Bulletin of Electrical Engineering and Informatics
Vol. 7, No. 1, March 2018, pp. 15~20
ISSN: 2302-9285, DOI: 10.11591/eei.v7i1.711  15
Journal homepage: http://journal.portalgaruda.org/index.php/EEI/index
Respiration Monitoring System of Lung Phantom Using
Magnetic Sensor
Imamul Muttakin1
*, David Abraham2
, Rocky Alfanz3
, Rohmadi4
1,2,3
University of Sultan Ageng Tirtayasa, Jl Jenderal Sudirman Km 3, Cilegon, Indonesia
4
C-Tech Labs Edwar Technology, Jl KH Hasyim Ashari No 79A Modern Business Park, Tangerang, Indonesia
Article Info ABSTRACT
Article history:
Received Nov 12, 2017
Revised Jan 13, 2018
Accepted Jan 27, 2018
Monitoring vital signs is substantial in healthcare to assist both diagnosis and
treatment. This work proposes a means of telemonitoring system at initial
stage to observe respiratory pattern on lung phantom. Magnetic sensor
module LDC1000 is used to read inductance value of conductive material in
relation to distance variation. Therefore, respiration pattern can be observed.
In continuous mode, the inspiration inductance value is 8 uH, while
expiration is 17 uH, with stoppage is 17 uH. For static measurement, the
inspiration inductance value is 7.80 uH, while expiration is 16.46 uH and
stoppage is 16.46 uH. Those values could be further referred for vital signs
telemonitoring system design based on contactless and portable devices.
Keywords:
LDC1000
Lung phantom
Magnetic sensor
Respiration monitoring
Telemonitoring Copyright © 2018 Institute of Advanced Engineering and Science.
All rights reserved.
Corresponding Author:
Imamul Muttakin,
University of Sultan Ageng Tirtayasa,
Jl Jenderal Sudirman Km 3, Cilegon, Indonesia.
Email: imamul@untirta.ac.id
1. INTRODUCTION
Vital signs monitoring is a way to detect system change of body. It is substantial in healthcare to
assist both diagnosis and treatment. There are needs to conduct vital signs monitoring conveniently with fast
but reliable measurement and less susceptible against object or environment generated artifact. This paves
way for telemonitoring research which has been increasingly demanded in health technology [1].
The work in [2] described wearable respiratory rate monitoring. This study used piezo-resistive
fabric sensor to measure respiration. Sensors are placed on clothes and the results will be read on the monitor
via bluetooth. This tool is still less effective because a lot of interference from environment. Another
wearable system was proposed in [3] to detect respiratory rate using an electromagnetic generator. The tool is
placed outside the body on the chest and the results will be read on the monitor through cables connecting
with the appliance. It is, however, less convenient and hindering user’s mobility. Optical sensor and
accelerometer placed on mouth was constructed as real-time wireless monitoring of respiratory diseases [4].
On the other hand, pressure sensitive array attached onto bed could be used as continuous monitoring system
with limited flexibility [5]. A portable and low-cost wireless solution for respiration monitoring employs
magnetic induction sensor that could be located on chest in contactless fashion [6].
In order to accommodate complexity of procedures and precision required in biomedical research,
phantom shaped like human’s body is designed [7]. Lung phantom resembling the original lung in terms of
dimension and respiration pattern will be used. This research employs LDC1000 magnetic sensor as a
contactless sensor, with short-range sensing, low-cost, and insensitivity towards environmental interference
[8]. This work aims to leverage advancement in medical technology, particularly respiration monitoring, at
initial stage. The system uses eddy current principle that can observe lung distance while inflate and deflate
without being affected by object’s movement.
 ISSN: 2302-9285
BEEI, Vol. 7, No. 1, March 2018 : 15 – 20
16
2. RESPIRATION SYSTEM DESIGN
Conductive material being moved towards magnetic field will invoke eddy current on its surface.
This current imposes another field that can be detected as relative position of the target object against
primary field’s source (see Figure 1). Coil can be arranged as ac magnetic field exciter; and simultaneously
employed as proximity sensor [9]. Magnetic sensor converts magnetic field to an equivalent electrical signal.
Due to its accuracy and robustness, this sensor is implemented for detection of position [10].
Figure 1. AC magnetic field interacts with conductor
Since biological tissue such as skin, particularly lung, has certain conductivity characteristics, its
movement can be monitored by using electromagnetic methods. Respiration pattern is then obtained from the
inferred proximity timing read by the sensor coil.
Hardware design comprises magnetic sensor and phantom (with pumping mechanism). Air
circulation from oxygen tank is regulated by electronics controlled solenoid valve. It manages human normal
respiration pattern and air volume in the lung phantom. Meanwhile, magnetic sensor will detect phantom’s
conductive surface and provide inductance value in accordance with proximity between the two. Those
measured results are then acquired as raw data to be monitored and analyzed. In addition, there is graphical
user interface (GUI) which comes with LDC1000 module as reference assessment. The instrument design is
shown in Figure 2.
Figure 2. Respiration monitoring system block diagram
Lung phantom is fabricated resembling the real human lung (male) according to reference of
pervious research with CT (Computed Tomography) images [11]. Table 1 shows detailed dimension.
Table 1. Lung Size by CT Images
Lung Male Female Combined
Peak to Peak 8.9 ±1.1 (8.9) cm 7.7 ±0.9 (7.6) cm 8.3 ±1.2 (8.2) cm
Height
Left 21 ±2.1 (21.8) cm 19 ±2.5 (20.3) cm 19.8 ±2.6 (20) cm
Right 21 ±2.1 (21.3) cm 19 ±2.5 (19.3) cm 20.6 ±2.6 (21) cm
Max. Height
Left 28.2 ±2.2 (27.4) cm 26 ±2.7 (25.6) cm 26.1 ±2.6 (26.5) cm
Right 21 ±2.1 (21.3) cm 26 ±2.7 (26.3) cm 26.9 ±2.7 (26.9) cm
Width
Left 12.3 ±1.1 (10.6) cm 11.1 ±1 (9.7) cm 10 ±1 (10) cm
Right 12.3 ±1.1 (12.3) cm 11.1 ±1 (11.2) cm 11.6 ±1.2 (11.4) cm
Depth
Left 18 ±1.5 (18.4) cm 16.2 ±1.7 (16.2) cm 17.1 ±2 (17.1) cm
Right 18 ±1.5 (17.6) cm 16.2 ±1.7 (15.9) cm 16.9 ±1.8 (17) cm
Volume
Left 2738 ±533 (2817) cm3
1968 ±505 (2028) cm3
2301 ±636 (2262) cm3
Right 3121 ±605 (3226) cm3
2300 ±547 (2332) cm3
2663 ±667 (2583) cm3
BEEI ISSN: 2302-9285 
Respiration Monitoring System of Lung Phantom Using Magnetic Sensor (Imamul Muttaqin)
17
The phantom is made of silicon material, fiberglass, color pigment, and plastic mold. Elasticity of
silicon is good for lung phantom to inflate and deflate. As this research needs conductive surface to impose
inductance, additional thin copper sheet envelope is also assembled. Figure 3 shows lung phantom design
along with its physical laboratory prototype.
Figure 3. Lung phantom design and prototype
3. MONITORING RESULTS AND ANALYSIS
Experiment was set involving oxygen tube (air supply), solenoid valve (controlled gate), relay,
Arduino Uno (controller electronics), and lung phantom (as shown in Figure 4). Programming on
microcontroller was composed for setting solenoid valve channeling oxygen tube and lung phantom. Initially,
both of solenoid valves are closed for 3 seconds. Then, solenoid valve on oxygen tube will open for 2
seconds; while on lung phantom is still closed. After 2 seconds, solenoid valve on oxygen tube will close for
2 seconds and on lung phantom is opened. Then, both of solenoid valves will close again for 1 second. These
operations are maintained during experiment.
Figure 4. Hardware assembly
Monitoring system incorporates magnetic sensor LDC1000 connected into laptop port. Lung
phantom can inflate and deflate automatically like human’s respiration pattern. Magnetic sensor will read
inductance value and proximity of lung phantom. The result will be shown on LDC1000 EVM GUI. Figure 5
shows the setup of monitoring system.
 ISSN: 2302-9285
BEEI, Vol. 7, No. 1, March 2018 : 15 – 20
18
Figure 5. Monitoring system setup
Inductance value in relationship with proximity between sensor and phantom’s surface was
observed. Distance was ranged from 0 to 10 mm incurring linear increase and decrease of inductance when
lung phantom is deflating and inflating respectively. Figure 6 plots graphs of the mechanism.
(a)
(b)
Figure 6. Inductance vs distance lung phantom is deflating (a) and inflating (b)
BEEI ISSN: 2302-9285 
Respiration Monitoring System of Lung Phantom Using Magnetic Sensor (Imamul Muttaqin)
19
Normal respiration static data which contains inspiration and expiration (with stop) was compiled to
produce inductance graph in Figure 7. Note that horizontal axis (distance in mm) is arranged reverse mirrored
so that the pattern can be obviously seen. For continuous monitoring in real-time operation, LDC1000 EVM
GUI was used providing proximity and inductance data as captured in Figure 8.
Figure 7. Lung Phantom static data of normal respiration
Figure 8. Continuous proximity and inductance monitoring via GUI
The LDC1000 is a sensitive sensor; hence small distance change will affect the result. Air flow was
also set to the time of real lung when inflate and deflate, i.e.: two seconds for inspiration and expiration,
whereas one second for stop. Comparison between static (manual) and continuous real-time (auto) inductance
values at respiration stages is described in Table 2.
 ISSN: 2302-9285
BEEI, Vol. 7, No. 1, March 2018 : 15 – 20
20
Table 2. Comparison Continuous vs Static Experiment
Stage Continous (uH) Static (uH)
Inspiration 8 7.805094
Ekspiration 17 16.469637
Stop 17 16.469637
There is difference of inductance level between continuous and static experiment. In continuous
auto mode, the volume of air supplied into lung phantom is different from the volume of air involved in the
real human respiration. This is due to simplicity for demonstrating inflate and deflate of lung phantom. For
static manual mode, the volume of air supplied into lung phantom is similar like the volume of air in the real
human respiration.
4. CONCLUSION
Prototype for laboratory scale respiration monitoring on lung phantom using magnetic sensor has
been assembled and tested. Inductance values related to proximity between lung phantom surface and sensor
are in proportion to inflate and deflate mechanism in respiration process. Further research could be done in
manipulating abnormal respiration pattern for case study of various lung diseases. The results will be referred
for vital signs telemonitoring system design based on contactless and portable devices.
ACKNOWLEDGEMENT
Authors would like to thank members of Medical Electronics Research Laboratory (MERC),
Department of Electrical Engineering, UNTIRTA, for their valuable supports to this work.
REFERENCES
[1] Sulaiman NH, Samsuri NA, Rahim MKA, Seman FC, Inam M. Design and Analysis of Optimum Performance
Pacemaker Telemetry Antenna. Telkomnika (Telecommunication Computing Electronics and Control). 2017; 15(2):
877-882.
[2] Jeong JW, Jang YW, Lee I, Shin S, Kim S. Wearable respiratory rate monitoring using piezo-resistive fabric
sensor. World congress on medical physics and biomedical engineering. Berlin. 2009: 282-284.
[3] Padasdao B, Boric-Lubecke O. Respiratory rate detection using a wearable electromagnetic generator. Annual
International Conference of the IEEE Engineering in Medicine and Biology Society. 2011: 3217-3220.
[4] Bagchi S, Chattopadhyay M. Real-time monitoring of respiratory diseases of distantly located elderly impaired
patients. Sensing Technology: Current Status and Future Trends I. 2014: 25-40.
[5] Liu JJ, Huang MC, Xu W, Zhang X, Stevens L, Alshurafa N, Sarrafzadeh M. Breathsens: A continuous on-bed
respiratory monitoring system with torso localization using an unobtrusive pressure sensing array. IEEE journal of
biomedical and health informatics. 2015; 19(5): 1682-1688.
[6] Ojarand J, Pille S, Min M, Land R, Oleitšuk J. Magnetic induction sensor for the respiration monitoring. In Proc.
of 10th Int. Conf. on Bioelectromagnetism (icBEM). Tallinn. 2015: 1-4.
[7] Muttakin I, Yeap SY, Mansor MM, Fathil MH, Ibrahim I, Ariffin I, Supriyanto E. Low cost design of precision
medical ultrasound power measurement system. International Journal of Circuits, Systems and Signal Processing.
2011; 5(6): 672-682.
[8] LDC1000 Inductance to Digital Converter. Texas Instrument. Datasheet SNOSCX2A. 2013.
[9] Oberhauser C. LDC Target Design. Texas Instrument. Application note SNOA957A. 2016.
[10] Benabdellah A, Abbassi Z, Nakheli A. New Electromagnetic Force-Displacement Sensor. Bulletin of Electrical
Engineering and Informatics (BEEI). 2016; 5(3): 334-339.
[11] Kramer GH, Capello K, Bearrs B, Lauzon A, Normandeau L. Linear dimensions and volumes of human lungs
obtained from ct images. Health physics. 2012; 102(4): 378-383.

More Related Content

PDF
Respiratory activity sensor
PDF
Wireless Body Area Networks for Healthcare: A Survey
PDF
Development of a Respiration Rate Meter –A Low-Cost Design Approach
PPTX
Respiratory Rate Measurement
PDF
Embedded system for upper-limb exoskeleton based on electromyography control
PDF
Advances and development in biomechatronics introduction to arm prosthesis
PDF
Reconstruction of Respiratory Signal from ECG
PDF
Patient Monitoring System
Respiratory activity sensor
Wireless Body Area Networks for Healthcare: A Survey
Development of a Respiration Rate Meter –A Low-Cost Design Approach
Respiratory Rate Measurement
Embedded system for upper-limb exoskeleton based on electromyography control
Advances and development in biomechatronics introduction to arm prosthesis
Reconstruction of Respiratory Signal from ECG
Patient Monitoring System

What's hot (20)

PDF
A prosthetic limb managed by sensors-based electronic system: Experimental re...
PDF
Application of EMG and Force Signals of Elbow Joint on Robot-assisted Arm Tra...
PDF
A Wireless Physiological Monitoring System for Hyperbaric Oxygen Chamber
PDF
C04611318
PDF
Smart Phone-Based Peak Expiratory Flow Meter
PDF
NET Paper no 63-Myoelectric Prosthetic Hand with Air muscles
PDF
40220140504002
PPTX
Mech biomechatronic hand ppt
PDF
Novel method of implementing spirometer using android
PDF
Desktop Tomography System using Planar ECT Device
PDF
CALIBRATION OF INERTIAL SENSOR BY USING PARTICLE SWARM OPTIMIZATION AND HUMAN...
PDF
Bionic arm using muscle sensor v3
PDF
An effective milli kelvin thermal management strategy for infrared imaging
PDF
Application ofasub–0.1-mm3 implantable mote forinvivo real-time wireless temp...
PDF
IRJET-Pedobarography Insoles with Wireless Data Transmission
PDF
Design and Implementation of Real Time Remote Supervisory System
PDF
25 Most Interesting Medical MEMS and Sensors Projects
PPTX
Bio mechatronics
PDF
683 690,tesma412,ijeast
A prosthetic limb managed by sensors-based electronic system: Experimental re...
Application of EMG and Force Signals of Elbow Joint on Robot-assisted Arm Tra...
A Wireless Physiological Monitoring System for Hyperbaric Oxygen Chamber
C04611318
Smart Phone-Based Peak Expiratory Flow Meter
NET Paper no 63-Myoelectric Prosthetic Hand with Air muscles
40220140504002
Mech biomechatronic hand ppt
Novel method of implementing spirometer using android
Desktop Tomography System using Planar ECT Device
CALIBRATION OF INERTIAL SENSOR BY USING PARTICLE SWARM OPTIMIZATION AND HUMAN...
Bionic arm using muscle sensor v3
An effective milli kelvin thermal management strategy for infrared imaging
Application ofasub–0.1-mm3 implantable mote forinvivo real-time wireless temp...
IRJET-Pedobarography Insoles with Wireless Data Transmission
Design and Implementation of Real Time Remote Supervisory System
25 Most Interesting Medical MEMS and Sensors Projects
Bio mechatronics
683 690,tesma412,ijeast
Ad

Similar to Respiration Monitoring System of Lung Phantom Using Magnetic Sensor (20)

PDF
DIY Ventilator using Arduino with Blood Oxygen Sensing for Covid Pandemic
PDF
Novel mems respiratory flow sensor
PPTX
Smart Ventilator.pptx hzjsjjzjsjsjejdjdjeke
PDF
Ijsea 2nd publication
PDF
Ijsea 2nd publication
PDF
Ijsea 2nd publication
PDF
IRJET-Intelligent ICU Monitoring System
PDF
State feedback control for human inspiratory system
PPTX
Sheela arokia mary
PPTX
EMT PPT.....final.pptx
PDF
Medicopter as a First Aid Tool
PDF
IRJET - A Novel Health Monitoring System for Trance Patients
PDF
Dual Mode Ventilator Integrated with Patient Monitoring System
PPT
DESIGN AND IMPLEMENTATION OF EMBEDDED MONITOR SYSTEM FOR DETECTION OF A PATIE...
PDF
A review on different technical specifications of respiratory rate monitors
DOCX
Designing an Automated Cardiopulmonary Resuscitation Device
PDF
Airflow vibration of diaphragmatic breathing: model and demonstration using o...
PDF
Thesis for research on automation of heart-lung machine
PPT
Critical Equip PPT.ppt
PDF
IoT Based Smart Ventilator & Patient Monitoring System
DIY Ventilator using Arduino with Blood Oxygen Sensing for Covid Pandemic
Novel mems respiratory flow sensor
Smart Ventilator.pptx hzjsjjzjsjsjejdjdjeke
Ijsea 2nd publication
Ijsea 2nd publication
Ijsea 2nd publication
IRJET-Intelligent ICU Monitoring System
State feedback control for human inspiratory system
Sheela arokia mary
EMT PPT.....final.pptx
Medicopter as a First Aid Tool
IRJET - A Novel Health Monitoring System for Trance Patients
Dual Mode Ventilator Integrated with Patient Monitoring System
DESIGN AND IMPLEMENTATION OF EMBEDDED MONITOR SYSTEM FOR DETECTION OF A PATIE...
A review on different technical specifications of respiratory rate monitors
Designing an Automated Cardiopulmonary Resuscitation Device
Airflow vibration of diaphragmatic breathing: model and demonstration using o...
Thesis for research on automation of heart-lung machine
Critical Equip PPT.ppt
IoT Based Smart Ventilator & Patient Monitoring System
Ad

More from journalBEEI (20)

PDF
Square transposition: an approach to the transposition process in block cipher
PDF
Hyper-parameter optimization of convolutional neural network based on particl...
PDF
Supervised machine learning based liver disease prediction approach with LASS...
PDF
A secure and energy saving protocol for wireless sensor networks
PDF
Plant leaf identification system using convolutional neural network
PDF
Customized moodle-based learning management system for socially disadvantaged...
PDF
Understanding the role of individual learner in adaptive and personalized e-l...
PDF
Prototype mobile contactless transaction system in traditional markets to sup...
PDF
Wireless HART stack using multiprocessor technique with laxity algorithm
PDF
Implementation of double-layer loaded on octagon microstrip yagi antenna
PDF
The calculation of the field of an antenna located near the human head
PDF
Exact secure outage probability performance of uplinkdownlink multiple access...
PDF
Design of a dual-band antenna for energy harvesting application
PDF
Transforming data-centric eXtensible markup language into relational database...
PDF
Key performance requirement of future next wireless networks (6G)
PDF
Noise resistance territorial intensity-based optical flow using inverse confi...
PDF
Modeling climate phenomenon with software grids analysis and display system i...
PDF
An approach of re-organizing input dataset to enhance the quality of emotion ...
PDF
Parking detection system using background subtraction and HSV color segmentation
PDF
Quality of service performances of video and voice transmission in universal ...
Square transposition: an approach to the transposition process in block cipher
Hyper-parameter optimization of convolutional neural network based on particl...
Supervised machine learning based liver disease prediction approach with LASS...
A secure and energy saving protocol for wireless sensor networks
Plant leaf identification system using convolutional neural network
Customized moodle-based learning management system for socially disadvantaged...
Understanding the role of individual learner in adaptive and personalized e-l...
Prototype mobile contactless transaction system in traditional markets to sup...
Wireless HART stack using multiprocessor technique with laxity algorithm
Implementation of double-layer loaded on octagon microstrip yagi antenna
The calculation of the field of an antenna located near the human head
Exact secure outage probability performance of uplinkdownlink multiple access...
Design of a dual-band antenna for energy harvesting application
Transforming data-centric eXtensible markup language into relational database...
Key performance requirement of future next wireless networks (6G)
Noise resistance territorial intensity-based optical flow using inverse confi...
Modeling climate phenomenon with software grids analysis and display system i...
An approach of re-organizing input dataset to enhance the quality of emotion ...
Parking detection system using background subtraction and HSV color segmentation
Quality of service performances of video and voice transmission in universal ...

Recently uploaded (20)

PPTX
Safety Seminar civil to be ensured for safe working.
PPTX
CH1 Production IntroductoryConcepts.pptx
PDF
Model Code of Practice - Construction Work - 21102022 .pdf
PDF
Operating System & Kernel Study Guide-1 - converted.pdf
PDF
Unit I ESSENTIAL OF DIGITAL MARKETING.pdf
PDF
737-MAX_SRG.pdf student reference guides
PDF
Automation-in-Manufacturing-Chapter-Introduction.pdf
PPTX
FINAL REVIEW FOR COPD DIANOSIS FOR PULMONARY DISEASE.pptx
DOCX
573137875-Attendance-Management-System-original
PPTX
MET 305 2019 SCHEME MODULE 2 COMPLETE.pptx
PPTX
Foundation to blockchain - A guide to Blockchain Tech
PPTX
CYBER-CRIMES AND SECURITY A guide to understanding
PDF
The CXO Playbook 2025 – Future-Ready Strategies for C-Suite Leaders Cerebrai...
PPTX
M Tech Sem 1 Civil Engineering Environmental Sciences.pptx
PDF
Embodied AI: Ushering in the Next Era of Intelligent Systems
PDF
III.4.1.2_The_Space_Environment.p pdffdf
PPTX
Engineering Ethics, Safety and Environment [Autosaved] (1).pptx
PDF
TFEC-4-2020-Design-Guide-for-Timber-Roof-Trusses.pdf
PDF
R24 SURVEYING LAB MANUAL for civil enggi
DOCX
ASol_English-Language-Literature-Set-1-27-02-2023-converted.docx
Safety Seminar civil to be ensured for safe working.
CH1 Production IntroductoryConcepts.pptx
Model Code of Practice - Construction Work - 21102022 .pdf
Operating System & Kernel Study Guide-1 - converted.pdf
Unit I ESSENTIAL OF DIGITAL MARKETING.pdf
737-MAX_SRG.pdf student reference guides
Automation-in-Manufacturing-Chapter-Introduction.pdf
FINAL REVIEW FOR COPD DIANOSIS FOR PULMONARY DISEASE.pptx
573137875-Attendance-Management-System-original
MET 305 2019 SCHEME MODULE 2 COMPLETE.pptx
Foundation to blockchain - A guide to Blockchain Tech
CYBER-CRIMES AND SECURITY A guide to understanding
The CXO Playbook 2025 – Future-Ready Strategies for C-Suite Leaders Cerebrai...
M Tech Sem 1 Civil Engineering Environmental Sciences.pptx
Embodied AI: Ushering in the Next Era of Intelligent Systems
III.4.1.2_The_Space_Environment.p pdffdf
Engineering Ethics, Safety and Environment [Autosaved] (1).pptx
TFEC-4-2020-Design-Guide-for-Timber-Roof-Trusses.pdf
R24 SURVEYING LAB MANUAL for civil enggi
ASol_English-Language-Literature-Set-1-27-02-2023-converted.docx

Respiration Monitoring System of Lung Phantom Using Magnetic Sensor

  • 1. Bulletin of Electrical Engineering and Informatics Vol. 7, No. 1, March 2018, pp. 15~20 ISSN: 2302-9285, DOI: 10.11591/eei.v7i1.711  15 Journal homepage: http://journal.portalgaruda.org/index.php/EEI/index Respiration Monitoring System of Lung Phantom Using Magnetic Sensor Imamul Muttakin1 *, David Abraham2 , Rocky Alfanz3 , Rohmadi4 1,2,3 University of Sultan Ageng Tirtayasa, Jl Jenderal Sudirman Km 3, Cilegon, Indonesia 4 C-Tech Labs Edwar Technology, Jl KH Hasyim Ashari No 79A Modern Business Park, Tangerang, Indonesia Article Info ABSTRACT Article history: Received Nov 12, 2017 Revised Jan 13, 2018 Accepted Jan 27, 2018 Monitoring vital signs is substantial in healthcare to assist both diagnosis and treatment. This work proposes a means of telemonitoring system at initial stage to observe respiratory pattern on lung phantom. Magnetic sensor module LDC1000 is used to read inductance value of conductive material in relation to distance variation. Therefore, respiration pattern can be observed. In continuous mode, the inspiration inductance value is 8 uH, while expiration is 17 uH, with stoppage is 17 uH. For static measurement, the inspiration inductance value is 7.80 uH, while expiration is 16.46 uH and stoppage is 16.46 uH. Those values could be further referred for vital signs telemonitoring system design based on contactless and portable devices. Keywords: LDC1000 Lung phantom Magnetic sensor Respiration monitoring Telemonitoring Copyright © 2018 Institute of Advanced Engineering and Science. All rights reserved. Corresponding Author: Imamul Muttakin, University of Sultan Ageng Tirtayasa, Jl Jenderal Sudirman Km 3, Cilegon, Indonesia. Email: imamul@untirta.ac.id 1. INTRODUCTION Vital signs monitoring is a way to detect system change of body. It is substantial in healthcare to assist both diagnosis and treatment. There are needs to conduct vital signs monitoring conveniently with fast but reliable measurement and less susceptible against object or environment generated artifact. This paves way for telemonitoring research which has been increasingly demanded in health technology [1]. The work in [2] described wearable respiratory rate monitoring. This study used piezo-resistive fabric sensor to measure respiration. Sensors are placed on clothes and the results will be read on the monitor via bluetooth. This tool is still less effective because a lot of interference from environment. Another wearable system was proposed in [3] to detect respiratory rate using an electromagnetic generator. The tool is placed outside the body on the chest and the results will be read on the monitor through cables connecting with the appliance. It is, however, less convenient and hindering user’s mobility. Optical sensor and accelerometer placed on mouth was constructed as real-time wireless monitoring of respiratory diseases [4]. On the other hand, pressure sensitive array attached onto bed could be used as continuous monitoring system with limited flexibility [5]. A portable and low-cost wireless solution for respiration monitoring employs magnetic induction sensor that could be located on chest in contactless fashion [6]. In order to accommodate complexity of procedures and precision required in biomedical research, phantom shaped like human’s body is designed [7]. Lung phantom resembling the original lung in terms of dimension and respiration pattern will be used. This research employs LDC1000 magnetic sensor as a contactless sensor, with short-range sensing, low-cost, and insensitivity towards environmental interference [8]. This work aims to leverage advancement in medical technology, particularly respiration monitoring, at initial stage. The system uses eddy current principle that can observe lung distance while inflate and deflate without being affected by object’s movement.
  • 2.  ISSN: 2302-9285 BEEI, Vol. 7, No. 1, March 2018 : 15 – 20 16 2. RESPIRATION SYSTEM DESIGN Conductive material being moved towards magnetic field will invoke eddy current on its surface. This current imposes another field that can be detected as relative position of the target object against primary field’s source (see Figure 1). Coil can be arranged as ac magnetic field exciter; and simultaneously employed as proximity sensor [9]. Magnetic sensor converts magnetic field to an equivalent electrical signal. Due to its accuracy and robustness, this sensor is implemented for detection of position [10]. Figure 1. AC magnetic field interacts with conductor Since biological tissue such as skin, particularly lung, has certain conductivity characteristics, its movement can be monitored by using electromagnetic methods. Respiration pattern is then obtained from the inferred proximity timing read by the sensor coil. Hardware design comprises magnetic sensor and phantom (with pumping mechanism). Air circulation from oxygen tank is regulated by electronics controlled solenoid valve. It manages human normal respiration pattern and air volume in the lung phantom. Meanwhile, magnetic sensor will detect phantom’s conductive surface and provide inductance value in accordance with proximity between the two. Those measured results are then acquired as raw data to be monitored and analyzed. In addition, there is graphical user interface (GUI) which comes with LDC1000 module as reference assessment. The instrument design is shown in Figure 2. Figure 2. Respiration monitoring system block diagram Lung phantom is fabricated resembling the real human lung (male) according to reference of pervious research with CT (Computed Tomography) images [11]. Table 1 shows detailed dimension. Table 1. Lung Size by CT Images Lung Male Female Combined Peak to Peak 8.9 ±1.1 (8.9) cm 7.7 ±0.9 (7.6) cm 8.3 ±1.2 (8.2) cm Height Left 21 ±2.1 (21.8) cm 19 ±2.5 (20.3) cm 19.8 ±2.6 (20) cm Right 21 ±2.1 (21.3) cm 19 ±2.5 (19.3) cm 20.6 ±2.6 (21) cm Max. Height Left 28.2 ±2.2 (27.4) cm 26 ±2.7 (25.6) cm 26.1 ±2.6 (26.5) cm Right 21 ±2.1 (21.3) cm 26 ±2.7 (26.3) cm 26.9 ±2.7 (26.9) cm Width Left 12.3 ±1.1 (10.6) cm 11.1 ±1 (9.7) cm 10 ±1 (10) cm Right 12.3 ±1.1 (12.3) cm 11.1 ±1 (11.2) cm 11.6 ±1.2 (11.4) cm Depth Left 18 ±1.5 (18.4) cm 16.2 ±1.7 (16.2) cm 17.1 ±2 (17.1) cm Right 18 ±1.5 (17.6) cm 16.2 ±1.7 (15.9) cm 16.9 ±1.8 (17) cm Volume Left 2738 ±533 (2817) cm3 1968 ±505 (2028) cm3 2301 ±636 (2262) cm3 Right 3121 ±605 (3226) cm3 2300 ±547 (2332) cm3 2663 ±667 (2583) cm3
  • 3. BEEI ISSN: 2302-9285  Respiration Monitoring System of Lung Phantom Using Magnetic Sensor (Imamul Muttaqin) 17 The phantom is made of silicon material, fiberglass, color pigment, and plastic mold. Elasticity of silicon is good for lung phantom to inflate and deflate. As this research needs conductive surface to impose inductance, additional thin copper sheet envelope is also assembled. Figure 3 shows lung phantom design along with its physical laboratory prototype. Figure 3. Lung phantom design and prototype 3. MONITORING RESULTS AND ANALYSIS Experiment was set involving oxygen tube (air supply), solenoid valve (controlled gate), relay, Arduino Uno (controller electronics), and lung phantom (as shown in Figure 4). Programming on microcontroller was composed for setting solenoid valve channeling oxygen tube and lung phantom. Initially, both of solenoid valves are closed for 3 seconds. Then, solenoid valve on oxygen tube will open for 2 seconds; while on lung phantom is still closed. After 2 seconds, solenoid valve on oxygen tube will close for 2 seconds and on lung phantom is opened. Then, both of solenoid valves will close again for 1 second. These operations are maintained during experiment. Figure 4. Hardware assembly Monitoring system incorporates magnetic sensor LDC1000 connected into laptop port. Lung phantom can inflate and deflate automatically like human’s respiration pattern. Magnetic sensor will read inductance value and proximity of lung phantom. The result will be shown on LDC1000 EVM GUI. Figure 5 shows the setup of monitoring system.
  • 4.  ISSN: 2302-9285 BEEI, Vol. 7, No. 1, March 2018 : 15 – 20 18 Figure 5. Monitoring system setup Inductance value in relationship with proximity between sensor and phantom’s surface was observed. Distance was ranged from 0 to 10 mm incurring linear increase and decrease of inductance when lung phantom is deflating and inflating respectively. Figure 6 plots graphs of the mechanism. (a) (b) Figure 6. Inductance vs distance lung phantom is deflating (a) and inflating (b)
  • 5. BEEI ISSN: 2302-9285  Respiration Monitoring System of Lung Phantom Using Magnetic Sensor (Imamul Muttaqin) 19 Normal respiration static data which contains inspiration and expiration (with stop) was compiled to produce inductance graph in Figure 7. Note that horizontal axis (distance in mm) is arranged reverse mirrored so that the pattern can be obviously seen. For continuous monitoring in real-time operation, LDC1000 EVM GUI was used providing proximity and inductance data as captured in Figure 8. Figure 7. Lung Phantom static data of normal respiration Figure 8. Continuous proximity and inductance monitoring via GUI The LDC1000 is a sensitive sensor; hence small distance change will affect the result. Air flow was also set to the time of real lung when inflate and deflate, i.e.: two seconds for inspiration and expiration, whereas one second for stop. Comparison between static (manual) and continuous real-time (auto) inductance values at respiration stages is described in Table 2.
  • 6.  ISSN: 2302-9285 BEEI, Vol. 7, No. 1, March 2018 : 15 – 20 20 Table 2. Comparison Continuous vs Static Experiment Stage Continous (uH) Static (uH) Inspiration 8 7.805094 Ekspiration 17 16.469637 Stop 17 16.469637 There is difference of inductance level between continuous and static experiment. In continuous auto mode, the volume of air supplied into lung phantom is different from the volume of air involved in the real human respiration. This is due to simplicity for demonstrating inflate and deflate of lung phantom. For static manual mode, the volume of air supplied into lung phantom is similar like the volume of air in the real human respiration. 4. CONCLUSION Prototype for laboratory scale respiration monitoring on lung phantom using magnetic sensor has been assembled and tested. Inductance values related to proximity between lung phantom surface and sensor are in proportion to inflate and deflate mechanism in respiration process. Further research could be done in manipulating abnormal respiration pattern for case study of various lung diseases. The results will be referred for vital signs telemonitoring system design based on contactless and portable devices. ACKNOWLEDGEMENT Authors would like to thank members of Medical Electronics Research Laboratory (MERC), Department of Electrical Engineering, UNTIRTA, for their valuable supports to this work. REFERENCES [1] Sulaiman NH, Samsuri NA, Rahim MKA, Seman FC, Inam M. Design and Analysis of Optimum Performance Pacemaker Telemetry Antenna. Telkomnika (Telecommunication Computing Electronics and Control). 2017; 15(2): 877-882. [2] Jeong JW, Jang YW, Lee I, Shin S, Kim S. Wearable respiratory rate monitoring using piezo-resistive fabric sensor. World congress on medical physics and biomedical engineering. Berlin. 2009: 282-284. [3] Padasdao B, Boric-Lubecke O. Respiratory rate detection using a wearable electromagnetic generator. Annual International Conference of the IEEE Engineering in Medicine and Biology Society. 2011: 3217-3220. [4] Bagchi S, Chattopadhyay M. Real-time monitoring of respiratory diseases of distantly located elderly impaired patients. Sensing Technology: Current Status and Future Trends I. 2014: 25-40. [5] Liu JJ, Huang MC, Xu W, Zhang X, Stevens L, Alshurafa N, Sarrafzadeh M. Breathsens: A continuous on-bed respiratory monitoring system with torso localization using an unobtrusive pressure sensing array. IEEE journal of biomedical and health informatics. 2015; 19(5): 1682-1688. [6] Ojarand J, Pille S, Min M, Land R, Oleitšuk J. Magnetic induction sensor for the respiration monitoring. In Proc. of 10th Int. Conf. on Bioelectromagnetism (icBEM). Tallinn. 2015: 1-4. [7] Muttakin I, Yeap SY, Mansor MM, Fathil MH, Ibrahim I, Ariffin I, Supriyanto E. Low cost design of precision medical ultrasound power measurement system. International Journal of Circuits, Systems and Signal Processing. 2011; 5(6): 672-682. [8] LDC1000 Inductance to Digital Converter. Texas Instrument. Datasheet SNOSCX2A. 2013. [9] Oberhauser C. LDC Target Design. Texas Instrument. Application note SNOA957A. 2016. [10] Benabdellah A, Abbassi Z, Nakheli A. New Electromagnetic Force-Displacement Sensor. Bulletin of Electrical Engineering and Informatics (BEEI). 2016; 5(3): 334-339. [11] Kramer GH, Capello K, Bearrs B, Lauzon A, Normandeau L. Linear dimensions and volumes of human lungs obtained from ct images. Health physics. 2012; 102(4): 378-383.