter starts taking a measurement when is pressed. Once the measurement is stable, the display freezes and the data is logged and printed. The AUTO-READ mode also controls the stirrer. The stirrer starts when is pressed and stops when the measurement becomes stable. • In the continuous mode, the meter is constantly taking measurements and updating the display. Press to log and print a measurement in this mode. • In the timed mode, the meter is constantly taking measurements and updating the display. The meter logs and prints the measurement at the selected time interval. Timed dissolved oxygen measurements with the RDO sensor are taken only at the selected time interval, which conserves the meter battery power. 1. In the measurement mode, press . 2. Press / to scroll through the setup menu until is displayed on the top line. 3. Press to accept the selection and move the arrow icon to the middle line. 4. Press / to scroll through for the measurement read type and for the timed reading interval. 5. Press to accept the selection and move the arrow icon to the bottom line. 6. To scroll through a list of options on the bottom line, press / until the desired option is displayed. To enter a numeric value for an option on the bottom line, press / to adjust each digit and to move to the next digit. 7. Press to accept the selection and move the arrow icon to the top line. 8. Repeat steps 3 through 7 to change another measurement setting or press to return to the measurement mode. EN-16 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 21. Meter Setup Selecting the Measurement Parameter In the measurement mode, the arrow icon on the left side of the display indicates the active line. Press to move the arrow icon to the desired measurement line and press / to scroll through the measurement parameters associated with the selected line. The measurement lines and icons for the 5-Star multi-parameter meter are shown below. The single and dual parameter meters will have fewer measurement lines and icons, depending on the meter capabilities. pH mV RmV ISE No icon for temperature No icon and no measurement – the measurement line is turned off μS/cm or mS/cm for conductivity mg/L for TDS ppt for salinity MΩ-cm for resistivity No icon for temperature No icon and no measurement – the measurement line is turned off % sat for dissolved oxygen percent saturation mg/L for dissolved oxygen concentration No icon for barometric pressure No icon for sample temperature No icon for membrane temperature (polarographic DO readings only) No icon and no measurement – the measurement line is turned off Note: If a measurement line is not needed, press to move the arrow icon to the measurement line that is not needed and press / until the measurement line is completely blank. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-17
  • 22. Meter Setup Method Setup The Orion 3-Star, 4-Star and 5-Star meters can save up to 10 methods when the GLP function is enabled. When a method is selected, the meter will use the last calibration performed in that method, so electrodes that share a common meter connection can be more easily interchanged. When using multiple methods, a calibration must be performed for each method that will be used. 1. To enable the GLP function: a. In the measurement mode, press . b. Press / until is displayed on top line. c. Press to move the arrow icon to the middle line and press / until is displayed. d. Press to move the arrow icon to the bottom line and press / until is displayed. e. Press to move the arrow icon to the top line. f. Press to exit the setup menu and return to the measurement mode. 2. To display and change the current method number: a. the measurement mode, press In . The current method number will be displayed. b. Press / to select a new method number. c. Press to save the method number and press to return to the measurement mode. EN-18 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 23. Chapter VI pH Technique pH Setup Menu Note: Refer to the Setup Menu section for the Setup Menu Table, which contains a complete list of meter setup options and descriptions. Refer to Appendix A for a description of the special setup menu features. 1. In the measurement mode, press . 2. Press / to scroll through the setup menu until is displayed on the top line. 3. Press to accept the selection and move the arrow icon to the middle line. 4. Press / to scroll through for pH measurement resolution and for the automatic buffer recognition setting. 5. Press to accept the selection and move the arrow icon to the bottom line. 6. To scroll through a list of options on the bottom line, press / until the desired option is displayed. 7 Press . to accept the selection and move the arrow icon to the top line. 8. Repeat steps 3 through 7 to change another pH setting or press to return to the measurement mode. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-19
  • 24. pH Technique pH Calibration 1. Prepare the electrode according to the electrode user guide. 2. In the setup mode, select the buffer set ( or ) that will be used for the automatic buffer recognition feature. 3. In the measurement mode, press until the arrow icon points to the top line, press until the pH icon is shown and press to begin the calibration. 4. Rinse the electrode, and ATC probe if being used, with distilled water and place into the buffer. 5. Wait for the pH icon to stop flashing. Automatic buffer recognition – When the pH icon stops flashing the meter a. will display the temperature-corrected pH value for the buffer. Manual calibration – When the pH icon stops flashing the meter will display b. the actual pH value read by the electrode. Press until the first digit to be changed is flashing, press / to change the value of the flashing digit and continue to change the digits until the meter displays the temperature-corrected pH value of the buffer. Once the pH buffer value is set, press until the decimal point is in the correct location. 6. Press to proceed to the next calibration point and repeat steps 4 and 5 or press to save and end the calibration. 7 The actual electrode slope, in percent, will be displayed in the main field and . will be displayed in the lower field. a. a one point calibration, press For and / to edit the slope and press to return to the measurement mode. b. or a two or more point calibration, the meter will automatically proceed to F the measurement mode after the slope is displayed. EN-20 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 25. pH Technique pH Measurement Note: Turn on the automatic datalog feature to send measurements to the meter datalog at the frequency specified in each measurement mode. Refer to the Data Archiving and Retrieval section for details. If the automatic datalog feature is off, connect the meter to a printer or computer to record the measurements. 1. Rinse the electrode with distilled or deionized water. Shake off any excess water and blot the electrode dry with lint-free tissue. 2. Place the electrode into the sample. a. the meter is in the continuous measurement mode, it will start reading If immediately and continuously update the display. The pH icon will flash until the reading is stable. Once the reading is stable, log and print the measurement by pressing . If a benchtop meter is used and the stirrer is enabled, press to start the stirrer. Press again to turn off the stirrer before removing the electrode and stirrer from the sample. b. the meter is in the AUTO-READ measurement mode, press If to start the reading. The AR icon will flash until the reading is stable. Once the reading is stable, the meter will log and print the measurement and freeze the display. If a benchtop meter is used and the stirrer is enabled, the stirrer will turn on when is pressed and turn off when the reading is stable. c. the meter is in the timed measurement mode, it will start reading If immediately and continuously update the display. The meter will log and print the measurement at the frequency specified in the setup menu. If a benchtop meter is used and the stirrer is enabled, press to start the stirrer. Press again to turn off the stirrer before removing the electrode and stirrer from the sample. 3. Remove the electrode from the sample, rinse it with distilled or deionized water, blot it dry, place it in the next sample and repeat step 2. 4. Once all of the samples have been measured, rinse the electrode with distilled or deionized water and blot it dry. Consult the electrode user guide for proper storage techniques. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-21
  • 26. pH Technique pH Temperature Display and Calibration pH Temperature Display Star Plus meters allow the temperature to be viewed on individual measurement lines in addition to the temperature display on the top, left of the screen. To view the temperature for the pH measurement line: 1. In the measurement mode, press to select the top display line. The arrow icon will point to the selected line. 2. Press / to change the value on the selected line. The top line can be changed to display pH (pH), millivolts (mV), relative millivolts (RmV), concentration (ISE), temperature (no icon) or a blank line. pH Temperature Calibration The temperature calibration mode of the Star Plus meter allows the temperature on each measurement line to be manually adjusted. To calibrate the temperature for the pH measurement line: 1. In the measurement mode, press to choose the top measurement line and press / until the temperature is shown for the selected line. 2. Press to begin the calibration. 3. When the reading stabilizes, the arrow icon and the first digit will flash. Enter the temperature by pressing / to adjust each digit and to move to the next digit. 4. Press to save and end the calibration. EN-22 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 27. mV, Relative mV and Chapter VII ORP Technique All meters with pH measurement capability include a mV, relative mV and ORP function. Measure the raw millivolt (mV) values of an electrode in the mV mode. Calibrate the relative millivolt (RmV) values of a redox electrode for oxidation- reduction potential (ORP) measurements in the relative mV and ORP mode. Refer to the pH Technique section for information on the temperature display and temperature calibration of the pH/mV/RmV/ISE measurement line. Note: The mV measurements are raw readings and cannot be calibrated. Use the relative mV mode to calibrate mV measurements. Relative mV and ORP Calibration 1. Prepare the electrode according to the electrode user guide. 2. In the measurement mode, press until the arrow icon points to the top line, press until the RmV icon is shown and press to begin the calibration. 3. Rinse the electrode with distilled water and place it into the standard. 4. Wait for the RmV icon to stop flashing. When the RmV icon stops flashing, the meter will display 000.0 RmV. Press until the first digit to be changed is flashing, press / to change the value of the flashing digit and continue to change the digits until the meter displays the millivolt value of the standard. To change the value to negative or positive number, press until none of the digits are blinking and the arrow icon is blinking and then press to change the sign of the millivolt value. 5. Press to save and end the calibration. 6. The millivolt offset will be displayed and the meter will automatically proceed to the measurement mode. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-23
  • 28. mV, Relative mV and ORP Technique mV, Relative mV and ORP Measurement Note: Turn on the automatic datalog feature to send measurements to the meter datalog at the frequency specified in each measurement mode. Refer to the Data Archiving and Retrieval section for details. If the automatic datalog feature is off, connect the meter to a printer or computer to record the measurements. 1. Rinse the electrode with distilled or deionized water. Shake off any excess water and blot the electrode dry with lint-free tissue. 2. Place the electrode into the sample. a. the meter is in the continuous measurement mode, it will start reading If immediately and continuously update the display. The mV or RmV icon will flash until the reading is stable. Once the reading is stable, log and print the measurement by pressing . If a benchtop meter is used and the stirrer is enabled, press to start the stirrer. Press again to turn off the stirrer before removing the electrode and stirrer from the sample. b. the meter is in the AUTO-READ measurement mode, press If to start the reading. The AR icon will flash until the reading is stable. Once the reading is stable, the meter will log and print the measurement and freeze the display. If a benchtop meter is used and the stirrer is enabled, the stirrer will turn on when is pressed and turn off when the reading is stable. c. the meter is in the timed measurement mode, it will start reading If immediately and continuously update the display. The meter will log and print the measurement at the frequency specified in the setup menu. If a benchtop meter is used and the stirrer is enabled, press to start the stirrer. Press again to turn off the stirrer before removing the electrode and stirrer from the sample. 3. Remove the electrode from the sample, rinse it with distilled or deionized water, blot it dry, place it in the next sample and repeat step 2. 4. Once all of the samples have been measured, rinse the electrode with distilled or deionized water and blot it dry. Consult the electrode user guide for proper storage techniques. EN-24 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 29. Chapter VIII Dissolved Oxygen Technique Dissolved Oxygen Setup Menu Note: Refer to the Setup Menu section for the Setup Menu Table, which contains a complete list of meter setup options and descriptions. Refer to Appendix A for a description of the special setup menu features. 1. In the measurement mode, press . 2. Press / to scroll through the setup menu until is displayed on the top line. 3. Press to accept the selection and move the arrow icon to the middle line. 4. Press / to scroll through for the % saturation resolution, for the mg/L concentration resolution, for the barometer type (automatic or manual), for the manual barometric pressure compensation value, for the salinity compensation type (automatic or manual), for the manual salinity correction value and for the dissolved oxygen calibration type. 5. Press to select the option and move the arrow icon to the bottom line. 6. To scroll through a list of options on the bottom line, press / until the desired option is displayed. To enter a numeric value for an option on the bottom line, press / to adjust each digit and to move to the next digit. 7. Press to accept the selection and move the arrow icon to the top line. 8. Repeat steps 3 through 7 to change another dissolved oxygen setting or press to return to the measurement mode. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-25
  • 30. Dissolved Oxygen Technique Dissolved Oxygen Calibration • Prior to calibration, the dissolved oxygen probe must be prepared and polarized. The probe is continuously polarized when it is connected to the meter. When the probe is first connected or if the probe is disconnected for more than 60 minutes, connect the probe to the meter, connect the meter to a power source and wait 30 to 60 minutes for the probe to polarize. Disconnecting the probe for less than one hour will require 5 to 25 minutes for polarization. • The meters will supply a polarization current to the dissolved oxygen probe even when the meter power is off. To maximize the meter battery life, unplug the probe if it will not be used for an extended period. 1. Select one of the following calibration modes in the setup menu. a. – An air calibration is performed in water saturated air using the calibration sleeve. This is the simplest and most accurate calibration. Due to the inherent differences between water saturated air and air saturated water, 102.3% saturation will be displayed when the calibration reading is stable. i. The highest possible accuracy is reached when calibration temperature is the same as the measuring temperature. ii. oisten the sponge or absorbent cloth in the calibration sleeve with M distilled water and insert the probe into the sleeve without touching the water saturated material. For BOD measurements, this calibration can be performed in a BOD bottle. b. – A water calibration is performed using water that is 100% saturated with air. Bubble air into a water sample and gently stir the sample to prevent the buildup of air bubbles on the dissolved oxygen probe membrane. c. – A manual calibration is performed using a water sample with a known concentration of dissolved oxygen. This method can be used to calibrate the dissolved oxygen probe to the value achieved by a Winkler titration. EN-26 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 31. Dissolved Oxygen Technique i. manual calibration involves performing a Winkler titration and using A that sample as a calibration standard. The oxygen level result from the titration is entered in a manual calibration as the dissolved oxygen value. This correlates the meter input to the Winkler titration. This method is inherently less accurate, due to the possibility of titration errors. d. – A zero point calibration is performed in an oxygen-free solution. A zero point calibration is not generally required unless measurements will be taken below 10% saturation or 1 mg/L. Zero the probe when using a new membrane, using fresh filling solution or when measuring dissolved oxygen levels below 1 mg/L. An air calibration should be performed prior to the zero point calibration. 2. Allow the probe and calibration standard (water saturated air, air saturated water, Winkler standard or oxygen-free solution) to reach equilibrium. 3. In the measurement mode, press until the arrow icon points to the bottom line, press until the % sat or mg/L icon is shown and press to begin the calibration. 4. Wait for the dissolved oxygen reading to stabilize. a. an air calibration is performed, the meter will display 102.3% and If automatically return to the measurement mode. b. a water calibration is performed, the meter will display 100.0% and If automatically return to the measurement mode. c. a manual calibration is performed, wait for the mg/L icon to stop flashing If and enter the dissolved oxygen value by pressing until the first digit to be changed is flashing, press / to change the value of the flashing digit and continue to change the digits until the meter displays the correct dissolved oxygen value. Once the dissolved oxygen value is set, press until the decimal point is in the correct location. d. a zero point calibration is performed, the meter will display 0.00 and If automatically return to the measurement mode. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-27
  • 32. Dissolved Oxygen Technique Dissolved Oxygen Measurement Note: Turn on the automatic datalog feature to send measurements to the meter datalog at the frequency specified in each measurement mode. Refer to the Data Archiving and Retrieval section for details. If the automatic datalog feature is off, connect the meter to a printer or computer to record the measurements. 1. Rinse the dissolved oxygen probe with distilled or deionized water. Shake off any excess water and blot the probe dry with lint-free tissue. 2. Place the dissolved oxygen probe into the sample. a. the meter is in the continuous measurement mode, it will start reading If immediately and continuously update the display. The mg/L or % sat icon will flash until the reading is stable. Once the reading is stable, log and print the measurement by pressing . If a benchtop meter is used and the stirrer is enabled, press to start the stirrer. Press again to turn off the stirrer before removing the probe and stirrer from the sample. b. the meter is in the AUTO-READ measurement mode, press If to start the reading. The AR icon will flash until the reading is stable. Once the reading is stable, the meter will log and print the reading and freeze the display. If a benchtop meter is used and the stirrer is enabled, the stirrer will turn on when is pressed and turn off when the reading is stable. If the BOD AUTO-STIR probe is used, press the button on the probe to start the AUTO-READ measurement. N ote: Benchtop Star Plus dissolved oxygen meters allow the measurement and stirring functions to be controlled by the AUTO-STIR probe when the meter is in the AUTO-READ measurement mode. Press the button on the AUTO-STIR probe to start and stop measurement and stirring functions. c. the meter is in the timed measurement mode, it will start reading If immediately and continuously update the display. The meter will log and print the measurement at the frequency specified in the setup menu. If a benchtop meter is used and the stirrer is enabled, press to start the stirrer. Press again to turn off the stirrer before removing the probe and stirrer from the sample. EN-28 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 33. Dissolved Oxygen Technique 3. Remove the dissolved oxygen probe from the sample, rinse it with distilled or deionized water, blot it dry, place it in the next sample and repeat step 2. 4. Once all of the samples have been measured, rinse the dissolved oxygen probe with distilled or deionized water and blot it dry. Consult the dissolved oxygen probe user guide for proper storage techniques. Dissolved Oxygen Temperature Display and Calibration Dissolved Oxygen Temperature Display Star Plus meters allow the temperature to be viewed on individual measurement lines in addition to the temperature display on the top, left of the screen. To view the temperature for the dissolved oxygen measurement line: 1. In the measurement mode, press to choose the bottom display line. The arrow icon will point to the selected line. 2. Press / to change the value on the selected line. The bottom line can be changed to display dissolved oxygen (% saturation), dissolved oxygen (mg/L), barometric pressure (no icon), sample temperature (no icon), electrolyte solution/membrane temperature (no icon, after number) or a blank line. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-29
  • 34. Dissolved Oxygen Technique Dissolved Oxygen Temperature Calibration The temperature calibration mode of the Star Plus meter allows the temperature on each measurement line to be manually adjusted. The dissolved oxygen measurement line displays the sample temperature and the electrolyte solution/ membrane temperature. To calibrate the sample temperature for the dissolved oxygen measurement line: 1. In the measurement mode, press to choose the bottom measurement line and press / until the sample temperature is displayed (i.e. ). 2. Press to begin the calibration. 3. When the reading stabilizes, the arrow icon and the first digit will flash. Enter the temperature by pressing / to adjust each digit and to move to the next digit. 4. Press to save and end the calibration. To calibrate the electrolyte solution/membrane temperature for the dissolved oxygen measurement line: 1. In the measurement mode, press to choose the bottom measurement line and press / until the membrane temperature is displayed (i.e. ). 2. Press to begin the calibration. 3. When the reading stabilizes, the arrow icon and the first digit will flash. Enter the temperature by pressing / to adjust each digit and to move to the next digit. 4. Press to save and end the calibration. EN-30 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 35. RDO® Optical Dissolved Chapter IX Oxygen Technique Note: The RDO optical dissolved oxygen meters have serial numbers that begin with R (i.e. R12345). Only the RDO meters are compatible with the RDO optical dissolved oxygen sensors. RDO Cap Overview The RDO sensor has an internal clock that counts down the 365 day lifespan of a new RDO cap. The countdown begins when the RDO cap is installed on the RDO sensor, the sensor is connected to the meter and the first measurement is taken. The 365 day countdown cannot be reset or changed once the first measurement is taken. Each RDO cap has a unique serial number that is recognized by the RDO sensor, so reinstalling the cap will not reset the countdown. Note: The power to the meter must be turned off when a new RDO cap is installed on the RDO sensor. Once the RDO cap is installed, turn the power to the meter on and the new cap information will be sent to the meter. To print the RDO cap information: 1. Connect the meter to a printer or computer and verify the meter baud rate and output settings in the setup menu. 2. From the measurement mode, press . 3. Press / to scroll through the setup menu until is displayed on the top line and is displayed on the middle line. 4. Press two times to move the arrow icon to the bottom line. 5. Press / until is displayed on the bottom line. 6. Press to print the RDO cap information. Press to return to the measurement mode. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-31
  • 36. RDO® Optical Dissolved Oxygen Technique RDO Optical Dissolved Oxygen Setup Menu Note: Refer to the Setup Menu section for the Setup Menu Table, which contains a complete list of meter setup options and descriptions. 1. From the measurement mode, press . 2. Press / to scroll through the setup menu until is displayed on the top line. 3. Press to accept the selection and move the arrow icon to the middle line. 4. Press / to scroll through for the % saturation resolution, for the mg/L concentration resolution, for the barometer type (automatic or manual), for the manual barometric pressure compensation value, for the salinity compensation type (automatic or manual), for the manual salinity correction value, for the dissolved oxygen calibration type and for the RDO cap replacement countdown in days. 5. Press to select the option and move the arrow icon to the bottom line. 6. To scroll through a list of options on the bottom line, press / until the desired option is displayed. To enter a numeric value for an option on the bottom line, press / to adjust each digit and to move to the next digit. 7. Press to accept the selection and move the arrow icon to the top line. 8. Repeat steps 3 through 7 to change another dissolved oxygen setting or press to return to the measurement mode. EN-32 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 37. RDO® Optical Dissolved Oxygen Technique Setup Menu for RDO Optical Dissolved Oxygen Sensor The RDO optical dissolved oxygen meters have a special setup menu feature that allows the user to print information about the RDO sensor. Top Line Middle Line Bottom Line Setup Menu Description Pr0b d0 tESt The tESt option initiates a 30 second test of the communication link between the RDO sensor and meter. The temperature display will show a 30 second countdown as the test progresses. The meter will send a report to the printer or computer when the test is done. Pr0b d0 CAL The CAL option prints the meter serial number, date, time and slope for the last five dissolved oxygen calibrations that were performed with the RDO sensor that is currently connected to the meter. Pr0b d0 dFLt The dFLt option prints the factory calibration information for the RDO sensor that is currently connected to the meter. Pr0b d0 SLP The SLP option prints a drift per minute value of the slope for the last five dissolved oxygen calibrations that were performed with the RDO sensor that is currently connected to the meter. Pr0b d0 mEtH The mEtH option prints the RDO specific method information of the last method used for the RDO sensor that is currently connected to the meter. Pr0b d0 InF0 The InF0 option prints the RDO sensor information, including the sensor serial number and revision, sensor and RDO cap date of manufacturing, cap start date, cap life and internal real time clock for the sensor that is currently connected to the meter. 1. Connect the meter to a printer or computer and verify the meter baud rate and output settings in the setup menu. 2. From the measurement mode, press . 3. Press / to scroll through the setup menu until is displayed on the top line and is displayed on the middle line. 4. Press two times to move the arrow icon to the bottom line. 5. Press / to scroll through for the communication link test, for the RDO sensor calibration information, for the RDO sensor factory calibration information, for the RDO sensor slope information, for the RDO method information and for the general RDO sensor information. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-33
  • 38. RDO® Optical Dissolved Oxygen Technique 6. Press to print the selected option. 7 Press . / to select another option from the bottom line and press to print the selected option or press to return to the measurement mode. Example RDO Sensor General Information Printout smartprobe info type 1 run_status 255 format_version 1 probe_SN 6 app_version 9 HW_version 1 cap_SN 129614 rtc 1210257372 (05-08-2008 14:36:12) mfg_time 1208371349 (04-16-2008 18:42:29) start_time 1210178953 (05-07-2008 16:49:13) expiration 1241736553 (05-07-2009 22:49:13) probe_mfg_time 1208440800 (04-17-2008 14:00:00) write_count 14 pass_count 12 expire 0x01 days_left 364.3 probe_reading 149.529243 temp_reading 24.942651 state 21 general_event 0x00 internal_event 0x00 internal_state 3 internal_retry 0x00 EN-34 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 39. RDO® Optical Dissolved Oxygen Technique RDO Optical Dissolved Oxygen Calibration 1. Select one of the following calibration modes in the setup menu. a. – An air calibration is performed in water saturated air using the calibration sleeve. This is the simplest and most accurate calibration method. i. The highest possible accuracy is reached when calibration temperature is the same as the measuring temperature. ii. oisten the sponge in the calibration sleeve with distilled water. M Insert the RDO sensor into the sleeve without touching the water saturated sponge. b. – A water calibration is performed using water that is 100% saturated with air. Bubble air into a water sample and gently stir the sample to prevent the buildup of air bubbles on the RDO cap. c. – A manual calibration is performed using a water sample with a known concentration of dissolved oxygen. This method can be used to calibrate the RDO sensor to the value achieved by a Winkler titration. i. manual calibration involves performing a Winkler titration and using A that sample as a calibration standard. The oxygen level result from the titration is entered in a manual calibration as the dissolved oxygen value. This correlates the meter input to the Winkler titration. This method is inherently less accurate, due to the possibility of titration errors. d. – A zero point calibration is performed in an oxygen-free solution. A zero point calibration is not generally required unless measurements will be taken below 10% saturation or 1 mg/L. 2. Allow the RDO sensor and calibration standard (water saturated air, air saturated water, Winkler standard or oxygen-free solution) to reach equilibrium. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-35
  • 40. RDO® Optical Dissolved Oxygen Technique 3. In the measurement mode, press until the arrow icon points to the bottom line, press until the % sat or mg/L icon is shown and press to begin the calibration. 4. Wait for the dissolved oxygen reading to stabilize. a. an air calibration is performed, the meter will display 100.0% and If automatically return to the measurement mode. b. a water calibration is performed, the meter will display 100.0% and If automatically return to the measurement mode. c. a manual calibration is performed, wait for the mg/L icon to stop flashing If and enter the dissolved oxygen value by pressing until the first digit to be changed is flashing, press / to change the value of the flashing digit and continue to change the digits until the meter displays the correct dissolved oxygen value. Once the dissolved oxygen value is set, press until the decimal point is in the correct location. d. a zero point calibration is performed, the meter will display 0.00 and If automatically return to the measurement mode. EN-36 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 41. RDO® Optical Dissolved Oxygen Technique RDO Optical Dissolved Oxygen Measurement Note: Turn on the automatic datalog feature to send measurements to the meter datalog at the frequency specified in each measurement mode. Refer to the Data Archiving and Retrieval section for details. If the automatic datalog feature is off, connect the meter to a printer or computer to record the measurements. 1. Rinse the RDO sensor with distilled or deionized water. Shake off any excess water and blot the sensor dry with lint-free tissue. 2. Place the RDO sensor into the sample. a. the meter is in the continuous measurement mode, it will start reading If immediately and continuously update the display. The mg/L or % sat icon will flash until the reading is stable. Once the reading is stable, log and print the measurement by pressing . This mode will use a large amount of meter battery power. b. the meter is in the AUTO-READ measurement mode, press If to start the reading. The AR icon will flash until the reading is stable. Once the reading is stable, the meter will log and print the measurement and freeze the display. This mode uses various amounts of meter battery power, depending on how often is pressed. c. the meter is in the timed measurement mode, it will take a dissolved If oxygen reading at the frequency specified in the setup menu. If parameters other than dissolved oxygen are measured, the meter will continuously take the other readings and update the display. The meter will log and print the measurement at the predetermined time interval. This mode can be used to extend the meter battery life, since the power consumption of the RDO sensor is reduced, depending on the set time interval. 3. Remove the RDO sensor from the sample, rinse it with distilled or deionized water, blot it dry, place it in the next sample and repeat step 2. 4. Once all of the samples have been measured, rinse the RDO sensor with distilled or deionized water and blot it dry. Consult the RDO sensor user guide for proper storage techniques. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-37
  • 42. RDO® Optical Dissolved Oxygen Technique RDO Optical Dissolved Oxygen Temperature Display and Calibration RDO Optical Dissolved Oxygen Temperature Display Star Plus meters allow the temperature to be viewed on individual measurement lines in addition to the temperature display on the top, left of the screen. To view the temperature for the RDO dissolved oxygen measurement line: 1. In the measurement mode, press to choose the bottom display line. The arrow icon will point to the selected line. 2. Press / to change the value on the selected line. The bottom line can be changed to display dissolved oxygen (% saturation), dissolved oxygen (mg/L), barometric pressure (no icon), temperature (no icon) or a blank line. RDO Optical Dissolved Oxygen Temperature Calibration The temperature calibration mode of the Star Plus meter allows the temperature on each measurement line to be manually adjusted. To calibrate the temperature for the RDO dissolved oxygen measurement line: 1. In the measurement mode, press to choose the bottom measurement line and press / until the temperature is displayed. 2. Press to begin the calibration. 3. When the reading stabilizes, the arrow icon and the first digit will flash. Enter the temperature by pressing / to adjust each digit and to move to the next digit. 4. Press to save and end the calibration. EN-38 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 43. Chapter X Conductivity Technique Conductivity Setup Menu Note: Refer to the Setup Menu section for the Setup Menu Table, which contains a complete list of meter setup options and descriptions. Refer to Appendix A for a description of the special setup menu features. 1. In the measurement mode, press . 2. Press / to scroll through the setup menu until is displayed on the top line. 3. Press to accept the selection and move the arrow icon to the middle line. 4. Press / to scroll through for the temperature compensation type, for the temperature coefficient value used for temperature compensation, for the TDS factor value used for total dissolved solids measurement, for the nominal cell constant value of the conductivity probe, for the reference temperature used for temperature compensation and for the conductivity cell type. 5. Press to accept the selection and move the arrow icon to the bottom line. 6. To scroll through a list of options on the bottom line, press / until the desired option is displayed. To enter a numeric value for an option on the bottom line, press / to adjust each digit and to move to the next digit. 7. Press to accept the selection and move the arrow icon to the top line. 8. Repeat steps 3 through 7 to change another conductivity setting or press to return to the measurement mode. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-39
  • 44. Conductivity Technique Conductivity Calibration Note: For an automatic calibration, the nominal cell constant of the conductivity probe must be entered in the setup menu before the calibration is performed. 1. In the measurement mode, press until the arrow icon points to the middle line, press until the µS/cm or mS/cm icon is shown and press to begin the calibration. 2. Rinse the probe with deionized water and place it into the conductivity standard. 3. To perform a manual calibration – The manual calibration screen will display the cell constant on the top line, the conductivity value of the calibration standard on the middle line and on the bottom line. To change the cell constant, press until the first digit to be changed is flashing, press / to change the value of the flashing digit and continue to change the digits until the displayed conductivity value matches the value of the standard at the measured temperature. Once the value is set, press until the decimal point is in the correct location. Press to save and end the calibration. Note: In the manual calibration screen, start changing the cell constant within five seconds or the meter will proceed to the automatic/direct calibration. If this occurs, press and hold to abort the calibration and repeat the calibration. 4. To perform an automatic or direct calibration – Wait for the meter to go from the manual calibration screen to the automatic/direct calibration screen. The automatic/direct calibration screen will display the conductivity value of the calibration standard on the middle line and on the bottom line. Automatic calibration – When the µS/cm or mS/cm icon stops flashing, the a. meter will display the temperature-corrected conductivity of the standard. Direct calibration – When the µS/cm or mS/cm icon stops flashing, the b. meter will display the actual conductivity value read by the probe. To change the conductivity value, press until the first digit to be changed is flashing, press / to change the value of the flashing digit and continue to change the digits until the correct conductivity value of the standard at the measured temperature is displayed. Once the value is set, press until the decimal point is in the correct location. EN-40 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 45. Conductivity Technique 5. Press to proceed to the next calibration point, rinse the conductivity probe with distilled or deionized water, place it into the next conductivity standard and repeat step 4a / 4b or press to save and end the calibration. 6. The cell constant will be displayed in the main field and the meter will automatically advance to the measurement mode. Conductivity Measurement Note: Turn on the automatic datalog feature to send measurements to the meter datalog at the frequency specified in each measurement mode. Refer to the Data Archiving and Retrieval section for details. If the automatic datalog feature is off, connect the meter to a printer or computer to record the measurements. 1. Rinse the conductivity probe with distilled or deionized water. Shake off any excess water and blot the probe dry with lint-free tissue. 2. Place the conductivity probe into the sample. a. the meter is in the continuous measurement mode, it will start reading If immediately and continuously update the display. The μS/cm or mS/cm icon will flash until the reading is stable. Once the reading is stable, log and print the measurement by pressing . If a benchtop meter is used and the stirrer is enabled, press to start the stirrer. Press again to turn off the stirrer before removing the probe and stirrer from the sample. b. the meter is in the AUTO-READ measurement mode, press If to start the reading. The AR icon will flash until the reading is stable. Once the reading is stable, the meter will log and print the measurement and freeze the display. If a benchtop meter is used and the stirrer is enabled, the stirrer will turn on when is pressed and turn off when the reading is stable. c. the meter is in the timed measurement mode, it will start reading If immediately and continuously update the display. The meter will log and print the measurement at the frequency specified in the setup menu. If a benchtop meter is used and the stirrer is enabled, press to start the stirrer. Press again to turn off the stirrer before removing the probe and stirrer from the sample. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-41
  • 46. Conductivity Technique 3. Remove the conductivity probe from the sample, rinse it with distilled or deionized water, blot it dry, place it in the next sample and repeat step 2. 4. Once all of the samples have been measured, rinse the conductivity probe with distilled or deionized water and blot it dry. Consult the conductivity probe user guide for proper storage techniques. Conductivity Temperature Display and Calibration Conductivity Temperature Display Star Plus meters allow the temperature to be viewed on individual measurement lines in addition to the temperature display on the top, left of the screen. 1. In the measurement mode, press to choose the middle display line. The arrow icon will point to the selected line. 2. Press / to change the value on the selected line. The middle line can be changed to display conductivity (µS/cm or mS/cm), total dissolved solids (mg/L), salinity (ppt), resistivity (MΩ-cm), temperature (no icon) or a blank line. Conductivity Temperature Calibration The temperature calibration mode of the Star Plus meter allows the temperature on each measurement line to be manually adjusted. 1. In the measurement mode, press to choose the middle measurement line and press / until the temperature is shown for the selected line. 2. Press to begin the calibration. 3. When the reading stabilizes, the arrow icon and the first digit will flash. Enter the temperature by pressing / to adjust each digit and to move to the next digit. 4. Press to save and end the calibration. EN-42 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 47. Chapter XI ISE Technique ISE Setup Menu Note: Refer to the Setup Menu section for the Setup Menu Table, which contains a complete list of meter setup options and descriptions. Refer to Appendix A for a description of the special setup menu features. 1. In the measurement mode, press . 2. Press / to scroll through the setup menu until is displayed on the top line. 3. Press to accept the selection and move the arrow icon to the middle line. 4. Press / to scroll through for the ISE measurement resolution, for the ISE measurement units, for the ISE calibration range and for the non-linear blank correction feature. 5. Press to accept the selection and move the arrow icon to the bottom line. 6. To scroll through a list of options on the bottom line, press / until the desired option is displayed. To enter a numeric value for an option on the bottom line, press / to adjust each digit and to move to the next digit. 7. Press to accept the selection and move the arrow icon to the top line. 8. Repeat steps 3 through 7 to change another ISE setting or press to return to the measurement mode. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-43
  • 48. ISE Technique ISE Calibration The calibration standards should be prepared in the same ISE units as the desired sample results. Start the calibration with the lowest concentration calibration standard and work up to the highest concentration calibration standard. Any reagents, such as ionic strength adjustors, should be added to samples and standards as specified in the electrode user guide. 1. Prepare the electrode, standards and any other required solutions for use according to the electrode user guide. 2. In the measurement mode, press until the arrow icon points to the top line, press until the ISE icon is shown and press to begin the calibration. 3. Rinse the electrode with distilled or deionized water, shake any excess water off, blot it dry and place the electrode into the least concentrated standard. 4. Wait for ISE icon to stop flashing. Press until the first digit to be changed is flashing, press / to change the value of the flashing digit and continue to change the digits until the meter displays the concentration value of the standard. Once the standard value is set, press until the decimal point is in the correct location. 5. Press to proceed to the next lowest calibration standard and repeat steps 3 and 4, working from the lowest concentration standard to the highest concentration standard, or press to save and end the calibration. 6. The actual electrode slope, in mV per decade concentration, will be displayed in the main field and will be displayed in the lower field. a. a one point calibration, press For and / to edit the slope. To change the sign of the slope to negative or positive, press until none of the digits are blinking and the arrow icon is blinking and press to change the sign of the slope. Press to return to the measurement mode. b. or a two or more point calibration, the meter will automatically proceed to F the measurement mode after the slope is displayed. EN-44 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 49. ISE Technique ISE Measurement Note: Turn on the automatic datalog feature to send measurements to the meter datalog at the frequency specified in each measurement mode. Refer to the Data Archiving and Retrieval section for details. If the automatic datalog feature is off, connect the meter to a printer or computer to record the measurements. 1. Rinse the electrode with distilled or deionized water. Shake off any excess water and blot the electrode dry with lint-free tissue. 2. Place the electrode into the sample. a. the meter is in the continuous measurement mode, it will start reading If immediately and continuously update the display. The ISE icon will flash until the reading is stable. Once the reading is stable, log and print the measurement by pressing . If a benchtop meter is used and the stirrer is enabled, press to start the stirrer. Press again to turn off the stirrer before removing the electrode and stirrer from the sample. b. the meter is in the AUTO-READ measurement mode, press If to start the reading. The AR icon will flash until the reading is stable. Once the reading is stable, the meter will log and print the measurement and freeze the display. If a benchtop meter is used and the stirrer is enabled, the stirrer will turn on when is pressed and turn off when the reading is stable. c. the meter is in the timed measurement mode, it will start reading If immediately and continuously update the display. The meter will log and print the measurement at the frequency specified in the setup menu. If a benchtop meter is used and the stirrer is enabled, press to start the stirrer. Press again to turn off the stirrer before removing the electrode and stirrer from the sample. 3. Remove the electrode from the sample, rinse it with distilled or deionized water, blot it dry, place it in the next sample and repeat step 2. 4. Once all of the samples have been measured, rinse the electrode with distilled or deionized water and blot it dry. Consult the electrode user guide for proper storage techniques. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-45
  • 50. ISE Technique ISE Temperature Display and Calibration ISE Temperature Display Star Plus meters allow the temperature to be viewed on individual measurement lines in addition to the temperature display on the top, left of the screen. To view the temperature for the ISE measurement line: 1. Press and hold until the meter displays the measurement mode. 2. Press to choose the top display line. The arrow icon will point to the selected line. 3. Press / to change the value on the selected line. The top line can be changed to display pH (pH), millivolts (mV), relative millivolts (RmV), concentration (ISE), temperature (no icon) or a blank line. ISE Temperature Calibration The temperature calibration mode of the Star Plus meter allows the temperature on each measurement line to be manually adjusted. To calibrate the temperature for the ISE measurement line: 1. In the measurement mode, press to choose the top measurement line and press / until the temperature is shown for the selected line. 2. Press to begin the calibration. 3. When the reading stabilizes, the arrow icon and the first digit will flash. Enter the temperature by pressing / to adjust each digit and to move to the next digit. 4. Press to save and end the calibration. EN-46 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 51. Chapter XII Data Archiving and Retrieval Datalog and Calibration Log All 3-Star Plus and the 4-Star Plus pH/ISE meters have a 1000 point datalog. The 4-Star Plus pH/conductivity and pH/dissolved oxygen meters have a 750 point datalog. The 5-Star Plus meters have a 500 point datalog. The 2-Star meter has a 50 point datalog and all Orion Star meters have a 200 point datalog. The Star Plus meter printouts have been enhanced to include additional information. The pH and ISE calibration printouts now include the average slope, the slope between points and the Eo per point. The polarographic dissolved oxygen printouts now include slope, membrane temperature and solution temperature. Automatic Datalog Feature The 3-Star, 4-Star and 5-Star Plus meters have been enhanced to include an automatic datalog feature that can be turned on or off. To enable the automatic recording of data into the datalog: 1. In the measurement mode, press . 2. Press / until is displayed on top line. 3. Press to accept the selection and move the arrow icon to the middle line and press / until is displayed. 4. Press to accept the selection and move the arrow icon to the bottom line and press / until is displayed. 5. Press to accept the selection and move the arrow icon to the top line. 6. Press to save the setup option and return to measurement mode. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-47
  • 52. Data Archival and Retrieval Datalog Deletion Setting The datalog deletion setting determines if the meter will automatically delete the datalog after it is downloaded to a printer or computer and if the meter will overwrite the datalog points when the datalog is full. If the datalog deletion setting is set to , the meter will automatically delete the datalog after the datalog is downloaded to a printer or computer. The meter will also display the error message when all 200 datalog points are filled and the datalog must be downloaded to a printer or computer to clear the error message. If the datalog deletion setting is set to , the meter will overwrite the oldest datalog point when all 200 datalog points are filled and will not delete the datalog after the datalog is downloaded to a printer or computer. 1. In the measurement mode, press . 2. Press / until is displayed on top line. 3. Press to accept the selection and move the arrow icon to the middle line and press / until is displayed. 4. Press to accept the selection and move the arrow icon to the bottom line and press / until or is displayed. 5. Press to accept the selection and move the arrow icon to the top line. 6. Press to save the setup option and return to measurement mode. Note: If the datalog is not required, set the datalog deletion setting to to prevent the error 038 (datalog full) message. EN-48 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 53. Viewing and Printing the Datalog and Calibration Log The 3-Star, 4-Star and 5-Star Plus meters have been enhanced to include a calibration log view feature in addition to the datalog view, datalog print and calibration log print features. The Orion Star meters include the datalog view, datalog print and calibration log print features. To view the datalog or calibration log: 1. In the measurement mode, press . 2. Press / to scroll through to view the datalog or to view the calibration log (Star Plus meters only). 3. Press . The meter will display the date/time screen. The log number will be on the top of the screen and the time, date and year the log was recorded will be on the top, middle and bottom display lines respectively. Press / to scroll through the log. 4. Press . The meter will display the data or calibration point associated with the selected date/time screen. a. Press to print the individual data point. b. Press / to scroll through the log. c. Press to return to the date/time screen. 5. To exit the log view mode, press until the meter displays the date/time screen and press . To send the datalog or calibration log to a printer or computer: 1. Connect the meter to a printer or computer and verify the meter baud rate and output settings in the setup menu. 2. In the measurement mode, press . Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-49
  • 54. 3. Press / to scroll through to print the datalog or to print the calibration log. The Orion Star meters will display instead of . 4. Press to send the selected data to the printer or computer. To interfacing the meter with a computer: The Orion Star and Star Plus meters can send measurement and calibration data to a computer in a comma delimited format that is easy to parse in computer programs like Excel. Select the , , output setting in the setup menu. To send data from the meter to a computer using HyperTerminal: 1. Connect the meter to a computer port using the computer interface cable, Cat. No. 1010053. 2. Click on the start button on the lower left side of the computer screen. Select All Programs, Accessories, Communications and HyperTerminal. 3. When the HyperTerminal window opens, enter a file name, select an icon for the connection and click on the OK button. 4. When a new window opens, go to the Connect Using drop-down menu, select the COM port that the meter is connected to and click on the OK button. 5. A window will open with the COM port properties listed. Select the following settings from the drop-down menus and then click on the OK button. Bits per second: 9600 Data bits: 8 Parity: None Stop bits: 1 Flow control: Hardware 6. Send data from the meter to HyperTerminal. EN-50 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 55. Chapter XIII Declaration of Conformity Manufacturer: Thermo Fisher Scientific Inc. Address: 1 66 Cummings Center Beverly, MA 01915 USA We declare that the following products described below conform to the Directive and Standard listed below: Product(s): eters for measuring pH, conductivity, dissolved oxygen, and/or ISE M Benchtop models are rated 100 to 240 VAC, 50/60 Hz, 0.5 A Handheld models use four non-rechargeable AA batteries Benchtop Meters Portable Meters 5-Star Plus pH/ISE/Conductivity/DO Meter 5-Star Plus pH/ISE/Conductivity/DO Meter 4-Star Plus pH/Conductivity Meter 5-Star Plus pH/Conductivity/DO Meter 4-Star Plus pH/DO 4-Star Plus pH/Conductivity Meter 4-Star Plus pH/ISE Meter 4-Star Plus pH/DO Meter 3-Star Plus Conductivity Meter 4-Star Plus pH/ISE Meter 3-Star Plus DO Meter 3-Star Plus Conductivity Meter 3-Star Plus pH Meter 3-Star Plus pH Meter 2-Star pH Meter 3-Star Plus DO Meter 5-Star Plus RDO® Optical DO/pH/Conductivity Meter 4-Star Plus RDO Optical DO/Conductivity Meter 3-Star Plus RDO Optical DO Meter Equipment Class: easurement, control and laboratory M Benchtop models are EMC Class A Portable models are EMC Class D Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-51
  • 56. Declaration of Conformity Directive(s) and Standard(s): • 89/336/EEC – Electromagnetic Compatibility (EMC Directive) • 61326:1997 + A1:1998 + A2:2001 – Electrical equipment for EN measurement, control, and laboratory use – EMC requirements • 73/23/EEC – Low Voltage Directive (LVD) • 61010-1:2001 – Safety requirements for electrical equipment for EN measurement, control, and laboratory use – general requirements Manufacturer’s Authorized Representative: Date: Patrick Chiu June 6, 2006 Senior Quality Assurance Engineering, Regulatory Compliance WEEE Compliance This product is required to comply with the European Union’s Waste Electrical Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol: We have contracted with one or more recycling/disposal companies in each EU Member State and this product should be disposed of or recycled through them. Further information on compliance with these Directives, the recyclers in your country, and information on Thermo Scientific Orion products which may assist the detection of substances subject to the RoHS Directive are available at www.thermo.com/WEEERoHS. EN-52 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 57. Chapter XIV Troubleshooting Meter Self Test 1. Disconnect all of the electrodes and probes from the meter and cover all of the meter inputs with the black caps. 2. Power on the meter, wait until the software revision is displayed and press . 3. All the segments on the display will turn on. Visually inspect the display segments to verify that all of the segments are lit and press . 4. All the segments on the display will turn off. Visually inspect the display segments to verify that all of the segments are not lit and press . 5. The display will read . Press every key on the keypad one at a time in any order. If the keys are not pressed within five seconds of one another, the display will read , which indicates a key failure. Press to clear the error 033 message and complete the self test. If all the keys are pressed and functioning, the meter will restart and proceed to the measurement mode. Note: If the meter reads during the self test, ensure that all of the electrodes are disconnected from the meter, all of the meter inputs are covered with the black caps and the BNC shorting cap is firmly attached to the BNC meter input. This error code usually occurs if the BNC shorting cap is missing or not fully connected to the BNC meter input during the meter self test. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-53
  • 58. Troubleshooting Meter Error Codes • If the reading on the screen is flashing , the value is out of range. Perform the meter self test, clean the electrode according to the electrode user guide and re-calibrate the electrode with new standards. • If the icon is lit and the reading is flashing, the sensor needs to be calibrated according to the user’s set calibration interval or the pH slope is outside the range of 85 % to 115%. • Press to clear an error code. Error codes show on the middle line and a set of three alphanumeric characters on the bottom line. Some of these codes are errors, some are warnings and some are purely informational. Error Code Description Troubleshooting 002, 026, Hardware or Memory Press to clear the error. If the error occurs again, contact E##, F## Error Technical Support. 005 Value Outside Press and re-enter the value. Check meter specifications for the Allowable Range allowable range of values. 033 Keypad Failure Repeat the self test. When the meter reads , press all the keys, including the power key, within five seconds of one another. 034 BNC Input Failure Disconnect all the electrodes from the meter, connect the BNC shorting cap to the meter and repeat the self test. 038 Datalog Full Download the datalog to a printer or computer, turn the automatic datalog feature off in the setup menu (Star Plus meters only) or change the datalog setting to , , in the setup menu so the meter deletes the datalog points when the datalog is full. D## Remote Control Error Check the programming instructions to verify the correct commands, names and values. 107 pH Calibration The millivolts measured during calibration are the same for two buffers. Standard Error Review the calibration procedure and verify that the electrode was placed in the buffers at the appropriate time. Clean the electrode according to the electrode user guide. Re-calibrate the electrode with fresh buffers. 109 Bad pH Slope or Clean the electrode according to the electrode user guide. Re-calibrate the Calibration Offset electrode with new buffers. 200 Autosampler The meter is unable to send a signal to the autosampler. Make sure that Interface Error the autosampler is properly connected to the meter. 201 Autosampler The autosampler is unable to receive a signal from the meter. Review the Signal Error meter setup parameters and make sure that the baud rate of the meter is set to 1200. EN-54 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 59. Troubleshooting Error Code Description Troubleshooting 202 Autosampler is Turn the autosampler off and wait 45 seconds before turning it back on. Jammed The autosampler should return to the home position. 203 Unstable Reading The measurements taken using the autosampler are unstable. Check the from Autosampler electrodes for proper function. Make sure that the electrode cables are properly connected. 306 ISE Automatic Disable the automatic blank feature in the setup menu and re-calibrate Blank Error the meter without using a zero concentration standard. 307 ISE Calibration The millivolts measured during calibration are the same for two Standard Error standards. Review the calibration procedure and verify that the electrode was placed in the standards at the appropriate times. Clean the electrode according to the electrode user guide. Re-calibrate the electrode with fresh standards. 309 Bad ISE Slope Clean the electrode according to the electrode user guide. Re-calibrate the electrode with freshly prepared standards. 707 Conductivity The conductivity value measured during calibration is the same for Calibration two standards. Review the calibration procedure and verify that the Standard Error conductivity probe was placed in the standards at the appropriate times. Clean the conductivity probe according to the probe user guide. Re- calibrate the probe with new standards. 709 Conductivity Cell The cell constant is not in the range of 0.001 to 199.0 cm-1. Clean the Constant Error conductivity probe according to the probe user guide. Re-calibrate the probe with new standards. 808 Bad Zero Point An air calibration should be performed before the zero point calibration. DO Slope Verify that a solution with zero oxygen is being used for the zero point calibration. A solution with 15 grams of Na2SO3 dissolved in 250 mL of distilled water is recommended. 809 Bad DO Slope For polarographic DO probes, connect the probe to the meter, power on the meter and let the probe to polarize for at least 30 minutes. For an air calibration, check that the sponge in the calibration sleeve is damp and there is no water on the probe membrane . For a water calibration, bubble air into the sample and stir to keep bubbles off the membrane. Clean the DO probe according to the probe user guide. Re-calibrate the DO probe. 880 RDO® Optical DO Verify that the RDO sensor is properly connected to the meter. Connect Sensor Not Attached another, known working RDO sensor to the meter. 881 RDO Optical DO The RDO cap attached to the RDO sensor has expired. Install a new RDO Sensor Expired cap according to the RDO sensor user guide. 882 RDO Optical DO Verify that the RDO cap on the RDO sensor was properly installed, not Sensor Failure expired and not tampered with. Connect another, known working RDO sensor to the meter. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-55
  • 60. Troubleshooting General Troubleshooting Problem: The display freezes and the measurement values will not change. Solution: The meter is in the AUTO-READ measurement mode (the AR icon will appear in the top, right corner of the display). Press to start a new reading or select another measurement mode in the setup menu. Problem: When I press the meter displays . Solution: The meter is printing and cannot enter the calibration mode until the printing is done. This should rarely occur if the meter is set to a 9600 baud rate. If the meter is at a lower baud rate, the delay will be longer. Problem: The meter did not accept the change I made in the setup menu. Solution: After making a change in the setup menu, press until the arrow icon points to the top line (confirms the change) and then press to save the change and return to the measurement mode. Problem: How do I abort a calibration? Solution: Press and hold to abort any meter operation and return to the measurement mode. Problem: The printout is a string of numbers and units with commas. Solution: The output format in the setup menu is set to the computer output or the printer baud rate is set incorrectly in the setup menu. Change the output format to the printer output in the setup menu. Change the baud rate to the correct value for the printer that is being used. Problem: When I press the stirrer button, the stirrer doesn’t work. Solution: The current stirrer setting is off. Set the speed to 1 through 7 in the setup menu. Problem: The timed measurement time entry screen does not appear in the setup menu. Solution: The meter is in the AUTO-READ or continuous mode. When the meter is set to the timed mode, the next setup screen will be for time entry. Problem: cannot tell if I have the Star Plus meter or the Orion Star meter. I Solution: When the meter is powered on, the Star Plus meters with enhanced features will display with the meter revision number ( or similar) and proceed to the measurement mode. EN-56 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 61. Troubleshooting pH Troubleshooting Problem: The meter does not recognize the pH buffer value during calibration. Solution: Verify that the correct buffer set was selected in the setup menu. The meter uses the raw mV reading of the electrode to recognize a buffer during calibration. As the electrode ages or becomes dirty, its mV readings will drift and you will need to manually enter the pH buffer value when calibrating. ISE Troubleshooting Problem: It takes several minutes for the readings to stabilize during a calibration. Solution: The concentration range in the setup menu is set to low. Change the concentration range to high. The ISE resolution is set to 3 digits in the setup menu. Change the ISE resolution to 2 digits for faster stabilization of the readings. Problem: When I use the automatic blank correction setting and calibrate an ISE, the meter gives a slope that is too low or cannot be manually checked. Solution: Turn the automatic blank correction setting off in the setup menu. Conductivity Troubleshooting Problem: The meter does not recognize the conductivity standard during calibration. Solution: Verify that the default cell constant was entered in the setup menu. The cell constant is usually printed on the conductivity probe cable. Verify that the conductivity standard is one that is programmed into the meter. Re-calibrate with a fresh standard. Problem: The temperature coefficient value does not appear in the setup menu. Solution: The current temperature compensation setting is nonlinear or off. Change the temperature compensation to linear and the next screen will be the temperature coefficient value entry screen. Problem: The measurement is out of range when it should be in range. Solution: Check that the conductivity probe is fully immersed in the solution. Verify that the cell constant is correct for the conductivity probe that is connected to the meter. Verify that the cell type selected in the setup menu is set to Std. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-57
  • 62. Troubleshooting Dissolved Oxygen Troubleshooting Problem: The manual barometric pressure entry screen does not appear in the setup menu. Solution: The barometric pressure compensation is set to automatic in the setup menu. Change the barometric pressure compensation to manual and the next screen will be the manual pressure entry screen. Problem: The manual salinity factor entry screen does not appear in the setup menu. Solution: The salinity correction is set to automatic in the setup menu. Change the salinity correction to manual and the next screen will be the salinity factor entry screen. Problem: The AUTO-STIR BOD probe does not turn on when the button on the probe is pressed. Solution: The read type must be set to AUTO-READ in the setup menu and the stirrer speed must be set from 1 to 7 to initiate a measurement and start stirring by pressing the button on the AUTO-STIR BOD probe. RDO® Optical Dissolved Oxygen Troubleshooting Problem: The meter displays an error 881 message and will not take a dissolved oxygen measurement. Solution: Turn the meter off, replace the old RDO cap with a new cap and turn the meter on. This should clear the error message. Assistance After troubleshooting all components of your measurement system, contact Technical Support. Within the United States call 1.800.225.1480 and outside the United States call 978.232.6000 or fax 978.232.6031. In Europe, the Middle East and Africa, contact your local authorized dealer. For the most current contact information, visit www.thermo.com/contactwater. For the latest application and technical resources for Thermo Scientific Orion products, visit www.thermo.com/waterapps. Warranty For the most current warranty information, visit www.thermo.com/water. EN-58 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 63. Chapter XV Meter Specifications Meter Specifications Environmental Operating Conditions Portable and Benchtop Meter Environmental Operating Conditions Operating Ambient Temperature 5 to 45 °C Operating Relative Humidity 5 to 85 %, non-condensing Storage Temperature -20 to +60 °C Storage Relative Humidity 5 to 85 %, non-condensing Pollution Degree 2 Overvoltage Category II Altitude Up to 2000 meters Weight Portable: 0.45 kg Benchtop: 0.91 kg Size Portable: 4.8 cm (H) x 9.7 cm (W) x 21.3 cm (D) Benchtop: 9.4 cm (H) x 17.0 cm (W) x 22.4 cm (D) AC Powered Meters Indoor use only Battery Operated Meters Indoor or outdoor use Regulatory and Safety CE, CSA, TÜV, UL, FCC Class limits* Case Material ABS Shock and Vibration Vibration, shipping/handling per ISTA #1A Shock, drop test in packaging per ISTA #1A Enclosure (designed to meet) IP67 (portable meter) IP54 (benchtop meter) * TÜV and UL certifications are pending for all Star Plus RDO® optical dissolved oxygen meters. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-59
  • 64. Meter Specifications Universal Power Adapter Environmental Operating Conditions Operating Ambient Temperature 0 to 50 °C Operating Relative Humidity 0 to 90 %, non-condensing Storage Temperature -20 to +75 °C Storage Relative Humidity 0 to 90 %, non-condensing Pollution Degree 2 Overvoltage Category II Operating Altitude Up to 2000 meters Benchtop Meters Indoor use only Meter Parameter Specifications The following meter parameter specifications are for the complete line of Orion Star and Star Plus series meters. Single parameter, dual parameter and some multi- parameter meters will not include all of the parameters listed in this section. pH Range -2.000 to 19.999 Resolution 0.1, 0.01, 0.001 Relative Accuracy ± 0.002 Calibration Points 1 to 5 pH (2-Star pH Meter Only) Range 0.000 to 14.999 Resolution 0.1, 0.01, 0.001 Relative Accuracy ± 0.002 Calibration Points 1 to 3 Millivolts, Relative Millivolts, ORP Range ± 1999.9 mV Resolution 0.1 mV Relative Accuracy ± 0.2 mV or 0.05 % of reading, whichever is greater EN-60 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 65. Meter Specifications ISE Range 0 to 19999 Resolution 1 to 3 significant figures Relative Accuracy ± 0.2 mV or 0.05 %, whichever is greater Displayed Units M, mg/L, %, ppb or no units Calibration Features Linear point to point, selectable non-linear automatic blank correction and low concentration range stability Dissolved Oxygen (Polarographic) Range 0.00 to 90.0 mg/L 0.0 to 600 % Resolution 0.1, 0.01 mg/L 0.1, 1 % Relative Accuracy ± 0.2 mg/L ±2% Salinity Factor 0 to 45 ppt Barometric Pressure 450 to 850 mm Hg Calibration Types Water saturated air, air saturated water, manual (Winkler), zero point Probe Type Polarographic RDO® Optical Dissolved Oxygen Range 0.00 to 20.0 mg/L 0.0 to 200 % Resolution 0.1, 0.01 mg/L 0.1, 1 % Relative Accuracy ± 0.1 mg/L up to 8 mg/L; ± 0.2 mg/L from 8 mg/L to 20 mg/L ±2% Salinity Factor 0 to 45 ppt Barometric Pressure 450 to 850 mm Hg Calibration Types Water saturated air, air saturated water, manual (Winkler), zero point Probe Type RDO optical Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-61
  • 66. Meter Specifications Conductivity Range 0.000 to 3000 mS/cm, auto-resolution with cell constant dependence Resolution 4 significant figures down to 0.001 µS/cm, cell constant dependant Relative Accuracy 0.5 % ± 1 digit or 0.01 µS/cm, whichever is greater Cell Constant 0.001 to 199.9 cm-1 Reference Temperature 5 °C, 10 °C, 15 °C, 20 °C or 25 °C Resistivity Range 0.0001 to 100 Megohm Resistivity Resolution Automatic Resistivity Relative Accuracy 0.5 % ± 1 digit Salinity Range 0.1 to 80.0 ppt NaCl equivalent, 0.1 to 42 ppt practical salinity Salinity Resolution 0.1 ppt Salinity Relative Accuracy 0.1 ± 1 digit TDS Range 0 to 19999 mg/L TDS Resolution 1 mg/L TDS Relative Accuracy 0.5 % ± 1 digit Temperature Range -5 to 105 °C Resolution 0.1 up to 99.9 °C, 1.0 over 99.9 °C Relative Accuracy ± 0.1 °C Temperature* (RDO® Optical Dissolved Oxygen Meter Only) Range 0 to 50 °C Resolution 0.1 °C Relative Accuracy ± 0.3 °C * his temperature specification is only for the Star Plus RDO optical dissolved T oxygen meter when it is used with the RDO optical dissolved oxygen sensor. The 4-Star and 5-Star RDO meters will have the standard temperature specifications when used with an electrode other than the RDO sensor. Note: Specifications are subject to change without notice. EN-62 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 67. Ordering Information Cat. No. Description 1111000 2-Star pH benchtop meter with universal power adapter and user guide 1112000 3-Star Plus pH benchtop meter with universal power adapter and user guide 1212000 3-Star Plus pH portable meter with batteries and user guide 1113000 3-Star Plus DO benchtop meter with universal power adapter and user guide 1213000 3-Star Plus DO portable meter with batteries and user guide 1114000 3-Star Plus conductivity benchtop meter with universal power adapter and user guide 1214000 3-Star Plus conductivity portable meter with batteries and user guide 1115000 4-Star Plus pH/ISE benchtop meter with universal power adapter and user guide 1215000 4-Star Plus pH/ISE portable meter with batteries and user guide 1116000 4-Star Plus pH/DO benchtop meter with universal power adapter and user guide 1216000 4-Star Plus pH/DO portable meter with batteries and user guide 1117000 4-Star Plus pH/conductivity benchtop meter with universal power adapter and user guide 1217000 4-Star Plus pH/conductivity portable meter with batteries and user guide 1218000 5-Star Plus pH/DO/conductivity portable meter with batteries and user guide 1119000 5-Star Plus pH/ISE/DO/conductivity benchtop meter with universal power adapter and user guide 1219000 5-Star Plus pH/ISE/DO/conductivity portable meter with batteries and user guide 1213300 3-Star Plus RDO® optical DO portable meter with batteries and user guide 1213310 4-Star Plus RDO optical DO/conductivity portable meter with batteries and user guide 1213320 5-Star Plus RDO optical DO/pH/conductivity portable meter with batteries and user guide Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-63
  • 68. Cat. No. Description 090043 Swing arm electrode stand 1010003 Universal power adapter 1010006 Star series printer with RS232 printer interface cable (Cat. No. 250302-001) 1010053 RS232 computer interface cable 096019 Stirrer probe with paddle, for 3-Star, 4-Star and 5-Star benchtop meters 8102BNUWP ROSS Ultra combination pH electrode with glass body 8165BNWP ROSS Sure-Flow combination pH electrode with epoxy body 8172BNWP ROSS Sure-Flow combination pH electrode with glass body 9107BNMD Gel-filled pH/ATC Triode with epoxy body 9157BNMD Refillable pH/ATC Triode with epoxy body 9165BNWP Sure-Flow combination pH electrode with epoxy body 9172BNWP Sure-Flow combination pH electrode with glass body 927005MD ATC probe with epoxy body 927007MD ATC probe with stainless steel body 9512HPBNWP High performance ammonia combination ion selective electrode 9609BNWP Fluoride combination ion selective electrode 9707BNWP Nitrate combination ion selective electrode 8611BNWP ROSS sodium combination ion selective electrode 083005MD Polarographic DO probe with calibration sleeve and 1.5 meter cable 083010MD Polarographic DO probe with calibration sleeve and 3 meter cable 086030MD Polarographic BOD AUTO-STIR DO probe with calibration sleeve 087010MD RDO® optical DO sensor with stainless steel guard, RDO cap and 3 meter cable 087020MD RDO optical DO sensor with stainless steel guard, RDO cap and 6 meter cable 011050MD Conductivity probe with 1 µS/cm to 20 mS/cm range and 1.5 meter cable 013005MD DuraProbe conductivity probe with 1 µS/cm to 200 mS/cm range and 1.5 meter cable 013010MD DuraProbe conductivity probe with 1 µS/cm to 200 mS/cm range and 3 meter cable 013016MD Conductivity probe with 0.01 µS/cm to 300 µS/cm range and 1.5 meter cable Visit www.thermo.com/water for additional meter kits, accessories, electrodes and solutions. EN-64 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 69. Appendix A Meter Setup Menu Features pH Setup Menu Features Automatic buffer recognition The Orion Star and Star Plus pH meters are capable of automatically recognizing pH 1.68, 4.01, 6.86, 7.00, 9.18, 10.01 and 12.46 buffers during a pH calibration. During a calibration, the meter uses the selected buffer set and the raw mV reading of the pH electrode in the buffer to recognize and display the buffer value at the measured temperature. The raw mV value must be about ± 30 mV from the theoretical mV reading of the buffer in order for the meter to automatically recognize the buffer. Buffer mV Range Buffer mV Range Buffer mV Range 1.68 +285 to +345 7.00 - 30 to + 30 10.01 -207 to -147 4.01 +207 to +147 9.18 -99 to -159 12.46 -293 to -353 6.86 +38 to -22 Dissolved Oxygen Setup Menu Features Barometric Pressure Compensation The Orion Star and Star Plus dissolved oxygen meters have an internal barometer that is used for pressure compensated dissolved oxygen readings. The meter can also use manual barometric pressure compensation if dissolved oxygen is measured with a submerged probe or in a pressurized vessel. The pressure must be entered as mm Hg. 1 mm Hg = 0.03937 inch Hg = 1.3332 hPa (mBar) = 0.01934 PSI. Salinity Correction Automatic salinity correction for dissolved oxygen readings is available on Orion Star and Star Plus dissolved oxygen meters that have a conductivity measurement mode. The meter uses the conductivity value read by the conductivity probe to calculate the salinity correction factor and applies the factor to dissolved oxygen readings reported as mg/L. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-65
  • 70. Meter Setup Menu Features The meter can also use manual salinity correction for dissolved oxygen readings reported as mg/L. The manual salinity correction factor must be entered as ppt (parts per thousand). Conductivity Salinity Conductivity Salinity Conductivity Salinity at 20 °C Correction at 20 °C Correction at 20 °C Correction (mS/cm) Value (ppt) (mS/cm) Value (ppt) (mS/cm) Value (ppt) 5 3 20 13 35 25 6 4 21 14 36 25 7 4 22 15 37 26 8 5 23 15 38 27 9 6 24 16 39 28 10 6 25 17 40 29 11 7 26 18 42 30 12 8 27 18 44 32 13 8 28 19 46 33 14 9 29 20 48 35 15 10 30 21 50 37 16 10 31 22 52 38 17 11 32 22 54 40 18 12 33 23 56 42 19 13 34 24 This table was calculated from the International Oceanographic Tables, Vol. 1, National Institute of Oceanography of Great Britain, Womley, Godaming, Surrey, England and Unesco, Paris 1971. Conductivity Setup Menu Features Temperature Compensation and Reference Temperature The Orion Star and Star Plus conductivity meters have the ability to use a temperature compensation feature that calculates and displays the conductivity measurements at a reference temperature of 5 °C, 10 °C, 15 °C, 20 °C or 25 °C (Orion Star meters have a reference temperature of 15 °C, 20 °C or 25 °C only). The temperature compensation can be set as linear for most aqueous samples, non-linear for ultra pure and low ionic strength samples or off for non-temperature compensated conductivity measurements. EN-66 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 71. Meter Setup Menu Features The closer the sample temperature is to the selected reference temperature, the more accurate the conductivity measurement will be, especially if the temperature compensation coefficient is estimated or inaccurate. The conductivity of a solution with a specific electrolyte concentration changes with temperature and this relationship is described by the temperature coefficient of the solution. The meter has a default temperature coefficient of 2.1 percent change in conductivity per ˚C, which is representative of many aqueous samples. Solution (25 °C to 50 °C) Temperature Coefficient (% / °C) Ultra Pure Water 4.55 Salt (NaCl) 2.12 5% NaOH 1.72 Dilute Ammonia 1.88 10% HCl 1.32 5% Sulfuric Acid 0.96 98% Sulfuric Acid 2.84 Sugar Syrup 5.64 Total Dissolved Solids (TDS) The Orion Star and Star Plus conductivity meters measure TDS as the total amount of dissolved inorganics in a solution. The dissolved inorganics carry a current that is measured by the conductivity probe. Since there is a direct relationship between conductivity and TDS, conductivity readings are used to estimate the presence of inorganics. The user must enter a TDS factor between 0.01 and 10 mg/L in the setup menu. The standard method of determining TDS involves evaporating a sample to dryness at 180 ˚C and weighing the residue. The TDS factor is calculated by taking the residue weight and dividing it by the sample conductivity. Subsequent conductivity readings are multiplied by the TDS factor to determine the TDS value of the sample. Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide EN-67
  • 72. Meter Setup Menu Features Automatic Calibration The Orion Star and Star Plus conductivity meters are capable of automatically recognizing 100 µS/cm, 1413 µS/cm and 12.9 mS/cm conductivity standards when the nominal cell constant of the conductivity probe is entered in the setup menu. For the meter to recognize the conductivity standard, the entered cell constant must be accurate within a factor of 3. For example, if the actual cell constant is 1.0 cm-1, entering a nominal cell constant in the range of 0.3 cm-1 to 3.0 cm-1 would allow the meter to identify the conductivity standard and perform the automatic calibration. ISE Setup Menu Features Concentration Range The Orion Star and Star Plus ISE meters can be set for a high or low ISE concentration range that is used to determine the calibration stability criteria. If a high ISE concentration range is selected, the meter will perform a normal calibration with no delay in displaying the calibration standard value. If a low ISE concentration range is selected, the meter will wait about three to five minutes before displaying a stable reading for the calibration standard values. The delay will depend on the species being measured and the concentration of the calibration standards. The low ISE concentration range is designed to improve the accuracy of low concentration measurements by allowing the electrode to have a longer amount of time to stabilize in the calibration standards. Automatic Blank Correction The Orion Star and Star Plus ISE meters have an ISE automatic blank correction feature that uses an algorithm to compensate for the non-linearity of the ion selective electrode in low-level standards and samples. Since the automatic blank correction feature requires the use of a set of non-linear equations that can only be calculated numerically, the user cannot analytically verify the calibration and the average slope value that is displayed on the meter may be outside of the slope range that is specified in the electrode user guide. In applications were analytical verification is required, the automatic blank correction feature should be turned off. EN-68 Thermo Scientific Orion Star™ and Star Plus Series Meter User Guide
  • 73. Environmental Instruments Water Analysis Instruments North America 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Int’l Fax: 978-232-6031 Europe P.O. Box 254, 3860 AG Nijkerk Wallerstraat 125K, 3862 BN Nijkerk, The Netherlands Tel: (31) 033-2463887 Fax: (31) 033-2460832 Asia Pacific Blk 55, Ayer Rajah Crescent #04-16/24, Singapore 139949 Tel: 65-6778-6876 Fax: 65-6773-0836 www.thermo.com/water © 2008 Thermo Fisher Scientific Inc. All rights reserved. 257250-001 Rev A 09-08