CSSD: DECONTAMINATION,
PACKAGING, STERILIZATION AND
MONITORING
CSSD.pdf
 The Central Sterile Supply Department (CSSD) is the service responsible
for receiving, sorting, processing, distributing and controlling the
professional supplies of sterile items for all users unit of hospital for the
care and safety of patient under strict quality control
 10sft/bed
 There is no international guidelines regarding space requirements for CSSD.
 For 600 bedded hospitals 6 or 7 sft per bed around 4000sft space
 100 bedded hospital, we may need 1000 sft space
 Number of operation theaters and number of surgical procedures to be done or being
done also taken into consideration. A CSSD of 300 bedded hospital having 7 operation
theaters may need lesser space than a CSSD of 250 bedded hospital having 10
operation theaters.
ISHA- Indian Sterilization Healthcare Association 2017
Used Articles Rinsing Assembling
Temporary storage in
user dept
Collection by
CSSD Staff
Transport to
CSSD
Transport to user
dept
Documentation Loading Place Indicators Packing & Labelling Trays Tray setting with
checklist
Decontamination
Manual / Machine
CSSD WORK FLOW
Drying
Unload & Storage in
sterial area
Sterilization
Handover items into
CSSD
Decontaminatio
n
Soiled Return
Prep and Pack
Sterilization
Sterile Storage
LOCATION PRESSURE TOTAL ACH ALL ROOMS AIR
EXHAUSTED
OUTSIDE
RELATIVE
HUMIDITY
TEMPERATURE
Decontamination Negative 6 Yes No requirement 22 -26°C
Pre pack/ packing Positive 4 No No requirement 22-26°C
Sterile storage Positive 4 No Max 60%
WHO 40-50%
22-26°C
Sterilizer equipment Negative 10 Yes No requirement No requirement
8
3M Health Care Academy
SM
Clotted Blood
Open Trolleys for
Instruments
transportation
Instruments transportation
through common usage
equipments
:
1) Instruments to be transported in dedicated
Closed trolleys
2) Count the number of instruments per set and
record in the Issue register in OT/wards also in
Issue register
3) Sort sharp items/ micro-instruments
4) Inspect for any damage/ dried blood stains
5) If damaged-report to the Manager & remove
from usage
6) Primary wash is to be done by the user
department to prevent organic matter from drying
on it
7) Possibly in a dedicated lift or by other means
CSSD.pdf
CSSD.pdf
 Cleaning: The physical removal of body materials, dust or foreign
material. Cleaning will reduce the number of microorganisms as well as
the soil, and reducing the risk of soil being fixed to the surface. Removal of
soil will reduce also the risk of inactivation of a chemical disinfectant and the
multiplication of microorganisms. The removal of contamination from
an item to the extent necessary for further processing or for
intended use. [ISO/TS 11139]
 The destruction or removal of microorganisms at a
level that is not harmful to health and safe to handle.
 This process does not necessarily include the destruction
of bacterial spores.
STERILIZATION
 The complete destruction or removal of microorganisms,
including bacterial spores.
A. Cleaning & Decontamination Process
Brushes for Cleaning
Instrument Sorting & Disassembly Manual Drying & Drying Cabinet
Compatible Cleaning Agent IFU
Instrument Checklist
Multiple Sinks for Cleaning
B. Methods of cleaning
Ultra-sonic Cleaning
• Use after gross soil is removed
• Usually used for Micro-instruments to
avoid their damage
• Conduct Aluminum foil test to check
the efficacy of the ultra-sonicator
Manual Cleaning
• Use brushes recommended for
washing medical devices
• Disinfect the brushes in the
Multi-enzyme cleaner after
every use or at least end of
the day
Washer Disinfector
• Do not overload
• Daily and preventive
• Correct chemical & lubricants
• Review and initial mechanical washer
cycle printouts
• Document results
Final rinse
by RO Water
FACTORS WHICH AFFECT CLEANING
 Effective cleaning and pre-cleaning of devices often requires chemicals, combined with
mechanical action and heat.
 It can be performed manually and/or with machines.
• Type and amount of soil
• Quality of water
• Type of detergent
• Concentration of detergent
• Type of instrument or device
• Amount of time permitted for the detergent to act
P
ASS
T
HROUGH
W
INDOW
A pass through window is
used to move clean items
into prep and pack area and
items that need to be
reprocessed back to
decontaminated area.
C. Instrumentation Inspection, Preparation & Packaging
Internal Pack Control CI
Cleaning Verification using
ATP
Instrument Checklist
Double Wrapping
Lubrication
Magnifying Lens Verification
Test Method0 Soil Component Tested Limitations
Length
of test
ATP
Swab device, extract ATP
from swab, and determine
ATP
ATP (present in
eukaryotic cells and live
bacteria)
- Need instrumentation
to read test
30 seconds
Protein
Swab device, immerse in
reagent and assess for color
development
Protein Not applicable to lumens minutes
 Equipment control -Bowie-Dick type(Autoclave)-class 2
 Exposure control- external indicator class 1
 Pack control- internal indicator class 4 or 5 or 6
 Load control- internal indicator class 4 or 5 or 6 or BI
 Equipment Monitoring is a way to find out whether or not your vacuum assisted
steam sterilizer is doing its job properly.
 To monitor dynamic air- removal steam sterilizers, you begin each day with a
Bowie-Dick type test to detect air leaks, inadequate air removal, inadequate
steam penetration and the presence of non-condensable gases, any of which can
compromise sterility.
CSSD.pdf
CSSD.pdf
Traceability can only be achieved if
the trays are recorded(Batch label)
Sterilization
Pre-vacuum Steam Sterilizer – High Speed ( Temperature 121 – 134 deg. C /
Pressure 1.1 to 2.1 Bar)
16 Towel BI PCD Pack
16 Towel Challenge Pack using
for Loaner and Implant load
Biological Indicator(BI) Challenge Pack
Steam sterilization loading
techniques
Following IFU
S.no Sterilization method Biological indicator Incubation
1. Steam Geobacillus stearothermophilus 60°C for 3hrs
2. EO B. Atrophaeus 37o C -4hrs
3. Plasma sterilisation Geobacillus stearothermophilus 58o C-24hrs
CSSD.pdf
 An internal indicator preferably class 5 (every pack)
 External class 1 indicator (every pack)
 Bowie dick-class 2 indicator-pre vacuum autoclave (every day empty load)
 Every load sterilized by EO
 Every load containing implantable devices in autoclave
 Every day for autoclave
 Every day for plasma sterilization
 Autoclave: run 3 empty cycles with BI PCD and subsequently 3
empty cycles with Bowie-dick
 Other sterilizers: run 3 empty cycles with BI PCD
 Advantages
Cost and environment
If chemically treated are resistant to
moisture and dust
 Disadvantages
More labor (laundry, inspect, delint
between each use and may need to repair)
Not lint free
If chemically treated may require
extended dry times
IAHCSMM Central Service Technical Manual, 7th edition, Chapter 13
Sterile Processing University, LLC The Basics of Sterile Processing, 3rd edition, Chapter 7
Central Services Association of Ontario, The Manual for Reprocessing Medical Devices, 1st Edition, 2009
)
Best wrap barrier
Water-repellent
No linting
 Wrapped basin sets should not exceed 3 Kg
 Containerized instrument sets do not
exceed 11kg
Tears easier than woven textiles
Has memory
 Advantages  Disadvantages
IAHCSMM Central Service Technical Manual, 7th edition, Chapter 13
Sterile Processing University, LLC The Basics of Sterile Processing, 3rd edition, Chapter 7
Central Services Association of Ontario, The Manual for Reprocessing Medical Devices, 1st Edition, 2009
Good to use for small, lightweight
instruments or single items
Contents easily visible
Paper/Plastic Pouches: labeling is done
on plastic side only
Must select the correct size for contents so
air can be expelled but not so large that the
contents can move around and damage the
seals
Seal rupture possible
Requires a heat seal
 Advantages  Disadvantages
IAHCSMM Central Service Technical Manual, 7th edition, Chapter 13
Sterile Processing University, LLC The Basics of Sterile Processing, 3rd edition, Chapter 7
Central Services Association of Ontario, The Manual for Reprocessing Medical Devices, 1st Edition, 2009
Excellent barrier to microorganisms
Water repellent
Easy-to-use
Eliminates torn wrappers
No linting
Protects instruments from damage
during processing, storage, and
transport
Ergonomics - adds weight to instruments
May need additional dry time
Additional space needed to store
containers but may be able to stack for
sterile storage
Additional labor for cleaning in between
use and inspection of gaskets, latches,
valves, etc
 Advantages  Disadvantages
IAHCSMM Central Service Technical Manual, 7th edition, Chapter 13
Sterile Processing University, LLC The Basics of Sterile Processing, 3rd edition, Chapter 7
Central Services Association of Ontario, The Manual for Reprocessing Medical Devices, 1st Edition, 2009
CSSD.pdf
LOADING OF ITEMS
CSSD.pdf
CSSD.pdf
©
3M
2014
All
Rights
Reserved.
39
INCORRECT LOADING
Maintain items on sterilizer cart until adequately cooled
Do not touch during the cooling process
Could wick bacteria from hands into packaging
Place sterilizer cart in a low traffic area, no air-conditioning or cold-air-vents
Do not transfer warm items to a cool cart which could result in condensate forming and
contamination
May open door slightly at end of cycle to reduce the potential for condensation
formation
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012, Section 8.8
Is event related and depends on
• Quality of packaging material
• Storage conditions
• Conditions during transportation
• Amount of handling
Label each product
• Expiration date or statement such as: “contents sterile unless package is open or
damaged. Please check before using.”
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012, Section 8.9.3 and 10.3.3
CSSD.pdf
CSSD.pdf
CSSD.pdf
WHO guidelines 2016 CSSD
Type of sterilizer Advantages Disadvantages
Autoclave Inexpensive
Fast
Effective, with a wide margin of safety
Penetrates medical packing, device lumens
Non-toxic – low environmental impact
Readily available
Sterilizers are available in many sizes for
many applications
Cycle easy to control and monitor
Unsuitable for anhydrous materials and for heat-
and moisture-sensitive materials
May leave instruments wet, causing them to rust
Expertise
Deleterious for heat-sensitive instruments
Microsurgical instruments damaged by repeated
exposure
Potential for burns
Hydrogen
Peroxide (Plasma)
Fast
Safe for the environment
Compatible with heat- and
moisture-sensitive
Absence of toxic waste
Easy installation and operation
No staff monitoring currently required
No aeration required
No chemical residues
Cannot sterilize materials that absorb hydrogen
peroxide (e.g. linen, gauze, cellulose/paper, wood)
Low penetration power
Medical devices must be dry before processing
Cellulose (paper), linens and liquids cannot be
processed
Limitations to length of lumens of medical
devices that can be effectively sterilized
Type of sterilizer Advantages Disadvantages
ETO Non-corrosive
Penetrates packaging materials, device
lumens
Has some ability to penetrate some synthetic
materials
Excellent material compatibility
Simple to operate and monitor
Toxic/carcinogenic to humans
Lengthy cycle due to aeration requirements
Requires monitoring of the work areas
Requires control and monitoring of discharge into
the environment
Flammable and explosive.
Expensive
 Implantable devices till the BI testing is satisfactory
 Non implantable devices BI/ class 5or 6/specific cycle physical
parameters in the load control
RAPID CYCLE OR FLASH STERILIZATION
 “Unwrapped” steam sterilization
 Should only be used when necessary
 Do not flash whole trays of instruments
 Items must be used immediately
 Avoid by keeping adequate supply of frequently dropped items
 Maintain records or “flash logs”
 Use to support need for additional instruments
Nurse from OT calls the CSSD to inform the Arthroplasty set is wet.
What to perform next?
1. Receive the wet set process sterilize and return
2. Receive all the sets sterilized in the same load
3. CSSD technician observes the external indicator is working and
tells to use the instrument
4. Its an issue between CSSD technician and the OT nurses, no
need to inform others
 Load contains positive BI- failed load and also all loads sterilized in that
sterilizer till last successfully sterilized load
 Incorrect reprocessing of device
 Print out of reprocessing equipment failure to reach correct parameters
 CI has not changed colour
 Doubt about sterility of equipment(torn, wet)
 Notification to all concerned departments
CSSD.pdf
Controlled temperature 24°C
≥ four air exchanges/hour
Relative humidity not to exceed
60%
Controlled traffic
Positive air flow (out)
Sterile items are stored at least 20-25cm (8-10”) above the floor, at least
45cm (18”) below the ceiling or sprinkler heads, and at least 5cm (2”) from
outside walls.
It should be located near a dock for the receiving of materials.
STERILE STORAGE
Supplies are distributed on a First In First Out
(FIFO) basis
CSSD.pdf
AUDIT- APSIC CHECK LIST
CSSD.pdf

More Related Content

PDF
CSSD Basics Chemical Indicators, Biological Indicators and Recall Process
PPTX
Central sterilization supply department
PPTX
central sterile supply department
PPTX
Central sterile supply department
PPTX
Bme and cssd
PPTX
CSSD & OT - June.pptx
DOCX
CSSD SOP
PPTX
Hospital CSSD
CSSD Basics Chemical Indicators, Biological Indicators and Recall Process
Central sterilization supply department
central sterile supply department
Central sterile supply department
Bme and cssd
CSSD & OT - June.pptx
CSSD SOP
Hospital CSSD

What's hot (20)

PPTX
Cssd (central sterile supply department)
PPT
Blood spills
PPTX
PPTX
History and developoment of cssd 2
PPTX
Central sterile supply department(cssd)
PPTX
SURGICAL ATTIRE.pptx
PPTX
Ot infection control rkch
PPTX
Central Sterile & Supply Department (CSSD)
PPTX
Central Sterile Supply Department (CSSD)
PDF
DHA_Anita_MHA2_Paper2_Unit 3.1.pdf
PPTX
Operation theatre design
PPTX
PPT
Aseptic technique
PPTX
PDF
Endoscope Cleaning
PPTX
Cleaning and Decontamination in Hospitals.pptx
PPTX
CSSD by Dr Zulfiquer Ahmed Amin
PDF
Appropriate-Operating-Room-Attire (1).pdf
PPT
Chapter 13 sterile packaging
PPTX
Introduction to CSSD
Cssd (central sterile supply department)
Blood spills
History and developoment of cssd 2
Central sterile supply department(cssd)
SURGICAL ATTIRE.pptx
Ot infection control rkch
Central Sterile & Supply Department (CSSD)
Central Sterile Supply Department (CSSD)
DHA_Anita_MHA2_Paper2_Unit 3.1.pdf
Operation theatre design
Aseptic technique
Endoscope Cleaning
Cleaning and Decontamination in Hospitals.pptx
CSSD by Dr Zulfiquer Ahmed Amin
Appropriate-Operating-Room-Attire (1).pdf
Chapter 13 sterile packaging
Introduction to CSSD
Ad

Similar to CSSD.pdf (20)

PPTX
Basics of Sterilization
PPTX
CSSD PPT.pptx
PDF
DHA_Anita_MHA2_Paper2_Unit 3.1.pdf
PPTX
Central sterile supply department in hospitals
PPTX
Sterilization & Disinfection
PPTX
sterilization & disinfectiono- lecture 5.pptx
PDF
CleaningDisinfectionandSterilization.pdf
PPTX
CSSD CENTRAL STERILE SUPPLY DEPARTMENT.pptx
PPTX
CENTRAL STERILE SUPPLY DEPARTMENT & FLASH STERILIZATION.pptx
PPT
Sterilization
PPTX
STERILIZATION OF ANAESTHESIA EQUIPMENT-1.pptx
PPTX
CSSD central sterile supply department and services
PPTX
CENTRAL STERILE SUPPLY DEPARTMENT (CSSD) By Sachin dwivedi.pptx
PPTX
INSTRUMENT PROCESSING PRESENTATION BY MASTER REBIRA.pptx
PPT
PPT
COMPLETE NOES ON STERILIZATION WITH METHODS.ppt
PPT
STERILIZATION.ppt meathods of sterlization
PPT
STERILIZATION it's PPT are amazing to learn about sterilization
PPTX
STERILIZATION.pptx
Basics of Sterilization
CSSD PPT.pptx
DHA_Anita_MHA2_Paper2_Unit 3.1.pdf
Central sterile supply department in hospitals
Sterilization & Disinfection
sterilization & disinfectiono- lecture 5.pptx
CleaningDisinfectionandSterilization.pdf
CSSD CENTRAL STERILE SUPPLY DEPARTMENT.pptx
CENTRAL STERILE SUPPLY DEPARTMENT & FLASH STERILIZATION.pptx
Sterilization
STERILIZATION OF ANAESTHESIA EQUIPMENT-1.pptx
CSSD central sterile supply department and services
CENTRAL STERILE SUPPLY DEPARTMENT (CSSD) By Sachin dwivedi.pptx
INSTRUMENT PROCESSING PRESENTATION BY MASTER REBIRA.pptx
COMPLETE NOES ON STERILIZATION WITH METHODS.ppt
STERILIZATION.ppt meathods of sterlization
STERILIZATION it's PPT are amazing to learn about sterilization
STERILIZATION.pptx
Ad

More from SHAIKHZUBAIR22 (12)

PDF
ecosystem-150606063938-lva1-app6892.pdf
PPTX
Demography PPt.pptx
PPT
SCHOOL HEALTH SERVICES.ppt
DOCX
communication pannel.docx
PPT
five year.ppt
PPTX
seminar on international health.pptx
PPT
national health prog..ppt
PPTX
hiap-ppt-module-1-part-2.pptx
PPTX
community diagnosis ppt.pptx
PPT
shock1.ppt
PPTX
tp zuber.pptx
PPTX
MILLENIUM DEVELOPMENT GOALS(MDGs).pptx
ecosystem-150606063938-lva1-app6892.pdf
Demography PPt.pptx
SCHOOL HEALTH SERVICES.ppt
communication pannel.docx
five year.ppt
seminar on international health.pptx
national health prog..ppt
hiap-ppt-module-1-part-2.pptx
community diagnosis ppt.pptx
shock1.ppt
tp zuber.pptx
MILLENIUM DEVELOPMENT GOALS(MDGs).pptx

Recently uploaded (20)

PPTX
Lesson notes of climatology university.
PDF
Complications of Minimal Access Surgery at WLH
PPTX
UV-Visible spectroscopy..pptx UV-Visible Spectroscopy – Electronic Transition...
PDF
ChatGPT for Dummies - Pam Baker Ccesa007.pdf
PDF
Computing-Curriculum for Schools in Ghana
PDF
Practical Manual AGRO-233 Principles and Practices of Natural Farming
PDF
RMMM.pdf make it easy to upload and study
PDF
Classroom Observation Tools for Teachers
PDF
Chapter 2 Heredity, Prenatal Development, and Birth.pdf
PDF
A systematic review of self-coping strategies used by university students to ...
PDF
The Lost Whites of Pakistan by Jahanzaib Mughal.pdf
PDF
01-Introduction-to-Information-Management.pdf
PPTX
History, Philosophy and sociology of education (1).pptx
PPTX
Cell Types and Its function , kingdom of life
PDF
grade 11-chemistry_fetena_net_5883.pdf teacher guide for all student
PDF
RTP_AR_KS1_Tutor's Guide_English [FOR REPRODUCTION].pdf
PPTX
Final Presentation General Medicine 03-08-2024.pptx
DOC
Soft-furnishing-By-Architect-A.F.M.Mohiuddin-Akhand.doc
PPTX
Cell Structure & Organelles in detailed.
PDF
Trump Administration's workforce development strategy
Lesson notes of climatology university.
Complications of Minimal Access Surgery at WLH
UV-Visible spectroscopy..pptx UV-Visible Spectroscopy – Electronic Transition...
ChatGPT for Dummies - Pam Baker Ccesa007.pdf
Computing-Curriculum for Schools in Ghana
Practical Manual AGRO-233 Principles and Practices of Natural Farming
RMMM.pdf make it easy to upload and study
Classroom Observation Tools for Teachers
Chapter 2 Heredity, Prenatal Development, and Birth.pdf
A systematic review of self-coping strategies used by university students to ...
The Lost Whites of Pakistan by Jahanzaib Mughal.pdf
01-Introduction-to-Information-Management.pdf
History, Philosophy and sociology of education (1).pptx
Cell Types and Its function , kingdom of life
grade 11-chemistry_fetena_net_5883.pdf teacher guide for all student
RTP_AR_KS1_Tutor's Guide_English [FOR REPRODUCTION].pdf
Final Presentation General Medicine 03-08-2024.pptx
Soft-furnishing-By-Architect-A.F.M.Mohiuddin-Akhand.doc
Cell Structure & Organelles in detailed.
Trump Administration's workforce development strategy

CSSD.pdf

  • 3.  The Central Sterile Supply Department (CSSD) is the service responsible for receiving, sorting, processing, distributing and controlling the professional supplies of sterile items for all users unit of hospital for the care and safety of patient under strict quality control
  • 4.  10sft/bed  There is no international guidelines regarding space requirements for CSSD.  For 600 bedded hospitals 6 or 7 sft per bed around 4000sft space  100 bedded hospital, we may need 1000 sft space  Number of operation theaters and number of surgical procedures to be done or being done also taken into consideration. A CSSD of 300 bedded hospital having 7 operation theaters may need lesser space than a CSSD of 250 bedded hospital having 10 operation theaters. ISHA- Indian Sterilization Healthcare Association 2017
  • 5. Used Articles Rinsing Assembling Temporary storage in user dept Collection by CSSD Staff Transport to CSSD Transport to user dept Documentation Loading Place Indicators Packing & Labelling Trays Tray setting with checklist Decontamination Manual / Machine CSSD WORK FLOW Drying Unload & Storage in sterial area Sterilization Handover items into CSSD
  • 6. Decontaminatio n Soiled Return Prep and Pack Sterilization Sterile Storage
  • 7. LOCATION PRESSURE TOTAL ACH ALL ROOMS AIR EXHAUSTED OUTSIDE RELATIVE HUMIDITY TEMPERATURE Decontamination Negative 6 Yes No requirement 22 -26°C Pre pack/ packing Positive 4 No No requirement 22-26°C Sterile storage Positive 4 No Max 60% WHO 40-50% 22-26°C Sterilizer equipment Negative 10 Yes No requirement No requirement
  • 8. 8 3M Health Care Academy SM Clotted Blood Open Trolleys for Instruments transportation Instruments transportation through common usage equipments
  • 9. : 1) Instruments to be transported in dedicated Closed trolleys 2) Count the number of instruments per set and record in the Issue register in OT/wards also in Issue register 3) Sort sharp items/ micro-instruments 4) Inspect for any damage/ dried blood stains 5) If damaged-report to the Manager & remove from usage 6) Primary wash is to be done by the user department to prevent organic matter from drying on it 7) Possibly in a dedicated lift or by other means
  • 12.  Cleaning: The physical removal of body materials, dust or foreign material. Cleaning will reduce the number of microorganisms as well as the soil, and reducing the risk of soil being fixed to the surface. Removal of soil will reduce also the risk of inactivation of a chemical disinfectant and the multiplication of microorganisms. The removal of contamination from an item to the extent necessary for further processing or for intended use. [ISO/TS 11139]
  • 13.  The destruction or removal of microorganisms at a level that is not harmful to health and safe to handle.  This process does not necessarily include the destruction of bacterial spores.
  • 14. STERILIZATION  The complete destruction or removal of microorganisms, including bacterial spores.
  • 15. A. Cleaning & Decontamination Process Brushes for Cleaning Instrument Sorting & Disassembly Manual Drying & Drying Cabinet Compatible Cleaning Agent IFU Instrument Checklist Multiple Sinks for Cleaning
  • 16. B. Methods of cleaning Ultra-sonic Cleaning • Use after gross soil is removed • Usually used for Micro-instruments to avoid their damage • Conduct Aluminum foil test to check the efficacy of the ultra-sonicator Manual Cleaning • Use brushes recommended for washing medical devices • Disinfect the brushes in the Multi-enzyme cleaner after every use or at least end of the day Washer Disinfector • Do not overload • Daily and preventive • Correct chemical & lubricants • Review and initial mechanical washer cycle printouts • Document results Final rinse by RO Water
  • 17. FACTORS WHICH AFFECT CLEANING  Effective cleaning and pre-cleaning of devices often requires chemicals, combined with mechanical action and heat.  It can be performed manually and/or with machines. • Type and amount of soil • Quality of water • Type of detergent • Concentration of detergent • Type of instrument or device • Amount of time permitted for the detergent to act
  • 18. P ASS T HROUGH W INDOW A pass through window is used to move clean items into prep and pack area and items that need to be reprocessed back to decontaminated area.
  • 19. C. Instrumentation Inspection, Preparation & Packaging Internal Pack Control CI Cleaning Verification using ATP Instrument Checklist Double Wrapping Lubrication Magnifying Lens Verification
  • 20. Test Method0 Soil Component Tested Limitations Length of test ATP Swab device, extract ATP from swab, and determine ATP ATP (present in eukaryotic cells and live bacteria) - Need instrumentation to read test 30 seconds Protein Swab device, immerse in reagent and assess for color development Protein Not applicable to lumens minutes
  • 21.  Equipment control -Bowie-Dick type(Autoclave)-class 2  Exposure control- external indicator class 1  Pack control- internal indicator class 4 or 5 or 6  Load control- internal indicator class 4 or 5 or 6 or BI
  • 22.  Equipment Monitoring is a way to find out whether or not your vacuum assisted steam sterilizer is doing its job properly.  To monitor dynamic air- removal steam sterilizers, you begin each day with a Bowie-Dick type test to detect air leaks, inadequate air removal, inadequate steam penetration and the presence of non-condensable gases, any of which can compromise sterility.
  • 25. Traceability can only be achieved if the trays are recorded(Batch label)
  • 26. Sterilization Pre-vacuum Steam Sterilizer – High Speed ( Temperature 121 – 134 deg. C / Pressure 1.1 to 2.1 Bar) 16 Towel BI PCD Pack 16 Towel Challenge Pack using for Loaner and Implant load Biological Indicator(BI) Challenge Pack Steam sterilization loading techniques Following IFU
  • 27. S.no Sterilization method Biological indicator Incubation 1. Steam Geobacillus stearothermophilus 60°C for 3hrs 2. EO B. Atrophaeus 37o C -4hrs 3. Plasma sterilisation Geobacillus stearothermophilus 58o C-24hrs
  • 29.  An internal indicator preferably class 5 (every pack)  External class 1 indicator (every pack)  Bowie dick-class 2 indicator-pre vacuum autoclave (every day empty load)  Every load sterilized by EO  Every load containing implantable devices in autoclave  Every day for autoclave  Every day for plasma sterilization
  • 30.  Autoclave: run 3 empty cycles with BI PCD and subsequently 3 empty cycles with Bowie-dick  Other sterilizers: run 3 empty cycles with BI PCD
  • 31.  Advantages Cost and environment If chemically treated are resistant to moisture and dust  Disadvantages More labor (laundry, inspect, delint between each use and may need to repair) Not lint free If chemically treated may require extended dry times IAHCSMM Central Service Technical Manual, 7th edition, Chapter 13 Sterile Processing University, LLC The Basics of Sterile Processing, 3rd edition, Chapter 7 Central Services Association of Ontario, The Manual for Reprocessing Medical Devices, 1st Edition, 2009
  • 32. ) Best wrap barrier Water-repellent No linting  Wrapped basin sets should not exceed 3 Kg  Containerized instrument sets do not exceed 11kg Tears easier than woven textiles Has memory  Advantages  Disadvantages IAHCSMM Central Service Technical Manual, 7th edition, Chapter 13 Sterile Processing University, LLC The Basics of Sterile Processing, 3rd edition, Chapter 7 Central Services Association of Ontario, The Manual for Reprocessing Medical Devices, 1st Edition, 2009
  • 33. Good to use for small, lightweight instruments or single items Contents easily visible Paper/Plastic Pouches: labeling is done on plastic side only Must select the correct size for contents so air can be expelled but not so large that the contents can move around and damage the seals Seal rupture possible Requires a heat seal  Advantages  Disadvantages IAHCSMM Central Service Technical Manual, 7th edition, Chapter 13 Sterile Processing University, LLC The Basics of Sterile Processing, 3rd edition, Chapter 7 Central Services Association of Ontario, The Manual for Reprocessing Medical Devices, 1st Edition, 2009
  • 34. Excellent barrier to microorganisms Water repellent Easy-to-use Eliminates torn wrappers No linting Protects instruments from damage during processing, storage, and transport Ergonomics - adds weight to instruments May need additional dry time Additional space needed to store containers but may be able to stack for sterile storage Additional labor for cleaning in between use and inspection of gaskets, latches, valves, etc  Advantages  Disadvantages IAHCSMM Central Service Technical Manual, 7th edition, Chapter 13 Sterile Processing University, LLC The Basics of Sterile Processing, 3rd edition, Chapter 7 Central Services Association of Ontario, The Manual for Reprocessing Medical Devices, 1st Edition, 2009
  • 40. Maintain items on sterilizer cart until adequately cooled Do not touch during the cooling process Could wick bacteria from hands into packaging Place sterilizer cart in a low traffic area, no air-conditioning or cold-air-vents Do not transfer warm items to a cool cart which could result in condensate forming and contamination May open door slightly at end of cycle to reduce the potential for condensation formation ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012, Section 8.8
  • 41. Is event related and depends on • Quality of packaging material • Storage conditions • Conditions during transportation • Amount of handling Label each product • Expiration date or statement such as: “contents sterile unless package is open or damaged. Please check before using.” ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012, Section 8.9.3 and 10.3.3
  • 46. Type of sterilizer Advantages Disadvantages Autoclave Inexpensive Fast Effective, with a wide margin of safety Penetrates medical packing, device lumens Non-toxic – low environmental impact Readily available Sterilizers are available in many sizes for many applications Cycle easy to control and monitor Unsuitable for anhydrous materials and for heat- and moisture-sensitive materials May leave instruments wet, causing them to rust Expertise Deleterious for heat-sensitive instruments Microsurgical instruments damaged by repeated exposure Potential for burns Hydrogen Peroxide (Plasma) Fast Safe for the environment Compatible with heat- and moisture-sensitive Absence of toxic waste Easy installation and operation No staff monitoring currently required No aeration required No chemical residues Cannot sterilize materials that absorb hydrogen peroxide (e.g. linen, gauze, cellulose/paper, wood) Low penetration power Medical devices must be dry before processing Cellulose (paper), linens and liquids cannot be processed Limitations to length of lumens of medical devices that can be effectively sterilized
  • 47. Type of sterilizer Advantages Disadvantages ETO Non-corrosive Penetrates packaging materials, device lumens Has some ability to penetrate some synthetic materials Excellent material compatibility Simple to operate and monitor Toxic/carcinogenic to humans Lengthy cycle due to aeration requirements Requires monitoring of the work areas Requires control and monitoring of discharge into the environment Flammable and explosive. Expensive
  • 48.  Implantable devices till the BI testing is satisfactory  Non implantable devices BI/ class 5or 6/specific cycle physical parameters in the load control
  • 49. RAPID CYCLE OR FLASH STERILIZATION  “Unwrapped” steam sterilization  Should only be used when necessary  Do not flash whole trays of instruments  Items must be used immediately  Avoid by keeping adequate supply of frequently dropped items  Maintain records or “flash logs”  Use to support need for additional instruments
  • 50. Nurse from OT calls the CSSD to inform the Arthroplasty set is wet. What to perform next? 1. Receive the wet set process sterilize and return 2. Receive all the sets sterilized in the same load 3. CSSD technician observes the external indicator is working and tells to use the instrument 4. Its an issue between CSSD technician and the OT nurses, no need to inform others
  • 51.  Load contains positive BI- failed load and also all loads sterilized in that sterilizer till last successfully sterilized load  Incorrect reprocessing of device  Print out of reprocessing equipment failure to reach correct parameters  CI has not changed colour  Doubt about sterility of equipment(torn, wet)  Notification to all concerned departments
  • 53. Controlled temperature 24°C ≥ four air exchanges/hour Relative humidity not to exceed 60% Controlled traffic Positive air flow (out) Sterile items are stored at least 20-25cm (8-10”) above the floor, at least 45cm (18”) below the ceiling or sprinkler heads, and at least 5cm (2”) from outside walls. It should be located near a dock for the receiving of materials.
  • 54. STERILE STORAGE Supplies are distributed on a First In First Out (FIFO) basis