The CT process
Mohamed Ibrahim
Medical physicist
Magdi Yaqoub Heart Foundation
Content
Brief History of CT
Principles of The CT process
Dual Energy CT
Modes of Acquisitions
CT Image Quality (Factors & Indices).
1895: Roentgen discovers x-ray.
1963: Cormack formulates x-ray absorption in tissues.
1972: Hounsfield demonstrates CT.
1979: Hounsfield and Cormack received Nobel Prize.
1983: EBCT demonstrated.
1989: Spiral CT demonstrated.
1991: MSCT introduced
Important events
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Conventional Radiography
Conventional Radiography
 A conventional X-ray image is basically a
shadow
 Shadows give you an incomplete picture
of an object's shape
 Collapsing of 3D structures onto a 2D image
(superimposition of all structure on the film)
Brief History of CT
 Computed Tomography (CT) imaging is also known as CAT Scanning (Computed Axial
Tomography).
 Tomography is from the Greek word "tomo" meaning "slice" or "section" and "graphia"
meaning "describing".
 Designed by Godfrey N. Hounsfield by collimating the X-ray beam and transmitting it only
through small cross-sections of the body, and Allan MacLeod Cormack, who developed
solutions to mathematical problems involved in CT.
Brief discussion about The CT scan process
Fourth generation
Fifth generation
The slip Ring
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Z-flying focal spot technique
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
CT Basic Principles
CT process creates cross sectional images with information collected when
an x-ray beam passes through an area of anatomy.
Data that form the CT slice are further sectioned into elements (2-D
square=pixel (x-axis=width ;y-axis= height)).
If z-axis is taken into consideration, it becomes a voxel.
Each pixel represents the linear attenuation value of X-rays by the structures
contained within the corresponding voxel.
The linear attenuation value or density of each voxel is represented by an
numerical value called CT number or Hounsfield units.
X-ray photons pass through/ absorbed/ scattered by structures in varying
amounts, depending on the average photon energy of the x-ray beam and
the characteristics of the structure in its path.
Beam Attenuation
The factors contributing to μ
The video monitor can display 256 shades of gray and the eye can
detect only approximately 20
Relationship between window level and width
 Window level
 Defined as the central value of the window
used for the display of the reconstructed CT
image.
 selected according to the attenuation
characteristics of the structure under
examination.
 Window Width
 The range of CT numbers converted into grey
levels and displayed on the image monitor.
 Selected by the operator according to the
clinical requirements.
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
CT requires at least 180° plus fan angle of projection data to perform
image reconstruction.
Brief discussion about The CT scan process
Single-source 64-slice CT with half-scan reconstruction resulting in
temporal resolution of 165 ms and dual-source 64-slice CT with half-scan
reconstruction resulting in temporal resolution of 83 ms .
Dual Energy CT
It produces better imaging of blood vessels and the images of with and without contrast
media can be obtained in a single examination image.
 Dual Source CT Scanners – A dual source CT Scanner consists of two X-ray tubes and two
detectors at right angles to each other.
The dual X-ray source doubles the speed of the scan(used for imaging of cardiac patients).
 Dual Energy CT Scanners – It consists of single X-ray tube. It produce different photon
energy by using the technique of kV switching and different filters .The tubes generate both
powerful and less powerful X-rays.
In a dual energy CT Scanner , two pictures are taken simultaneously of the same slice at
different energies .
High energy scan are obtained at 140 kVp, and low energy scans are obtained at 80 kVp.
Brief discussion about The CT scan process
Modes of Acquisitions
Scout ,Surview , Topogram, Scanogram
Axial , Sequential Scanning
Spiral , Helical Scanning
Factors affecting image quality
X-ray production
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
kilovolt-peak (kVp)
How KV affect image quality
 Increase Max Energy
 Increase quantity of X Ray
 Increase the mean energy
 Increase patient dose
 No change in the minimum energy
 No change in the scan time
Increasing KV
Why low Kv increase image contrast
Low KV :
 Increase Photoelectric effect.
 At low kv, the mean energy is slightly higer
K-edge binding energy (33.2 Kev) of Iodine
Brief discussion about The CT scan process
Interactions with matter
Interactions with matter
Milliampere-Second (mAs)
Increasing Tube current :
 Increase intensity of X Ray
 Increasing patient dose
 Reduce the image noise
 No change in the scan time
 No change in the max Energy or the mean
energy and minimum Energy
Noise ∝ 1 / √mAs
Dose ∝ mAs
mAs eff (calcolated mAs per acquired slice)
mAs eff = mAs × exposure time
(exposure time = gantry rotation time / pitch)
mAs eff = mAs × gantry rotation time / pitch
High mAs has higher contrast resolution than Low mAs
Noise 1/ √mAs
∝
Lowering mA by 50 % reduces dose by 50 % but increases noise by 41%
mAs Modulation
mAs Modulation
mAs Modulation
Different mAs with different body dimensions >>
 Reduce the patient dose
 Maintain the noise level
Angular-Longitudinal Modulation (ap vs. lat and changing anatomy
over the z-axis).
 Smart mA - GE
 Z-DOM – Phillips
 CareDose 4D - Siemens
 SureExposure - Toshiba
Brief discussion about The CT scan process
Angular-Longitudinal Modulation (ap vs. lat and
changing anatomy over the z-axis)
The patient must be centered in the gantry isocenter for accurate imaging of
the anatomy (If the patient is not centered correctly the image quality is
degraded and the patient dose increases).
Centering of patient effect on the patient dose using mAs modulation
Centering of patient
Centering of patient
Centering of patient
Noise = 1/√slice thickness
Dose depends on beam thickness and not on the thickness of the
reconstructed image.
Slice thickness & Beam collimation
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
Brief discussion about The CT scan process
 The inverse pitch 1/P measures how many rotations contribute to each z-
position.
 For example with P = 1 we find that a given z-position is covered by exactly
one rotation and for P = 0.5 two rotations contribute to the z-position.
Pitch = 1: Implies contiguous slice similar to conventional step-and-shoot
scan, for example: 10-mm slice thickness with 10-mm slice interval.
 Pitch > 1: Implies extended imaging and reduced patient dose with lower
axial spatial resolution.
Pitch < 1: Implies overlapping and higher patient dose with higher axial
spatial resolution.
Increase Rot. time :
Increase scan time
Increase patient dose
Increase the probability of motion artifacts
Rotation time
Brief discussion about The CT scan process
Image Quality indices
The CT numbers vary randomly and the image is mottled or grainy.
Noise is reduced by :
 Using a higher mA, longer rotation time, higher kVp , a larger slice thickness and
large pixel.
Noise
Brief discussion about The CT scan process
The ability of the image to differentiate between
structures with very different attenuation
characteristics (CT numbers ) when they are very
close together.
 A high level of spatial resolution can be achieved by
having small pixels and thin slices.
 Smaller focal spot produce high spatial resolution
(increase the sharpness).
 decreasing the pitch increase resolution.
CT spatial resolution (high contrast resolution)
For two objects to be seen as separate the detectors must be able to
identify a gap between them.
Brief discussion about The CT scan process
Brief discussion about The CT scan process
10-mm 1.5-mm
A standard kernel A bone kernel
Utilizing a bone, sharp, high frequency or high pass algorithm during reconstruction
can improve the spatial resolution.
Field of view & The spatial resolution
Field of view
Images reconstructed with different FOVs.
A, Reconstructed with 50-cm FOV.
B, Reconstructed with 50-cm FOV and interpolated to 10-cm FOV.
C, Targeted reconstruction to 10-cm FOV.
Field of view & The spatial resolution
The ability to discriminate between structures
with very similar attenuation characteristics (CT
numbers).
The ability to differentiate a structure that varies
only slightly in density from its surrounding.
The limiting factor for contrast resolution is noise.
 Any factor that reduces the amount or visibility of noise
improves contrast resolution, choice of mA, kVp , slice
thickness, matrix size, FOV, and reconstruction filter
affects contrast resolution.
CT contrast resolution (low contrast resolution)
Brief discussion about The CT scan process
Temporal Resolution
 Temporal resolution is still the major limitation of coronary CT and the main
cause of non diagnostic images, and therefore all possible measures must be
taken to improve this parameter.
 CT requires at least 180° plus fan angle of projection data to perform image
reconstruction.
 This implies that the intrinsic temporal resolution of a standard CT scan is in
the order of gantry rotation time /2.
 The temporal resolution is most dependent on the speed of the gantry
rotation.
Brief discussion about The CT scan process
To be continued,,,,,
Cardiac CT Imaging

More Related Content

PPTX
Image Quality And Artifacts in Computed Tomography.pptx
PDF
Computed Tomography
PPTX
Ct image quality artifacts and it remedy
PPTX
Hrct i
PPTX
Comparison of ct and cbct
PPT
Basics in ct
PPT
Ct Basics
Image Quality And Artifacts in Computed Tomography.pptx
Computed Tomography
Ct image quality artifacts and it remedy
Hrct i
Comparison of ct and cbct
Basics in ct
Ct Basics

Similar to Brief discussion about The CT scan process (20)

PPTX
Basic principle of CT and generation of CT.pptx
PPTX
CT scanner
PPT
CT ITS BASIC PHYSICS
PPT
Physics of Multidetector CT Scan
PPTX
CT PHysics for the medical gradutaes in india
PPTX
medical equipment lec 6 CT scan
PPT
CT PHYSICS for first year radiology residents
PPTX
Ct brain presentation
PPT
Ct radiation dose &amp; safety lecture 1
PPTX
CBCT IN ORTHODONTICS
PPTX
Computed tomography
PDF
computedtomography-170122041110.pdf
PPTX
Computed tomography
PPT
principals of computed tomography / dental implant courses
PPT
ct physics final jhlgkuyfkytfukyguktfkytfkytfiyt
PPTX
Cone beam ct
PPTX
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
PPTX
Basic principle of ct and ct generations
PPTX
Basic principle of ct and ct generations
PDF
lecture continue.pdf
Basic principle of CT and generation of CT.pptx
CT scanner
CT ITS BASIC PHYSICS
Physics of Multidetector CT Scan
CT PHysics for the medical gradutaes in india
medical equipment lec 6 CT scan
CT PHYSICS for first year radiology residents
Ct brain presentation
Ct radiation dose &amp; safety lecture 1
CBCT IN ORTHODONTICS
Computed tomography
computedtomography-170122041110.pdf
Computed tomography
principals of computed tomography / dental implant courses
ct physics final jhlgkuyfkytfukyguktfkytfkytfiyt
Cone beam ct
LCU RDG 402 PRINCIPLES OF COMPUTED TOMOGRAPHY.pptx
Basic principle of ct and ct generations
Basic principle of ct and ct generations
lecture continue.pdf
Ad

Recently uploaded (20)

PDF
medical_surgical_nursing_10th_edition_ignatavicius_TEST_BANK_pdf.pdf
PPTX
Onco Emergencies - Spinal cord compression Superior vena cava syndrome Febr...
PDF
احياء السادس العلمي - الفصل الثالث (التكاثر) منهج متميزين/كلية بغداد/موهوبين
PDF
FOISHS ANNUAL IMPLEMENTATION PLAN 2025.pdf
PPTX
CHAPTER IV. MAN AND BIOSPHERE AND ITS TOTALITY.pptx
PDF
Indian roads congress 037 - 2012 Flexible pavement
PDF
David L Page_DCI Research Study Journey_how Methodology can inform one's prac...
PDF
OBE - B.A.(HON'S) IN INTERIOR ARCHITECTURE -Ar.MOHIUDDIN.pdf
PDF
Practical Manual AGRO-233 Principles and Practices of Natural Farming
PDF
AI-driven educational solutions for real-life interventions in the Philippine...
PDF
IGGE1 Understanding the Self1234567891011
PPTX
Virtual and Augmented Reality in Current Scenario
PDF
RTP_AR_KS1_Tutor's Guide_English [FOR REPRODUCTION].pdf
PPTX
A powerpoint presentation on the Revised K-10 Science Shaping Paper
DOC
Soft-furnishing-By-Architect-A.F.M.Mohiuddin-Akhand.doc
PDF
Weekly quiz Compilation Jan -July 25.pdf
PDF
A GUIDE TO GENETICS FOR UNDERGRADUATE MEDICAL STUDENTS
PDF
Τίμαιος είναι φιλοσοφικός διάλογος του Πλάτωνα
PDF
1.3 FINAL REVISED K-10 PE and Health CG 2023 Grades 4-10 (1).pdf
PPTX
B.Sc. DS Unit 2 Software Engineering.pptx
medical_surgical_nursing_10th_edition_ignatavicius_TEST_BANK_pdf.pdf
Onco Emergencies - Spinal cord compression Superior vena cava syndrome Febr...
احياء السادس العلمي - الفصل الثالث (التكاثر) منهج متميزين/كلية بغداد/موهوبين
FOISHS ANNUAL IMPLEMENTATION PLAN 2025.pdf
CHAPTER IV. MAN AND BIOSPHERE AND ITS TOTALITY.pptx
Indian roads congress 037 - 2012 Flexible pavement
David L Page_DCI Research Study Journey_how Methodology can inform one's prac...
OBE - B.A.(HON'S) IN INTERIOR ARCHITECTURE -Ar.MOHIUDDIN.pdf
Practical Manual AGRO-233 Principles and Practices of Natural Farming
AI-driven educational solutions for real-life interventions in the Philippine...
IGGE1 Understanding the Self1234567891011
Virtual and Augmented Reality in Current Scenario
RTP_AR_KS1_Tutor's Guide_English [FOR REPRODUCTION].pdf
A powerpoint presentation on the Revised K-10 Science Shaping Paper
Soft-furnishing-By-Architect-A.F.M.Mohiuddin-Akhand.doc
Weekly quiz Compilation Jan -July 25.pdf
A GUIDE TO GENETICS FOR UNDERGRADUATE MEDICAL STUDENTS
Τίμαιος είναι φιλοσοφικός διάλογος του Πλάτωνα
1.3 FINAL REVISED K-10 PE and Health CG 2023 Grades 4-10 (1).pdf
B.Sc. DS Unit 2 Software Engineering.pptx
Ad

Brief discussion about The CT scan process

  • 1. The CT process Mohamed Ibrahim Medical physicist Magdi Yaqoub Heart Foundation
  • 2. Content Brief History of CT Principles of The CT process Dual Energy CT Modes of Acquisitions CT Image Quality (Factors & Indices).
  • 3. 1895: Roentgen discovers x-ray. 1963: Cormack formulates x-ray absorption in tissues. 1972: Hounsfield demonstrates CT. 1979: Hounsfield and Cormack received Nobel Prize. 1983: EBCT demonstrated. 1989: Spiral CT demonstrated. 1991: MSCT introduced Important events
  • 7. Conventional Radiography  A conventional X-ray image is basically a shadow  Shadows give you an incomplete picture of an object's shape  Collapsing of 3D structures onto a 2D image (superimposition of all structure on the film)
  • 8. Brief History of CT  Computed Tomography (CT) imaging is also known as CAT Scanning (Computed Axial Tomography).  Tomography is from the Greek word "tomo" meaning "slice" or "section" and "graphia" meaning "describing".  Designed by Godfrey N. Hounsfield by collimating the X-ray beam and transmitting it only through small cross-sections of the body, and Allan MacLeod Cormack, who developed solutions to mathematical problems involved in CT.
  • 16. Z-flying focal spot technique
  • 20. CT Basic Principles CT process creates cross sectional images with information collected when an x-ray beam passes through an area of anatomy. Data that form the CT slice are further sectioned into elements (2-D square=pixel (x-axis=width ;y-axis= height)). If z-axis is taken into consideration, it becomes a voxel. Each pixel represents the linear attenuation value of X-rays by the structures contained within the corresponding voxel. The linear attenuation value or density of each voxel is represented by an numerical value called CT number or Hounsfield units.
  • 21. X-ray photons pass through/ absorbed/ scattered by structures in varying amounts, depending on the average photon energy of the x-ray beam and the characteristics of the structure in its path. Beam Attenuation
  • 23. The video monitor can display 256 shades of gray and the eye can detect only approximately 20
  • 24. Relationship between window level and width  Window level  Defined as the central value of the window used for the display of the reconstructed CT image.  selected according to the attenuation characteristics of the structure under examination.  Window Width  The range of CT numbers converted into grey levels and displayed on the image monitor.  Selected by the operator according to the clinical requirements.
  • 32. CT requires at least 180° plus fan angle of projection data to perform image reconstruction.
  • 34. Single-source 64-slice CT with half-scan reconstruction resulting in temporal resolution of 165 ms and dual-source 64-slice CT with half-scan reconstruction resulting in temporal resolution of 83 ms .
  • 35. Dual Energy CT It produces better imaging of blood vessels and the images of with and without contrast media can be obtained in a single examination image.  Dual Source CT Scanners – A dual source CT Scanner consists of two X-ray tubes and two detectors at right angles to each other. The dual X-ray source doubles the speed of the scan(used for imaging of cardiac patients).  Dual Energy CT Scanners – It consists of single X-ray tube. It produce different photon energy by using the technique of kV switching and different filters .The tubes generate both powerful and less powerful X-rays. In a dual energy CT Scanner , two pictures are taken simultaneously of the same slice at different energies . High energy scan are obtained at 140 kVp, and low energy scans are obtained at 80 kVp.
  • 38. Scout ,Surview , Topogram, Scanogram
  • 39. Axial , Sequential Scanning
  • 40. Spiral , Helical Scanning
  • 52. How KV affect image quality
  • 53.  Increase Max Energy  Increase quantity of X Ray  Increase the mean energy  Increase patient dose  No change in the minimum energy  No change in the scan time Increasing KV
  • 54. Why low Kv increase image contrast Low KV :  Increase Photoelectric effect.  At low kv, the mean energy is slightly higer K-edge binding energy (33.2 Kev) of Iodine
  • 58. Milliampere-Second (mAs) Increasing Tube current :  Increase intensity of X Ray  Increasing patient dose  Reduce the image noise  No change in the scan time  No change in the max Energy or the mean energy and minimum Energy Noise ∝ 1 / √mAs Dose ∝ mAs mAs eff (calcolated mAs per acquired slice) mAs eff = mAs × exposure time (exposure time = gantry rotation time / pitch) mAs eff = mAs × gantry rotation time / pitch High mAs has higher contrast resolution than Low mAs
  • 59. Noise 1/ √mAs ∝ Lowering mA by 50 % reduces dose by 50 % but increases noise by 41%
  • 62. mAs Modulation Different mAs with different body dimensions >>  Reduce the patient dose  Maintain the noise level Angular-Longitudinal Modulation (ap vs. lat and changing anatomy over the z-axis).  Smart mA - GE  Z-DOM – Phillips  CareDose 4D - Siemens  SureExposure - Toshiba
  • 64. Angular-Longitudinal Modulation (ap vs. lat and changing anatomy over the z-axis)
  • 65. The patient must be centered in the gantry isocenter for accurate imaging of the anatomy (If the patient is not centered correctly the image quality is degraded and the patient dose increases). Centering of patient effect on the patient dose using mAs modulation Centering of patient
  • 68. Noise = 1/√slice thickness Dose depends on beam thickness and not on the thickness of the reconstructed image. Slice thickness & Beam collimation
  • 75.  The inverse pitch 1/P measures how many rotations contribute to each z- position.  For example with P = 1 we find that a given z-position is covered by exactly one rotation and for P = 0.5 two rotations contribute to the z-position.
  • 76. Pitch = 1: Implies contiguous slice similar to conventional step-and-shoot scan, for example: 10-mm slice thickness with 10-mm slice interval.  Pitch > 1: Implies extended imaging and reduced patient dose with lower axial spatial resolution. Pitch < 1: Implies overlapping and higher patient dose with higher axial spatial resolution.
  • 77. Increase Rot. time : Increase scan time Increase patient dose Increase the probability of motion artifacts Rotation time
  • 80. The CT numbers vary randomly and the image is mottled or grainy. Noise is reduced by :  Using a higher mA, longer rotation time, higher kVp , a larger slice thickness and large pixel. Noise
  • 82. The ability of the image to differentiate between structures with very different attenuation characteristics (CT numbers ) when they are very close together.  A high level of spatial resolution can be achieved by having small pixels and thin slices.  Smaller focal spot produce high spatial resolution (increase the sharpness).  decreasing the pitch increase resolution. CT spatial resolution (high contrast resolution) For two objects to be seen as separate the detectors must be able to identify a gap between them.
  • 86. A standard kernel A bone kernel
  • 87. Utilizing a bone, sharp, high frequency or high pass algorithm during reconstruction can improve the spatial resolution.
  • 88. Field of view & The spatial resolution Field of view
  • 89. Images reconstructed with different FOVs. A, Reconstructed with 50-cm FOV. B, Reconstructed with 50-cm FOV and interpolated to 10-cm FOV. C, Targeted reconstruction to 10-cm FOV. Field of view & The spatial resolution
  • 90. The ability to discriminate between structures with very similar attenuation characteristics (CT numbers). The ability to differentiate a structure that varies only slightly in density from its surrounding. The limiting factor for contrast resolution is noise.  Any factor that reduces the amount or visibility of noise improves contrast resolution, choice of mA, kVp , slice thickness, matrix size, FOV, and reconstruction filter affects contrast resolution. CT contrast resolution (low contrast resolution)
  • 92. Temporal Resolution  Temporal resolution is still the major limitation of coronary CT and the main cause of non diagnostic images, and therefore all possible measures must be taken to improve this parameter.  CT requires at least 180° plus fan angle of projection data to perform image reconstruction.  This implies that the intrinsic temporal resolution of a standard CT scan is in the order of gantry rotation time /2.  The temporal resolution is most dependent on the speed of the gantry rotation.