Basic Electron Microscopy
Arthur Rowe
The Knowledge Base at a Simple Level
Introduction
These 3 presentations cover the
fundamental theory of electron microscopy
In presentation #2 we cover:
– lens aberrations and their importance
– how we correct for lens astigmatism
– limits to ultimate resolution of the TEM
– Interactions of electrons with matter
aberrations of electromagnetic
lenses
the most important ones to consider
are:
• spherical aberration
• chromatic aberration
• astigmatism
spherical aberration
object plane
• arises because a simple lens is more powerful at the edge than
at the centre
• is not a problem with glass lenses (can be ground to shape)
• disc of minimum confusion results instead of point focus:
• is not correctable for electromagnetic lenses
coping with spherical aberration
• disc of minimum confusion has diameter given by:
d = C 
{C = constant}
• hence reducing  gives a large reduction in d
• . . . but for optimal resolution we need large  !
• best compromise is with  = 10-3 radians (= f/500)
• gives resolution = 0.1 nm - can not be bettered
chromatic aberration
• light of differentbrought to different focal positions
• for electrons can be controlled by fixed KV and lens currents
• but  of electrons can change by interaction with specimen !
• rule of thumb: resolution >= (specimen thickness)/10
astigmatism
minimal confusion
• arises when the lens is more powerful in one plane
than in the plane normal to it
• causes points to be imaged as short lines, which ‘flip’ through
90 degrees on passing through ‘focus’ (minimal confusion)
astigmatism - arises from:
• inherent geometrical defects in ‘circular’ bore of lens
• inherent inhomogeneities in magnetic properties of pole piece
• build-up of contamination on bore of pole-piece and on
apertures gives rise to non-conducting deposits which become
charged as electron strike them
• hence astigmatism is time-dependent
• and cannot be ‘designed out’
• inevitably requires continuous correction
astigmatism - correction:
• with glass optics (as in spectacles) astigmatism is corrected
using an additional lens of strength & asymmetry
opposed to the asymmetry of the basic (eye) lens
• with electron optics, same principle employed:
electrostatic stigmator lens apposed to main lens
strength & direction of its asymmetry user-variable
• only the OBJECTIVE lens needs accurate correction
• correction usually good for 1-2 hours for routine work
The TEM Column
_ Gun emits electrons
_ Electric field accelerate
_ Magnetic (and electric) field
control path of electrons
_ Electron wavelength @ 200KeV
 2x10-12 m
_ Resolution normally achievable
@ 200KeV  2 x 10-10 m  2Å
depth of focus - depth of field
• depth of useful focus (in the specimen) is primarily limited by
chromatic aberration effects
• the absolute depth of focus is larger than this: for all practical
purposes, everything is in focus to same level
• . . . So one cannot rack through focus (as in a light or even
scanning electron) microscope
• depth of field (in the image plane) is - for all practical purposes
infinite
when electrons hit matter ..
when electrons hit matter ..
(1) they may collide with an inner shell electron, ejecting same
> the ejected electron is a low-energy, secondary electron
- detected & used to from SEM images
> the original high-energy electron is scattered
- known as a ‘back-scattered’ electron (SEM use)
> an outer-shell electron drops into the position formerly
occupied by the ejected electron
> this is a quantum process, so a X-ray photon of precise
wavelength is emitted - basis for X-ray
microanalysis
when electrons hit matter ..
when electrons hit matter ..
(2) they may collide or nearly collide with an atomic nucleus
> undergo varying degree ofdeflection (inelastic scattering)
> undergo loss of energy - again varying
> lost energy appears as X-rays of varying wavelength
> this X-ray continuum is identical to that originating from
an X-ray source/generator (medical, XRC etc)
> original electrons scattered in a forward direction will
enter the imaging system, but with ‘wrong’ 
> causes a ‘haze’ and loss of resolution in image
when electrons hit matter ..
when electrons hit matter ..
(3) they may collide with outer shell electrons
> either removing or inserting an electron
> results in free radical formation
> this species is extremely chemically active
> reactions with neighbouring atoms induce massive change
in the specimen, especially in the light atoms
> this radiation damage severely limits possibilities of EM
> examination of cells in the live state NOT POSSIBLE
> all examinations need to be as brief (low dose) as possible
when electrons hit matter ..
when electrons hit matter ..
(4) they may pass through unchanged
> these transmitted electrons can be used to form an image
> this is called imaging by subtractive contrast
> can be recorded by either
(a) TV-type camera (CCD) - very expensive
(b) photographic film - direct impact of electrons
Photographic film
> silver halide grains detect virtually every electron
> at least 50x more efficient than photon capture !
when electrons hit matter ..
‘beam damage’occurs:
• light elements (H, O) lost very rapidly
• change in valency shell means free radicals formed
• . . .& consequent chemical reactions causing further damage
• beam damage is minimised by use of
• low temperatures (-160°)
• high beam voltages
• minimal exposure times

More Related Content

PPT
BASICS OF SEM , INCLUDES ALL THE DEFINITIONS AND WORKING MODES OF SEM
PPT
sem application and properties and tem applications and properties
PDF
Otago 2019 1
PPTX
Electron microscopy
PPT
PPTX
Transmission electron microscopy donload pptx
PPT
Electron Microscopy
PPT
SEM1.ppt BBB HA BAN ABD AJJD ASSK JD AK
BASICS OF SEM , INCLUDES ALL THE DEFINITIONS AND WORKING MODES OF SEM
sem application and properties and tem applications and properties
Otago 2019 1
Electron microscopy
Transmission electron microscopy donload pptx
Electron Microscopy
SEM1.ppt BBB HA BAN ABD AJJD ASSK JD AK

Similar to basicem2.ppt (20)

PPTX
Sem mahfooz
PPTX
Sem mahfooz
PPT
overview on scanning electron microscope
PPTX
Scanning Electron Microscopy - SEM
PDF
Imaging And Electron Physicsaberrationcorrected Electron Microscopy Peter W H...
PPTX
TEM.pptx
PPTX
Transmission Electron Microscopy
PPTX
scanning electron microscope for analysis
PDF
Introduction to Electron Microscopyy.pdf
PPTX
Sem mahfooz
PPT
Transmission and electron microscope Tem ppt
PPTX
Transmission electron microscope (TEM)
PPTX
Microscopy and Micrometry
PDF
Scanning Electron Microscopy
PPT
Transmission electron microscopy- Copy (2).ppt
PPT
Electron microscope
PPT
PPTX
SEM-for MS.pptx
PPT
SCANNING ELECTRON MICROSCOPE MITHILESH CHOUDHARY
Sem mahfooz
Sem mahfooz
overview on scanning electron microscope
Scanning Electron Microscopy - SEM
Imaging And Electron Physicsaberrationcorrected Electron Microscopy Peter W H...
TEM.pptx
Transmission Electron Microscopy
scanning electron microscope for analysis
Introduction to Electron Microscopyy.pdf
Sem mahfooz
Transmission and electron microscope Tem ppt
Transmission electron microscope (TEM)
Microscopy and Micrometry
Scanning Electron Microscopy
Transmission electron microscopy- Copy (2).ppt
Electron microscope
SEM-for MS.pptx
SCANNING ELECTRON MICROSCOPE MITHILESH CHOUDHARY
Ad

Recently uploaded (20)

PDF
Warm, water-depleted rocky exoplanets with surfaceionic liquids: A proposed c...
PPT
1. INTRODUCTION TO EPIDEMIOLOGY.pptx for community medicine
PDF
CHAPTER 3 Cell Structures and Their Functions Lecture Outline.pdf
PDF
The Land of Punt — A research by Dhani Irwanto
PDF
Cosmic Outliers: Low-spin Halos Explain the Abundance, Compactness, and Redsh...
PPTX
Pharmacology of Autonomic nervous system
PPTX
perinatal infections 2-171220190027.pptx
PDF
Phytochemical Investigation of Miliusa longipes.pdf
PPTX
TOTAL hIP ARTHROPLASTY Presentation.pptx
PDF
CAPERS-LRD-z9:AGas-enshroudedLittleRedDotHostingaBroad-lineActive GalacticNuc...
PDF
Placing the Near-Earth Object Impact Probability in Context
PDF
Looking into the jet cone of the neutrino-associated very high-energy blazar ...
PDF
Unveiling a 36 billion solar mass black hole at the centre of the Cosmic Hors...
PPTX
Seminar Hypertension and Kidney diseases.pptx
PPTX
Introduction to Cardiovascular system_structure and functions-1
PPTX
Application of enzymes in medicine (2).pptx
PDF
An interstellar mission to test astrophysical black holes
PPTX
Science Quipper for lesson in grade 8 Matatag Curriculum
PPTX
Microbes in human welfare class 12 .pptx
PPT
Heredity-grade-9 Heredity-grade-9. Heredity-grade-9.
Warm, water-depleted rocky exoplanets with surfaceionic liquids: A proposed c...
1. INTRODUCTION TO EPIDEMIOLOGY.pptx for community medicine
CHAPTER 3 Cell Structures and Their Functions Lecture Outline.pdf
The Land of Punt — A research by Dhani Irwanto
Cosmic Outliers: Low-spin Halos Explain the Abundance, Compactness, and Redsh...
Pharmacology of Autonomic nervous system
perinatal infections 2-171220190027.pptx
Phytochemical Investigation of Miliusa longipes.pdf
TOTAL hIP ARTHROPLASTY Presentation.pptx
CAPERS-LRD-z9:AGas-enshroudedLittleRedDotHostingaBroad-lineActive GalacticNuc...
Placing the Near-Earth Object Impact Probability in Context
Looking into the jet cone of the neutrino-associated very high-energy blazar ...
Unveiling a 36 billion solar mass black hole at the centre of the Cosmic Hors...
Seminar Hypertension and Kidney diseases.pptx
Introduction to Cardiovascular system_structure and functions-1
Application of enzymes in medicine (2).pptx
An interstellar mission to test astrophysical black holes
Science Quipper for lesson in grade 8 Matatag Curriculum
Microbes in human welfare class 12 .pptx
Heredity-grade-9 Heredity-grade-9. Heredity-grade-9.
Ad

basicem2.ppt

  • 1. Basic Electron Microscopy Arthur Rowe The Knowledge Base at a Simple Level
  • 2. Introduction These 3 presentations cover the fundamental theory of electron microscopy In presentation #2 we cover: – lens aberrations and their importance – how we correct for lens astigmatism – limits to ultimate resolution of the TEM – Interactions of electrons with matter
  • 3. aberrations of electromagnetic lenses the most important ones to consider are: • spherical aberration • chromatic aberration • astigmatism
  • 4. spherical aberration object plane • arises because a simple lens is more powerful at the edge than at the centre • is not a problem with glass lenses (can be ground to shape) • disc of minimum confusion results instead of point focus: • is not correctable for electromagnetic lenses
  • 5. coping with spherical aberration • disc of minimum confusion has diameter given by: d = C  {C = constant} • hence reducing  gives a large reduction in d • . . . but for optimal resolution we need large  ! • best compromise is with  = 10-3 radians (= f/500) • gives resolution = 0.1 nm - can not be bettered
  • 6. chromatic aberration • light of differentbrought to different focal positions • for electrons can be controlled by fixed KV and lens currents • but  of electrons can change by interaction with specimen ! • rule of thumb: resolution >= (specimen thickness)/10
  • 7. astigmatism minimal confusion • arises when the lens is more powerful in one plane than in the plane normal to it • causes points to be imaged as short lines, which ‘flip’ through 90 degrees on passing through ‘focus’ (minimal confusion)
  • 8. astigmatism - arises from: • inherent geometrical defects in ‘circular’ bore of lens • inherent inhomogeneities in magnetic properties of pole piece • build-up of contamination on bore of pole-piece and on apertures gives rise to non-conducting deposits which become charged as electron strike them • hence astigmatism is time-dependent • and cannot be ‘designed out’ • inevitably requires continuous correction
  • 9. astigmatism - correction: • with glass optics (as in spectacles) astigmatism is corrected using an additional lens of strength & asymmetry opposed to the asymmetry of the basic (eye) lens • with electron optics, same principle employed: electrostatic stigmator lens apposed to main lens strength & direction of its asymmetry user-variable • only the OBJECTIVE lens needs accurate correction • correction usually good for 1-2 hours for routine work
  • 10. The TEM Column _ Gun emits electrons _ Electric field accelerate _ Magnetic (and electric) field control path of electrons _ Electron wavelength @ 200KeV  2x10-12 m _ Resolution normally achievable @ 200KeV  2 x 10-10 m  2Å
  • 11. depth of focus - depth of field • depth of useful focus (in the specimen) is primarily limited by chromatic aberration effects • the absolute depth of focus is larger than this: for all practical purposes, everything is in focus to same level • . . . So one cannot rack through focus (as in a light or even scanning electron) microscope • depth of field (in the image plane) is - for all practical purposes infinite
  • 12. when electrons hit matter ..
  • 13. when electrons hit matter .. (1) they may collide with an inner shell electron, ejecting same > the ejected electron is a low-energy, secondary electron - detected & used to from SEM images > the original high-energy electron is scattered - known as a ‘back-scattered’ electron (SEM use) > an outer-shell electron drops into the position formerly occupied by the ejected electron > this is a quantum process, so a X-ray photon of precise wavelength is emitted - basis for X-ray microanalysis
  • 14. when electrons hit matter ..
  • 15. when electrons hit matter .. (2) they may collide or nearly collide with an atomic nucleus > undergo varying degree ofdeflection (inelastic scattering) > undergo loss of energy - again varying > lost energy appears as X-rays of varying wavelength > this X-ray continuum is identical to that originating from an X-ray source/generator (medical, XRC etc) > original electrons scattered in a forward direction will enter the imaging system, but with ‘wrong’  > causes a ‘haze’ and loss of resolution in image
  • 16. when electrons hit matter ..
  • 17. when electrons hit matter .. (3) they may collide with outer shell electrons > either removing or inserting an electron > results in free radical formation > this species is extremely chemically active > reactions with neighbouring atoms induce massive change in the specimen, especially in the light atoms > this radiation damage severely limits possibilities of EM > examination of cells in the live state NOT POSSIBLE > all examinations need to be as brief (low dose) as possible
  • 18. when electrons hit matter ..
  • 19. when electrons hit matter .. (4) they may pass through unchanged > these transmitted electrons can be used to form an image > this is called imaging by subtractive contrast > can be recorded by either (a) TV-type camera (CCD) - very expensive (b) photographic film - direct impact of electrons Photographic film > silver halide grains detect virtually every electron > at least 50x more efficient than photon capture !
  • 20. when electrons hit matter .. ‘beam damage’occurs: • light elements (H, O) lost very rapidly • change in valency shell means free radicals formed • . . .& consequent chemical reactions causing further damage • beam damage is minimised by use of • low temperatures (-160°) • high beam voltages • minimal exposure times