Bili Chek:
 Transcutaneous
Bilirubin Analyzer



              NLC P/P# 6600.0216
Purpose:
   To analyze total serum bilirubin (TSB) in a noninvasive and
    pain-free manner.
Policy:

• The Bili Chek Analyzer has been clinically proven to be accurate in
  patients within the following parameters:
       • Gestational Age of 27 to 42 weeks
       • Postnatal Age of 0 to 20 days
       • Infant Weight of 950 to 4995 grams

• The Bili-Chek should not be used in the following situations:
       • Following exchange transfusion
       • If the measurement site on the forehead contains excessive
           bruising, birthmarks, hematomas, or excessive hairiness
           due to erroneous results.
       • If unable to use the forehead, it is recommended to use the
           sternum as an alternative measurement site.
Testing procedure

   1.   Initializing the Unit: Install a fully charged battery pack into the
        battery compartment of the unit. Press and release either the F1 (blue)
        or F2 (gray) button on the front of the Bili Chek unit to turn the device
        on. The device will perform a self-test, momentarily displaying all
        LCD indicators. When the self-test is complete, the home screen will be
        visible, displaying the last measurement, time and date (or an error
        code message, if applicable).

   2.   The first time the Bili Chek is used, it is necessary to enter the set-up
        mode and program the display settings before proceeding. Refer to the
        User Instruction Manual for complete set up instructions.

   3.   The Bili Chek does not have an off switch, and will automatically turn
        off if is idle for a 60 seconds.
Calibration
   1.   Remove a new BiliCal disposable tip from its foil pouch and apply it to the
        optical sensor on the Bili Chek handheld device before each use.

   2.   Firmly press the BiliCal on the Bili Chek handheld device to ensure proper
        seating of the tip.

   3.   With the home screen displayed, press and release the trigger button (the blue
        button located on the hand grip) to start calibration.

   4.   Three dashes (---) will flash in the display window, and the Measurement
        Status Indicator (MSI) will be amber if the BiliCal is properly seated,
        indicating that the device is ready to calibrate. (If the BiliCal is not properly
        seated, the mSI will be red, and an EO1 error message will be displayed).

   5.   Press and release the trigger button again. The dashed lines will stop flashing,
        indicating that the Bili Chek is calibrating.

   6.   The MSI will be amber colored and a beep will be heard. The display window
        will read (005) to indicate that the calibration was completed successfully. If
        there is a failure in calibration, and error message will be displayed, and one
        will be unable to proceed with the testing. Refer to the trouble shooting
        section of the User Instruction Manual.
Performing the Patient Test

   1.   After performing the calibration, pull on the BiliCal tab and peel away
        the protective covering from the disposable tip and discard.

   2.   Press and release the trigger button. The device is now activated and
        ready to take a measurement (005) will be displayed and blinking.

   3.   Gently press the bili cal against the infant’s forehead or sternum. The
        MSI on the display will change from amber to green, and the 005 will
        stop blinking when the proper pressure is applied.

   4.   Hold the Bili Chek handheld device steady until the measurement is
        complete (for about 1-3 seconds). The device will beep if the audible
        alarm is enabled.
5.   Perform a series of five (5) measurements by lifting and replacing the
     BiliCal on the center of the infant’s forehead of sternum. Press and
     release the trigger button before each measurement. The current
     measurement will be indicated on the display (005…003..001).


6.   Upon completion of the five (5) measurements, a final beep will
     sound, and the test result will be displayed along with the current time
     and date. Remove and discard the disposable tip.


7.   Place the blue protective tip cover onto the Bili Chek handheld device.


8.   The Bili Chek will turn off automatically.
Documenting the Results

   1. Document the infant test result, date, and time in the infant’s chart, on
      Jaundice Teaching Sheet and the electronic Newborn Discharge
      Information Sheet.

   2. Notify the infant’s physician as appropriate.

   3. Obtain follow-up measurements in accordance with physician orders.

   4. QA documentation Procedures: The BiliChek performs internal
      calibration controls prior to each patient test. The device will not permit
      testing to occur if the calibration does not meet the control
      specifications. In order to document completion of the calibration prior
      to each test, a “Quality Documentation Record” should be completed
      (refer to Appendix A). An individual record should be maintained for
      each BiliChek device, and tracked by serial number.
Cleaning and Maintenance

   1.   Cavicide wipes are used to clean the Bili Chek after each use.


   2.   To clean, wipe the Bili Chek with a Cavicide wipe, being sure to clean the
        display window carefully.


   3.   Do not immerse the Bili Chek in water or any liquid. If liquids spill onto
        the unit, wipe immediately with a cloth and let unit dry before use.


   4.   Do not attempt to clean and/or reuse the BiliCal Disposable Tip.
Use During Phototherapy

   1.   Prior to the initiation of Phototherapy, the measurement site on the
        infant’s forehead or sternum must be covered with a photo-opaque
        patch material. The patch must remain in place throughout
        phototherapy.


   2.   To take a measurement during phototherapy: Turn off the
        phototherapy lights (fiber optic or overhead), remove the photo-
        opaque patch, perform the Bili Chek measurement, replace the photo-
        opaque patch, resume phototherapy as ordered.


   3.   Clinical studies indicate that up to 48 hours may be required before
        the skin treated by phototherapy returns to the bilirubin level of an
        unexposed site; therefore measurements from an unprotected site are
        not reliable.

More Related Content

PPTX
Exchange transfusion
PPTX
Radiant Warmers and Incubators
PPTX
Weight Recording
PPTX
PPTX
Cne on bishop’s score
PPTX
Care of a ventilated child
PPTX
care of child on ventilator
PPT
Clinical Monitoring of Sick Newborn LSD
Exchange transfusion
Radiant Warmers and Incubators
Weight Recording
Cne on bishop’s score
Care of a ventilated child
care of child on ventilator
Clinical Monitoring of Sick Newborn LSD

What's hot (20)

PPT
Exchange Transfusion PPT
PPTX
NEONATAL RESUSCITATION 2022.pptx
PPTX
Neonatal warning signs
PPT
PPTX
NURSING MANAGEMENT OF CHILDREN WITH COMMUNICABLE DISEASES
PPTX
Neonatal hypothermia
PPTX
Nursing care of a normal newborn.pptx
PPTX
Head injury and its management
PPTX
APGAR and BISHOP Scoring.pptx
PPTX
Assessment of newborn
PPTX
Value and selection of play
PPT
Neonatal jaundice, cause, Etiology, Treatment, nursing Care of baby in Photo-...
PPTX
Neonatal transport
PPTX
Characteristics of a Healthy Newborn
PPTX
Phototherapy in jaundice
PPTX
Play in children ppt presentation
PPTX
Management of LBW Babies
PPTX
KMC.pptx
PPTX
National health policy for children
PPTX
Nursing management of low birth weight(lbw) babies
Exchange Transfusion PPT
NEONATAL RESUSCITATION 2022.pptx
Neonatal warning signs
NURSING MANAGEMENT OF CHILDREN WITH COMMUNICABLE DISEASES
Neonatal hypothermia
Nursing care of a normal newborn.pptx
Head injury and its management
APGAR and BISHOP Scoring.pptx
Assessment of newborn
Value and selection of play
Neonatal jaundice, cause, Etiology, Treatment, nursing Care of baby in Photo-...
Neonatal transport
Characteristics of a Healthy Newborn
Phototherapy in jaundice
Play in children ppt presentation
Management of LBW Babies
KMC.pptx
National health policy for children
Nursing management of low birth weight(lbw) babies
Ad

More from wcmc (20)

PPTX
Dress code
PPTX
Suicide precautions
PPTX
Social work services
PPTX
Rounding
PPTX
Restraints
PPTX
Preparing for surgery
PPTX
Pharmacy
PPTX
Peritoneal dialysis
PPTX
Peak & trough
PPTX
Pca
PPTX
Pain management
PPTX
Organ tissue donation
PPTX
Medication reconciliation
PPTX
Medication administration
PPTX
Iv therapy
PPTX
Isolation
PPTX
Insulin drips
PPTX
Hemodialysis
PPTX
Haldol drips
PPTX
Epidural
Dress code
Suicide precautions
Social work services
Rounding
Restraints
Preparing for surgery
Pharmacy
Peritoneal dialysis
Peak & trough
Pca
Pain management
Organ tissue donation
Medication reconciliation
Medication administration
Iv therapy
Isolation
Insulin drips
Hemodialysis
Haldol drips
Epidural
Ad

Recently uploaded (20)

PPTX
svnfcksanfskjcsnvvjknsnvsdscnsncxasxa saccacxsax
PPTX
basic introduction to research chapter 1.pptx
PPTX
Astra-Investor- business Presentation (1).pptx
PDF
NISM Series V-A MFD Workbook v December 2024.khhhjtgvwevoypdnew one must use ...
PDF
ANALYZING THE OPPORTUNITIES OF DIGITAL MARKETING IN BANGLADESH TO PROVIDE AN ...
PDF
Keppel_Proposed Divestment of M1 Limited
PPT
Lecture notes on Business Research Methods
PDF
income tax laws notes important pakistan
PPTX
IITM - FINAL Option - 01 - 12.08.25.pptx
PPT
Lecture 3344;;,,(,(((((((((((((((((((((((
PDF
Kishore Vora - Best CFO in India to watch in 2025.pdf
DOCX
80 DE ÔN VÀO 10 NĂM 2023vhkkkjjhhhhjjjj
PPTX
Board-Reporting-Package-by-Umbrex-5-23-23.pptx
PPTX
Project Management_ SMART Projects Class.pptx
PDF
NEW - FEES STRUCTURES (01-july-2024).pdf
PPTX
operations management : demand supply ch
PDF
Environmental Law Communication: Strategies for Advocacy (www.kiu.ac.ug)
PDF
PMB 401-Identification-of-Potential-Biotechnological-Products.pdf
PDF
THE COMPLETE GUIDE TO BUILDING PASSIVE INCOME ONLINE
PDF
Booking.com The Global AI Sentiment Report 2025
svnfcksanfskjcsnvvjknsnvsdscnsncxasxa saccacxsax
basic introduction to research chapter 1.pptx
Astra-Investor- business Presentation (1).pptx
NISM Series V-A MFD Workbook v December 2024.khhhjtgvwevoypdnew one must use ...
ANALYZING THE OPPORTUNITIES OF DIGITAL MARKETING IN BANGLADESH TO PROVIDE AN ...
Keppel_Proposed Divestment of M1 Limited
Lecture notes on Business Research Methods
income tax laws notes important pakistan
IITM - FINAL Option - 01 - 12.08.25.pptx
Lecture 3344;;,,(,(((((((((((((((((((((((
Kishore Vora - Best CFO in India to watch in 2025.pdf
80 DE ÔN VÀO 10 NĂM 2023vhkkkjjhhhhjjjj
Board-Reporting-Package-by-Umbrex-5-23-23.pptx
Project Management_ SMART Projects Class.pptx
NEW - FEES STRUCTURES (01-july-2024).pdf
operations management : demand supply ch
Environmental Law Communication: Strategies for Advocacy (www.kiu.ac.ug)
PMB 401-Identification-of-Potential-Biotechnological-Products.pdf
THE COMPLETE GUIDE TO BUILDING PASSIVE INCOME ONLINE
Booking.com The Global AI Sentiment Report 2025

Transcutaneous bilirubin monitoring (bilichek)

  • 1. Bili Chek: Transcutaneous Bilirubin Analyzer NLC P/P# 6600.0216
  • 2. Purpose: To analyze total serum bilirubin (TSB) in a noninvasive and pain-free manner.
  • 3. Policy: • The Bili Chek Analyzer has been clinically proven to be accurate in patients within the following parameters: • Gestational Age of 27 to 42 weeks • Postnatal Age of 0 to 20 days • Infant Weight of 950 to 4995 grams • The Bili-Chek should not be used in the following situations: • Following exchange transfusion • If the measurement site on the forehead contains excessive bruising, birthmarks, hematomas, or excessive hairiness due to erroneous results. • If unable to use the forehead, it is recommended to use the sternum as an alternative measurement site.
  • 4. Testing procedure 1. Initializing the Unit: Install a fully charged battery pack into the battery compartment of the unit. Press and release either the F1 (blue) or F2 (gray) button on the front of the Bili Chek unit to turn the device on. The device will perform a self-test, momentarily displaying all LCD indicators. When the self-test is complete, the home screen will be visible, displaying the last measurement, time and date (or an error code message, if applicable). 2. The first time the Bili Chek is used, it is necessary to enter the set-up mode and program the display settings before proceeding. Refer to the User Instruction Manual for complete set up instructions. 3. The Bili Chek does not have an off switch, and will automatically turn off if is idle for a 60 seconds.
  • 5. Calibration 1. Remove a new BiliCal disposable tip from its foil pouch and apply it to the optical sensor on the Bili Chek handheld device before each use. 2. Firmly press the BiliCal on the Bili Chek handheld device to ensure proper seating of the tip. 3. With the home screen displayed, press and release the trigger button (the blue button located on the hand grip) to start calibration. 4. Three dashes (---) will flash in the display window, and the Measurement Status Indicator (MSI) will be amber if the BiliCal is properly seated, indicating that the device is ready to calibrate. (If the BiliCal is not properly seated, the mSI will be red, and an EO1 error message will be displayed). 5. Press and release the trigger button again. The dashed lines will stop flashing, indicating that the Bili Chek is calibrating. 6. The MSI will be amber colored and a beep will be heard. The display window will read (005) to indicate that the calibration was completed successfully. If there is a failure in calibration, and error message will be displayed, and one will be unable to proceed with the testing. Refer to the trouble shooting section of the User Instruction Manual.
  • 6. Performing the Patient Test 1. After performing the calibration, pull on the BiliCal tab and peel away the protective covering from the disposable tip and discard. 2. Press and release the trigger button. The device is now activated and ready to take a measurement (005) will be displayed and blinking. 3. Gently press the bili cal against the infant’s forehead or sternum. The MSI on the display will change from amber to green, and the 005 will stop blinking when the proper pressure is applied. 4. Hold the Bili Chek handheld device steady until the measurement is complete (for about 1-3 seconds). The device will beep if the audible alarm is enabled.
  • 7. 5. Perform a series of five (5) measurements by lifting and replacing the BiliCal on the center of the infant’s forehead of sternum. Press and release the trigger button before each measurement. The current measurement will be indicated on the display (005…003..001). 6. Upon completion of the five (5) measurements, a final beep will sound, and the test result will be displayed along with the current time and date. Remove and discard the disposable tip. 7. Place the blue protective tip cover onto the Bili Chek handheld device. 8. The Bili Chek will turn off automatically.
  • 8. Documenting the Results 1. Document the infant test result, date, and time in the infant’s chart, on Jaundice Teaching Sheet and the electronic Newborn Discharge Information Sheet. 2. Notify the infant’s physician as appropriate. 3. Obtain follow-up measurements in accordance with physician orders. 4. QA documentation Procedures: The BiliChek performs internal calibration controls prior to each patient test. The device will not permit testing to occur if the calibration does not meet the control specifications. In order to document completion of the calibration prior to each test, a “Quality Documentation Record” should be completed (refer to Appendix A). An individual record should be maintained for each BiliChek device, and tracked by serial number.
  • 9. Cleaning and Maintenance 1. Cavicide wipes are used to clean the Bili Chek after each use. 2. To clean, wipe the Bili Chek with a Cavicide wipe, being sure to clean the display window carefully. 3. Do not immerse the Bili Chek in water or any liquid. If liquids spill onto the unit, wipe immediately with a cloth and let unit dry before use. 4. Do not attempt to clean and/or reuse the BiliCal Disposable Tip.
  • 10. Use During Phototherapy 1. Prior to the initiation of Phototherapy, the measurement site on the infant’s forehead or sternum must be covered with a photo-opaque patch material. The patch must remain in place throughout phototherapy. 2. To take a measurement during phototherapy: Turn off the phototherapy lights (fiber optic or overhead), remove the photo- opaque patch, perform the Bili Chek measurement, replace the photo- opaque patch, resume phototherapy as ordered. 3. Clinical studies indicate that up to 48 hours may be required before the skin treated by phototherapy returns to the bilirubin level of an unexposed site; therefore measurements from an unprotected site are not reliable.