Unit 4: Special Oscilloscopes
Delayed time base oscilloscope:
        In a conventional oscilloscope, the signal to the vertical plates is delayed by time, using1
delay line circuit. The waveform before the delay circuit is used to trigger the time base. This
allows the study of all the leading or lagging edges of a pulse type waveform. This is shown in
the Figure to triggering of time base by input signal, sweep starts well in time and when input
appears at vertical sections, the sweep is triggered and delayed waveform is displayed.

        The delay ensures that no part of the waveform gets lost. In delayed time base
oscilloscope, a variable time delay circuit is used in the basic time base circuit. This allows the
triggering of sweep time after the delay time. Thus the delay time is variable. This time is
denoted as td. After this, the sweep is triggered for the time t, Then the portion of the waveform
for the time t x gets expanded on the complete oscilloscope screen, for the detail study.




        If input is pulse waveform and leading edge is used to trigger the delay time, then lagging
edge can be displayed to fill the entire oscilloscope screen. This is shown in the Fig (a). Similarly
if the lagging edge is used to trigger the delay time then leading edge will be displayed on the
entire screen for the time tx. This is shown in the Fig.(b). If the time delay is perfectly adjusted,
then any portion of the waveform can be extended to fill the entire screen. This is shown in the
Fig. (c).
The normal time base circuit is main time base (MTB) circuit which functions same as c1
conventional oscilloscope. The function of MTB blanking circuit is to produce an unblanking
pulse which is applied to CRT grid to turn on an electron beam in the CRT, during the display
sweep time. The ramp output of MTB is given to the horizontal deflection amplifier via switch S.
It is also given as one input to the voltage comparator.

        The other input to the voltage comparator is derived from the potentiometer whose level
is adjustable. The unblanking pulses from MTB and DTB are added by summing circuit and
given to the CRT grid. The unblanking pulse of MTB produces a trace of uniform intensity. But
during ramp time of DTB, the addition of two pulses decides the intensity of the trace on the
screen. Hence during DTB time, the voltage applied to CRT grid is almost twice than the voltage
corresponding to MTB time.

         This increases the brightness of the displayed waveform for the DTB time. When the
part of the waveform to be brightened is identified, then the DTB ramp output is connected to the
input of the horizontal deflection amplifier through switch S. The DTB ramp time is much
smaller than MTB period but its amplitude (- V to + V) is same as MTB ramp.

        hence it causes the oscilloscope electron beam to be deflected from one side of the screen
to the other, during short DTB time. By adjusting DTB time/ div control, the brightened portion
can be extended, so as to fill the entire screen of the oscilloscope. The horizontal deflection starts
only after the delay time t d from the beginning of the MTB sweep. Thus very small part of the
waveform can be extended on the entire screen. Waveforms of the Delayed Time Base
Oscilloscope The waveforms of the delayed time base oscilloscope are shown in the Fig.
Analog storage oscilloscope:
        The conventional cathode ray tube has the persistence of the phosphor ranging from a
Few millisecond to several seconds. But sometimes it is necessary to retain the image for much
'longer periods, upto several hours. It requires storing of a waveform for a certain duration,'
independent of phosphor persistence. Such a retention property helps to display the waveforms
of very low frequency. Mainly two types of storage techniques are used in cathode ray tubes
which are:
i) Mesh storage and ii) Phosphor storage

Sampling Time Base: The time base circuit of the sampling oscilloscope is different than the
conventional oscilloscope. The time base of sampling oscilloscope has two functions: i) To move
the dots across the screen ii) To generate the sampling command pulses for the sampling circuit.

       It consists of synchronous circuit, which determines the sampling rate and establishes a
reference point in time with respect to the input signal. The time base generates a triggering
pulse which activates the oscillator to generate a ramp voltage. Similarly it generates a stair case
waveform. The ramp generation is based on the output of the synchronizing circuit.

        Both the ramp as well as staircase waveforms are applied to a voltage comparator. This
comparator compares the two voltages and whenever these two voltages are equal, it generates a
samppng pulse. This pulse then momentarily bias the diodes of the sampling gate in the forward
direction and thus diode switch gets closed for short duration of time.
The capacitor charges but for short time hence, it can charge to only a small percentage
of the input signal value at that instant. This voltage is amplified by the vertical amplifier and
then applied to the vertical deflecting plates. This is nothing but a sample. At the same time, the
comparator gives a signal to the staircase generator to advance through one step. This is applied
to horizontal deflecting plates, thus during each step of the stair case waveform, the spot moves
across the screen. Thus the sampling time base is called a staircase-ramp generator in case of a
sampling oscillosope.

Block diagram of Sampling Oscilloscope:

The block diagram of sampling oscilloscope is shown in the Fig.




        The input signal is applied to the diode sampling gate. At the start of each sampling cycle
a trigger input pulse is generated which activates the blocking oscillator. The oscillator output is
given to the ramp generator which generates the linear ramp signal. Since the sampling must be
synchronized with the input signal frequency, the signal is delayed in the vertical amplifier.

       The staircase generator produces a staircase waveform which is applied to an attenuator.
The attenuator controls the magnitude of the staircase signal and then it is applied to a voltage
comparator. Another input to the voltage comparator is the output of the ramp generator. The
voltage comparator compares the two signals and produces the output pulse when the two
voltages are equal. This is nothing but a sampling pulse which is applied to sampling gate
through the gate control circuitry. This pulse opens the diode gate and sample is taken in.

        This sampled signal is then applied to the vertical amplifier and the vertical deflecting
plates. The output of the staircase generator is also applied to the horizontal deflecting plates.
During each step of staircase the spot moves on the screen. The comparator output advances the
staircase output through one step. After certain number of pulses about thousand or so, the
staircase generator resets. The smaller the size of the steps of the staircase generator, larger is the
number of samples and higher is the resolution of the image.

Digital Storage Oscilloscope:
In this digital storage oscilloscope, the waveform to be stored is digitized and then stored in a
digital memory. The conventional cathode ray tube is used in this oscilloscope hence the cost is
less. The power to be applied to memory is small and can be supplied by small battery. Due to
this the stored image can be displayed indefinitely as long ,15 power is supplied to memory.
Once the waveform is digitized then it can be further loaded into the computer and can be
analysed in detail.

Block Diagram: The block diagram of digital storage oscilloscope is shown in the Fig.




        As done in all the oscilloscopes, the input signal is applied to the amplifier and attenuator
section. The oscilloscope uses same type of amplifier and attenuator circuitry as used in the
conventional oscilloscopes. The attenuated signal is then applied to the vertical amplifier.
The vertical input, after passing through the vertical amplifier, is digitised by an analog to
digital converter to create a data set that is stored in the memory. The data set is processed by the
microprocessor and then sent to the display. To digitise the analog signal, analog to digital (AID)
converter is used. The output of the vertical amplifier is applied to the AID converter section.
The main requirement of AID converter in the digital storage oscilloscope is its speed, while in
digital voltmeters accuracy and resolution were the main requirements. The digitised output
needed only in the binary form and not in BCD. The successive approximation type of AID
converter is most oftenly used in the digital storage oscilloscopes.

Modes of operation: The digital storage oscilloscope has three modes of operation:

   1. Roll mode ii) Store mode iii) Hold or save mode.

Roll mode: This mode is used to display very fast varying signals, clearly on the screen. The fast
varying signal is displayed as if it is changing slowly, on the screen. In this mode, the input
signal is not triggered at all.

More Related Content

PDF
Lecture 6 Tachogenerators
PPT
Harmonic reduction
PPTX
Closed loop speed control
PPT
Servomotor
PPTX
Module 5 :Voltage regulator IC 78XX 79XX LM317
PPT
slew rate in opamp
PDF
Op amp comparator
Lecture 6 Tachogenerators
Harmonic reduction
Closed loop speed control
Servomotor
Module 5 :Voltage regulator IC 78XX 79XX LM317
slew rate in opamp
Op amp comparator

What's hot (20)

PDF
Ec6412 linear-integrated-circuit-laboratory
PPTX
Rectifiers BY TALHA MIRZA
PPT
Oscillators
PPTX
Hartley oscillator circuit theory working and application
PPTX
Satellite communication lecture7
PPTX
Oscillators
PPTX
Ac Potentiometer (EMMI)
PPTX
Position control system block diagram
PPTX
digita storage oscilloscope (dso)
PPTX
Hays bridge
PPT
differential amplifier for electronic
PPTX
Tuned amplifire
PPTX
Cathode Ray Oscilloscope CRO & Digital Oscilloscope 'S WORKING
PPTX
Z parameters
DOCX
Li-Fi Audio Transmission Project Documentation/Report
PPTX
Tv transmitters & receiver
PPTX
Power Converters for AC drives CH-1.pptx
PPT
Power amplifire analog electronics
PPTX
Phase locked loop
PDF
Types of bio amplifiers
Ec6412 linear-integrated-circuit-laboratory
Rectifiers BY TALHA MIRZA
Oscillators
Hartley oscillator circuit theory working and application
Satellite communication lecture7
Oscillators
Ac Potentiometer (EMMI)
Position control system block diagram
digita storage oscilloscope (dso)
Hays bridge
differential amplifier for electronic
Tuned amplifire
Cathode Ray Oscilloscope CRO & Digital Oscilloscope 'S WORKING
Z parameters
Li-Fi Audio Transmission Project Documentation/Report
Tv transmitters & receiver
Power Converters for AC drives CH-1.pptx
Power amplifire analog electronics
Phase locked loop
Types of bio amplifiers
Ad

Viewers also liked (20)

PPT
Hv instructions
PDF
Eltherm ELSR-Ramp Heat Trace Cable - Spec Sheet
PPT
DC Power Analyzer, Modular, 600W and Four Slots: N6705A
DOCX
ADC - Dual Slope Integrator
PDF
tca785
PPT
ABB Instrumentation Overview
PPTX
Function Generator
PPT
Back corona model
PPTX
Cro (emmi) (1)
PDF
Embarquer Linux et des systèmes libres, méthodes et apports
PDF
Tracabilite & Identification Automatique par Bernard JEANNE-BEYLOT @JB Thèque
PDF
Séminaire Captronic Yocto 24 février 2015
PDF
Concevoir un système Linux embarqué avec Yocto Project
PPTX
Tasich-Koyani, Mindy, Conestoga-Rovers Associates, Waste Reduction Strategies...
PPSX
Lng as marine fuel norway
PDF
Meet ROUSH CleanTech - BNA Airport Pres.
PPTX
AIR FLOW DETECTOR SYSTEM
PDF
Hydrogen energy
Hv instructions
Eltherm ELSR-Ramp Heat Trace Cable - Spec Sheet
DC Power Analyzer, Modular, 600W and Four Slots: N6705A
ADC - Dual Slope Integrator
tca785
ABB Instrumentation Overview
Function Generator
Back corona model
Cro (emmi) (1)
Embarquer Linux et des systèmes libres, méthodes et apports
Tracabilite & Identification Automatique par Bernard JEANNE-BEYLOT @JB Thèque
Séminaire Captronic Yocto 24 février 2015
Concevoir un système Linux embarqué avec Yocto Project
Tasich-Koyani, Mindy, Conestoga-Rovers Associates, Waste Reduction Strategies...
Lng as marine fuel norway
Meet ROUSH CleanTech - BNA Airport Pres.
AIR FLOW DETECTOR SYSTEM
Hydrogen energy
Ad

Similar to Unit 4 ei (20)

PPTX
Basic electrical and electronics engineering
PPTX
DSO.pptx
PPTX
CRO
PPTX
1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...
DOCX
Lissajous pattern
DOCX
Lissajous pattern
DOCX
Types of cro
PPT
emi 3rd ye 2semhapter-14 instrunments.ppt
PPT
Chapter-14 (1) measurementys and instruments.ppt
PPT
PE&I Chapter-14. ...
PPTX
PPT
CRO and FG.ppt
DOCX
CRO and functional generator physics
PPT
Cathode-ray-Oscilloscopendjknddnilndoid.ppt
PDF
I & M Part-2.pdfbuibuibiubbvybbbibbiuhuhi
PPT
Cathode ray oscilloscope for engineering .ppt
PPTX
SAMPLING AND STORAGE OSCILLOSCOPE
PDF
Lecture Unit_3_CRO(Cathode Rays osicolscope.pdf
PPT
cro oscilloscope for physics classes and students.ppt
PDF
EMI Unit 3 CRO
Basic electrical and electronics engineering
DSO.pptx
CRO
1.Oscilloscope. 2.Block diagram of Oscilloscope. 3.Types of Oscilloscope. 4.A...
Lissajous pattern
Lissajous pattern
Types of cro
emi 3rd ye 2semhapter-14 instrunments.ppt
Chapter-14 (1) measurementys and instruments.ppt
PE&I Chapter-14. ...
CRO and FG.ppt
CRO and functional generator physics
Cathode-ray-Oscilloscopendjknddnilndoid.ppt
I & M Part-2.pdfbuibuibiubbvybbbibbiuhuhi
Cathode ray oscilloscope for engineering .ppt
SAMPLING AND STORAGE OSCILLOSCOPE
Lecture Unit_3_CRO(Cathode Rays osicolscope.pdf
cro oscilloscope for physics classes and students.ppt
EMI Unit 3 CRO

Unit 4 ei

  • 1. Unit 4: Special Oscilloscopes Delayed time base oscilloscope: In a conventional oscilloscope, the signal to the vertical plates is delayed by time, using1 delay line circuit. The waveform before the delay circuit is used to trigger the time base. This allows the study of all the leading or lagging edges of a pulse type waveform. This is shown in the Figure to triggering of time base by input signal, sweep starts well in time and when input appears at vertical sections, the sweep is triggered and delayed waveform is displayed. The delay ensures that no part of the waveform gets lost. In delayed time base oscilloscope, a variable time delay circuit is used in the basic time base circuit. This allows the triggering of sweep time after the delay time. Thus the delay time is variable. This time is denoted as td. After this, the sweep is triggered for the time t, Then the portion of the waveform for the time t x gets expanded on the complete oscilloscope screen, for the detail study. If input is pulse waveform and leading edge is used to trigger the delay time, then lagging edge can be displayed to fill the entire oscilloscope screen. This is shown in the Fig (a). Similarly if the lagging edge is used to trigger the delay time then leading edge will be displayed on the entire screen for the time tx. This is shown in the Fig.(b). If the time delay is perfectly adjusted, then any portion of the waveform can be extended to fill the entire screen. This is shown in the Fig. (c).
  • 2. The normal time base circuit is main time base (MTB) circuit which functions same as c1 conventional oscilloscope. The function of MTB blanking circuit is to produce an unblanking pulse which is applied to CRT grid to turn on an electron beam in the CRT, during the display sweep time. The ramp output of MTB is given to the horizontal deflection amplifier via switch S. It is also given as one input to the voltage comparator. The other input to the voltage comparator is derived from the potentiometer whose level is adjustable. The unblanking pulses from MTB and DTB are added by summing circuit and given to the CRT grid. The unblanking pulse of MTB produces a trace of uniform intensity. But during ramp time of DTB, the addition of two pulses decides the intensity of the trace on the screen. Hence during DTB time, the voltage applied to CRT grid is almost twice than the voltage corresponding to MTB time. This increases the brightness of the displayed waveform for the DTB time. When the part of the waveform to be brightened is identified, then the DTB ramp output is connected to the input of the horizontal deflection amplifier through switch S. The DTB ramp time is much smaller than MTB period but its amplitude (- V to + V) is same as MTB ramp. hence it causes the oscilloscope electron beam to be deflected from one side of the screen to the other, during short DTB time. By adjusting DTB time/ div control, the brightened portion can be extended, so as to fill the entire screen of the oscilloscope. The horizontal deflection starts only after the delay time t d from the beginning of the MTB sweep. Thus very small part of the waveform can be extended on the entire screen. Waveforms of the Delayed Time Base Oscilloscope The waveforms of the delayed time base oscilloscope are shown in the Fig.
  • 3. Analog storage oscilloscope: The conventional cathode ray tube has the persistence of the phosphor ranging from a Few millisecond to several seconds. But sometimes it is necessary to retain the image for much 'longer periods, upto several hours. It requires storing of a waveform for a certain duration,' independent of phosphor persistence. Such a retention property helps to display the waveforms of very low frequency. Mainly two types of storage techniques are used in cathode ray tubes which are: i) Mesh storage and ii) Phosphor storage Sampling Time Base: The time base circuit of the sampling oscilloscope is different than the conventional oscilloscope. The time base of sampling oscilloscope has two functions: i) To move the dots across the screen ii) To generate the sampling command pulses for the sampling circuit. It consists of synchronous circuit, which determines the sampling rate and establishes a reference point in time with respect to the input signal. The time base generates a triggering pulse which activates the oscillator to generate a ramp voltage. Similarly it generates a stair case waveform. The ramp generation is based on the output of the synchronizing circuit. Both the ramp as well as staircase waveforms are applied to a voltage comparator. This comparator compares the two voltages and whenever these two voltages are equal, it generates a samppng pulse. This pulse then momentarily bias the diodes of the sampling gate in the forward direction and thus diode switch gets closed for short duration of time.
  • 4. The capacitor charges but for short time hence, it can charge to only a small percentage of the input signal value at that instant. This voltage is amplified by the vertical amplifier and then applied to the vertical deflecting plates. This is nothing but a sample. At the same time, the comparator gives a signal to the staircase generator to advance through one step. This is applied to horizontal deflecting plates, thus during each step of the stair case waveform, the spot moves across the screen. Thus the sampling time base is called a staircase-ramp generator in case of a sampling oscillosope. Block diagram of Sampling Oscilloscope: The block diagram of sampling oscilloscope is shown in the Fig. The input signal is applied to the diode sampling gate. At the start of each sampling cycle a trigger input pulse is generated which activates the blocking oscillator. The oscillator output is given to the ramp generator which generates the linear ramp signal. Since the sampling must be synchronized with the input signal frequency, the signal is delayed in the vertical amplifier. The staircase generator produces a staircase waveform which is applied to an attenuator. The attenuator controls the magnitude of the staircase signal and then it is applied to a voltage comparator. Another input to the voltage comparator is the output of the ramp generator. The voltage comparator compares the two signals and produces the output pulse when the two voltages are equal. This is nothing but a sampling pulse which is applied to sampling gate through the gate control circuitry. This pulse opens the diode gate and sample is taken in. This sampled signal is then applied to the vertical amplifier and the vertical deflecting plates. The output of the staircase generator is also applied to the horizontal deflecting plates. During each step of staircase the spot moves on the screen. The comparator output advances the staircase output through one step. After certain number of pulses about thousand or so, the
  • 5. staircase generator resets. The smaller the size of the steps of the staircase generator, larger is the number of samples and higher is the resolution of the image. Digital Storage Oscilloscope: In this digital storage oscilloscope, the waveform to be stored is digitized and then stored in a digital memory. The conventional cathode ray tube is used in this oscilloscope hence the cost is less. The power to be applied to memory is small and can be supplied by small battery. Due to this the stored image can be displayed indefinitely as long ,15 power is supplied to memory. Once the waveform is digitized then it can be further loaded into the computer and can be analysed in detail. Block Diagram: The block diagram of digital storage oscilloscope is shown in the Fig. As done in all the oscilloscopes, the input signal is applied to the amplifier and attenuator section. The oscilloscope uses same type of amplifier and attenuator circuitry as used in the conventional oscilloscopes. The attenuated signal is then applied to the vertical amplifier.
  • 6. The vertical input, after passing through the vertical amplifier, is digitised by an analog to digital converter to create a data set that is stored in the memory. The data set is processed by the microprocessor and then sent to the display. To digitise the analog signal, analog to digital (AID) converter is used. The output of the vertical amplifier is applied to the AID converter section. The main requirement of AID converter in the digital storage oscilloscope is its speed, while in digital voltmeters accuracy and resolution were the main requirements. The digitised output needed only in the binary form and not in BCD. The successive approximation type of AID converter is most oftenly used in the digital storage oscilloscopes. Modes of operation: The digital storage oscilloscope has three modes of operation: 1. Roll mode ii) Store mode iii) Hold or save mode. Roll mode: This mode is used to display very fast varying signals, clearly on the screen. The fast varying signal is displayed as if it is changing slowly, on the screen. In this mode, the input signal is not triggered at all.