2. 2
Measurements
Measurement of a given quantity is
essentially an act or result of comparison
between the quantity (whose magnitude is
unknown) and predetermined or predefined
standards.
Two quantities are compared the result is
expressed in numerical values.
3. 3
Basic requirements for a
meaningful measurement
The standard used for comparison purposes
must be accurately defined and should be
commonly accepted.
The apparatus used and the method adopted
must be provable (verifiable).
4. 4
Significance Of Measurement
Importance of Measurement is simply and
eloquently expressed in the following
statement of famous physicist Lord
Kelvin: ”I often say that when you can
measure what you are speaking about and
can express it in numbers, you know
something about it; when you cannot
express in it numbers your knowledge is
of meager and unsatisfactory kind”
5. 5
Two major functions of all branch
of engineering
Design of equipment and processes
Proper Operation and maintenance of
equipment and processes.
7. 7
DIRECT METHODS: In these methods, the
unknown quantity (called the measurand ) is
directly compared against a standard.
INDIRECT METHOD: Measurements by direct
methods are not always possible, feasible
and practicable. In engineering applications
measurement systems are used which require
need of indirect method for measurement
purposes.
8. 8
Instruments and Measurement
Systems.
Measurement involve the use of
instruments as a physical means of
determining quantities or variables.
Because of modular nature of the
elements within it, it is common to refer
the measuring instrument as a
MEASUREMENT SYSTEM.
9. 9
Evolution of Instruments.
a) Mechanical
b)Electrical
c) Electronic Instruments.
MECHANICAL: These instruments are
very reliable for static and stable conditions.
But their disadvantage is that they are unable
to respond rapidly to measurements of
dynamic and transient conditions.
10. 10
Contd
ELECTRICAL: It is faster than mechanical,
indicating the output are rapid than mechanical
methods. But it depends on the mechanical
movement of the meters. The response is 0.5 to
24 seconds.
ELECTRONIC: It is more reliable than other
system. It uses semiconductor devices and weak
signal can also be detected.
11. 11
Classification Of Instruments
Absolute Instruments.
Secondary Instruments.
ABSOLUTE: These instruments give the
magnitude if the quantity under
measurement terms of physical constants
of the instrument.
SECONDARY: These instruments are
calibrated by the comparison with absolute
instruments which have already been
calibrated.
13. 13
Functions of instrument and measuring
system can be classified into three. They are:
i) Indicating function.
ii) Recording function.
iii) Controlling function.
Application of measurement systems are:
i) Monitoring of process and operation.
ii) Control of processes and operation.
iii) Experimental engineering analysis.
14. 14
Types Of Instrumentation
System
Intelligent Instrumentation (data has been
refined for the purpose of presentation )
Dumb Instrumentation (data must be
processed by the observer)
15. 15
Elements of Generalized
Measurement System
Primary sensing element.
Variable conversion element.
Data presentation element.
PRIMARY SENSING ELEMENT: The
quantity under measurement makes its first
contact with the primary sensing element of a
measurement system.
VARIABLE CONVERSION ELEMENT: It
converts the output of the primary sensing
element into suitable form to preserve the
information content of the original signal.
16. 16
Contd..
DATA PRESENTATION ELEMENT:
The information about the quantity under
measurement has to be conveyed to the
personnel handling the instrument or the
system for monitoring, control or analysis
purpose.
17. 17
Functional Elements of an
Instrumentation System
PRIMARY
SENSING
ELEMENT
VARIABLE
CONVER
-SION
ELEMENT
VARIABLE
MANIPULATI-
ON ELEMENT
DATA
TRANSMISSIO
-N ELEMENT
DATA CONDITIONING ELEMENT
INTERMEDIATE STAGE
DETECTOR
TRANSDUCER
STAGE
TERMINATING
STAGE
QUANTITY
TO BE
MEASURED
DATA
PRESENTA
TION
ELEMENT
18. 18
Calibration
Calibration of all instruments is important since it
affords the opportunity to check the instruments
against a known standard and subsequently to find
errors and accuracy.
Calibration Procedure involve a comparison of the
particular instrument with either
a Primary standard
a secondary standard with a higher accuracy than
the instrument to be calibrated.
an instrument of known accuracy.
19. 19
Standards
A standard is a physical representation of
a unit of measurement. The term ‘standard’
is applied to a piece of equipment having a
known measure of physical quantity.
20. 20
Types of Standards
–International Standards (defined based
on international agreement )
–Primary Standards (maintained by
national standards laboratories)
–Secondary Standards ( used by industrial
measurement laboratories)
–Working Standards ( used in general
laboratory)
22. MOVING COIL INSTRUMENT – PERMANENT
MAGNET TYPE
Principle :
when a current carrying conductor is placed in
magnetic field it is acted upon by a force which tends
to move it to one side and out of the field. This
movement of coil is used to measure current or
voltage.
23. Construction
This instrument consists of a
permanent magnet and a rectangular
coil of many turns wound on a light
aluminum or copper former inside
which is an iron core
• The sides of the coil are free to
move in the two air gaps
between the poles and core
• To the moving coil spindle is
attached, a pointer is attached to the
spindle to move over a calibrated
scale.
24. Working
A magnetic field of sufficient density is produced
by the permanent magnet.
The moving coil carries the current or a current
proportional to the voltage to be measured.
Hence, an electromagnetic force is produced
which tends to act on the moving coil and moves it
away from the field.
This movement makes the spindle move and so the
pointer gives a proportionate deflection
25. Deflecting torque : It is directly proportional to the
current or the voltage to be measured. So, the instrument can
be used to measure direct current and dc voltage.
Control torque : Spring control.
Damping torque : Eddy current damping.Damping is
electromagnetic by eddy currents induced in the metal frame
over which the coil is wound. Since the frame moves in an
intense magnetic field, the induced eddy currents are large and
damping is very effective.
26. The permanent-magnet moving coil (PMMC) type instruments
have the following advantage and disadvantages:
ADVANTAGES
1. They have low power consumption
2. Their scales are uniform and can be designed to extend over and arc of
1700 degree or so
3. They possess high (torque/weight) ratio.
4. They can be modified what the help o f shunts and resistances to cover a
wide range of currents and voltages.
5. They have no hysteresis loss.
DISADVANTAGES
1. Due to delicate construction and the necessary accurate machining and
assembly of various parts, such instruments are somewhat costlier as
compared to moving iron instruments.
2. Some errors are set in due to the ageing of control springs and the
permanent magnets.
29. MOVING IRON INSTRUMENTS – ATTRACTION
TYPE
Principle
A soft iron piece gets magnetized when it is brought into a magnetic
field produced by a permanent magnet.
The same phenomenon happens when the soft iron piece is brought
near either of the ends of a coil carrying current.
The iron piece is attracted towards that portion where the magnetic flux
density is more.
This movement of soft iron piece is used to measure the current or
voltage which produces the magnetic field.
30. Construction
A soft iron disc is attached
to the spindle
To the spindle, a pointer is
also attached, which is
made to move over
calibrated scale
The moving iron is
pivoted such that it is
attracted towards the
center of the coil where
the magnetic field is
maximum
31. Principle
When the current to be measured is passed
through the coil or solenoid, field is
produced which attracts the eccentrically
mounted disc inwards, thereby deflection
the pointer which moves over a calibrated
scale
32. Deflecting Torque
Produced by the current or the voltage to be measured.
It is proportional to the square of the voltage or current.
Hence, the instrument can be used to measure d.c. or a.c.
Scale is non- uniform
Control torque : Spring or gravity
Damping : Air friction damping
33. MOVING IRON INSTRUMENT - REPULSION
TYPE
Principle
Two iron piece kept with close proximity in a
magnetic field get magnetized to the same polarity.
Hence, a repulsive force is produced.
If one of the two piece is made movable, the
repulsive force will act on it and move it on to one
side.
This movement is used to measure the current or
voltage which produces the magnetic field.
34. Construction
There are two iron pieces-fixed and moving.
The moving iron is connected to the spindle to
which is attached a pointer. It is made to move over a
calibrated scale.
35. Working
When the current to be measured is passed through the
fixed coil it sets up its own magnetic field which
magnetizes the two rods similarly the adjacent points on the
lengths of the rods will have the same magnetic polarity.
Hence, they repel each other with the result that the pointer
is deflected against the controlling torque of a spring or
gravity.
The force of repulsion is approximately proportional to the
square of the current passing through the coil
Whatever be the direction of current in the coil, the two
irons are always similarly magnetised.
36. Deflecting torque
Produced by the current or the voltage to be measured.
It is proportional to the square of the voltage or current.
Hence, the instrument can be used to measure d.c. or a.c.
Control torque : Spring or gravity
Damping : Air friction damping
39. Advantages and disadvantages:
• The instruments are cheap ,reliable and
robust
The instruments can be used on both A.C
and D.C
They cannot be calibrated with high degree
of precision with D.C on account of the
effect of hysteresis in the iron rods or
vanes .
41. The dynamometer type 3-phase wattmeter consist
of two separate wattmeter movements mounted
together in one case with the two moving coils
mounted on the same spindle.
The current and the pressure coil mounted
together is known as an element. Therefore a
3phase wattmeter has two elements.
The connection of two elements of 3 phase
wattmeter is shown in diagram.
41
42. Two Element Energy Meter
42
The construction of a three-phase
induction type energy meter appears
as an assembly of two single-phase
induction type energy meters in one
case, having a common spindle and
registering mechanism.
The schematic diagram of the meter
is shown in the below figure.
44. Working of Three Phase
Energy Meter :
The principle of working of a 3-phase energy meter
is similar to the single-phase energy meter. When
the load is connected to the meter both the
pressure coil and the current coil of the two
elements mounted on the shunt and series magnet
produces magnetic flux. This flux when links with
the discs cause an eddy current to flow in it.
Interaction of eddy currents with the flux imposed
by the two coils causes the production of torque on
discs. Since two discs are attached to one spindle,
the torque exerted on the two discs added
mechanically. Hence, rotation of the shaft gives the
3-phase energy consumed.
44
45. • The essential parts of the 3-phase
energy meter are,
• Driving system
• Moving system
• Braking system
• Registering system.
45
46. Driving System:
The combination of a shunt and series magnet is called an
element. So, it consists of two elements. The connections for
the winding of these electromagnets are shown in the figure.
The shunt magnet windings are provided such that, during
no-load conditions, the torque developed by both the shunt
magnets is opposite in nature.
In order to make the resultant driving torque equal to zero, a
magnetic shunt is provided in the meter. Its position is
adjusted until the discs stop rotating at no-load. The
necessary driving torque is obtained due to the interaction
between the shunt and series magnetic fields on the disc in
each element.
46
Moving System :
The moving system consists of two aluminum discs (one for
each element) which are mounted over a single spindle. The
torque developed by each disc will be added up and as a
result, the total torque will be proportional to the 3-phase
power consumed by the load.
47. Braking System :
This system provides the necessary braking action on the
discs. An individual permanent magnet is provided for each
disc. There is a provision for adjusting the position of each
brake magnet in order to vary the braking torque.
Registering System :
This system is attached to the moving system through the
pinion and a gear train. It continuously counts or registers
the number of revolutions made by the discs. It means it
integrates the power consumed by the three-phase load over
a considered period of time which is nothing but energy.
47
50. 50
What is Instrument Transformer ?
A transformer that is used in conjunction with a
measuring instrument.
It utilizes the current-transformation and voltage
transformation properties to measure high ac current and
voltage.
51. 51
Importance of Instrument
transformers
In dc circuits for current and voltage measurement we use
ammeters and voltmeters.
For measurement of high current ,it is usual to use low
range ammeter with suitable shunt.
For measurement of high voltage, low range voltmeter
are used with high resistance connected in series.
But for measurement of high A.C. current and voltage we
cannot use these methods.
We use specially constructed instrument transformers.
52. 52
Types of instrument transformers
These instrument transformers are of two types:-
1. Current transformers
2. Potential transformers
53. 53
Current Transformers
Current transformer normally known as c.t. is a step up transformer.
These are used with low range ammeter to measure current in high
voltage alternating circuits where it is not practical to connect
instrument and meters directly to lines.
This is step up transformer because when we step up the, voltage
increases and current decreases.
The current is step down in a known ratio called current ratio.
54. 54
Construction of C.T.
C.T. has a primary coil of
one or more turns of thick
wire connected in series
with the line whose current
is to be measured.
The secondary consist of
large number of turns of
fine wire, is connected
across the ammeter
terminals.
55. 55
Working
If a current transformer has primary to secondary current
ratio of 100:5 then it step up the voltage 20 times and
step down the current 1/20 times of its actual value.
If we know the current ratio I1/I2 and the reading of a.c.
ammeter, the current can be calculated.
Current = ratio × ammeter reading
56. 56
Importance of short ckt.
Ammeter resistance is very low ,the
current transformer normally works
short circuited.
If for any reason the ammeter is taken
out of secondary winding then the
secondary winding must be short ckted
with the help of short ckt switch s.
If this is not done, then due to high
m.m.f. will set up high flux in the core
and it will produces excessive core
loss which produce heat and high
voltage across the secondary terminals
Hence the secondary of current
transformer is never left open.
58. 58
Potential transformer
A PT is a step down transformer having many primary
turns but few secondary turns.
In step down the voltage decreases and current increases,
thus voltage can be easily measured by using low range
voltmeter.
The voltage is stepped down in known ratio called
voltage ratio.
59. 59
Construction and working of P.T.
Construction
A potential transformer has many primary windings but few number of
secondary windings that makes it step down transformer.
Voltmeter is connected to secondary winding usually voltmeter of 150 v is
suitable.
Working
Primary terminals are connected across the line to which the voltage is to be
measured.
The voltmeter gives the transformed value of voltage at secondary.
The deflection of voltmeter when divided by transformed ratio gives the actual
voltage at primary.
Line voltage = deflection / trasf. Ratio
Where transformation ratio = V2/V1
60. 60
Precaution for P.T.
Since the secondary of p.t. is connected to
relays, their ratings are usually 40 to 100
Watts.
For safety purpose the secondary should be
completely insulated from the high voltage
primary and should be in addition grounded.
61. 61
Types of P.T.
Some types of p.t. are
1. Shell type
2. Dry type
3. Oil type
Rating Type
below 5000 v Shell type
5000-13800 v Dry type and
oil type
above 13800 v only oil type
63. 63
Digital Storage Oscilloscope
Definition: The digital storage oscilloscope is defined as the
oscilloscope which stores and analysis the signal digitally,
i.e. in the form of 1 or 0 preferably storing them as analogue
signals. The digital oscilloscope takes an input signal, store
them and then display it on the screen. The digital
oscilloscope has advanced features of storage, triggering and
measurement. Also, it displays the signal visually as well as
numerically.
64. 64
Working Principle of Digital Storage
Oscilloscope
The digital oscilloscope digitizes and stores the input
signal. This can be done by the use of CRT (Cathode ray
tube) and digital memory. The block diagram of the basic
digital oscilloscope is shown in the figure below. The
digitization can be done by taking the sample input
signals at periodic waveforms.
65. 65
Digital Storage Oscilloscope
The maximum frequency of the signal
which is measured by the digital
oscilloscope depends on the two factors.
Theses factors are the
1. Sampling rate
2. Nature of converter.
66. Sampling Rate – For safe analysis of input signal the sampling
theory is used. The sampling theory states that the sampling rate of
the signal must be twice as fast as the highest frequency of the input
signal. The sampling rate means analogue to digital converter has a
high fast conversion rate.
Converter – The converter uses the expensive flash whose
resolution decreases with the increases of a sampling rate. Because
of the sampling rate, the bandwidth and resolution of the
oscilloscope are limited. The need of the analogue to digital signal
converters can also be overcome by using the shift register. The
input signal is sampled and stored in the shift register. From the shift
register, the signal is slowly read out and stored in the digital form.
This method reduces the cost of the converter and operates up to
100 mega sample per second.
66
The only disadvantage of the digital oscilloscope is that
it does not accept the data during digitization, so it had a
blind spot at that time.
67. 67
Waveform Reconstruction
For visualizing the final wave, the oscilloscopes use the technique of
inter-polarization. The inter-polarization is the process of creating
the new data points with the help of known variable data points.
Linear interpolation and sinusoidal interpolation are the two
processes of connecting the points together.
68. Applications
The applications of the DSO are
• It checks faulty components in circuits
• Used in the medical field
• Used to measure capacitor, inductance, time interval between
signals, frequency and time period
• Used to observe transistors and diodes V-I characteristics
• Used to analyze TV waveforms
• Used in video and audio recording equipment’s
• Used in designing
• Used in the research field
• For comparison purpose, it displays 3D figure or multiple
waveforms
• It is widely used an oscilloscope 68
69. Advantages
The advantages of the DSO are
• Portable
• Have the highest bandwidth
• The user interface is simple
• Speed is high
Disadvantages
The disadvantages of the DSO are
• Complex
• High cost
69
70. Data Acquisition System:
A typical Data Acquisition System consists of individual
sensors with the necessary signal conditioning, data
conversion, data processing, multiplexing, data handling
and associated transmission, storage and display systems.
In order to optimize the characteristics of the system in
terms of performance, handling capacity and cost, the
relevant sub systems can be combined together. Analog
Data Acquisition System is generally acquired and
converted into digital form for the purpose of processing,
transmission, display and storage.
70
71. Introduction
A data acquisition system is a collection of software and
hardware that allows one to measure or control physical
characteristics of something in the real world. Data
acquisition systems can be classified into the following two
types. Analog Data Acquisition Systems The data
acquisition systems, which can be operated with analog
signals are known as analog data acquisition systems.
Digital Data Acquisition Systems The data acquisition
systems, which can be operated with digital signals are
known as digital data acquisition systems. So, they use
digital components for storing or displaying the information
71
72. The components of DAS include:
– Sensors that convert physical parameters to
electrical signals.
– Signal conditioning circuitry to convert sensor
signals into a form that can be converted into
digital values.
– Analog-to-digital converters that convert
conditioned sensor signals to digital values.
72