SlideShare a Scribd company logo
1
College of American
Pathologists
Laboratory Accreditation Program
Laboratory General Checklist:
How to Validate a New Test
D. Robert Dufour, MD, FCAP, FACB
Consultant, Veterans Affairs Medical Center, Washington DC
Emeritus Professor of Pathology, George Washington University
September 17, 2008
Copyright © 2008 College of American Pathologists (CAP). All rights are reserved. Participants are permitted to
duplicate the materials for educational use only within their own institution. These materials may not be used for
commercial purposes or altered in any way.
2
Learning Objectives
• As a result of participating in this
activity, you will be able to:
– Describe the necessary steps to validating
new test performance
– Assure that all steps are performed for all
new methods introduced into your
laboratory
3
What is required?
• GEN 42020 – Verify analytic accuracy
and precision
• GEN 42025 – Verify and document
analytic sensitivity (lower detection limit)
• GEN 42030 – Verify and document
analytic interferences
• GEN 42085 – Verify reportable range
4
What is required?
• GEN 42160 – Explain significant
differences in results to clients
• GEN 42162 – Verify or establish
reference intervals
• GEN 42163 – Evaluate appropriateness
of reference intervals; take corrective
action if necessary
5
What is required?
• Questions also appear in discipline
specific checklists when more than one
instrument is used to evaluate the same
parameter (e.g., HEM.28000,
CHM.13800)
• Comparability of results should be
checked at initial setup, as well as
rechecked at least twice yearly
6
WHY DO WE DO THIS?
• CLIA and CAP require that laboratories
validate the performance of tests (GEN
42020-42163)
• Manufacturer’s data, and even work done by
manufacturer’s representatives on set-up,
may not meet all requirements
• Need to be familiar with what is required and
how to perform needed studies
7
What is Required?
• Checklist indicates that, for unmodified
FDA cleared tests, can use
manufacturer’s or published information
for some of these
• However, lab must verify data on
accuracy, precision, and reportable
range, as well as reference intervals
8
Analytic Accuracy
• Agreement between test result and
“true” result
• Can be done in two main ways:
– Comparison of results between new
method and “reference” method
– Results using new method on certified
reference materials (Recovery)
• First approach almost always used
9
Method Comparison
• Simplest procedure involves testing 20
samples that span the entire testing range
(but do not exceed measurement range)
(CLSI-EP15-A2)
• Run samples by both new and comparative
method (your current method, reference
method)
• Evaluate bias (difference between new,
comparative method) in one of several ways
10
Method Comparison
• Calculate average bias (difference between
your result and comparison method); is this
within allowable limits?
• Review standard texts (Tietz, Kaplan and
Pesce) for clinically allowable limits of
difference (for example, for albumin, clinically
allowable bias 0.06 g/dL; round to 0.1), also
check CLIA limits
• Average is poorest way to compare tests
11
Method Comparison
• This same approach can be used to
compare two methods currently in use;
needs to be done at least every 6
months
• We run 10 samples on both
instruments, compare results to
clinically allowable bias; if each
acceptable, approve comparison study
12
Method Comparison
• Can compare using linear regression
analysis; while better than average bias,
relatively insensitive to biases in only
part of the measurement range
• Standard software (e.g., Excel) can plot
linear regression and calculate
regression equation; ideally, slope = 1,
intercept = 0, r = 0.99
13
Method Comparison
• Slope = proportional bias; often related to
calibration differences between methods
• Intercept = constant bias; may be related to
calibration or set point issues
• r = how close points are to the line drawn; low
r may indicate inadequate range of values
studied, individual samples with
interferences, or poor correlation between
methods; need to exclude any outliers and
repeat
14
0
100
200
300
400
500
600
700
800
0 100 200 300 400 500
pO2 Radiometer
pO2
Gem
y = 1.28x - 44.6
r = 0.911
15
Method Comparison
• In this case, slope = 1.28, implying 28%
higher results with new method; intercept = -
44, indicating absolute difference is 44 mm
Hg lower with new instrument; r is low
(0.911), indicating scatter of data
• This would imply poor agreement between
new, old method
• Graph also identified two data outliers; in
reviewing data, there was a significant delay
between testing for these two samples, so
they were excluded
16
Method Comparison
• A better approach is to prepare bias plots
(difference between results versus value from
comparison method); difference (in %) should
be similar at all values; can also be done in
Excel
• Difference at extremes indicates need to
consider that measurement range needs to
be tightened, or that other approaches to
assure correlation are needed.
17
-120.00
-100.00
-80.00
-60.00
-40.00
-20.00
0.00
20.00
40.00
0 100 200 300 400 500
pO2 Reference
pO2
Difference
(%)
18
Method Comparison
• Data in bias plot show no difference in pO2
up to values of 300 mm Hg
• At pO2 over 300, however, there is a clear
negative bias with new method; thus, should
not report pO2 > 300 (yet)
• Need to proceed to analytical measurement
range studies to see if problem is with new
method or current method
19
Method Comparison Problems
• When data suggest a bias, a larger
study is needed (typically 100 samples)
• CLSI EP9-A2 provides suggested
sample selection criteria and more
elaborate interpretation information than
provided in EP15-A2
20
Method Comparison Problems
• We recently brought a new instrument from
same manufacturer in house
• Correlation study for BUN seemed to show
good agreement (next slide)
• Inspection of results and bias plot, however,
showed that while results were comparable at
high values, they were significantly higher
with the new method at values in the
reference range and slightly above
21
BUN LXi vs DXc
y = 0.9702x + 2.9581
R
2
= 0.9983
0
20
40
60
80
100
120
0 20 40 60 80 100 120
BUN LXi (mg/dL)
BUN
DXc
(mg/dL)
22
BUN LXi vs DXc
-2
-1
0
1
2
3
4
5
0 20 40 60 80 100 120
BUN LXi (mg/dL)
Bias
(mg/dL)
23
Method Comparison Problems
• What should be done?
• Requires a more careful study to
evaluate the extent of the bias and
suggest approaches to fix it
• We performed bias study with 100
samples having BUN below 25 mg/dL
by our current assay
24
BUN LXi vs DXc
-6
-4
-2
0
2
4
6
8
0 5 10 15 20 25 30
BUN LXi (mg/dL)
Bias
(mg/dL)
25
Method Comparison Problems
• Results showed a constant bias of about 4
mg/dL (average) in this range
• Evaluated new lot of calibrator – same
findings
• Adjusted set point of instrument and re-
tested; results now within allowable bias at
values < 30 mg/dL
• Must remember to take out correction when
performing proficiency testing!!!
26
Verify Precision
• Precision is repeatability
• Analyze repeatedly, determine variation
• CLSI protocol (EP15-A2) suggests two levels,
run 3 times per run for 5 days (15 replicates
in all)
• Use a spreadsheet to calculate s.d. for the
measurements, compare to manufacturer’s
claim; if higher, need to evaluate for cause
27
Verify Precision
• Ideally take samples with clinically important
results (e.g., for TSH, should select low and
high values, not both in reference range)
• Compare precision to clinically acceptable
variation (as described earlier) to assure meets
clinical needs
• For samples that are not stable, need to adjust
protocol; may not be able to test long term
precision using patient samples, may have to
use controls or calibrators that have been
stabilized
28
Verify Analytic Sensitivity
• Also called lower detection limit
• Involves two steps: determination of values
obtained with blank samples, and values
obtained with low level positive samples
• Blank samples often use the zero calibrator
for an assay, as it can be difficult to find
samples truly missing the substance of
interest
29
Verify Analytic Sensitivity
• Low level positive samples identified at or only
slightly above the manufacturer’s stated analytical
sensitivity; for uncommon tests, may need to contact
another laboratory doing the test for help in finding
low positive samples
• CLSI EP17-A describes actual procedure
– Run 20 blanks; if < 3 exceed stated blank value, accept that
value
– Run 20 low patient samples near the detection limit; if at
least 17 are above the blank value, the detection limit is
verified
30
Verify Analytic Interferences
• No approved protocol for performing this task
• Most commonly, involves listing stated
interferences from manufacturer, and
evaluating samples in correlation studies for
differences (outliers), investigation of causes
of interference
• Often difficult to determine exact cause of
interference except for common causes
(hemolysis, lipemia, icterus, related
compounds)
31
Verify Analytic Interferences
• New method for serum iron from same
manufacturer
• Ran 20 samples with old, new method
• All samples but one within acceptable
limits
• One sample from dialysis patient: Fe
123 mcg/dL with old method, 452
mcg/dL with new method
32
Verify Analytic Interferences
• Ran several additional samples from dialysis
patients; no differences noted
• Sent sample to another lab for measurement
by different method; result 125 mcg/dL
• Sent sample to manufacturer with list of drugs
patient taking; unable to determine cause of
interference
• Noted patient’s name; all future requests for
iron on this patient sent to another lab using a
different method
33
Verify Reportable Range
• Reportable range includes:
– Analytical measurement range (AMR) –
range of values that instrument can report
directly (less accurately called linearity)
– Clinically reportable range (CRR) – range
of values that can be reported with dilution
or concentration of samples
• Retain records while method in use, at
least 2 years after discontinued
34
Verify Reportable Range
• AMR must be verified before method
introduced, and at least every 6 months (and
after recalibration or major maintenance)
while in use
• CRR is a clinical decision by lab director, and
does not require experiments or re-validation;
however, dilution or concentration protocols
must be specified in methods
35
AMR Verification
• Must include three levels (low, midpoint,
high)
• Can use commercial linearity materials,
PT or patient samples with known
results, standards or calibrators
• Can also be done by calibration
verification if three samples that span
measurement range used
36
AMR Verification
• Lab must specify criteria for
acceptability (can be as absolute values
or percent differences); should be
specific for each method, based on
clinical criteria, as discussed earlier
• More formal protocols for linearity
evaluation are found in CLSI EP6-A, but
are not required
37
Practical Example
• If no commercial material available, will
need to create own materials
• Select high, low sample; can mix to create
mid-point sample
• If high above measurement limit, do
dilution with low sample to create level
near limit
• If stable, can aliquot, freeze for future use
38
Reference Limits
• Not required for laboratory to establish
its own reference limits, but must verify
that limits it uses are appropriate for
patient population served
• Should review how any published limits
were established before considering
whether to try to adopt them
39
Adopted Reference Limits
• Manufacturer suggested
• Reference laboratory
• Published articles
• Neighboring laboratory
• Previous reference limits in own lab
40
Reference Limits
• Validation protocol outlined in CLSI
C28-A2
• Select 20 representative healthy
individuals and do test; if ≤ 2 outside
proposed limits, validated
• If > 2 outside, can repeat with another
20, and accept if ≤ 2 outside (not worth
repeat if > 5 outside proposed limits)
41
105 103 107 102 106
101 103 107 101 100
102 108 104 105 103
100 107 102 105 101
Example
• Old Cl reference limits 98-104; data
below from 20 patients. What would
you do?
42
Reference Limits
• If not validated, will need to establish
own reference limits
• Briefly, need 120 individuals (200 if
does not follow gaussian distribution) to
be confident in accuracy
• Need to have criteria for exclusion (may
differ for certain tests)
43
Summary
• While many steps are involved in
method validation, not all have to be
performed by the laboratory
• Lab must verify accuracy, precision,
AMR, CRR, and reference intervals
before methods are adopted (and must
be signed by lab director)
44
Summary
• Simple protocols are provided to
perform these steps, but more formal
procedures are available in CLSI
documents
• Most critical step is having good criteria
established for what is acceptable,
based on patient care needs
45
Resources
• CAP Laboratory General Checklist
• CLSI EP5-A2: Evaluation of Precision
Performance of Quantitative
Measurement Methods (2004)
• CLSI EP6-A: Evaluation of the Linearity
of Quantitative Measurement
Procedures: A Statistical Approach
(2003)
46
Resources
• CLSI EP9-A2: Method Comparison and
Bias Estimation Using Patient Samples
(2002)
• CLSI EP10-A3: Preliminary Evaluation
of Quantitative Clinical Laboratory
Measurement Procedures (2006)
47
Resources
• CLSI EP15-A2: User Verification of
Performance for Precision and Trueness
(2005)
• CLSI EP17-A: Protocols for Determination
of Limits of Detection and Limits of
Quantitation (2004)
• CLSI C28-A2: How to Define and
Determine Reference Intervals in the
Clinical Laboratory (2000)
48
Technical Assistance
http://www.cap.org
Email: accred@cap.org
800-323-4040, ext. 6065
49
Questions?
50
Past Audioconferences
• Missed part/all of an audioconference?
• Want to hear it again?
• Want to tell a co-worker?
• Virtual Library of Past Audioconferences
– Available 24/7 on www.cap.org 4 weeks post
session (see Attachment A for steps on how to
access post audioconferences).
– Laboratory Improvement and Accreditation tab
– Preparing to Inspect
51
#3
#2
#1
Attachment A: Steps to Access the Virtual Library of Past Audioconferences

More Related Content

PPTX
Journal club presentation @ Rxvichu!!
PDF
Declaration of-helsinki presantation
PPTX
Journal club presentation, Dr Usama Ragab
PDF
Steps to Writing an Effective Case Report
PPTX
TYPES OF ANOVA.pptx
PDF
Minimize the Impact of E2B(R3) on Drug Safety Operations with Argus Safety
PPTX
Blinding in clinical trials
Journal club presentation @ Rxvichu!!
Declaration of-helsinki presantation
Journal club presentation, Dr Usama Ragab
Steps to Writing an Effective Case Report
TYPES OF ANOVA.pptx
Minimize the Impact of E2B(R3) on Drug Safety Operations with Argus Safety
Blinding in clinical trials

What's hot (20)

PPTX
Research protocol writting
PPTX
Calibration and Qualification of Analytical Instruments
PDF
Research methodology in medical research
PPTX
Cross over design, Placebo and blinding techniques
PPTX
Accuracy and precision presentation
PPTX
Calibration
PDF
Data Management in Clinical Research
 
PPTX
HPLC Trouble Shotting
PPTX
CPCSEA GUIDELINE
DOCX
From Chef to Actor: 8 Fascinating Facts About Lucas Bravo’s Journey
PPTX
Declaration of Helsinki
PPTX
Vendor relationship...fatality.
PPTX
Internal Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
PPTX
Validation of utility system (water system)
PPTX
Mass spectrometry
PPTX
Bragg’s law
PPTX
Qualification of Dry Heat Sterilizer or Tunnel.pptx
PPTX
Journal Club: Thrombin-Receptor Antagonist Vorapaxar in Acute Coronary Syndromes
PDF
Pipette Calibration Worksheet & Guidelines -M. Asif
PPTX
Method Validation: What Are Its Key Parameters
Research protocol writting
Calibration and Qualification of Analytical Instruments
Research methodology in medical research
Cross over design, Placebo and blinding techniques
Accuracy and precision presentation
Calibration
Data Management in Clinical Research
 
HPLC Trouble Shotting
CPCSEA GUIDELINE
From Chef to Actor: 8 Fascinating Facts About Lucas Bravo’s Journey
Declaration of Helsinki
Vendor relationship...fatality.
Internal Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
Validation of utility system (water system)
Mass spectrometry
Bragg’s law
Qualification of Dry Heat Sterilizer or Tunnel.pptx
Journal Club: Thrombin-Receptor Antagonist Vorapaxar in Acute Coronary Syndromes
Pipette Calibration Worksheet & Guidelines -M. Asif
Method Validation: What Are Its Key Parameters
Ad

Similar to How to Validate a New Test.pdf (20)

PPTX
Validation of lab instruments and quantitative test methods
PPTX
Validation of qualitative lab test methods
PPT
Clinical lab-method validation.ppt
PDF
Cap blood gas
PPT
Method validation for calibration measure.ppt
PDF
method_validation_molinaro_laboratory.pdf
 
PPTX
Validation my ppt
PPTX
Quality control.pptx
PPTX
(IQC) - Internal Quality Control.pptx
PDF
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
 
PPT
HPLC validation.ppt
PPTX
QA QC Program for Waste Water Analysis ppt
PPTX
Control of analytical quality using stable control materials postgrad
PPTX
Analytical Method Validation
PPT
2. Quality Control Notes
PPTX
New Microsoft PowerPoint Presentation.pptx
PPT
Validation of Analytical method.ppt
PPTX
Analytical Method Validation.pptx
PDF
Analytical Errors and Validation of Analytical procedures.pdf
PDF
Analytical Errors and Validation of Analytical procedures.pdf
Validation of lab instruments and quantitative test methods
Validation of qualitative lab test methods
Clinical lab-method validation.ppt
Cap blood gas
Method validation for calibration measure.ppt
method_validation_molinaro_laboratory.pdf
 
Validation my ppt
Quality control.pptx
(IQC) - Internal Quality Control.pptx
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
 
HPLC validation.ppt
QA QC Program for Waste Water Analysis ppt
Control of analytical quality using stable control materials postgrad
Analytical Method Validation
2. Quality Control Notes
New Microsoft PowerPoint Presentation.pptx
Validation of Analytical method.ppt
Analytical Method Validation.pptx
Analytical Errors and Validation of Analytical procedures.pdf
Analytical Errors and Validation of Analytical procedures.pdf
Ad

Recently uploaded (20)

PPTX
AI_in_Pharmaceutical_Technology_Presentation.pptx
PPTX
Nursing Care Aspects for High Risk newborn.pptx
PPTX
ABG advance Arterial Blood Gases Analysis
PPTX
Vaginal Bleeding and Uterine Fibroids p
PPT
Parental-Carer-mental-illness-and-Potential-impact-on-Dependant-Children.ppt
PPT
Microscope is an instrument that makes an enlarged image of a small object, t...
PPTX
Bronchial_Asthma_in_acute_exacerbation_.pptx
PDF
Dr Masood Ahmed Expertise And Sucess Story
PPTX
Immunity....(shweta).................pptx
PDF
2E-Learning-Together...PICS-PCISF con.pdf
PPT
Pyramid Points Lab Values Power Point(11).ppt
PPTX
Rheumatic heart diseases with Type 2 Diabetes Mellitus
PPTX
HEMODYNAMICS - I DERANGEMENTS OF BODY FLUIDS.pptx
PPTX
Galactosemia pathophysiology, clinical features, investigation and treatment ...
PDF
Structure Composition and Mechanical Properties of Australian O.pdf
PPTX
3. Adherance Complianace.pptx pharmacy pci
PPT
KULIAH UG WANITA Prof Endang 121110 (1).ppt
PPTX
different types of Gait in orthopaedic injuries
PDF
Dr. Jasvant Modi - Passionate About Philanthropy
PDF
Khaled Sary- Trailblazers of Transformation Middle East's 5 Most Inspiring Le...
AI_in_Pharmaceutical_Technology_Presentation.pptx
Nursing Care Aspects for High Risk newborn.pptx
ABG advance Arterial Blood Gases Analysis
Vaginal Bleeding and Uterine Fibroids p
Parental-Carer-mental-illness-and-Potential-impact-on-Dependant-Children.ppt
Microscope is an instrument that makes an enlarged image of a small object, t...
Bronchial_Asthma_in_acute_exacerbation_.pptx
Dr Masood Ahmed Expertise And Sucess Story
Immunity....(shweta).................pptx
2E-Learning-Together...PICS-PCISF con.pdf
Pyramid Points Lab Values Power Point(11).ppt
Rheumatic heart diseases with Type 2 Diabetes Mellitus
HEMODYNAMICS - I DERANGEMENTS OF BODY FLUIDS.pptx
Galactosemia pathophysiology, clinical features, investigation and treatment ...
Structure Composition and Mechanical Properties of Australian O.pdf
3. Adherance Complianace.pptx pharmacy pci
KULIAH UG WANITA Prof Endang 121110 (1).ppt
different types of Gait in orthopaedic injuries
Dr. Jasvant Modi - Passionate About Philanthropy
Khaled Sary- Trailblazers of Transformation Middle East's 5 Most Inspiring Le...

How to Validate a New Test.pdf

  • 1. 1 College of American Pathologists Laboratory Accreditation Program Laboratory General Checklist: How to Validate a New Test D. Robert Dufour, MD, FCAP, FACB Consultant, Veterans Affairs Medical Center, Washington DC Emeritus Professor of Pathology, George Washington University September 17, 2008 Copyright © 2008 College of American Pathologists (CAP). All rights are reserved. Participants are permitted to duplicate the materials for educational use only within their own institution. These materials may not be used for commercial purposes or altered in any way.
  • 2. 2 Learning Objectives • As a result of participating in this activity, you will be able to: – Describe the necessary steps to validating new test performance – Assure that all steps are performed for all new methods introduced into your laboratory
  • 3. 3 What is required? • GEN 42020 – Verify analytic accuracy and precision • GEN 42025 – Verify and document analytic sensitivity (lower detection limit) • GEN 42030 – Verify and document analytic interferences • GEN 42085 – Verify reportable range
  • 4. 4 What is required? • GEN 42160 – Explain significant differences in results to clients • GEN 42162 – Verify or establish reference intervals • GEN 42163 – Evaluate appropriateness of reference intervals; take corrective action if necessary
  • 5. 5 What is required? • Questions also appear in discipline specific checklists when more than one instrument is used to evaluate the same parameter (e.g., HEM.28000, CHM.13800) • Comparability of results should be checked at initial setup, as well as rechecked at least twice yearly
  • 6. 6 WHY DO WE DO THIS? • CLIA and CAP require that laboratories validate the performance of tests (GEN 42020-42163) • Manufacturer’s data, and even work done by manufacturer’s representatives on set-up, may not meet all requirements • Need to be familiar with what is required and how to perform needed studies
  • 7. 7 What is Required? • Checklist indicates that, for unmodified FDA cleared tests, can use manufacturer’s or published information for some of these • However, lab must verify data on accuracy, precision, and reportable range, as well as reference intervals
  • 8. 8 Analytic Accuracy • Agreement between test result and “true” result • Can be done in two main ways: – Comparison of results between new method and “reference” method – Results using new method on certified reference materials (Recovery) • First approach almost always used
  • 9. 9 Method Comparison • Simplest procedure involves testing 20 samples that span the entire testing range (but do not exceed measurement range) (CLSI-EP15-A2) • Run samples by both new and comparative method (your current method, reference method) • Evaluate bias (difference between new, comparative method) in one of several ways
  • 10. 10 Method Comparison • Calculate average bias (difference between your result and comparison method); is this within allowable limits? • Review standard texts (Tietz, Kaplan and Pesce) for clinically allowable limits of difference (for example, for albumin, clinically allowable bias 0.06 g/dL; round to 0.1), also check CLIA limits • Average is poorest way to compare tests
  • 11. 11 Method Comparison • This same approach can be used to compare two methods currently in use; needs to be done at least every 6 months • We run 10 samples on both instruments, compare results to clinically allowable bias; if each acceptable, approve comparison study
  • 12. 12 Method Comparison • Can compare using linear regression analysis; while better than average bias, relatively insensitive to biases in only part of the measurement range • Standard software (e.g., Excel) can plot linear regression and calculate regression equation; ideally, slope = 1, intercept = 0, r = 0.99
  • 13. 13 Method Comparison • Slope = proportional bias; often related to calibration differences between methods • Intercept = constant bias; may be related to calibration or set point issues • r = how close points are to the line drawn; low r may indicate inadequate range of values studied, individual samples with interferences, or poor correlation between methods; need to exclude any outliers and repeat
  • 14. 14 0 100 200 300 400 500 600 700 800 0 100 200 300 400 500 pO2 Radiometer pO2 Gem y = 1.28x - 44.6 r = 0.911
  • 15. 15 Method Comparison • In this case, slope = 1.28, implying 28% higher results with new method; intercept = - 44, indicating absolute difference is 44 mm Hg lower with new instrument; r is low (0.911), indicating scatter of data • This would imply poor agreement between new, old method • Graph also identified two data outliers; in reviewing data, there was a significant delay between testing for these two samples, so they were excluded
  • 16. 16 Method Comparison • A better approach is to prepare bias plots (difference between results versus value from comparison method); difference (in %) should be similar at all values; can also be done in Excel • Difference at extremes indicates need to consider that measurement range needs to be tightened, or that other approaches to assure correlation are needed.
  • 18. 18 Method Comparison • Data in bias plot show no difference in pO2 up to values of 300 mm Hg • At pO2 over 300, however, there is a clear negative bias with new method; thus, should not report pO2 > 300 (yet) • Need to proceed to analytical measurement range studies to see if problem is with new method or current method
  • 19. 19 Method Comparison Problems • When data suggest a bias, a larger study is needed (typically 100 samples) • CLSI EP9-A2 provides suggested sample selection criteria and more elaborate interpretation information than provided in EP15-A2
  • 20. 20 Method Comparison Problems • We recently brought a new instrument from same manufacturer in house • Correlation study for BUN seemed to show good agreement (next slide) • Inspection of results and bias plot, however, showed that while results were comparable at high values, they were significantly higher with the new method at values in the reference range and slightly above
  • 21. 21 BUN LXi vs DXc y = 0.9702x + 2.9581 R 2 = 0.9983 0 20 40 60 80 100 120 0 20 40 60 80 100 120 BUN LXi (mg/dL) BUN DXc (mg/dL)
  • 22. 22 BUN LXi vs DXc -2 -1 0 1 2 3 4 5 0 20 40 60 80 100 120 BUN LXi (mg/dL) Bias (mg/dL)
  • 23. 23 Method Comparison Problems • What should be done? • Requires a more careful study to evaluate the extent of the bias and suggest approaches to fix it • We performed bias study with 100 samples having BUN below 25 mg/dL by our current assay
  • 24. 24 BUN LXi vs DXc -6 -4 -2 0 2 4 6 8 0 5 10 15 20 25 30 BUN LXi (mg/dL) Bias (mg/dL)
  • 25. 25 Method Comparison Problems • Results showed a constant bias of about 4 mg/dL (average) in this range • Evaluated new lot of calibrator – same findings • Adjusted set point of instrument and re- tested; results now within allowable bias at values < 30 mg/dL • Must remember to take out correction when performing proficiency testing!!!
  • 26. 26 Verify Precision • Precision is repeatability • Analyze repeatedly, determine variation • CLSI protocol (EP15-A2) suggests two levels, run 3 times per run for 5 days (15 replicates in all) • Use a spreadsheet to calculate s.d. for the measurements, compare to manufacturer’s claim; if higher, need to evaluate for cause
  • 27. 27 Verify Precision • Ideally take samples with clinically important results (e.g., for TSH, should select low and high values, not both in reference range) • Compare precision to clinically acceptable variation (as described earlier) to assure meets clinical needs • For samples that are not stable, need to adjust protocol; may not be able to test long term precision using patient samples, may have to use controls or calibrators that have been stabilized
  • 28. 28 Verify Analytic Sensitivity • Also called lower detection limit • Involves two steps: determination of values obtained with blank samples, and values obtained with low level positive samples • Blank samples often use the zero calibrator for an assay, as it can be difficult to find samples truly missing the substance of interest
  • 29. 29 Verify Analytic Sensitivity • Low level positive samples identified at or only slightly above the manufacturer’s stated analytical sensitivity; for uncommon tests, may need to contact another laboratory doing the test for help in finding low positive samples • CLSI EP17-A describes actual procedure – Run 20 blanks; if < 3 exceed stated blank value, accept that value – Run 20 low patient samples near the detection limit; if at least 17 are above the blank value, the detection limit is verified
  • 30. 30 Verify Analytic Interferences • No approved protocol for performing this task • Most commonly, involves listing stated interferences from manufacturer, and evaluating samples in correlation studies for differences (outliers), investigation of causes of interference • Often difficult to determine exact cause of interference except for common causes (hemolysis, lipemia, icterus, related compounds)
  • 31. 31 Verify Analytic Interferences • New method for serum iron from same manufacturer • Ran 20 samples with old, new method • All samples but one within acceptable limits • One sample from dialysis patient: Fe 123 mcg/dL with old method, 452 mcg/dL with new method
  • 32. 32 Verify Analytic Interferences • Ran several additional samples from dialysis patients; no differences noted • Sent sample to another lab for measurement by different method; result 125 mcg/dL • Sent sample to manufacturer with list of drugs patient taking; unable to determine cause of interference • Noted patient’s name; all future requests for iron on this patient sent to another lab using a different method
  • 33. 33 Verify Reportable Range • Reportable range includes: – Analytical measurement range (AMR) – range of values that instrument can report directly (less accurately called linearity) – Clinically reportable range (CRR) – range of values that can be reported with dilution or concentration of samples • Retain records while method in use, at least 2 years after discontinued
  • 34. 34 Verify Reportable Range • AMR must be verified before method introduced, and at least every 6 months (and after recalibration or major maintenance) while in use • CRR is a clinical decision by lab director, and does not require experiments or re-validation; however, dilution or concentration protocols must be specified in methods
  • 35. 35 AMR Verification • Must include three levels (low, midpoint, high) • Can use commercial linearity materials, PT or patient samples with known results, standards or calibrators • Can also be done by calibration verification if three samples that span measurement range used
  • 36. 36 AMR Verification • Lab must specify criteria for acceptability (can be as absolute values or percent differences); should be specific for each method, based on clinical criteria, as discussed earlier • More formal protocols for linearity evaluation are found in CLSI EP6-A, but are not required
  • 37. 37 Practical Example • If no commercial material available, will need to create own materials • Select high, low sample; can mix to create mid-point sample • If high above measurement limit, do dilution with low sample to create level near limit • If stable, can aliquot, freeze for future use
  • 38. 38 Reference Limits • Not required for laboratory to establish its own reference limits, but must verify that limits it uses are appropriate for patient population served • Should review how any published limits were established before considering whether to try to adopt them
  • 39. 39 Adopted Reference Limits • Manufacturer suggested • Reference laboratory • Published articles • Neighboring laboratory • Previous reference limits in own lab
  • 40. 40 Reference Limits • Validation protocol outlined in CLSI C28-A2 • Select 20 representative healthy individuals and do test; if ≤ 2 outside proposed limits, validated • If > 2 outside, can repeat with another 20, and accept if ≤ 2 outside (not worth repeat if > 5 outside proposed limits)
  • 41. 41 105 103 107 102 106 101 103 107 101 100 102 108 104 105 103 100 107 102 105 101 Example • Old Cl reference limits 98-104; data below from 20 patients. What would you do?
  • 42. 42 Reference Limits • If not validated, will need to establish own reference limits • Briefly, need 120 individuals (200 if does not follow gaussian distribution) to be confident in accuracy • Need to have criteria for exclusion (may differ for certain tests)
  • 43. 43 Summary • While many steps are involved in method validation, not all have to be performed by the laboratory • Lab must verify accuracy, precision, AMR, CRR, and reference intervals before methods are adopted (and must be signed by lab director)
  • 44. 44 Summary • Simple protocols are provided to perform these steps, but more formal procedures are available in CLSI documents • Most critical step is having good criteria established for what is acceptable, based on patient care needs
  • 45. 45 Resources • CAP Laboratory General Checklist • CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods (2004) • CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach (2003)
  • 46. 46 Resources • CLSI EP9-A2: Method Comparison and Bias Estimation Using Patient Samples (2002) • CLSI EP10-A3: Preliminary Evaluation of Quantitative Clinical Laboratory Measurement Procedures (2006)
  • 47. 47 Resources • CLSI EP15-A2: User Verification of Performance for Precision and Trueness (2005) • CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation (2004) • CLSI C28-A2: How to Define and Determine Reference Intervals in the Clinical Laboratory (2000)
  • 50. 50 Past Audioconferences • Missed part/all of an audioconference? • Want to hear it again? • Want to tell a co-worker? • Virtual Library of Past Audioconferences – Available 24/7 on www.cap.org 4 weeks post session (see Attachment A for steps on how to access post audioconferences). – Laboratory Improvement and Accreditation tab – Preparing to Inspect
  • 51. 51 #3 #2 #1 Attachment A: Steps to Access the Virtual Library of Past Audioconferences