A Seminar as a part of curricular requirement
for I year M.Pharm II Semester
Presented by
Ms. T.Jayasree
(Reg.No 20L81S0709)
Pharmaceutical Analysis
QUALIFICATION OF
GAS CHROMATOGRAPHY
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 2
Qualification
Gas Chromatography (GC)
 Design Qualification (DQ)
 Installation Qualification (IQ)
 Operational Qualification (OQ)
 Performance Qualification (PQ)
References
Contents
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 3
• Qualification- It is defined as the “action of proving that any
equipment works correctly and leads to the expected results”.
Qualification is also part of validation and is product specific.
• Qualification is often a part (the intial stage) of validation ,but the
individual qualification steps alone do not constitute process
validation.
Qualification
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 4
• There are four stages of qualification
1) Design Qualification (DQ)
2) Installation Qualification (IQ)
3) Operational Qualification (OQ)
4) Performance Qualification (PQ)
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 5
• validation and qualification are essentially components of the same
concept.
• The term qualification is normally used for equipment , utilities and
systems.
• Validation is used for processes.
• In this sense, qualification is part of validation.
Relationship between validation and qualification
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 6
GAS CHROMATOGRAPHY (GC)
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 7
• Supplier must provide documented evidence that the product has
been designed, developed and manufactured in a quality
environment e.g. ISO 9001:2000 certification.
• Supplier must provide phone and on-site support in case of defects.
• Supplier must provide information through the internet on
availability of new firmware upgrades.
Design Qualification
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 8
• Equipment is compared as received with purchase order, including
software, accessories, spare parts and consumables.
• Documentation checked for completeness of operating, maintenance
instructions, standard operating procedures for testing and safety,
validation certificates and health and safety instruments.
• Equipment is checked for any damage.
• The supplier’s instruction for installation is read.
Installation Qualification
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 9
• The supplier’s safety instructions are read, if there are any.
• Hardware (computer, equipment, fittings and tubings for fluid
connections, power cable, data flow and instrument control cables) is
installed following the manufacturer’s recommendation.
• The instruments are switched on and ensured that all modules power
up and an electronic self test is performed.
• Any deviations are recorded.
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 10
• Software is installed on computer following the manufacturer’s
recommendation.
• Correct software installation is verified.
• A backup copy of software is made.
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 11
Precision of peak retention times
Precision of peak areas
Operational Qualification
Test procedure Acceptance limits Test frequency
Five injections of a
standard.
Calculation of relative
standard deviation.
<1% RSD (Relative
Standard Deviation)
Yearly
Test procedure Acceptance limits Test frequency
Five injections of a
standard.
Calculation of relative
standard deviation.
<2% RSD Yearly
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 12
Test
procedure
Acceptance
limits
Test frequency Remarks
Temperature in
column oven is
measured and
compared with
set point.
±1°C Yearly
The
temperature
measurement
device should
be calibrated
and traceable to
a national
standard.
Accuracy of the temperature of the column oven
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 13
Precision of heated zone temperature
Test
procedure
Acceptance
Limits
Test
frequency Remarks
Actual
temperature is
measured and
compared with
set points at
50,70 and
90°C.
2°C Yearly
The
temperature
measurement
device should
be calibrated
and traceable to
a national
standard.
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 14
Precision of injection
Test procedure Acceptance limits Test frequency
Five injections of a
standard.
Calculation of
relative
Standard deviation.
<2% RSD Yearly
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 15
Test procedure Acceptance limits Test frequency
Blank solvent is
injected after
standard.
Peak ratio is
measured between
blank and standard
injection.
<5% Yearly
Carryover of Injection
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 16
Solution A 10 ml of methanol, ethanol, acetone make upto 100ml
with ethyl acetate.
Solution B 15ml of methanol, ethanol, acetone make upto 100ml
with ethyl acetate.
Solution C 20 ml of methanol, ethanol, acetone make upto 100ml
with ethyl acetate.
Solution D 25 ml of methanol, ethanol, acetone make upto 100ml
with ethyl acetate.
Solution E 30 ml of methanol, ethanol, acetone make upto 100ml
with ethyl acetate.
Performance Qualification
Detector Linearity
Detector linearity solutions:
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 17
Procedure Acceptance Criteria
Blank is injected followed by
detector linearity solutions and
the peak responses are recorded.
A standard plot is drawn between
the concentrations vs the peak
responses.
The plot should be linear and
regression coefficient (R2) should
not be less than 0.9999.
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 18
• The digital flow meter is connected to the detector outlet port.
• The carrier gas flow is set and wait till it reaches the set flow.
• The observed flow in replicate is noted.
• The procedure is repeated for other carrier gases such as hydrogen
and air.
• The flow rate of carrier gas should be ±10% of set flow.
Flow rate accuracy
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 19
• 20ml of methanol, ethanol and acetone is transferred into 100ml
volumetric flask and made up with ethyl acetate.
• Blank is injected followed by standard preparation in six replicates.
• The areas and retention times are noted down.
• The %RSD (Relative standard deviation) of retention time should not
be more than 1.0% and peak area not more than 5.0%.
System precision
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 20
• After GC is ready, the system is run up to 15 minutes through single
run.
• When the run is completed noise and drift is calculated through
software.
• The acceptance values are:
a) Noise not more than 100µV.
b) Drift not more than 2500µV/hr.
Detector noise and Drift test
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 21
1. D. Gowrisankar, K. Abbulu , O. Bala souri, K. Sujana, Validation
and Calibration of Analytical instruments. 2010;2(2).
2. McDowall RD. Method validation in gas chromatography. Gas
chromatographic techniques and applications, 2001:201-2.
3. USP 30 NF 25 “The official compendia of standards”, Asian
edition, 2007, vol: 1.
References
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 22
4. Chan CC, Lam H, Zhang XM. Practical approaches to method
validation and essential instrument qualification. John Wiley &
Sons; 2011 Mar 1.
5. A.H.Beckett, J.B.Stenlake, Practical Pharmaceutical chemistry, CBS
Publishers and distributors, IV-Edition, part two, 2004, p.no.326.
6. Y.Anjaneyulu ,R.Marayya ,Quality Assurance and Quality
Management in Pharmaceutical industry , Pharmabook Syndicate ,
2005.
RIPER
AUTONOMOUS
NAAC &
NBA (UG)
SIRO- DSIR
Raghavendra Institute of Pharmaceutical Education and Research - Autonomous
K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 23

More Related Content

PPTX
Qualification of gc equipment
PDF
Notes for the subject 'Pharmaceutical Validation'
PPTX
Qualification of GC & FTIR
PPTX
who guidelines for herbal drugs
PPTX
Pesticide analysis
PPTX
Validation of dry heat sterilization
PPTX
Process validation
Qualification of gc equipment
Notes for the subject 'Pharmaceutical Validation'
Qualification of GC & FTIR
who guidelines for herbal drugs
Pesticide analysis
Validation of dry heat sterilization
Process validation

What's hot (20)

PPTX
validation of analytical method used in cleaning
PPTX
PPTX
Concept &amp; evolution of qa &amp; qc
PPTX
Cleaning validation
PPTX
Pharmaceutical validation, Quantification, streamlining, validation masterplan.
PPTX
Qualification of Dissolution Test Apparatus and Validation of Utility System
PPTX
Validation of pharaceutical water system and pure steam
PPTX
Analytical method validation as per ich and usp
PPTX
Manufacturing operations and controls.pptx
PPTX
Validation utility system
PPTX
Qualification of laboratory equipments
PPTX
Qualification of analytical instruments
PPTX
Qualification of high performance liquid chromatography
PPTX
PPTX
Qualification of analytical instruments
validation of analytical method used in cleaning
Concept &amp; evolution of qa &amp; qc
Cleaning validation
Pharmaceutical validation, Quantification, streamlining, validation masterplan.
Qualification of Dissolution Test Apparatus and Validation of Utility System
Validation of pharaceutical water system and pure steam
Analytical method validation as per ich and usp
Manufacturing operations and controls.pptx
Validation utility system
Qualification of laboratory equipments
Qualification of analytical instruments
Qualification of high performance liquid chromatography
Qualification of analytical instruments
Ad

Similar to Qualification of Gas Chromatography (20)

PPTX
Pharmaceutical validation of water system
PPTX
PPTX
Computerized System Validation (CSV) PPT
PPTX
Quality-by-Design in Pharmaceutical Development
PPTX
ilovepdf_merged (1).pptx
PPTX
ipqc seminor.pptx
PDF
Cleaning validation and regulatory guidelines
PDF
Process validation of liquid orals and aerosals
PPTX
Pharmaceutical validation of water system
Computerized System Validation (CSV) PPT
Quality-by-Design in Pharmaceutical Development
ilovepdf_merged (1).pptx
ipqc seminor.pptx
Cleaning validation and regulatory guidelines
Process validation of liquid orals and aerosals
Ad

More from Raghavendra institute of pharmaceutical education and research . (20)

PPTX
JOURNAL CLUB PRESENTATION (20L81S0710-PA)
PPTX
JOURNAL CLUB PRESENTATION (20L81S0714-PA )
PPTX
JOURNAL CLUB PRESENTATION (20L81S0706-PA)
PPTX
JOURNAL CLUB PRESENTATION (20L81S0712-PA)
PPTX
JOURNAL CLUB PRESENTATION (20L81S0713-PA)
PPTX
JOURNAL CLUB PRESENTATION (20L81S0707-PA)
PPTX
JOURNAL CLUB PRESENTATION (20L81S0702-PA)
PPT
JOURNAL CLUB PRESENTATION (20L81S0705-PA)
PPTX
JOURNAL CLUB PRESENTATION (20L81S0402-PA & QA)
PPTX
JOURNAL CLUB PRESENTATION (20L81S0401-PA & QA)
PPTX
JOURNAL CLUB PRESENTATION (20L81S0703-PA)
PPTX
JOURNAL CLUB PRESENTATION (20L81S0701-PA)
PPTX
JOURNAL CLUB PRESENTATION (20L81S0709-PA)
PPTX
JOURNAL CLUB PRESENTATION (20L81S0711-PA)
PPTX
JOURNAL CLUB PRESENTATION (20L81S0704-PA )
PPTX
JOURNAL CLUB PRESENTATION (20L81S0708-PA)
PPTX
JOURNAL CLUB PRESENTATION (20L81S0403-PA & QA)
PPTX
Study of Quality of Raw Materials and General methods of analysis of Raw mate...
JOURNAL CLUB PRESENTATION (20L81S0710-PA)
JOURNAL CLUB PRESENTATION (20L81S0714-PA )
JOURNAL CLUB PRESENTATION (20L81S0706-PA)
JOURNAL CLUB PRESENTATION (20L81S0712-PA)
JOURNAL CLUB PRESENTATION (20L81S0713-PA)
JOURNAL CLUB PRESENTATION (20L81S0707-PA)
JOURNAL CLUB PRESENTATION (20L81S0702-PA)
JOURNAL CLUB PRESENTATION (20L81S0705-PA)
JOURNAL CLUB PRESENTATION (20L81S0402-PA & QA)
JOURNAL CLUB PRESENTATION (20L81S0401-PA & QA)
JOURNAL CLUB PRESENTATION (20L81S0703-PA)
JOURNAL CLUB PRESENTATION (20L81S0701-PA)
JOURNAL CLUB PRESENTATION (20L81S0709-PA)
JOURNAL CLUB PRESENTATION (20L81S0711-PA)
JOURNAL CLUB PRESENTATION (20L81S0704-PA )
JOURNAL CLUB PRESENTATION (20L81S0708-PA)
JOURNAL CLUB PRESENTATION (20L81S0403-PA & QA)
Study of Quality of Raw Materials and General methods of analysis of Raw mate...

Recently uploaded (20)

PPTX
Introduction to Fisheries Biotechnology_Lesson 1.pptx
PPTX
GEN. BIO 1 - CELL TYPES & CELL MODIFICATIONS
PDF
. Radiology Case Scenariosssssssssssssss
PPTX
famous lake in india and its disturibution and importance
PPTX
SCIENCE10 Q1 5 WK8 Evidence Supporting Plate Movement.pptx
PDF
SEHH2274 Organic Chemistry Notes 1 Structure and Bonding.pdf
PPTX
Protein & Amino Acid Structures Levels of protein structure (primary, seconda...
PPTX
microscope-Lecturecjchchchchcuvuvhc.pptx
PPTX
Classification Systems_TAXONOMY_SCIENCE8.pptx
PDF
diccionario toefl examen de ingles para principiante
PPTX
ognitive-behavioral therapy, mindfulness-based approaches, coping skills trai...
PPTX
G5Q1W8 PPT SCIENCE.pptx 2025-2026 GRADE 5
PDF
Sciences of Europe No 170 (2025)
PPTX
2. Earth - The Living Planet Module 2ELS
PPT
protein biochemistry.ppt for university classes
PDF
The scientific heritage No 166 (166) (2025)
PPTX
DRUG THERAPY FOR SHOCK gjjjgfhhhhh.pptx.
PDF
MIRIDeepImagingSurvey(MIDIS)oftheHubbleUltraDeepField
PPTX
ECG_Course_Presentation د.محمد صقران ppt
PPTX
The KM-GBF monitoring framework – status & key messages.pptx
Introduction to Fisheries Biotechnology_Lesson 1.pptx
GEN. BIO 1 - CELL TYPES & CELL MODIFICATIONS
. Radiology Case Scenariosssssssssssssss
famous lake in india and its disturibution and importance
SCIENCE10 Q1 5 WK8 Evidence Supporting Plate Movement.pptx
SEHH2274 Organic Chemistry Notes 1 Structure and Bonding.pdf
Protein & Amino Acid Structures Levels of protein structure (primary, seconda...
microscope-Lecturecjchchchchcuvuvhc.pptx
Classification Systems_TAXONOMY_SCIENCE8.pptx
diccionario toefl examen de ingles para principiante
ognitive-behavioral therapy, mindfulness-based approaches, coping skills trai...
G5Q1W8 PPT SCIENCE.pptx 2025-2026 GRADE 5
Sciences of Europe No 170 (2025)
2. Earth - The Living Planet Module 2ELS
protein biochemistry.ppt for university classes
The scientific heritage No 166 (166) (2025)
DRUG THERAPY FOR SHOCK gjjjgfhhhhh.pptx.
MIRIDeepImagingSurvey(MIDIS)oftheHubbleUltraDeepField
ECG_Course_Presentation د.محمد صقران ppt
The KM-GBF monitoring framework – status & key messages.pptx

Qualification of Gas Chromatography

  • 1. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 1 A Seminar as a part of curricular requirement for I year M.Pharm II Semester Presented by Ms. T.Jayasree (Reg.No 20L81S0709) Pharmaceutical Analysis QUALIFICATION OF GAS CHROMATOGRAPHY
  • 2. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 2 Qualification Gas Chromatography (GC)  Design Qualification (DQ)  Installation Qualification (IQ)  Operational Qualification (OQ)  Performance Qualification (PQ) References Contents
  • 3. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 3 • Qualification- It is defined as the “action of proving that any equipment works correctly and leads to the expected results”. Qualification is also part of validation and is product specific. • Qualification is often a part (the intial stage) of validation ,but the individual qualification steps alone do not constitute process validation. Qualification
  • 4. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 4 • There are four stages of qualification 1) Design Qualification (DQ) 2) Installation Qualification (IQ) 3) Operational Qualification (OQ) 4) Performance Qualification (PQ)
  • 5. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 5 • validation and qualification are essentially components of the same concept. • The term qualification is normally used for equipment , utilities and systems. • Validation is used for processes. • In this sense, qualification is part of validation. Relationship between validation and qualification
  • 6. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 6 GAS CHROMATOGRAPHY (GC)
  • 7. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 7 • Supplier must provide documented evidence that the product has been designed, developed and manufactured in a quality environment e.g. ISO 9001:2000 certification. • Supplier must provide phone and on-site support in case of defects. • Supplier must provide information through the internet on availability of new firmware upgrades. Design Qualification
  • 8. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 8 • Equipment is compared as received with purchase order, including software, accessories, spare parts and consumables. • Documentation checked for completeness of operating, maintenance instructions, standard operating procedures for testing and safety, validation certificates and health and safety instruments. • Equipment is checked for any damage. • The supplier’s instruction for installation is read. Installation Qualification
  • 9. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 9 • The supplier’s safety instructions are read, if there are any. • Hardware (computer, equipment, fittings and tubings for fluid connections, power cable, data flow and instrument control cables) is installed following the manufacturer’s recommendation. • The instruments are switched on and ensured that all modules power up and an electronic self test is performed. • Any deviations are recorded.
  • 10. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 10 • Software is installed on computer following the manufacturer’s recommendation. • Correct software installation is verified. • A backup copy of software is made.
  • 11. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 11 Precision of peak retention times Precision of peak areas Operational Qualification Test procedure Acceptance limits Test frequency Five injections of a standard. Calculation of relative standard deviation. <1% RSD (Relative Standard Deviation) Yearly Test procedure Acceptance limits Test frequency Five injections of a standard. Calculation of relative standard deviation. <2% RSD Yearly
  • 12. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 12 Test procedure Acceptance limits Test frequency Remarks Temperature in column oven is measured and compared with set point. ±1°C Yearly The temperature measurement device should be calibrated and traceable to a national standard. Accuracy of the temperature of the column oven
  • 13. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 13 Precision of heated zone temperature Test procedure Acceptance Limits Test frequency Remarks Actual temperature is measured and compared with set points at 50,70 and 90°C. 2°C Yearly The temperature measurement device should be calibrated and traceable to a national standard.
  • 14. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 14 Precision of injection Test procedure Acceptance limits Test frequency Five injections of a standard. Calculation of relative Standard deviation. <2% RSD Yearly
  • 15. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 15 Test procedure Acceptance limits Test frequency Blank solvent is injected after standard. Peak ratio is measured between blank and standard injection. <5% Yearly Carryover of Injection
  • 16. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 16 Solution A 10 ml of methanol, ethanol, acetone make upto 100ml with ethyl acetate. Solution B 15ml of methanol, ethanol, acetone make upto 100ml with ethyl acetate. Solution C 20 ml of methanol, ethanol, acetone make upto 100ml with ethyl acetate. Solution D 25 ml of methanol, ethanol, acetone make upto 100ml with ethyl acetate. Solution E 30 ml of methanol, ethanol, acetone make upto 100ml with ethyl acetate. Performance Qualification Detector Linearity Detector linearity solutions:
  • 17. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 17 Procedure Acceptance Criteria Blank is injected followed by detector linearity solutions and the peak responses are recorded. A standard plot is drawn between the concentrations vs the peak responses. The plot should be linear and regression coefficient (R2) should not be less than 0.9999.
  • 18. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 18 • The digital flow meter is connected to the detector outlet port. • The carrier gas flow is set and wait till it reaches the set flow. • The observed flow in replicate is noted. • The procedure is repeated for other carrier gases such as hydrogen and air. • The flow rate of carrier gas should be ±10% of set flow. Flow rate accuracy
  • 19. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 19 • 20ml of methanol, ethanol and acetone is transferred into 100ml volumetric flask and made up with ethyl acetate. • Blank is injected followed by standard preparation in six replicates. • The areas and retention times are noted down. • The %RSD (Relative standard deviation) of retention time should not be more than 1.0% and peak area not more than 5.0%. System precision
  • 20. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 20 • After GC is ready, the system is run up to 15 minutes through single run. • When the run is completed noise and drift is calculated through software. • The acceptance values are: a) Noise not more than 100µV. b) Drift not more than 2500µV/hr. Detector noise and Drift test
  • 21. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 21 1. D. Gowrisankar, K. Abbulu , O. Bala souri, K. Sujana, Validation and Calibration of Analytical instruments. 2010;2(2). 2. McDowall RD. Method validation in gas chromatography. Gas chromatographic techniques and applications, 2001:201-2. 3. USP 30 NF 25 “The official compendia of standards”, Asian edition, 2007, vol: 1. References
  • 22. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 22 4. Chan CC, Lam H, Zhang XM. Practical approaches to method validation and essential instrument qualification. John Wiley & Sons; 2011 Mar 1. 5. A.H.Beckett, J.B.Stenlake, Practical Pharmaceutical chemistry, CBS Publishers and distributors, IV-Edition, part two, 2004, p.no.326. 6. Y.Anjaneyulu ,R.Marayya ,Quality Assurance and Quality Management in Pharmaceutical industry , Pharmabook Syndicate , 2005.
  • 23. RIPER AUTONOMOUS NAAC & NBA (UG) SIRO- DSIR Raghavendra Institute of Pharmaceutical Education and Research - Autonomous K.R.Palli Cross, Chiyyedu, Anantapuramu, A. P- 515721 23