www.s-can.at
spectro::lyser V3
	
∙ measuring principle: UV-Vis spectrometry over the total
range (190-750 nm)
	
∙ web server on board - IoT enabled, no user software is
needed to configure the probe
	
∙ communicates directly with your mobile device via WLAN
	
∙ choose exactly the parameters you want to measure –
unlimited number of parameters possible
	
∙ 8 GB onboard memory - capacity for logging data for many
years
	
∙ improved optical performance - revolutionary precision
	
∙ fast measurement interval - every 10 seconds possible
	
∙ extremely power efficient - sleep mode for low energy
consumption
	
∙ multiparameter probe with 1 mm, 5 mm or 35 mm optical
path length, ideal for waste water, surface water and
drinking water
	
∙ long term stable and maintenance free in operation
	
∙ factory precalibrated, local multi-point calibration possible
	
∙ automatic cleaning with compressed air or brush/ruck::sack
44
42
5
266,5
457
44,5
~
Messgeräte Sonstige Daten
recommended accessories
part number article name
B-32-xxx s::can compressor
B-33-012 con::nect V3
B-44
B-44-2
cleaning valve
C-32-V3 Adapter cable to connect a V3 spectrometer (M12) to V2
Terminal (MIL Plug)
D-330-xxx con::cube V3
F-110-V3 carrier s::can spectrometer V3 & V2 probe, 45°
F-48-V3 spectrometer V3 & V2 flow-cell (bypass setup), PVC
S-11-xx-moni moni::tool Software
spectro::lyser® UV-Vis monitors depending on the application an
individual selection of: TSS, TS, turbidity, color, TOC, DOC, BOD,
COD, NO3-N, NO3, HS-, O3, CLD, UV254, fingerprints, spectral alarms
and temperature
© s::can GmbH (2020)
www.s-can.at
technical specification
measuring principle UV-Vis spectrometry 190 - 750 nm
measuring principle detail xenon flash lamp, pixel array
detector
measurement interval 10 sec (configurable, depending on
application)
automatic compensation instrument real dual beam measurement
for compensation and detailed
diagnostics
automatic compensation cross
sensitivities
turbidity / solids / organic
substances
precalibrated ex-works all parameters
accuracy standard solution (>1 mg/l) NO3-N: +/- 2% +1/OPL[mg/l]*
COD-KHP: +/-2% +10/OPL[mg/l]*
(* OPL … optical pathlength in mm)
access to raw signals access to spectral information
reference standard distilled water
onboard memory 8 GB
integrated temperature sensor 0 ... 45 °C
resolution temperature sensor 0.1 °C
integration via con::cube V3
con::nect V3
con::lyte V5 (D-320-pro2) and
adapter cable (C-32-V3)
power supply 10 ... 18 VDC
power consumption (typical) 3 W
power consumption (sleep model) 60 mW
power consumption (max.) 20 W
interface to s::can terminals M12 RSTS 8Y (IP67), RS485,
Ethernet
interface to third party terminals con::nect V3 incl. Modbus RTU,
REST API, Modbus TCP/IP
digital interface (for cleaning
devices)
1 digital in/out
1 digital out
network connection 100Base-T Ethernet, WLAN
status information RGB LED ring
internal sensors supply voltage sensor, tilt sensor,
rotation sensor
cable length 1 m fixed cable (-010) or
7.5 m fixed cable (-075) or
15 m fixed cable (-150)
cable type PU jacket
housing material stainless steel 1.4404
window material optical path length 5 and 1 mm:
sapphire
optical path length 35 mm:
fused silica (UV-grade)
weight (min.) 3.4 kg (incl. cable)
dimensions (Ø x l) optical path length 35 mm:
44 x 473 mm / 517.5 mm
optical path length 5 mm:
44 x 457 mm / 501.5 mm
optical path length 1 mm:
44 x 453 mm / 497.5 mm
operating temperature 0 ... 45 °C
operating pressure 0 ... 3 bar
high pressure specification
(optional)
10 bar
installation / mounting submersed or in a flow cell
flow velocity 3 m/s (max.)
mechanical stability 30 Nm
ingress protection class IP68
automatic cleaning media: compressed air or autobrush
permissible pressure: 3 ... 6 bar
storage temperature -10 ... 65 °C
conformity - environmental testing EN 60721-3
conformity - EMC EN 61326-1
conformity - RoHS 2 EN 50581
standard warranty 2 years
extended warranty (optional) 3 years
© s::can GmbH (2020)
www.s-can.at
© s::can GmbH (2020)
www.s-can.at
municipal WWTP influent & sewer
parameter
TSS
[mg/l]
color (app)
[Hazen]
color (tru)
[Hazen]
TOC
[mg/l]
DOC
[mg/l]
BOD
[mg/l]
COD
[mg/l]
COD f
[mg/l]
NO3-N
[mg/l]
HS-
[mg/l]
UV254
[Abs/m]
UV254 f
[Abs/m]
part number
spectro::lyser™ V3
(1 mm OPL, UV-Vis)
min. 0 0 0 0 0 0 0 0 0 0 0 0 SP3-1-01-NO-xxx
max. 8000 23000 14000 3300 2600 5300 10000 5300 100 80 3300 2800
spectro::lyser™ V3
(5 mm OPL, UV-Vis)
min. 0 0 0 0 0 0 0 0 0 0 0 0 SP3-1-05-NO-xxx
max. 1200 3500 2100 500 400 800 1500 800 16 12 500 420
municipal WWTP aeration
parameter
TS
[g/l]
COD f
[mg/l]
NO3-N
[mg/l]
UV254
[Abs/m]
UV254 f
[Abs/m]
part number
spectro::lyser™ V3
(1 mm OPL, UV-Vis)
min. 0 0 0 0 0 SP3-1-01-NO-xxx
max. 20 530 26 3300 2800
municipal WWTP effluent
parameter
TSS
[mg/l]
turbidity
[NTU/
FTU]
color
(app)
[Hazen]
color
(tru)
[Hazen]
TOC
[mg/l]
DOC
[mg/l]
BOD
[mg/l]
COD
[mg/l]
COD f
[mg/l]
NO3-N
[mg/l]
O3
[mg/l]
UV254
[Abs/m]
UV254
f
[Abs/m]
part number
spectro::lyser™ V3
(1 mm OPL, UV-Vis)
min. 0 0 0 0 0 0 0 0 0 0 0 0 0 SP3-1-01-NO-xxx
max. 4000 8000 23000 14000 2600 2000 2000 3300 2000 300 1200 3300 2800
spectro::lyser™ V3
(5 mm OPL, UV-Vis)
min. 0 0 0 0 0 0 0 0 0 0 0 0 0 SP3-1-05-NO-xxx
max. 600 1200 3500 2100 400 300 300 500 300 45 180 500 420
paper mill WWTP influent
parameter
TSS
[mg/l]
COD
[mg/l]
COD f
[mg/l]
NO3-N
[mg/l]
UV254
[Abs/m]
UV254 f
[Abs/m]
part number
spectro::lyser™ V3
(1 mm OPL, UV-Vis)
min. 0 0 0 0 0 0 SP3-1-01-NO-xxx
max. 8000 13000 11000 100 3300 2800
spectro::lyser™ V3
(5 mm OPL, UV-Vis)
min. 0 0 0 0 0 0 SP3-1-05-NO-xxx
max. 1200 2000 1700 16 500 420
paper mill WWTP effluent
parameter
TSS
[mg/l]
COD
[mg/l]
COD f
[mg/l]
NO3-N
[mg/l]
UV254
[Abs/m]
UV254 f
[Abs/m]
part number
spectro::lyser™ V3
(1 mm OPL, UV-Vis)
min. 0 0 0 0 0 0 SP3-1-01-NO-xxx
max. 4000 5300 3300 100 3300 2800
spectro::lyser™ V3
(5 mm OPL, UV-Vis)
min. 0 0 0 0 0 0 SP3-1-05-NO-xxx
max. 600 790 490 16 500 420
brewery WWTP influent
parameter
TSS
[mg/l]
COD
[mg/l]
COD f
[mg/l]
NO3-N
[mg/l]
UV254
[Abs/m]
UV254 f
[Abs/m]
part number
spectro::lyser™ V3
(1 mm OPL, UV-Vis)
min. 0 0 0 0 0 0 SP3-1-01-NO-xxx
max. 13000 60000 53000 100 3300 2800
spectro::lyser™ V3
(5 mm OPL, UV-Vis)
min. 0 0 0 0 0 0 SP3-1-05-NO-xxx
max. 2000 9000 7900 16 500 420
dairy WWTP influent
parameter
TSS
[mg/l]
COD
[mg/l]
COD f
[mg/l]
NO3-N
[mg/l]
UV254
[Abs/m]
UV254 f
[Abs/m]
part number
spectro::lyser™ V3
(1 mm OPL, UV-Vis)
min. 0 0 0 0 0 0 SP3-1-01-NO-xxx
max. 8000 33000 16000 210 3300 2800

More Related Content

PDF
12.Eksploatowanie maszyn do drukowania wypukłego
DOCX
календарне 3 клас лобова
PDF
Multilyser v3 ww_en_wastewater multi parameter sensor_scan
PDF
Carbolyser v3 ww_en - sensor for online monitoring station - S::CAN
PDF
Spectrolyser v3 dw_en_multi parameter_scan
PDF
Nitrolyser v3 ww_en_no3 tss parameters sensor_scan
PDF
Spectrolyser ww en_multi parameter_scan
PDF
Multilyser v3 dw_en_water multi parameter sensor_scan
12.Eksploatowanie maszyn do drukowania wypukłego
календарне 3 клас лобова
Multilyser v3 ww_en_wastewater multi parameter sensor_scan
Carbolyser v3 ww_en - sensor for online monitoring station - S::CAN
Spectrolyser v3 dw_en_multi parameter_scan
Nitrolyser v3 ww_en_no3 tss parameters sensor_scan
Spectrolyser ww en_multi parameter_scan
Multilyser v3 dw_en_water multi parameter sensor_scan

Similar to Spectrolyser v3 ww_en_multi parameter_scan (20)

PDF
Uvlyser v3 ww_en_tss uv254 in water_scan
PDF
Nitrolyser v3 dw_en_no3 turbidity parameters sensor_scan
PDF
Uvlyser v3 dw_en_turbidity uv254 in water_scan
PDF
Ozolyser v3 ww_en_ozone tss sensor_scan
PDF
Ozolyser v3 dw_en_ozone turbidity sensor_scan
PDF
Spectrolyser dw en_multi parameter_scan
PDF
Spectrolyser industrial ww_en_multi parameter_scan
PDF
Spectrolyser titanium dw_en_multi parameter_scan
PDF
Spectrolyser titanium ww_en_multi parameter_scan
PDF
CAS74 stip scan cam74-cas74-endress+hauser datasheet-BOD sensor
PDF
Redolyser dw en_orp temp sensor_scan
PDF
Iscan dw en_turbidity sensor_scan
PDF
UV400-ATEX online water analyser-tethys datasheet
PDF
FOT Catalog 2016
PDF
FOT Catalog 2016
PDF
Coriolis rct1000 datasheet badger meter rct1000 with rcsguatda.com/cmx.p018...300 sensors
PDF
Iscan ww en_cod tss color sensor_scan
PDF
Redolyser ww en_orp temp sensor_scan
PDF
Drager Polytron 5700 Fixed Gas Detector - Spec Sheet
PDF
Oxilyser ww en_oxygen do sensor_scan
Uvlyser v3 ww_en_tss uv254 in water_scan
Nitrolyser v3 dw_en_no3 turbidity parameters sensor_scan
Uvlyser v3 dw_en_turbidity uv254 in water_scan
Ozolyser v3 ww_en_ozone tss sensor_scan
Ozolyser v3 dw_en_ozone turbidity sensor_scan
Spectrolyser dw en_multi parameter_scan
Spectrolyser industrial ww_en_multi parameter_scan
Spectrolyser titanium dw_en_multi parameter_scan
Spectrolyser titanium ww_en_multi parameter_scan
CAS74 stip scan cam74-cas74-endress+hauser datasheet-BOD sensor
Redolyser dw en_orp temp sensor_scan
Iscan dw en_turbidity sensor_scan
UV400-ATEX online water analyser-tethys datasheet
FOT Catalog 2016
FOT Catalog 2016
Coriolis rct1000 datasheet badger meter rct1000 with rcsguatda.com/cmx.p018...300 sensors
Iscan ww en_cod tss color sensor_scan
Redolyser ww en_orp temp sensor_scan
Drager Polytron 5700 Fixed Gas Detector - Spec Sheet
Oxilyser ww en_oxygen do sensor_scan
Ad

More from ENVIMART (20)

PDF
Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
PDF
Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
PDF
Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
PDF
Lead Continuous Analysis from SEIBOLD-Wasser.pdf
PDF
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
PDF
Iron Continuous Analysis from SEIBOLD-Wasser.pdf
PDF
Chromium Continuous Analysis from SEIBOLD-Wasser.pdf
PDF
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
PDF
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
PDF
Copper Continuous Analysis from SEIBOLD-Wasser.pdf
PDF
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdf
PDF
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdf
PDF
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdf
PDF
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdf
PDF
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
PDF
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
PDF
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
PDF
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
PDF
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
PDF
B12-2-Wire-Gas-Transmitter.pdf
Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Lead Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Continuous Analysis from SEIBOLD-Wasser.pdf
Chromium Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Continuous Analysis from SEIBOLD-Wasser.pdf
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdf
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdf
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdf
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
B12-2-Wire-Gas-Transmitter.pdf
Ad

Recently uploaded (20)

PPT
Compliance Monitoring report CMR presentation.ppt
PDF
Lecture 2 investigation of renal diseses.pdf
DOCX
Epoxy Coated Steel Bolted Tanks for Farm Digesters Supports On-Farm Organic W...
PPTX
Biodiversity of nature in environmental studies.pptx
DOCX
Epoxy Coated Steel Bolted Tanks for Crude Oil Large-Scale Raw Oil Containment...
DOCX
Epoxy Coated Steel Bolted Tanks for Beverage Wastewater Storage Manages Liqui...
PDF
Earthquake, learn from the past and do it now.pdf
PPTX
FIRE SAFETY SEMINAR SAMPLE FOR EVERYONE.pptx
PPTX
Biodiversity.udfnfndrijfreniufrnsiufnriufrenfuiernfuire
PPTX
Green Modern Sustainable Living Nature Presentation_20250226_230231_0000.pptx
PPTX
Concept of Safe and Wholesome Water.pptx
PPTX
Green and Cream Aesthetic Group Project Presentation.pptx
PPTX
Disposal Of Wastes.pptx according to community medicine
PDF
Bai bao Minh chứng sk2-DBTrong-003757.pdf
PPT
PPTPresentation3 jhsvdasvdjhavsdhsvjcksjbc.jasb..ppt
PDF
Ornithology-Basic-Concepts.pdf..........
PPT
Environmental pollution for educational study
DOCX
Epoxy Coated Steel Bolted Tanks for Fish Farm Water Provides Reliable Water f...
PDF
Urban Hub 50: Spirits of Place - & the Souls' of Places
PPTX
Arugula. Crop used for medical plant in kurdistant
Compliance Monitoring report CMR presentation.ppt
Lecture 2 investigation of renal diseses.pdf
Epoxy Coated Steel Bolted Tanks for Farm Digesters Supports On-Farm Organic W...
Biodiversity of nature in environmental studies.pptx
Epoxy Coated Steel Bolted Tanks for Crude Oil Large-Scale Raw Oil Containment...
Epoxy Coated Steel Bolted Tanks for Beverage Wastewater Storage Manages Liqui...
Earthquake, learn from the past and do it now.pdf
FIRE SAFETY SEMINAR SAMPLE FOR EVERYONE.pptx
Biodiversity.udfnfndrijfreniufrnsiufnriufrenfuiernfuire
Green Modern Sustainable Living Nature Presentation_20250226_230231_0000.pptx
Concept of Safe and Wholesome Water.pptx
Green and Cream Aesthetic Group Project Presentation.pptx
Disposal Of Wastes.pptx according to community medicine
Bai bao Minh chứng sk2-DBTrong-003757.pdf
PPTPresentation3 jhsvdasvdjhavsdhsvjcksjbc.jasb..ppt
Ornithology-Basic-Concepts.pdf..........
Environmental pollution for educational study
Epoxy Coated Steel Bolted Tanks for Fish Farm Water Provides Reliable Water f...
Urban Hub 50: Spirits of Place - & the Souls' of Places
Arugula. Crop used for medical plant in kurdistant

Spectrolyser v3 ww_en_multi parameter_scan

  • 1. © s::can GmbH (2020) www.s-can.at spectro::lyser V3 ∙ measuring principle: UV-Vis spectrometry over the total range (190-750 nm) ∙ web server on board - IoT enabled, no user software is needed to configure the probe ∙ communicates directly with your mobile device via WLAN ∙ choose exactly the parameters you want to measure – unlimited number of parameters possible ∙ 8 GB onboard memory - capacity for logging data for many years ∙ improved optical performance - revolutionary precision ∙ fast measurement interval - every 10 seconds possible ∙ extremely power efficient - sleep mode for low energy consumption ∙ multiparameter probe with 1 mm, 5 mm or 35 mm optical path length, ideal for waste water, surface water and drinking water ∙ long term stable and maintenance free in operation ∙ factory precalibrated, local multi-point calibration possible ∙ automatic cleaning with compressed air or brush/ruck::sack 44 42 5 266,5 457 44,5 ~ Messgeräte Sonstige Daten recommended accessories part number article name B-32-xxx s::can compressor B-33-012 con::nect V3 B-44 B-44-2 cleaning valve C-32-V3 Adapter cable to connect a V3 spectrometer (M12) to V2 Terminal (MIL Plug) D-330-xxx con::cube V3 F-110-V3 carrier s::can spectrometer V3 & V2 probe, 45° F-48-V3 spectrometer V3 & V2 flow-cell (bypass setup), PVC S-11-xx-moni moni::tool Software spectro::lyser® UV-Vis monitors depending on the application an individual selection of: TSS, TS, turbidity, color, TOC, DOC, BOD, COD, NO3-N, NO3, HS-, O3, CLD, UV254, fingerprints, spectral alarms and temperature
  • 2. © s::can GmbH (2020) www.s-can.at technical specification measuring principle UV-Vis spectrometry 190 - 750 nm measuring principle detail xenon flash lamp, pixel array detector measurement interval 10 sec (configurable, depending on application) automatic compensation instrument real dual beam measurement for compensation and detailed diagnostics automatic compensation cross sensitivities turbidity / solids / organic substances precalibrated ex-works all parameters accuracy standard solution (>1 mg/l) NO3-N: +/- 2% +1/OPL[mg/l]* COD-KHP: +/-2% +10/OPL[mg/l]* (* OPL … optical pathlength in mm) access to raw signals access to spectral information reference standard distilled water onboard memory 8 GB integrated temperature sensor 0 ... 45 °C resolution temperature sensor 0.1 °C integration via con::cube V3 con::nect V3 con::lyte V5 (D-320-pro2) and adapter cable (C-32-V3) power supply 10 ... 18 VDC power consumption (typical) 3 W power consumption (sleep model) 60 mW power consumption (max.) 20 W interface to s::can terminals M12 RSTS 8Y (IP67), RS485, Ethernet interface to third party terminals con::nect V3 incl. Modbus RTU, REST API, Modbus TCP/IP digital interface (for cleaning devices) 1 digital in/out 1 digital out network connection 100Base-T Ethernet, WLAN status information RGB LED ring internal sensors supply voltage sensor, tilt sensor, rotation sensor cable length 1 m fixed cable (-010) or 7.5 m fixed cable (-075) or 15 m fixed cable (-150) cable type PU jacket housing material stainless steel 1.4404 window material optical path length 5 and 1 mm: sapphire optical path length 35 mm: fused silica (UV-grade) weight (min.) 3.4 kg (incl. cable) dimensions (Ø x l) optical path length 35 mm: 44 x 473 mm / 517.5 mm optical path length 5 mm: 44 x 457 mm / 501.5 mm optical path length 1 mm: 44 x 453 mm / 497.5 mm operating temperature 0 ... 45 °C operating pressure 0 ... 3 bar high pressure specification (optional) 10 bar installation / mounting submersed or in a flow cell flow velocity 3 m/s (max.) mechanical stability 30 Nm ingress protection class IP68 automatic cleaning media: compressed air or autobrush permissible pressure: 3 ... 6 bar storage temperature -10 ... 65 °C conformity - environmental testing EN 60721-3 conformity - EMC EN 61326-1 conformity - RoHS 2 EN 50581 standard warranty 2 years extended warranty (optional) 3 years
  • 3. © s::can GmbH (2020) www.s-can.at
  • 4. © s::can GmbH (2020) www.s-can.at municipal WWTP influent & sewer parameter TSS [mg/l] color (app) [Hazen] color (tru) [Hazen] TOC [mg/l] DOC [mg/l] BOD [mg/l] COD [mg/l] COD f [mg/l] NO3-N [mg/l] HS- [mg/l] UV254 [Abs/m] UV254 f [Abs/m] part number spectro::lyser™ V3 (1 mm OPL, UV-Vis) min. 0 0 0 0 0 0 0 0 0 0 0 0 SP3-1-01-NO-xxx max. 8000 23000 14000 3300 2600 5300 10000 5300 100 80 3300 2800 spectro::lyser™ V3 (5 mm OPL, UV-Vis) min. 0 0 0 0 0 0 0 0 0 0 0 0 SP3-1-05-NO-xxx max. 1200 3500 2100 500 400 800 1500 800 16 12 500 420 municipal WWTP aeration parameter TS [g/l] COD f [mg/l] NO3-N [mg/l] UV254 [Abs/m] UV254 f [Abs/m] part number spectro::lyser™ V3 (1 mm OPL, UV-Vis) min. 0 0 0 0 0 SP3-1-01-NO-xxx max. 20 530 26 3300 2800 municipal WWTP effluent parameter TSS [mg/l] turbidity [NTU/ FTU] color (app) [Hazen] color (tru) [Hazen] TOC [mg/l] DOC [mg/l] BOD [mg/l] COD [mg/l] COD f [mg/l] NO3-N [mg/l] O3 [mg/l] UV254 [Abs/m] UV254 f [Abs/m] part number spectro::lyser™ V3 (1 mm OPL, UV-Vis) min. 0 0 0 0 0 0 0 0 0 0 0 0 0 SP3-1-01-NO-xxx max. 4000 8000 23000 14000 2600 2000 2000 3300 2000 300 1200 3300 2800 spectro::lyser™ V3 (5 mm OPL, UV-Vis) min. 0 0 0 0 0 0 0 0 0 0 0 0 0 SP3-1-05-NO-xxx max. 600 1200 3500 2100 400 300 300 500 300 45 180 500 420 paper mill WWTP influent parameter TSS [mg/l] COD [mg/l] COD f [mg/l] NO3-N [mg/l] UV254 [Abs/m] UV254 f [Abs/m] part number spectro::lyser™ V3 (1 mm OPL, UV-Vis) min. 0 0 0 0 0 0 SP3-1-01-NO-xxx max. 8000 13000 11000 100 3300 2800 spectro::lyser™ V3 (5 mm OPL, UV-Vis) min. 0 0 0 0 0 0 SP3-1-05-NO-xxx max. 1200 2000 1700 16 500 420 paper mill WWTP effluent parameter TSS [mg/l] COD [mg/l] COD f [mg/l] NO3-N [mg/l] UV254 [Abs/m] UV254 f [Abs/m] part number spectro::lyser™ V3 (1 mm OPL, UV-Vis) min. 0 0 0 0 0 0 SP3-1-01-NO-xxx max. 4000 5300 3300 100 3300 2800 spectro::lyser™ V3 (5 mm OPL, UV-Vis) min. 0 0 0 0 0 0 SP3-1-05-NO-xxx max. 600 790 490 16 500 420 brewery WWTP influent parameter TSS [mg/l] COD [mg/l] COD f [mg/l] NO3-N [mg/l] UV254 [Abs/m] UV254 f [Abs/m] part number spectro::lyser™ V3 (1 mm OPL, UV-Vis) min. 0 0 0 0 0 0 SP3-1-01-NO-xxx max. 13000 60000 53000 100 3300 2800 spectro::lyser™ V3 (5 mm OPL, UV-Vis) min. 0 0 0 0 0 0 SP3-1-05-NO-xxx max. 2000 9000 7900 16 500 420 dairy WWTP influent parameter TSS [mg/l] COD [mg/l] COD f [mg/l] NO3-N [mg/l] UV254 [Abs/m] UV254 f [Abs/m] part number spectro::lyser™ V3 (1 mm OPL, UV-Vis) min. 0 0 0 0 0 0 SP3-1-01-NO-xxx max. 8000 33000 16000 210 3300 2800