SCAVENGING SYSTEM
INTRODUCTION :
Trace level - conc of anesthetic gas that is far below the one
that is needed for clinical anesthesia.
Varies greatly – depending on FGF, ventilation system , the
length of time of anesthesia, measurement site and other
variables
Degree of pollution is more during paediatric surgeries
Workplace exposures :
- Hospital-based and stand-alone operating rooms,
- Recovery rooms,
- Dental operatories,
- Veterinary facilities .
Two different classes of
chemicals:
nitrous oxide and halogenated
agents
In 1977 National Institute for
Occupational Safety and Health
(NIOSH) –
recommended exposure limits
(RELs) for both nitrous oxide and
halogenated agent
Methods of studies
Animal Investigations
Human Volunteer studies
Epidemiological studies of exposed humans
Mortality studies
Problems
Spontaneous abortions :
Epidemiological studies - Higher rates in OR personnel
than in women in other settings.
study by Viera et al. (1980), spontaneous abortion was
observed in rats at 1000 ppm or more of N2O
In summary with possible exception of N2O the conc of
halogenated agents should be some 100 times more than
what is present usually
Infertility : Studies found higher than expected rates of
involuntary infertility among exposed
Birth Defects : Studies show that conc of inhalation agents
are well above those found in even unscavenged OR
Impaired Skilled performance : many studies failed to
establish , but one study showed that neuropsychological
symptoms and tiredness were reported more by individuals
in OR that are less scavenged
Carcinogenicity : A large study found higher risk of cancer in
females than males who are exposed, but data has been
questioned.
Liver disease : recurrent hepatitis – halothane reported in
few individuals
Renal Disease
Hematological : N2O Inactivates Vit B12 – impaired DNA
sysnthesis
Cardiac Disease : higher Freq of HTN and dysrthythmias
Scavenging and Waste Anesthetic
Gases (WAGs)
SCAVENGING
Definition
 Scavenging is the collection and
removal of vented anesthetic gases
from the OR.
 Since the amount of anesthetic gas
supplied usually far exceeds the
amount necessary for the patient, OR
pollution is decreased by scavenging.
Scavenger and operating room
ventilation efficiency are the two
most important factors in reduction of
waste anesthetic gases (WAGs).
The use of scavenging devices with anesthesia delivery systems is the
most effective way to decrease waste anesthetic gases.
 An efficient scavenging system is capable of reducing ambient
concentrations of waste gases by up to 90%.
 Anesthesia machines and breathing systems delivering
halogenated hydrocarbon anesthetics and/or nitrous
oxide should not be operated unless they are equipped
with a functional scavenging system.
To be effective, a scavenging system must not leak and must control
the concentration of trace anesthetics in ambient air.
 Waste gases should not be discharged into the outside air in an area
where reentry into the building is likely.
Types
Scavenging may be active (suction applied) or passive (waste gases proceed
passively down corrugated tubing through the room ventilation exhaust
grill of the OR).
 Active systems require a means to protect the patient's airway from the
application of suction, or buildup of positive pressure.
 Passive systems require that the patient be protected from positive
pressure buildup only.
The 2 major causes of waste gas contamination in the O.R :
1. The anesthetic technique used and
2. The equipment issues
Regarding anesthesia technique the following will increase anesthetic gas
pollution when inhalant anesthetics are used:
Administering inhalant anesthetics by open drop (eg, periodically
dripping liquid volatile anesthetic onto a gauze sponge) or insufflation
(eg, delivery of a relatively high flow of anesthetic in oxygen into the
trachea or pharynx through a catheter) techniques.
Turning on flowmeters and vaporizers before attaching the breathing
system to the patient.
Allowing flowmeters and vaporizers to remain on after the patient is
disconnected from the breathing system.
Using uncuffed endotracheal tubes that do not create a completely
sealed airway or using cuffed tubes without inflating the cuff.
Disconnecting a patient from a breathing system without eliminating as
much of the residual gases as reasonably possible through the scavenging
system .
Spilling liquid anesthetic during the filling of vaporizers, especially
during an anesthetic procedure.
Poorly fitting masks.
Flushing the circuit.
Use of breathing circuits that are difficult to scavenge, such as Jackson-
Rees.
NIOSH recommendations for trace gas levels
anesthetic gas Max. TWA conc.[ppm]
Halogenated agent alone 2
Nitrous oxide 25
Halogenated gas +nitrous oxide 0.5 + 25
Dental facilities[nitrous oxide alone] 50
Components of the scavenger system:
Gas collection assembly.
Transfer tubing.
Scavenging interface .
Gas disposal tubing .
Gas disposal assembly.
1.Gas collecting assembly
Captures excess anesthetic gases
and delivers it to the transfer
tubing.
WAG are vented from anesthesia
system through either adjustable
pressure limiting valve or
ventilator relief valve.
Gas passing through these valves
accumulates in the gas collecting
assembly and is directed to the
transfer means.
2.Transfer tubing
The transfer tubes carries excess gas from gas collecting assembly to
scavenging interface.
The tubes must be either 19 or 30 mm, sometimes yellow color-coded .
The tubes should be sufficiently rigid to prevent kinking and as short as
possible to minimize the chance of occlusion.
Separate tubes from the APL valve and ventilator relief valve merge into a
single hose before they enter scavenging interface.
If the transfer means is occluded, baseline breathing circuit pressure will
increase and barotrauma can occur.
3.Scavenging interface
The scavenger interface is the most important component because it
protects the breathing circuit or ventilator from excess positive or
negative pressure.
Positive-pressure relief is mandatory, irrespective of the type of
disposable system used, to vent excess gas in case of occlusion distal to
interface.
scavenger interfaces may be open
closed
B&C closed interfaces
A&B are active systems
C is passive system
OPEN INTERFACE:
It contains no valves and is open to
the atmosphere, allowing positive and
negative pressure relief.
Open interfaces should be used only
with active disposable systems that
have a central evacuation system.
 open interfaces require a reservoir
because waste gases are intermittently
discharged in surges whereas flow
from the evacuation system is
continuous.
The efficiency of it depends on several factors:
A.The vacuum flow rate per min must equal or exceed the
minute volume of excess gases to prevent spillage.
B.Spillage will occur if the volume of a single exhaled
breath exceeds the capacity of reservoir.
Open interfaces
are safer for the patients.
CLOSED INTERFACE:
It communicates with the atmosphere through valves.
Two types of closed interfaces are commercially available:
positive pressure relief only
positive and negative pressure relief
POSITIVE PRESSURE RELIEF ONLY:
 It has a single positive pressure relief valve and is designated to be used
only with passive disposable systems.
Transfer of the waste gas from the interface to the disposable system
relies on weight or pressure of the waste gas itself because a negative
pressure evacuation system is not used.
In this system reservoir bag is not required.
Scavenging system in operating room
POSITIVE AND NEGATIVE PRESSURE RELIEF:
 It has positive and negative pressure relief valve in addition to a reservoir
bag.
It is used with active disposable systems.
The effectiveness of a closed system in preventing spillage depends on:
the rate of waste gas inflow
the evacuation flow rate
the size of the reservoir
Leakage occurs only when the reservoir bag becomes fully inflated and
pressure increases sufficiently to open the positive pressure relief valve .
In contrast, the effectiveness of a open system in preventing spillage
depends not only on the volume of the reservoir but also on the flow
characteristics within the interface.
4.Gas disposal tubing
The gas disposable tubing conducts waste gas from the Scavenging
interface to the gas disposable assembly.
It should be collapse proof and should run overhead, if possible to
minimize the chance of accidental occlusion.
5.Gas disposal assembly
It ultimately eliminates excess waste gas.
It is of two types: Active
Passive
Active assembly:
most commonly used. It uses central evacuation system.
A vacuum pump serves as mechanical flow inducing device that removes
the waste gases.
An interface with a negative pressure relief valve is mandatory because
the pressure within the system is negative.
A reservoir is very desirable and the larger the reservoir, the lower the
suction flow rate needed.
- Piped Vacuum
- Active duct system - employs flow inducing devices (fans ,
pumps , blowers etc..) to move even at low pressures
Advantages
 Convenient in large hospitals, where many machines are in
use in different locations
Disadvantages
  Vacuum system and pipe work is a major expense 
 Needle valve may need continual adjustment
Passive disposable system:
 does not use a mechanical flow inducing device.
The weight or pressure from the heavier-than-air anesthetic gases
produces flow through the system.
Positive pressure relief is mandatory, but a negative pressure relief and
reservoir are unnecessary.
Passive – simpler and less expensive but not effective
- Room ventilation system
- Nonrecirculating or – Recirculating
- Piping direct to atmosphere
- Adsorption devices
- Catalyst decomposition
Charcoal canisters :
activated charcoal
connected to the outlet of the breathing
system
removes halogenated anaesthetics by
filtration
Catalytic Decomposition :
• Catalytic decomposition can be used
to convert nitrous oxide to nitrogen
and oxygen, reducing its contribution
to the greenhouse effect
Advantages
  No set-up cost
 Mobile - moves with the machine
 
Disadvantages
  Continuing cost of replacement.
 Continuos replacement is messy
  Does not remove nitrous oxide
  Heating the canister causes the release of the
inhalational agents
Evaluation of Anesthetic Equipment
Each piece of equipment involved in the delivery of inhalant anesthetics
should be evaluated regularly to assure its function and integrity.
Procedures for checkout of anesthesia equipment, depending on
the equipment to be used, should include the following:
Status of the high-pressure system, including the oxygen supply and nitrous oxide
supply - The nitrous oxide supply should not leak when the cylinder valve
is on and the nitrous oxide flowmeter is off.
Status of the low-pressure system (flowmeter function) - A negative-pressure
leak test should be performed at the common gas outlet or the outlet of
the vaporizer immediately upstream from the breathing system.
Status of the breathing system - An appropriate leak test for a circle
system and Noncircle systems can usually be done. The quantity of
leakage can be measured by determining the flow rate of oxygen
necessary to maintain a constant pressure in the system, and the
leak rate should be less than 300 ml/min at 30 cm of h2O.
Status of the scavenging system - The scavenging system should be
properly attached at all connectors, and the appropriate vacuum
should be assured for active systems. If charcoal canisters are
employed for scavenging, they should be changed at appropriate
intervals.
Status of mechanical ventilators - Ventilators should be connected
properly to the anesthesia machine, and an absence of leaks should
be assured.
Monitoring of the Effectiveness of
Antipollution Techniques
Monitoring trace-gas concentrations in the workplace provides a
quantitative assessment of the effectiveness of a waste-gas control
program.
Measuring the concentration of anesthetic in the breathing zone of the
most heavily exposed workers is the usual procedure.
An air-monitoring program is most appropriately started after anesthesia
delivery systems have been equipped with scavenging systems and after
other techniques for minimizing waste gas pollution are in place.
 An ideal approach would include frequent air monitoring, preferably at least
semiannual evaluations.
Effectiveness
Unscavenged operating rooms show 10-70 ppm halothane, and 400-3000
ppm N2O.
 scavenging brings these levels down to 1 and 60 ppm respectively.
 Adding careful attention to leaks and technique can yield levels as low as
0.005 and 1 ppm.
HAZARDS
Scavenging system functionally extends the anesthesia circuit all the way
from the anesthesia machine to the disposable site.
Obstruction of scavenging pathway can cause excessive positive pressure in
the breathing circuit and barotrauma can occur.
 excessive vacuum applied can result in undesirable negative pressures
within breathing system.
Loss of Monitoring Input – it may mask the strong odor of a volatile
anesthetic agent, delaying recognition of an overdose
Alarm failure – neg. pressure from the scavenging system interface prevent
the bellows from collapsing when breathing system is disconnected . Low
airway pressure alarm not activated.
Avoiding waste gas exposure
Good mask fit.
Avoid unscavengeable techniques if possible (insufflation).
Prevent flow from breathing system into room air (only turn on agent and
nitrous oxide after mask is on face, turn them off before suctioning).
Washout anesthetics (into the breathing circuit) at the end of the
anesthetia.
Don’t spill liquid agent.
Use low flows .
Use cuffed tracheal tubes when possible.
Check the machine regularly for leaks.
Disconnect nitrous oxide pipeline connection at wall at the day’s end.
Total intravenous anesthesia.
THANK U….THANK U….

More Related Content

PPTX
Scavenging of waste anaesthetic gases.
PPTX
Anaesthesia Vaporizers
PPTX
scavenging system
DOCX
Ambulance service
PPTX
Airway assessment
PPTX
Pulmonary hypertension
PPTX
Safety features in anesthesia machine
PPTX
Artificial intelligence in anesthesiology by dr tushar chokshi
Scavenging of waste anaesthetic gases.
Anaesthesia Vaporizers
scavenging system
Ambulance service
Airway assessment
Pulmonary hypertension
Safety features in anesthesia machine
Artificial intelligence in anesthesiology by dr tushar chokshi

What's hot (20)

PPTX
Anaesthesia Workstation checklist and safety features
PPTX
Low pressure system in anaesthesia machine
PPTX
Bain’s circuit (Mapelson D) by Zikrullah mallick
PPTX
SAFETY FEATURES OF ANAESTHESIA MACHINE
PPTX
Double lumen tubes
PPTX
Ayre t piece
PPTX
High pressure system- Anaesthesia Machine
PPTX
Anaesthesia ventilators
PPTX
anaesthesia Breathing circuits and its classification and functional analysis
PDF
Day case anesthesia
PPTX
Breathing systems
PPTX
CIRCLE SYSTEM.pptx
PPTX
Breathing circuits
PPTX
Low flow Anesthesia system
PPTX
Circle system low flow anesthesia
PPTX
Anaesthesia gas cylinders & pipeline gas supply
PPTX
Medical gas cylinders
PPTX
The canister (the absorber)
PPTX
Rapid sequence spinal anesthesia (RSS).pptx
PPTX
Baska mask
Anaesthesia Workstation checklist and safety features
Low pressure system in anaesthesia machine
Bain’s circuit (Mapelson D) by Zikrullah mallick
SAFETY FEATURES OF ANAESTHESIA MACHINE
Double lumen tubes
Ayre t piece
High pressure system- Anaesthesia Machine
Anaesthesia ventilators
anaesthesia Breathing circuits and its classification and functional analysis
Day case anesthesia
Breathing systems
CIRCLE SYSTEM.pptx
Breathing circuits
Low flow Anesthesia system
Circle system low flow anesthesia
Anaesthesia gas cylinders & pipeline gas supply
Medical gas cylinders
The canister (the absorber)
Rapid sequence spinal anesthesia (RSS).pptx
Baska mask
Ad

Viewers also liked (7)

PPTX
Muscle relaxant
PPTX
Anesthesia Ventilator
 
PPT
13236530 Anesthesia Pharmacology
PPTX
Face masks, laryngeal tube, airways yuvaraj
PPT
Breathing systems (2)
PPTX
hypertension anesthesia, general management. antihypertensive pharmacology
PPTX
Anaesthesia machine
Muscle relaxant
Anesthesia Ventilator
 
13236530 Anesthesia Pharmacology
Face masks, laryngeal tube, airways yuvaraj
Breathing systems (2)
hypertension anesthesia, general management. antihypertensive pharmacology
Anaesthesia machine
Ad

Similar to Scavenging system in operating room (20)

PPTX
Scavanging system
PPTX
scavenging system and hazards
PDF
Breathing systems - Mapleson Classification
PPT
Vets 238 Anesthetic Equipment Final
PDF
Anesthesia machine basics_V1.0_EN讲 - Copy.pdf
PPTX
Low Pressure System anaesthesia apparatus. pptx
PPTX
Anaesthetic Machines and Breathing Systems .pptx
PPTX
Máquina de Anestesia
PPTX
Anesthestic Breathing Systems by Dr. Mohammad abdeljawad
PPTX
Trace anesthetic exposure
PPTX
Anesthesia machine
PPSX
Good containment practices
PPTX
breathing circuits (1) (1).pptx BY MOHAMED ANWER RIFKY
PPTX
Anesthesia Machine.pptx
PDF
Brochure MedAir AHU & HVAC Services
PPTX
Trouble Shooting of Equipment.pptx
PPTX
PIPELINE.pptx
PPTX
Leak Tests.pptx
PDF
Pharmaceutical Engineering Downflow Booth Testing
Scavanging system
scavenging system and hazards
Breathing systems - Mapleson Classification
Vets 238 Anesthetic Equipment Final
Anesthesia machine basics_V1.0_EN讲 - Copy.pdf
Low Pressure System anaesthesia apparatus. pptx
Anaesthetic Machines and Breathing Systems .pptx
Máquina de Anestesia
Anesthestic Breathing Systems by Dr. Mohammad abdeljawad
Trace anesthetic exposure
Anesthesia machine
Good containment practices
breathing circuits (1) (1).pptx BY MOHAMED ANWER RIFKY
Anesthesia Machine.pptx
Brochure MedAir AHU & HVAC Services
Trouble Shooting of Equipment.pptx
PIPELINE.pptx
Leak Tests.pptx
Pharmaceutical Engineering Downflow Booth Testing

More from Dr Kumar (20)

PPTX
Anaesthesia for robotic surgery
PPTX
Telemedicine in Anaesthesia
PPTX
TOURNIQUET AND ITS EFFECTS
PPTX
Supra glottic airway children
PPTX
shock
PPT
Renal replacement therapy
PPT
Pheochromocytoma and its anaesthetic management
PPTX
Neuromuscular weakness related to critical illness
PPTX
Mechanisms of cerebral injury and cerebral protection
PPTX
Latex allergy and its management
PPTX
Intra operative hypoxia and hypercarbia
PPTX
Hypotensive anesthesia
PPTX
Hypertension , crf post renal transplant patient for surgery
PPTX
Extubation problems and its management
PPTX
Effects of anaesthesia on immune system
PPTX
Buffers in the body
PPTX
Anaesthetic management of conjoined twins’
PPTX
Anaesthesia for interventional neuroradiology
PPT
Ambulatory anaesthesia
PPTX
Acute pain and its management
Anaesthesia for robotic surgery
Telemedicine in Anaesthesia
TOURNIQUET AND ITS EFFECTS
Supra glottic airway children
shock
Renal replacement therapy
Pheochromocytoma and its anaesthetic management
Neuromuscular weakness related to critical illness
Mechanisms of cerebral injury and cerebral protection
Latex allergy and its management
Intra operative hypoxia and hypercarbia
Hypotensive anesthesia
Hypertension , crf post renal transplant patient for surgery
Extubation problems and its management
Effects of anaesthesia on immune system
Buffers in the body
Anaesthetic management of conjoined twins’
Anaesthesia for interventional neuroradiology
Ambulatory anaesthesia
Acute pain and its management

Recently uploaded (20)

PPTX
NEET PG 2025: Memory-Based Recall Questions Compiled by Dr. Shivankan Kakkar, MD
PPT
Breast Cancer management for medicsl student.ppt
PPTX
post stroke aphasia rehabilitation physician
PPTX
CME 2 Acute Chest Pain preentation for education
PDF
Human Health And Disease hggyutgghg .pdf
PPTX
15.MENINGITIS AND ENCEPHALITIS-elias.pptx
PPTX
CEREBROVASCULAR DISORDER.POWERPOINT PRESENTATIONx
PDF
Therapeutic Potential of Citrus Flavonoids in Metabolic Inflammation and Ins...
PPT
genitourinary-cancers_1.ppt Nursing care of clients with GU cancer
PPTX
Pathophysiology And Clinical Features Of Peripheral Nervous System .pptx
PPT
1b - INTRODUCTION TO EPIDEMIOLOGY (comm med).ppt
DOCX
NEET PG 2025 | Pharmacology Recall: 20 High-Yield Questions Simplified
PPTX
Fundamentals of human energy transfer .pptx
PPTX
LUNG ABSCESS - respiratory medicine - ppt
PPTX
NEET PG 2025 Pharmacology Recall | Real Exam Questions from 3rd August with D...
PPT
CHAPTER FIVE. '' Association in epidemiological studies and potential errors
PPTX
Note on Abortion.pptx for the student note
PPTX
POLYCYSTIC OVARIAN SYNDROME.pptx by Dr( med) Charles Amoateng
PDF
Khadir.pdf Acacia catechu drug Ayurvedic medicine
PPTX
SKIN Anatomy and physiology and associated diseases
NEET PG 2025: Memory-Based Recall Questions Compiled by Dr. Shivankan Kakkar, MD
Breast Cancer management for medicsl student.ppt
post stroke aphasia rehabilitation physician
CME 2 Acute Chest Pain preentation for education
Human Health And Disease hggyutgghg .pdf
15.MENINGITIS AND ENCEPHALITIS-elias.pptx
CEREBROVASCULAR DISORDER.POWERPOINT PRESENTATIONx
Therapeutic Potential of Citrus Flavonoids in Metabolic Inflammation and Ins...
genitourinary-cancers_1.ppt Nursing care of clients with GU cancer
Pathophysiology And Clinical Features Of Peripheral Nervous System .pptx
1b - INTRODUCTION TO EPIDEMIOLOGY (comm med).ppt
NEET PG 2025 | Pharmacology Recall: 20 High-Yield Questions Simplified
Fundamentals of human energy transfer .pptx
LUNG ABSCESS - respiratory medicine - ppt
NEET PG 2025 Pharmacology Recall | Real Exam Questions from 3rd August with D...
CHAPTER FIVE. '' Association in epidemiological studies and potential errors
Note on Abortion.pptx for the student note
POLYCYSTIC OVARIAN SYNDROME.pptx by Dr( med) Charles Amoateng
Khadir.pdf Acacia catechu drug Ayurvedic medicine
SKIN Anatomy and physiology and associated diseases

Scavenging system in operating room

  • 1. Presentor : Dr.T.Kumar Moderator : Dr.Venkatesh (Asst. professor) SCAVENGING SYSTEM
  • 2. INTRODUCTION : Trace level - conc of anesthetic gas that is far below the one that is needed for clinical anesthesia. Varies greatly – depending on FGF, ventilation system , the length of time of anesthesia, measurement site and other variables Degree of pollution is more during paediatric surgeries
  • 3. Workplace exposures : - Hospital-based and stand-alone operating rooms, - Recovery rooms, - Dental operatories, - Veterinary facilities .
  • 4. Two different classes of chemicals: nitrous oxide and halogenated agents In 1977 National Institute for Occupational Safety and Health (NIOSH) – recommended exposure limits (RELs) for both nitrous oxide and halogenated agent
  • 5. Methods of studies Animal Investigations Human Volunteer studies Epidemiological studies of exposed humans Mortality studies
  • 6. Problems Spontaneous abortions : Epidemiological studies - Higher rates in OR personnel than in women in other settings. study by Viera et al. (1980), spontaneous abortion was observed in rats at 1000 ppm or more of N2O In summary with possible exception of N2O the conc of halogenated agents should be some 100 times more than what is present usually
  • 7. Infertility : Studies found higher than expected rates of involuntary infertility among exposed Birth Defects : Studies show that conc of inhalation agents are well above those found in even unscavenged OR Impaired Skilled performance : many studies failed to establish , but one study showed that neuropsychological symptoms and tiredness were reported more by individuals in OR that are less scavenged
  • 8. Carcinogenicity : A large study found higher risk of cancer in females than males who are exposed, but data has been questioned. Liver disease : recurrent hepatitis – halothane reported in few individuals Renal Disease Hematological : N2O Inactivates Vit B12 – impaired DNA sysnthesis Cardiac Disease : higher Freq of HTN and dysrthythmias
  • 9. Scavenging and Waste Anesthetic Gases (WAGs) SCAVENGING Definition  Scavenging is the collection and removal of vented anesthetic gases from the OR.  Since the amount of anesthetic gas supplied usually far exceeds the amount necessary for the patient, OR pollution is decreased by scavenging. Scavenger and operating room ventilation efficiency are the two most important factors in reduction of waste anesthetic gases (WAGs).
  • 10. The use of scavenging devices with anesthesia delivery systems is the most effective way to decrease waste anesthetic gases.  An efficient scavenging system is capable of reducing ambient concentrations of waste gases by up to 90%.  Anesthesia machines and breathing systems delivering halogenated hydrocarbon anesthetics and/or nitrous oxide should not be operated unless they are equipped with a functional scavenging system. To be effective, a scavenging system must not leak and must control the concentration of trace anesthetics in ambient air.  Waste gases should not be discharged into the outside air in an area where reentry into the building is likely.
  • 11. Types Scavenging may be active (suction applied) or passive (waste gases proceed passively down corrugated tubing through the room ventilation exhaust grill of the OR).  Active systems require a means to protect the patient's airway from the application of suction, or buildup of positive pressure.  Passive systems require that the patient be protected from positive pressure buildup only.
  • 12. The 2 major causes of waste gas contamination in the O.R : 1. The anesthetic technique used and 2. The equipment issues Regarding anesthesia technique the following will increase anesthetic gas pollution when inhalant anesthetics are used: Administering inhalant anesthetics by open drop (eg, periodically dripping liquid volatile anesthetic onto a gauze sponge) or insufflation (eg, delivery of a relatively high flow of anesthetic in oxygen into the trachea or pharynx through a catheter) techniques. Turning on flowmeters and vaporizers before attaching the breathing system to the patient. Allowing flowmeters and vaporizers to remain on after the patient is disconnected from the breathing system.
  • 13. Using uncuffed endotracheal tubes that do not create a completely sealed airway or using cuffed tubes without inflating the cuff. Disconnecting a patient from a breathing system without eliminating as much of the residual gases as reasonably possible through the scavenging system . Spilling liquid anesthetic during the filling of vaporizers, especially during an anesthetic procedure. Poorly fitting masks. Flushing the circuit. Use of breathing circuits that are difficult to scavenge, such as Jackson- Rees.
  • 14. NIOSH recommendations for trace gas levels anesthetic gas Max. TWA conc.[ppm] Halogenated agent alone 2 Nitrous oxide 25 Halogenated gas +nitrous oxide 0.5 + 25 Dental facilities[nitrous oxide alone] 50
  • 15. Components of the scavenger system: Gas collection assembly. Transfer tubing. Scavenging interface . Gas disposal tubing . Gas disposal assembly.
  • 16. 1.Gas collecting assembly Captures excess anesthetic gases and delivers it to the transfer tubing. WAG are vented from anesthesia system through either adjustable pressure limiting valve or ventilator relief valve. Gas passing through these valves accumulates in the gas collecting assembly and is directed to the transfer means.
  • 17. 2.Transfer tubing The transfer tubes carries excess gas from gas collecting assembly to scavenging interface. The tubes must be either 19 or 30 mm, sometimes yellow color-coded . The tubes should be sufficiently rigid to prevent kinking and as short as possible to minimize the chance of occlusion. Separate tubes from the APL valve and ventilator relief valve merge into a single hose before they enter scavenging interface. If the transfer means is occluded, baseline breathing circuit pressure will increase and barotrauma can occur.
  • 18. 3.Scavenging interface The scavenger interface is the most important component because it protects the breathing circuit or ventilator from excess positive or negative pressure. Positive-pressure relief is mandatory, irrespective of the type of disposable system used, to vent excess gas in case of occlusion distal to interface. scavenger interfaces may be open closed
  • 19. B&C closed interfaces A&B are active systems C is passive system
  • 20. OPEN INTERFACE: It contains no valves and is open to the atmosphere, allowing positive and negative pressure relief. Open interfaces should be used only with active disposable systems that have a central evacuation system.  open interfaces require a reservoir because waste gases are intermittently discharged in surges whereas flow from the evacuation system is continuous.
  • 21. The efficiency of it depends on several factors: A.The vacuum flow rate per min must equal or exceed the minute volume of excess gases to prevent spillage. B.Spillage will occur if the volume of a single exhaled breath exceeds the capacity of reservoir. Open interfaces are safer for the patients.
  • 22. CLOSED INTERFACE: It communicates with the atmosphere through valves. Two types of closed interfaces are commercially available: positive pressure relief only positive and negative pressure relief POSITIVE PRESSURE RELIEF ONLY:  It has a single positive pressure relief valve and is designated to be used only with passive disposable systems. Transfer of the waste gas from the interface to the disposable system relies on weight or pressure of the waste gas itself because a negative pressure evacuation system is not used. In this system reservoir bag is not required.
  • 24. POSITIVE AND NEGATIVE PRESSURE RELIEF:  It has positive and negative pressure relief valve in addition to a reservoir bag. It is used with active disposable systems. The effectiveness of a closed system in preventing spillage depends on: the rate of waste gas inflow the evacuation flow rate the size of the reservoir Leakage occurs only when the reservoir bag becomes fully inflated and pressure increases sufficiently to open the positive pressure relief valve . In contrast, the effectiveness of a open system in preventing spillage depends not only on the volume of the reservoir but also on the flow characteristics within the interface.
  • 25. 4.Gas disposal tubing The gas disposable tubing conducts waste gas from the Scavenging interface to the gas disposable assembly. It should be collapse proof and should run overhead, if possible to minimize the chance of accidental occlusion.
  • 26. 5.Gas disposal assembly It ultimately eliminates excess waste gas. It is of two types: Active Passive Active assembly: most commonly used. It uses central evacuation system. A vacuum pump serves as mechanical flow inducing device that removes the waste gases. An interface with a negative pressure relief valve is mandatory because the pressure within the system is negative. A reservoir is very desirable and the larger the reservoir, the lower the suction flow rate needed.
  • 27. - Piped Vacuum - Active duct system - employs flow inducing devices (fans , pumps , blowers etc..) to move even at low pressures Advantages  Convenient in large hospitals, where many machines are in use in different locations Disadvantages   Vacuum system and pipe work is a major expense   Needle valve may need continual adjustment
  • 28. Passive disposable system:  does not use a mechanical flow inducing device. The weight or pressure from the heavier-than-air anesthetic gases produces flow through the system. Positive pressure relief is mandatory, but a negative pressure relief and reservoir are unnecessary. Passive – simpler and less expensive but not effective - Room ventilation system - Nonrecirculating or – Recirculating - Piping direct to atmosphere - Adsorption devices - Catalyst decomposition
  • 29. Charcoal canisters : activated charcoal connected to the outlet of the breathing system removes halogenated anaesthetics by filtration Catalytic Decomposition : • Catalytic decomposition can be used to convert nitrous oxide to nitrogen and oxygen, reducing its contribution to the greenhouse effect
  • 30. Advantages   No set-up cost  Mobile - moves with the machine   Disadvantages   Continuing cost of replacement.  Continuos replacement is messy   Does not remove nitrous oxide   Heating the canister causes the release of the inhalational agents
  • 31. Evaluation of Anesthetic Equipment Each piece of equipment involved in the delivery of inhalant anesthetics should be evaluated regularly to assure its function and integrity. Procedures for checkout of anesthesia equipment, depending on the equipment to be used, should include the following: Status of the high-pressure system, including the oxygen supply and nitrous oxide supply - The nitrous oxide supply should not leak when the cylinder valve is on and the nitrous oxide flowmeter is off. Status of the low-pressure system (flowmeter function) - A negative-pressure leak test should be performed at the common gas outlet or the outlet of the vaporizer immediately upstream from the breathing system.
  • 32. Status of the breathing system - An appropriate leak test for a circle system and Noncircle systems can usually be done. The quantity of leakage can be measured by determining the flow rate of oxygen necessary to maintain a constant pressure in the system, and the leak rate should be less than 300 ml/min at 30 cm of h2O. Status of the scavenging system - The scavenging system should be properly attached at all connectors, and the appropriate vacuum should be assured for active systems. If charcoal canisters are employed for scavenging, they should be changed at appropriate intervals. Status of mechanical ventilators - Ventilators should be connected properly to the anesthesia machine, and an absence of leaks should be assured.
  • 33. Monitoring of the Effectiveness of Antipollution Techniques Monitoring trace-gas concentrations in the workplace provides a quantitative assessment of the effectiveness of a waste-gas control program. Measuring the concentration of anesthetic in the breathing zone of the most heavily exposed workers is the usual procedure. An air-monitoring program is most appropriately started after anesthesia delivery systems have been equipped with scavenging systems and after other techniques for minimizing waste gas pollution are in place.  An ideal approach would include frequent air monitoring, preferably at least semiannual evaluations.
  • 34. Effectiveness Unscavenged operating rooms show 10-70 ppm halothane, and 400-3000 ppm N2O.  scavenging brings these levels down to 1 and 60 ppm respectively.  Adding careful attention to leaks and technique can yield levels as low as 0.005 and 1 ppm.
  • 35. HAZARDS Scavenging system functionally extends the anesthesia circuit all the way from the anesthesia machine to the disposable site. Obstruction of scavenging pathway can cause excessive positive pressure in the breathing circuit and barotrauma can occur.  excessive vacuum applied can result in undesirable negative pressures within breathing system. Loss of Monitoring Input – it may mask the strong odor of a volatile anesthetic agent, delaying recognition of an overdose Alarm failure – neg. pressure from the scavenging system interface prevent the bellows from collapsing when breathing system is disconnected . Low airway pressure alarm not activated.
  • 36. Avoiding waste gas exposure Good mask fit. Avoid unscavengeable techniques if possible (insufflation). Prevent flow from breathing system into room air (only turn on agent and nitrous oxide after mask is on face, turn them off before suctioning). Washout anesthetics (into the breathing circuit) at the end of the anesthetia. Don’t spill liquid agent. Use low flows . Use cuffed tracheal tubes when possible. Check the machine regularly for leaks. Disconnect nitrous oxide pipeline connection at wall at the day’s end. Total intravenous anesthesia.