High-throughput capillary-flow LC-MS proteomics
with maximum MS utilization
2
Content
• The Thermo Fisher LC Portfolio: The right system for every application
• The need for speed in low-flow LC-MS proteomics: high-throughput
low-flow LC-MS solutions
• Capillary-flow LC-MS with 150 µm ID columns
• High-throughput LC-MS with 75 µm ID columns
• 100% MS Utilization with Tandem LC-MS
• Summary
3
The Thermo Fisher LC Portfolio: The right system for every application
4
Column ID: 50 µm 75 µm 150 µm 300 µm 0.5 mm 1.0 mm …2.1–4.6 mm
Nano: <1.5 µL/min
Capillary: 1–15 µL/min
Micro: 10–100 µL/min
Analytical: >100 µL/min
50 µL/min–5 mL/min
100 µL/min–8 mL/min100–1000
nL/min
50 nL/min–50 µL/min
Sensitivity
Throughput
Based on recommended flow rate
Thermo Scientific™
UltiMate™ 3000 RSLCnano System
Thermo Scientific™
Vanquish™ Horizon System
Thermo
Scientific™
UltiMate™ 3000
Systems
Thermo Scientific™
Vanquish™ Flex and
Vanquish™ Duo UHPLC
Systems
Thermo Scientific™
EASY-nLC™ 1200 System
Thermo Scientific HPLC and UHPLC Systems: LC Portfolio for LC-MS Applications
5
Why do low-flow LC-MS: what are the benefits?
0
20
40
60
80
100
120
140
160
180
200
0 100 200 300 400 500
0
10
20
30
40
50
60
0 20 40 60 80 100
CapLC-MS
MicroLC-MS
NanoLC-MS
Analytical LC-MS
Flow rate, µL/min
Sensitivitygainincomparisonwith
analyticalLC-MSat450µL/min
Low flow LC-MS
The sensitivity gains were measured as a relative peak area
averaged for Cytochrome C tryptic peptides
ESI-MS exhibits a mixed behavior of both
concentration and mass-sensitive
detectors
Sensitivity gains (experimental,
relative to 2.1 mm ID)
microLC-MS: 2-4
capLC-MS: 4-50
nanoLC-MS: >50
sub-nanoLC-MS: >100
Reasons
• Improved ionization efficiency
• Smaller droplets
• Increased surface to volume ratio
• Increase of analyte concentration with
decrease of column ID
6
Sensitivity: why is it improved with low-flow LC?
HESI Source Thermo Scientific™
EASY-Spray™ Source
Significant improvements in sensitivity for LC-MS analysis using an
ESI source are observed at flow rates below 5 µL/min
7
Capillary-flow LC with the Thermo Scientific UltiMate 3000 RSLCnano system
Micro flow ternary pump for fast sample
loading in trap and elute experiments
Flow up to 2.5 mL/min, Pressure 620 bar
Near zero sample loss with unique μL
pick-up injections
High precision for sub-μL volume
injections
UHPLC binary capillary flow pump
for robust separations and high
retention time precision
Settable flow up to 50 µL/min
Two low-dispersion snap-in valves
for standardized workflows and
custom column switching
applications
~100 nL port to port volume
Thermo Scientific™ nanoViper™
Fingertight fittings for ease-of-use
and leak prevention
8
The need for speed in low-flow LC-MS proteomics: high-throughput solutions
9
“Triangular” and “rectangular” biomarker discovery and validation pipeline
Throughput
Sensitivity
Discovery
• High number of samples
• Targeted quantification
• Minimal sample preparation
• LC-MS vs. ELISA
• Low number sample
• Multiplexed labelling
• Pooled samples
• Fractionation for deep
quantitative profiling
Analytical flow
LC-MS on 2.1
mm ID column
Nano-flow LC-
MS method on
50-75 µm ID
column
Sub-nano-flow
LC-MS
<100 nL/min
Validation
1000’s samples
1000’s analytes
Discovery
1000’s samples
1000’s analytes
• Similar large cohort size in discovery
and validation
• Same sensitivity and number of targets
• Same robust LC-MS technology for
discovery and validation
• High analytical throughput
10
How to acheive high-throughput with low-flow LCMS?
UltiMate 3000 RSLCnano system
Thermo Scientific™ Q Exactive™
HF-X Hybrid Quadrupole-Orbitrap™
mass spectrometer
HESI probe with 50 µm ID
stainless steel needle
EASY-Spray source
Ion Max source
EASY-Spray transfer line
with 20 µm ID silica
emitter
Capillary column
Q Exactive HF
Capillary flow
UltiMate 3000 RSLCnano
Trap cartridge
Recommended
AlternativeThermo Scientific™ EASY-Spray™
column
150 µm x 150 mm, 2 µm – ES806
75 µm x 150 mm, 3 µm – ES800
Robust & SensitiveEasy-to-use &
• Versatile platform – safe investment
• Fast sample loading with integrated micro-flow
loading pump
• Fast injection routines
• UHPLC capabilities
• Optimized fluidics
• Simple installation and operation
• Low ID and short columns with low internal
volumes
• Temperature control to accommodate
increased flow rates
• High loading and peak capacity
• Fast scan speed to detect very narrow
peaks
• Improved sensitivity at capillary and nano-
flow rates
• Advanced precursor detection and charge
state recognition (APD)
Solutions
introduced
@ASMS2017
@ASMS2018
11
• Translational proteomics and biomarker
validation
• Population monitoring
• Samples characterization for Biobanking
• Food and beverage authentication
• …
Demands for high-throughput low-flow LCMS
Applications Requirements
1.Robustness
2.Sensitivity
3.Throughput
4.Easy handling and setup of fluidic connections
5.Plug & play column consumable with UHPLC
capability
Flow rate, µL/min
Throughput
75 µm ID
150 µm ID
Flow rate, µL/min
Sensitivity
75 µm ID
150 µm ID
• Balance between sensitivity and
throughput
• Higher linear velocity results in higher
throughput, but lower sensitivity
• Ultra-fast low flow methods requires the
operation of low ID columns at maximum
possible linear velocity
12
Capillary-flow LC-MS with 150 µm ID columns
150 µm x 150
mm, 2 µm
ES806
Q Exactive HF-X
UltiMate 3000
RSLCnano system Flow rate, µL/min
75 µm
150 µm
Sensitivity
13
Direct injection: how to get it right?
3 1.2 3
Flow rate, µL/min
Sample loading Column washing
and equilibration
0.0E+00
2.0E+08
4.0E+08
6.0E+08
8.0E+08
1.0E+09
1.2E+09
1.4E+09
0
20
40
60
80
100
0 10 20 30 40 50 60
Signal
%B
Time, min
Sample loop is switched offline
to reduce delay volume and is
washed to avoid carryover
Peptide separation
14
Single shot capLC-MS analysis
3,507
3,813
4,392
4,060
4,301
4,826
0
1000
2000
3000
4000
5000
6000
1 ug 2 ug 5ug
Proteingroups
21002
23918
2573426708
28515
30529
0
5000
10000
15000
20000
25000
30000
35000
1 ug 2 ug 5ug
Peptidegroups
• Deep proteome coverage with single shot analysis
(>4000 protein groups in 60 min)
• Comparable to nanoLC-MS performance with
increased sample loading
0 20 40 60 80
Signal
Time, min
Direct injection
60 and 90 min cycle time
24 and 18 samples per day
60min
90min
60min
90min
60min
90min
60min
90min
60min
90min
60min
90min
15
Quantitative reproducibility
4578 4550 4553
Column 1 Column 2 Column 3
29307 29552 30448
Column 1 Column 2 Column 3
RSD=1.8% RSD=1.9%
Durable response factor
4578 4520 4209
85%
79%
87%
Injection
1
Injection
350
Injection
1000
Long-term quantitative performance
Protein Groups Peptide Groups
Quantified,
CV <20%
Column to column reproducibility
Protein Groups
• Excellent column-to-column consistency was
observed, including coverage of >4500 proteins,
and ~30000 peptides all with RSD <2% between
columns
• Comparable proteome coverage and quantitative
precision was observed throughout the 1000
injection experiment
16
Where to find more information about capLC-MS?
Link
Link
Link
Link
Link
Link
More information on our web-page
https://www.thermofisher.com/nanoLCMS
17
High-throughput LC-MS with 75 µm ID columns
75 µm x 150 mm, 3 µm
PN ES800
Q Exactive HF-X Flow rate, µL/min
Throughput
75 µm
150 µm
18
How to maximize low-flow LC-MS throughput?
Delay volume is around
1.0 µL
100nL
70nL
300nL
70nL
100nL
150nL
400nL
Back
Flush
100nL
Gradient delay volume is
below 300 nL
Total internal volume ~ 1 µL
ES800
75 µm x 150 mm, 3 µm
Separation pump
+
Column compartment
+
Micro-flow loading pump
Autosampler
• Low gradient delay volume
• Low total internal volume
• High linear velocity – low delay time
• Fast and precise gradient formation
Nano-flow standard Pre-concentration kit
19
High-throughput low-flow LC method explained
Fast injection routine (< 1 min)
Fast sample loading onto the column
with micro-flow loading pump (6 sec
@150 µL/min)
Parallel sample elution and fluidic
washing steps
Parallel column washing/equilibration
and sampler injection routine
20
Typical BPC and TIC chromatograms
Wide peptide elution window Extremely low carryover
Zoom in
21
Robustness of High-throughput LC-MS Method
0
2000
4000
6000
8000
10000
12000
0 30 60 90 120 150 180
Injection No.
Protein groups PSMs
MS/MS Peptide groups
0
1
2
3
4
5
6
0 30 60 90 120 150 180
Retentiontime,min
Injection No.
0
1
2
3
4
0 30 60 90 120 150 180
PWHM,sec
Injection No.
[K].LTDcVVMR.[D] [K].SLTNDWEDHLAVK.[H]
[K].STELLIR.[K] [K].TVTAMDVVYALK.[R]
[K].VDNDENEHQLSLR.[T] [K].VNQIGSVTESIQAcK.[L]
[R].EDSQRPGAHLTVK.[K] [R].TVSLGAGAK.[D]
Protein groups
Peptide groups PSMs
MS/MS
0.0E+00
1.0E+07
2.0E+07
3.0E+07
4.0E+07
5.0E+07
0.E+00
2.E+06
4.E+06
6.E+06
8.E+06
1.E+07
0 30 60 90 120 150 180
Area
Area
Injection No.
Retention time Peak area
PWHM
SD < 0.1 min
PWHM < 3 sec
RSD < 10%
Maximum speed
TOP 40
MS2 7.5K
IT MS2 14 sec
22
How do we outperform competitors?
14.4 minutes
8 minutes
Vendor “E”
UltiMate 3000 RSLCnano
180 samples/day
MS utilization – 75% per run
100 samples/day
MS utilization – 42% per run
SignalSignal
0 71 62 53 4
Previous
injection
Next
injection
Previous
injection
Next
injection
1 min
3 min
UltiMate 3000 RSLCnano system
compared to Vendor “E”:
Almost double the throughput
Higher peak capacity
More equal peptide distribution
23
ASMS2018 poster
High-throughput LC-MS: where to find more information?
Post 1
Post 2
Post 3
Tweet 3
Tweet 2
Tweet 1
Link
In review
24
2 x 75 µm x 150 mm,
3 µm; PN 164568
Q Exactive HF-XTandem Pre-concentration setup with
UltiMate 3000 RSLCnano system
Flow rate, µL/min
Throughput
75 µm
150 µm
100% MS Utilization with Tandem LC-MSTandem low-flow LCMS for maximum MS utilization
25
Enhanced Productivity with Tandem low-flow LC
• Tandem nano LC is applied to increase throughput.
• Analyte separation is carried out on two columns in an alternating fashion
• While one column is used for separation the other is washed and re-equilibrated for the next run
• Tandem nano LC employed for long gradient / deep dive proteomics applications run on long columns,
enables multiple washing steps to be carried out on the offline column.
Tandem operation – Separation matches washing and
equilibration
Cleaning operation -Long gradient for separation
on Column A is combined with consecutive short
gradients for washing column B
0
20
40
60
80
100
1 11 21 31 41 51 61
71
81
91
101
111
121
Column A Column B
0
20
40
60
80
100
1 11 21 31 41 51 61
71
81
91
101
111
121
Column A Column B
%,B
%,B
26
High throughput low-flow Tandem LC setup
to ESI MS
waste
waste
Separation pump Equilibration pump
Separation pump
+
Column compartment
+
Micro-flow loading pump
Equilibration pump
Autosampler
3 pumps
2 nano columns
2 trap cartridges
2 switching valves
One
LC-MS
method
27
Tandem low-flow LC-MS method explained
Peptide elution window: ~ 6.5 min
Column 1, cycle time: 7 min
Method duration and raw file length: 7 min
Prepare for
next injection
routine
Sample loop
strong solvent
wash
Trap 2 strong
solvent wash
& equilibration
Done in parallel
with loading pump,
syringe, 2nd nano
pump
Sampler wash
between
reinjections
Sample load
onto trap 2
Sample
transfer to 2nd
column
Column 2 washing Column 2 equilibration
Trap 2 washing
Next gradient
on column 2
Previous
gradient on
column 2
Look ahead injections: sample is
loaded onto the second trap
cartridge in parallel with current
run
Trap cartridge is washed with
strong wash solvent using micro-
flow pump
Injection fluidics is washed in
parallel with peptide separation
The delay in peptide elution is
removed by intelligent use of
post column flow diversion
Method can be fully adapted
according to the required
gradient length
The same pump is used for
separation on column 1 and
column 2
28
High throughput, Tandem low-flow LC: what are the benefits?
RT: 0.00-7.01
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0
Time(min)
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
3.63
3.65
3.71
4.50
3.13
1.80
2.401.82 5.74
3.43 5.77
1.78
2.82
5.80
3.78 4.17
2.42
1.84
2.55
0.820.80 1.04 1.13
4.94
3.95 4.971.470.490.34 1.97 5.38
5.854.600.30 5.15 6.345.99
6.856.87
6.46
NL: 9.46E8
BasePeakF: FTMS+p
NSIFull ms
[350.0000-2000.0000]
MS
7min_HeLa_200ng_MS1_
column1_2018042022021
4
RT: 0.00-7.01
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0
Time(min)
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
RelativeAbundance
3.62
3.60
3.68
4.543.70
4.52
4.56
3.06
3.05
5.96
3.42 5.93
2.24
4.16
1.60
1.59 2.71
1.62
4.212.26
2.98
2.280.15 3.28 5.571.65
0.66 5.020.16 5.031.41 6.003.950.81 4.99
0.99 5.590.34
6.025.234.742.021.02 6.09 6.481.99 6.50
6.60
NL: 7.91E8
BasePeakF: FTMS+p
NSIFull ms
[350.0000-2000.0000]
MS
7min_HeLa_200ng_MS1_
column2_2018042022092
5
Column 1 Column 2
7 min 7 min
972
5374
6129
10399
1175
5587
6689
12948
0
2000
4000
6000
8000
10000
12000
14000
Pre-concentration Tandem
Protein
groups
Peptide
groups
PSMs
0
200
400
600
800
Count
0.500
1.000
1.500
2.000
2.500
3.000
3.500
4.000
4.500
5.000
5.500
6.000
6.500
7.000
PSMs-RT[min]
PSMs
> 200 samples per 24 hours
~100% MS utilization
Good separation performance
Reproducible peptide separation
profiles on column 1 and column 2
Maximum MS speed utilization
29
Summary
30
Conclusions
• Time is Money!
• There is an increasing demand for speed, sensitivity and MS utility in LC-MS applications in industrial and translational
proteomics and beyond
• The UltiMate solution for all your low-flow needs
• The UltiMate 3000 RSLCnano is uniquely capable of delivering ultra high throughput, low-flow applications due to its:
• Integrated micro-flow pump for high speed online sample loading and desalting and trap and loop washing
• Wide pressure / flow footprint of 50 nL/min to 50 µL/min at up to 900 bar column pressure
• Robust pressure stable tool free industry leading nanoViper connections
• Offering high throughput AND sensitivity
• High through-put low-flow LC-MS methods using nano columns and capillary flow rates yield uncompromising levels of
speed and sensitivity whilst delivering as yet unseen levels of MS utility
• Make the most of your MS with Tandem low-flow LCMS
• Tandem LC-MS is possible through the simple addition of a low-flow pump and column set
• Intelligent method control combined with look ahead injections enable continuous back-to-back MS data generation for
the first time in the history of low-flow LC-MS
31
• Complete with schematics, method
parameters and example data
Last but not least…. The ALL NEW
• Comprehensive and
easy to follow
• Detailed information from
instrument setup to best
practices
• Available for free download on the RSLCnano webpage
• https://www.thermofisher.com/order/catalog/product/ULTIM3000RSLCNANO
• And as a hard copy shipped with each application kit
• UltiMate 3000 RSLCnano Standard Applications Manual (V 3.0)
32
Thank you!

More Related Content

PPTX
Electrophoresis
PPTX
Ion pair , reversed pair liquid chromatography
PPTX
HPLC[ HIGH PERPROMANCE LIQUID CHROMATOGRAPHY OR HIGH PRESSURE LIQUID CHROMAT...
PPTX
Ion Selective Electrode
PPTX
Ion selective electrodes(ise)
PPTX
Fluorometry
PPTX
Fluorometry
PPTX
Simplifying Chromatographic Methods Transfer: Novel Tools for Replicating You...
Electrophoresis
Ion pair , reversed pair liquid chromatography
HPLC[ HIGH PERPROMANCE LIQUID CHROMATOGRAPHY OR HIGH PRESSURE LIQUID CHROMAT...
Ion Selective Electrode
Ion selective electrodes(ise)
Fluorometry
Fluorometry
Simplifying Chromatographic Methods Transfer: Novel Tools for Replicating You...

What's hot (20)

PPTX
Fluorimetry presentation
PDF
Ion exchange chromatography
PPTX
Glucose sensor
PPTX
Introduction to UV-based detectors
PDF
Flame Photometric Estimation
PPTX
Size exclusion chromatography
PPTX
chemicalionization
DOCX
Report on Lab on a chip
PPTX
Ion exchange chromatography PPT.
PPTX
Derivitization of gc
PPTX
Flourimetry
PDF
UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application...
PDF
Near Infrared & Raman Spectroscopy.pdf
PPTX
Capillary electrophoresis
PPTX
Flame photometer (Atomic Emission Spectroscopy) Flame emission spectroscopy
PPT
Interfaces in chromatography [LC-MS, GC-MS, HPTLC, LC, GC]
PPTX
Lcms Basics
PPTX
PPTX
Atomic absorption spectrophotometry
PPT
Non protein nitrogen
Fluorimetry presentation
Ion exchange chromatography
Glucose sensor
Introduction to UV-based detectors
Flame Photometric Estimation
Size exclusion chromatography
chemicalionization
Report on Lab on a chip
Ion exchange chromatography PPT.
Derivitization of gc
Flourimetry
UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application...
Near Infrared & Raman Spectroscopy.pdf
Capillary electrophoresis
Flame photometer (Atomic Emission Spectroscopy) Flame emission spectroscopy
Interfaces in chromatography [LC-MS, GC-MS, HPTLC, LC, GC]
Lcms Basics
Atomic absorption spectrophotometry
Non protein nitrogen
Ad

Similar to High-throughput capillary-flow LC-MS proteomics with maximum MS utilisation (20)

PPTX
Separate Your Science LC tour introduction
PDF
New Innovations in Ultra High Performance Liquid Chromatography and Liquid Ch...
PPTX
Recent advances in hplc and gc
PPTX
Mayuri shitre
PPTX
Peptide_Bioanalysis (1)
PPTX
Strategies for bioanalysis of proteins using LC-MS
PPTX
LC-MS
PPTX
PPTX
Nano liquid chromatography (ncl)
PPT
LC-MS in bioactivity screening and proteomics
PDF
From sample-to-spray: high performance workflow for top down protein analysis
PDF
Enhanced bioseparations peptide mapping and m abs
PDF
Pesticide Residue Analysis Webinar Series: Tips and Tricks for the Whole Work...
PPTX
Liquid Chromatoghraphy / Mass Spectroscopy(LCMS)
PPTX
Milton lee lecture-2
PPTX
LC MS INSTUMENTION PRINCIPLE WORKING HOW THEY CONECT TO EACH OTHER
PPTX
Microfluidics
PDF
Lcms 2020
PDF
Peptide and Protein Bioanalysis: Infographic
PPT
HPLC2005
Separate Your Science LC tour introduction
New Innovations in Ultra High Performance Liquid Chromatography and Liquid Ch...
Recent advances in hplc and gc
Mayuri shitre
Peptide_Bioanalysis (1)
Strategies for bioanalysis of proteins using LC-MS
LC-MS
Nano liquid chromatography (ncl)
LC-MS in bioactivity screening and proteomics
From sample-to-spray: high performance workflow for top down protein analysis
Enhanced bioseparations peptide mapping and m abs
Pesticide Residue Analysis Webinar Series: Tips and Tricks for the Whole Work...
Liquid Chromatoghraphy / Mass Spectroscopy(LCMS)
Milton lee lecture-2
LC MS INSTUMENTION PRINCIPLE WORKING HOW THEY CONECT TO EACH OTHER
Microfluidics
Lcms 2020
Peptide and Protein Bioanalysis: Infographic
HPLC2005
Ad

Recently uploaded (20)

PDF
Placing the Near-Earth Object Impact Probability in Context
PDF
Cosmic Outliers: Low-spin Halos Explain the Abundance, Compactness, and Redsh...
PPTX
Microbes in human welfare class 12 .pptx
PPTX
GREEN FIELDS SCHOOL PPT ON HOLIDAY HOMEWORK
PDF
Worlds Next Door: A Candidate Giant Planet Imaged in the Habitable Zone of ↵ ...
PPTX
Hypertension_Training_materials_English_2024[1] (1).pptx
PPT
LEC Synthetic Biology and its application.ppt
PPTX
Lesson-1-Introduction-to-the-Study-of-Chemistry.pptx
PPTX
BODY FLUIDS AND CIRCULATION class 11 .pptx
PPTX
Welcome-grrewfefweg-students-of-2024.pptx
PDF
Unit 5 Preparations, Reactions, Properties and Isomersim of Organic Compounds...
PPT
THE CELL THEORY AND ITS FUNDAMENTALS AND USE
PDF
BET Eukaryotic signal Transduction BET Eukaryotic signal Transduction.pdf
PPTX
BIOMOLECULES PPT........................
PPTX
Seminar Hypertension and Kidney diseases.pptx
PPTX
Understanding the Circulatory System……..
PPTX
SCIENCE 4 Q2W5 PPT.pptx Lesson About Plnts and animals and their habitat
PDF
GROUP 2 ORIGINAL PPT. pdf Hhfiwhwifhww0ojuwoadwsfjofjwsofjw
PDF
S2 SOIL BY TR. OKION.pdf based on the new lower secondary curriculum
PDF
Communicating Health Policies to Diverse Populations (www.kiu.ac.ug)
Placing the Near-Earth Object Impact Probability in Context
Cosmic Outliers: Low-spin Halos Explain the Abundance, Compactness, and Redsh...
Microbes in human welfare class 12 .pptx
GREEN FIELDS SCHOOL PPT ON HOLIDAY HOMEWORK
Worlds Next Door: A Candidate Giant Planet Imaged in the Habitable Zone of ↵ ...
Hypertension_Training_materials_English_2024[1] (1).pptx
LEC Synthetic Biology and its application.ppt
Lesson-1-Introduction-to-the-Study-of-Chemistry.pptx
BODY FLUIDS AND CIRCULATION class 11 .pptx
Welcome-grrewfefweg-students-of-2024.pptx
Unit 5 Preparations, Reactions, Properties and Isomersim of Organic Compounds...
THE CELL THEORY AND ITS FUNDAMENTALS AND USE
BET Eukaryotic signal Transduction BET Eukaryotic signal Transduction.pdf
BIOMOLECULES PPT........................
Seminar Hypertension and Kidney diseases.pptx
Understanding the Circulatory System……..
SCIENCE 4 Q2W5 PPT.pptx Lesson About Plnts and animals and their habitat
GROUP 2 ORIGINAL PPT. pdf Hhfiwhwifhww0ojuwoadwsfjofjwsofjw
S2 SOIL BY TR. OKION.pdf based on the new lower secondary curriculum
Communicating Health Policies to Diverse Populations (www.kiu.ac.ug)

High-throughput capillary-flow LC-MS proteomics with maximum MS utilisation

  • 1. The world leader in serving science Alexander Boychenko and Chris Pynn 2018-07-26 High-throughput capillary-flow LC-MS proteomics with maximum MS utilization
  • 2. 2 Content • The Thermo Fisher LC Portfolio: The right system for every application • The need for speed in low-flow LC-MS proteomics: high-throughput low-flow LC-MS solutions • Capillary-flow LC-MS with 150 µm ID columns • High-throughput LC-MS with 75 µm ID columns • 100% MS Utilization with Tandem LC-MS • Summary
  • 3. 3 The Thermo Fisher LC Portfolio: The right system for every application
  • 4. 4 Column ID: 50 µm 75 µm 150 µm 300 µm 0.5 mm 1.0 mm …2.1–4.6 mm Nano: <1.5 µL/min Capillary: 1–15 µL/min Micro: 10–100 µL/min Analytical: >100 µL/min 50 µL/min–5 mL/min 100 µL/min–8 mL/min100–1000 nL/min 50 nL/min–50 µL/min Sensitivity Throughput Based on recommended flow rate Thermo Scientific™ UltiMate™ 3000 RSLCnano System Thermo Scientific™ Vanquish™ Horizon System Thermo Scientific™ UltiMate™ 3000 Systems Thermo Scientific™ Vanquish™ Flex and Vanquish™ Duo UHPLC Systems Thermo Scientific™ EASY-nLC™ 1200 System Thermo Scientific HPLC and UHPLC Systems: LC Portfolio for LC-MS Applications
  • 5. 5 Why do low-flow LC-MS: what are the benefits? 0 20 40 60 80 100 120 140 160 180 200 0 100 200 300 400 500 0 10 20 30 40 50 60 0 20 40 60 80 100 CapLC-MS MicroLC-MS NanoLC-MS Analytical LC-MS Flow rate, µL/min Sensitivitygainincomparisonwith analyticalLC-MSat450µL/min Low flow LC-MS The sensitivity gains were measured as a relative peak area averaged for Cytochrome C tryptic peptides ESI-MS exhibits a mixed behavior of both concentration and mass-sensitive detectors Sensitivity gains (experimental, relative to 2.1 mm ID) microLC-MS: 2-4 capLC-MS: 4-50 nanoLC-MS: >50 sub-nanoLC-MS: >100 Reasons • Improved ionization efficiency • Smaller droplets • Increased surface to volume ratio • Increase of analyte concentration with decrease of column ID
  • 6. 6 Sensitivity: why is it improved with low-flow LC? HESI Source Thermo Scientific™ EASY-Spray™ Source Significant improvements in sensitivity for LC-MS analysis using an ESI source are observed at flow rates below 5 µL/min
  • 7. 7 Capillary-flow LC with the Thermo Scientific UltiMate 3000 RSLCnano system Micro flow ternary pump for fast sample loading in trap and elute experiments Flow up to 2.5 mL/min, Pressure 620 bar Near zero sample loss with unique μL pick-up injections High precision for sub-μL volume injections UHPLC binary capillary flow pump for robust separations and high retention time precision Settable flow up to 50 µL/min Two low-dispersion snap-in valves for standardized workflows and custom column switching applications ~100 nL port to port volume Thermo Scientific™ nanoViper™ Fingertight fittings for ease-of-use and leak prevention
  • 8. 8 The need for speed in low-flow LC-MS proteomics: high-throughput solutions
  • 9. 9 “Triangular” and “rectangular” biomarker discovery and validation pipeline Throughput Sensitivity Discovery • High number of samples • Targeted quantification • Minimal sample preparation • LC-MS vs. ELISA • Low number sample • Multiplexed labelling • Pooled samples • Fractionation for deep quantitative profiling Analytical flow LC-MS on 2.1 mm ID column Nano-flow LC- MS method on 50-75 µm ID column Sub-nano-flow LC-MS <100 nL/min Validation 1000’s samples 1000’s analytes Discovery 1000’s samples 1000’s analytes • Similar large cohort size in discovery and validation • Same sensitivity and number of targets • Same robust LC-MS technology for discovery and validation • High analytical throughput
  • 10. 10 How to acheive high-throughput with low-flow LCMS? UltiMate 3000 RSLCnano system Thermo Scientific™ Q Exactive™ HF-X Hybrid Quadrupole-Orbitrap™ mass spectrometer HESI probe with 50 µm ID stainless steel needle EASY-Spray source Ion Max source EASY-Spray transfer line with 20 µm ID silica emitter Capillary column Q Exactive HF Capillary flow UltiMate 3000 RSLCnano Trap cartridge Recommended AlternativeThermo Scientific™ EASY-Spray™ column 150 µm x 150 mm, 2 µm – ES806 75 µm x 150 mm, 3 µm – ES800 Robust & SensitiveEasy-to-use & • Versatile platform – safe investment • Fast sample loading with integrated micro-flow loading pump • Fast injection routines • UHPLC capabilities • Optimized fluidics • Simple installation and operation • Low ID and short columns with low internal volumes • Temperature control to accommodate increased flow rates • High loading and peak capacity • Fast scan speed to detect very narrow peaks • Improved sensitivity at capillary and nano- flow rates • Advanced precursor detection and charge state recognition (APD) Solutions introduced @ASMS2017 @ASMS2018
  • 11. 11 • Translational proteomics and biomarker validation • Population monitoring • Samples characterization for Biobanking • Food and beverage authentication • … Demands for high-throughput low-flow LCMS Applications Requirements 1.Robustness 2.Sensitivity 3.Throughput 4.Easy handling and setup of fluidic connections 5.Plug & play column consumable with UHPLC capability Flow rate, µL/min Throughput 75 µm ID 150 µm ID Flow rate, µL/min Sensitivity 75 µm ID 150 µm ID • Balance between sensitivity and throughput • Higher linear velocity results in higher throughput, but lower sensitivity • Ultra-fast low flow methods requires the operation of low ID columns at maximum possible linear velocity
  • 12. 12 Capillary-flow LC-MS with 150 µm ID columns 150 µm x 150 mm, 2 µm ES806 Q Exactive HF-X UltiMate 3000 RSLCnano system Flow rate, µL/min 75 µm 150 µm Sensitivity
  • 13. 13 Direct injection: how to get it right? 3 1.2 3 Flow rate, µL/min Sample loading Column washing and equilibration 0.0E+00 2.0E+08 4.0E+08 6.0E+08 8.0E+08 1.0E+09 1.2E+09 1.4E+09 0 20 40 60 80 100 0 10 20 30 40 50 60 Signal %B Time, min Sample loop is switched offline to reduce delay volume and is washed to avoid carryover Peptide separation
  • 14. 14 Single shot capLC-MS analysis 3,507 3,813 4,392 4,060 4,301 4,826 0 1000 2000 3000 4000 5000 6000 1 ug 2 ug 5ug Proteingroups 21002 23918 2573426708 28515 30529 0 5000 10000 15000 20000 25000 30000 35000 1 ug 2 ug 5ug Peptidegroups • Deep proteome coverage with single shot analysis (>4000 protein groups in 60 min) • Comparable to nanoLC-MS performance with increased sample loading 0 20 40 60 80 Signal Time, min Direct injection 60 and 90 min cycle time 24 and 18 samples per day 60min 90min 60min 90min 60min 90min 60min 90min 60min 90min 60min 90min
  • 15. 15 Quantitative reproducibility 4578 4550 4553 Column 1 Column 2 Column 3 29307 29552 30448 Column 1 Column 2 Column 3 RSD=1.8% RSD=1.9% Durable response factor 4578 4520 4209 85% 79% 87% Injection 1 Injection 350 Injection 1000 Long-term quantitative performance Protein Groups Peptide Groups Quantified, CV <20% Column to column reproducibility Protein Groups • Excellent column-to-column consistency was observed, including coverage of >4500 proteins, and ~30000 peptides all with RSD <2% between columns • Comparable proteome coverage and quantitative precision was observed throughout the 1000 injection experiment
  • 16. 16 Where to find more information about capLC-MS? Link Link Link Link Link Link More information on our web-page https://www.thermofisher.com/nanoLCMS
  • 17. 17 High-throughput LC-MS with 75 µm ID columns 75 µm x 150 mm, 3 µm PN ES800 Q Exactive HF-X Flow rate, µL/min Throughput 75 µm 150 µm
  • 18. 18 How to maximize low-flow LC-MS throughput? Delay volume is around 1.0 µL 100nL 70nL 300nL 70nL 100nL 150nL 400nL Back Flush 100nL Gradient delay volume is below 300 nL Total internal volume ~ 1 µL ES800 75 µm x 150 mm, 3 µm Separation pump + Column compartment + Micro-flow loading pump Autosampler • Low gradient delay volume • Low total internal volume • High linear velocity – low delay time • Fast and precise gradient formation Nano-flow standard Pre-concentration kit
  • 19. 19 High-throughput low-flow LC method explained Fast injection routine (< 1 min) Fast sample loading onto the column with micro-flow loading pump (6 sec @150 µL/min) Parallel sample elution and fluidic washing steps Parallel column washing/equilibration and sampler injection routine
  • 20. 20 Typical BPC and TIC chromatograms Wide peptide elution window Extremely low carryover Zoom in
  • 21. 21 Robustness of High-throughput LC-MS Method 0 2000 4000 6000 8000 10000 12000 0 30 60 90 120 150 180 Injection No. Protein groups PSMs MS/MS Peptide groups 0 1 2 3 4 5 6 0 30 60 90 120 150 180 Retentiontime,min Injection No. 0 1 2 3 4 0 30 60 90 120 150 180 PWHM,sec Injection No. [K].LTDcVVMR.[D] [K].SLTNDWEDHLAVK.[H] [K].STELLIR.[K] [K].TVTAMDVVYALK.[R] [K].VDNDENEHQLSLR.[T] [K].VNQIGSVTESIQAcK.[L] [R].EDSQRPGAHLTVK.[K] [R].TVSLGAGAK.[D] Protein groups Peptide groups PSMs MS/MS 0.0E+00 1.0E+07 2.0E+07 3.0E+07 4.0E+07 5.0E+07 0.E+00 2.E+06 4.E+06 6.E+06 8.E+06 1.E+07 0 30 60 90 120 150 180 Area Area Injection No. Retention time Peak area PWHM SD < 0.1 min PWHM < 3 sec RSD < 10% Maximum speed TOP 40 MS2 7.5K IT MS2 14 sec
  • 22. 22 How do we outperform competitors? 14.4 minutes 8 minutes Vendor “E” UltiMate 3000 RSLCnano 180 samples/day MS utilization – 75% per run 100 samples/day MS utilization – 42% per run SignalSignal 0 71 62 53 4 Previous injection Next injection Previous injection Next injection 1 min 3 min UltiMate 3000 RSLCnano system compared to Vendor “E”: Almost double the throughput Higher peak capacity More equal peptide distribution
  • 23. 23 ASMS2018 poster High-throughput LC-MS: where to find more information? Post 1 Post 2 Post 3 Tweet 3 Tweet 2 Tweet 1 Link In review
  • 24. 24 2 x 75 µm x 150 mm, 3 µm; PN 164568 Q Exactive HF-XTandem Pre-concentration setup with UltiMate 3000 RSLCnano system Flow rate, µL/min Throughput 75 µm 150 µm 100% MS Utilization with Tandem LC-MSTandem low-flow LCMS for maximum MS utilization
  • 25. 25 Enhanced Productivity with Tandem low-flow LC • Tandem nano LC is applied to increase throughput. • Analyte separation is carried out on two columns in an alternating fashion • While one column is used for separation the other is washed and re-equilibrated for the next run • Tandem nano LC employed for long gradient / deep dive proteomics applications run on long columns, enables multiple washing steps to be carried out on the offline column. Tandem operation – Separation matches washing and equilibration Cleaning operation -Long gradient for separation on Column A is combined with consecutive short gradients for washing column B 0 20 40 60 80 100 1 11 21 31 41 51 61 71 81 91 101 111 121 Column A Column B 0 20 40 60 80 100 1 11 21 31 41 51 61 71 81 91 101 111 121 Column A Column B %,B %,B
  • 26. 26 High throughput low-flow Tandem LC setup to ESI MS waste waste Separation pump Equilibration pump Separation pump + Column compartment + Micro-flow loading pump Equilibration pump Autosampler 3 pumps 2 nano columns 2 trap cartridges 2 switching valves One LC-MS method
  • 27. 27 Tandem low-flow LC-MS method explained Peptide elution window: ~ 6.5 min Column 1, cycle time: 7 min Method duration and raw file length: 7 min Prepare for next injection routine Sample loop strong solvent wash Trap 2 strong solvent wash & equilibration Done in parallel with loading pump, syringe, 2nd nano pump Sampler wash between reinjections Sample load onto trap 2 Sample transfer to 2nd column Column 2 washing Column 2 equilibration Trap 2 washing Next gradient on column 2 Previous gradient on column 2 Look ahead injections: sample is loaded onto the second trap cartridge in parallel with current run Trap cartridge is washed with strong wash solvent using micro- flow pump Injection fluidics is washed in parallel with peptide separation The delay in peptide elution is removed by intelligent use of post column flow diversion Method can be fully adapted according to the required gradient length The same pump is used for separation on column 1 and column 2
  • 28. 28 High throughput, Tandem low-flow LC: what are the benefits? RT: 0.00-7.01 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 Time(min) 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 3.63 3.65 3.71 4.50 3.13 1.80 2.401.82 5.74 3.43 5.77 1.78 2.82 5.80 3.78 4.17 2.42 1.84 2.55 0.820.80 1.04 1.13 4.94 3.95 4.971.470.490.34 1.97 5.38 5.854.600.30 5.15 6.345.99 6.856.87 6.46 NL: 9.46E8 BasePeakF: FTMS+p NSIFull ms [350.0000-2000.0000] MS 7min_HeLa_200ng_MS1_ column1_2018042022021 4 RT: 0.00-7.01 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 Time(min) 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 RelativeAbundance 3.62 3.60 3.68 4.543.70 4.52 4.56 3.06 3.05 5.96 3.42 5.93 2.24 4.16 1.60 1.59 2.71 1.62 4.212.26 2.98 2.280.15 3.28 5.571.65 0.66 5.020.16 5.031.41 6.003.950.81 4.99 0.99 5.590.34 6.025.234.742.021.02 6.09 6.481.99 6.50 6.60 NL: 7.91E8 BasePeakF: FTMS+p NSIFull ms [350.0000-2000.0000] MS 7min_HeLa_200ng_MS1_ column2_2018042022092 5 Column 1 Column 2 7 min 7 min 972 5374 6129 10399 1175 5587 6689 12948 0 2000 4000 6000 8000 10000 12000 14000 Pre-concentration Tandem Protein groups Peptide groups PSMs 0 200 400 600 800 Count 0.500 1.000 1.500 2.000 2.500 3.000 3.500 4.000 4.500 5.000 5.500 6.000 6.500 7.000 PSMs-RT[min] PSMs > 200 samples per 24 hours ~100% MS utilization Good separation performance Reproducible peptide separation profiles on column 1 and column 2 Maximum MS speed utilization
  • 30. 30 Conclusions • Time is Money! • There is an increasing demand for speed, sensitivity and MS utility in LC-MS applications in industrial and translational proteomics and beyond • The UltiMate solution for all your low-flow needs • The UltiMate 3000 RSLCnano is uniquely capable of delivering ultra high throughput, low-flow applications due to its: • Integrated micro-flow pump for high speed online sample loading and desalting and trap and loop washing • Wide pressure / flow footprint of 50 nL/min to 50 µL/min at up to 900 bar column pressure • Robust pressure stable tool free industry leading nanoViper connections • Offering high throughput AND sensitivity • High through-put low-flow LC-MS methods using nano columns and capillary flow rates yield uncompromising levels of speed and sensitivity whilst delivering as yet unseen levels of MS utility • Make the most of your MS with Tandem low-flow LCMS • Tandem LC-MS is possible through the simple addition of a low-flow pump and column set • Intelligent method control combined with look ahead injections enable continuous back-to-back MS data generation for the first time in the history of low-flow LC-MS
  • 31. 31 • Complete with schematics, method parameters and example data Last but not least…. The ALL NEW • Comprehensive and easy to follow • Detailed information from instrument setup to best practices • Available for free download on the RSLCnano webpage • https://www.thermofisher.com/order/catalog/product/ULTIM3000RSLCNANO • And as a hard copy shipped with each application kit • UltiMate 3000 RSLCnano Standard Applications Manual (V 3.0)