DataColor standard operating procedures ( SOP )
Abdullah-Al-Mamun (Porag)
NITER 1st Batch, 01792-556070 Working-With-Data-Color Page 1
The following calibration procedure is generally followed-
1. Click the Calibrate button on the button bar, or Instrument Menu, Calibrate.
2. Click Calibrate. The Prepare for Calibration dialog box displays.
3. Place the black trap over the measurement port, and click the Ready button. If the
instrument does not have a black trap, clear the port of any obstruction and click the Ready
button. The Prepare for Calibration window redisplays when the measurements completed.
4. Place the white calibration tile over the measurement port, and click the Ready button.
When the second measurement is completed, the Control Measurement dialog box displays.
5. Click OK. If you have the diagnostic (green) tile test enabled, the program will prompt you
for the diagnostic tile.
6. Place the green tile at the port and click Ready. When the measurement is complete, the
results of the diagnostic tile test will be displayed.
7. The DECISION should be "Pass". Click OK to close the instrument.
Data color Tools Does the following Functions:
Measurement Procedure for Standard and Sample/Batch:
After completing the calibration procedure, the data color is ready for use. The measurement
procedure for standard sample-
1. Place the standard sample on the probe/port of the data color carefully(Usually four folded
is done)
2. Click the New Std button on the desktop. This clears the Name field, and places the cursor
in the field.
3. Enter a name for the sample. Sample data can come from the instrument, the database or
the keyboard. On the desktop there are two buttons that identify the origin of the standard
and batch measurements. (Here from Instrument).
4. The measurement method is set independently for each sample type. In this example, both
the standard and the batch measurement come from the instrument (Inst). Click the Std:
Inst button to start the measurement.
The sample just measured becomes the current standard on the desktop. Data fields on the desktop
will be updated to display the data for the sample just measured.
Data color Match’s does the following Functions: Colorant Sets:
A colorant set is a set of color information about the substrate and dyes the system uses to produce
match and correction recipes. It contains...
• Information about the overall colorant set, e.g., the substrate and process that will be used with
the dyes;
• Product information about each dye, e.g., strength, minimum and maximum concentrates;
• Color information about each dye;
Colorant Set procedure:
1. Open the “Colorant Set List” window.
2. On the “Colorant Set” or the context sensitive menu, select New Textile.
DataColor standard operating procedures ( SOP )
Abdullah-Al-Mamun (Porag)
NITER 1st Batch, 01792-556070 Working-With-Data-Color Page 2
3. Specify the name and the identification of the new colorant set.
4. A click in the „Dye Process“, „Substrate Delivery“ and „Operation“ field opens the
corresponding selection box. Select Input Form on the context sensitive menu of the
selection box to specify a new object.
5. Click Store.
Recipe Calculation (Matching): Selection process as follows
1. Quality/style (data of the substrate)
2. Combined process
3. Substrate delivery (only for deliveries with data different to the blank dyeing substrate)
4. Dyed substrate (over-dyeing only)
5. Dyestuff group with dyes pre-selected from the assigned colorant set. The dyestuff group is
used to optimize the recipe calculation.
6. Selection criteria: • Dyes from the list
7. Parameter values, e.g., fastness information
8. Concentration values, e.g., min., max., conc.
9. Settings (parameters for calculation control) • standard: Color to be matched.
Recipe Calculation procedure:
1. Open the “Recipe List” window.
2. On the “Recipe” or the context-sensitive menu, select Calculate, or press F5.
3. Select the quality/style and the combined process.
4. If necessary, select the substrate delivery.
5. Select the colorant set(s)
6. Select the “Standard” to be matched.
7. If you have to re-dye, select the dyed substrate or re-measure it.
8. In the “Colorant Set” tab select the dyestuffs to used and select or specify a dyestuff group.
9. In the “Settings” tab, select the parameters for the matching process.
10. Click Calculate to immediately start the recipe calculation. The recipe table appears.
11. Select the recipes you want to dye.
12. Click close, or Save to save the recipe.
Laboratory Correction recipe procedure:
1. Select the recipe to be corrected in the “Recipe List” window.
2. On the context-sensitive menu, select Pass Fail and Laboratory Correction, or pressF6.
3. Click Pass Fail and Correction.
4. In the “Batch and Color Difference” field, measure or select the sample.
5. If necessary, alter the data in the dyestuffs table.
6. Click Save to save a manual correction. The correction recipe is saved.
7. Click Laboratories. The “Recipe Correction” table appears.
8. In the “Recipe Correction” dialog box, you can look at the result of the matching. The color
differences between “Standard” and “Batch” are displayed.
9. If finished, close the recipe table.. Click Yes.
10. Select the recipe output(s) according your needs, and/or close the” Show Full Recipe”
dialog box.
For standards, do the following:
DataColor standard operating procedures ( SOP )
Abdullah-Al-Mamun (Porag)
NITER 1st Batch, 01792-556070 Working-With-Data-Color Page 3
1. Click the New Std button on the desktop. This clears the Name field, and places the cursor in
the field:
2. Sample data can come from the instrument, the database or the keyboard. On the desktop
there are two buttons that identify the origin of the standard and batch measurements.
3. The measurement method is set independently for each sample type. In this example, both
the standard and the batch measurement come from the instrument (Inst).
4. Place the standard at the instrument port, and click the Std:Inst button to start the
measurement. The sample just measured becomes the current standard on the desktop.
5. Data fields on the desktop will be updated to display the data for the sample just measured.
To change the measurement method for either the standard or batch:
1. Click on the down arrow to the view the measurement methods:
2. Click on the method to be used. The button will refresh, and display the measurement method
selected. In the example below, the standard measurement method has been changed to %R 31
points.
3. The next time you measure a standard; a window will open, and prompt you to enter 31-point
spectral data for the standard.
Color Evaluations Once you have selected a standard or batch, there are numerous color
evaluations you can perform.
Typically, the Datacolor TOOLS desktop includes several color evaluations including color
differences, pass/fail, color plots etc. The color evaluation displayed is always based on the current
standard and batch data on the desktop.
Basic Components of the Data color/Spectrophotometer:
All spectrophotometer instruments designed to measure the absorption of radiant Energy have the
basic components as follows:
1. A stable source of radiant energy (Light);
2. A wavelength selector to isolate a desired wavelength from the source (filter or
monochromatic);
3. Transparent container (curette) for the sample and the blank.
4. A radiation detector (phototube) to convert the radiant energy received to measurable
signal; and a readout device that displays the signal from the detector.
Recipe Calculation
Match: The recipes are calculated according to the selections and the results are displayed.
Review: The recipes can be reviewed according to the different criteria (various color difference
values, coordinates, price, etc.).
Further use: The recipes can be saved, printed and/or sent to a dispenser.
DataColor standard operating procedures ( SOP )
Abdullah-Al-Mamun (Porag)
NITER 1st Batch, 01792-556070 Working-With-Data-Color Page 4

More Related Content

PPTX
Diabetes Mellitus
PPTX
Hypertension
PPTX
Republic Act No. 11313 Safe Spaces Act (Bawal Bastos Law).pptx
PPTX
Power Point Presentation on Artificial Intelligence
PDF
Caça palavras - Bullying
PPTX
PDF
Atividade ortográfica - Caçada aos erros
Diabetes Mellitus
Hypertension
Republic Act No. 11313 Safe Spaces Act (Bawal Bastos Law).pptx
Power Point Presentation on Artificial Intelligence
Caça palavras - Bullying
Atividade ortográfica - Caçada aos erros

What's hot (20)

PPT
Reactive dye and disperse dye
PPTX
Fabric inspections
PPTX
Mercerization
PDF
LABORATORY CALCULATIONS & PROCEDRES
PPT
Mercerisation
PPTX
Color matching cabinet
DOCX
PC blend dyeing
PDF
Wet Testing of textile materials
PDF
Calculations in weft knitting machines
DOCX
One bath dyeing process of polyester,cotton(pc) blend fabric.
DOCX
Computer color matching system in textile industry
PPTX
Color matching
PDF
Burn-out printing
PPTX
package dyeing machine
PDF
Garment dyeing technique
PPTX
mass colaration
PPTX
Assesment of colour fastness
PDF
Dyeing of knitted fabrics
DOC
Garments washing imran personal
PPTX
Denim warp sheet dyeing
Reactive dye and disperse dye
Fabric inspections
Mercerization
LABORATORY CALCULATIONS & PROCEDRES
Mercerisation
Color matching cabinet
PC blend dyeing
Wet Testing of textile materials
Calculations in weft knitting machines
One bath dyeing process of polyester,cotton(pc) blend fabric.
Computer color matching system in textile industry
Color matching
Burn-out printing
package dyeing machine
Garment dyeing technique
mass colaration
Assesment of colour fastness
Dyeing of knitted fabrics
Garments washing imran personal
Denim warp sheet dyeing
Ad

Similar to datacolor spectrophotometer working procedure (20)

PDF
Spectrophotometer wroking procedur
PPT
Soft And Handling
PPTX
The 5 Steps of the Color Management Process
DOC
Spss notes
DOCX
SPSS User Guide
PPTX
Introduction to spss 2
PDF
Sampling
PDF
Cn mx3000 p_20120808e
PPTX
Color management
PDF
Eppendorf mastercyclertechnote
PDF
Cn abi vii_a7setup_20120808e
PDF
Training Outline HMLS Auto Calcimeter
PDF
InnerSoft STATS - Analyze
PPTX
QbD harsh new ggu central university .pptx
PPTX
Advanced Filter Concepts in MS-Excel
PPTX
How to use a Radionic Computer System
PDF
Som lc480 setup_e
PDF
Roche lc4801.2.9.11e
PDF
Abi7900 ht2.1e
Spectrophotometer wroking procedur
Soft And Handling
The 5 Steps of the Color Management Process
Spss notes
SPSS User Guide
Introduction to spss 2
Sampling
Cn mx3000 p_20120808e
Color management
Eppendorf mastercyclertechnote
Cn abi vii_a7setup_20120808e
Training Outline HMLS Auto Calcimeter
InnerSoft STATS - Analyze
QbD harsh new ggu central university .pptx
Advanced Filter Concepts in MS-Excel
How to use a Radionic Computer System
Som lc480 setup_e
Roche lc4801.2.9.11e
Abi7900 ht2.1e
Ad

Recently uploaded (20)

PDF
Computer organization and architecuture Digital Notes....pdf
PPTX
Information Storage and Retrieval Techniques Unit III
PPTX
PRASUNET_20240614003_231416_0000[1].pptx
PPTX
ai_satellite_crop_management_20250815030350.pptx
PDF
UEFA_Embodied_Carbon_Emissions_Football_Infrastructure.pdf
PPTX
Module 8- Technological and Communication Skills.pptx
PDF
UEFA_Carbon_Footprint_Calculator_Methology_2.0.pdf
PDF
Applications of Equal_Area_Criterion.pdf
PDF
Prof. Dr. KAYIHURA A. SILAS MUNYANEZA, PhD..pdf
PDF
Computer System Architecture 3rd Edition-M Morris Mano.pdf
PDF
Influence of Green Infrastructure on Residents’ Endorsement of the New Ecolog...
PDF
MLpara ingenieira CIVIL, meca Y AMBIENTAL
PPT
Chapter 1 - Introduction to Manufacturing Technology_2.ppt
PDF
Unit I -OPERATING SYSTEMS_SRM_KATTANKULATHUR.pptx.pdf
PPTX
Principal presentation for NAAC (1).pptx
PPTX
Sorting and Hashing in Data Structures with Algorithms, Techniques, Implement...
PPTX
CyberSecurity Mobile and Wireless Devices
PDF
Design of Material Handling Equipment Lecture Note
DOC
T Pandian CV Madurai pandi kokkaf illaya
PPTX
Amdahl’s law is explained in the above power point presentations
Computer organization and architecuture Digital Notes....pdf
Information Storage and Retrieval Techniques Unit III
PRASUNET_20240614003_231416_0000[1].pptx
ai_satellite_crop_management_20250815030350.pptx
UEFA_Embodied_Carbon_Emissions_Football_Infrastructure.pdf
Module 8- Technological and Communication Skills.pptx
UEFA_Carbon_Footprint_Calculator_Methology_2.0.pdf
Applications of Equal_Area_Criterion.pdf
Prof. Dr. KAYIHURA A. SILAS MUNYANEZA, PhD..pdf
Computer System Architecture 3rd Edition-M Morris Mano.pdf
Influence of Green Infrastructure on Residents’ Endorsement of the New Ecolog...
MLpara ingenieira CIVIL, meca Y AMBIENTAL
Chapter 1 - Introduction to Manufacturing Technology_2.ppt
Unit I -OPERATING SYSTEMS_SRM_KATTANKULATHUR.pptx.pdf
Principal presentation for NAAC (1).pptx
Sorting and Hashing in Data Structures with Algorithms, Techniques, Implement...
CyberSecurity Mobile and Wireless Devices
Design of Material Handling Equipment Lecture Note
T Pandian CV Madurai pandi kokkaf illaya
Amdahl’s law is explained in the above power point presentations

datacolor spectrophotometer working procedure

  • 1. DataColor standard operating procedures ( SOP ) Abdullah-Al-Mamun (Porag) NITER 1st Batch, 01792-556070 Working-With-Data-Color Page 1 The following calibration procedure is generally followed- 1. Click the Calibrate button on the button bar, or Instrument Menu, Calibrate. 2. Click Calibrate. The Prepare for Calibration dialog box displays. 3. Place the black trap over the measurement port, and click the Ready button. If the instrument does not have a black trap, clear the port of any obstruction and click the Ready button. The Prepare for Calibration window redisplays when the measurements completed. 4. Place the white calibration tile over the measurement port, and click the Ready button. When the second measurement is completed, the Control Measurement dialog box displays. 5. Click OK. If you have the diagnostic (green) tile test enabled, the program will prompt you for the diagnostic tile. 6. Place the green tile at the port and click Ready. When the measurement is complete, the results of the diagnostic tile test will be displayed. 7. The DECISION should be "Pass". Click OK to close the instrument. Data color Tools Does the following Functions: Measurement Procedure for Standard and Sample/Batch: After completing the calibration procedure, the data color is ready for use. The measurement procedure for standard sample- 1. Place the standard sample on the probe/port of the data color carefully(Usually four folded is done) 2. Click the New Std button on the desktop. This clears the Name field, and places the cursor in the field. 3. Enter a name for the sample. Sample data can come from the instrument, the database or the keyboard. On the desktop there are two buttons that identify the origin of the standard and batch measurements. (Here from Instrument). 4. The measurement method is set independently for each sample type. In this example, both the standard and the batch measurement come from the instrument (Inst). Click the Std: Inst button to start the measurement. The sample just measured becomes the current standard on the desktop. Data fields on the desktop will be updated to display the data for the sample just measured. Data color Match’s does the following Functions: Colorant Sets: A colorant set is a set of color information about the substrate and dyes the system uses to produce match and correction recipes. It contains... • Information about the overall colorant set, e.g., the substrate and process that will be used with the dyes; • Product information about each dye, e.g., strength, minimum and maximum concentrates; • Color information about each dye; Colorant Set procedure: 1. Open the “Colorant Set List” window. 2. On the “Colorant Set” or the context sensitive menu, select New Textile.
  • 2. DataColor standard operating procedures ( SOP ) Abdullah-Al-Mamun (Porag) NITER 1st Batch, 01792-556070 Working-With-Data-Color Page 2 3. Specify the name and the identification of the new colorant set. 4. A click in the „Dye Process“, „Substrate Delivery“ and „Operation“ field opens the corresponding selection box. Select Input Form on the context sensitive menu of the selection box to specify a new object. 5. Click Store. Recipe Calculation (Matching): Selection process as follows 1. Quality/style (data of the substrate) 2. Combined process 3. Substrate delivery (only for deliveries with data different to the blank dyeing substrate) 4. Dyed substrate (over-dyeing only) 5. Dyestuff group with dyes pre-selected from the assigned colorant set. The dyestuff group is used to optimize the recipe calculation. 6. Selection criteria: • Dyes from the list 7. Parameter values, e.g., fastness information 8. Concentration values, e.g., min., max., conc. 9. Settings (parameters for calculation control) • standard: Color to be matched. Recipe Calculation procedure: 1. Open the “Recipe List” window. 2. On the “Recipe” or the context-sensitive menu, select Calculate, or press F5. 3. Select the quality/style and the combined process. 4. If necessary, select the substrate delivery. 5. Select the colorant set(s) 6. Select the “Standard” to be matched. 7. If you have to re-dye, select the dyed substrate or re-measure it. 8. In the “Colorant Set” tab select the dyestuffs to used and select or specify a dyestuff group. 9. In the “Settings” tab, select the parameters for the matching process. 10. Click Calculate to immediately start the recipe calculation. The recipe table appears. 11. Select the recipes you want to dye. 12. Click close, or Save to save the recipe. Laboratory Correction recipe procedure: 1. Select the recipe to be corrected in the “Recipe List” window. 2. On the context-sensitive menu, select Pass Fail and Laboratory Correction, or pressF6. 3. Click Pass Fail and Correction. 4. In the “Batch and Color Difference” field, measure or select the sample. 5. If necessary, alter the data in the dyestuffs table. 6. Click Save to save a manual correction. The correction recipe is saved. 7. Click Laboratories. The “Recipe Correction” table appears. 8. In the “Recipe Correction” dialog box, you can look at the result of the matching. The color differences between “Standard” and “Batch” are displayed. 9. If finished, close the recipe table.. Click Yes. 10. Select the recipe output(s) according your needs, and/or close the” Show Full Recipe” dialog box. For standards, do the following:
  • 3. DataColor standard operating procedures ( SOP ) Abdullah-Al-Mamun (Porag) NITER 1st Batch, 01792-556070 Working-With-Data-Color Page 3 1. Click the New Std button on the desktop. This clears the Name field, and places the cursor in the field: 2. Sample data can come from the instrument, the database or the keyboard. On the desktop there are two buttons that identify the origin of the standard and batch measurements. 3. The measurement method is set independently for each sample type. In this example, both the standard and the batch measurement come from the instrument (Inst). 4. Place the standard at the instrument port, and click the Std:Inst button to start the measurement. The sample just measured becomes the current standard on the desktop. 5. Data fields on the desktop will be updated to display the data for the sample just measured. To change the measurement method for either the standard or batch: 1. Click on the down arrow to the view the measurement methods: 2. Click on the method to be used. The button will refresh, and display the measurement method selected. In the example below, the standard measurement method has been changed to %R 31 points. 3. The next time you measure a standard; a window will open, and prompt you to enter 31-point spectral data for the standard. Color Evaluations Once you have selected a standard or batch, there are numerous color evaluations you can perform. Typically, the Datacolor TOOLS desktop includes several color evaluations including color differences, pass/fail, color plots etc. The color evaluation displayed is always based on the current standard and batch data on the desktop. Basic Components of the Data color/Spectrophotometer: All spectrophotometer instruments designed to measure the absorption of radiant Energy have the basic components as follows: 1. A stable source of radiant energy (Light); 2. A wavelength selector to isolate a desired wavelength from the source (filter or monochromatic); 3. Transparent container (curette) for the sample and the blank. 4. A radiation detector (phototube) to convert the radiant energy received to measurable signal; and a readout device that displays the signal from the detector. Recipe Calculation Match: The recipes are calculated according to the selections and the results are displayed. Review: The recipes can be reviewed according to the different criteria (various color difference values, coordinates, price, etc.). Further use: The recipes can be saved, printed and/or sent to a dispenser.
  • 4. DataColor standard operating procedures ( SOP ) Abdullah-Al-Mamun (Porag) NITER 1st Batch, 01792-556070 Working-With-Data-Color Page 4