SlideShare a Scribd company logo
9
Most read
10
Most read
17
Most read
Pulse Oximeter Sensor
Our Goal
We hope to quantify the dynamics of an active
muscle.
LED spectroscopy provides a way to track fatigue
and oxygen trends of an area of interest.
pulse oximetry final
pulse oximetry final
pulse oximetry final
Properties of the Red and IR LED
• Near-IR light has much better penetration into
biological tissue than visible light
• NIR light absorption affected by: hemoglobin,
myoglobin and cytochrome c oxidase molecules
(90%Hb)
• To track specifically muscle activity, it appears
that Infrared light will be the most appropriate
signal to track
IR signal
We know from Beer-Lambert’s law that the absorption of
an LED is a function of each absorbed substance given
their:
1. extinction factor
2. concentration
3. path length.
Our IR light detects mostly Hb/Mb-deoxygenation in
muscle tissue but we cannot account for path length
using 1 LED.
Path Length
For a fixed optode (forehead) we could assume
path length to be constant.
However, we are interested in quantifying
muscle movement within exercise, and our
signal includes an initial contraction of muscle
(change of path length)
Accounting for path length
The optical path length was accounted for by
measuring the absorbance changes obtained during
cuff ischemia.
The percent deoxygenation during exercise is then
calculated from the maximal deoxygenation of the
ischemic response.
A way of normalizing NIRS signals across subjects.
Oxygen saturations
Boushel claims one must use multiple
wavelengths to generate algorithms for NIRS.
• Coefficients used at each wavelength
deconvolute the overlapping absorption
spectra.
• Algorithms are used for either trend
monitoring or to assess relative
concentrations.
Below are wavelengths and coefficients used in
the laboratory at Duke for assessing relative
concentrations.
Measurements with Two
Wavelengths
Other NIRS applications use two near-IR LEDs
(760 and 850nm) and use the difference
between the signals to determine relative
oxygen saturation. The sum of the signals to
provide a reference to blood volume changes.
Red-IR= relative O2
While I am using a
Red LED and IR LED,
the data shows
plausible oxygen
trends in that the
bicep oxygen content
desaturates with
increased load in
isometric exercise.
Infrared Trends
We changed the experiment set up
Infrared signals have shown a promising fatigue
trend in almost all of our samples.
pulse oximetry final
Analyzing IR trends
Static load in some types of activity compromise
blood flow to working muscle.
Are our IR decline trends due to decreased
blood flow, or is the use of oxygen in the muscle
exceeding a continuous supply?
Moving Forward
Ultrasound & IR signal of Nellcor sensor in
cooperation. Separate muscle influence and oxygen
Classify experiment activity, static load or active
Use multiple NIRS LEDs?
• Overlapping spectra= trend monitoring/ find
relative concentrations of each molecule.

More Related Content

PPTX
Perioperative Arrythmias and management
PPTX
Pulseoximeter and Plethysmography by Pandian M
PPT
11 capnography
PPT
Pulseoximetry
PPT
cardiac output measurment
PPTX
Pulse oximetry journal
PPT
Autonomic function tests
PPTX
ANAESTHESIA IN SPACE.pptx
Perioperative Arrythmias and management
Pulseoximeter and Plethysmography by Pandian M
11 capnography
Pulseoximetry
cardiac output measurment
Pulse oximetry journal
Autonomic function tests
ANAESTHESIA IN SPACE.pptx

What's hot (20)

PPT
Pulseoximetry fianl
PPTX
Capnography
PPT
Ivc guided fluid management in the icu
PPTX
ABG Analysis
PPTX
Monitoring in anaesthesia by stanley ogochukwu okerulu
PPTX
Body plethesmography
PPTX
High altitude Physiology
PPTX
Newer modes of ventilation
PDF
How Pulse oximeter works- Principle and Limitations I SpO2 I Respiratory Phys...
PPTX
Capnography
PPT
Systemic oxygenation
PPTX
O2 cascade flux n odc
PPTX
Pulmonary function tests
PPT
Liver function test
PPTX
Bed side pulmonary function tests 7
PPTX
Stellate ganglion block
PPT
Fick principle
PPT
Basic Of Mechanical Ventilation
PPTX
Gas laws and its implications in Anaesthesiology
PPTX
Oxygen cascade
Pulseoximetry fianl
Capnography
Ivc guided fluid management in the icu
ABG Analysis
Monitoring in anaesthesia by stanley ogochukwu okerulu
Body plethesmography
High altitude Physiology
Newer modes of ventilation
How Pulse oximeter works- Principle and Limitations I SpO2 I Respiratory Phys...
Capnography
Systemic oxygenation
O2 cascade flux n odc
Pulmonary function tests
Liver function test
Bed side pulmonary function tests 7
Stellate ganglion block
Fick principle
Basic Of Mechanical Ventilation
Gas laws and its implications in Anaesthesiology
Oxygen cascade
Ad

Viewers also liked (14)

PPTX
Pulse oximeter - SpO2
PPT
Pulseoximetry
PPT
Pulse Oximetry
PPTX
Pulse oximeter
PPTX
use of pulse oximetry
PPTX
Diagnosis of testicular torsion using near infrared spectroscopy
PPTX
Monitoring of bladder filling to capacity
PPT
Monitoring the injured brain
PPTX
Cerebral oximetry
PPT
Near Infrared Spectroscopy In Off Line Biomass Monitoring Of Candida Utilis C...
PPTX
PDF
Project Pelican Pulse Oximeter Presentations
PPTX
Pulse oximetry slidefinal abc
Pulse oximeter - SpO2
Pulseoximetry
Pulse Oximetry
Pulse oximeter
use of pulse oximetry
Diagnosis of testicular torsion using near infrared spectroscopy
Monitoring of bladder filling to capacity
Monitoring the injured brain
Cerebral oximetry
Near Infrared Spectroscopy In Off Line Biomass Monitoring Of Candida Utilis C...
Project Pelican Pulse Oximeter Presentations
Pulse oximetry slidefinal abc
Ad

Similar to pulse oximetry final (20)

PPTX
Pulse oximetry ppt
PPTX
Mr. Ali mushtaq General - pulse oximetry.pptx
PPTX
New Microsoft PowerPoint Presentation.pptx
PDF
International Journal of Computational Engineering Research(IJCER)
PPTX
An Introduction to Near Infared Spectroscopy (NIRS)
PDF
Co32559562
PPTX
LECTURE 3- PULSE OXIMETER FOR BIOMEDICALTECHNICIAN.pptx
PDF
humon lactato
PPT
photoplethysmogram
PPTX
Working of pulse oxymetry ( oxygen saturation).pptx
PDF
Near infrared spectroscopy for medical applications current status and future...
PPT
Notes on the Pulse Oximetry in medical engineering
PDF
PulseOxymetry_AN_Finalv3p1
PDF
Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based...
PPTX
Gas measurement O2,Co2,& ph) 04/2024.pptx
PPTX
Pulse oximetry.
PPTX
PULSE OXIMETRY
PPTX
RESPIRATORY MONITORING - Dr. JAGADEESH.pptx
PPTX
Pulse oximeter
DOCX
DOC-20230516..docx
Pulse oximetry ppt
Mr. Ali mushtaq General - pulse oximetry.pptx
New Microsoft PowerPoint Presentation.pptx
International Journal of Computational Engineering Research(IJCER)
An Introduction to Near Infared Spectroscopy (NIRS)
Co32559562
LECTURE 3- PULSE OXIMETER FOR BIOMEDICALTECHNICIAN.pptx
humon lactato
photoplethysmogram
Working of pulse oxymetry ( oxygen saturation).pptx
Near infrared spectroscopy for medical applications current status and future...
Notes on the Pulse Oximetry in medical engineering
PulseOxymetry_AN_Finalv3p1
Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based...
Gas measurement O2,Co2,& ph) 04/2024.pptx
Pulse oximetry.
PULSE OXIMETRY
RESPIRATORY MONITORING - Dr. JAGADEESH.pptx
Pulse oximeter
DOC-20230516..docx

pulse oximetry final

  • 2. Our Goal We hope to quantify the dynamics of an active muscle. LED spectroscopy provides a way to track fatigue and oxygen trends of an area of interest.
  • 6. Properties of the Red and IR LED • Near-IR light has much better penetration into biological tissue than visible light • NIR light absorption affected by: hemoglobin, myoglobin and cytochrome c oxidase molecules (90%Hb) • To track specifically muscle activity, it appears that Infrared light will be the most appropriate signal to track
  • 7. IR signal We know from Beer-Lambert’s law that the absorption of an LED is a function of each absorbed substance given their: 1. extinction factor 2. concentration 3. path length. Our IR light detects mostly Hb/Mb-deoxygenation in muscle tissue but we cannot account for path length using 1 LED.
  • 8. Path Length For a fixed optode (forehead) we could assume path length to be constant. However, we are interested in quantifying muscle movement within exercise, and our signal includes an initial contraction of muscle (change of path length)
  • 9. Accounting for path length The optical path length was accounted for by measuring the absorbance changes obtained during cuff ischemia. The percent deoxygenation during exercise is then calculated from the maximal deoxygenation of the ischemic response. A way of normalizing NIRS signals across subjects.
  • 10. Oxygen saturations Boushel claims one must use multiple wavelengths to generate algorithms for NIRS. • Coefficients used at each wavelength deconvolute the overlapping absorption spectra. • Algorithms are used for either trend monitoring or to assess relative concentrations.
  • 11. Below are wavelengths and coefficients used in the laboratory at Duke for assessing relative concentrations.
  • 12. Measurements with Two Wavelengths Other NIRS applications use two near-IR LEDs (760 and 850nm) and use the difference between the signals to determine relative oxygen saturation. The sum of the signals to provide a reference to blood volume changes.
  • 13. Red-IR= relative O2 While I am using a Red LED and IR LED, the data shows plausible oxygen trends in that the bicep oxygen content desaturates with increased load in isometric exercise.
  • 14. Infrared Trends We changed the experiment set up Infrared signals have shown a promising fatigue trend in almost all of our samples.
  • 16. Analyzing IR trends Static load in some types of activity compromise blood flow to working muscle. Are our IR decline trends due to decreased blood flow, or is the use of oxygen in the muscle exceeding a continuous supply?
  • 17. Moving Forward Ultrasound & IR signal of Nellcor sensor in cooperation. Separate muscle influence and oxygen Classify experiment activity, static load or active Use multiple NIRS LEDs? • Overlapping spectra= trend monitoring/ find relative concentrations of each molecule.

Editor's Notes

  • #4: The basis for our sensor are two LED signals, Red and Infrared. They have wavelengths of 660 and 910nm, respectively.
  • #5:   When there is high oxygen saturation, the IR pulse amplitude is higher, and Red pulse amplitude greater for lower oxygen saturation. A red-to-infrared pulse modulation ratio is needed to determine Oxygen saturation.
  • #6: We have a Nellcor SpO2 Forehead Sensor that was designed to track poor pulse perfusion as fast as possible as arterial blood reaches the head before the fingers. Nellcor products have monitors that account for calibration curves that correlate the pulse Modulation Ratio to oxygen saturation. Using that system, oxygen content for an area of interest is easily monitored. However, we are using a TI SP02-FE EVM dac that allows us to capture raw Infrared and Red LED signals from the sensor.
  • #7: A chromophore is the part of a molecule responsible for its color. The color arises when a molecule absorbs certain wavelengths of visible light and transmits or reflects others.
  • #10: Cut off the supply of oxygen to the area of interest and use the max deoxygenation measurement to calculate the oxygenation during exercise
  • #14: (scaled to the maximum and minimum values). However, without accounting for the influence of tissue and the change within the muscle, there exists the problem of normalizing the data between people, and determining true oxygen saturation changes. You can also see some activity in the signal with the contraction in the first few seconds, and it doesn’t seem intuitive to assume that is due to oxygen saturation changes over milliseconds.
  • #15: The experiments were taken over 5 minutes with the subject maintaining a steady force of 30% of their maximum contraction against a table top. They were seated with their arm at about 90 degrees, without any support and the maximum force was calibrated for each person. HOPES: to track fatigue over a longer time period, paying attention to IR and less on initial contraction fluctuations in the data— put aside path length
  • #18: We hope to get an ultrasound that will be able to track muscle diameter in cooperation with the Nellcor sensor’s Infrared signal. Ideally, the ultrasound will be a useful tool to provide a real-time picture of the muscle, and I think we will be able to better account for the muscle thickness in the individual signal. Something we may want to consider is using multiple Near-Infrared LEDs as they penetrate to the muscular level well. We could explore how to deconvolute the overlapping absorption spectra to trend monitoring or to assess relative concentrations. Keep in mind the type of activity we are tracking– keep in mind the intramuscular pressure and it’s affect on blood flow.